| Date Initiated by Firm | June 29, 2018 |
| Create Date | October 18, 2018 |
| Recall Status1 |
Terminated 3 on August 12, 2021 |
| Recall Number | Z-0194-2019 |
| Recall Event ID |
80774 |
| 510(K)Number | K151767 |
| Product Classification |
Analyzer, chemistry (photometric, discrete), for clinical use - Product Code JJE
|
| Product | Atellica CH 930 Analyzer (SMN 11067000) .
The Atellica" Solution is a multi-component system for in vitro diagnostic testing of clinical specimens. |
| Code Information |
Software (SW) version 1.13 and lower |
| FEI Number |
2432235
|
Recalling Firm/ Manufacturer |
Siemens Healthcare Diagnostics, Inc. 511 Benedict Ave Tarrytown NY 10591-5005
|
| For Additional Information Contact | Same 914-631-8000 |
Manufacturer Reason for Recall | Atellica Solution system software (SW) version 1.13 and lower, five (5) behavioral issues were identified:
1. Reagent Lock may not work as intended with Primary Reagent ReadyPacks and Ancillary Packs when more than one (1) lot of specific Reagent kits are loaded on the analyzer at the same time.
2. Sample barcodes may be incorrectly read
3. The Atellica IM 1300 Analyzer and Atellica IM 1600 Analyzer may not run CEA dilutions at 50x and 100x. The Analyzer may post a Sample Integrity Error when 50x or 100x dilutions are ordered for Carcinoembryonic Antigen (CEA). Dilution factors lower than 50x and neat samples are not affected by this behavior.
4. The Atellica CH 930 Analyzer may continue to display the default reporting units after the reporting units have been changed.
5. The Atellica CH 930 Analyzer may not display results for sample if certain assays are configured to auto-dilute results above or below the assay measuring interval.
|
FDA Determined Cause 2 | Software design |
| Action | Siemens issued an Urgent medical Device Correction, (UMDC ASW18-02.A.US) and Urgent Field Safety Notification (UFSN ASW18-02.A.OUS) on 6/29/18, via FedEx to the US Customer and via email distribution to countries Outside US (OUS).
The field action letters explain the behaviors and informed Customers of action to be taken to prevent the behaviors from occurring. Siemens has released an updated software (SW) version (V), 1.14, which corrects the reported behaviors. |
| Quantity in Commerce | 115 |
| Distribution | Nationwide
Foreign:
Australia
Austria
Belgium
Brazil
Canada
Croatia
Czech Republic
Denmark
France
Germany
Greece
Italy
Japan
Malaysia
Mexico
Netherlands
Norway
Poland
Republic Korea
Singapore
Spain
Sweden
Switzerland
U.A.E.
United Kingdom |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = JJE
|