|
Class 2 Device Recall VITROS 350 Chemistry System, |
|
Date Initiated by Firm |
April 08, 2019 |
Create Date |
July 03, 2019 |
Recall Status1 |
Terminated 3 on November 12, 2020 |
Recall Number |
Z-1936-2019 |
Recall Event ID |
82784 |
510(K)Number |
K922072
|
Product Classification |
Analyzer, chemistry (photometric, discrete), for clinical use - Product Code JJE
|
Product |
VITROS 350 Chemistry System, Product Code 6802153, (UDI # 10758750002054)
Product Usage: Performs discrete clinical tests on serum, urine, and cerebral spinal fluid specimens. Methodologies include colorimetric (CM), potentiometric (PM), rate (RT), and immuno-rate (IR) tests using multi-layered VITROS Slides.
|
Code Information |
All lots |
Recalling Firm/ Manufacturer |
Ortho-Clinical Diagnostics 100 Indigo Creek Dr Rochester NY 14626-5101
|
Manufacturer Reason for Recall |
Potential negative drift in Creatine Kinase (CK) results when used on the VITROS 250 and 350 Systems.
|
FDA Determined Cause 2 |
Device Design |
Action |
On 08 April 2019, a customer letter (Ref. CL2019-090) was sent via FedEx overnight courier and/ or ORTHO PLUS e-Communications and and/or US Postal Service Priority Mail (for PO Boxes only) to either all customers who were shipped affected VITROS 250 or 350 Systems. This communication provided the following required actions:
- As per the VITROS CK Slide Instructions for Use, perform Quality Control (QC) testing daily to help ensure the slides are performing within acceptable limits.
- If QC results are outside of the established ranges, discard the cartridge and repeat QC testing using a fresh cartridge of VITROS CK Slides.
- If multiple cartridges are loaded on the VITROS 250/350 System, daily QC testing must be performed for each cartridge prior to use.
- Consider loading one cartridge of VITROS CK Slides on board your system at a time.
- Complete and return the request form via fax to 1-888-557-3759.
Foreign affiliates were informed by email on 08 April 2019 of the issue and instructed to notify affected consignees of the issue and appropriate actions to take. |
Quantity in Commerce |
3889 units |
Distribution |
US Nationwide Distribution |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = JJE and Original Applicant = EASTMAN KODAK COMPANY
|
|
|
|