| | Class 2 Device Recall Gang Cath Lab Set |  |
| Date Initiated by Firm | March 31, 2021 |
| Date Posted | April 21, 2021 |
| Recall Status1 |
Terminated 3 on March 20, 2023 |
| Recall Number | Z-1474-2021 |
| Recall Event ID |
87647 |
| 510(K)Number | K961404 |
| Product Classification |
Filter, infusion line - Product Code FPB
|
| Product | Gang Cath Lab Set, Model Number M20858R1.for the administration of IV fluids. Stopcocks and manifolds are used to control the direction of IV fluid flow. |
| Code Information |
Lot Numbers: 4007165 |
| FEI Number |
3012307300
|
Recalling Firm/ Manufacturer |
Smiths Medical ASD Inc. 6000 Nathan Ln N Minneapolis MN 55442-1690
|
| For Additional Information Contact | Dave Halverson 763-383-3072 |
Manufacturer Reason for Recall | Specific lots of Medex High Pressure Adaptors, Rotators, Manifolds, Stopcocks and Tubing and Medex LogiCal Kits may have been manufactured with insufficient or incomplete welds. These are components for various cardiac kits including pressure monitoring devices and interventional imaging devices.
|
FDA Determined Cause 2 | Process control |
| Action | The firm initiated the action by email on 03/31/2021. The Recall Notice instructs consignees to identify, and quarantine affected product in their possession. A list of affected product has been included on the Recall Notice for consignees to identify affected product. The consignees have been instructed to complete the response form, and return it to fieldactions@smiths-medical.com within 10 days of receipt. After the completed Recall Response Form has been submitted to fieldactions@smiths-medical.com, the consignee will be contacted to arrange return of any affected product in their possession. Distributors have been instructed that if they have distributed potentially affected product to their customers to immediately notify them of the Recall. |
| Quantity in Commerce | 400 devices |
| Distribution | Distribution US nationwide including Puerto Rico, Austria, Canada, Chile, Ecuador, Hong Kong, Mexico, Slovenia, Brazil, South Africa, Colombia, and Japan. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = FPB
|
|
|
|