 Class 2 Device Recall FLOW COUPLER Monitorsee related information
Date Initiated by FirmDecember 28, 2020
Create DateJanuary 30, 2021
Recall Status1 Open3, Classified
Recall NumberZ-0939-2021
Recall Event ID 87013
510(K)NumberK191252 
Product Classification Device, anastomotic, microvascular - Product Code MVR
ProductFLOW COUPLER Monitor Flow Coupler devices are sold in 5 sizes (2.0 mm, 2.5 mm, 3.0 mm, 3.5 mm, 4.0 mm). Implantable device for vascular use.
Code Information UDI: 00844735006148; Product Code: 5156-00000-011; Serial Numbers: 135306000279, 135306000280, 135306000281, 135306000282, 135306000283, 135306000284, 135306000286, 135306000287, 135306000288, 135306000289, 135306000290, 135306000291, 135306000292, 135306000293, 135306000294, 135306000295, 135306000298, 135306000299, 135306000300, 135306000301, 135306000303, 135306000304, 135306000305, 135591000306, 135591000307, 136959000309, 136959000310, 136959000311, 136959000312, 136959000313, 136959000314, 136959000315, 136959000316, 136959000317, 136959000318, 136959000319, 136959000320, 136959000321, 136959000322, 136959000323, 136959000324, 136959000325, 136959000326, 136959000327, 136959000328, 136959000329, 136959000330, 136959000331, 136959000332, 136959000333, 136959000334, 136959000335, 136959000336, 136959000337, 136959000338, 136959000339, 136959000340, 136959000341, 136959000342, 136959000344, 136959000345, 136959000346, 136959000347, 136959000348, 136959000349, 136959000350, 136959000351, 136959000352, 136959000353, 136959000354, 136959000355, 136959000357, 136959000358, 136959000359, 136959000360, 136959000361, 136959000362, 136959000364, 136959000365, 136959000366, 137645000392, 137645000393, 137645000394, 137645000395, 137645000396, 137645000397, 137645000399, 137645000400, 137645000401, 139608000552, 139608000555, 139608000571, 139609000583, 139609000589, 139609000596, 139609000597, 139609000598, 139609000599, 139609000600, 139609000601, 139609000602, 139609000603, 139609000604, 139609000605, 139609000606, 139609000607, 139609000608, 139609000609, 139609000610
FEI Number 1417572
Recalling Firm/
Manufacturer
Baxter Healthcare Corporation
1 Baxter Pkwy
Deerfield IL 60015-4625
For Additional Information ContactCenter for One Baxter
800-422-9837
Manufacturer Reason
for Recall
Potential for the battery to lose its ability to be recharged.
FDA Determined
Cause 2
Software design
ActionWritten notification letter titled, "Urgent Medical Device Recall" was sent to all consignees on 12/28/2020 via first class mail. The letter instructs the consignee to locate all affected devices at their facility; contact Synovis MCA to arrange for product return for the firmware upgrade, power button update, and battery replacement, if applicable; complete the enclosed Baxter customer reply form and return it to Baxter by e-mailing it to fca@baxter.com, even if the consignees does not have any inventory; and if the consignee distributes this product to other facilities or departments within their institution, to please forward a copy of this communication to them.
Quantity in Commerce109 units
DistributionDistributed nationwide to AZ, FL, OH, NY, GA, VA, MA, WI, MO, MT, WV, IL, NC, NJ, AR, CA, CO, CT, IA, ID, IN, KS, LA, MD, MI, MN, MS, NE, NH, NM, NV, OK, OR, PA, RI, WY, WA, VT, TX, TN, SC, SD
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.