| Date Initiated by Firm | September 04, 2018 |
| Date Posted | September 14, 2018 |
| Recall Status1 |
Terminated 3 on August 17, 2021 |
| Recall Number | Z-3175-2018 |
| Recall Event ID |
80979 |
| Product Classification |
Microscope, surgical - Product Code EPT
|
| Product | Leica M525 F20 Surgical Microscope System |
| Code Information |
Serial Numbers: 10714001 10813001 11010001 11010002 20312001 20316001 20513001 20518001 20813001 20911002 20916001 30112001 30412001 30616001 30813001 30813002 40112001 40310001 40510001 40610001 40916001 41215001 50218001 50813001 51011001 51015001 51116001 60218001 60317001 60415001 60813001 70116002 70414001 70714001 71010001 71011001 80114001 80212002 80316001 80513001 80611001 80817001 81014001 90317001 90610001 91117001 91215001 100310001 100317001 100616001 101211001 101216001 110110001 110112001 110116001 110310001 110517001 110809001 110816001 120116001 121216001 130313001 130813001 131010002 131011001 131115001 131211001 131212001 131216001 131216002 140214001 140413001 140612001 140813001 141209001 150415001 150612001 150813001 151013001 151116002 151210001 151211001 151211002 151215001 160216001 160313001 160418001 160810001 160911001 161210002 170310001 170809001 170811001 170916001 170917001 180116001 180117001 180213001 180216001 180511001 180610001 180612001 181111001 190214002 190309001 190316001 190813001 200418001 200910001 210115001 210214001 210316001 211010001 220715001 220909001 220916001 221010002 230909001 230910001 230916001 231109001 231116001 231213001 240712001 240713001 250218001 250414001 250510001 250516001 250814001 250816001 250914001 260514001 260815002 260914001 270810001 280118001 280217001 280411001 280414001 280513001 280910002 280911001 281111001 281113001 290116001 290212001 290515001 290713001 290817001 290817002 290912001 300414001 300418001 300516001 300713001 300816001 310514001 310810001 310815001 The following serial numbers were added in October, 2018. 140415001 170415001 170415002 40714001 170212001 70116001 70513001 90513001 |
| FEI Number |
3000210133
|
Recalling Firm/ Manufacturer |
Leica Microsystems, Inc.
1700 Leider Ln
Buffalo Grove
IL
60089-6622
|
| For Additional Information Contact | Jason Battle 847-876-5705 |
Manufacturer Reason for Recall | The firm received complaints that the instrument tipped over when the swing arm was fully extended. |
FDA Determined Cause 2 | Device Design |
| Action | The firm disseminated their Field Corrective Action Notices by mail beginning on 09/04/2018 to the US and Canadian consignees. The letter explained the problem and informed the consignee that an update package with enhanced instructions for use and additional marking will be provided within the next 3 months.
Additional consignees found were notified by email on 10/01/2018. |
| Quantity in Commerce | 160 units |
| Distribution | US and Canada |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|