Date Initiated by Firm |
October 22, 2019 |
Create Date |
March 02, 2020 |
Recall Status1 |
Completed |
Recall Number |
Z-0928-2020 |
Recall Event ID |
84464 |
510(K)Number |
K141133
|
Product Classification |
Analyzer, chemistry (photometric, discrete), for clinical use - Product Code JJE
|
Product |
BIOMERIEUX VIDAS 3, Compact immunoanalyzer with full traceability and automation |
Code Information |
software versions 1.2 and higher |
Recalling Firm/ Manufacturer |
Biomerieux Inc 595 Anglum Rd Hazelwood MO 63042-2320
|
For Additional Information Contact |
bioMerieux Customer Support Operations 800-682-2666
|
Manufacturer Reason for Recall |
Following Customers complaints, investigations have been initiated on potentially false results obtained on VIDAS 3 with an expired calibration.
Indeed, the calibrations of assays were valid in the calibration menu whereas in fact the calibrations were expired and no alarm displayed to warn the users about the expiration date of the calibrations. Software computes results of analysis for assays with expired calibrations.
The anomaly is due to an incorrect update of the calibration status by the software.
|
FDA Determined Cause 2 |
Software design |
Action |
The firm initiated their field action by an email to their subsidiaries on 10/22/2019. The subsidiaries then notified their customers by 2-day express via Fed Ex. In the US customer letter, customers are instructed to check their software version. If the VIDAS 3 software version is between 1.2.0 and 1.3.1, customers are instructed to check the archiving mode. If the archiving mode is Automatic When Results are Reported then customers are instructed how to configure the workaround, and to contact bioMerieux if the customer has the incorrect software configuration. bioMerieux then instructs impacted customers to review and interpret results in the context of the overall clinical procedure. Once this is completed to complete the acknowledgment form in Attachment A and return it the local bioMerieux representative and confirm receipt of this notice. |
Quantity in Commerce |
738 units |
Distribution |
US |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
|
510(K) Database |
510(K)s with Product Code = JJE and Original Applicant = BIOMERIEUX
|