| Date Initiated by Firm | April 18, 2019 |
| Date Posted | May 31, 2019 |
| Recall Status1 |
Terminated 3 on July 28, 2025 |
| Recall Number | Z-1600-2019 |
| Recall Event ID |
82703 |
| 510(K)Number | K080592 |
| Product Classification |
Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass - Product Code DWF
|
| Product | Maquet Getinge-BEQ-TOP 50903 Custom ECC Pack-Pedriatric
Material: 709000078 |
| Code Information |
Batch Numbers: 3000047631 3000054421 3000063290 3000071804 3000073166 3000075337 |
| FEI Number |
2242352
|
Recalling Firm/ Manufacturer |
Maquet Cardiovascular, LLC
45 Barbour Pond Dr
Wayne
NJ
07470-2094
|
Manufacturer Reason for Recall | Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured |
FDA Determined Cause 2 | Device Design |
| Action | Maquet Getinge issued recall notification letter on April 18, 2019 via FedEx 2-day delivery identifies issue, health risk and action to take: quarantine and return to Maquet/Getinge recalled product. Complete response form and return to acknowledge this recall. Questions contact Getinge Customer Service at 1 888 627 8383 (press option 2, then option 2) between the hours of 8 AM and 6 PM Eastern Standard Time.
On 3/26/20 firm expanded recall to include an additional lot for one consignee,product: "BEQ TOP 25003 3/8 3/8 up to 7 lpm", item number 701063255R02, lot 3000087051. |
| Distribution | Nationwide |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = DWF
|