Class 2 Device Recall VENOVO Venous Stent System |
|
Date Initiated by Firm |
December 14, 2020 |
Create Date |
June 17, 2021 |
Recall Status1 |
Open3, Classified |
Recall Number |
Z-1898-2021 |
Recall Event ID |
87208 |
PMA Number |
P180037 |
Product Classification |
Stent, iliac vein - Product Code QAN
|
Product |
VENOVO Venous Stent System 8F
Product Description(REF)/dimension: VENEL10040 /Venovo 8F 10/40/1200mm OUS; VENEL10060 /Venovo 8F 10/60/1200mm OUS; VENEL10080 /Venovo 8F 10/80/1200mm OUS; VENEL10100 /Venovo 8F 10/100/1200mm OUS; VENEL10120 /Venovo 8F 10/120/1200mm OUS; VENEL10140 /Venovo 8F 10/140/1200mm OUS; VENEL10160 /Venovo 8F10/160/1200mm OUS; VENEL12040 /Venovo 8F 12/40/1200mm OUS; VENEL12060 /Venovo 8F 12/60/1200mm OUS; VENEL12080 /Venovo 8F 12/80/1200mm OUS; VENEL12100 /Venovo 8F 12/100/1200mm OUS; VENEL12120 /Venovo 8F 12/120/1200mm OUS; VENEL12140 /Venovo 8F 12/140/1200mm OUS; VENEL12160 /Venovo 8F 12/160/1200mm OUS; VENEM10040 /Venovo 8F 10/40/800mm OUS; VENEM10060 /Venovo 8F 10/60/800mm OUS; VENEM10080 /Venovo 8F 10/80/800mm OUS; VENEM10100 /Venovo 8F 10/100/800mm OUS; VENEM10120 /Venovo 8F 10/120/800mm OUS; VENEM10140 /Venovo 8F 10/140/800mm OUS; VENEM10160 /Venovo 8F 10/160/800mm OUS; VENEM12040 /Venovo 8F 12/40/800mm OUS; VENEM12060 /Venovo 8F 12/60/800mm OUS; VENEM12080 /Venovo 8F 12/80/800mm OUS; VENEM12100 /Venovo 8F 12/100/800mm OUS; VENEM12120 /Venovo 8F 12/120/800mm OUS; VENEM12140 /Venovo 8F 12/140/800mm OUS; VENEM12160 /Venovo 8F 12/160/800mm OUS; VENUL10040 /Venovo 8F 10/40/1200mm US; VENUL10060 /Venovo 8F 10/60/1200mm US; VENUL10080 /Venovo 8F 10/80/1200mm US; VENUL10100 /Venovo 8F 10/100/1200mm US; VENUL10120 /Venovo 8F 10/120/1200mm US; VENUL10140 /Venovo 8F 10/140/1200mm US; VENUL10160 /Venovo 8F 10/160/1200mm US; VENUL12040 /Venovo 8F 12/40/1200mm US; VENUL12060 /Venovo 8F 12/60/1200mm US; VENUL12080 /Venovo 8F 12/80/1200mm US; VENUL12100 /Venovo 8F 12/100/1200mm US; VENUL12120 /Venovo 8F 12/120/1200mm US; VENUL12140 /Venovo 8F 12/140/1200mm US; VENUL12160 /Venovo 8F 12/160/1200mm US; VENUM10040 /Venovo 8F 10/40/800mm US; VENUM10060 /Venovo 8F 10/60/800mm US; VENUM10080 /Venovo 8F 10/80/800mm US; VENUM10100 /Venovo 8F 10/100/800mm US; VENUM10120 /Venovo 8F 10/120/800mm US; VENUM10140 /Venovo 8F 10/140/800mm US; VENUM10160 /Venovo 8F 10/160/800mm US; VENUM12040 /Venovo 8F 12/40/800mm US; VENUM12060 /Venovo 8F 12/60/800mm US; VENUM12080 /Venovo 8F 12/80/800mm US; VENUM12100 /Venovo 8F 12/100/800mm US; VENUM12120 /Venovo 8F 12/120/800mm US; VENUM12140 /Venovo 8F 12/140/800mm US; VENUM12160 /Venovo 8F 12/160/800mm US;
|
Code Information |
Product Code/GTIN: VENEL10040 /00801741102004; VENEL10060 /00801741104664; VENEL10080 /00801741104671; VENEL10100 /00801741102035; VENEL10120 /00801741102042; VENEL10140 /00801741102059; VENEL10160 /00801741102066; VENEL12040 /00801741102141; VENEL12060 /00801741102158; VENEL12080 /00801741102172; VENEL12100 /00801741102189; VENEL12120 /00801741102196; VENEL12140 /00801741102202; VENEL12160 /00801741102219; VENEM10040 /00801741101939; VENEM10060 /00801741101946; VENEM10080 /00801741101953; VENEM10100 /00801741101960; VENEM10120 /00801741101977; VENEM10140 /00801741101984; VENEM10160 /00801741101991; VENEM12040 /00801741102073; VENEM12060 /00801741102080; VENEM12080 /00801741102097; VENEM12100 /00801741102103; VENEM12120 /00801741102110; VENEM12140 /00801741102127; VENEM12160 /00801741102134; VENUL10040 /00801741103209; VENUL10060 /00801741103216; VENUL10080 /00801741103223; VENUL10100 /00801741103230; VENUL10120 /00801741103247; VENUL10140 /00801741103254; VENUL10160 /00801741103261; VENUL12040 /00801741103346; VENUL12060 /00801741103353; VENUL12080 /00801741103360; VENUL12100 /00801741103377; VENUL12120 /00801741103384; VENUL12140 /00801741103391; VENUL12160 /00801741103407; VENUM10040 /00801741103131; VENUM10060 /00801741103148; VENUM10080 /00801741103155; VENUM10100 /00801741103162; VENUM10120 /00801741103179; VENUM10140 /00801741103186; VENUM10160 /00801741103193; VENUM12040 /00801741103278; VENUM12060 /00801741103285; VENUM12080 /00801741103292; VENUM12100 /00801741103308; VENUM12120 /00801741103315; VENUM12140 /00801741103322; VENUM12160 /00801741103339; U.S. Device; Serial Numbers: ANEQ4079; ANEQ4090; ANEQ4091; ANEQ4101; ANEQ4104; ANEQ4105; ANEQ4109; ANEQ4978; ANEQ4979; ANEQ4980; ANEQ4996; ANEQ4997; ANEQ5023; ANER0477; ANER0478; ANER0490; ANER0491; ANER0510; ANER0515; ANER0532; ANER1502; ANER1528; ANER1529; ANER1530; ANER1533; ANER2795; ANER2796; ANER2797; ANER2812; ANER2813; ANER2814; ANER2815; ANER2816; ANER2838; ANER2841; ANER2855; ANER4217; ANER4228; ANER4236; ANER4238; ANER4239; ANER5126; ANER5127; ANER5136; ANER5141; ANER5142; ANER5153; ANER5159; ANER5160; ANER5169; ANER5175; ANES2223; ANES2224; ANES2233; ANES2234; ANES2235; ANES2236; ANES3045; ANES3050; ANES3067; ANES4102; ANES4103; ANES4111; ANES4112; ANES4113; ANES4114; ANES4115; ANES4116; ANES4117; ANES4118; ANET0561; ANET0566; ANET0567; ANET0568; ANET1490; ANET1508; ANET1509; ANET2533; ANET2534; ANET2535; ANET2536; ANET2537; ANET2540; ANET2541; ANET3797; ANET3798; ANET3799; ANET3800; ANET3814; ANET3815; ANET3816; ANET3817; ANET3818; ANET3877; ANET3878; ANEU0748; ANEU1594; ANEU1601; ANEU1602; ANEU1603; ANEU1604; ANEU1605; ANEU1606; ANEU2442; ANEU3275; ANEU3276; ANEU3277; ANEU3278; ANEU3284; ANEU3285; ANEU3286; ANEU3287; ANEU3288; ANEU3289; ANEU3290; ANEU3306; ANEU4231; ANEU4232; ANEU4238; ANEU4239; ANEU4240; ANEU4246; ANEU4253; ANEU4256; ANEU4260; ANEU4261; ANEU4264; ANEU4265; ANEV0586; ANEV0587; ANEV0591; ANEV0594; ANEV0598; ANEV0599; ANEV0603; ANEV0606; ANEV0611; ANEV0614; ANEV1714; ANEV1715; ANEV1716; ANEV1717; ANEV1718; ANEV1737; ANEV1738; ANEV1739; ANEV1740; ANEV3078; ANEV3079; ANEV3080; ANEV3081; ANEV3082; ANEV3494; ANEV4019; ANEV4020; ANEV4021; ANEV4022; ANEV4023; ANEV4024; ANEV4025; ANEV4043; ANEV4044; ANEV4045; ANEV4046; ANEV4071; ANEV4076; ANEW0553; ANEW0554; ANEW0560; ANEW0561; ANEW0562; ANEW0563; ANEW0564; ANEW0565; ANEW0566; ANEW0599; ANEW0600; ANEW2005; ANEW2006; ANEW2007; ANEW2028; ANEW2029; ANEW2030; ANEW2031; ANEW2032; ANEW2033; ANEW2034; ANEW2239; ANEW2772; ANEW2773; ANEW3378; ANEW3379; ANEW3380; ANEW3381; ANEW3394; ANEW3395; ANEW3396; ANEW3397; ANEW3422; ANEW4229; ANEW4230; ANEW4231; ANEW4232; ANEW4233; ANEW4242; ANEW4243; ANEW4244; ANEW4245; ANEW4246; ANEW4247; ANEW4276; ANEW4280; ANEW4282; ANEW4291; ANEW4297; ANEW4298; ANEW4300; ANEX0704; ANEX0711; ANEX0713; ANEX0720; ANEX0729; ANEX0732; ANEX0735; ANEX0736; ANEX0739; ANEX0740; ANEX0741; ANEX0742; ANEX0743; ANEX0744; ANEX0745; ANEX1737; ANEX1743; ANEX1744; ANEX1746; ANEX1747; ANEX1750; ANEX1752; ANEX1753; ANEX1754; ANEX1755; ANEX3120; ANEX3121; ANEX3123; ANEX3131; ANEX3135; ANEX3136; ANEX3137; ANEX3144; ANEX3145; ANEX3147; ANEX3148; ANEX3149; ANEX3151; ANEX3153; ANEX3154; ANEX3156; ANEX4737; ANEX4740; ANEX4744; ANEX4745; ANEX4751; ANEX4753; ANEX4761; ANEX4762; ANEX4764; ANEX4773; ANEX4779; ANEX4780; ANEX4782; ANEX4783; ANEX4784; ANEX4785; ANEX4786; ANEX4787; ANEX4788; ANEX4789; ANEX6042; ANEX6046; ANEX6049; ANEX6056; ANEX6057; ANEX6065; ANEX6066; ANEX6068; ANEY1246; ANEY1247; ANEY1251; ANEY1257; ANEY1258; ANEY1259; ANEY1262; ANEY1263; ANEY1264; ANEY1265; ANEY1266; ANEY1267; ANEY1268; ANEY1269; ANEY1278; ANEY1279; ANEY1280; ANEY2049; ANEY2050; ANEY2051; ANEY2052; ANEY2053; ANEY2054; ANEY2055; ANEY2056; ANEY3241; ANEY3243; ANEY3245; ANEY3255; ANEY3256; ANEY3257; ANEY3258; ANEY3261; ANEY3537; ANEY3538; ANEY4346; ANEY4354; ANEY4355; ANEY4356; ANEY4357; ANEY4358; ANEY4562; ANEZ0957; ANEZ0958; ANEZ0959; ANEZ0967; ANEZ0972; ANEZ0973; ANEZ0974; ANEZ0980; ANEZ0996; ANEZ0997; ANEZ1000; ANEZ1001; ANEZ1002; ANEZ1003; ANEZ1004; ANEZ2107; ANEZ2112; ANEZ2113; ANEZ2121; ANEZ2122; ANEZ2123; ANEZ2124; ANEZ2137; ANEZ2138; ANEZ2139; ANEZ2140; ANEZ3453; ANEZ3454; ANEZ3455; ANEZ3502; ANEZ3939; ANEZ3949; ANFN1482; ANFN1483; ANFN1484; ANFN1485; ANFN1486; ANFN1487; ANFN1488; ANFN1490; ANFN1492; ANFN1493; ANFN2809; ANFN2817; ANFN2821; ANFN2822; ANFN2824; ANFN2825; ANFN2835; ANFN2836; ANFN2837; ANFN2838; ANFN2839; ANFN2840; ANFN2841; ANFN2842; ANFN2843; ANFN3989; ANFN3990; ANFN3991; ANFN3992; ANFN3993; ANFN3994; ANFN3999; ANFN4000; ANFN4001; ANFN4002; ANFP0891; ANFP0896; ANFP0897; ANFP0915; ANFP0923; ANDS0730; ANDS0722; ANDS0729; ANDS1275; ANDS1312; ANDS1302; ANDS2454; ANDS4842; ANDS4851; ANDS4854; ANDS4852; ANDT1073; ANDT1713; ANDU2241; ANDT1721; ANDS4853; ANDT2908; ANDT2907; ANDT2906; ANDT2905; ANDU0121; ANDU1126; ANDU1125; ANDU0123; ANDU0122; ANDV0714; ANDU1894; ANDU1906; ANDU1861; ANDU1862; ANDU1890; ANDU1856; ANDU1884; ANDU1873; ANDU1124; ANDU1114; ANDU1905; ANDU1878; ANDU1867; ANDU3084; ANDU3144; ANDU3145; ANDU3139; ANDU3140; ANDU3104; ANDU3098; ANDU3128; ANDU3103; ANDU3127; ANDU3089; ANDV3781; ANDU3090; ANDV1215; ANDV1227; ANDU4041; ANDV1261; ANDU3111; ANDV3603; ANDV3602; ANDV3606; ANDV3576; ANDV3604; ANDV3601; ANDV4566; ANDV4554; ANDV4551; ANDV4567; ANDV4546; ANDV4552; ANDV4549; ANDV4550; ANDV4553; ANDV4568; ANDV4545; ANDV3600; ANDV4555; ANDV1234; ANDV3605; ANDV4547; ANDW2044; ANDW2042; ANDW2028; ANDW2043; ANDW2046; ANDW2047; ANDW2045; ANDX1371; ANDX1388; ANDX1429; ANDX1428; ANDW3626; ANDX1427; ANDX1423; ANDX1426; ANDV4569; ANDX1425; ANDW3625; ANDW4360; ANDW4367; ANDW4383; ANDX1430; ANDW4369; ANDW4400; ANDX1424; ANDW4340; ANDX0708; ANDW4419; ANDX0725; ANDX0720; ANDW4402; ANDW4353; ANDX2191; ANDX2172; ANDX2171; ANDX2192; ANDX2190; ANDX3082; ANDX3103; ANDX3105; ANDX3102; ANDX3101; ANDX3086; ANDX3081; ANDX2173; ANDX3100; ANDX3085; ANDX3083; ANDX2193; ANDX3104; ANDX3099; ANDX3084; ANDX3106; ANDX3107; ANDX5400; ANDX5399; ANDY1247; ANDZ0784; ANDY2488; ANDY2486; ANDY2487; ANDY2470; ANDZ2797; ANDZ2796; ANDZ1376; ANDZ2803; ANDY3323; ANDY3310; ANDZ1255; ANDY3311; ANDY3324; ANDZ1243; ANDZ1254; ANDZ1372; ANDZ1357; ANDZ1373; ANDY3325; ANDZ1358; ANDZ1375; ANDY3326; ANDZ1256; ANDZ1244; ANDZ1377; ANDZ1258; ANDZ1374; ANDZ2326; ANDZ1257; ANDZ2325; ANDZ3877; ANDZ3844; ANDZ3870; ANDZ3861; ANEN1090; ANEN1107; ANEN1112; ANEN1088;
ANEN1111; ANEN1093; ANEN1108; ANEN1109; ANEN1094; ANEN1089; ANEN1110; ANEN2194; ANEN2193; ANEN2189; ANEN2190; ANEN2185; ANEN2182; ANEN2186; ANEN2187; ANEN3444; ANEN3421; ANEN3438; ANEN2188; ANEN3430; ANEN3436; ANEN3422; ANEN3441; ANEN3433; ANEN2180; ANEN3423; ANEN3431; ANEN4928; ANEN4916; ANEP0914; ANEP0919; ANEP0915; ANEP0912; ANEP0906; ANEP0918; ANEP2441; ANEP2455; ANEP2450; ANEP2442; ANEP2461; ANEQ3077; ANEP2443; ANEP3957; ANEP3963; ANEP3944; ANEP3974; ANEP3967; ANEP3984; ANEP3969; ANEP3942; ANEP3949; ANEQ3078; ANEP3985; ANEP5356; ANEP5334; ANEP5340; ANEP5386; ANEP5359; ANEP5358; ANEP5338; ANEP5365; ANEP5380; ANEP5385; ANEP5361; ANEP5374; ANEP5341; ANEP5335; ANEP5352; ANEQ1378; ANEQ1397; ANEQ1379; ANEQ1399; ANEQ1398; ANEQ1376; ANEQ1377; ANEQ1401; ANEQ1396; ANEQ1380; ANEQ1375; ANEQ2773; ANEQ2774; ANEQ1400; ANEQ2776; ANEQ2775; ANEQ2756; ANEQ2757; ANEQ2755; ANEQ4078; ANFQ2793; O.U.S. Device; Serial Numbers: ANDU2240; ANES2208; ANDU2240; ANDY3304; ANEP2429; ANEP5328; ANEP2429; ANDZ3827; ANDN0820; ANDU2240; ANEP2429; ANDU2240; ANEP2429; ANEP2429; ANDU2240; ANEP5328; ANEP2429; ANES1042; ANDZ2288; ANEP5328; ANDU2240; ANDY3304; ANDZ3827; ANEP2429; ANDN0820; ANES1042; ANDZ2288; ANEP2429; ANEP5328; ANEP2429; ANEP5328; ANES1042; ANDU2240; ANDY3304; ANDZ2288; ANDZ3827; ANDZ3827; ANCV0493; ANDU2240; ANDU2240; ANES1042; ANES1042; ANEP5328; ANEP2429; ANEP2429; ANDN0820; ANDU2240; ANEP5328; ANDY3304; ANDN0820; ANDZ2288; ANDN0820; ANEP5328; ANCV0493; ANDZ3827; ANEP2429; ANEP5328; ANES2208; ANEP5328; ANES1042; ANES2208; ANEY3214; ANEY4323; ANFP2666; ANFP2666; ANDU2240; ANEY3214; ANEY3214; ANEY3214; ANEY3214; ANEP5328; ANEP2429; ANDR1257; ANDR2270; ANDR2270; ANDU1102; ANDV1222; ANDY2458; ANDY2458; ANDY3301; ANDY3301; ANDZ1338; ANDZ3819; ANEN1079; ANEN3405; ANER5167; ANEX0679; ANEZ2095; ANEZ3410; ANFN3972; ANCR2676; ANDR2270; ANDV1222; ANDV2517; ANDY2458; ANEZ3410; ANEZ3410; ANEZ2095; ANDX5352; ANEQ4065; ANEQ4961; ANEN3405; ANDV1222; ANDX5352; ANDZ1338; ANDV2517; ANDV1222; ANFN3972; ANEN3405; ANFN3972; ANFN3972; ANFN3972; ANEN3405; ANDV2517; ANDY2458; ANDR2270; ANEQ4065; ANDV2517; ANEQ4065; ANEN3405; ANEQ4961; ANFN3972; ANDV2517; ANEZ2095; ANCV1579; ANEN3405; ANDR2270; ANEN3405; ANDX5352; ANDZ3819; ANEN1079; ANEQ4961; ANER5167; ANEX0679; ANEX4720; ANEY4324; ANEX4720; ANEY4324; ANEZ0921; ANEZ0921; ANEZ2095; ANEZ2095; ANEZ3410; ANDR2270; ANEQ4065; ANEQ4961; ANDR2270; ANEQ4961; ANEQ4961; ANEQ4961; ANEQ4065; ANEQ4065; ANDR1257; ANCV1579; ANDZ1338; ANEQ4065; ANER5167; ANDR1257; ANDV1222; ANDV2517; ANDX5352; ANDX5352; ANDY2458; ANDR1257; ANDR1257; ANDR1257; ANDZ1338; ANDZ1338; ANCV1579; ANDY2458; ANEQ4065; ANFN3972; ANEX4720; ANEZ0921; ANFN3972; ANEQ4961; ANDV2517; ANDY2458; ANDQ3035; ANDQ3035; ANDQ3035; ANDQ3877; ANDU1105; ANDW4412; ANDX0715; ANDX5308; ANDZ2292; ANDZ3826; ANEN1080; ANEQ4066; ANES2209; ANET2531; ANET2531; ANEX1699; ANEX1699; ANEY4325; ANEZ0982; ANBY2401; ANCT0397; ANCT0397; ANDZ1330; ANES2209; ANDX5308; ANEY4325; ANDW4412; ANCW4225; ANDV0287; ANEQ2730; ANEQ2730; ANEQ2730; ANEQ4066; ANEX1699; ANEZ0982; ANEX1699; ANFP0878; ANFP0878; ANET2531; ANDQ3035; ANDQ3035; ANDQ3877; ANEQ4066; ANET2531; ANEN1080; ANEY4325; ANDX5308; ANDX5308; ANDQ3035; ANEQ2730; ANCW3270; ANES2209; ANDX5308; ANDZ1330; ANDZ2292; ANDZ3826; ANEN1080; ANEX1699; ANEY3212; ANEY4325; ANEZ0982; ANFP0878; ANET2531; ANEX1699; ANEX1699; ANEY3212; ANEY4325; ANEZ0982; ANDQ3035; ANDQ3035; ANEZ0982; ANEY4325; ANCW3270; ANCW4225; ANEY4325; ANET2531; ANES2209; ANET2531; ANDQ3035; ANDQ3877; ANDZ3826; ANDQ3877; ANES2209; ANET2531; ANDV0287; ANDX0715; ANDX5308; ANDZ1330; ANDZ2292; ANDQ3877; ANDX5308; ANEN1080; ANEY4325; ANDQ3877; ANEQ4066; ANEZ0982; ANCW3270; ANEQ4066; ANES2209; ANEQ2730; ANEY4325; ANDQ3041; ANCU2726; ANCX3902; ANDQ3880; ANCX3902; ANDQ3041; ANDQ3041; ANEP5323; ANEP5323; ANEQ2731; ANEP5323; ANEP5323; ANEQ4067; ANDQ3880; ANDY3305; ANDQ3880; ANCX3902; ANDY3305; ANEP0856; ANEP0856; ANEP0856; ANEP5323; ANEQ2731; ANES2210; ANES2210; ANEZ0922; ANFN3973; ANFN3973; ANEZ0922; ANFN1442; ANFN3973; ANFN3973; ANDQ3041; ANCX3902; ANDY3305; ANDY3305; ANEQ2731; ANDY3305; ANEP5323; ANDQ3880; ANDY3305; ANCX3902; ANDQ3041; ANDQ3041; ANDY3305; ANDQ3880; ANDQ3880; ANDQ3880; ANEQ2731; ANER0516; ANCX3902; ANDQ3880; ANEP5323; ANEP5323; ANEP5323; ANDQ3880; ANEP0856; ANEP5323; ANEZ0922; ANFN3973; ANDQ3023; ANDT1058; ANDT1058; ANDT1058; ANDU4075; ANDU4075; ANDZ1348; ANEN1081; ANEN1081; ANEP0857; ANEP0857; ANEP3924; ANET3794; ANEX0680; ANEX0680; ANEX0680; ANEZ2096; ANEP3924; ANEZ2096; ANEP3924; ANEP3924; ANDQ3023; ANDQ3023; ANEP3924; ANEP3924; ANDU4075; ANDQ3023; ANDU4075; ANDZ1348; ANCU3706; ANDZ2290; ANCU3706; ANDZ2290; ANDZ2290; ANDZ2290; ANDZ2296; ANEP5332; ANDZ2296; ANCU3716; ANEP5332; ANDZ2296; ANDZ2296; ANDR1253; ANDU2239; ANEQ2716; ANDU2239; ANES2187; ANCU0847; ANDX5374; ANDW4406; ANDW4406; ANES2187; ANDU2239; ANCU0847; ANDR1253; ANEQ2716; ANDR1253; ANDW4406; ANES2187; ANCY3323; ANDU2239; ANDU2239; ANDW4406; ANDU2239; ANES2187; ANDW4406; ANEP3938; ANDQ3029; ANDQ3029; ANDW4406; ANDX5374; ANDQ3029; ANDR1253; ANDR1253; ANDX5374; ANES2187; ANDR1253; ANDR1253; ANEQ2716; ANDW4406; ANCU0847; ANCU0847; ANDR1253; ANDU2239; ANDU2239; ANDW4406; ANDW4406; ANEQ2716; ANCU0847; ANCY3323; ANDR1253; ANDR1253; ANDX5374; ANDX5374; ANDU2239; ANEQ2716; ANES2187; ANDR1253; ANDU2239; ANDW4406; ANEP3938; ANDR1253; ANDU2239; ANDQ3037; ANDQ3037; ANDT1057; ANDT1057; ANDW4338; ANDW4338; ANDX5309; ANDX5309; ANDZ2308; ANEP0852; ANEP0852; ANER5149; ANET3785; ANEZ0906; ANDT1057; ANDQ3037; ANDX5309; ANEQ4945; ANDT1057; ANDT1057; ANDX5309; ANCW3260; ANDT1057; ANEP3936; ANEP3936; ANDX5309; ANEQ4945; ANDT1057; ANEX3092; ANEQ4036; ANCY3332; ANDU4074; ANDX5309; ANDZ2308; ANEQ4036; ANEZ0906; ANDZ2308; ANEP3936; ANDX5309; ANEP0852; ANEQ4945; ANEQ4945; ANET3785; ANEZ0906; ANFP2368; ANEQ4945; ANET3785; ANEX0671; ANEX0671; ANEX3092; ANEX3092; ANEZ0906; ANDQ3037; ANDX5309; ANDQ3037; ANDT1057; ANDX5309; ANDX5309; ANDZ2308; ANEP3936; ANEP3936; ANEP3936; ANEQ4036; ANEQ4036; ANEQ4036; ANDT1057; ANDT1057; ANEZ0906; ANEQ4945; ANDT1057; ANEQ4036; ANDW4338; ANCR1284; ANDZ2308; ANER5149; ANET3785; ANEX0671; ANER5149; ANDX5309; ANEX3092; ANDX5309; ANER5149; ANEP3936; ANDQ3037; ANEQ4945; ANEQ4945; ANDX5309; ANDU4074; ANEZ3430; ANEZ0906; ANCY3303; ANDP3952; ANDP3952; ANDP3952; ANDU4027; ANDU4027; ANDV4526; ANDW4339; ANDW4339; ANDX5324; ANDX5385; ANDZ3818; ANDZ3818; ANEN1086; ANEQ2717; ANEQ4946; ANEQ4946; ANER5165; ANET2530; ANEU0741; ANEU3303; ANEX0749; ANEZ0907; ANEZ0907; ANCQ1427; ANCQ1427; ANET2530; ANFN3944; ANDZ3818; ANEQ2717; ANEQ4037; ANCQ1427; ANEQ2717; ANEQ2717; ANER5165; ANCQ1427; ANDW4339; ANDX5385; ANCQ1427; ANCU1844; ANEX0749; ANDZ3818; ANEQ2717; ANEQ2717; ANEQ2717; ANET2530; ANFN3944; ANEU0741; ANEU3303; ANEX0749; ANEX0749; ANEZ0907; ANDP3952; ANDQ3871; ANDQ3871; ANDX5324; ANDX5324; ANDX5324; ANDX5385; ANDZ3818; ANEQ4037; ANEQ4037; ANER5165; ANER5165; ANEZ0907; ANFN3944; ANCY3303; ANEZ0907; ANEU0741; ANEU3303; ANDU4027; ANDX5324; ANDX5324; ANEN1086; ANDX5385; ANDZ3818; ANDX5385; ANEQ2717; ANET2530; ANEQ4037; ANDX5324; ANEQ4037; ANEQ2717; ANER5165; ANFN3944; ANER5165; ANEX0749; ANDX5324; ANDZ3818; ANEQ2717; ANEQ4037; ANDT0505; ANDT0505; ANDT1054; ANDT1054; ANDY2469; ANDY2469; ANDY4963; ANDY4963; ANDY4963; ANDY4963; ANDZ1345; ANEW2047; ANEW4217; ANEY4322; ANEY4322; ANFN1431; ANDT1054; ANEP2417; ANDZ1345; ANDT1054; ANDT1054; ANDT0505; ANDT1054; ANDT1054; ANDT1054; ANDY2469; ANEP2417; ANDY4963; ANEP0862; ANDY2469; ANEP2417; ANEW2047; ANEW4217; ANEY3209; ANEY3209; ANEY4322; ANDN0807; ANDQ0320; ANDQ1722; ANDQ3030; ANDQ3030; ANDT0505; ANDT0505; ANDT1054; ANDT1054; ANDT1054; ANDY2469; ANDY4963; ANEP0862; ANEP2417; ANDT1054; ANDT1054; ANEY4322; ANDT1054; ANEP0862; ANEP0862; ANEW4217; ANDT0505; ANDT1054; ANCU3703; ANDT1054; ANDY4963; ANDY4963; ANEP0862; ANFN1431; ANDT0505; ANCR2675; ANDQ0320; ANDQ1722; ANDT0505; ANEP0862; ANEP0862; ANEP0862; ANEP2417; ANFN1431; ANDN0808; ANDQ1726; ANEQ1347; ANEQ1347; ANDQ3031; ANEQ1347; ANES2188; ANDQ1726; ANDQ3031; ANDN0808; ANDQ1726; ANDQ3031; ANEQ1347; ANEQ1347; ANEQ1347; ANDQ3031; ANDQ3031; ANDQ3031; ANEQ1347; ANEQ1347; ANDQ3031; ANDQ3031; ANDN0808; ANES2188; ANDQ1726; ANEQ1347; ANDU1108; ANDY3280; ANCY1571; ANDU1108; ANEN3385; ANEN3385; ANEQ1348; ANEN3385; ANCY1571; ANEN3385; ANCY1571; ANCY1571; ANCY1571; ANEN3385; ANDU1108; ANEN3385; ANDW1996; ANCW3261; ANET1855; ANDN0606; ANDR1266; ANDR1266; ANDW1996; ANDW1996; ANDW1996; ANDY2448; ANDW1996; ANET1855; ANDW1996; ANDY2448; ANCW3261; ANCW3261; ANDY2448; ANET1855; ANDY2448; ANDS4772; ANDS4772; ANDS4772; ANDS4772; ANDT0508; ANDW4345; ANDY4968; ANDY4968; ANDY4968; ANEQ2732; ANEZ3411; ANDQ3024; ANDR1268; ANDS4772; ANDT0508; ANDX5315; ANDX5384; ANDY4968; ANDZ3817; ANDZ3817; ANEN4877; ANEN4877; ANEP0861; ANER5152; ANET0555; ANET3795; ANEW3410; ANEZ3411; ANFN3974; ANDT0508; ANCT4323; ANCY2200; ANER5152; ANDZ3817; ANDX5384; ANEQ2732; ANET0555; ANFN1443; ANFN3974; ANDT0508; ANEQ2732; ANCS1161; ANDS4772; ANEW3410; ANEY1223; ANDS4772; ANDX5315; ANEQ2732; ANDT0508; ANDZ3817; ANDX5315; ANEQ2732; ANEQ2732; ANDX5384; ANDX5384; ANDX5315; ANFN1443; ANEQ2732; ANDZ3817; ANEP0861; ANEQ2732; ANEW3410; ANEY1223; ANFP0911; ANDZ3817; ANFP2377; ANFP0911; ANEQ2732; ANER5152; ANEQ2732; ANDZ3817; ANDY4968; ANEQ2732; ANCY3322; ANDS4772; ANDQ3024; ANEQ2732; ANET0555; ANER5152; ANER5152; ANEY1223; ANFP0911; ANDQ3024; ANER5152; ANEW3410; ANEQ2732; ANDX5384; ANER5152; ANEY1223; ANER5152; ANCS1161; ANEY1223; ANDQ3024; ANDS4772; ANDX5315; ANEQ2732; ANEQ2732; ANEQ2732; ANER5152; ANEW3410; ANEQ2732; ANFN3974; ANFP2377; ANDX5315; ANFP0911; ANDX5384; ANEQ2732; ANEP0861; ANDZ3817; ANFN1443; ANDQ3024; ANDQ3024; ANDR1268; ANDT0508; ANDT0508; ANDX5315; ANDX5315; ANDX5384; ANEN4877; ANER5152; ANET0555; ANEY1223; ANEZ3411; ANEZ3411; ANFN1443; ANFP0911; ANDQ1716; ANDX5384; ANDX5384; ANET0555; ANET0555; ANET3795; ANEY1223; ANEZ3411; ANDT0508; ANDT0508; ANER5152; ANEN4877; ANEQ2732; ANEQ2732; ANEW3410; ANEZ3411; ANCS1161; ANDS4772; ANDX5315; ANDX5384; ANER5152; ANFN1443; ANCY3322; ANEQ2732; ANEQ2732; ANER5152; ANER5152; ANDX5315; ANEW3410; ANDY4968; ANDR1268; ANDW4345; ANDX5315; ANDX5384; ANDX5384; ANDY4968; ANDZ2301; ANDZ3817; ANDS4772; ANFP2377; ANDX5315; ANCY2200; ANER5152; ANEY1223; ANFN1443; ANDT0508; ANDZ3817; ANEW3410; ANEZ3411; ANEP0861; ANDX5384; ANDX5384; ANDQ3024; ANEP0861; ANDQ1716; ANEQ2732; ANFP2377; ANEQ2732; ANEW3410; ANDS3973; ANDS3973; ANDS4774; ANDS4774; ANDS4774; ANDS4774; ANDT2946; ANDW1997; ANDW4361; ANDX0760; ANDX0760; ANDX0760; ANDY1234; ANEP2420; ANDQ5251; ANDS2451; ANDS3973; ANDS4774; ANDW4361; ANDX2782; ANDZ3820; ANEN1082; ANEN2140; ANER1507; ANEZ0923; ANFN1444; ANFN1444; ANFN2844; ANDQ5251; ANDS3973; ANFN3975; ANDZ3820; ANEN4896; ANER1507; ANER1507; ANDQ5251; ANDQ3872; ANDX5373; ANEQ4962; ANER1507; ANDQ3025; ANDZ2291; ANDX5373; ANEQ2733; ANER1507; ANCN1543; ANEQ1353; ANDT2946; ANDT2946; ANDX2782; ANDX5373; ANEQ2733; ANDT2946; ANDX3060; ANDX5373; ANEQ1353; ANEQ1353; ANEQ4962; ANER1507; ANDS3973; ANDW1997; ANEQ2733; ANFN1444; ANFN2844; ANDT2946; ANEQ2733; ANER0517; ANDS3973; ANDS4774; ANER1507; ANER1507; ANDZ2291; ANFP2318; ANES4086; ANDS3973; ANDX0760; ANDX5373; ANER1507; ANFN1444; ANES2211; ANER1507; ANDS3973; ANFP2318; ANDT2946; ANDZ3820; ANEQ2733; ANFN2844; ANEN4896; ANER1507; ANDT2946; ANDW4361; ANDP1356; ANEN4896; ANFN1444; ANEQ2733; ANER1507; ANDZ3820; ANER1507; ANDT2946; ANFN3975; ANDX2782; ANDX5373; ANDW4361; ANEZ2098; ANFN1444; ANDZ2291; ANER1507; ANEQ1353; ANDQ5251; ANFN1444; ANDX5373; ANER1507; ANER1507; ANER1507; ANER1507; ANFN3975; ANDS3973; ANEN4896; ANDQ3872; ANDW4361; ANDX2782; ANEQ4962; ANDT2946; ANER1507; ANDQ5251; ANDX5373; ANDY1234; ANDX5373; ANER1507; ANDS4774; ANER1507; ANDS4774; ANDX0760; ANDY1234; ANER1507; ANER1507; ANDW1997; ANDX0760; ANDX5373; ANDZ3820; ANER1507; ANER1507; ANDV0687; ANDY1234; ANER1507; ANFN2844; ANFP2318; ANEZ2098; ANDU1877; ANDW1997; ANDX5373; ANDW4361; ANDS3973; ANEN4896; ANDW4361; ANDS2451; ANDP1356; ANDW4361; ANER1507; ANEQ4962; ANDQ3025; ANDS2451; ANES2211; ANES4086; ANES4086; ANEZ0923; ANEZ2098; ANER0517; ANFN1444; ANDQ3025; ANDQ3872; ANDS2451; ANDS2451; ANDS4774; ANDT2946; ANDT2946; ANDV0687; ANDW1997; ANDW3561; ANDW4361; ANDX0760; ANDX0760; ANDX2782; ANDX2782; ANDX2782; ANDX3060; ANDX5373; ANDY1234; ANDZ2291; ANDZ3820; ANDZ3820; ANEN2140; ANEQ1353; ANER1507; ANES1043; ANES1043; ANES2211; ANEZ0923; ANFN1444; ANFN1444; ANFN2844; ANFN3975; ANFN3975; ANFP0879; ANEN2140; ANEN4896; ANEP2420; ANEQ1353; ANEQ2733; ANES2211; ANES2211; ANEZ2098; ANEZ2098; ANEZ3412; ANFN1444; ANFP0879; ANFP2318; ANDQ3025; ANDQ5251; ANDS3973; ANDS4774; ANDT2946; ANDX0760; ANDX2782; ANDX3060; ANER1507; ANER1507; ANER1507; ANER1507; ANES2211; ANDW4361; ANDS3973; ANDV0687; ANES4086; ANDX5373; ANEP2420; ANES4086; ANDQ3872; ANER1507; ANDX5373; ANDS2451; ANDT2946; ANEN4896; ANER1507; ANDQ3025; ANDS4774; ANER1507; ANDQ3025; ANEP2420; ANER1507; ANDQ3025; ANDS2451; ANEN4896; ANEQ1353; ANEQ4962; ANER1507; ANFN1444; ANDS2451; ANER0517; ANER1507; ANDT2946; ANER1507; ANER1507; ANFP0879; ANDT2946; ANDY1234; ANDQ3025; ANDU1877; ANDV0687; ANDX5373; ANDY1234; ANES4086; ANEN2140; ANFP2317; ANFP2318; ANDS4774; ANDT2946; ANDW1997; ANEQ1353; ANDT2946; ANDX5373; ANCW1737; ANDQ3025; ANER1507; ANDT2946; ANDS4774; ANFN1444; ANDX3060; ANDY1234; ANDX5373; ANDT2946; ANDY1234; ANER0517; ANDP0508; ANDP0508; ANDP1356; ANDP1356; ANEP2420; ANEQ1353; ANEQ2733; ANEQ4962; ANER0517; ANER1507; ANEZ2098; ANFP2318; ANER0517; ANDT2946; ANDT2946; ANES4086; ANCX3883; ANCX3883; ANDV0702; ANDV2519; ANDV2519; ANDX0727; ANDX0727; ANDX2426; ANDX2426; ANDX3063; ANDX3063; ANDY3291; ANDY3291; ANDZ1331; ANDQ3033; ANDS0753; ANDS0753; ANDT4130; ANDV0702; ANDX2032; ANDY3291; ANDZ1331; ANDZ2297; ANDZ3824; ANDZ3824; ANEN4886; ANEP0860; ANER5168; ANEZ2099; ANEZ3952; ANFN3976; ANDW4378; ANEP0860; ANDV0702; ANDZ3824; ANEQ1354; ANFN3976; ANDX5307; ANDY1221; ANEQ1354; ANEP3934; ANDX2426; ANDY3291; ANEN2145; ANEP3934; ANEQ1354; ANDQ0322; ANDS0753; ANDX2426; ANEQ4069; ANEZ3952; ANEQ4069; ANFP2336; ANDZ2297; ANEQ4069; ANDV1233; ANDY1221; ANEP0860; ANEQ1354; ANFN3976; ANEQ1354; ANEP0860; ANFN3976; ANFN3976; ANEQ4069; ANFN3976; ANDT4130; ANFP0912; ANEQ4069; ANEP3934; ANEP3934; ANEQ4069; ANEP0860; ANEZ3952; ANDQ3033; ANEQ1354; ANCS3348; ANDT4130; ANEP0860; ANEP3934; ANEQ4069; ANDV0702; ANDY2464; ANEP0860; ANDV1233; ANEQ4069; ANCS3348; ANDV0702; ANDV1233; ANDX2426; ANDY3291; ANEN2145; ANEN4886; ANEP0860; ANDW4378; ANDS0753; ANEP0860; ANEQ1354; ANEQ1354; ANDS0753; ANDV0702; ANDX2426; ANDX2426; ANEN2145; ANEP0860; ANEP0860; ANEQ1354; ANER5168; ANDQ0322; ANDQ3033; ANDS0753; ANDS0753; ANDU3110; ANDV0300; ANDW4378; ANDW4378; ANDX0727; ANDX2426; ANDX5307; ANDX5307; ANDY1221; ANDY2464; ANDZ1331; ANDZ2297; ANEN2145; ANEN3389; ANEN4886; ANEP0860; ANEP3934; ANER5168; ANES2212; ANES2212; ANEZ0924; ANEZ2099; ANEZ2099; ANEZ2099; ANEZ3952; ANEZ3952; ANFN2787; ANFN3976; ANFN3976; ANFP0912; ANEZ0924; ANEZ3952; ANFP0912; ANDQ3033; ANDS0753; ANDT4130; ANDV1233; ANDY1221; ANEP3934; ANEP3934; ANEQ1354; ANEQ1354; ANEQ4069; ANEQ4069; ANEZ3952; ANDZ2297; ANEQ4069; ANEQ4069; ANDS0753; ANDT4130; ANDV0702; ANDX2032; ANDX5307; ANEN2145; ANEN3389; ANEQ4069; ANDQ0322; ANDS0753; ANDT4130; ANDU3110; ANEQ1354; ANEQ1354; ANER5168; ANDX5307; ANEQ1354; ANES2212; ANEN2145; ANEQ4069; ANDV2519; ANDX2426; ANDX5307; ANDY2464; ANDZ1331; ANDZ1331; ANDZ3824; ANEN1064; ANEN2145; ANEN2145; ANEQ1354; ANDZ1331; ANER5168; ANFP0912; ANFP2336; ANDV2519; ANEQ1354; ANES2212; ANDT4130; ANCX3883; ANDV0702; ANDY2464; ANER5168; ANDZ2297; ANEQ1354; ANEN3389; ANDT4130; ANDT4130; ANES2212; ANDQ0322; ANDT4130; ANDZ3824; ANEN3389; ANDV0702; ANDV1233; ANEZ0924; ANEZ3952; ANFN3976; ANCT4331; ANDQ0322; ANDU3110; ANEQ4069; ANEP0860; ANEQ4069; ANER5168; ANDV2519; ANEQ1354; ANDY1221; ANEQ1354; ANCX3884; ANDR0754; ANDS2461; ANDS3976; ANDS3976; ANDS3976; ANDS4839; ANDT4131; ANDY1231; ANDY3284; ANDY3284; ANDY3284; ANDR0754; ANDS2461; ANDS3976; ANDX2031; ANDX3062; ANDX5363; ANDX5363; ANDY1231; ANDZ3823; ANEN2147; ANEN3398; ANER1508; ANET0556; ANEZ3429; ANCS0655; ANEQ1355; ANER0518; ANEQ1355; ANDS3976; ANDS3976; ANDS4839; ANDX5323; ANDX5363; ANDX5363; ANER0518; ANDX5363; ANER0518; ANDX5363; ANER0518; ANER0518; ANET0556; ANER0518; ANER1508; ANFP2365; ANDS4839; ANDX5323; ANDX5323; ANDX5363; ANEQ1355; ANDU1889; ANDX2031; ANER1508; ANER1508; ANET0556; ANER0518; ANEQ1355; ANDX5363; ANDX5363; ANDT4131; ANER0518; ANDX5323; ANER1508; ANDS3976; ANDQ0321; ANDW1999; ANET0556; ANDS2461; ANDR0754; ANEQ1355; ANER1508; ANES4087; ANEZ2100; ANDQ0321; ANDQ3876; ANDS2461; ANDS2461; ANDS3976; ANDT4131; ANDT4131; ANDU1889; ANDW1999; ANDW1999; ANDX0739; ANDX0739; ANDX3062; ANDX5323; ANDX5323; ANDX5323; ANDX5363; ANDY3284; ANDY4974; ANDZ3823; ANEN2147; ANEN2147; ANEQ1355; ANEQ2734; ANEQ4963; ANER1508; ANES4087; ANES4087; ANET0556; ANET0556; ANEY2031; ANEY2031; ANEZ3429; ANEZ3429; ANFN1445; ANFN1445; ANFN3977; ANFN3977; ANFN3977; ANFP2365; ANEN3398; ANEQ1355; ANEQ4963; ANER1508; ANET0556; ANEY2031; ANEZ2100; ANFP2365; ANDQ3876; ANDX2031; ANDX5323; ANER0518; ANER0518; ANER0518; ANER1508; ANER1508; ANDU1889; ANDT4131; ANER0518; ANER0518; ANDQ0321; ANDS3976; ANDS4839; ANER1508; ANDX5363; ANDX5363; ANDS3976; ANDX5323; ANDQ3876; ANDT4131; ANDU1889; ANDW1999; ANDW4395; ANDW4395; ANDX5363; ANDX5363; ANDY3284; ANDY4974; ANDZ3823; ANET0556; ANDY3284; ANDR0754; ANDU1889; ANEN3398; ANER1508; ANET0556; ANFP2365; ANDS4839; ANDX2031; ANET0556; ANER0518; ANDX5363; ANDR0754; ANDX2031; ANDS4839; ANDS3976; ANDR0754; ANDX2031; ANDX5323; ANFN1445; ANDZ3823; ANDR0754; ANDS4839; ANDU1889; ANEN3398; ANER1508; ANDU1889; ANDS4839; ANDX5363; ANEN2147; ANER0518; ANDR0754; ANDR0754; ANDS4839; ANDZ3823; ANEN3398; ANEQ4963; ANER0475; ANER0518; ANER1508; ANER0518; ANDU4028; ANDV0275; ANDV0275; ANDW4413; ANDW4413; ANDX5301; ANDX5301; ANDY4972; ANDY4972; ANDU4028; ANDV0275; ANEN1083; ANEN1083; ANEQ2735; ANDW4413; ANEQ2735; ANER5120; ANEQ2735; ANDQ3028; ANES4088; ANEQ2735; ANES4088; ANER5120; ANES4088; ANES4088; ANDV0275; ANDQ3028; ANDU4028; ANEP5329; ANES4088; ANES4088; ANES4088; ANFP0880; ANFP0880; ANDQ3028; ANDW4413; ANEP5329; ANEP5329; ANES4088; ANES4088; ANDX5301; ANEN1083; ANEQ2735; ANEQ2735; ANES4088; ANDX5301; ANDV0275; ANDQ3028; ANDW4413; ANDW4413; ANDX0751; ANDX5301; ANDX5301; ANDY4972; ANER5120; ANDU4028; ANDV0275; ANDX5301; ANDX5301; ANER5120; ANER5120; ANES4088; ANER5120; ANDU4028; ANDX5301; ANDV0275; ANDV0275; ANDV0275; ANEQ2735; ANEQ2735; ANEN1083; ANEN1083; ANEN4883; ANEN4883; ANEP5329; ANEQ2735; ANEQ2735; ANFP0880; ANFP0880; ANFP0880; ANFP2331; ANER5120; ANDQ3027; ANDQ3027; ANDU4040; ANDU4040; ANDX5327; ANDX5327; ANDX5389; ANDX5389; ANEP5309; ANEP5309; ANEP5309; ANEQ1356; ANEQ2736; ANEP5309; ANEP5309; ANER0519; ANDQ3874; ANDX5327; ANER1535; ANEP5309; ANDQ3027; ANDX5327; ANEP5309; ANEQ1356; ANDX5327; ANDQ3874; ANDU1103; ANDU4040; ANDX5389; ANDZ1326; ANEP5309; ANER1535; ANER1535; ANER1535; ANDQ3874; ANDQ3874; ANDU1103; ANER1535; ANDU1103; ANCX3885; ANEP5309; ANEP5309; ANEQ1356; ANDU1103; ANEP5309; ANEP5309; ANEQ2736; ANDU1104; ANDY2452; ANDY2452; ANDY3299; ANEN4894; ANDQ3034; ANDQ3034; ANEN1084; ANEN1084; ANEQ4964; ANEN1084; ANEN1084; ANDY2452; ANDZ3825; ANEN1084; ANCQ4294; ANDQ3034; ANDU1104; ANDY2452; ANDZ3825; ANEN1084; ANER0520; ANER0520; ANER0520; ANER0520; ANCV1583; ANER0520; ANDU1104; ANEN1084; ANDT1055; ANDR1267; ANDT1055; ANDW1998; ANDW1998; ANDW1998; ANDX1402; ANDY4964; ANDY4964; ANES2189; ANET0542; ANEW4218; ANEX1697; ANEZ2081; ANEZ3398; ANFN3946; ANFP0857; ANDR1267; ANDT1055; ANEN2153; ANEP2424; ANDU4039; ANDT1055; ANDT0506; ANEZ3398; ANCW3262; ANDT0506; ANDU2242; ANEZ3398; ANDZ2307; ANDR1267; ANDW1998; ANEN2153; ANDU2242; ANDZ2307; ANEX6010; ANEZ0908; ANDU4039; ANFN3946; ANDZ2307; ANDT0506; ANDT1055; ANFP0857; ANDU4039; ANDT1055; ANDZ2307; ANEZ2081; ANEX6010; ANDT1055; ANEW4218; ANDT1055; ANDW1998; ANDT0506; ANDU2242; ANDU2242; ANDZ2307; ANEN2153; ANCS1158; ANFP0857; ANDT1055; ANEZ3398; ANDW1998; ANEP2424; ANDW1998; ANDY3289; ANDY3289; ANDY4964; ANDZ2307; ANET0542; ANEW3366; ANEX1697; ANEX6010; ANEX6010; ANEZ0908; ANEZ3398; ANFP0857; ANDT0506; ANDX1402; ANDY3289; ANDR1267; ANDT0506; ANDU2242; ANDU4039; ANEP2424; ANEP2424; ANEW4218; ANEX6010; ANEX6010; ANEZ2081; ANEZ3398; ANFP0857; ANDU4039; ANDT0506; ANDT1055; ANDW1998; ANEW4218; ANDU4039; ANDZ2307; ANEN2153; ANDZ2307; ANES2189; ANDU2242; ANDU2242; ANES2189; ANDP1357; ANDR1267; ANEN2153; ANEN2153; ANDR1267; ANDZ2307; ANFN3946; ANDY3289; ANDZ2307; ANDX1402; ANDZ2307; ANDR1267; ANDT1055; ANDX1402; ANEN2153; ANEN2153; ANDU4039; ANDU4039; ANEP2424; ANET0542; ANFP0857; ANCW3262; ANDT0506; ANDU2242; ANDU4039; ANDU4039; ANDZ2307; ANDU2242; ANDT0506; ANDU4039; ANDR1267; ANEZ2081; ANDR1267; ANDX1402; ANEP2424; ANES2189; ANDP2690; ANDU2242; ANCR1285; ANDY4964; ANEX6010; ANDU4039; ANEP2424; ANEX1697; ANEP2421; ANDQ3868; ANDS0752; ANDT1059; ANDX1401; ANDX2425; ANDX5362; ANDY1230; ANDZ1322; ANDZ2286; ANEN4897; ANEN4897; ANEP0867; ANEP2421; ANEV1711; ANEW4219; ANFN1432; ANDN0810; ANDS0752; ANDS0752; ANDV4534; ANDX5362; ANDZ1322; ANDZ2286; ANEQ4039; ANEN4897; ANEQ2718; ANCW1424; ANDT0507; ANDX1401; ANDQ3022; ANDV4534; ANFP0858; ANFN3947; ANFN3947; ANEW4219; ANEP3930; ANES4062 ANDX2425; ANDQ1713; ANEP2421; ANEP3930; ANEZ2082; ANDQ3022; ANFP2358; ANDV4534; ANEQ4039; ANER1514; ANDT0507; ANDV4534; ANDY1230; ANDV4534; ANDT1059; ANEP3930; ANEW4219; ANCW1424; ANER1514; ANCX1867; ANER1514; ANDX2425; ANFN3947; ANDY1230; ANDY1230; ANDZ1322; ANDY1230; ANEP0867; ANEQ4039 ANES2190; ANDZ1322; ANEP0867; ANEW3367; ANEW4219; ANES4062; ANES4062; ANEV1711; ANEW3367; ANEW4219; ANEX0746; ANEX1708; ANEY1228; ANFN1432; ANFN3947; ANFP0858; ANDQ1713; ANDQ3022; ANDQ3022; ANDQ3868; ANDX3057; ANDX5362; ANCV0494; ANDN0810; ANDT0507; ANDT0507; ANDX1401; ANDX1401; ANDX2425; ANDX5362; ANDY1230; ANDZ2286; ANEN4897; ANEP0867; ANEP3930; ANEQ2718; ANEQ4039; ANEQ4039; ANEQ4039; ANER1514; ANER1514; ANER1514; ANES2190; ANES4062; ANEX1708; ANEX3091; ANDT1059; ANDQ3868; ANEP3930; ANCU0848; ANDX1401; ANER1514; ANDQ3868; ANDX2425; ANDY1230; ANEQ4039; ANER1514; ANES4062; ANEY1228; ANDT0507; ANDV4534; ANDY1230; ANDN0810; ANDT0507; ANDT1059; ANDX2425; ANDZ1322; ANER1514; ANFP0858; ANDT0507; ANEW4219; ANEZ2082; ANER1514; ANEX3091; ANEW4219; ANCU0848; ANDX3057; ANDZ1322; ANER1514; ANES2190; ANDS0752; ANER1514; ANDT0507; ANDQ1713; ANDS0752; ANDX2425; ANDX3057; ANER1514; ANEX3091; ANFN3947; ANDX2425; ANDX2425; ANER1514; ANEX3091; ANDT0507; ANDX1401; ANDX2425; ANDV4534 ANEP3930; ANES4062; ANFP0858; ANDS0752; ANDV4534; ANDX1401; ANEQ4039; ANEQ4039; ANDX3057; ANEZ2082; ANES4062; ANEQ4039; ANDT0507; ANDX2425; ANES4062; ANEW4219; ANEP3930; ANEQ2718; ANCV0494; ANDQ1713; ANDQ1713; ANFN3947; ANDV4534; ANDX2425; ANEQ4039; ANFN3947; ANDQ1713; ANDT1059; ANEQ4039; ANER1514; ANDS0752; ANDS0752; ANDX3057; ANES2190; ANDZ1322; ANEP3930; ANES4062; ANFN3947; ANEQ4039; ANEX3091; ANER1514; ANES2190; ANDX2425; ANEQ4039; ANEP3930; ANES4062; ANDX3057; ANDX3057; ANDT0507; ANEX1708; ANEP5317; ANDR2271; ANDR2271; ANDV3567; ANDX1405; ANDX2034; ANDY4960; ANDZ1332; ANEP3939; ANES2191; ANET0543; ANEW4260; ANFN2783; ANFN3948; ANCW4216; ANDN0821; ANDN0821; ANDV4528; ANDZ2304; ANEN1070; ANEN3402; ANEQ4040; ANBY2393; ANCQ2599; ANDV4528; ANEQ4040; ANFN3948; ANFN3948; ANCW3263; ANEQ4040; ANEQ4040; ANDR2271; ANFN3948; ANFN3948; ANDP3957; ANEX3093; ANCW3263; ANEP3939; ANFN3948; ANDV4528; ANDV4528; ANDV4528; ANEQ4040; ANES2191; ANDV4528; ANDR2271; ANDV4528; ANCW4216; ANEP5317; ANCT1573; ANDV4528; ANDZ1332; ANEP5317; ANDZ2304; ANDZ2304; ANDZ2304; ANEP5317; ANEP5317; ANFN3948; ANDQ3039; ANDV4528; ANDZ1332; ANDZ1332; ANEP5317; ANET0543; ANET3786; ANET3786; ANEW4260; ANEW4260; ANEX3093; ANEX3093; ANEY1221; ANEY3210; ANEY3210; ANEY3210; ANFN2783; ANFN3948; ANDP3957; ANDQ1717; ANDR2271; ANCT4314; ANCV0496; ANDP3957; ANDQ3039; ANDV4528; ANDX1374; ANDY1238; ANDY4960; ANDZ1332; ANEP3939; ANEQ2719; ANEQ4040; ANEQ4040; ANEQ4040; ANEQ4040; ANER5118; ANER5118; ANES2191; ANEX3093; ANEX6157; ANEY1221; ANFN3948; ANDZ2304; ANDY1238; ANEQ2719; ANDX1374; ANDV4528; ANEQ4040; ANDP3957; ANDV4528; ANDX1374; ANDY1238; ANDP3987; ANDV3567; ANEP3939; ANER5118; ANEY3210; ANDZ2304; ANDV4528; ANEY3210; ANEZ3431; ANDZ2304; ANDV4528; ANDP3957; ANES2191; ANEX6157; ANEZ3431; ANER5118; ANDQ1717; ANET3786; ANEY1221; ANEX6157; ANFN3948; ANDV4528; ANCW4216; ANDQ1717; ANDX1405; ANDQ3039; ANDX1374; ANDX2034; ANDY1238; ANDZ1332; ANEN1070; ANER5118; ANEQ4040; ANDR2271; ANDQ1717; ANDR2271; ANDV4528; ANDV4528; ANEQ4040; ANEX6157; ANEX6157; ANET3786; ANEY1221; ANEP5317; ANEQ2719; ANEQ4040; ANDP3957; ANEZ3431; ANDQ3039; ANDX1405; ANEP5317; ANDX1374; ANDX1374; ANEQ4040; ANEQ4040; ANFN3948; ANEQ4040; ANCW3263; ANEQ2719; ANEQ4040; ANER5118; ANDQ1717; ANEP5317; ANEQ4040; ANES2191; ANDR2271; ANDV3567; ANEP5317; ANEQ4040; ANDV3567; ANDX2034; ANEN3402; ANDR2271; ANET3786; ANEX6157; ANET3786; ANDY1238; ANDZ2304; ANET3786; ANFN3948; ANEP2435; ANDQ1723; ANDQ3040; ANDU2243; ANDW4355; ANDW4355; ANDW4355; ANDX1406; ANDX2035; ANDX5316; ANDX5390; ANDZ2283; ANEN1071; ANEN4895; ANER1515; ANES4063; ANEX0672; ANEX6011; ANEZ3399; ANFN1433; ANCX0889; ANDP3959; ANDP3959; ANDP3959; ANDP3959; ANDX2035; ANDZ2283; ANEN1071; ANEQ2720; ANEQ4041; ANDW4355; ANEX0672; ANFN1433; ANDQ3040; ANDV1244; ANDX2035; ANDX2035; ANET3787; ANDQ3040; ANEQ2720; ANER1515; ANEY3211; ANEQ2720; ANDX5316; ANFN1433; ANCX0889; ANCX3891; ANDQ3879; ANDU2243; ANDU2243; ANDV1244; ANDX2035; ANDU2243; ANDX2035; ANDV1244; ANES4063; ANDQ1723; ANDV1244; ANEP2435; ANEQ2720; ANEX6011; ANDV1244; ANDQ1723; ANDQ3040; ANDZ2283; ANEQ2720; ANDQ5218; ANEN4895; ANEX3081; ANER1515; ANES4063; ANET3787; ANEW3412; ANEX3081; ANEX6011; ANEY3211; ANEZ3399; ANFN1433; ANFN3949; ANDQ0326; ANDQ1723; ANDQ3040; ANDQ3879; ANDP3959; ANDQ0326; ANDX5316; ANDX5316; ANDX5316; ANDX5390; ANDY3308; ANDZ2283; ANEQ2720; ANEQ4947; ANEQ4947; ANER1515; ANES4063; ANEX3081; ANEX6011; ANDU2243; ANDU2243; ANEQ4041; ANCX3891; ANDW4355; ANEN4895; ANDQ3879; ANDQ5218; ANDU2243; ANDV1244; ANDX5390; ANDZ2283; ANEN4895; ANEP2435; ANEQ4947; ANER1515; ANER1515; ANER1515; ANES4063; ANFN3949; ANDQ1723; ANDQ5218; ANDV1244; ANCQ1429; ANDZ2283; ANEQ4947; ANER1515; ANEY3211; ANDX1406; ANDX2035; ANDQ3040; ANCQ1429; ANDX5390; ANDX5390; ANEX6011; ANEY3211; ANDU2243; ANDX2035; ANER1515; ANDQ1723; ANDW4355; ANDX1406; ANER1515; ANEQ2720; ANCQ2600; ANDX1406; ANEQ4947; ANEQ4947; ANCN3578; ANDX2035; ANDP3959; ANDQ1723; ANDU2243; ANDX2035; ANDX2035; ANEQ2720; ANEQ4041; ANEQ4041; ANEQ4041; ANER1515; ANEX6011; ANES4063; ANEX6011; ANDZ2283; ANEQ2720; ANDX2035; ANEP2435; ANDX5316; ANEQ4041; ANEX3081; ANDW4355; ANER1515; ANEX6011; ANFN1433; ANDX5390; ANDQ3879; ANDX5390; ANEN1071; ANES4063; ANEX0672; ANEY3211; ANEQ4947; ANEX6011; ANDX1406; ANES4063; ANFN1433; ANEX6011; ANEZ3399; ANFN3949; ANEN2150; ANDS0751; ANDS0751; ANDS4788; ANDU2244; ANDV4533; ANDX3056; ANDX5361; ANDY3283; ANDY4965; ANDZ1327; ANDZ2287; ANEW4220; ANEW4220; ANEX1707; ANFN1434; ANFN2784; ANDP3950; ANDP3950; ANDX1373; ANDY4965; ANEN4880; ANDS4788; ANDS4823; ANEQ4042; ANCQ1430; ANDN0809; ANDS4788; ANDW4372; ANEQ4042; ANDY3283; ANEQ4042; ANER1516; ANEN2150; ANEN4880; ANCS1563; ANCW4217; ANDS4823; ANEN2150; ANFN2784; ANDS4788; ANDY4965; ANEP0877; ANER1516; ANDS4823; ANDS4788; ANDY3283; ANFN2784; ANDW4372; ANDZ2287; ANDZ2287; ANEX1707; ANDP3950; ANDX1373; ANDX3056; ANDY1229; ANDY2456; ANCW4217; ANDS0751; ANDS4788; ANDS4823; ANDU2244; ANDV4533; ANDW4372; ANDW4372; ANDW4372; ANDY2456; ANDY2456; ANEP0877; ANEQ4042; ANER1516; ANEX1707; ANDY3283; ANDS4788; ANEN2150; ANDX1373; ANEW4220; ANDS4788; ANEX1707; ANEQ4042; ANDV4533; ANDS4823; ANER1516; ANER1516; ANEX1707; ANDS4788; ANDU2244; ANEW4220; ANEW4220; ANCS3345; ANCS3345; ANDS4788; ANDX1373; ANDZ2287; ANDU2244; ANDU2244; ANEW4220; ANER1516; ANER1516; ANFN2784; ANER1516; ANFN2784; ANER1516; ANDW4372; ANDS4823; ANEP0877; ANDV4533; ANEN4880; ANEX1707; ANEN2150; ANER1516; ANCS3345; ANDN0809; ANEN4880; ANDY3283; ANEN4880; ANEN4880; ANEW4220; ANEW4220; ANEN2150; ANDS4823; ANEP0877; ANEQ4042; ANEX1707; ANDS0751; ANDU2244; ANER1516; ANDS4823; ANDU2244; ANDS4788; ANDZ2287; ANEP0877; ANDY4965; ANDZ2287; ANER1516; ANEW4220; ANEX1707; ANEX1707; ANEX1707; ANCQ1430; ANER1516; ANFN1434; ANEN3387; ANEN1072; ANEP2430; ANBZ1288; ANDX5376; ANEP2430; ANCS3433; ANEN1072; ANER1517; ANDS1282; ANDS1282; ANDU2245; ANEP2430; ANDW2002; ANDW4389; ANDX5376; ANDY3288; ANCS3433; ANDU2245; ANDU2245; ANDU2245; ANDW2002; ANDY3288; ANDY3288; ANEN1072; ANEP2430; ANEQ4043; ANER1517; ANDS1282; ANEQ4043; ANEQ4043; ANER1517; ANDU2245; ANDU2245; ANDW2002; ANDW2002; ANEN1072; ANEN3387; ANBZ1288; ANDS1282; ANDW4389; ANDS1282; ANDS1282; ANEP2430; ANDX5376; ANER1517; ANDS1282; ANEP2430; ANEN3387; ANEQ4043; ANDX5376; ANEN1072; ANEN3387; ANDU2245; ANEN3387; ANDX5376; ANEN1072; ANEN3387; ANFN3952; ANEN1072; ANDY3287; ANDX5375; ANDY3287; ANEP2412; ANDV4536; ANDY3287; ANEQ4948; ANES4065; ANDS4776; ANDS4776; ANDV4536; ANDX5375; ANDY3287; ANEP2412; ANEP2412; ANEP2412; ANEQ4948; ANDY3287; ANEP2412; ANDV4536; ANDY3287; ANDX1403; ANDX5375; ANCY3335; ANDV4536; ANDX1403; ANDX1403; ANDX5375; ANDS4776; ANDY3287; ANEQ4948; ANES4065; ANEQ4948; ANES4065; ANDS4776; ANDT1686; ANEZ3427; ANEZ3427; ANCY3335; ANDS4776; ANDT1686; ANEQ4948; ANDX1403; ANCU1841; ANEQ4948; ANEQ4948; ANEP2412; ANDY3287; ANES4065; ANES4065; ANEZ3427; ANEZ3427; |
Recalling Firm/ Manufacturer |
Bard Peripheral Vascular Inc 1625 W 3rd St Bldg 1 Tempe AZ 85281-2438
|
For Additional Information Contact |
844-823-5433
|
Manufacturer Reason for Recall |
Reports of the proximal end of the stent not immediately expanding upon deployment and remaining connected to the delivery system.
|
FDA Determined Cause 2 |
Under Investigation by firm |
Action |
On 01/13/2021, the firm sent an "URGENT MEDICAL DEVICE SAFETY NOTICE" via FedEx to customers informing them that the Recalling Firm has received reports regarding the venous stent system 9F where the proximal end of the stent does not immediately expand upon deployment, and remains connected to the delivery system, and that over-manipulation or forcing the catheter delivery system in attempts to assist the stent's expansion, could potentially have a varying degree of harm associated with it. potential harm ranges from prolonging the procedure, damaging or deformity of the stent, potential vascular injury and / or hemodynamic disruption affecting the blood flow and / or a thrombotic event.
The Recalling Firm is informing customers that the affected product can continue to be used, but are instructing them to:
1. Share this communication with all physicians using the Venovo" Venous Stent Systems product within their organization to ensure that they are aware of this Safety Notice.
2. Continue the use of these Venovo" Venous Stent Systems 9F, 14mm products following the Safety Notice guidance.
3. Report any adverse health consequences experienced with the use of this product to the Recalling Firm.
A list of the affected Product Codes and Lot/Serial Numbers are included as Appendix 1 attached to the Recall Notification Letter.
For questions/inquiries/assistance contact the Recalling Firm's Customer/Technical Support at 1-888-237-2762 (Opt 4, Opt 2) Monday thru Friday 8:00 a.m. - 5:00 p.m. CST
Updated Information for Recall Expansion (dated 05/11/2021):
On 05/07/2021, the firm sent additional " "URGENT MEDICAL DEVICE SAFETY NOTICE" via FedEx to customers informing them that the scope of the recall (initiated on 01/13/2021) was being expanded to include all sizes and lots within expiry date.
Customer are instructed to:
1. Check all inventory locations within their institution for the affected Venous Stent System listed in Attachment A and immediately |
Quantity in Commerce |
Total Devices=12,258 (U.S.= 7,234; O.U.S.=5,024) |
Distribution |
Worldwide distribution. US Nationwide, ALGERIA,
ANDORRA,
ARGENTINA,
AUSTRALIA,
AUSTRIA,
AZERBAIJAN,
BAHRAIN,
BELGIUM,
BRAZIL,
BRUNEI,
BULGARIA,
CHILE,
COSTA RICA,
CYPRUS,
CZECH REPUBLIC,
DENMARK,
DEUTSCHLAND,
EGYPT,
FINLAND,
FRANCE,
GEORGIA,
GREECE,
HONG KONG,
HUNGARY,
INDIA,
IRAN,
IRAQ,
IRELAND,
ISRAEL,
ITALY,
JORDAN,
KAZAKHSTAN,
MALAYSIA,
MEXICO,
NETHERLANDS,
NORWAY,
OMAN,
PANAMA,
POLAND,
PORTUGAL,
ROMANIA,
RUSSIA,
SAUDIA ARABIA,
SINGAPORE,
SLOVAKIA (SLOVAK REPUBLIC),
SLOVENIA,
SOUTH AFRICA,
SOUTH KOREA,
SPAIN,
SWEDEN,
SWITZERLAND,
TAIWAN,
TURKEY,
UNITED ARAB EMIRATES (UAE),
UNITED KINGDOM,
VIETNAM |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
PMA Database |
PMAs with Product Code = QAN and Original Applicant = Bard Peripheral Vascular, Inc.
|
|
|
|