| | Class 2 Device Recall Ethicon |  |
| Date Initiated by Firm | July 20, 2021 |
| Create Date | August 26, 2021 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2356-2021 |
| Recall Event ID |
88434 |
| 510(K)Number | K191555 |
| Product Classification |
Instrument, ultrasonic surgical - Product Code LFL
|
| Product | HARMONICHD 1000i Shears (36cm Shaft Length)-indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired.
Product Code: HARHD36 |
| Code Information |
Lot Codes:
U94Y9V
U95C2A
U94Z98
U95427
U9516E
U95599
U9503G
U95856
U94Y9W
U95L0A
U9518D
U95526
U9507E
U9548R
U95127
U95525
U9530P
U95T2Z
U95366
U95R4F
U9521Z
U95F2T
U9554U
U95F8W
U9564G
U95U6A
U95524
U95E6F
U9512D
U95K4D
U94Z49
U95F07
U95A9X
U95F6N
U9569W
U95E1U
U9571P
U95E72
U9587D
U95E29
U95815
U95T90
U95814
U95Z67
U95754
U95Y7G
U95D4T
U9523L
U95566
U9530X
U9543R
GTIN: 10705036015055 |
| FEI Number |
1527736
|
Recalling Firm/ Manufacturer |
Ethicon Endo-Surgery Inc 4545 Creek Rd Blue Ash OH 45242-2803
|
| For Additional Information Contact | SAME 513-337-7000 |
Manufacturer Reason for Recall | An internal component may be cracked and become lodged behind the energy button potentially resulting in continuous activation of the device, may result in burns |
FDA Determined Cause 2 | Process control |
| Action | Ethicon issued Urgent Medical Device Removal letter to Operating Room Supervisors, Materials Management Personnel, and Chief of Surgery, on July 20, 2021.
Letter states reason for recall, health risk and action to take:
1. Examine your inventory immediately to determine if you have product subject to this recall on hand and quarantine such product(s).
2. Remove the product subject to this recall and communicate the issue to relevant operating room or materials management personnel, or anyone else in your facility who needs to be informed.
3. If any product subject to this recall has been forwarded to another facility, contact that facility to arrange return. Please consider including a copy of this recall letter when communicating.
4. Complete the Business Reply Form (BRF) (Attachment 3) confirming receipt of this notice and fax or email it to Stericycle at 877-884-7828 or Ethicon7616@stericycle.com within three (3) business days. Please return the BRF even if you do not have product subject to this recall.
5. Customers are required to return unused affected HARMONIC HD 1000i Shears subject to this recall that are in their inventory immediately. To receive replacement reimbursement, customers must return product subject to this recall no later than October 31, 2021. Any non-affected product and any product returned after the date specified will not receive replacement.
6. Keep this notice visibly posted for awareness until all product subject to this recall has been returned to Stericycle. While processing your returns, please maintain a copy of this notice with the product subject to this recall and keep a copy for your records.
7. To return product subject to this recall, photocopy the completed BRF, place it in the box with the product, and affix the pre-paid authorized shipping label included with the recall notification letter. Ethicon will pay for the shipping charges only if the authorized label is used. Extra shipping labels may be obtained by calling 888-3 |
| Quantity in Commerce | 68960 units |
| Distribution | Worldwide distribution - US Nationwide distribution and the countries of Austria, Argentina, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Czechia,
Denmark, Estonia, Finland, France, Germany, Greece, Guadeloupe, Hong Kong, Ireland,
Israel, Italy, Japan, Jordan, Latvia, Lebanon, Luxembourg, Malaysia, Mexico, Mongolia,
Netherlands, New Zealand, Norway, Peru, Philippines, Poland, Portugal, Puerto Rico,
Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovenia, South Korea, Spain, Sweden,
Switzerland, Taiwan, Thailand, Turkey, United Kingdom, United Arab Emirates. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = LFL
|
|
|
|