| Date Initiated by Firm | April 04, 2019 |
| Date Posted | July 03, 2019 |
| Recall Status1 |
Terminated 3 on May 12, 2020 |
| Recall Number | Z-1928-2019 |
| Recall Event ID |
82651 |
| Product Classification |
Calculator/data processing module, for clinical use - Product Code JQP
|
| Product | cobas p 701 post-analytical units, Part Number 05349966001. For diagnostic sample storage and retrieval. |
| Code Information |
Serial Numbers - 1001 through 1037, and 502 |
| FEI Number |
1823260
|
Recalling Firm/ Manufacturer |
Roche Diagnostics Corporation
9115 Hague Rd
Indianapolis
IN
46256-1025
|
| For Additional Information Contact | Roche Customer Support Center 800-428-2336 |
Manufacturer Reason for Recall | There is a potential for the tray input flap on the post analytical units to become loose, potentially detaching from the instrument. |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | On April 4, 2019, Roche issued Urgent Medical Device Correction notices (UMDC TP-00620) to customers via UPS Ground (receipt required).
Customers were advised to take the following actions:
1) Inspect the tray input flap on the cobas p 501 or p 701 post-analytical unit to determine if it is loose. If it is not fastened securely or becomes loose prior to the part being replaced, contact your Roche Field Engineer Specialist (FES) to repair the instrument.
2) Complete and return the fax form (TP-00621) via fax or email
3) File the UMDC for future reference.
Customers with questions may contact the Roche Support Network Customer Support Center 24 hours a day, seven days a week at 1-800-428-2336 if you have questions about the information contained in the UMDC. |
| Quantity in Commerce | 9 units |
| Distribution | US Nationwide Distribution in the states of AL, CA, FL, IA, IL, KY, MD, OR & PA |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|