| | Class 2 Device Recall GE Healthcare Optima CT660 |  |
| Date Initiated by Firm | April 27, 2018 |
| Create Date | August 17, 2018 |
| Recall Status1 |
Terminated 3 on August 21, 2020 |
| Recall Number | Z-2871-2018 |
| Recall Event ID |
80563 |
| 510(K)Number | K110227 |
| Product Classification |
System, x-ray, tomography, computed - Product Code JAK
|
| Product | GE Healthcare Optima CT660. Sold under the following product names:
BEIJING OPTIMA CT660 PRO GT170, BEIJING OPTIMA CT660 PRO GT200,
BEIJING OPTIMA CT660S PRO 32 S, CT GOLDSEAL,
CT GOLDSEAL OPTIMA 660, CT GOLDSEAL OPTIMA 660 UT-BUC, CT OPTIMA 660 LNG HN P20, CT OPTIMA 660 MID HN P20, CT OPTIMA 660 PRO, CT OPTIMA 660 PRO LONG BJG P2., CT OPTIMA 660 PRO MID, CT OPTIMA 660 PRO MID HN P20, CT OPTIMA 660S LNG HN P20, CT OPTIMA CT660 GT 1700, CT OPTIMA CT660 GT1700, CT OPTIMA CT660 GT1700V,
CT OPTIMA CT660 PRO GT1700, CT OPTIMA CT660 PRO LNG P1.5 H, CT OPTIMA CT660 PRO MID P1.5 H, CT OPTIMA CT660 SE GT 1700, CT OPTIMA CT660 SE GT2000, CT660, CT660 ASIR OPT, OPT CT660 FREEDOM GT1700 MKE25, OPT CT660 FREEDOM GT1700 MKE25, OPT CT660 FREEDOM GT2000 MKE25, OPTI660_A_G_1.5, OPTI660_A_L_2.0, OPTI660_N_G_1.5, OPTI660_N_G_2.0, OPTI660_N_L_2.0, OPTIMA 660 DIVA, OPTIMA 660 LNG HN P20, OPTIMA 660 LNG HN P20, OPTIMA 660 LONG, OPTIMA 660 LONG BJG P2.0, OPTIMA 660 LONG WSO, OPTIMA 660 MID, OPTIMA 660 MID BJG P1.5, OPTIMA 660 MID HN P20, OPTIMA 660 MID HN P20, OPTIMA 660 MID WSO, OPTIMA 660 PRO LGN HN P20, OPTIMA 660 PRO LONG BJG P1.5_, OPTIMA 660 PRO LONG BJG P2, OPTIMA 660 PRO LONG BJG P2.0, OPTIMA 660 PRO LONG BJG P2.0, OPTIMA 660 PRO LONG WSO, OPTIMA 660 PRO MID BJG P1.5, OPTIMA 660 PRO MID BJG P2.0, OPTIMA 660 PRO MID H, OPTIMA 660 PRO MID HN P20, OPTIMA 660 PRO MID WSO, OPTIMA 660 WSO, OPTIMA 660S 32
OPTIMA 660S 32 SLICE LONG WSO, OPTIMA 660S 32 SLICE MID WSO,
OPTIMA 660S LNG HN P20, OPTIMA 660S MID BJG P2.0, OPTIMA 660S PRO 32 SLICE M WSO, OPTIMA 660S PRO LONG BJG P2.0, OPTIMA 660S PRO MID BJG P1.5, OPTIMA 660S WSO, OPTIMA CT 660 FREEDOM, OPTIMA CT660, OPTIMA CT660 2ND EDITION, OPTIMA CT660 ASIR, OPTIMA CT660 ASIR ADV 2ND, OPTIMA CT660 FREEDOM, OPTIMA CT660 FREEDOM LG BJG2.5, OPTIMA CT660 FREEDOM LG BJG2.5, OPTIMA CT660 FREEDOM LG HN2, OPTIMA CT660 FREEDOM LG HN2.5, OPTIMA CT660 FREEDOM MD, OPTIMA CT660 FREEDOM MD BJG2., OPTIMA CT660 GS, OPTIMA CT660 GT 1700 BJG2, OPTIMA CT660 GT 1700 BJG2.5, OPTIMA CT660 GT 1700 HN P2, OPTIMA CT660 GT 1700 HN P2., OPTIMA CT660 GT 1700 HN P2.5, OPTIMA CT660 GT 2000 HN P2.5, OPTIMA CT660 GT1700, OPTIMA CT660 GT1700 MKE2.5, OPTIMA CT660 GT1700 PHASE 1.5, OPTIMA CT660 GT2000, OPTIMA CT660 GT2000, OPTIMA CT660 GT2000 PHASE 1.5, OPTIMA CT660 LITE, OPTIMA CT660 LITE HN P2.5, OPTIMA CT660 MID P1.5 HINO, OPTIMA CT660 MID P20, OPTIMA CT660 MID P2000, OPTIMA CT660 PRO, OPTIMA CT660 PRO GT 2000, OPTIMA CT660 PRO GT1700, OPTIMA CT660 PRO GT1700 ASIR, OPTIMA CT660 PRO GT1700 P20, OPTIMA CT660 PRO GT1700 PHASE, OPTIMA CT660 PRO GT1700 PHASE 1.5, OPTIMA CT660 PRO GT1700V, OPTIMA CT660 PRO GT1700V ASIR, OPTIMA CT660 PRO GT2000, OPTIMA CT660 PRO GT2000 PHASE, OPTIMA CT660 PRO GT2000 PHASE 1.5, OPTIMA CT660 PRO LNG, OPTIMA CT660 PRO LNG P1.5, OPTIMA CT660 PRO LNG P1.5 HINO, OPTIMA CT660 PRO LNG P1.5 HINO, OPTIMA CT660 PRO LONG BJG P2.0, OPTIMA CT660 PRO LONG P2.0, OPTIMA CT660 PRO MID P1.5,
OPTIMA CT660 PRO MID P1.5 H, OPTIMA CT660 PRO MID P1.5 HINO, OPTIMA CT660 SE, OPTIMA CT660 SE GT 1700 BJG, OPTIMA CT660 SE GT 1700 BJG2.5, OPTIMA CT660 SE GT 1700 BJG2.5, OPTIMA CT660 SE GT 1700 HN2, OPTIMA CT660 SE GT 2000, OPTIMA CT660 SE GT 2000 BJG, OPTIMA CT660 SE GT 2000 BJG2., OPTIMA CT660 SE GT 2000 BJG2.5, OPTIMA CT660 SE GT 2000 HN2, OPTIMA CT660 SE GT 2000 HN2.5, OPTIMA CT660 SE GT2000 MKE2.5, OPTIMA CT660 SE WITH ASIR, OPTIMA CT660 W/PRO PKG-BR, OPTIMA CT660 W/PRO PKG-LA, OPTIMA CT660-64 USA, OPTIMA CT660S, OPTIMA CT660S, OPTIMA CT660S 32 SLICE GT1700, OPTIMA CT660S 32 SLICE HINO, OPTIMA CT660S GT, OPTIMA CT660S GT 1700 HN2.5, OPTIMA CT660S GT 2000, OPTIMA CT660S GT 2000 BJG2.5, OPTIMA CT660S GT 2000 HN2.5, OPTIMA CT660S GT1700 BJG P2.5, OPTIMA CT660S GT1700 HN P2.5, OPTIMA CT660S PRO 32 HINO, OPTIMA CT660S PRO 32 SLICE, OPTIMA CT660S PRO 32 SLICE GT2, OPTIMA CT660S PRO P1.5 HINO, OPTIMACT660_A_G,
OPTIMACT660_A_L, OPTIMACT660_N_G, OPTIMACT660_N_L, REVOLUTION EVO, REVOLUTION EVO WITH GT2000 TABLE (H |
| Code Information |
Model Number, Serial Number (System ID): 5454001-220, 00000296364HM3 (910212220); 5454001-222, 00000316775HM6 (910216062); 5454001-222, 00000316560HM2 (GF1001CT02); 5454001-222, 00000312787HM5 (190163CT01); 5454001-200, 00000296053HM2 (M2842586); 5454001-222, 00000312788HM3 (M9024415); 5454001-220, 00000302025HM2 (M4189263); 5454001-222, 00000316776HM4 (600118CT01); 5454001-222, 00000305880HM7 (X140985807); 5454001-100, 427331CN4 (412357CT); 5454001-100, 424963CN7 (334826CT); 5454001-100, 373759GE0 (303LA660CT); 5454001-100, 428253CN9 (SHICT01432); 5454001-22, 51941YC3 (406434660); 5454001-100, 423901CN8 (201488OPT660); 5454001-100, 424944CN7 (516422CT660); 5454001-122, 00000431995CN0 (405473CT1); 5454001-222, 379369HM2 (864797OPT); 5454001-100, 424968CN6 (864482CT660); 5454001-122, 432338CN2 (703825OP660); 5454001-222, 325767HM2 (757388CT); 5454001-222, 00000316792HM1 (910212229); 5454001, 00000026201YC4 (83021000510515); 5454001, 00000026844YC1 (IL1000CT02); 5454001, 00000033401YC1 (A5634107); 5454001, 00000039355YC3 (850210791); 5454001, 00000029121YC1 (850210732); 5454001, 00000037278YC9 (850210764); 5454001-22, 00000058550YC5 (850210858); 5454001-22, 00000066104YC1 (850210865); 5454001-22, 00000066105YC8 (850210872); 5454001, 00000029510YC5 (850210727); 5454001, 00000029657YC4 (850210735); 5454001, 00000030118YC4 (850210745); 5454001, 00000030498YC0 (850210750); 5454001-22, 00000065974YC8 (850210876); 5454001, 00000030761YC1 (850210746); 5454001-22, 00000058893YC9 (850210860); 5454001-22, 00000064918YC6 (850210877); 5454001-22, 00000065975YC5 (850210875); 5454001-22, 00000055311YC5 (850210849); 5454001-22, 00000062112YC8 (850210874); 5454001, 00000027868YC9 (850210722); 5454001, 00000030116YC8 (850210730); 5454001-22, 00000044851YC4 (850210823); 5454001-22, 00000051792YC0 (850210840); 5454001, 00000030762YC9 (850210751); 5454001, 00000039839YC6 (850210800); 5454001-22, 00000046274YC7 (850210782); 5454001-22, 00000053859YC5 (850210844); 5454001, 00000027381YC3 (850210712); 5454001, 00000027682YC4 (850210724); 5454001-22, 00000042974YC6 (850210808); 5454001-22, 00000049666YC1 (850210830); 5454001-22, 00000052852YC1 (850210841); 5454001-22, 00000054304YC1 (850210845); 5454001-22, 00000055220YC8 (850210848); 5454001-22, 00000063232YC3 (850210870); 5454001-22, 00000042006YC7 (850210807); 5454001-22, 00000059361YC6 (850210859); 5454001-22, 00000065976YC3 (850210878); 5454001, 00000027382YC1 (850210714); 5454001, 00000030759YC5 (850210743); 5454001, 00000025757YC6 (850210705); 5454001, 00000035909YC1 (850210773); 5454001-22, 00000042972YC0 (850210813); 5454001-22, 00000056741YC2 (850210854); 5454001, 00000029387YC8 (850210728); 5454001, 00000034255YC0 (850210771); 5454001, 00000038527YC8 (850210792); 5454001-22, 00000044853YC0 (850210822); 5454001-22, 00000048331YC3 (850210828); 5454001-22, 00000050432YC4 (850210836); 5454001-22, 00000051114YC7 (850210839); 5454001-22, 00000054301YC7 (850210846); 5454001-22, 00000055750YC4 (850210847); 5454001-22, 00000056738YC8 (850210851); 5454001-22, 00000059229YC5 (850210862); 5454001, 00000030763YC7 (850210752); 5454001-22, 00000044852YC2 (RU7619CT01); 5454001-22, 00000040079YC6 (RU1820CT03); 5454001-22, 00000047318YC1 (83021000035812); 5454001-22, 00000043473YC8 (RU1203CT02); 5454001-22, 00000042795YC5 (RU1100CT03); 5454001-22, 00000046469YC3 (RU8065CT01); 5454001-22, 00000039798YC4 (RU7362CT01); 5454001-22, 00000041139YC7 (RU7525CT01); 5454001-22, 00000045309YC2 (RU7837CT01); 5454001-22, 00000045307YC6 (RU7818CT01); 5454001-22, 00000040363YC4 (UK1127CT01); 5454001-222, 00000311087HM1 (NP666104); 5454001-222, 00000315025HM7 (PL2854CT01); 5454001-222, 00000310572HM3 (YP065822); 5454001-22, 00000045310YC0 (KA2839CT01); 5454001-22, 00000045600YC4 (RU7866CT01); 5454001-22, 00000045311YC8 (RU7750CT01); 5454001-22, 00000041825YC1 (RU2635CT02); 5454001-22, 00000044517YC1 (RU1550CT02); 5454001-22, 00000044856YC3 (RU3883CT01); 5454001-22, 00000047517YC8 (RU7827CT01); 5454001-22, 00000064661YC2 (RU9128CT01); 5454001-22, 00000049252YC0 (RU8146CT01); 5454001-22, 00000063233YC1 (RU9064CT01); 5454001, 00000056737YC0 (82421111086); 5454001, 00000058082YC9 (82421111087); 5454001, 00000035650YC1 (82421040266); 5454001, 00000041975YC4 (82421040282); 5454001, 00000045056YC9 (82421040288); 5454001, 00000048836YC1 (82421040293); 5454001, 00000050685YC7 (82421040276); 5454001, 00000050997YC6 (82421040299); 5454001, 00000052083YC3 (82421040280); 5454001, 00000053643YC3 (82421040307); 5454001, 00000054321YC5 (82421040314); 5454001, 00000054683YC8 (82421040310); 5454001, 00000054867YC7 (82421040312); 5454001, 00000054868YC5 (82421040319); 5454001, 00000056278YC5 (82421040323); 5454001, 00000057294YC1 (82421040320); 5454001, 00000058083YC7 (82421040325); 5454001, 00000059882YC1 (82421040338); 5454001, 00000064067YC2 (82421040349); 5454001, 00000065168YC7 (82421040358); 5454001, 00000044569YC2 (82421220097); 5454001, 00000055357YC8 (82421220106); 5454001, 00000056544YC0 (82421220107); 5454001, 00000057359YC2 (82421220109); 5454001, 00000058510YC9 (82421220113); 5454001, 00000052261YC5 (82421150132); 5454001, 00000061567YC4 (82421150155); 5454001, 00000061568YC2 (82421150148); 5454001, 00000050050YC4 (82421150129); 5454001, 00000065761YC9 (82421150170); 5454001, 00000054484YC1 (82421150138); 5454001, 00000040379YC0 (82421150109); 5454001, 00000045055YC1 (82421150124); 5454001, 00000047314YC0 (82421150126); 5454001, 00000056128YC2 (82421150139); 5454001, 00000038729YC0 (82421150111); 5454001, 00000059209YC7 (82421150149); 5454001, 00000061431YC3 (82421150152); 5454001, 00000053644YC1 (82421290080); 5454001, 00000038881YC9 (82421290068); 5454001, 00000065945YC8 (82421290099); 5454001, 00000046243YC2 (82421290071); 5454001, 00000051552YC8 (82421290078); 5454001, 00000048363YC6 (82421290076); 5454001, 00000044676YC5 (82421290072); 5454001, 00000051430YC7 (82421160340); 5454001, 00000065765YC0 (82421160428); 5454001, 00000065596YC9 (82421160413); 5454001, 00000039621YC8 (82421160304); 5454001, 00000046484YC2 (82421160316); 5454001, 00000065440YC0 (82421160402); 5454001, 00000059679YC1 (82421160379); 5454001, 00000055780YC1 (82421160326); 5454001, 00000058509YC1 (82421160363); 5454001, 00000059883YC9 (82421160377); 5454001, 00000058515YC8 (82421160368); 5454001, 00000060052YC8 (82421160373); 5454001, 00000061273YC9 (82421160384); 5454001, 00000039836YC2 (82421160297); 5454001, 00000051803YC5 (82421160312); 5454001, 00000065592YC8 (82421160403); 5454001, 00000058216YC3 (82421160361); 5454001, 00000065766YC8 (82421180148); 5454001, 00000041968YC9 (82421180121); 5454001, 00000056275YC1 (82421180133); 5454001, 00000061904YC9 (82421180143); 5454001, 00000037864YC6 (82421240112); 5454001, 00000041972YC1 (82421240122); 5454001, 00000044394YC5 (82421240123); 5454001, 00000058214YC8 (82421240150); 5454001, 00000036417YC4 (82421240110); 5454001, 00000056543YC2 (82421240148); 5454001, 00000065171YC1 (82421240167); 5454001, 00000065597YC7 (82421240172); 5454001, 00000062065YC8 (82421240161); 5454001, 00000048146YC5 (82421240131); 5454001, 00000052406YC6 (82421240136); 5454001, 00000047317YC3 (82421240129); 5454001, 00000050385YC4 (82421240124); 5454001, 00000052407YC4 (82421240138); 5454001, 00000057292YC5 (82421240151); 5454001, 00000058512YC5 (82421240145); 5454001, 00000049256YC1 (82421170041); 5454001, 00000036176YC6 (82421170038); 5454001, 00000036041YC2 (82421170037); 5454001, 00000043467YC0 (82421070297); 5454001, 00000055963YC3 (82421070356); 5454001, 00000058213YC0 (82421070355); 5454001, 00000065757YC7 (82421070411); 5454001, 00000045924YC8 (82421070306); 5454001, 00000053398YC4 (82421070349); 5454001, 00000040861YC7 (82421070290); 5454001, 00000036718YC5 (82421070302); 5454001, 00000054487YC4 (82421070352); 5454001, 00000065942YC5 (82421070450); 5454001, 00000060270YC6 (82421070380); 5454001, 00000060464YC5 (82421070429); 5454001, 00000047049YC2 (82421070308); 5454001, 00000050383YC9 (82421070314); 5454001, 00000051137YC8 (82421070332); 5454001, 00000053980YC9 (82421070353); 5454001, 00000058675YC0 (82421070372); 5454001, 00000045057YC7 (82421070305); 5454001, 00000050996YC8 (82421070330); 5454001, 00000036899YC3 (82421070285); 5454001, 00000037863YC8 (82421070288); 5454001, 00000052566YC7 (82421070335); 5454001, 00000059678YC3 (82421070369); 5454001, 00000065759YC3 (82421070423); 5454001, 00000064845YC1 (82421070403); 5454001, 00000061903YC1 (82421070388); 5454001, 00000058673YC5 (82421070374); 5454001, 00000052410YC8 (82421070343); 5454001, 00000058370YC8 (82421010320); 5454001, 00000051683YC1 (82421010303); 5454001, 00000045362YC1 (82421010273); 5454001, 00000050682YC4 (82421010292); 5454001, 00000041156YC1 (82421010266); 5454001, 00000052679YC8 (82421010297); 5454001, 00000054486YC6 (82421010310); 5454001, 00000057811YC2 (82421010319); 5454001, 00000062059YC1 (82421010329); 5454001, 00000065167YC9 (82421010340); 5454001, 00000056736YC2 (82421010315); 5454001, 00000060273YC0 (82421010325); 5454001, 00000047728YC1 (82421010283); 5454001, 00000060462YC9 (82421010307); 5454001, 00000043464YC7 (82421010272); 5454001, 00000038536YC9 (82421090379); 5454001, 00000043211YC2 (82421090413); 5454001, 00000044677YC3 (82421090419); 5454001, 00000036614YC6 (82421090361); 5454001, 00000041157YC9 (82421090399); 5454001, 00000040621YC5 (82421090381); 5454001, 00000045363YC9 (82421090420); 5454001, 00000039838YC8 (82421090385); 5454001, 00000038529YC4 (82421090383); 5454001, 00000039622YC6 (82421090387); 5454001, 00000057808YC8 (82421090490); 5454001, 00000061569YC0 (82421090537); 5454001, 00000039837YC0 (82421090391); 5454001, 00000064492YC2 (82421090532); 5454001, 00000043208YC8 (82421090407); 5454001, 00000049257YC9 (82421090426); 5454001, 00000036716YC9 (82421090357); 5454001, 00000037409YC0 (82421090378); 5454001, 00000043752YC5 (82421090410); 5454001, 00000047625YC9 (82421090414); 5454001, 00000060051YC0 (82421090483); 5454001, 00000044675YC7 (82421090406); 5454001, 00000043210YC4 (82421090402); 5454001, 00000037033YC8 (82421090364); 5454001, 00000048144YC0 (82421090395); 5454001, 00000053834YC8 (82421090457); 5454001, 00000050384YC7 (82421090440); 5454001, 00000037862YC0 (82421090380); 5454001, 00000060623YC6 (82421090495); 5454001, 00000048145YC7 (82421090424); 5454001, 00000055962YC5 (82421090478); 5454001, 00000055577YC1 (82421090480); 5454001, 00000058080YC3 (82421210291); 5454001, 00000064894YC9 (82421210334); 5454001, 00000054681YC2 (82421210282); 5454001, 00000050815YC0 (82421210263); 5454001, 00000055960YC9 (82421210286); 5454001, 00000059010YC9 (82421210302); 5454001, 00000059560YC3 (82421210309); 5454001, 00000060272YC2 (82421210311); 5454001, 00000053833YC0 (82421210279); 5454001, 00000061430YC5 (82421210317); 5454001, 00000043466YC2 (82421210251); 5454001, 00000038374YC5 (82421210236); 5454001, 00000047047YC6 (82421210254); 5454001, 00000053240YC8 (82421210267); 5454001, 00000053642YC5 (82421210258); 5454001, 00000053645YC8 (82421210271); 5454001, 00000053837YC1 (82421210294); 5454001, 00000053983YC3 (82421210275); 5454001, 00000057809YC6 (82421210301); 5454001, 00000062064YC1 (82421210322); 5454001, 00000065436YC8 (82421210336); 5454001, 00000064220YC7 (82421210320); 5454001, 00000048635YC7 (82421210261); 5454001, 00000060271YC4 (82421210259); 5454001, 00000040863YC3 (82421210240); 5454001, 00000065166YC1 (82421190313); 5454001, 00000059562YC9 (82421190287); 5454001, 00000065170YC3 (82421190310); 5454001, 00000064892YC3 (82421190302); 5454001, 00000055097YC0 (82421190261); 5454001, 00000044162YC6 (82421190218); 5454001, 00000047624YC2 (82421190200); 5454001, 00000064493YC0 (82421190305); 5454001, 00000057680YC1 (82421190273); 5454001, 00000041159YC5 (82421190215); 5454001, 00000051684YC9 (82421120411); 5454001, 00000058677YC6 (82421120449); 5454001, 00000048512YC8 (82421120386); 5454001, 00000062060YC9 (82421120471); 5454001, 00000061428YC9 (82421120465); 5454001, 00000048634YC0 (82421120379); 5454001, 00000050998YC4 (82421120405); 5454001, 00000065944YC1 (82421120483); 5454001, 00000059208YC9 (82421120452); 5454001, 00000061272YC1 (82421120459); 5454001, 00000062242YC3 (82421120475); 5454001, 00000065442YC6 (82421120492); 5454001, 00000046245YC7 (82421120342); 5454001, 00000038373YC7 (82421120340); 5454001, 00000056129YC0 (82421120429); 5454001, 00000058212YC2 (82421120436); 5454001, 00000059677YC5 (82421120453); 5454001, 00000060053YC6 (82421120461); 5454001, 00000048362YC8 (82421120373); 5454001, 00000051139YC4 (82421120406); 5454001, 00000038372YC9 (82421120345); 5454001, 00000038376YC0 (82421120347); 5454001, 00000048838YC7 (82421120389); 5454001, 00000050818YC4 (82421120388); 5454001, 00000051804YC3 (82421120397); 5454001, 00000053647YC4 (82421120414); 5454001, 00000058671YC9 (82421120446); 5454001, 00000060624YC4 (82421120467); 5454001, 00000049045YC8 (82421200218); 5454001, 00000060465YC2 (82421200249); 5454001, 00000065441YC8 (82421200256); 5454001, 00000065443YC4 (82421200271); 5454001, 00000062063YC3 (82421200243); 5454001, 00000057810YC4 (82421200237); 5454001, 00000057293YC3 (82421200232); 5454001, 00000056130YC8 (82421200234); 5454001, 00000065758YC5 (82421200268); 5454001, 00000051551YC0 (82421200222); 5454001, 00000054323YC1 (82421200225); 5454001, 00000054682YC0 (82421200231); 5454001, 00000058081YC1 (82421200241); 5454001, 00000045925YC5 (82421200205); 5454001, 00000036419YC0 (82421200197); 5454001, 00000065755YC1 (82421020209); 5454001, 00000056276YC9 (82421020195); 5454001, 00000065593YC6 (82421020204); 5454001, 00000038731YC6 (82421020164); 5454001, 00000058514YC1 (82421020197); 5454001, 00000065760YC1 (82421020045); 5454001, 00000050233YC6 (82421020186); 5454001, 00000053836YC3 (82421030388); 5454001, 00000064893YC1 (82421030430); 5454001, 00000046244YC0 (82421030361); 5454001, 00000053649YC0 (82421030387); 5454001, 00000049044YC1 (82421030376); 5454001, 00000054320YC7 (82421030192); 5454001, 00000052678YC0 (82421030385); 5454001, 00000055961YC7 (82421030396); 5454001, 00000064895YC6 (82421030436); 5454001, 00000065165YC3 (82421030197); 5454001, 00000047046YC8 (82421030370); 5454001, 00000059561YC1 (82421030191); 5454001, 00000057807YC0 (82421030394); 5454001, 00000042809YC4 (82421030357); 5454001, 00000043468YC8 (82421030360); 5454001, 00000044395YC2 (82421030362); 5454001, 00000048511YC0 (82421030186); 5454001, 00000054869YC3 (82421030390); 5454001, 00000064219YC9 (82421030194); 5454001, 00000064218YC1 (82421030203); 5454001, 00000048637YC3 (82421030373); 5454001, 00000039623YC4 (82421320001); 5454001, 00000045923YC0 (82421060118); 5454001, 00000059009YC1 (82421060150); 5454001, 00000056734YC7 (82421060141); 5454001, 00000041970YC5 (82421060112); 5454001, 00000053396YC8 (82421060134); 5454001, 00000043753YC3 (82421300065); 5454001, 00000057357YC6 (82421300068); 5454001, 00000038730YC8 (82421300063); 5454001, 00000049686YC9 (82421280046); 5454001, 00000052565YC9 (82421280047); 5454001, 00000036175YC8 (82421100546); 5454001, 00000050683YC2 (82421100629); 5454001, 00000058672YC7 (82421100690); 5454001, 00000043207YC0 (82421100590); 5454001, 00000058674YC3 (82421100712); 5454001, 00000064065YC6 (82421100777); 5454001, 00000050814YC3 (82421100641); 5454001, 00000053835YC5 (82421100668); 5454001, 00000044884YC5 (82421100614); 5454001, 00000055356YC0 (82421100665); 5454001, 00000065946YC6 (82421100798); 5454001, 00000041971YC3 (82421100584); 5454001, 00000043209YC6 (82421100595); 5454001, 00000048636YC5 (82421100628); 5454001, 00000050684YC0 (82421100632); 5454001, 00000050816YC8 (82421100598); 5454001, 00000065599YC3 (82421100800); 5454001, 00000064846YC9 (82421100758); 5454001, 00000052568YC3 (82421100653); 5454001, 00000056545YC7 (82421100688); 5454001, 00000058513YC3 (82421100720); 5454001, 00000062061YC7 (82421100763); 5454001, 00000040862YC5 (82421100587); 5454001, 00000041155YC3 (82421100580); 5454001, 00000044161YC8 (82421100605); 5454001, 00000053982YC5 (82421100646); 5454001, 00000056279YC3 (82421100686); 5454001, 00000060463YC7 (82421100749); 5454001, 00000046483YC4 (82421100596); 5454001, 00000058371YC6 (82421100716); 5454001, 00000061432YC1 (82421100759); 5454001, 00000047048YC4 (82421100620); 5454001, 00000040622YC3 (82421100560); 5454001, 00000053981YC7 (82421100673); 5454001, 00000052680YC6 (82421100658); 5454001, 00000054162YC3 (82421100667); 5454001, 00000054483YC3 (82421100672); 5454001, 00000065598YC5 (82421100770); 5454001, 00000045633YC5 (82421100602); 5454001, 00000058369YC0 (82421100718); 5454001, 00000062243YC1 (82421100774); 5454001, 00000058215YC5 (82421100713); 5454001, 00000065435YC0 (82421100792); 5454001, 00000038531YC0 (82421100565); 5454001, 00000059210YC5 (82421100728); 5454001, 00000065762YC7 (82421100842); 5454001, 00000035908YC3 (82421130135); 5454001, 00000046485YC9 (82421130145); 5454001, 00000058508YC3 (82421130149); 5454001, 00000049866YC7 (82421310225); 5454001, 00000052084YC1 (82421310231); 5454001, 00000058676YC8 (82421310252); 5454001, 00000065594YC4 (82421310279); 5454001, 00000036715YC1 (82421080114); 5454001, 00000041974YC7 (82421310210); 5454001, 00000051431YC5 (82421310229); 5454001, 00000048837YC9 (82421080130); 5454001, 00000050995YC0 (82421080132); 5454001, 00000056277YC7 (82421310240); 5454001, 00000046486YC7 (82421080126); 5454001, 00000051553YC6 (82421080135); 5454001, 00000060055YC1 (82421080146); 5454001, 00000061274YC7 (82421310245); 5454001, 00000050051YC2 (82421310227); 5454001, 00000064066YC4 (82421310266); 5454001, 00000065756YC9 (82421310275); 5454001, 00000052408YC2 (82421310226); 5454001, 00000059211YC3 (82421310255); 5454001, 00000065439YC2 (82421310274); 5454001, 00000062241YC5 (82421310250); 5454001, 00000041969YC7 (82421310206); 5454001, 00000045634YC3 (82421310213); 5454001, 00000065591YC0 (82421310256); 5454001, 00000049046YC6 (82421230253); 5454001, 00000051138YC6 (82421230243); 5454001, 00000057358YC4 (82421230309); 5454001, 00000058511YC7 (82421230326); 5454001, 00000044163YC4 (82421230236); 5454001, 00000052682YC2 (82421230295); 5454001, 00000055576YC3 (82421230262); 5454001, 00000057681YC9 (82421230320); 5454001, 00000054485YC8 (82421230308); 5454001, 00000050234YC4 (82421230256); 5454001, 00000065754YC4 (82421230369); 5454001, 00000047727YC3 (82421230265); 5454001, 00000052411YC6 (82421230291); 5454001, 00000061427YC1 (82421230328); 5454001, 00000061429YC7 (82421230347); 5454001, 00000049865YC9 (82421230278); 5454001, 00000047316YC5 (82421230246); 5454001, 00000039027YC8 (82421230234); 5454001, 00000060054YC4 (82421230333); 5454001, 00000044885YC2 (82421230248); 5454001, 00000050817YC6 (82421230286); 5454001, 00000056127YC4 (82421230288); 5454001, 00000048513YC6 (82421050092); 5454001, 00000052681YC4 (82421050098); 5454001, 00000057361YC8 (82421050101); 5454001, 00000065169YC5 (82421050110); 5454001, 00000054322YC3 (82421270131); 5454001, 00000051432YC3 (82421270123); 5454001, 00000043465YC4 (82421270116); 5454001, 00000051550YC2 (82421270130); 5454001, 00000040378YC2 (82421250129); 5454001, 00000053646YC6 (82421250145); 5454001, 00000052262YC3 (82421250146); 5454001, 00000065164YC6 (82421250168); 5454001, 00000039620YC0 (82421250125); 5454001, 00000039122YC7 (82421250124); 5454001, 00000053648YC2 (82421140276); 5454001, 00000056735YC4 (82421140296); 5454001, 00000065764YC3 (82421140375); 5454001, 00000027680YC8 (828210086); 5454001, 00000037719YC2 (828210089); 5454001-222, 00000350390HM1 (MP126218); 5454001, 00000025754YC3 (FPG30533); 5454001, 00000026668YC4 (UE260523); 5454001, 00000027866YC3 (DE528478); 5454001, 00000029495YC9 (FE443892); 5454001, 00000025890YC5 (MEX68903); 5454001, 00000034381YC4 (FPG13901); 5454001, 00000028845YC6 (DE528480); 5454001-222, 00000359521HM2 (YP158321); 5454001-222, 00000335045HM1 (FP182A11); 5454001-222, 00000350392HM7 (O063CT01); 5454001, 00000026202YC2 (A5739945); 5454001-122, 00000438204CN0 (479964S660); 5454001-122, 00000436687CN8 (314768MCT01); 5454001-122, 00000433437CN1 (EG1024CT03); 5454001-22, 00000052448YC8 (83093458744); 5454001-122, 00000437363CN5 (913541ER660); 5454001-122, 00000434006CN3 (314205CT3); 5454001-122, 00000433107CN0 (816282CT660); 5454001-122, 00000432898CN5 (SJRCT660128); 5454001-122, 00000435503CN8 (212255CT); 5454001-122, 00000437279CN3 (484884CT660); 5454001-122, 437815CN4 (806775OPT); 5454001, 00000029656YC6 (OH0008); 5454001, 00000034256YC8 (OH0039); 5454001, 00000038532YC8 (OH0038); 5454001, 00000034144YC6 (OH0069); 5454001, 00000038375YC2 (OH0034); 5454001, 00000034787YC2 (OH0018); 5454001, 00000035288YC0 (OH0022); 5454001, 00000026659YC3 (OH0012); 5454001, 00000029507YC1 (OH0002); 5454001, 00000033403YC7 (OH0021); 5454001, 00000029503YC0 (OH0006); 5454001, 00000029509YC7 (OH0005); 5454001, 00000034380YC6 (OH0015); 5454001, 00000027683YC2 (OH0003); 5454001, 00000029508YC9 (OH0019); 5454001, 00000036902YC5 (OH0033); 5454001-22, 00000055749YC6 (OH0050); 5454001, 00000037412YC4 (OH0029); 5454001, 00000036901YC7 (OH0026); 5454001, 00000034786YC4 (OH0016); 5454001, 00000038149YC1 (OH0032); 5454001, 00000030760YC3 (OH0009); 5454001, 00000036900YC9 (OH0024); 5454001, 00000031030YC0 (OH0013); 5454001, 00000029793YC7 (OH0007); 5454001, 00000031719YC8 (OH0014); 5454001, 00000037411YC6 (OH0027); 5454001, 00000037035YC3 (OH0025); 5454001, 00000037589YC9 (OH0031); 5454001, 00000029795YC2 (OH0004); 5454001, 00000034788YC0 (OH0030); 5454001, 00000035021YC5 (OH0020); 5454001, 00000036420YC8 (OH0023); 5454001, 00000037865YC3 (OH0035); 5454001, 00000037410YC8 (OH0028); 5454001-22, 00000042443YC2 (OH0115); 5454001-22, 00000048620YC9 (OH0131); 5454001-22, 00000051661YC7 (OH0202); 5454001-22, 00000042446YC5 (OH0112); 5454001-22, 00000049250YC4 (OH0133); 5454001-22, 00000052572YC5 (OH0160); 5454001-22, 00000053860YC3 (OH0153); 5454001-22, 00000043214YC6 (OH0114); 5454001-22, 00000047321YC5 (OH0124); 5454001-22, 00000050822YC6 (OH0141); 5454001-22, 00000052227YC6 (OH0145); 5454001-22, 00000039802YC4 (OH0111); 5454001-22, 00000052571YC7 (OH0210); 5454001-22, 00000053010YC5 (OH0211); 5454001-22, 00000047930YC3 (OH0155); 5454001-22, 00000050011YC6 (OH0200); 5454001-22, 00000054834YC7 (OH0161); 5454001-22, 00000055081YC4 (OH0159); 5454001-22, 00000055964YC1 (OH0162); 5454001-22, 00000042439YC0 (OH0113); 5454001-22, 00000051944YC7 (OH0144); 5454001-22, 00000047515YC2 (OH0121); 5454001-22, 00000042444YC0 (OH0116); 5454001-122, 00000432824CN1 (OH0142); 5454001-22, 00000056143YC1 (OH0180); 5454001-22, 00000050014YC0 (OH0139); 5454001-22, 00000050183YC3 (OH0135); 5454001-22, 00000039801YC6 (OH0123); 5454001-22, 00000052225YC0 (OH0151); 5454001-22, 00000047932YC9 (OH0126); 5454001-22, 00000042971YC2 (OH0118); 5454001-22, 00000043469YC6 (OH0122); 5454001-22, 00000047053YC4 (OH0125); 5454001-22, 00000056144YC9 (OH0166); 5454001-22, 00000056500YC2 (OH0163); 5454001-22, 00000054116YC9 (OH0154); 5454001-22, 00000054673YC9 (OH0149); 5454001-22, 00000048618YC3 (OH0129); 5454001-22, 00000055965YC8 (OH0157); 5454001-22, 00000040080YC4 (OH0110); 5454001-22, 00000053125YC1 (OH0148); 5454001-22, 00000053415YC6 (OH0156); 5454001-22, 00000051117YC0 (OH0143); 5454001-22, 00000049856YC8 (OH0140); 5454001-22, 00000051662YC5 (OH0146); 5454001-22, 00000043215YC3 (OH0117); 5454001-22, 00000047051YC8 (OH0150); 5454001-22, 00000049069YC8 (OH0132); 5454001-22, 00000049670YC3 (OH0138); 5454001-22, 00000051115YC4 (OH0170); 5454001-22, 00000055551YC6 (OH0165); 5454001-22, 00000056502YC8 (OH0171); 5454001-22, 00000041137YC1 (OH0130); 5454001-22, 00000042005YC9 (OH0120); 5454001-22, 00000052226YC8 (OH0152); 5454001-22, 00000052570YC9 (OH0199); 5454001-22, 00000045915YC6 (OH0119); 5454001-22, 00000054833YC9 (OH0158); 5454001-22, 00000048334YC7 (OH0127); 5454001-22, 00000039803YC2 (OH0108); 5454001-22, 00000040813YC8 (OF0102); 5454001-22, 00000058373YC2 (OJ1002); 5454001, 00000032055YC6 (OG0100); 5454001, 00000031143YC1 (OG0003); 5454001, 00000033397YC1 (OG0006); 5454001, 00000034379YC8 (OG0008); 5454001, 00000037413YC2 (OG0013); 5454001, 00000038528YC6 (OG0015); 5454001-22, 00000040362YC6 (OG0018); 5454001, 00000030119YC2 (OG0002); 5454001, 00000038533YC6 (OG0016); 5454001, 00000029502YC2 (OG0001); 5454001, 00000039354YC6 (OG0020); 5454001, 00000035019YC9 (OG0009); 5454001, 00000035020YC7 (OG0010); 5454001-22, 00000056318YC9 (OG0135); 5454001-22, 00000048142YC4 (OG0202); 5454001-22, 00000053414YC9 (OG0222); 5454001-22, 00000040812YC0 (OG0103); 5454001-22, 00000053011YC3 (OG0131); 5454001-22, 00000048621YC7 (OG0121); 5454001-22, 00000046273YC9 (OG0115); 5454001-22, 00000050179YC1 (OG0122); 5454001-22, 00000053522YC9 (OG0128); 5454001-22, 00000052569YC1 (OG0126); 5454001-22, 00000041142YC1 (OG0104); 5454001-22, 00000040815YC3 (OG0107); 5454001-22, 00000046720YC9 (OG0120); 5454001-22, 00000049858YC4 (OG0124); 5454001-22, 00000045917YC2 (OG0113); 5454001-22, 00000053216YC8 (OG0140); 5454001-22, 00000040814YC6 (OG0106); 5454001-22, 00000052714YC3 (OG0127); 5454001-22, 00000043901YC8 (OG0109); 5454001-22, 00000052229YC2 (OG0130); 5454001-22, 00000055967YC4 (OG0206); 5454001-22, 00000054508YC7 (OG0132); 5454001-22, 00000045305YC0 (OG0112); 5454001-22, 00000042007YC5 (OG0111); 5454001-22, 00000054671YC3 (OG0129); 5454001-22, 00000048333YC9 (OG0119); 5454001-22, 00000055312YC3 (OG0150); 5454001-22, 00000043213YC8 (OG0110); 5454001-22, 00000047319YC9 (OG0117); 5454001-22, 00000039800YC8 (OG0105); 5454001-22, 00000046466YC9 (OG0116); 5454001-22, 00000054831YC3 (OG0133); 5454001-22, 00000041138YC9 (OG0108); 5454001-22, 00000046465YC1 (OG0114); 5454001-22, 00000047516YC0 (OG0123); 5454001-22, 00000049068YC0 (OG0118); 5454001-22, 00000050182YC5 (OG0125); 5454001-22, 00000053520YC3 (OE0105); 5454001-22, 00000045916YC4 (OE0103); 5454001-22, 00000053009YC7 (OE0104); 5454001-100, 00000429355CN1 (CT429355CN1); 5454001-222, 00000353971HM5 (M40480218); 5454001-22, 00000041826YC9 (DZ1279CT02); 5454001-22, 00000044147YC7 (AZ1003CT02); 5454001-22, 00000048529YC2 (902678CT1); 5454001-22, 00000041533YC1 (M2004604); 5454001-222, 00000312790HM9 (M4166100); 5454001-222, 00000305878HM1 (X6002817); 5454001-222, 00000306953HM1 (X141385604); 5454001-222, 00000306952HM3 (A1656608); 5454001-220, 00000298211HM4 (M4499034); 5454001-222, 00000310018HM7 (M2844535); 5454001-222, 00000304752HM9 (M1279740); 5454001-22, 00000042442YC4 (835210197); 5454001-22, 00000042445YC7 (IQ1309CT01); 5454001-22, 00000042794YC8 (IQ1361CT01); 5454001-22, 00000042796YC3 (IQ1361CT02); 5454001-22, 00000048336YC2 (SA2308CT01); 5454001-222, 00000312789HM1 (O020CT03); 5454001-222, 00000320008HM6 (00366CTS04); 5454001-200, 00000302026HM0 (00572CTS03); 5454001-222, 00000322997HM8 (00573CTS01); 5454001-22, 00000046467YC7 (859210105); 5454001-22, 00000044632YC8 (910213167); 5454001-22, 00000044634YC4 (910213173); 5454001-22, 00000044631YC0 (910213168); 5454001-122, 433448CN8 (740633OPT660); 5454001-122, 433789CN5 (304234OPT660); 5454001-222, 00000316559HM4 (A5690005); 5454001-222, 00000318143HM5 (PL2863CT01); 5454001-122, 00000432767CN2 (KAISCT660); 5454001-122, 00000429888CN1 (904CLAYCT); 5454001-100, 00000425716CN8 (301891OP64); 5454001-22, 00000052849YC7 (908859CT660); 5454001-22, 00000058549YC7 (918494ERCT2); 5454001-122, 00000431926CN5 (971KWCT1); 5454001-22, 00000048830YC4 (971KWCT2); 5454001-122, 00000435362CN9 (908859CT1); 5454001-122, 00000430370CN7 (215710EDCT1); 5454001-100, 00000426556CN7 (304243OPT660); 5454001-100, 00000424967CN8 (416438CT1); 5454001-100, 00000430436CN6 (514262CT1); 5454001-222, 00000360236HM4 (NG1309CT01); 5454001-222, 00000361578HM8 (BA4008CT01); 5454001-222, 00000379366HM8 (83021280031214); 5454001-22, 00000060868YC7 (LY1071CT01); 5454001-22, 00000059228YC7 (83113353642); 5454001-22, 00000059029YC9 (83113601124); 5454001-122, 00000438836CN9 (PK1414CT01); 5454001-122, 428651CN4 (480301CT660); 5454001-122, 432048CN7 (718226STHCT); 5454001-122, 00000430693CN2 (516719CT660); 5454001-200, 00000294756HM2 (A5803716); 5454001-222, 00000311885HM8 (M4484743); 5454001-22, 00000042440YC8 (DZ1035CT03); 5454001-220, 00000296363HM5 (M4500619); 5454001-222, 00000030462HM2 (M2634401); 5454001-22, 00000040618YC1 (83021821002812); 5454001, 00000047315YC7 (83021200037512); 5454001-220, 00000296362HM7 (A5327212); 5454001-220, 00000296360HM1 (A5805627); 5454001-220, 00000302024HM5 (A5650212); 5454001, 00000033557YC0 (NG1127CT01); 5454001-22, 00000045308YC4 (RU1875CT03); 5454001-22, 00000045602YC0 (856217009); 5454001-22, 00000047711YC7 (856210164); 5454001-22, 00000040078YC8 (856210156); 5454001-220, 00000297491HM3 (00617CTS03); 5454001-200, 00000295311HM5 (00165CTS03); 5454001-222, 00000310017HM9 (00094CTS02); 5454001-222, 00000304645HM5 (10331CTS50); 5454001-22, 00000042447YC3 (UZ1011CT01); 5454001-22, 00000044637YC7 (910213177); 5454001-22, 00000044854YC8 (910213171); 5454001-100, 00000428917CN9 (501257CT64); 5454001-122, 00000428650CN6 (480342CT660); 5454001-122, 00000430166CN9 (707521OP660); 5454001-122, 00000432107CN1 (760830CT1); 5454001-22, 59667YC6 (386254TLC3); 5454001-100, 00000427728CN1 (386F660CT2); 5454001-122, 00000431277CN3 (561882CTO1); 5454001-122, 00000431384CN7 (815942CT2); 5454001-100, 00000427181CN3 (847292CT660); 5454001-122, 00000434754CN8 (847872CT660); 5454001-122, 00000430274CN1 (620285LCT); 5454001-122, 00000431671CN7 (225768CT1); 5454001-100, 00000427333CN0 (318330CT2); 5454001-122, 00000430360CN8 (504894CT1); 5454001-122, 00000431482CN9 (978466CT1); 5454001-100, 00000427115CN1 (207487CT660); 5454001-122, 00000438153CN9 (828696CT1); 5454001-122, 00000430063CN8 (516674OPTIMA); 5454001-122, 00000432573CN4 (516MHCTER); 5454001-122, 00000430595CN9 (LIJ044434); 5454001-122, 00000433449CN6 (513MWCT2); 5454001-122, 00000434198CN8 (513MWCT1); 5454001-122, 00000437219CN9 (814938CT); 5454001-222, 00000358392HM9 (830500RHCT660S); 5454001-122, 00000430158CN6 (801357CT5); 5454001-100, 00000429510CN1 (262767CT1); 5454001-100, 00000428574CN8 (450771CT1); 5454001-100, 00000429149CN8 (819324CT1); 5454001-100, 00000426555CN9 (A5704204); 5454001-100, 00000430596CN7 (CT430596CN7); 5454001-222, 00000322996HM0 (O009CT05); 5454001-222, 00000319296HM0 (R001CT02); 5454001-222, 00000325773HM0 (R001CT01); 5454001-22, 00000043472YC0 (859210101); 5454001-200, 00000291313HM5 (M4174143); 5454001-222, 00000305881HM5 (00182CTS01); 5454001-222, 00000306527HM3 (M4481023); 5454001-222, 00000308957HM0 (GP1000CT03); 5454001-222, 00000311088HM9 (A5135919); 5454001-222, 00000318142HM7 (B5580615); 5454001-222, 00000310614HM3 (HGV0121); 5454001-222, 00000311884HM1 (UDN0124); 5454001-222, 00000316774HM9 (PL1580CT02); 5454001-222, 00000318144HM3 (PL1196CT01); 5454001-222, 00000318969HM3 (PL1050CT01); 5454001-222, 00000328752HM1 (PL2975CT01); 5454001-222, 00000368072HM5 (00133CTS04); 5454001-200, 00000298632HM1 (00689CTS01); 5454001-22, 00000045060YC1 (910216063); 5454001-100, 00000428576CN3 (502226OPT); 5454001-100, 00000428451CN9 (207777CT660); 5454001-100, 00000426680CN5 (207474CT660); 5454001-100, 00000424970CN2 (816691OPT1); 5454001-22, 00000045599YC8 (816691EROP); 5454001-122, 00000432450CN5 (601968OPCT); 5454001-100, 00000424971CN0 (614MSK660); 5454001-100, 00000427754CN7 (605333660CT); 5454001-200, 00000298631HM3 (A5806329); 5454001-222, 00000315835HM9 (910213170); 5454001-222, 00000337373HM5 (D5363106); 5454001-22, 00000045601YC2 (ET1012CT01); 5454001-22, 00000044633YC6 (EG1757CT02); 5454001-22, 00000047713YC3 (83021060044213); 5454001-22, 00000047712YC5 (83021860044813); 5454001-22, 00000043736YC8 (83021240023512); 5454001-22, 00000043734YC3 (83021000037712); 5454001-22, 00000045597YC2 (83021000037612); 5454001-22, 00000046721YC7 (835210208); 5454001-22, 00000040810YC4 (835210190); 5454001-22, 00000040365YC9 (835210204); 5454001-22, 00000047514YC5 (835210220); 5454001, 00000036717YC7 (835210187); 5454001-22, 00000040620YC7 (835210189); 5454001-222, 00000307771HM6 (B5801733); 5454001-22, 00000046468YC5 (NG1163CT01); 5454001-22, 00000043212YC0 (RU4929CT01); 5454001-22, 00000043900YC0 (RU7564CT01); 5454001-22, 00000041141YC3 (856210160); 5454001-22, 00000049857YC6 (856210153); 5454001-22, 00000044855YC5 (856210159); 5454001-22, 00000050184YC1 (859210112); 5454001-100, 00000428965CN8 (205481OPTIMA); 5454001-122, 00000436963CN3 (530876CT660); 5454001-22, 00000050576YC8 (970SRMCT1); 5454001-122, 00000433788CN7 (954987OPT64); 5454001-122, 00000430874CN8 (352401CT3); 5454001-100, 00000429301CN5 (706278H660); 5454001-22, 00000043737YC6 (706282CT660); 5454001-122, 00000434466CN9 (671USNHCT660); 5454001-100, 00000426955CN1 (773525CT660); 5454001-122, 00000431483CN7 (618288CT2); 5454001-100, 00000429244CN7 (816655OPT); 5454001-100, 00000423878CN8 (816561OPT); 5454001-122, 00000429753CN7 (816455NCT); 5454001-22, 00000046470YC1 (212238CT); 5454001-100, 00000427609CN3 (605725CT2); 5454001-122, 00000429758CN6 (608873CT660); 5454001-100, 00000424452CN1 (304526OPT128); 5454001-122, 00000429604CN2 (814676OP660); 5454001-122, 00000430797CN1 (416756CT1); 5454001-122, 431481CN1 (314251MRCT); 5454001-100, 00000424969CN4 (303SANCT1); 5454001-122, 00000434199CN6 (563422CT66); 5454001-122, 00000432555CN1 (COR522695); 5454001-100, 00000428452CN7 (864529OPTI1); 5454001-100, 00000426911CN4 (907264CT2); 5454001-100, 00000429890CN7 (907443OPT660); 5454001-122, 00000434007CN1 (479BVOPT660); 5454001-100, 00000427263CN9 (501257CT32); 5454001-22, 00000054302YC5 (209SRMCT); 5454001-100, 00000427753CN9 (239LEECT1); 5454001-222, 00000348980HM4 (727588CT660); 5454001-122, 00000434185CN5 (305223CT32); 5454001-100, 00000429529CN1 (808243MLCT); 5454001-122, 00000429553CN1 (641446CT2); 5454001-122, 00000436383CN4 (712722CT66); 5454001-122, 00000433276CN3 (319272ERCT); 5454001-122, 00000431185CN8 (815777CT1); 5454001-100, 00000428022CN8 (815462NLCT); 5454001-100, 00000428021CN0 (574537OP660); 5454001-122, 00000431786CN3 (574753OP660); 5454001-100, 00000427296CN9 (913334OPTCT); 5454001-100, 00000430135CN4 (620221WCT); 5454001-100, 00000429439CN3 (504842BS); 5454001-122, 00000431702CN0 (504894CT2); 5454001-100, 00000425273CN0 (774442CT); 5454001-100, 00000423431CN6 (207532CT660); 5454001-100, 00000427180CN5 (207723CT660); 5454001-100, 00000424454CN7 (207523CT); 5454001-100, 00000428756CN1 (586558OPT); 5454001-222, 00000349592HM6 (417820CT5); 5454001-100, 00000427114CN4 (601883OCT1); 5454001-100, 00000424453CN9 (601928SCT); 5454001-122, 00000431434CN0 (662751OPCT); 5454001-100, 00000428575CN5 (828213CT660); 5454001-22, 00000052076YC7 (828213OCT2); 5454001-122, 00000433859CN6 (GVA660CT); 5454001-100, 00000429150CN6 (910HARNETTCT1); 5454001-22, 00000051428YC1 (910FHHOKECT660); 5454001-100, 00000430167CN7 (WILMVACT); 5454001-100, 00000429528CN3 (603629CT660); 5454001-122, 00000436182CN0 (775STMACT660); 5454001-122, 00000432574CN2 (516796CT660); 5454001-122, 00000431127CN0 (RADNETCT1526); 5454001-100, 00000429509CN3 (518579BS660); 5454001-100, 00000426413CN1 (RADNETCT1515); 5454001-22, 00000050577YC6 (HJCOPT660); 5454001-100, 00000428092CN1 (516256OPTIMA); 5454001-22, 00000051000YC8 (419332CT); 5454001-100, 00000427805CN7 (740532SMCT); 5454001-122, 00000431706CN1 (412942CT660); 5454001-122, 00000430873CN0 (843692GSCT); 5454001-22, 00000049249YC6 (605696660CT); 5454001-100, 00000429699CN2 (817HEBCT660); 5454001-122, 00000436493CN1 (210SLB660); 5454001-122, 00000431927CN3 (817FP660CT); 5454001-100, 00000427163CN1 (434MJOCCCT); 5454001-100, 00000428316CN4 (434MJ29NCT); 5454001-100, 00000428254CN7 (757388CT1); 5454001-100, 00000425220CN1 (425640CT); 5454001-22, 00000051223YC6 (360805CT); 5454001-100, 00000428964CN1 (920996CTOP); 5454001-122, 00000434272CN1 (414219OPTCT2); 5454001-22, 00000043738YC4 (309263MCT); 5454001-22, 00000054675YC4 (83021805544113); 5454001-22, 00000055221YC6 (IQ1394CT02); 5454001-222, 00000308956HM2 (A5702009); 5454001-122, 00000432653CN4 (330305OPT); 5454001-222, 00000315833HM4 (FI1130CT02); 5454001-222, 00000317420HM8 (M4501210); 5454001-200, 00000278512HM9 (M5481247); 5454001-122, 430151CN1 (262741CT1); 5454001-22, 00000047052YC9 (83021281980912); 5454001-22, 00000053523YC7 (853210264); 5454001, 00000038880YC1 (853210251); 5454001-22, 00000054114YC4 (853210269); 5454001-22, 00000052074YC2 (853210263); 5454001, 00000034143YC8 (853210249); 5454001-22, 00000044516YC3 (853210254); 5454001-22, 00000056315YC5 (853210271); 5454001-22, 00000045306YC8 (853210255); 5454001-22, 00000049667YC9 (853210259); 5454001-22, 00000047627YC5 (856217008); 5454001-22, 00000047709YC1 (856217007); 5454001-222, 00000329542HM5 (190171CT02); 5454001-122, 432014CN9 (631968CT660); 5454001-22, 00000047626YC7 (CT47626YC7); 5454001, 00000029504YC8 (910213144); 5454001-222, 00000331759HM1 (080038CT02); 5454001-222, 00000324511HM5 (110059CT02); 5454001-100, 00000426351CN3 (CT426351CN3); 5454001-100, 00000428317CN2 (CT428317CN2); 5454001, 00000033973YC9 (CT33973YC9); 5454001, 00000033404YC5 (CT275691HM4); 5454001-122, 00000437168CN8 (CT437168CN8); 5454001-122, 00000436106CN9 (CT436106CN9); 5454001-122, 00000432537CN9 (CT432537CN9); 5454001-122, 00000432715CN1 (CT432715CN1); 5454001-122, 00000430157CN8 (CT430157CN8); 5454001-122, 00000431276CN5 (CT431276CN5); 5454001, 00000033048YC0 (33048YC0); 5454001-22, 00000047320YC7 (83021841017512); 5454001, 00000038535YC1 (83021280008612); 5454001-22, 00000040809YC6 (83021820008312); 5454001-22, 00000044635YC1 (83021821951912); 5454001, 00000033716YC2 (83021826006817); 5454001-22, 00000044146YC9 (83021820008712); 5454001, 00000038732YC4 (83021821188412); 5454001, 00000032393YC1 (83021200844217); 5454001-22, 00000043899YC4 (83021600023312); 5454001-22, 00000047710YC9 (835210212); 5454001, 00000033051YC4 (A5106406); 5454001-122, 00000438389CN9 (CT59360YC8); 5454001, 00000031297YC5 (PK1248CT01); 5454001-122, 00000440774CN8 (CT440774CN8); 5454001, 00000027882YC0 (PL1639CT02); 5454001, 00000026660YC1 (PL1715CT02); 5454001-100, 00000429806CN3 (CT429806CN3); 5454001-222, 00000332384HM7 (O006CT01); 5454001-222, 00000319295HM2 (P003CT02); 5454001-200, 00000295310HM7 (O009CT04); 5454001-222, 00000313522HM5 (10673CTS01); 5454001, 00000027881YC2 (00243CTS10); 5454001-200, 00000299339HM2 (00362CTS02); 5454001-200, 00000298630HM5 (00557CTS02); 5454001-222, 00000326338HM1 (00510CTS02); 5454001-200, 00000298633HM9 (10223CTS01); 5454001-222, 00000331562HM9 (00181CTS01); 5454001-100, 00000430275CN8 (CT430275CN8); 5454001-122, 00000431703CN8 (CT431703CN8); 5454001, 00000037718YC4 (CT37718YC4); 5454001-122, 00000435500CN4 (7947); 5454001-122, 00000435572CN3 (7948); 5454001-122, 00000437327CN0 (7744); 5454001-122, 431705CN3 (504897CT1); 5454001-222, 00000305879HM9 (CS1339CT01); 5454001, 00000026841YC7 (PC0815CT06); 5454001-222, 00000344557HM4 (01065CTS01); 5454001, 00000032057YC2 (00084CTS01); 5454001, 00000033838YC4 (70431CTS06); 5454001-200, 00000288270HM2 (00226CTS04); 5454001-222, 00000308327HM6 (00632CTS03); 5454001-222, 00000322648HM7 (00249CTS02); 5454001-222, 00000305384HM0 (00094CTS03); 5454001-22, 00000044636YC9 (910213175); 5454001-222, 00000330500HM0 (10260CTS02); 5454001-222, 00000347976HM3 (543201CT02); 5454001-222, 00000350389HM3 (584237CT02); 5454001-222, 00000360841HM1 (IE1164CT01); 5454001-222, 00000332386HM2 (BY4002CT01); 5454001-222, 00000347524HM1 (BY4035CT02); 5454001-122, G56587171P1 (902742CTOPT); 5454001-222, 00000333250HM9 (M4046579); 5454001-222, 00000327471HM9 (125607CT02); 5454001, 00000050382YC1 (82421030185); 5454001-22, 00000051224YC4 (6073CT0001); 5454001-22, 00000051940YC5 (83021860022613); 5454001-22, 00000056739YC6 (83021801024614); 5454001-222, 00000374879HM5 (83021801754914); 5454001-22, 00000048140YC8 (835210211); 5454001-222, 00000380363HM2 (83021826635614); 5454001-222, 00000347522HM5 (110017CT02); 5454001-222, 00000334388HM6 (786090CT06); 5454001, 00000025893YC9 (83021847082316); 5454001-22, 00000054117YC7 (859210120); 5454001-22, 00000055966YC6 (83104558262); 5454001-22, 00000053861YC1 (83093535530); 5454001-22, 00000058715YC4 (83021060469714); 5454001-222, 00000361637HM2 (83021860835514); 5454001-222, 00000381059HM5 (83021862060814); 5454001-22, 00000057383YC2 (83021860011214); 5454001-22, 00000055751YC2 (83021200048614); 5454001-222, 00000380364HM0 (83021201532614); 5454001-22, 00000048826YC2 (83021600048413); 5454001-22, 00000048622YC5 (83021800047813); 5454001-22, 00000054509YC5 (83021803915314); 5454001-22, 00000054510YC3 (83021844434613); 5454001-22, 00000005150YC9 (83021020019813); 5454001-22, 00000051505YC6 (83021021009713); 5454001-22, 00000048335YC4 (83021000047913); 5454001-22, 00000051939YC7 (RU8410CT01); 5454001-22, 00000048827YC0 (859210111); 5454001, 00000033050YC6 (360966CT01); 5454001-222, 00000325268HM1 (910213178); 5454001-222, 00000348979HM6 (318798OPT); 5454001-222, 00000325267HM3 (910212234); 5454001-222, 00000318175HM7 (910212232); 5454001-222, 00000321958HM1 (910215045); 5454001-222, 00000322644HM6 (910215046); 5454001-222, 00000356106HM5 (AZ4029CT01); 5454001-222, 00000334389HM4 (EG1537CT02); 5454001-222, 00000342015HM5 (108502CT01); 5454001-222, 00000338338HM7 (H4118CT04); 5454001-222, 00000343754HM8 (H10638CT03); 5454001-222, 00000330497HM9 (H42261CT03); 5454001-222, 00000360234HM9 (B5167330); 5454001-222, 00000338339HM5 (B5810531); 5454001-22, 00000347521HM7 (B5258018); 5454001-222, 00000356481HM2 (A5730101); 5454001-222, 00000340665HM9 (A5293310); 5454001-222, 00000354748HM6 (A34802401); 5454001-222, 00000332387HM0 (A51043196); 5454001-222, 00000356480HM4 (A99155709); 5454001-222, 00000321690HM0 (A5649014); 5454001-222, 00000331626HM2 (A58080142); 5454001-222, 00000385300HM9 (KA1035CT01); 5454001-222, 00000396564HM7 (KA1149CT01); 5454001-222, 00000382292HM1 (KA2875CT01); 5454001-222, 00000344559HM0 (KA2849CT01); 5454001-222, 00000385143HM3 (KA1144CT01); 5454001-222, 00000345654HM8 (KA2842CT01); 5454001-222, 00000359520HM4 (671645CT01); 5454001-222, 00000350391HM9 (UA4295CT01); 5454001-22, 00000057273YC5 (835210251); 5454001-22, 00000059665YC0 (PK1415CT01); 5454001-22, 00000059873YC0 (PK1406CT01); 5454001-22, 00000060985YC9 (856217013); 5454001-22, 00000051429YC9 (856210168); 5454001-22.00000053217YC6 (835210230); 5454001-22, 00000061905YC6 (5512CT0001); 5454001-22, 00000054835YC4 (835210242); 5454001-22, 00000060869YC5 (835210238); 5454001-22, 00000051118YC8 (835210222); 5454001-22, 00000054300YC9 (835210233); 5454001-22, 00000053127YC7 (835210227); 5454001-22, 00000055550YC8 (835210243); 5454001-22, 00000059874YC8 (859210141); 5454001-22, 00000054115YC1 (859210118); 5454001-22, 00000061478YC4 (760241CT); 5454001-22, 60102YC1 (904819CT1); 5454001-22, 50820YC0 (718830CTMAIN); 5454001-22, 00000058086YC0 (570622CT660); 5454001-22, 52851YC3 (972566ERCT); 5454001-222, 00000330498HM7 (83089988618); 5454001-22, 00000059365YC7 (CY1015CT01); 5454001-22, 00000047714YC1 (83021820045913); 5454001-22, 00000047929YC5 (835210210); 5454001-22, 00000051793YC8 (835210224); 5454001-22, 00000051427YC3 (JO1147CT01); 5454001, DUMFMI25462002 (KA2882CT01); 5454001-22, 00000056316YC3 (KA2857CT01); 5454001, 00000055230YC7 (CT55230YC7); 5454001-22, 00000049251YC2 (RU8071CT01); 5454001-22, 00000053858YC7 (RU5425CT01); 5454001-22, 00000049665YC3 (RU3885CT01); 5454001-22, 00000051116YC2 (RU7189CT01); 5454001-22, 00000060986YC7 (RU1031CT02); 5454001-22, 00000055315YC6 (600157CT01); 5454001-22, 00000051943YC9 (RU1274CT02); 5454001-22, 00000053128YC5 (RU8471CT01); 5454001-22, 00000051942YC1 (RU8411CT01); 5454001-22, 00000050015YC7 (RU8151CT01); 5454001-22, 00000055752YC0 (ZA2614CT01); 5454001-222, 00000341586HM6 (BY4161CT01); 5454001-222, 00000333249HM1 (BY4035CT03); 5454001-222, 00000347016HM8 (BY4173CT01); 5454001-222, 00000347017HM6 (BY4062CT01); 5454001-222, 00000330913HM5 (BY4172CT01); 5454001-100, 00000430176CN8 (250368CT1); 5454001-222, 00000326339HM9 (M4206629); 5454001-222, 00000339676HM9 (PC0114CT05); 5454001-222, 00000342992HM5 (A5311013); 5454001-222, 00000343753HM0 (A5360319); 5454001-222, 00000350393HM5 (A5156841); 5454001-222, 00000333904HM1 (A5816630); 5454001-222, 00000348041HM5 (A5156902); 5454001-222, 00000334386HM0 (A5131868); 5454001-222, 00000335044HM4 (C5304410); 5454001-222, 00000333902HM5 (A5816716); 5454001-222, 00000335047HM7 (B5199701); 5454001-222, 00000335043HM6 (A51581133); 5454001-222, 00000335046HM9 (A51581128); 5454001-222, 00000332385HM4 (A51529111); 5454001-222, 00000335768HM8 (B5346356); 5454001-222, 00000359792HM9 (A5804429); 5454001-222, 00000333248HM3 (A5354505); 5454001-222, 00000333903HM3 (A5298906); 5454001-222, 00000348780HM8 (A5261110); 5454001-222, 00000345655HM5 (B5346007); 5454001-222, 00000329546HM6 (C5294650); 5454001-222, 00000320791HM7 (A5814920); 5454001-222, 00000321954HM0 (B5140625); 5454001-222, 00000328754HM7 (C5373407); 5454001-222, 00000320790HM9 (C5808609); 5454001-222, 00000327469HM3 (A5356260); 5454001-222, 00000321955HM7 (B5740302); 5454001-222, 00000329543HM3 (A5116066); 5454001-222, 00000348402HM9 (KA2847CT01); 5454001-222, 00000327472HM7 (LV4056CT01); 5454001-222, 00000342924HM8 (MT1019CT01); 5454001-222, 00000356816HM9 (PL1398CT01); 5454001-222, 00000341584HM1 (PL3039CT01); 5454001-222, 00000342014HM8 (PL2342CT01); 5454001-222, 00000360235HM6 (PL3131CT01); 5454001-222, 00000353230HM6 (RO1084CT03); 5454001-22, 00000057978YC9 (208743CT660); 5454001-22, 00000056317YC1 (IQ1540CT02); 5454001-22, 00000053521YC1 (PK1385CT01); 5454001-22, 00000051222YC8 (843210122); 5454001-22, 00000050180YC9 (RU8159CT01); 5454001-22, 00000047934YC5 (RU3303CT01); 5454001-22, 00000048141YC6 (RU6261CT01); 5454001, 00000026842YC5 (GOLD011235); 5454001, 00000026836YC7 (M4169033); 5454001, 00000027384YC7 (X4482707); 5454001, 00000030499YC8 (519546CT01); 5454001, 00000025762YC6 (A5764003); 5454001, 00000027865YC5 (RU1716CT01); 5454001, 00000027864YC8 (SA1133CT02); 5454001-122, 00000436148CN1 (256517CT1); 5454001-222, 00000350425HM5 (870364CT64); 5454001-122, 00000431843CN2 (928729CT); 5454001-122, 00000434793CN6 (805489CT1); 5454001-122, 438389CN9 (661KMCCT); 5454001-122, Not Available (20957CT1); 5454001-122, 437738CN8 (203732CT660); 5454001-122, 00000434829CN8 (352596CT660); 5454001-22, 00000061297YC8 (712265CT); 5454001-122, 00000436837CN9 (815544OPT1); 5454001-122, 00000434214CN3 (815968OPT1); 5454001-122, 00000436855CN1 (815968OPT2); 5454001-122, 00000438154CN7 (207795CT02); 5454001-122, 00000434986CN6 (906932OPT660); 5454001-222, 00000349563HM7 (406327AI660); 5454001-122, 00000437085CN4 (910642CT1); 5454001-122, 00000436306CN5 (MHCT660); 5454001-122, 00000429717CN2 (716845OPCT1); 5454001-122, 00000437130CN8 (423639CT660); 5454001-122, 00000432369CN7 (615769SKYCT); 5454001-122, 00000431670CN9 (713393EHCT); 5454001-122, 00000434271CN3 (435716CT); 5454001-122, 00000434183CN0 (801314CT2); 5454001-122, 437030CN0 (276236CT660); 5454001-122, 00000430998CN5 (608263CT660); 5454001-122, 00000442211CN9 (CT442211CN9); 5454001-122, 00000436590CN4 (CT436590CN4); 5454001-122, 00000434901CN5 (CT434901CN5); 5454001-122, 00000441968CN5 (CT441968CN5); 5454001-122, 00000437278CN5 (CT437278CN5); 5454001-122, 00000434985CN8 (CT434985CN8); 5454001-122, 00000436266CN1 (CT436266CN1); 5454001, 00000052567YC5 (082421070337A); 5454001, 00000031145YC6 (A5566930); 5454001, 00000030115YC0 (LT4022CT02); 5454001-22, 00000058244YC5 (870845CT660); 5454001-100, 00000423056CN1 (231547CT660); 5454001-22, 00000054832YC1 (803799CT); 5454001-122, 00000437867CN5 (304526COPT660); 5454001-122, 00000441555CN0 (CT441555CN0); 5454001-122, 00000435839CN6 (CT435839CN6); 5454001-122, 00000442453CN7 (CT442453CN7); 5454001-100, 00000426412CN3 (902567CT2); 5454001, 00000025755YC0 (HR4189CT01); 5454001, 00000029655YC8 (83021800040711); 5454001-200, 00000274513HM1 (B5357650); 5454001-200, 00000291312HM7 (A5153626); 5454001, 00000027871YC3 (NG1080CT01); 5454001, 00000027872YC1 (NG1081CT01); 5454001, 00000031140YC7 (ZA2195CT02); 5454001-222, 00000331758HM3 (A020CT06); 5454001-122, 00000440628CN6 (CT440628CN6); 5454001-122, 00000438991CN2 (CT438991CN2); 5454001-222, 00000325767HM2 (GOLD011214); 5454001-22, 00000052713YC5 (83092799894); 5454001-222, 00000349593HM4 (562799OPTM); 5454001-122, 00000435838CN8 (408259EDCT); 5454001-122, 00000431595CN8 (303KRCT1); 5454001-122, 00000437877CN4 (386FNSB660C1); 5454001-122, 00000435209CN2 (561882CT3); 5454001-122, 00000435571CN5 (912466SER); 5454001-122, 00000433450CN4 (808HRAWCT); 5454001-122, 00000435300CN9 (812GW660); 5454001-122, 00000435274CN6 (765747CT128); 5454001-122, 00000435733CN1 (270522CT); 5454001-222, 00000349564HM5 (617355ER660); 5454001-122, 00000436637CN3 (269966CT660); 5454001-122, 00000434345CN5 (231832OPT660); 5454001-122, 00000435237CN3 (228575CT660); 5454001-222, 00000353922HM8 (732840OPTIMA); 5454001-122, 00000435950CN1 (513624OPT); 5454001-122, 00000433446CN2 (484503CT660); 5454001-122, 00000435948CN5 (864560ED660); 5454001-122, 00000434265CN5 (281348KCTO); 5454001-122, 00000436732CN2 (214648CT); 5454001-22, 57385YC7 (281MHSLCT); 5454001-122, 00000439344CN3 (CT439344CN3); 5454001-122, 00000441866CN1 (CT441866CN1); 5454001-122, 434751CN4 (867975CT); 5454001-122, 438486CN3 (902258CT); 5454001-122, 436686CN0 (416864CT2); 5454001-122, 437801CN4 (647439CT); 5454001-122, 00000437362CN7 (CT437362CN7); 5454001, 00000039026YC0 (SA2125CT02); 5454001-122, 00000438309CN7 (CT438309CN7); 5454001-22, 00000057272YC7 (CT57272YC7); 5454001-22, 00000057271YC9 (CT57271YC9); 5454001-22, 00000049669YC5 (CT49669YC5); 5454001-22, 00000061614YC4 (CT61614YC4); 5454001, 00000030501YC1 (DZ1269CT02); 5454001, 00000030504YC5 (910212207); 5454001, 00000030117YC6 (BH1037CT01); 5454001, 00000031714YC9 (600102CT03); 5454001, 00000031722YC2 (83021801106111); 5454001, 00000035910YC9 (83021040029711); 5454001-22, 00000039799YC2 (83021600005612); 5454001, 00000031142YC3 (83021601101411); 5454001, 00000038378YC6 (83021020045712); 5454001, 00000029391YC0 (A5385833); 5454001, 00000029505YC5 (M5423102); 5454001, 00000031291YC8 (A5356929); 5454001, 00000031296YC7 (A5364454); 5454001-220, 00000297492HM1 (D5137202); 5454001-200, 00000278511HM1 (A5146408); 5454001, 00000033968YC9 (A5190311); 5454001, 00000031294YC2 (A5355215); 5454001, 00000033839YC2 (A5355219); 5454001, 00000028846YC4 (A5419866); 5454001, 00000031295YC9 (A5419879); 5454001, 00000033402YC9 (A5801307); 5454001, 00000029506YC3 (A5403908); 5454001, 00000036898YC5 (KA2792CT01); 5454001, 00000039619YC2 (834210043); 5454001, 00000053397YC6 (CT53397YC6); 5454001-22, 00000056145YC6 (CT56145YC6); 5454001, 00000033715YC4 (MA1113CT04); 5454001-222, 00000325428HM1 (BON0120); 5454001, 00000030505YC2 (NG1043CT01); 5454001, 00000036177YC4 (NG1097CT01); 5454001, 00000029390YC2 (RU2283CT01); 5454001, 00000029119YC5 (RU3896CT01); 5454001, 00000035906YC7 (RU1017CT02); 5454001, 00000034382YC2 (RU3998CT01); 5454001, 00000031144YC9 (SK1111CT01); 5454001, 00000029120YC3 (ZA2372CT01); 5454001-22, 00000042009YC1 (ZA2101CT02); 5454001, 00000036613YC8 (856210149); 5454001, 00000032058YC0 (07027CTS01); 5454001, 00000033192YC6 (34338CTS01); 5454001-222, 00000305385HM7 (439560CT02); 5454001-222, 00000310574HM9 (AP6264CT02); 5454001-222, 00000345518HM5 (M2533226); 5454001-220, 00000301351HM3 (PC1401CT02); 5454001, 00000033554YC7 (489691CT01); 5454001-222, 00000382291HM3 (83021284452214); 5454001, 00000029389YC4 (PL1517CT03); 5454001, 00000025753YC5 (150210CT01); 5454001-122, Not Available (408259MNCT); 5454001-100, 00000427083CN1 (760806SCOCT); 5454001-122, 00000430490CN3 (912530WCT); 5454001-100, 427663CN0 (816LSMISH660); 5454001-22, 00000057980YC5 (281METHSPCT); 5454001-222, 00000320009HM4 (910215044); 5454001, 00000028982YC7 (910214070); 5454001, 00000065943YC3 (82421100911); 5454001, 00000031141YC5 (856210132); 5454001-22, 00000061296YC0 (CT61296YC0); 5454001, 00000037588YC1 (910213159); 5454001, 00000025894YC7 (M4160463); 5454001, 00000027224YC5 (M4143826); 5454001, 00000027376YC3 (M4175530); 5454001, 00000025892YC1 (X1299801); 5454001, 00000031293YC4 (M4198095); 5454001, 00000026667YC6 (83021041009610); 5454001, 00000025758YC4 (IE1025CT03); 5454001, 00000026198YC2 (00488CTS03); 5454001, 00000026838YC3 (662455CT01); 5454001-222, 00000315834HM2 (910213166); 5454001-220, 00000296361HM9 (910213162); 5454001, 00000032395YC6 (910213153); 5454001, 00000033049YC8 (910216058); 5454001, 00000027879YC6 (110013CT02); 5454001, 00000031721YC4 (M4480257); 5454001, 00000033348YC4 (X128968003); 5454001, 00000032397YC2 (M4189437); 5454001, 00000031716YC4 (M4191035); 5454001, 00000029511YC3 (M4490510); 5454001, 00000031290YC0 (M4854018); 5454001, 00000025316YC1 (M4497261); 5454001, 00000035647YC7 (910212217); 5454001, 00000033187YC6 (910212211); 5454001, 00000034782YC3 (910212212); 5454001, 00000031569YC7 (910212208); 5454001, 00000035649YC3 (910212219); 5454001, 00000026664YC3 (910216055); 5454001, 00000031713YC1 (543500CT02); 5454001, 00000027379YC7 (PC3312CT03); 5454001, 00000029658YC2 (83021002085011); 5454001, 00000029388YC6 (83021000040411); 5454001, 00000026837YC5 (C5807433); 5454001, 00000027387YC0 (A5228511); 5454001, 00000033398YC9 (SPK0121); 5454001, 00000027377YC1 (SA1052CT02); 5454001, 00000027383YC9 (W003CT01); 5454001, 00000030758YC7 (W003CT02); 5454001, 00000026840YC9 (16017CTS01); 5454001, 00000027681YC6 (10638CTS01); 5454001, 00000028981YC9 (10433CTS02); 5345001, 00000426073CN3 (00039CTS02); 5454001, 00000027225YC2 (361416CT01); 5454001, 00000034784YC9 (SA2161CT02); 5454001-200, 00000288269HM4 (M4161645); 5454001, 00000034254YC3 (M4161650); 5454001, 00000034257YC6 (M4166996); 5454001, 00000031717YC2 (M4467032); 5454001, 00000033975YC4 (M4174528); 5454001, 00000032394YC9 (M4184870); 5454001, 00000034781YC5 (M4400634); 5454001, 00000033974YC7 (P4211492); 5454001-22, 00000042793YC0 (910215043); 5454001-222, 00000308329HM2 (AH5212CT01); 5454001-200, 00000280836HM8 (HC0220CT05); 5454001-200, 00000280835HM0 (HC0881CT06); 5454001, 00000033047YC2 (JO1138CT01); 5454001, 00000034780YC7 (AH5372CT02); 5454001, 00000034378YC0 (FI1003CT06); 5454001, 00000038147YC5 (83021240049012); 5454001-200, 00000299980HM3 (B5300712); 5454001-220, 00000301352HM1 (A5137711); 5454001-200, 00000299981HM1 (A5128419); 5454001, 00000034783YC1 (OM1005CT01); 5454001, 00000034145YC3 (RO1095CT01); 5454001, 00000036040YC4 (RO1093CT01); 5454001-222, 00000311089HM7 (RO1098CT01); 5454001, 00000034785YC6 (RU2551CT01); 5454001, 00000035907YC5 (RU1089CT05); 5454001, 00000036038YC8 (RU2451CT01); 5454001, 00000035769YC9 (RU3906CT01); 5454001, 00000035768YC1 (RU8412CT01); 5454001-200, 00000299341HM8 (U001CT04); 5454001-200, 00000294755HM4 (150010CT01); 5454001, 00000033840YC0 (UZ1018CT01); 5454001-22, 00000043471YC2 (910214079); 5454001-22, 00000045059YC3 (910214080); 5454001-22, 00000043735YC0 (910213165); 5454001-222, 00000333251HM7 (122002CT01); 5454001-222, 00000325772HM2 (475430CT06); 5454001, 00000032054YC9 (83021204997411); 5454001-22, 00000041140YC5 (835210195); 5454001-122, 00000441652CN5 (CT441652CN5); 5454001, 00000031570YC5 (835210172); 5454001, 00000035765YC7 (835210177); 5454001, 00000031571YC3 (DZ1406CT01); 5454001, 00000033400YC3 (DZ1064CT02); 5454001, 00000032396YC4 (910213151); 5454001, 00000033555YC4 (910216059); 5454001-100, 00000429530CN9 (CT429530CN9); 5454004-22, 00000340293HM0 (082421100640); 5454001, 00000035022YC3 (83021000028711); 5454001, 00000028983YC5 (83021245356411); 5454001, 00000035289YC8 (83021600028811); 5454001, 00000038377YC8 (83021600048912); 5454001, 00000031720YC6 (83021061103811); 5454001, 00000036416YC6 (83021060031511); 5454001-22, 00000042438YC2 (83021802764013); 5454001-22, 00000042008YC3 (83021020014912); 5454001-22, 00000045598YC0 (83021600033912); 5454001-220, 00000302027HM8 (B5810110); 5454001-220, 00000301349HM7 (B5810213); 5454001, 00000038530YC2 (834210042); 5454001, 00000037279YC7 (RO1223CT01); 5454001, 00000036714YC4 (RU3894CT01); 5454001, 00000036037YC0 (RU2652CT01); 5454001-200, 00000299340HM0 (SK1004CT03); 5454001, 00000033971YC3 (ZA2119CT02); 5454001, 00000033399YC7 (A043CT02); 5454001, 00000028980YC1 (426020CT03); 5454001-22, 00000044145YC1 (856210157); 5454001, 00000033349YC2 (856210137); 5454001-22, 00000042973YC8 (856210163); 5454001, 00000035767YC3 (TN1003CT02); 5454001, 00000035766YC5 (TN1530CT01); 5454001, 00000030502YC9 (19002CTS04); 5454001, 00000029659YC0 (00521CTS01); 5454001-222, 00000304753HM7 (10448CTS03); 5454001-222, 00000310573HM1 (PL1731CT02); 5454001-222, 00000335769HM6 (10683CTS01); 5454001-222, 00000342926HM3 (00242CTS04); 5454001-222, 00000345274HM5 (00242CTS02); 5454001-222, 00000359793HM7 (21023CTS01); 5454001-222, 00000360227HM3 (21023CTS02); 5454001-222, 00000356356HM6 (10537CTS01); 5454001-222, 00000362607HM4 (20297CTS02); 5454001-222, 00000342018HM9 (10618CTS01); 5454001-222, 00000330912HM7 (80040CTS03); 5454001-22, 00000051426YC5 (83108645065); 5454001-22, 00000352854HM4 (83021005576014); 5454001-22, 00000054672YC1 (83021205319814); 5454001-22, 00000050430YC8 (835210215); 5454001-222, 00000329544HM1 (MA1493CT01); 5454001-222, 00000353227HM2 (BG4555CT02); 5454001-222, 00000340227HM8 (A6047918); 5454001-222, 00000356179HM2 (M41439101); 5454001-222, 00000335766HM2 (M9624247); 5454001-222, 00000335767HM0 (M2875114); 5454001-222, 00000356354HM1 (M348992901); 5454001-222, 00000321956HM5 (409868CT03); 5454001-222, 00000356355HM8 (A5375131); 5454001-222, 00000363931HM7 (A5399602); 5454001-222, 00000336752HM1 (475150CT03); 5454001-222, 00000328083HM1 (476780CT01); 5454001-222, 00000347523HM3 (34380CTS01); 5454001-222, 00000359093HM2 (34388CTS01); 5454001-222, 00000360842HM9 (34386CTS02); 5454001-222, 00000360843HM7 (34387CTS02); 5454001-222, 00000360844HM5 (34389CTS01); 5454001-222, 00000327367HM9 (23012CTS01); 5454001-22, 00000051002YC4 (83101807198); 5454001-22, 00000051502YC3 (83097189427); 5454001-22, 00000051945YC4 (835210231); 5454001-22, 00000050431YC6 (835210216); 5454001-22, 00000059664YC3 (835210254); 5454001-22, 00000047933YC7 (DZ1027CT02); 5454001-22, 00000055082YC2 (H2014CT08); 5454001-22, 00000055548YC2 (83021008838914); 5454001-22, 00000045913YC1 (83021868323513); 5454001-22, 00000047931YC1 (83021000260813); 5454001-22, 00000050999YC1 (83021000027013); 5454001-22, 00000050181YC7 (83021800007213); 5454001-22, 00000050823YC4 (83021802605013); 5454001-22, 00000052850YC5 (83021605570514); 5454001-22, 00000048829YC6 (83021800050213); 5454001-22, 00000051791YC2 (83021840415814); 5454001-22, 00000050999YC2 (83021000013613); 5454001-22, 00000058243YC7 (834210048); 5454001-22, 00000053219YC2 (85099024386); 5454001-22, 00000057979YC7 (RU8755CT01); 5454001-22, 00000046723YC3 (RU4809CT01); 5454001-22, 00000048623YC3 (RU1516CT01); 5454001-22, 00000046722YC5 (RU4806CT01); 5454001-22, 00000048143YC2 (RU8036CT01); 5454001-22, 00000053220YC0 (856210177); 5454001-22, 00000055313YC1 (856210172); 5454001-22, 00000056501YC0 (856210020); 5454001-22, 00000052449YC6 (856210171); 5454001-22, 00000055316YC4 (TN1641CT01); 5454001-22, 00000056503YC6 (06135CTS02); 5454001-222, 00000357155HM1 (M4475675); 5454001-222, 00000345651HM4 (HC3055CT02); 5454001-222, 00000327470HM1 (LV4005CT02); 5454001-222, 00000326863HM8 (M1082421); 5454001-222, 00000353228HM0 (110901CT03); 5454001-222, 00000333905HM8 (M4471526); 5454001-222, 00000354747HM8 (G4200409); 5454001-222, 00000339678HM5 (M9409547); 5454001-222, 00000340666HM7 (M185279205); 5454001-222, 00000337372HM7 (M2265923); 5454001-222, 00000336131HM8 (A128275402); 5454001-222, 00000336753HM9 (M9483444); 5454001-222, 00000353229HM8 (M4464422); 5454001-222, 00000342017HM1 (C4200445); 5454001-222, 00000334387HM8 (M4201119); 5454001-222, 00000336134HM2 (M4206223); 5454001-222, 00000328082HM3 (M94838138); 5454001-222, 00000330911HM9 (M4144632); 5454001-222, 00000339677HM7 (GP1007CT02); 5454001-222, 00000329545HM8 (A5103348); 5454001-222, 00000331627HM0 (600062CT03); 5454001-222, 00000330499HM5 (RO1104CT02); 5454001-222, 00000348781HM6 (BD007CT02); 5454001-222, 00000352685HM2 (O010CT02); 5454001-222, 00000353969HM9 (P007CT01); 5454001-222, 00000365892HM9 (O001CT01); 5454001-22, 00000049668YC7 (83091661063); 5454001-22, 00000057384YC0 (MA1521CT01); 5454001-22, 00000053218YC4 (834210045); 5454001-22, 00000050013YC2 (PL1030CT01); 5454001-22, 00000061613YC6 (RU9155CT01); 5454001-22, 00000055083YC0 (RU8615CT01); 5454001-22, 00000051794YC6 (SA2352CT01); 5454001-22, 00000050012YC4 (SA2339CT01); 5454001-22, 00000051119YC6 (SA2396CT01); 5454001-22, 00000057382YC4 (SA1339CT02); 5454001-22, 00000053008YC9 (AE1009CT04); 5454001-122, 00000433447CN0 (600151CT02); 5454001-222, 00000328753HM9 (00615CTS01); 5454001, 00000030120YC0 (1364533); 5454001, 00000035290YC6 (CT35290YC6); 5454001, 00000030500YC3 (CT30500YC3); 5454001, 00000033969YC7 (CT33969YC7); 5454001-100, 00000425274CN8 (CT425274CN8); 5454001, 00000033970YC5 (PL1321CT01); 5454001-122, 430670CN0 (716845OPCT2); 5454001, 00000031718YC0 (M9202623); 5454001, 00000033188YC4 (M24481301); 5454001, 00000030503YC7 (M4459617); 5454001, 00000033190YC0 (B5343113); 5454001-200, 00000294437HM9 (SK1126CT01); 5454001-200, 00000280833HM5 (SK1044CT01); 5454001-22, 00000059366YC5 (864255660); 5454001-222, 00000358066HM9 (561969FFCT); 5454001-222, 00000353923HM6 (469608CT660); 5454001-222, 00000365999HM2 (ET1028CT01); 5454001-222, 00000342016HM3 (X201514809); 5454001-222, 00000337370HM1 (M2470513); 5454001-222, 00000340664HM2 (M2312407); 5454001-222, 00000338337HM9 (A1588535); 5454001-222, 00000340667HM5 (X226587705); 5454001-222, 00000340668HM3 (M161913207); 5454001-222, 00000340226HM0 (M9960539); 5454001-222, 00000345653HM0 (SK1056CT03); 5454001-22, 00000058087YC8 (954327P64); 5454001-22, 59875YC5 (843660CT); 5454001-22, 60614YC5 (615328CT2); 5454001-22, 60615YC2 (512901PCT); 5454001-22, 00000051225YC1 (540636OP660); 5454001-22, 00000058374YC0 (715930CT); 5454001-22, 00000051659YC1 (RU8279CT01); 5454001-22, 00000051660YC9 (RU8278CT01); 5454001-222, 00000357156HM9 (M4215006); 5454001-222, Not Available (940384CT660); 5454001-222, 00000357100HM7 (MHPEARCT); 5454001-222, 00000344558HM2 (060055CT01); 5454001-22, 00000347977HM1 (FI1016CT03); 5454001-222, 00000341585HM8 (X307050801); 5454001-222, 00000342925HM5 (X171897206); 5454001-222, 00000347015HM0 (M4480021); 5454001-222, 00000348042HM3 (M334214301); 5454001-222, 00000342923HM0 (M2748733); 5454001-222, 00000370165HM3 (M5172571); 5454001-222, 00000336751HM3 (M4482625); 5454001-222, 00000368073HM3 (M2746023); 5454001-222, 00000348401HM1 (M4491118); 5454001-222, 00000336133HM4 (M2717510); 5454001-222, 00000365991HM9 (S41600105); 5454001-222, 00000368067HM5 (S41600109); 5454001-222, 00000327468HM5 (XM5480901); 5454001-222, 00000327368HM7 (GF1006CT02); 5454001-222, 00000337371HM9 (E5384603); 5454001-222, 00000322645HM3 (NO1276CT01); 5454001-222, 00000345652HM2 (K001CT02); 5454001-22, 00000048619YC1 (83090635243); 5454001-22, 00000051425YC7 (703737OP660A); 5454001-222, 00000336132HM6 (X4209906); 5454001-222, 00000361688HM5 (M2848445); 5454001-222, 00000321957HM3 (M4487525); 5454001-22, 00000053126YC9 (AE1456CT01); 5454001-122, 00000437084CN7 (909464CT); 5454001-122, 00000437228CN0 (RADNETCT467); 5454001-22, 59666YC8 (303341CT); 5454001-122, 00000436778CN5 (719383CT); 5454001-122, 00000438070CN5 (303NWMCT1); 5454001-22, 00000059362YC4 (941362660CT); 5454001-122, 00000437195CN1 (333NF3); 5454001-122, 00000437328CN8 (ODC660CT1); 5454001-122, 00000435732CN3 (561798CT64); 5454001-122, 436423CN8 (850CT660); 5454001-22, 00000058085YC2 (850696WCT2); 5454001-222, 00000353734HM7 (478274CT660); 5454001-222, 00000358065HM1 (912576CT); 5454001-122, 00000433880CN2 (478289ECCT); 5454001-122, 00000434344CN8 (808591CT); 5454001-122, 00000433349CN8 (641843CT); 5454001-122, 00000435761CN2 (641394CT66); 5454001-122, 00000435598CN8 (563568CT66); 5454001-222, 00000348978HM8 (WVMC660CT); 5454001-22, 00000054303YC3 (847842CT66); 5454001-122, 436502CN9 (815547OPT1); 5454001-122, 00000430692CN4 (773292CT660); 5454001-22, 00000050821YC8 (708687CT660); 5454001-122, 00000431184CN1 (309655CT3); 5454001-222, 00000357099HM1 (765655PCCT); 5454001-222, 00000353921HM0 (812238UNCT2); 5454001-122, 00000434750CN6 (620365ACCT); 5454001-122, 00000436345CN3 (508548FAL1); 5454001-122, 00000436685CN2 (508548FAL2); 5454001-122, 00000436099CN6 (508CAPECTER); 5454001-122, 00000436425CN3 (508CAPECT1); 5454001-122, 00000436730CN6 (508CAPECT128); 5454001-122, 00000436421CN2 (781329DMCT); 5454001-22, 00000054118YC5 (410968CT660); 5454001-122, 00000430799CN7 (507794CT); 5454001-122, 00000436424CN6 (WC231CT); 5454001-222, 00000349565HM2 (417820CT6); 5454001-122, 00000435208CN4 (828369CT); 5454001-122, 00000431704CN6 (336760GECT); 5454001-122, 00000433884CN4 (505332CT); 5454001-122, 00000434693CN8 (716487CT660); 5454001-122, 00000432049CN5 (330834OPT); 5454001-122, 00000436666CN2 (580497O660); 5454001-122, 00000430872CN2 (918825CT660); 5454001-122, 00000436108CN5 (814274CT); 5454001-22, 00000054119YC3 (412373OPT660); 5454001-122, 00000434902CN3 (803313OPT); 5454001-122, 00000434792CN8 (423844CT660); 5454001-122, 00000438310CN5 (615758MJCT); 5454001-222, 00000381062HM9 (972747OP66); 5454001-122, 00000436183CN8 (817BURED660); 5454001-122, 00000434264CN8 (512ARACPCT2); 5454001-222, 00000361685HM1 (915521CT2); 5454001-122, 00000436940CN1 (817FPFW660); 5454001-222, 00000353925HM1 (512942GTCT); 5454001-222, 00000353924HM4 (972969OPT); 5454001-122, 00000437031CN8 (817255CT660); 5454001-122, 00000438203CN2 (703698OP660); 5454001-122, 00000437280CN1 (434447CT660); 5454001-122, 00000436836CN1 (804HANCT660); 5454001-122, 00000430669CN2 (360482CT660); 5454001-122, 00000430435CN8 (60877660CT); 5454001-222, 00000353970HM7 (608847CT); 5454001-22, 00000058084YC5 (307783CT); 5454001-122, 00000434320CN8 (510797CT4); 5454001-122, 00000433152CN6 (719STM660); 5454001-122, 00000431785CN5 (860676CT660); 5454001-122, 00000436984CN9 (706858CT); 5454001-122, 00000434322CN4 (847990EROPT); 5454001-122, 00000435434CN6 (618288CT1); 5454001-222, 377438HM7 (260665CT660); 5454001-122, 00000436343CN8 (318240CT660); 5454001-122, 00000436966CN6 (517543660CT); 5454001-122, 00000431996CN8 (MH693CT); 5454001-222, 00000357101HM5 (WH231CT); 5454001-122, 00000435161CN5 (816455NKCNLCT); 5454001-122, 00000435570CN7 (828213SJCT); 5454001-122, 00000432108CN9 (SJRCT66064); 5454001-122, 00000434147CN5 (212989CT); 5454001-222, 00000350426HM3 (914654CT660); 5454001-122, 00000434346CN3 (440960CT660); 5454001-122, 00000434984CN1 (440816ER660); 5454001-122, 00000433858CN8 (405717CT); 5454001-122, 00000435762CN0 (METHPEARCT); 5454001-122, 00000437467CN4 (801299CT3); 5454001-122, 00000437061CN5 (435734CT3); 5454001-122, 431925CN7 (703737OP660); 5454001-122, 00000436379CN2 (304329OPT660); 5454001-200, 00000278513HM7 (NO1157CT01); 5454001-222, 00000315026HM5 (852016CT01); 5454001, 00000033972YC1 (M4173326); 5454001-200, 00000274512HM3 (M4173213); 5454001-220, 00000301350HM5 (X24481301); 5454001-222, 00000306528HM1 (M2865815); 5454001, 00000037034YC6 (83021601121212); 5454001, 00000026832YC6 (OA0008); 5454001, 00000026663YC5 (OA0003); 5454001, 00000026831YC8 (OA0010); 5454001, 00000027875YC4 (OA0014); 5454001, 00000027869YC7 (OA0015); 5454001, 00000036418YC2 (OA0020); 5454001, 00000026830YC0 (OA0009); 5454001, 00000027679YC0 (OA0017); 5454001, 00000024839YC3 (OA0001); 5454001, 00000026835YC9 (OA0013); 5454001, 00000027385YC4 (OA0012); 5454001, 00000026834YC2 (OA0011); 5454001, 00000027386YC2 (OA0022); 5454001, 00000026199YC0 (OA0007); 5454001, 00000027876YC2 (OA0027); 5454001, 00000027870YC5 (OA0016); 5454001, 00000029792YC9 (OB0004); 5454001, 00000027878YC8 (OB0002); 5454001, 00000027877YC0 (OB0003); 5454001, 00000026843YC3 (OB0001); 5454001, 00000026200YC6 (OC0003); 5454001, 00000026307YC9 (OC0002); 5454001, 00000027380YC5 (OC0007); 5454001, 00000026833YC4 (OC0005); 5454001, 00000025759YC2 (OC0001); 5454001, 00000027867YC1 (OC0006); 5454001, 00000026665YC0 (OC0004); 5454001, 00000025116YC5 (OD0001); 5454001, 00000038534YC4 (OD0002); 5454001-22, 00000041824YC4 (910214078); 5454001-22, 00000040619YC9 (920219056); 5454001-22, 00000040811YC2 (910214077); |
| FEI Number |
3002808529
|
Recalling Firm/ Manufacturer |
GE Healthcare Japan Corporation Ge Herusukea.Japan ( Kabu -4 Chome 7-127 Asahigaoka Hino Japan
|
| For Additional Information Contact | 800-437-1171 |
Manufacturer Reason for Recall | GE Healthcare has identified that some CT systems may have a damaged cable that can expose 120VAC to
Service Engineers working inside the gantry. |
FDA Determined Cause 2 | Employee error |
| Action | The firm, GE Healthcare, sent an " URGENT MEDICAL DEVICE CORRECTION" letter dated 4/27/2018 to its Customers on about 04/27/2018. The letter describes the product, problem and actions to be taken. The customers were Instructed to: download and provide the updated Service Manual to the organization that services the equipment on site and informed customers that GE Healthcare would be correcting all affected devices at no cost.
Questions or concerns can be directed to GE Healthcare Service at 1-800-437-1171 or the local Sales Representative.
GE Healthcare expanded the recall to include an additional product and model number. Customers were notified on about 08/09/2018 via URGENT MEDICAL DEVICE CORRECTION letter. Instructions were the same as the original notification. |
| Quantity in Commerce | 4,007 devices total (products 1, 2, 3) |
| Distribution | Worldwide Distribution: US (Nationwide) distribution to states of: AK, AL, CA, CO, CT, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, RI, SC, SD, TN, TX, UT, WI, WV, WY; and countries of:: ALGERIA, ARGENTINA, ARMENIA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BELARUS, BELGIUM, BOLIVIA, BOSNIA AND HERZEGOVINA, BRAZIL, BULGARIA, CAMEROON, Canada, CHILE, CHINA, COLOMBIA, Cote D'Ivoire, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, ECUADOR, EGYPT, ETHIOPIA, FINLAND, FRANCE, French Guiana, GERMANY, GHANA, GREECE, GUADELOUPE, GUATEMALA, HONDURAS, Hong Kong, HUNGARY, INDIA, INDONESIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KAZAKHSTAN, Korea (Republic of), KUWAIT, LATVIA, LEBANON, LIBYA, LITHUANIA, MALAYSIA, Mali, MALTA, MARTINIQUE, MEXICO, MOROCCO, NETHERLANDS, NEW , ZEALAND, NIGERIA, NORWAY, OMAN, PAKISTAN, Palestine, PANAMA, PERU, PHILIPPINES, POLAND, PORTUGAL, PUERTO RICO, REUNION, ROMANIA, Russia, RWANDA, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, TANZANIA, THAILAND, TUNISIA, TURKEY, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY, UZBEKISTAN, VENEZUELA, VIETNAM, and YEMEN. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = JAK
|
|
|
|