| | Class 2 Device Recall Leica ARveo and M530 OH6 microscope systems. |  |
| Date Initiated by Firm | May 20, 2020 |
| Create Date | June 17, 2020 |
| Recall Status1 |
Terminated 3 on May 13, 2021 |
| Recall Number | Z-2399-2020 |
| Recall Event ID |
85722 |
| Product Classification |
Microscope, surgical - Product Code EPT
|
| Product | Leica ARveo and M530 OH6 microscope systems. |
| Code Information |
M530 OH6 Serial Numbers for Model # 10448680 OH6-016215, OH6-042317, OH6-032016, OH6-009815,OH6-044518 and OH6-030016. Arveo Serial Numbers for Model # 10448999 OHA-006318, OHA-010719, OHA-011919, OHA-000818, OHA-009119, OHA-000718, OHA-014919, OHA-016619, OHA-015919, OHA-007818, OHA-013519, OHA-001218, OHA-012219, OHA-017319, OHA-000418, OHA-017119, OHA-013719, OHA-009919, OHA-017219, OHA-017419, OHA-002218, OHA-000318, OHA-014819, OHA-015219, OHA-003718, OHA-001018, OHA-016019, OHA-012919, OHA-012619, OHA-011719, OHA-012119, OHA-010819, OHA-016119, OHA-009219, OHA-002618, OHA-015719, OHA-012019, OHA-009519, OHA-000618, OHA-008518, OHA-004818, OHA-012719, OHA-010619, OHA-001118, OHA-010419, OHA-016319, OHA-016519, OHA-010219, OHA-012519, OHA-015419, OHA-009018, OHA-016419, OHA-016719, OHA-008918, OHA-007618, OHA-009619, OHA-012319, OHA-010519, OHA-005418. |
| FEI Number |
1423337
|
Recalling Firm/ Manufacturer |
Leica Microsystems, Inc.
1700 Leider Ln
Buffalo Grove
IL
60089-6622
|
| For Additional Information Contact | Jason Battle 800-225-8867 |
Manufacturer Reason for Recall | Possible unintended interruption of surgical procedures due to equipment shutdown, related to changes in voltage. |
FDA Determined Cause 2 | Other |
| Action | The firm sent letters to consignees on May 20, 2020 advising them that they can continue to use the system pending this field update. Based on our data analysis, the probability of occurrence for this issue (excessive voltage fluctuation and a subsequent microscope shut off) is remote. However, please ensure that a functional check is performed prior to each surgery as described in section 16.1, directions for the operator of the instrument of the user manual. |
| Quantity in Commerce | Domestic distribution 65 devices. International distribution 12 devices. Total of devices distributed 77. |
| Distribution | Domestic distribution: AL,AR,AZ, CA, CT, DC, DE, FL, IL, KY, LA, MI, MN, MO, NC, NY, OH. OR. TN, TX, UT, VA, WA, WI ,WV.
International distribution: Japan |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
|
|
|