  Class 1 Device Recall ArrowTrerotola PTD see related information
Date Initiated by Firm January 11, 2022
Create Date February 15, 2022
Recall Status1 Open3, Classified
Recall Number Z-0624-2022
Recall Event ID 89390
510(K)Number K011056  
Product Classification Peripheral mechanical thrombectomy with aspiration - Product Code QEW
Product Arrow-Trerotola Over-The-Wire PTD Kit, Percutaneous Thrombolytic Device, REF PT-65709-HFWC containing components: PT-65709-C, PT-03009, PT-65709-HFS

The Arrow-Trerotola Percutaneous Thrombolytic Device (PTD) Catheter, in conjunction with the Arrow Rotator Drive Unit, permits mechanical declotting of native arterio-venous (AV) fistulae and synthetic dialysis grafts.
Code Information Lot Number / UDI: 13F20A0323 / (01)108019022121292 (17)211231 (10)13F20A0323, Components: Code PT-65709-C, Lot number 13C19M0499; Code PT-65709-HFS, Lot number 14C19J0342; Code PT-03009, Lot number 13C19J1873; Code PT-03009, Lot number 13C19L0622; Code PT-03009, Lot number 13C19J0572; Code PT-03009, Lot number 13C19K0832; Lot Number / UDI: 13F20B0139 / (01)10801902121292 (17)220131 (10)13F20B0139, Components: Code PT-65709-C, Lot number 13C20A0970; Code PT-65709-HFS, Lot number 14C20A0611; Code PT-65709-HFS, Lot number 14C19G0671; Code PT-03009, Lot number 13C19L0622; Lot Number / UDI: 13F20C0594 / (01)10801902121292 (17)220331 (10)13F20C0594, Components: Code PT-65709-C, Lot number 13C20C2082; Code PT-65709-HFS, Lot number 14C20A0611; Code PT-65709-HFS, Lot number 14C19J0342; Code PT-03009, Lot number 13C20B1435; Lot Number / UDI: 13F20F008 3/ (01)10801902121292 (17)220531 (10)13F20F0083, Components: Code PT-65709-C, Lot number 13C20F0572; Code PT-65709-HFS, Lot number 14C20D1060; Code PT-65709-HFS, Lot number 14C20A0611; Code PT-03009, Lot number 13C20C1558; Code PT-03009, Lot number 13C20B1435; Lot Number / UDI: 13F20F0230 / (01)10801902121292 (17)220531 (10)13F20F0230, Components: Code PT-65709-C, Lot number 13C20F1139; Code PT-65709-HFS, Lot number 14C20F0586; Code PT-65709-HFS, Lot number 14C20D1060; Code PT-03009, Lot number 13C20F0570; Lot Number / UDI: 13F20F0577 / (01)10801902121292 (17)220630 (10)13F20F0577, Components: Code PT-65709-C, Lot number 13C20F2175; Code PT-65709-HFS, Lot number 14C20F0586; Code PT-03009, Lot number 13C20F1138; Code PT-03009, Lot number 13C20F0570; Lot Number / UDI: 13F20H0756 / (01)10801902121292 (17)220831 (10)13F20H0756, Components: Code PT-65709-C, Lot number 13C20G1976; Code PT-65709-HFS, Lot number 14C20F0274; Code PT-65709-HFS, Lot number 14C20F0586; Code PT-03009, Lot number 13C20H0604; Code PT-03009, Lot number 13C20G1198; Lot Number / UDI: 13F20K0849 / (01)10801902121292 (17)221031 (10)13F20K0849, Components: Code PT-65709-C, Lot number 13C20H1675; Code PT-65709-C, Lot number 13C20G1976; Code PT-65709-HFS, Lot number 14C20F0586; Code PT-65709-HFS, Lot number 14C20F0274; Code PT-03009, Lot number 13C20H0604; Lot Number / UDI: 13F20L0283 / (01)10801902121292 (17)221031 (10)13F20L0283, Components: Code PT-65709-C, Lot number 13C20H1675; Code PT-65709-C, Lot number 13C20J2689; Code PT-65709-HFS, Lot number 14C20J1446; Code PT-65709-HFS, Lot number 14C20F0586; Code PT-03009, Lot number 13C20L0872; Lot Number / UDI: 13F20M0182 / (01)10801902121292 (17)221231 (10)13F20M0182, Components: Code PT-65709-C, Lot number 13C20J2689; Code PT-65709-C, Lot number 13C20H1675; Code PT-65709-HFS, Lot number 14C20L0665; Code PT-65709-HFS, Lot number 14C20J1446; Code PT-03009, Lot number 13C20L0871; Lot Number / UDI: 13F21A0498 / (01)10801902121292 (17)221231 (10)13F21A0498, Components: Code PT-65709-C, Lot number 13C20M0639; Code PT-65709-HFS, Lot number 14C20J1446; Code PT-03009, Lot number 13C21A0855; Lot Number / UDI: 13F21B0158 / (01)10801902121292 (17)230131 (10)13F21B0158, Components: Code PT-65709-C, Lot number 13C21A1348; Code PT-65709-HFS, Lot number 14C20M0932; Code PT-65709-HFS, Lot number 14C20L0665; Code PT-65709-HFS, Lot number 14C20J1446; Code PT-03009, Lot number 13C21A0855; Lot Number / UDI: 13F21C0747 / (01)10801902121292 (17)230331 (10)13F21C0747, Components: Code PT-65709-C, Lot number 13C21C0326; Code PT-65709-C, Lot number 13C21A1348; Code PT-65709-HFS, Lot number 14C20L0665; Code PT-65709-HFS, Lot number 14C20M0932; Code PT-03009, Lot number 13C21B1533; Lot Number / UDI: 13F21E0555 / (01)108019021212292 (17)230430 (10)13F21E0555, Components: Code PT-65709-C, Lot number 13C21E0328; Code PT-65709-C, Lot number 13C21D0717; Code PT-65709-C, Lot number 13C21C2412; Code PT-65709-HFS, Lot number 14C21C0895; Code PT-65709-HFS, Lot number 14C20M0932; Code PT-03009, Lot number 13C21C2413; Code PT-03009, Lot number 13C21C2397
Recalling Firm/
Manufacturer
TELEFLEX MEDICAL INC
3015 Carrington Mill Blvd
Morrisville NC 27560-5437
Manufacturer Reason
for Recall
The firm received reports indicating PTD tip separation during use.
FDA Determined
Cause 2
Device Design
Action Arrow International LLC/Teleflex notified their consignees by letter on 01/11/2022. The letter explained the issue and requested that the consignee cease use and distribution. Distributors were directed to notify their customers. The firm is seeking return of the affected product. The following instructions were provided: Action list number 1 Medical facilities 1. We request that you check your inventory for product within the scope of this FSCA. Users should cease use and distribution of impacted product and quarantine immediately. 2. If you do not have stock in scope of this FSCA mark the according checkbox on the Acknowledgement Form (Appendix 1) and return the form to the fax number or e-mail address mentioned below. 3. If you do have stock in scope of this FSCA, complete the Acknowledgement Form (Appendix 1) and fax it to 1-855-419-8507, Attn: Customer Service or e-mail to recalls@teleflex.com. This will allow us to document the amount of product you have on hand for return. A customer service representative will contact you with a Return Goods Authorization (RGA) Number and will provide instructions for the return of products to Teleflex Medical. Action list number 2 Distributors 1. Provide this field safety notice to all customers who have received product in scope of this FSCA. Your customer is then required to complete the Acknowledgement Form and return to you. 2. We request that you check your inventory for product within the scope of this FSCA. Cease use and distribution of impacted product and quarantine immediately. You may then return all product in scope to Teleflex. 3. As a distributor, you are then required to confirm to Teleflex that you have completed the field activity outlined above. Upon completion of your actions, please forward the completed Acknowledgement Form to Customer Service. Important - Please ensure you only list batch numbers in scope of this Field Safety Notice when completing this form. 4. If you have further distr
Quantity in Commerce 1400 units
Distribution Worldwide distribution: US Nationwide and the countries of AU, BE, CA, CL, CZ, DE, ES, FR, GB, IL, IT, KR, KW, NL, PA, PH, PT, SA, TR.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database 510(K)s with Product Code = QEW and Original Applicant = ARROW INTL., INC.
-
-