|
Class 2 Device Recall Philips Respironics V60 Ventilator |
![see related information](../images/related-info.gif) |
Date Initiated by Firm |
October 30, 2020 |
Create Date |
January 22, 2021 |
Recall Status1 |
Completed |
Recall Number |
Z-0888-2021 |
Recall Event ID |
86646 |
510(K)Number |
K082660
|
Product Classification |
Ventilator, continuous, minimal ventilatory support, facility use - Product Code MNT
|
Product |
Philips Respironics V60 Ventilator Part Number 1053616 - Product Usage: intended to be used only with various combinations of Respironics-recommended patient circuits, interfaces (masks), humidifiers, and other accessories. |
Code Information |
Model: V60 Ventilator, Part Number 1053616 V60 Ventilator, English Opt: NONE Serial Numbers: 100087733 100087738 100087932 100087999 100088155 100004031 100004071 100004073 100005244 100005350 100005351 100005599 100006209 100006210 100006217 100006406 100006701 100006702 100006881 100006883 100006886 100006888 100006891 100006929 100006933 100006939 100006940 100006941 100009417 100009501 100009512 100009513 100009549 100009550 100009553 100009554 100009556 100009566 100009679 100009760 100009761 100009762 100009768 100009769 100009770 100009777 100009812 100009816 100009854 100010206 100010242 100010390 100010514 100010515 100010517 100010519 100011308 100011343 100011344 100011468 100011740 100011741 100011744 100011749 100011750 100011751 100011754 100011791 100011794 100011799 100011801 100011820 100011821 100011822 100011823 100011824 100011825 100011826 100011827 100011828 100011829 100011830 100011831 100012694 100012884 100012887 100012890 100012891 100013021 100013023 100013027 100013032 100013034 100013083 100013667 100013669 100013791 100014073 100014088 100014142 100014147 100014153 100014283 100014286 100015076 100015209 100015213 100015281 100015667 100016078 100016597 100016599 100016736 100016746 100016747 100016749 100017103 100017258 100017287 100017293 100017301 100017302 100019586 100020214 100020215 100020217 100020219 100020220 100020227 100020312 100020314 100020316 100020317 100020445 100020885 100021037 100021042 100021874 100021877 100021968 100023238 100024476 100024484 100025989 100026618 100026620 100026805 100026806 100026807 100026816 100027758 100029842 100029849 100030985 100031674 100033326 100033327 100033328 100033329 100033330 100033332 100033333 100033334 100033335 100033336 100033338 100033339 100033340 100033342 100033454 100034684 100034703 100035358 100035365 100035742 100035744 100035755 100037043 100037044 100037049 100037051 100037057 100037059 100037061 100037257 100038935 100038936 100038939 100040185 100042229 100042475 100043439 100043440 100044800 100044801 100047830 100048873 100049578 100049922 100050003 100050633 100050634 100050638 100050891 100052079 100052289 100052290 100052639 100052645 100053142 100053143 100053337 100053356 100053971 100055007 100056058 100056891 100056894 100058686 100058693 100058698 100063885 100064427 100064431 100064439 100064682 100064684 100064784 100064787 100065219 100066227 100071766 100072295 100079339 100079474 100080118 100080120 100080121 100080122 100080123 100080124 100080126 100080127 100081004 100084640 100091078 100095091 100095180 100095333 100095545 100095633 100095634 100096642 100096690 100096788 100096789 100097541 100097922 100097924 100097926 100097927 100104724 201001172 201001189 201001254 201001255 201001256 201001259 201001260 201001682 201001683 201001696 201001699 201001700 201001720 201001729 201001808 201001812 201001814 201001823 201001825 201001826 201001827 201001829 201001830 201001843 201001849 201001873 201001884 201001939 201001944 201002117 201002148 201002184 201002193 201002426 201002427 201002443 201002445 201002448 201002552 201002566 201002574 201002587 201002770 201002778 201002834 201002868 201002873 201002894 201002899 201002900 201003014 201003145 201003149 201003179 201003183 201003185 201003187 201003216 201003219 201003233 201003238 201003263 201003266 201003268 201003269 201003272 201003273 201006557 |
Recalling Firm/ Manufacturer |
Respironics California, LLC 2271 Cosmos Ct Carlsbad CA 92011-1517
|
For Additional Information Contact |
800-722-9377
|
Manufacturer Reason for Recall |
Ventilator may experience potential issues with the backup battery failures: 1) if the battery is allowed to over discharge, or 2) if the ventilator is stored for an extended period of time without being plugged into AC power.
|
FDA Determined Cause 2 |
Device Design |
Action |
On 10/30/2020, the firm sent an "URGENT-MEDICAL DEVICE CORRECTION" Notification Letter via priority mail informing customer that the Ventilators may experience two potential issues with the battery (see table below for additional
information).
1. The ventilator may exhibit a high priority Check Vent: Battery Failed alarm if the battery is allowed to over discharge.
2. The battery may fail to charge if the ventilator is stored for an extended period of time without being plugged into AC power.
If the ventilator exhibits a high priority Battery Failed alarm while connected to
AC power, the V60 ventilator will annunciate an audible and visual alarm. The
ventilator will continue to deliver prescribed therapy; however, the audible and
visual alarm will continue until the device is powered down; therefore, the patient
should be transferred to an alternate source of ventilation as soon as one
becomes available. The hazard associated with transferring a patient to an
alternate source of ventilation is a temporary loss of therapy leading to potential
hypercarbia or hypoxemia (moderate).
If AC power fails and the backup battery is not functioning, an audible and visual
alarm will annunciate for at least two minutes. The ventilator will no longer be
operative resulting in a total loss of therapy. The patient will require an alternate
source of ventilation. The hazard associated with a ventilator shutdown is a total
loss of therapy leading to potential hypercarbia or hypoxemia (severe).
Customers are instructed:
Follow the instructions in the V60 user manual and the instructions below to
reduce any risk associated with potential battery issues.
1) Avoid allowing the ventilator battery to become completely discharged.
Otherwise, the battery may become over-discharged and require long
recharge times of up to 16 hours or more. The over-discharged condition
may permanently damage the battery so that it is unable to recharge. To
prevent the occurrence o |
Quantity in Commerce |
330 systems |
Distribution |
Worldwide distribution - U.S. Nationwide distribution including in the states of AK, AL, AR, AS, AZ, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE,NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Afghanistan, Albania, Algeria, Andorra, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bhutan, Bolivia, Brazil, Bulgaria, Burkina Faso, Cambodia, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Ethiopia, Finland, France, French Guiana, French Polynesia, Germany, Ghana, Gibraltar, Greece, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kazakhstan, Kenya, Latvia, Lebanon, Libya, Luxembourg, Malaysia, Maldives, Mali, Malta, Mexico, Monaco, Mongolia, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Fed., Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, ST. Pier Miquel, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, Venezuela, Vietnam, Wallis Futuna, and Zambia. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
|
510(K) Database |
510(K)s with Product Code = MNT and Original Applicant = RESPIRONICS CALIFORNIA, INC.
|
|
|
|