• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall KWire Trochar Tip

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall KWire Trochar Tipsee related information
Date Initiated by FirmMay 28, 2019
Create DateJuly 04, 2019
Recall Status1 Terminated 3 on November 19, 2021
Recall NumberZ-1948-2019
Recall Event ID 83024
Product Classification Orthopedic manual surgical instrument - Product Code LXH
ProductK-Wire Trochar Tip, Model Number 212000008 Product Usage: These DVR Express Disposable Instruments are used to facilitate implantation of orthopedic medical devices in trauma.
Code Information Lot Numbers: 187764 193497 193567 232280 232683 232684 232685 232686 233858 234049 235477 BFI0LJ8 BFI0M5B BFI0MAV BFI0N16 BFI0NNN I0N17 I0N18 BFI0LJ8R BFI0MAVR BF10LJ8 BF10MAV BFIOLI8 I0NNN 
Recalling Firm/
Manufacturer
Zimmer Biomet, Inc.
56 E Bell Dr
Warsaw IN 46582-6989
For Additional Information Contact411 Technical Services
574-371-3071
Manufacturer Reason
for Recall
There is a potential for weak seals of the sterile packaging.
FDA Determined
Cause 2
Under Investigation by firm
ActionUrgent Medical Device Recall notification letters dated 5/28/19 were sent to customers. The letter identified the affected product, problem and actions to be taken. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency.
Quantity in Commerce3022 total
DistributionWorldwide Distribution - US Nationwide: AL, AR, AZ, CA, CO, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, NC, NE, NJ, NM, NY, OH OK, OR, PA, SC, TN, TX, UT, WA, WI, WV, and Puerto Rico. Products were distributed to the following foreign countries: Australia, Canada, Chile, Japan, Netherlands.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-