 Class 1 Device Recall CONFIRM antiProgesterone Receptor (PR) (1E2) Rabbit Monoclonal Primary Antibodysee related information
Date Initiated by FirmAugust 02, 2018
Date PostedSeptember 14, 2018
Recall Status1 Terminated 3 on November 22, 2021
Recall NumberZ-3006-2018
Recall Event ID 80807
510(K)NumberK103818 
Product Classification Immunohistochemistry assay, antibody, progesterone receptor - Product Code MXZ
ProductCONFIRM anti-Progesterone Receptor (PR) (1E2) Rabbit Monoclonal Primary Antibody, Catalog Number 05278392001, model 790-4296 Immunohistochemistry (IHC) for in vitro diagnostic use.
Code Information Lot Numbers: Y12992, Y18852, Y23051, UDI: 4015630972203
FEI Number 1000125183
Recalling Firm/
Manufacturer
Ventana Medical Systems Inc
1910 E Innovation Park Dr
Oro Valley AZ 85755-1962
For Additional Information Contact
800-227-2155
Manufacturer Reason
for Recall
Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers. A false negative IHC result may occur.
FDA Determined
Cause 2
Under Investigation by firm
ActionThe firm disseminated their field safety notices by letter on 08/02/2018. The firm requested the destruction of the product.
Quantity in Commerce1655 units
Distributionworldwide
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = MXZ
-
-