Date Initiated by Firm | April 03, 2019 |
Create Date | May 10, 2019 |
Recall Status1 |
Terminated 3 on May 04, 2020 |
Recall Number | Z-1313-2019 |
Recall Event ID |
82616 |
510(K)Number | K180438 |
Product Classification |
Devices detecting influenza A, B, and C virus antigens - Product Code PSZ
|
Product | BD Veritor Plus Analyzer- In-Vitro Diagnostic intended to use with the BD Veritor System Test Devices.
Catalog Number: 256066 |
Code Information |
Serial Numbers: Serial Number 1807138J1E050 1710067J160C0 1703027J07FD0 1810268J2D370 1710217J18470 1703027J084C0 1703307J0C630 1710067J16190 1709187J12360 1710267J19D40 1710067J16230 1710067J161D0 1710067J161F0 1710067J16260 1807138J1E0B0 1609076J04270 1802138J07B40 1803298J11070 1802078J05D80 1704067J0DA00 1612166J070C0 1710257J18F50 1802028J044C0 1703167J0B0B0 1612026J07010 1710067J16110 1802058J05210 1811128J2FF50 1710067J16200 1710067J16270 1710237J185C0 1702147J04360 |
Recalling Firm/ Manufacturer |
Becton Dickinson & Co. 7 Loveton Cir Sparks MD 21152-9212
|
Manufacturer Reason for Recall | Background calibration has the potential to cause false positive or false negative test results. An incorrect test result with the BD Veritor System could lead to possible misdiagnosis and inappropriate treatment |
FDA Determined Cause 2 | Under Investigation by firm |
Action | BD notified customers by letter on 4/3/19 via FedX providing problem, health risk and action to take: BD is requesting that customers discard the affected instrument(s) for replacement. BD will replace your affected Veritor instrument(s) upon receipt of the Customer Response Form.
BD Customer/Technical Support 800-638-8663 Monday Friday between 7:00am and 7:00pm (EST) in the United States.For customers outside the US, contact your local BD representative or distributor. |
Quantity in Commerce | 32 units |
Distribution | Nationwide
Foreign: Canada, Chile, Hong Kong, Japan, Taiwan |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = PSZ
|