| | Class 2 Device Recall Micross |  |
| Date Initiated by Firm | April 26, 2021 |
| Create Date | June 02, 2021 |
| Recall Status1 |
Terminated 3 on September 12, 2024 |
| Recall Number | Z-1777-2021 |
| Recall Event ID |
87866 |
| Product Classification |
File, bone, surgical - Product Code EMI
|
| Product | REF 4049 MICROSS, CE0123, STERILE EO. Used in oral surgery. |
| Code Information |
Lot number: 2-01121, 23-28920, 21-26520, 18-24819, 24-29118, 21-24818, 13-12818, 26-33118 |
Recalling Firm/ Manufacturer |
Meta C.G.M. Spa Via Modena 22-24 Correggio Italy
|
Manufacturer Reason for Recall | Sterility issues; single use devices labeled as sterile may not have been adequately sterilized |
FDA Determined Cause 2 | Vendor change control |
| Action | On April 26, 2021 C.G.M. S.p.A. sent a "Urgent Field Safety Notice" to all affected consignees via email. On May 10, 2021, the firm followed up this communication with an "Urgent: Medical Device Recall" notification to all affected consignee. In addition to informing the customers about the recall, the firm asked consignees to take the following actions:
1. Stop the use of all devices in your possession that have the lot number listed in Attachment 1
2. Circulate this medical device Recall Notice to everyone in your organization who need to be informed;
3. Identify and segregate all items listed in Attachment 01, if still available on the premises.
4. Fill in the Medical device Recall Return REsponse (acknowledgement and receipt form), including the number of received devices, used or sold devices, remaining and segregated devices
5. If the product has been distributed, identify these facilities and immediately submit this notification.
6. Within 20 working days from receiving the official notification, return involved items to the distributor.
7. Within 30 working days from receiving the official notification, the distributor shall return involved items to manufacturer C.g.M. spa divisione medicale META
8. Contact Information for Distributor: C.G.M. spa divisione medicale META
Days/Hours Available for call: Monday -Friday, 9:00 AM to 4:30 PM, European Time (UTC/GMT +2 h)
Email Address:helpdesk@metahosp.com
9. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch adverse Event Reporting program either online, by regular mail or by fax. |
| Quantity in Commerce | 1,692 |
| Distribution | Distribution in US states of Texas and Florida
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|