|
Class 2 Device Recall |
 |
Date Initiated by Firm |
March 18, 2020 |
Date Posted |
April 29, 2020 |
Recall Status1 |
Open3, Classified |
Recall Number |
Z-1819-2020 |
Recall Event ID |
85297 |
Product Classification |
Electrocardiograph - Product Code DPS
|
Product |
Replaceable lead set for the Wireless Acquisition Module (WAM), which is the ECG front-end for various Electrocardiographs. Sold under the following device names: LEAD SET WAM 10 WIRE BANANA AHA GRAY, LEAD SET WAM 10 WIRE BANANA IEC GRAY, RPLCE LD SET WAM/AM12 LIMBS BANA AHA GRY, RPLCE LD SET WAM/AM12 LIMBS BANA IEC GRY, RPLCE LD SET WAM/AM12 V1-V3 BANA AHA GRY, RPLCE LD SET WAM/AM12 C1-C3 BANA IEC GRY, RPLCE LD SET WAM/AM12 V4-V6 BANA AHA GRY, RPLCE LD SET WAM/AM12 C4-C6 BANA IEC GRY, LEAD SET AM15E 13-WIRE BANANA AHA GRAY, LEAD SET AM15E 13-WIRE BANANA IEC GRAY, RPLCE LEAD SET AM15E E2-E4 BANANA AHA GRAY, LEAD SET WAM 10 WIRE CLIPS AHA GRAY, LEAD SET WAM 10 WIRE CLIPS IEC GRAY, RPLCE LD SET WAM/AM12 LIMBS CLIP AHA GRY, RPLCE LD SET WAM/AM12 LIMBS CLIP IEC GRY, RPLCE LD SET WAM/AM12 V1-V3 CLIP AHA GRY, RPLCE LD SET WAM/AM12 C1-C3 CLIP IEC GRY, RPLCE LD SET WAM/AM12 V4-V6 CLIP AHA GRY, RPLCE LD SET WAM/AM12 C4-C6 CLIP IEC GRY, LEAD SET WAM/AM12 10-WIRE BANANA AHA GRY, LEAD SET WAM/AM12 10-WIRE BANANA IEC GRY, LD SET WAM/AM12 10 WRE SHORT AHA CLIP, LD SET WAM/AM12 10 WRE SHORT IEC CLIP. |
Code Information |
Model Numbers: 9293-046-60 - 9293-047-71; 9293-017-50 - 9293-017-51; 9293-026-50 - 9293-026-51; 9293-061-50 - 9293-061-53; 9293-033-52 - 9293-033-53 and 9293-034-50. |
Recalling Firm/ Manufacturer |
WELCH ALLYN, INC/MORTARA 7865 N 86th St Milwaukee WI 53224-3431
|
For Additional Information Contact |
888-667-8272
|
Manufacturer Reason for Recall |
Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.
|
FDA Determined Cause 2 |
Device Design |
Action |
Welch Allyn notified customers on 2/19/2020 via "Voluntary Medical Device Correction Welch Allyn Patient Cables or Lead Sets". The recall letter identified the affected product and stated the reason for the recall. The firm also requested the distributors to immediately notify accounts if the product was further distributed. |
Quantity in Commerce |
78,521 units total |
Distribution |
US nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, ON, OR, PA, PR, RI, SC, SD, TN,UT, VA, VT, WA, WI, WV and WY. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|