| Class 2 Device Recall Vereos PET/CT | |
Date Initiated by Firm | June 03, 2019 |
Create Date | July 03, 2019 |
Recall Status1 |
Terminated 3 on April 26, 2021 |
Recall Number | Z-1929-2019 |
Recall Event ID |
83187 |
510(K)Number | K123599 |
Product Classification |
System, tomography, computed, emission - Product Code KPS
|
Product | Vereos PET/CT, Model No. 882446
Product Usage:
Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems. The CT subsystem images anatomical cross-sections by computer reconstruction of X-ray transmission data. The PET subsystem images the distribution of PET anatomy-specific radiopharmaceuticals in the patient. The PET/CT system is used for the purposes of detecting, localizing, diagnosing, staging, re-staging, and follow-up for monitoring therapy response of various diseases in oncology, cardiology, and neurology. The system is intended to image the whole body, heart, brain, lung, gastrointestinal, bone, lymphatic, and other major organs for a wide range of patient types, sizes, and extent of diseases. Both subsystems can also be operated as fully functional, independent diagnostic tools, including application of the CT scanner for diagnosis and for use in radiation therapy planning. |
Code Information |
Serial Nos. 900050 900036 900047 900042 900019 900055 900063 900048 900045 900011 900037 900044 900007 900049 900017 900067 900058 900020 900035 900032 900024 900064 900040 900054 900025 900030 900005 900028 900059 900009 900052 900053 900046 900026 900051 900013 900065 900015 900006 900027 900075 900031 900073 900041 900016 900012 900057 900043 900021 900062 900033 900038 900010 900029 900066 900023 900008 900039 900060 900018 900022 900014 900004 900034 900056 900061 900069 900070 900068 900071 |
Recalling Firm/ Manufacturer |
Philips Medical Systems (Cleveland) Inc 595 Miner Rd Cleveland OH 44143-2131
|
For Additional Information Contact | Ms. Holly Wright Lee 440-483-7600 |
Manufacturer Reason for Recall | The Vereos PET/CT may have misaligned front and rear covers, which may result in gaps between cover parts and/or protrusion of the cover into the bore diameter. |
FDA Determined Cause 2 | Device Design |
Action | On June 3, 2019, the firm notified their customers of the recall via an "Urgent Field Safety Notice Medical Device Correction" letter.
Customers were advised to take the following actions:
* The operator is instructed to observe the patient during motorized motion. In case of collision, the operator should use the E-Stop to abort
all motions.
" The operator is instructed to ensure accessories do not collide with
gantry covers. Additionally, ensure that sheets, blankets and/or IV lines
are not loose or hanging outside the Patient Pallet.
" Ensure the patients arms/hands are positioned appropriately at their
side or overhead. If necessary, use the arm restraint strap as
described in the IFU.
" Instruct the patient to not move their hands during the scan and during
any table motions until the operator removes the patient from the
scanner at the completion of the study.
Philips will deploy Field Change Order (FCO) 88200520 free of charge to address overall cover fit on the affected systems. A Philips Field Service Engineer will contact you to schedule the updates at your site.
If you need any further information or support concerning this issue, please contact your local Philips representative: For North America and Canada, contact the Customer Care Solutions
Center (1-800-722-9377, follow the prompts). |
Quantity in Commerce | 70 |
Distribution | Worldwide Distribution - US Nationwide
Domestic distribution to CA
GA
LA
MD
MI
MN
MO
NE
NJ
NY
OH
TN
TX
VT
WI.
Foreign distribution to Mexico, Argentina
Austria
Belgium
China
Colombia
France
Germany
Israel
Italy
Japan
Latvia
Luxembourg
Netherlands
Portugal
Singapore
Spain
Switzerland |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = KPS
|
|
|
|