| | Class 2 Device Recall Endoscopic Cleaning Brush |  |
| Date Initiated by Firm | June 15, 2018 |
| Date Posted | October 11, 2018 |
| Recall Status1 |
Terminated 3 on May 18, 2020 |
| Recall Number | Z-3210-2018 |
| Recall Event ID |
80758 |
| Product Classification |
Accessories, cleaning brushes, for endoscope - Product Code MNL
|
| Product | Endoscopic Cleaning Brush, single use, sold in kits and labeled as the following:
a. Item Number DYK1002DCB, BRUSH, CLEANING, DOUBLE ENDED, W/ HANDLE - 300 EA/CS, 50EA/BX, 6BX/CS;
b. Item Number DYK1002DBDE, BRUSH, CLEANING, DBL, DIF ENDS - 500 EA/CS, 50 EA/BX, 10 BX/CS;
c. Item Number DYK1002DBSE, BRUSH, CLEANING, DBL, SAME END - 500 EA/CS, 50 EA/BX, 10 BX/CS;
d. Item Number DYK1002SBC, BRUSH, CLEANING, SINGLE END - 500 EA/CS, 50 EA/BX, 10 BX/CS |
| Code Information |
Endoscopic cleaning brush finished good lot numbers: GM10, GJ25, HA05, HA20. Kits: a. Item Number DYK1002DCB, kit Number (Kit Lot Number): DYK1001DCB (18BA0735), DYK1002144B (18BA0644, 18DA0124), DYK1008972E (18BA2199, 18DA1418, 18EAA023, 18EAA653), DYK1012268E1 (18BA1667, 18CA0872, 18CA1289, 18DA1732), DYK1012414E1 (18CA0793, 18CA1644, 18DA0413), DYK1017922E (18CA1300), DYK1018542E (18CA0622, 18DA1159, 18EAB507), DYK1044978E (18EBE085, 18NB1289), DYK1067817E1 (18AA1969, 18EAA004, 18EAB365), DYK1072097E (18DA0437, 18EAA790), DYK1097615E (18CA0687, 18EAA084), DYK1177017E (18BA2352, 18DA1481), DYK1188722E (18EAA016, 18EAB529), DYK1253578E (18AA2092, 18DA0331), DYK1253578F (18DA1595), DYK1266757E (18BA0203, 18DA0062, 18DF0085), DYKE1003A (18DA1488, 18FAA025), DYKE1011B (18AA1276), DYKE1012C (18DA0090, 18EAA521), DYKE1013 (18CB4388, 18EBL745), DYKE1014B (17KA1195, 18AA0984, 18CA0691, 18CA1969, 18DA0344, 18DA1167), DYKE1022A (18EBE830), DYKE1029C (18BA0279, 18CA0842), DYKE1032C (18BB1877, 18BB6935, 18EBH874, 18EBS890, 18OB9988), DYKE1033B (18MB9361), DYKE1033C (18CB1145, 18DB0018, 18EBH875), DYKE1037 (18CA1072, 18DA0377), DYKE1043C (18DA0411, 18EAA198), DYKE1048 (18BA0550), DYKE1050 (18BA0830, 18CA1744, 18EAA936), DYKE1056C (18DA0952, 18EBF206, 18EBF513, 18EBN896, 18PB3439), DYKE1058A (18BB6939, 18CB5355, 18PB4193), DYKE1059A (18BB582118EBF514), DYKE1062A (18CA1504), DYKE1062B (18EAB341), DYKE1063 (18BB4018, 18EBL836, 18OB2537, 18PB6224), DYKE1064 (18BB2161, 18DB8246, 18EBL833), DYKE1069A (18CA0564, 18CA1222), DYKE1069B (18DA0510), DYKE1072 (18BA0524, 18DA0412, 18EAA070), DYKE1077 (18BA1594, 18DA0096, 18EAB672), DYKE1096A (18BA1239, 18DA0345, 18FAA190), DYKE1099 (18BA1707), DYKE1108C (17JA2124, 17KA1225), DYKE1108D (18AA2499, 18BA1296, 18CA0336), DYKE1112 (18CA0139), DYKE1120 (18BA2330, 18DA0523, 18EAB011), DYKE1123 (18CA0181, 18DA0779, 18EAB549), DYKE1126A (18BA0589, 18CA1597), DYKE1138A (18BA1559, 18CA2056), DYKE1143A (18BA0473), DYKE1152 (18DA1187, 18EAA714), DYKE1158A (18BA2208, 18EAA253, 18EAB695, 18FAA253), DYKE1159A (18FAA227), DYKE1161A (18BA1435, 18EAB349), DYKE1162G (17JA2795), DYKE1162I (18AA1616, 18AA2211, 18BA0354), DYKE1162J (18BA1753, 18DA1469), DYKE1164 (18EBD023, 18OB2534), DYKE1172C (18EAA129), DYKE1176 (18CA1092, 18DA1719), DYKE1184B (17IA1038, 17JA1424, 17LA1730, 18BA1937, 18DA1028), DYKE1185 (18CA1513), DYKE1203A (18DA0680, 18FAA307), DYKE1207 (18BA1174, 18CA0313, 18EAA076, 18EAA397), DYKE1224C (18DA0018, 18EAA709), DYKE1232 (18CA0337, 18EAA730), DYKE1233 (18AA1392, 18BA0937, 18DA0390, 18DB7237, 18EBG997, 18EBT332), DYKE1234 (18FAA404), DYKE1245 (18BA0307, 18BA0908), DYKE1249 (18BA1440, 18DA1612, 18FAA258), DYKE1250 (18BA2009, 18EAA141), DYKE1258 (18BA1480, 18EAA325), DYKE1265 (18DA0068, 18DA0206, 18EAB322, 18EAB747), DYKE1274A (18CA0987, 18EAB342), DYKE1293 (18BA0492, 18CA1224, 18DA1377, 18EAB042), DYKE1295 (18BA0770, 18CA1963, 18DA1479), DYKE1303 (18MB9887, 18OB2532), DYKE1304 (18BB5870, 18DB7236), DYKE1308 (18CA0292, 18DA0841), DYKE1313B (18CA1559, 18EAB036), DYKE1314A (18CA1327), DYKE1318 (18BA0888, 18DA1352, 18FAA368), DYKE1332A (18EAA061), DYKE1333 (18BA0763, 18EAA432), DYKE1334B (18BA0597, 18EAA447), DYKE1345 (18CA1207, 18DA1486), DYKE1350A (18BB2129, 18DB7371, 18EBO982), DYKE1357B (18CA0267), DYKE1357C (18DA1208, 18FAA100), DYKE1364A (17LA1129, 18BA0029, 18DB7239, 18EBD050), DYKE1366 (18BA0796, 18CA1707, 18EAB410), DYKE1369C (17JA2209, 17LA2210, 18BA0882, 18CA1422), DYKE1372 (18NB4423), DYKE1375A (18BA1778, 18CA2032), DYKE1385 (18BA0562, 18CA1193, 18EAA304), DYKE1386 (18CA0393, 18DA0589, 18FAA222), DYKE1391 (18EAB593), DYKE1393A (18BA0591, 18CA1406, 18DA0833, 18EAA789), DYKE1395A (18CA0886, 18DA0717, 18FAA293), DYKE1399 (18DA0204, 18EAA422, 18FAA144), DYKE1402 (18BB5817, 18DB7373, 18EBS803), DYKE1406 (18CA0328, 18DA0091), DYKE1409 (18DA0216, 18EAB769), DYKE1410 (18BA0708, 18CA0837, 18DA1564, 18EAA453), DYKE1415 (18CA1723, 18DA1887, 18EAB265), DYKE1416 (18DA1072, 18EAA886), DYKE1422A (18BA1978, 18DA1468), DYKE1423A (18CA1662, 18DA1174, 18EAA686), DYKE1431B (18BA0503, 18DA1586, 18EAA301, 18EAB901), DYKE1438A (18AA1176), DYKE1438B (18CA0630, 18EAB038), DYKE1441B (18DA1320, 18DA1745, 18EAB829), DYKE1442A (18DB7372, 18NB2540), DYKE1448 (18BA1463, 18CA1544), DYKE1455 (18CA0109, 18EAA273, 18EAB548), DYKE1456 (18BA0145, 18CA0268, 18DA0642, 18EAA706), DYKE1457 (18BA0626, 18EAA078), DYKE1460 (18AA2138, 18CA1928, 18EAB528), DYKE1466 (18BA1634, 18CA1187, 18DA1776), DYKE1468 (18BA1335, 18EAB838), DYKE1469A (18BA0868, 18DA1474), DYKE1475A (17LA0444, 18AA2019, 18BA1162, 18CA1545), DYKE1480 (18BA0028, 18BA0211, 18BB6704, 18EBQ945, 18OB9926, 18PB5260), DYKE1481 (18BA1471, 18CA1402, 18DA1569, 18EAB051), DYKE1487A (18AA2145, 18BA1736), DYKE1487B (18DA0599), DYKE1488 (18DB7238, 18EBE095), DYKE1491 (18DA0948), DYKE1494 (18BA1450, 18DA1778), DYKE1498 (18CA0824), DYKE1503B (17LA0480, 18AA0783, 18BA0551, 18CA0587, 18DA0778, 18EAA371, 18FAA192), DYKE1506 (18FAA157), DYKE1507 (18DA0591, 18EAB736, 18FAA039), DYKE1516A (18AA1029, 18BA0055, 18BA1034, 18EAB049), DYKE1519 (17IA0093, 18BA1540, 18CA1650, 18EAA320), DYKE1521 (17IA2305, 17JA2046), DYKE1521A (17LA1124, 17LA2004, 18AA2110, 18DA0535), DYKE1526 (17JA0857, 17KA1032, 18AA2090), DYKE1526A (18BA1336, 18CA1705, 18DA1783, 18EAA523, 18FAA247), DYKE1527 (17IA1501, 17KA1034, 18AA0950, 18CA1957, 18DA1731, 18EAB046), DYKE1535 (17JA1057, 17LA0481, 18AA2114, 18DA1728), DYKE1536 (17JA0776, 17KA1944, 18AA1963, 18BA1970, 18DA0211, 18EAA468), DYKE1537 (17JA1066), DYKE1538 (17LA0810, 18BA0475, 18BA2349, 18DA1031, 18EAB415), DYKE1543 (17KA0386, 17LA1373, 18AA2292), DYKE1554 (18AA0653, 18CA1845), DYKE1555 (17LA1505), DYKE1559 (18AA1924, 18BA0444), DYKE1561 (18BA1796, 18DA0372, 18EAB826), DYKE1564 (18CA1183, 18EAA246), DYKE1567 (18CA0739), DYKE1573 (18EAA049, 18FAA288), DYKE1574 (18DA1130, 18DA1594), DYKE1579 (18EAB289, 18EAB583), DYKE1586 (18EAB222), DYKE1588 (18FAA342), DYKE18CMPLY (18CA1467), DYKENDOBASIC (18CA0995, 18DA1811, 18FAA369), DYKENDOBPLUS (18BA1853, 18DA1190, 18FAA207), DYKENDOCLEAN (18DA0382, 18EAB587), DYKENDOCMPLY (18BA1056, 18CA0124, 18DA0129, 18EAB667), DYKENDODBCMP (18BA0916, 18DA0383, 18EAA260, 18EAB752), DYKENDOLWCMP (18DA1812, 18FAA202), DYKENDOUPCMP (18BA1854, 18DA0718, 18EAA508), DYKM1404 (18CA1636, 18DA0654), DYKM1555 (17JA1068, 17LA1482, 18AA1177, 18AA2029, 18AA2612, 18CA0565, 18DA0069, 18EAA433), DYSAMP200458 (18DA1963), DYSAMP200460 (18FAA265), P880995 (18EKD315), P881009 (18EAA498), P971311 (18BA0897, 18CA1642, 18DA0859, 18EAA859), PHS347468000 (18DK2120, 18FKB069); b. Item Number DYK1002DBDE, kit Number (Kit Lot Number): DYK1001DBDE (17IA0598, 17LA0129), DYK1002DBDE (18BA0041), DYK1276244E1 (17JA0713), DYKE1122C (17JA0877, 17JA1514, 17JA2093, 17LA0460, 17LA1984, 18BA2299, 18CA0795, 18CA1927, 18EAA705), DYKE1446A (17JA2156, 17LA1648, 18BA0062, 18CA1226, 18DA1726), DYKE1447A (17IA1474, 17LA1178, 18BA0942), DYKE1516 (17HA1857, 17JA2263), DYKE1539 (17JA1281, 17KA1644, 18AA2157), DYKE1540 (17KA1547), DYKE1540A (18BA0740, 18CA1706, 18DA1020, 18EAB360), DYKE1552 (17LA1291, 18CA0448, 18CA0843), DYKE1580 (18EAA623, 18FAA191); c. Item Number DYK1002DBSE, kit Number (Kit Lot Number): DYK1091457E1 (17LA2255, 18CA0974, 18EAA258), DYK1245641E (17LA1486, 18CA1307), DYKE1000A (17LA0061, 17LA1462, 18CA1093, 18EAA814), DYKE1071 (17LA1174, 18AA0941, 18AA1383), DYKE1110 (18AA1490, 18CA0971, 18DA1557), DYKE1131 (18CA0972, 18EAA321), DYKE1135 (18AA0928, 18BA2275), DYKE1136B (18AA0807, 18AA2529, 18CA0970), DYKE1142 (18AA1125, 18DA0283, 18EAA075), DYKE1144 (18AA0905, 18BA0411, 18CA1924), DYKE1156 (18AA1587, 18CA0976), DYKE1157 (18BA1683, 18CA0975, 18DA1464), DYKE1193 (18AA2159, 18DA0924), DYKE1200B (18AA0869, 18BA2046, 18DA1022), DYKE1201C (17LA2024), DYKE1222B (18AA1458, 18DA0221), DYKE1307A (17JA0617, 17LA1157, 18AA0914, 18CA0967, 18CA1964, 18DA1473), DYKE1392 (17LA1416, 18AA0963, 18DA0376), DYKE1424A (18AA2437, 18CA0985, 18DA1013), DYKE1462A (18BA1818, 18DA1309, 18EAA071), DYKE1505A (17LA2183, 18DA1017), DYKE1533 (17JA0051, 18CA1355), DYKE1581 (18EAA459); d. Item Number DYK1002SBC, kit Number (Kit Lot Number): DYKE1098A (17KA0949, 17LA0422, 18BA0284, 18CA1966, 18DA1296, 18EAA950), DYKE1105 (17JA2744, 17KA1969, 18BB5818, 18EBU560), DYKE1294 (18AA0754), DYKE1363 (17LA0251), DYKE1493 (17KA2116), DYKE1560 (18BA1585, 18DA0003) |
| FEI Number |
1417592
|
Recalling Firm/ Manufacturer |
Medline Industries Inc
Three Lakes Drive
Northfield
IL
60093
|
| For Additional Information Contact | Kassandra Cotner 866-359-1704 |
Manufacturer Reason for Recall | There is potential risk of brush detachment during cleaning. If brush detachment is not identified, there is the potential for the brush to be left within the scope. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | Medline notified customers on about 06/15/2018 via URGENT RECALL MEDLINE INDUSTRIES, INC. IMMEDIATE ACTION REQUIRED letter. Accounts that purchased a custom kit or standard Pack were instructed to immediately check inventory for affected devices, quarantine any affected devices, affix the provided stickers on affected kits and continue use but were instructed to not use the Endoscopic Cleaning Brush in the kit and to instead pull another approved product from inventory. They were also instructed to complete and return the Verification Form.
Accounts that just purchased the Endoscopic Cleaning Brush were instructed to immediately check inventory for affected devices, destroy any affected devices, notify customers if the product has been further distributed, and complete and return the provided destruction form listing the quantity of affected products destroyed or that there was no product on hand to destroy. |
| Quantity in Commerce | 183,606 devices total |
| Distribution | Nationwide distribution to AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NH, NJ, NM,
NV, NY, OH, OK, PA, RI, SC, TN, TX, VA, VT, WA, WI, WV, WY |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
|
|
|