 Class 2 Device Recall Maestro Orthodontic Bracketsee related information
Date Initiated by FirmApril 02, 2019
Create DateJune 30, 2020
Recall Status1 Terminated 3 on March 24, 2022
Recall NumberZ-2478-2020
Recall Event ID 85468
Product Classification Bracket, metal, orthodontic - Product Code EJF
ProductMaestro Orthodontic Bracket REF711-298Q QTY. 20 Rx Only (01)00190707017064 (10)889437 (422)484 Maestro Orthodontic Bracket REF711-299MCQ QTY. 20 Rx Only (01)00190707017071(10)889437 (422)484
Code Information Maestro Orthodontic Bracket ,.018 Kit U/L SxS P/N711-298Q (01)00190707017064 (10)889437 (422)484 Lot# 878285  Maestro Orthodontic Bracket .018 Kit U/L SxS MC Hooks P/N711-299MCQ UDI (01)00190707017071(10)889437 (422)484 Lot#872900    
FEI Number 2081322
Recalling Firm/
Manufacturer
Ortho Organizers, Inc.
1822 Aston Ave
Carlsbad CA 92008-7306
For Additional Information ContactCustomer Service
760-448-8600
Manufacturer Reason
for Recall
Orthodontic Bracket is mislabeled
FDA Determined
Cause 2
Process control
ActionThe firm notified their consignees of the recall by letter sent out 1st class mail on 4/2/2019. The recall was initiated due to a mislabeling error. The Upper Right bracket was erroneously color coded as an Upper Left brackets and then comingled within the same bag. The firm is asking recipients of affected product to please contact Customer Service at 1-800-547-2000 to receive an RMA number. The RMA will allow for a quick return and replacement or credit. The firm is requesting that distributors and physicians complete and E-mail the acknowledgment form back to the firm in order to confirm receipt of the field correction letter. Additionally they are asking authorized distributor,to identify those customers that may have been shipped the affected product and contact those customers to inform them of the issue within fortyeight (48) hours of receipt of this field correction notification.
Quantity in Commerce147 kits
Distribution13 US Consignees in FL, GA, CA, MI, ID, TX 1 = Italy 1 = Japan 1 = Greece 1 = Saudi Arabia 1 = Russia 1 = Czech Republic
Total Product Life CycleTPLC Device Report