Date Initiated by Firm |
September 04, 2018 |
Create Date |
October 03, 2018 |
Recall Status1 |
Terminated 3 on July 10, 2020 |
Recall Number |
Z-0019-2019 |
Recall Event ID |
80986 |
Product Classification |
Device for sealing microsections - Product Code KIM
|
Product |
Leica HistoCore SPECTRA CV Coverslipper, used in preparation of histological and cytological tissue samples for microscopic analysis for medical diagnostics. |
Code Information |
Serial Number 102 |
Recalling Firm/ Manufacturer |
Leica Microsystems, Inc. 1700 Leider Ln Buffalo Grove IL 60089-6622
|
For Additional Information Contact |
Adam Blunier 847-821-3462
|
Manufacturer Reason for Recall |
Product was assembled with an isolating fiberglass hose within the oven in the wrong position. This results in improper shielding from potential electrical shock.
|
FDA Determined Cause 2 |
Process design |
Action |
The firm issued a Field Action Notice by letter on 09/05/2018. The notice explained the issue and the possible hazard. The firm will schedule a visit with a service engineer to perform the correction. |
Quantity in Commerce |
1 unit |
Distribution |
Distribution in US state of MO. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|