|
Class 2 Device Recall |
|
Date Initiated by Firm |
January 15, 2020 |
Date Posted |
January 24, 2020 |
Recall Status1 |
Terminated 3 on March 17, 2023 |
Recall Number |
Z-1185-2020 |
Recall Event ID |
84726 |
Product Classification |
Catheter, Conduction, Anesthetic - Product Code BSO
|
Product |
FlexTip Plus Epidural Catheterization Kit, Product Code ASK-17019-MSC
|
Code Information |
Lots: 13F18K0094 13F19B0335 13F19C0594 13F19G0068 13F19G0591 13F19J0119 |
Recalling Firm/ Manufacturer |
Arrow International Inc 2400 Bernville Rd Reading PA 19605-9607
|
For Additional Information Contact |
Customer Service 866-396-2111
|
Manufacturer Reason for Recall |
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
|
FDA Determined Cause 2 |
Other |
Action |
An Urgent Medical Device Recall notification letter dated 1/15/20 was sent to customers.
Our records indicate you have received products that are subject to this field action. We are now notifying our customers to take the following actions:
1. If you have affected stock, immediately identify product, discontinue use and quarantine any products with the product codes and lot numbers listed above.
2. To acknowledge receipt of this Urgent Medical Device Recall Notice or to return product, complete the enclosed Recall Acknowledgement Form and fax it to 1-855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. This will allow us to document the amount of product you have on hand for return. A customer service representative will contact you with a Return Goods Authorization (RGA) Number and will provide instructions for the return of products to Teleflex Medical.
3. If you have no affected stock, please complete the enclosed Recall Acknowledgment Form and fax it to 1-855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. This will allow us to document your receipt of this letter. |
Quantity in Commerce |
343,105 total |
Distribution |
The products were distributed to the following US states: AK, AR, AZ, CA, CO, CT, FL, GA, ID, IL, IN, LA, MA, MD, MI, MN, MO, MT, NC, NE, NH, NJ, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WY. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|