| | Class 2 Device Recall FLOW COUPLER Monitor |  |
| Date Initiated by Firm | December 28, 2020 |
| Create Date | January 30, 2021 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0939-2021 |
| Recall Event ID |
87013 |
| 510(K)Number | K191252 |
| Product Classification |
Device, anastomotic, microvascular - Product Code MVR
|
| Product | FLOW COUPLER Monitor
Flow Coupler devices are sold in 5 sizes (2.0 mm, 2.5 mm, 3.0 mm, 3.5 mm, 4.0 mm).
Implantable device for vascular use. |
| Code Information |
UDI: 00844735006148; Product Code: 5156-00000-011; Serial Numbers: 135306000279, 135306000280, 135306000281, 135306000282, 135306000283, 135306000284, 135306000286, 135306000287, 135306000288, 135306000289, 135306000290, 135306000291, 135306000292, 135306000293, 135306000294, 135306000295, 135306000298, 135306000299, 135306000300, 135306000301, 135306000303, 135306000304, 135306000305, 135591000306, 135591000307, 136959000309, 136959000310, 136959000311, 136959000312, 136959000313, 136959000314, 136959000315, 136959000316, 136959000317, 136959000318, 136959000319, 136959000320, 136959000321, 136959000322, 136959000323, 136959000324, 136959000325, 136959000326, 136959000327, 136959000328, 136959000329, 136959000330, 136959000331, 136959000332, 136959000333, 136959000334, 136959000335, 136959000336, 136959000337, 136959000338, 136959000339, 136959000340, 136959000341, 136959000342, 136959000344, 136959000345, 136959000346, 136959000347, 136959000348, 136959000349, 136959000350, 136959000351, 136959000352, 136959000353, 136959000354, 136959000355, 136959000357, 136959000358, 136959000359, 136959000360, 136959000361, 136959000362, 136959000364, 136959000365, 136959000366, 137645000392, 137645000393, 137645000394, 137645000395, 137645000396, 137645000397, 137645000399, 137645000400, 137645000401, 139608000552, 139608000555, 139608000571, 139609000583, 139609000589, 139609000596, 139609000597, 139609000598, 139609000599, 139609000600, 139609000601, 139609000602, 139609000603, 139609000604, 139609000605, 139609000606, 139609000607, 139609000608, 139609000609, 139609000610 |
| FEI Number |
1417572
|
Recalling Firm/ Manufacturer |
Baxter Healthcare Corporation 1 Baxter Pkwy Deerfield IL 60015-4625
|
| For Additional Information Contact | Center for One Baxter 800-422-9837 |
Manufacturer Reason for Recall | Potential for the battery to lose its ability to be recharged. |
FDA Determined Cause 2 | Software design |
| Action | Written notification letter titled, "Urgent Medical Device Recall" was sent to all consignees on 12/28/2020 via first class mail. The letter instructs the consignee to locate all affected devices at their facility; contact Synovis MCA to arrange for product return for the firmware upgrade, power button update, and battery replacement, if applicable; complete the enclosed Baxter customer reply form and return it to Baxter by e-mailing it to fca@baxter.com, even if the consignees does not have any inventory; and if the consignee distributes this product to other facilities or departments within their institution, to please forward a copy of this communication to them. |
| Quantity in Commerce | 109 units |
| Distribution | Distributed nationwide to AZ, FL, OH, NY, GA, VA, MA, WI, MO, MT, WV, IL, NC, NJ, AR, CA, CO, CT, IA, ID, IN, KS, LA, MD, MI, MN, MS, NE, NH, NM, NV, OK, OR, PA, RI, WY, WA, VT, TX, TN, SC, SD |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|