| Date Initiated by Firm | September 04, 2018 |
| Create Date | October 03, 2018 |
| Recall Status1 |
Terminated 3 on July 10, 2020 |
| Recall Number | Z-0019-2019 |
| Recall Event ID |
80986 |
| Product Classification |
Device for sealing microsections - Product Code KIM
|
| Product | Leica HistoCore SPECTRA CV Coverslipper, used in preparation of histological and cytological tissue samples for microscopic analysis for medical diagnostics. |
| Code Information |
Serial Number 102 |
| FEI Number |
3000210133
|
Recalling Firm/ Manufacturer |
Leica Microsystems, Inc.
1700 Leider Ln
Buffalo Grove
IL
60089-6622
|
| For Additional Information Contact | Adam Blunier 847-821-3462 |
Manufacturer Reason for Recall | Product was assembled with an isolating fiberglass hose within the oven in the wrong position. This results in improper shielding from potential electrical shock. |
FDA Determined Cause 2 | Process design |
| Action | The firm issued a Field Action Notice by letter on 09/05/2018. The notice explained the issue and the possible hazard. The firm will schedule a visit with a service engineer to perform the correction. |
| Quantity in Commerce | 1 unit |
| Distribution | Distribution in US state of MO. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|