• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall BOND Slide Labeler Zebra Printer Model TLP3842 and Model GX4301

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall BOND Slide Labeler Zebra Printer Model TLP3842 and Model GX4301see related information
Date Initiated by FirmSeptember 17, 2018
Create DateOctober 26, 2018
Recall Status1 Terminated 3 on August 21, 2020
Recall NumberZ-0281-2019
Recall Event ID 81036
Product Classification Slide stainer, automated - Product Code KPA
ProductBOND Slide Labeler Zebra Printer Model TLP3842 and Model GX4301, for use with BOND-III and BOND-MAX instruments
Code Information Zebra Printer Model TLP3842 - Affected PSU Part Number 808113-001 or P1028888-001 with date codes ranging from 0639XX through 1253XX (YYWWXX, where YY=year and WW=week);    Zebra Printer Model GX4301 - Affected PSU Part Number 808101-001 with date codes ranging from 0639XX through 1253XX (YYWWXX, where YY=year and WW=week).
FEI Number 3000210133
Recalling Firm/
Manufacturer
Leica Microsystems, Inc.
1700 Leider Ln
Buffalo Grove IL 60089-6622
For Additional Information ContactMr. Adam Blunier
847-821-3462
Manufacturer Reason
for Recall
The power supply unit for the printer used with the device can potentially overheat or cause a fire hazard because it could contain an improperly-mixed phosphorus compound.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionThe recalling firm issued letters dated 9/3/2018 via email on 9/17/2018. A second letter dated 9/24/2018 was issued to update the acknowledgment form to include the number of affected power supply units the consignee destroyed.
Quantity in Commerce876 devices (in U.S. and Canada)
DistributionDistribution was nationwide. There was also government/military distribution. Foreign distribution was made to Canada.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
-
-