| | Class 2 Device Recall Universal Drainage Set |  |
| Date Initiated by Firm | May 28, 2020 |
| Create Date | July 20, 2020 |
| Recall Status1 |
Terminated 3 on September 20, 2024 |
| Recall Number | Z-2654-2020 |
| Recall Event ID |
85751 |
| Product Classification |
Thoracentesis tray - Product Code PXI
|
| Product | Thora-Para 8 Fr Catheter Drainage Tray w/o Lidocaine |
| Code Information |
Catalog: TPT1000CDF UDI: 50885403231900; Batch 0001333309 |
| FEI Number |
3008002452
|
Recalling Firm/ Manufacturer |
Carefusion 2200 Inc 75 N Fairway Dr Vernon Hills IL 60061-1845
|
| For Additional Information Contact | BD Customer Support 888-237-2762 |
Manufacturer Reason for Recall | The Universal Drainage Set (UDS) may contain valves that leak internally which may result in the introduction of up to 30cc of air into the patient. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | All affected consignees are being notified via FedEx mailed letter and/or email beginning 6/15/2020. BD requests a sub-recall to be conducted if product has been further distributed. Consignees are asked to complete the acknowledgement form included with the notification letter and return to BD. |
| Distribution | 370 US customers and 13 distributors.
Product distributed in: AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MD, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY
Customer listed, including Federal distribution, uploaded. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|