• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Ultra/Phonic Focus BP Conforming Gel Pads

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Ultra/Phonic Focus BP Conforming Gel Padssee related information
Date Initiated by FirmAugust 13, 2020
Create DateAugust 20, 2020
Recall Status1 Terminated 3 on September 16, 2021
Recall NumberZ-2830-2020
Recall Event ID 86138
Product Classification Media, coupling, ultrasound - Product Code MUI
ProductUltra/Phonic Focus BP Conforming Gel Pads, UDI 00300361301021 (Primary); 10300361301028 (Package) - Product Usage: A gel pad for use on woman s areola during an ultrasound examination. Its primary purpose is for patient comfort over the loose gel and helps with nipple location during the scan, making the areola location easier to identify.
Code Information Lots  021805-02 030805-02 050405-02 051106-02 052406-02 101006-02 101806-02 103406-02 052708-02 053308-02 060308-02 111208-02 111908-02 112108-02 050309-02 050809-02 101009-02 101309-02 061310-02 061710-02 061910-02 021411-02 021911-02 030211-02 030711-02 070711-02 113211-02 113311-02 011312-02 022112-02 030412-02 030812-02 061712-02 061912-02 070512-02 120612-02 011013-02 011313-02 021413-02 030313-02 051313-02 052713-02 060113-02 120613-02 011414-02 021314-02 022614-02 030614-02 060914-02 061314-02 062014-02 070714-02 100314-02 101614-02 110914-02 120314-02 020115-02 030115-02 060615-02 061315-02 070515-02 070815-02 100315-02 100615-02 110315-02 110715-02 120316-02 010416-02 011316-02 020216-02 021016-02 030516-02 052016-02 060316-02 091416-02 092316-02 100616-02 101516-02 102116-02 020517-02 021617-02 030817-02 031517-02 032417-02 040717-02 041017-02 050617-02 060317-02 071217-02 071917-02 080417-02 081717-02 091017-02 091617-02 092317-02 100517-02 021018-02 031318-02 040418-02 040718-02 041118-02 041318-02 041818-02 051218-02 060718-02 020619-02 020919-02 
FEI Number 2241583
Recalling Firm/
Manufacturer
Pharmaceutical Innovations, Inc.
897 Frelinghuysen Ave
Newark NJ 07114-2122
For Additional Information ContactFirm Contact
973-242-2900
Manufacturer Reason
for Recall
The firm registered the product as a Class I device, but has since been informed by FDA that the device is not Class I and requires a 510(k) submission.
FDA Determined
Cause 2
No Marketing Application
ActionOn 8/13/2020, the firm distributed Urgent Medical Device Recall letters to its distributor, with instructions to notify all affected customers who received the product. Customers were informed that the product had been misclassified by the firm and required a 510(k) submission to FDA. Customers were instructed to identify and quarantine the products, and to cease use and sales of the product. Customers are to complete the included response form and return it to the firm. The firm will issue a Return Material Authorization for the return of the product. No option will be given to destroy product in the field. Customers should notify all users within their facility as well as any downstream accounts if the product has been further distributed/ transferred.
Quantity in Commercetbd
DistributionUS Nationwide distribution including in the state of Nevada.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
-
-