| | Class 2 Device Recall LW Scientific ZIP IQ Combo Centrifuge |  |
| Date Initiated by Firm | February 06, 2019 |
| Create Date | March 23, 2019 |
| Recall Status1 |
Terminated 3 on May 07, 2021 |
| Recall Number | Z-1049-2019 |
| Recall Event ID |
82214 |
| Product Classification |
Centrifuges (micro, ultra, refrigerated) for clinical use - Product Code JQC
|
| Product | LW Scientific ZIP IQ Combo Centrifuge, ZIC-CMBO-2RT3 Rev B
Product Usage:
used as a general purpose centrifuge to spin down blood in hematocrit tubes with additional rotor to spin other fluids in test tubes (15ml) |
| Code Information |
Serial Numbers: 6C1712002, 6C1712008, 6C1803021, 6C1803022, 6C1803053, 6C1803053, 6C1803054, 6C1803055, 6C1803057, 6C1803058, 6C1803059, 6C1803060, 6C1803097, 6C1803124, 6C1803128, 6C1803129, 6C1803134, 6C1803135, 6C1803136, 6C1803140, 6C1805146, 6C1805147, 6C1805148, 6C1805149, 6C1805150, 6C1805151, 6C1805152, 6C1805153, 6C1805154, 6C1805155, 6C1805156, 6C1805157, 6C1805158, 6C1805159, 6C1805160, 6C1805161, 6C1805162, 6C1805163, 6C1805164, 6C1805165, 6C1805166, 6C1805167, 6C1805168, 6C1805169, 6C1805170, 6C1805171, 6C1805172, 6C1805173, 6C1805174, 6C1805175, 6C1805176, 6C1805177, 6C1805177, 6C1805178, 6C1805179, 6C1805180, 6C1805181, 6C1805182, 6C1805183, 6C1805184, 6C1805185, 6C1805186, 6C1805187, 6C1805188, 6C1805189, 6C1805190, 6C1805191, 6C1805192, 6C1805193, 6C1805194, 6C1805195, 6C1805196, 6C1805197, 6C1805198, 6C1805199, 6C1805200, 6C1805201, 6C1805202, 6C1805203, 6C1805204, 6C1805205, 6C1805206, 6C1805207, 6C1805208, 6C1805209, 6C1805209, 6C1805210, 6C1805211, 6C1805212, 6C1805213, 6C1805214, 6C1805218, 6C1805219, 6C1805220, 6C1805221, 6C1805222, 6C1805223, 6C1805224, 6C1805225, 6C1805226, 6C1805227, 6C1805228, 6C1805229, 6C1805230, 6C1805231, 6C1805232, 6C1805233, 6C1805234, 6C1805235, 6C1805236, 6C1805237, 6C1805238, 6C1805239, 6C1805240, 6C1805241, 6C1805242, 6C1805243, 6C1805244, 6C1805245, 6C1810346, 6C1810347, 6C1810348, 6C1810349, 6C1810350, 6C1810351, 6C1810352, 6C1810353, 6C1810354, 6C1810355, 6C1810356, 6C1810357, 6C1810358, 6C1810359, 6C1810360, 6C1810361, 6C1810362, 6C1810363, 6C1810364, 6C1810365, 6C1810366, 6C1810370, 6C1810371, 6C1810372, 6C1810373, 6C1810374, 6C1810375, 6C1810376, 6C1810378, 6C1810379, 6C1810380, 6C1810381, 6C1810386, 6C1810387, 6C1810388, 6C1810389, 6C1810390, 6C1810391, 6C1810392, 6C1810393, 6C1810394, 6C1810395, 6C1810396, 6C1810397, 6C1810398, 6C1810399, 6C1810400, 6C1810401, 6C1810404, 6C1810405, 6C1810406, 6C1810407, 6C1810408, 6C1810409, 6C1810410, 6C1810411, 6C1810412, 6C1810413, 6C1810414, 6C1810415, 6C1810416, 6C1810417, 6C1810418, 6C1810419, 6C1810420, 6C1810421, 6C1810422, 6C1810423, 6C1810424, 6C1810425, 6C1810426, 6C1810427, 6C1810428, 6C1810429, 6C1810434, 6C1810435, 6C1810436, 6C1810437, 6C1810438, 6C1810439, 6C1810440, 6C1811448, 6C1811449, 6C1811450, 6C1811451, 6C1811452, 6C1811453, 6C1811454, 6C1811455, 6C1811464, 6C1811465, 6C1811466, 6C1811467, 6C1811504, 6C1811505, 6C1811506, 6C1811507, 6C1811508, 6C1811509, 6C1811510, 6C1811511, 6C1811540, 6C1811541, 6C1811542, 6C1811543, 6C183137, 6C183138 |
| FEI Number |
3000251573
|
Recalling Firm/ Manufacturer |
LW Scientific, Inc.
865 Marathon Pkwy
Lawrenceville
GA
30046-2857
|
| For Additional Information Contact | Amanda Vander Iest 770-270-1394 Ext. 43453 |
Manufacturer Reason for Recall | The clear, plastic, polycarbonate hematocrit rotor may break apart during use and be ejected through the plastic housing, posing a safety hazard to the technician and potentially damaging the surrounding area. |
FDA Determined Cause 2 | Device Design |
| Action | The recall was initiated by telephone on 02/06/2019. Initial requesting that the consignee immediately cease use of the product. Recall notices were disseminated by telephone and email requesting the return of the product beginning 02/13/2019. The recall is to the user level. The firm has posted a notice on their website stating there is a voluntary recall due to a safety issue and asking users to call if they have a unit purchased after 9/28/18. |
| Quantity in Commerce | 226 units |
| Distribution | Worldwide Distribution - US Nationwide and the countries of Philippines, Honduras, El Salvador, Australia, Guatemala, Canada, Singapore, Australia, Peru, Nicaragua, Trinidad and Tobago, and New Zealand |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
|
|
|