• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall KWIKSTIK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall KWIKSTIKsee related information
Date Initiated by FirmOctober 06, 2014
Create DateMarch 03, 2022
Recall Status1 Terminated 3 on January 17, 2023
Recall NumberZ-0714-2022
Recall Event ID 89451
510(K)NumberK861022 
Product Classification Kit, quality control for culture media - Product Code JTR
ProductKWIK-STIK, Quality control kit for culture media, sold as: KWIK-STIK (single-pack) Enterococcus faecalis derived from ATCC 51299
Code Information Catalog Number HL163263 OEM Private Label with Healthlink; UDI: N/A (OEM agreement with Healthlink ended in Q1 2016 before UDI was required for class 1 IVDs) Lot Number (Expiration Date): 959-74-1 (10/08/2014), 959-75-2 (04/22/2015), 959-75-3 (04/22/2015), 959-75-6 (04/22/2015), 959-75-7 (04/22/2015), 959-75-10 (04/22/2015), and 959-76-2 (07/08/2015)
FEI Number 2150138
Recalling Firm/
Manufacturer
Microbiologics Inc
200 Cooper Ave N
Saint Cloud MN 56303-4440
For Additional Information ContactBecky Neu
320-229-7073
Manufacturer Reason
for Recall
This strain should be resistant to high levels of Gentamicin and be resistant to Gentamicin Synergy. A specific lot was found susceptible to Gentamicin Synergy during in-house testing.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionMicrobiologics notified customers on about 10/06/2014 via letter titled "E.faecalis Gentamicin Susceptibility Notification." Instructions included to inspect inventory for affected products. If customers have received the product identified within the recall, it is recommended that the intended use of the organism be reviewed to determine if Gentamicin synergy results have an impact on use. It is recommended to discard these lots if an impact is to be expected. Questions or concerns can be directed to the Technical Support team at 320.229.7064 or techsupport@microbiologics.com.
Quantity in Commerce150 units
DistributionWorldwide distribution - US Nationwide and the countries of Argentina, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia/Hrvatska, Ecuador, Finland, France, Germany, Guam, Guatemala, Hong Kong, India, Ireland, Israel, Italy, Japan, Kenya, Republic of Korea, Latvia, Macedonia, Malaysia, Netherlands, Panama, Peru, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, Singapore, South Africa, Switzerland, Taiwan, Tanzania, Thailand, Turkey, Uganda, United Arab Emirates, and Vietnam.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = JTR
-
-