|
Class 1 Device Recall ENDOLOGIX AFX2 Bifurcated Endograft System |
|
Date Initiated by Firm |
July 31, 2018 |
Date Posted |
October 03, 2018 |
Recall Status1 |
Terminated 3 on April 09, 2024 |
Recall Number |
Z-0007-2019 |
Recall Event ID |
80812 |
PMA Number |
P040002S060 P040002S061 |
Product Classification |
System, endovascular graft, aortic aneurysm treatment - Product Code MIH
|
Product |
ENDOLOGIX AFX2 Bifurcated Endograft System, Stent Graft System, Rx Only, Sterile EO, REF: BEA22-120/I20-40, BEA22-100/I20-40, BEA22-80/I20-40, BEA22-60/I20-40, BEA22-40/I20-40, BEA22-120/I16-40, BEA22-100/I16-40, BEA22-80/I16-40, BEA22-60/I16-40, BEA22-40/I16-40, BEA22-120/I13-40, BEA22-100/I13-40, BEA22-80/I13-40, BEA22-60/I13-40, BEA22-40/I13-40, BEA22-110/I20-30, BEA22-90/I20-30, BEA22-70/I20-30, BEA22-50/I20-30, BEA22-110/I16-30, BEA22-90/I16-30, BEA22-70/I16-30, BEA22-50/I16-30, BEA22-100/I20-55, BEA22-80/I20-55, BEA22-100/I16-55, BEA22-80/I16-55, BEA25-120/I20-40, BEA25-100/I20-40, BEA25-80/I20-40, BEA25-60/I20-40, BEA25-40/I20-40, BEA25-120/I16-40, BEA25-100/I16-40, BEA25-80/I16-40, BEA25-60/I16-40, BEA25-40/I16-40, BEA25-120/I13-40, BEA25-100/I13-40, BEA25-80/I13-40, BEA25-60/I13-40, BEA25-40/I13-40, BEA25-110/I20-30, BEA25-90/I20-30, BEA25-70/I20-30, BEA25-50/I20-30, BEA25-110/I16-30, BEA25-90/I16-30, BEA25-70/I16-30, BEA25-50/I16-30, BEA25-100/I20-55, BEA25-80/I20-55, BEA25-100/I16-55, BEA25-80/I16-55, BEA28-120/I20-40, BEA28-100/I20-40, BEA28-80/I20-40, BEA28-60/I20-40, BEA28-40/I20-40, BEA28-120/I16-40, BEA28-100/I16-40, BEA28-80/I16-40, BEA28-60/I16-40, BEA28-40/I16-40, BEA28-120/I13-40, BEA28-100/I13-40, BEA28-80/I13-40, BEA28-60/I13-40, BEA28-40/I13-40, BEA28-110/I20-30, BEA28-90/I20-30, BEA28-70/I20-30, BEA28-50/I20-30, BEA28-110/I16-30, BEA28-90/I16-30, BEA28-70/I16-30, BEA28-50/I16-30, BEA28-100/I20-55, BEA28-80/I20-55, BEA28-100/I16-55, BEA28-80/I16-55
|
Code Information |
All lots distributed from 03/2011 to present. |
Recalling Firm/ Manufacturer |
Endologix 35 Hammond Irvine CA 92618-1607
|
Manufacturer Reason for Recall |
Endologix is notifying physicians of patient tailored surveillance recommendations as well as general warnings/precautions for interventions to or through an existing AFX device. This is a follow-up notification to the one that was sent out in December 2016. The updates are related to the observation of Type III endoleaks.
|
FDA Determined Cause 2 |
Device Design |
Action |
On 07/31/2018, the firm mailed Urgent Important Safety Updates via FedEx ground service with return delivery confirmation. Physicians were informed about the following: 1) Type III endoleak rates, 2) Refined patient-tailored surveillance recommendations, 3) Sizing recommendations, 4) Recommendations for device interventions/ re-interventions. No product return is required. Customers with questions are encouraged to call Customer service 800-983-2284 (5:00 A.M. - 6:00 P.M. Pacific Time).
Updated 03/01/2023:
Because of the ongoing concerns regarding Type III endoleaks with Endologix AFX Endovascular AAA Graft Systems, the FDA has publicly communicated concerns:
https://public4.pagefreezer.com/browse/FDA/21-08-2022T08:51/https:/www.fda.gov/medical-devices/letters-health-care-providers/update-type-iii-endoleaks-associated-endovascular-graft-systems-letter-health-care-providers
Update on Risk of Type III Endoleaks with Use of Endologix AFX Endovascular AAA Graft Systems: FDA Safety Communication | FDA (pagefreezer.com)
UPDATE: The FDA Reminds Patients and Health Care Providers of the Importance of At Least Yearly, Lifelong Follow-Up with Use of Endologix AFX Endovascular AAA Graft Systems: FDA Safety Communication | FDA (pagefreezer.com)
On November 2, 2021, the FDA convened a public meeting of the CDRH Circulatory System Devices Panel of the Medical Devices Advisory Committee to share information and perspectives from interested parties on the benefit-risk profile of the Endologix AFX endovascular graft system focused on the risk of Type III endoleaks.
|
Quantity in Commerce |
13178 |
Distribution |
U.S.:
PA, VA, CA, AZ, IN, GA, FL, IL, OK, MN, OH, KY, MS, NC, MI, AL, CT, LA, SC, TN, NY, WI, MD, WV, AR, NV, NJ, TX, CO, NE, NM, MA, IA, AK, OR, ME, SD, MO, MT, KS, DE, WA, UT, NH, ND, DC, RI, HI, WY, ID, VT, PR;
Foreign (OUS):
Canada, Argentina, Australia, Brazil, CHILE , Hong Kong, Japan, South Korea, Philippines, Austria, Belgium, Czech Republic, France, Germany, Greece, Ireland, Israel, Italy, Latvia, Luxembourg, Monaco, Netherlands, Poland, Portugal, Romania, Spain, Sweden, Switzerland, Turkey, United Kingdom |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
PMA Database |
PMAs with Product Code = MIH and Original Applicant = Endologix, LLC
|
|
|
|