• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall VITROS 5600 Integrated System

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 2 Device Recall VITROS 5600 Integrated System see related information
Date Initiated by Firm February 20, 2019
Create Date May 06, 2019
Recall Status1 Terminated 3 on September 15, 2021
Recall Number Z-1288-2019
Recall Event ID 82311
510(K)Number K081543  
Product Classification Analyzer, chemistry (photometric, discrete), for clinical use - Product Code JJE
Product VITROS¿ 5600 Integrated System Refurbished-Software V3.3.2 & below
Product Code: 6802915

For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Immunodiagnostic Products Reagents
Code Information Serial Numbers: Affects systems that had the Luminometer replaced during a service-repair UDI: 10758750007110
Recalling Firm/
Manufacturer
Ortho-Clinical Diagnostics
100 Indigo Creek Dr
Rochester NY 14626-5101
Manufacturer Reason
for Recall
Luminometer Malfunction May Cause Inability to Process MicroWell Assays on VITROS¿ Systems
FDA Determined
Cause 2
Software design
Action Ortho Clinical Diagnostics (Ortho) issued an "URGENT PRODUCT CORRECTION NOTIFICATION" , a customer letter (Ref. CL2019-052) dated 21 February 2019 via FedEx overnight courier and/or US Postal Service Priority Mail (for PO Boxes only) to customers who were shipped affected VITROS¿ Systems/or whose VITROS System has a luminometer parts replacement. The customers were notified of the issue and instructed to notify Ortho upon experiencing the issue described: -Avoid performing a shutdown/restart on your systems unless directed by an Ortho representative or prompted by condition code help text. -If you shutdown/restart and encounter any of the above Luminometer condition codes and are unable to process MicroWell assays, immediately contact the Ortho Care" Technical Solutions Center for service to your system. -Complete the enclosed Confirmation of Receipt form no later than March 1, 2019 by Fax: 1.888.557.3759 or 1.585.453.4110 or e-Mail Address: RA-OCDUS-CONFIRMAD@ORTHOCLINICALDIAGNOSTICS.COM. -Please forward this notification if the product was distributed outside of your facility. Foreign affiliates were informed by email on 21 February 2019 of the issue and instructed to notify affected consignees of the issue and appropriate actions to take. Questions contact Ortho Care" Technical Solutions Center at 1-800-421-3311.
Quantity in Commerce 4 units
Distribution Worldwide Distribution: US (Nationwide) and countries of: Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Poland, Portugal, Russia, Singapore, Spain, Sweden,The Netherlands, and United Kingdom.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = JJE and Original Applicant = Ortho-Clinical Diagnostics, Inc.
-
-