• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall MedSchenker Smart Transport Medium (STM) System

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall MedSchenker Smart Transport Medium (STM) Systemsee related information
Date Initiated by FirmFebruary 16, 2021
Create DateApril 13, 2021
Recall Status1 Terminated 3 on December 08, 2022
Recall NumberZ-1417-2021
Recall Event ID 87575
Product Classification Culture media, non-selective and non-differential - Product Code JSG
ProductMedSchenker Smart Transport Medium (STM) System: CULTURE MEDIA, NON-SELECTIVE AND NON-DIFFERENTIAL - MULTIPURPOSE CULTURE MEDIUM - Product Usage: intended for the collection and transport of clinical specimens containing viruses, from the collection site to the testing laboratory. I
Code Information Serial #: STM-UNV-0001-Q121, Lot #: JAH161, KAG076
Recalling Firm/
Manufacturer
Vectornate USA Inc
10 Industrial Ave Ste 4
Mahwah NJ 07430-2284
For Additional Information ContactShreya Patil
201-393-2973
Manufacturer Reason
for Recall
Discoloration and turbidity were present in the specified lots of STMs because of Longer Monsoon season at the South Korean Manufacturing Facility.
FDA Determined
Cause 2
Process control
ActionThe firm did not give out a public statement of a recall or recall strategy as it only affected one client and was specific to two identified lots. the firm did notify the customer of the issue via phone and email.
Quantity in Commerce130,000 devices
DistributionUS Nationwide distribution in the state of NJ.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-