Date Initiated by Firm | September 22, 2020 |
Date Posted | October 13, 2020 |
Recall Status1 |
Open3, Classified |
Recall Number | Z-0101-2021 |
Recall Event ID |
86504 |
510(K)Number | K131307 |
Product Classification |
Calibrator, secondary - Product Code JIT
|
Product | ADVIA Chemistry Systems-Fructosamine (FRUC)-For in vitro diagnostic use in the quantitative measurement of glycated protein (fructosamine) in human serum or plasma on the ADVIA Chemistry systems
Siemens Material Number (SMN): 10361941 |
Code Information |
All lots UDI: (01)00630414558073(10)501874(17)20201128 (01)00630414558073(10)514594(17)20210221 (01)00630414558073(10)528971(17)20210630 |
Recalling Firm/ Manufacturer |
Siemens Healthcare Diagnostics, Inc. 511 Benedict Ave Tarrytown NY 10591-5005
|
For Additional Information Contact | SAME 914-631-8000 |
Manufacturer Reason for Recall | Fructosamine QC and patient samples may exhibit a 60 umol/L positive bias across the measuring range compared to predicate assay reference leading to a potential adjustment of hyperglycemic therapy. Positively biased fructosamine results may prompt changes in glucose management that may affect glycemia. In extreme situations, more aggressive management of glycemia may occur and contribute to hypoglycemia. |
FDA Determined Cause 2 | Device Design |
Action | Siemens Urgent Medical Device Recall (ACHC20-18.A.US.CHC) distributed to customers on 9/22/20, in the United States and an Urgent Field Safety Notice (ACHC20-18.A.OUS.CHC) was distributed to customers outside the United States who have received ADVIA Chemistry Fructosamine in-date reagent lots.
For Atellica, an Urgent Medical Device Recall (ACHC20-18.A.US) was distributed to l customers in the United States and an Urgent Field Safety Notice (ACHC20-1.OUS) was distributed to all customers outside the United States who have received Atellica CH in date reagent lots.
These letters instruct the customers to:
Perform the following actions on the ADVIA Chemistry System
1. Enter the Real-time correction factor in the Analytical Parameters (Chemistry) window.
2. Once the Real-time correction factor has been configured, update QC ranges as described above in the FRUC
QC Definition following your laboratory procedures.
3. Update the reference range following your laboratory procedures.
4. Perform a FRUC calibration and process QC.
5. Perform a system back-up.
Complete and return the Field Correction Effectiveness Check attached to this letter within 30 days.
" Review this letter with your Medical Director.
" If you have received any complaints of illness or adverse events associated with the products listed in Table 1, immediately contact your local Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative |
Quantity in Commerce | US: 23 units; OUS: 204 units |
Distribution | CA, DE, MD, NC, NJ, NY, PA
Foreign:
Argentina¿¿¿¿¿¿
Australia¿¿¿¿¿¿
Brazil¿¿¿¿
Canada¿¿¿¿¿
Czech¿Republic¿
Denmark¿¿¿¿¿¿¿¿
Germany¿¿¿¿¿¿¿¿
Italy¿¿¿¿¿¿¿¿¿¿
Netherlands¿¿¿¿
P.R.¿China¿¿¿¿¿
Paraguay¿¿¿¿¿¿¿
Slovakia¿¿¿¿¿¿¿
Turkey¿¿¿¿¿¿¿¿¿
United¿Kingdom¿
Vietnam¿¿¿¿¿¿¿¿ |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
510(K) Database | 510(K)s with Product Code = JIT
|