 Class 2 Device Recall Getinge Meera Mobile Operating Tablesee related information
Date Initiated by FirmMay 06, 2019
Date PostedJune 04, 2019
Recall Status1 Terminated 3 on July 06, 2022
Recall NumberZ-1734-2019
Recall Event ID 82792
Product Classification Table, examination, medical, powered - Product Code LGX
ProductGETINGE MEERA MOBILE OPERATING TABLE: (1) Part 720001B2 (2) Part 720001B0 (3) Part 720001F2 (4) Part 720001F0 (5) Part 710001B2 (6) Part 710001B0 (7) Part 710001F0 Product Usage: The Meera Mobile Operating Table is designed to support and position the patient immediately before, during and after surgical interventions as well as for examination and treatment.
Code Information (1) Part 720001B2 S/N 1 to S/N 416 (2) Part 720001B0 S/N 1 to S/N 1416 (3) Part 720001F2 S/N 1 to S/N 91 (4) Part 720001F0 S/N 1 to S/N 220 (5) Part 710001B2 S/N 1 to S/N 8 (6) Part 710001B0 S//N 1 to S/N 11 (7) Part 710001F0 S/N 1  Expanded lots: 720001B2 SN 417 to SN 503 720001B0 SN 1417 to SN 1689 720001F2 SN 92 to SN 119 720001F0 SN 221 to SN 226 and SN 701 to 711 710001B2 SN 9 to SN 14 710001B0 SN 12 to SN 88
FEI Number 3008355164
Recalling Firm/
Manufacturer
Maquet Cardiovascular Us Sales, Llc
45 Barbour Pond Dr
Wayne NJ 07470-2094
For Additional Information ContactSAME
973-709-7000
Manufacturer Reason
for Recall
User may suffer an electric shock when coming in contact with the exposed electronic components
FDA Determined
Cause 2
Process control
ActionMaquet /Getinge issued URGENT MEDICAL DEVICE CORRECTION Notice on 5/6/19, identified issued, health risk and action to take: Per the IFU, section 8.1, a visual and functional check is required to be performed by a trained person before operation; If the plastic socket of the cold-device plug is pulled out do not try to put it back into the housing and if there are any exposed electronic components: do not touch them. Call your in-house technical service and arrange the repair of the table. Per IFU section 8.5 a defective product may not be used and may not be repaired by yourself. Complete/sign and return the MEDICAL DEVICE RECALL  CORRECTION RESPONSE FORM. Questions contact Technical Support Department at 1-888-627-8383 (press option 3 then option 1), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. EST. Expanded Recall: Maquet/Getinge issued letters on 9/24/20 to consignees via FedEx 2 day Delivery with Signature Proof of Delivery (SPOD).
Quantity in CommerceU.S: 163; Foreign: 1864. Expanded: US:27 Foreign: 455
DistributionWorldwide - US Nationwide Distribution Foreign: ALGERIA ANGOLA ARGENTINA AUSTRALIA AUSTRIA BAHRAIN BANGLADESH BELGIUM BOLIVIA BOTSWANA BOSNIA AND HERZEGOVINA BRAZIL BULGARIA CANADA CHILE CHINA COLOMBIA COSTA RICA CROATIA CZECH REPUBLIC DENMARK ECUADOR EGYPT FINLAND FRANCE GABON GERMANY GREECE HONG KONG HONDURAS HUNGARY ICELAND INDIA IRAN IRELAND ISRAEL ITALY JAPAN JORDAN KUWAIT LATVIA LEBANON LITHUANIA MACEDONIA MALAYSIA MAURITIUS MEXICO MOLDAVA MONTENEGRO MOROCCO MYANMAR NETHERLANDS NAMIBIA NORWAY NEW ZEALAND PAKISTAN PARAGUAY PERU PHILIPPINES POLAND PORTUGAL ROMANIA SAUDI ARABIA SERBIA SINGAPORE SLOVAKIA SLOVENIA SOUTH AFRICA SOUTH KOREA SPAIN SWEDEN SWITZERLAND SYRIA TAIWAN TANZANIA THAILAND TURKEY TURKMENISTAN UKRAINE UNITED ARAB EMIRATES URUGUAY UNITED KINGDOM VIETNAM