| Class 2 Device Recall Atellica CH Fructosamine |  |
Date Initiated by Firm | February 09, 2022 |
Create Date | April 14, 2022 |
Recall Status1 |
Open3, Classified |
Recall Number | Z-0930-2022 |
Recall Event ID |
89740 |
510(K)Number | K131307 |
Product Classification |
Assay, glycosylated hemoglobin - Product Code LCP
|
Product | Atellica CH Fructosamine (Fruc), Siemens Material Number 11097637 |
Code Information |
UDI: 00630414595580;
Lot Numbers: 211359
211451 |
Recalling Firm/ Manufacturer |
Siemens Healthcare Diagnostics, Inc. 511 Benedict Ave Tarrytown NY 10591-5005
|
For Additional Information Contact | 914-631-8000 |
Manufacturer Reason for Recall | Siemens Healthcare Diagnostics Inc. has confirmed the potential for reagent carryover impacting quality control (QC), patient samples, and calibrator results. |
FDA Determined Cause 2 | Device Design |
Action | Urgent Medical Device Correction notification letters dated 2/9/22 were sent to customers.
Actions to be Taken by the Customer
Please review this letter with your Medical Director.
Perform the instructions provided in the Additional Information section below.
Complete and return the Field Correction Effectiveness Check Form attached to this letter
within 30 days.
If you have received any complaints of illness or adverse events associated with the
products listed in Table 1, immediately contact your local Siemens Healthineers Customer
Care Center or your local Siemens Healthineers technical support representative.
Please retain this letter with your laboratory records and forward this letter to those who may have received these products.
Additional Information
If your laboratory has multiple Atellica CH 930 Analyzers, separate the assays as follows:
-LITH_2 from CA_2 and TBil_2
-A1c_E/A1c_H from CA_2
- CA_2 from UN_c (urine only); UN_c serum and plasma are not impacted and can be run on the same analyzer as CA_2
-Fruc from AAT, IgA_2, IgM_2 and Theo
If you choose not to separate the assays as indicated above, batch testing of A1c_E, Fruc, LITH_2, and UN_c (urine only) may be considered.
Note: An RPC2 wash mitigation must be initiated after completion of A1c_E, Fruc, LITH_2, and
prior to UN_c (urine only). Any of the following will initiate the RPC2 wash:
-After the Atellica CH 930 Analyzer has been in standby for 12 minutes.
- After completion of any Open Channel assay.
- Restarting the Atellica CH 930 Analyzer. Refer to the Atellica Solution Online Help for
instructions on system restart.
If you have any questions, contact your local Siemens Healthineers technical support representative. |
Quantity in Commerce | 199 |
Distribution | Global Distribution. US Nationwide. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
510(K) Database | 510(K)s with Product Code = LCP
|
|
|
|