 Class 2 Device Recall Leica ARveo and M530 OH6 microscope systems.see related information
Date Initiated by FirmMay 20, 2020
Create DateJune 17, 2020
Recall Status1 Terminated 3 on May 13, 2021
Recall NumberZ-2399-2020
Recall Event ID 85722
Product Classification Microscope, surgical - Product Code EPT
ProductLeica ARveo and M530 OH6 microscope systems.
Code Information M530 OH6 Serial Numbers for Model # 10448680  OH6-016215, OH6-042317, OH6-032016, OH6-009815,OH6-044518 and OH6-030016.  Arveo Serial Numbers for Model # 10448999  OHA-006318, OHA-010719, OHA-011919, OHA-000818, OHA-009119, OHA-000718, OHA-014919, OHA-016619, OHA-015919, OHA-007818, OHA-013519, OHA-001218, OHA-012219, OHA-017319, OHA-000418, OHA-017119, OHA-013719, OHA-009919, OHA-017219, OHA-017419, OHA-002218, OHA-000318, OHA-014819, OHA-015219, OHA-003718, OHA-001018, OHA-016019, OHA-012919, OHA-012619, OHA-011719, OHA-012119, OHA-010819, OHA-016119, OHA-009219, OHA-002618, OHA-015719, OHA-012019, OHA-009519, OHA-000618, OHA-008518, OHA-004818, OHA-012719, OHA-010619, OHA-001118, OHA-010419, OHA-016319, OHA-016519, OHA-010219, OHA-012519, OHA-015419, OHA-009018, OHA-016419, OHA-016719, OHA-008918, OHA-007618, OHA-009619, OHA-012319, OHA-010519, OHA-005418.
FEI Number 1423337
Recalling Firm/
Manufacturer
Leica Microsystems, Inc.
1700 Leider Ln
Buffalo Grove IL 60089-6622
For Additional Information ContactJason Battle
800-225-8867
Manufacturer Reason
for Recall
Possible unintended interruption of surgical procedures due to equipment shutdown, related to changes in voltage.
FDA Determined
Cause 2
Other
ActionThe firm sent letters to consignees on May 20, 2020 advising them that they can continue to use the system pending this field update. Based on our data analysis, the probability of occurrence for this issue (excessive voltage fluctuation and a subsequent microscope shut off) is remote. However, please ensure that a functional check is performed prior to each surgery as described in section 16.1, directions for the operator of the instrument of the user manual.
Quantity in CommerceDomestic distribution 65 devices. International distribution 12 devices. Total of devices distributed 77.
DistributionDomestic distribution: AL,AR,AZ, CA, CT, DC, DE, FL, IL, KY, LA, MI, MN, MO, NC, NY, OH. OR. TN, TX, UT, VA, WA, WI ,WV. International distribution: Japan
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
-
-