• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall LYFO DISK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall LYFO DISKsee related information
Date Initiated by FirmOctober 01, 2020
Create DateFebruary 08, 2022
Recall Status1 Terminated 3 on July 07, 2022
Recall NumberZ-0583-2022
Recall Event ID 89391
510(K)NumberK861022 
Product Classification Kit, quality control for culture media - Product Code JTR
ProductLYFO DISK(TM), LYFO DISK(TM) Streptococcus pneumoniae derived from ATCC(R) 49619(TM). LYFO DISK microorganisms are packaged in a resealable vial that contains 6 lyophilized microorganism pellets and a desiccant to prevent adverse moisture accumulation. The LYFO DISK microorganisms are 3 passages or fewer from the reference culture and are guaranteed to recover when processed using the recommended media and incubation requirements.
Code Information Catalog Number: 0947L, Lot Number: 947-126-1, UDI: 10845357022964
FEI Number 2150138
Recalling Firm/
Manufacturer
Microbiologics Inc
200 Cooper Ave N
Saint Cloud MN 56303-4440
For Additional Information ContactBecky Neu
320-229-7073
Manufacturer Reason
for Recall
Potential contamination with Escherichia coli, Staphylococcus epidermidis and S. warneri.
FDA Determined
Cause 2
Unknown/Undetermined by firm
ActionMicrobiologics notified customers on about 10/01/2020 via letter titled "Urgent Medical Device Recall." Instructions included to inspect inventory for affected products. If found, instructions included the following: 1. REVIEW lab procedures to understand how this information affects your usage; 2. USE OR DISCARD depending on your lab procedures and how this information affects your usage; 3. COMPLETE the response form provided; 4. RETURN the response form to recall@microbiologics.com; 5. KEEP this letter for your records; 6. CONTACT Microbiologics if a replacement kit is needed. Questions or concerns can be directed to Recall Support team at 320.229.7073 or recall@microbiologics.com.
Quantity in Commerce42 units
DistributionWorldwide distribution - US Nationwide distribution in the states of AR, AZ, CA, CO, DE, FL, GA, IA, IL, IN, KS, LA, MA, MD, ME, MI, MN, MS, MT, NC, ND, NE, NM, NY, OH, OK, OR, PR, SC, TN, TX, VA, VT, WA, WV and the countries of AUSTRALIA, AUSTRIA, BAHRAIN, BELARUS, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, COLOMBIA, FRANCE, GEORGIA, GERMANY, GUATAMALA, HONG KONG, INDIA, INDONESIA, IRELAND, ITALY, JAPAN, JORDAN, LATVIA, MEXICO, NETHERLANDS, PERU, PHILIPPINES, POLAND, ROMANIA, SAUDI ARABIA, SOUTH AFRICA, SWITZERLAND, UNITED ARAB EMIRATES, VIETNAM.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.