| Date Initiated by Firm | September 17, 2018 |
| Create Date | October 26, 2018 |
| Recall Status1 |
Terminated 3 on August 21, 2020 |
| Recall Number | Z-0281-2019 |
| Recall Event ID |
81036 |
| Product Classification |
Slide stainer, automated - Product Code KPA
|
| Product | BOND Slide Labeler Zebra Printer Model TLP3842 and Model GX4301, for use with BOND-III and BOND-MAX instruments |
| Code Information |
Zebra Printer Model TLP3842 - Affected PSU Part Number 808113-001 or P1028888-001 with date codes ranging from 0639XX through 1253XX (YYWWXX, where YY=year and WW=week); Zebra Printer Model GX4301 - Affected PSU Part Number 808101-001 with date codes ranging from 0639XX through 1253XX (YYWWXX, where YY=year and WW=week). |
| FEI Number |
3000210133
|
Recalling Firm/ Manufacturer |
Leica Microsystems, Inc.
1700 Leider Ln
Buffalo Grove
IL
60089-6622
|
| For Additional Information Contact | Mr. Adam Blunier 847-821-3462 |
Manufacturer Reason for Recall | The power supply unit for the printer used with the device can potentially overheat or cause a fire hazard because it could contain an improperly-mixed phosphorus compound. |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | The recalling firm issued letters dated 9/3/2018 via email on 9/17/2018. A second letter dated 9/24/2018 was issued to update the acknowledgment form to include the number of affected power supply units the consignee destroyed. |
| Quantity in Commerce | 876 devices (in U.S. and Canada) |
| Distribution | Distribution was nationwide. There was also government/military distribution. Foreign distribution was made to Canada. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|