| Date Initiated by Firm | May 05, 2022 |
| Create Date | June 06, 2022 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1242-2022 |
| Recall Event ID |
90103 |
| 510(K)Number | K153778 |
| Product Classification |
Device, Vascular, for Promoting Embolization - Product Code KRD
|
| Product | Tornado Embolization Microcoil,
RPN GPN
MWCE-18S-5/2-TORNADO G08356
MWCE-18S-6/2-TORNADO G08259
MWCE-18S-3/2-TORNADO-01 G47416
MWCE-18S-4/2-TORNADO-01 G47417
MWCE-18S-3/2-TORNADO-081800 G13102
MWCE-18S-3/2-TORNADO G08261
MWCE-18S-4/2-TORNADO-081800 G13103
MWCE-18S-4/2-TORNADO G08357
MWCE-18S-5/2-TORNADO-01 G47418
MWCE-18S-6/2-TORNADO-01 G47419
MWCE-18S-6/2-TORNADO-081800 G13105
MWCE-18S-5/2-TORNADO-081800 G13104
MWCE-18S-2/4-TORNADO-LEF G09945
MWCE-18S-2/5-TORNADO-LEF-081800 G13108
MWCE-18S-4/3-TORNADO G12924
MWCE-18S-2/3-TORNADO-LEF G09944
MWCE-18S-2/3-TORNADO-LEF-081800 G13106
MWCE-18S-2/4-TORNADO-LEF-081800 G13107 |
| Code Information |
Lots:
14220596
14220940
14266927
14266928
14266929
14266930
14266931
14270156
14270688
NS14220942
14173005
14173006
14173009
14174725
14183341
14183343
14183344
14183345
14183346
14183347
14185759
14185762
14185763
14185764
14185765
NS14183348
14144260
14144261
14149518
14151428
14165309
14176799
14176801
14176802
14176804
14181260
14181262
14181263
14195340
14195341
14195342
14198726
14198727
14198728
14198729
14198730
14198731
14198732
14202396
14202397
14202398
14205449
NS14202553
14149487
14158173
14159388
14173007
14202382
14205450
14208005
14208006
14208007
14208008
14208009
14210749
14210750
14210751
NS14202380
NS14202381
14158174
14159387
14181258
14181259
14188217
14188363
14188366
14195343
14195344
14195345
14196666
14196669
14202556
14210752
14210753
14210754
14212795
14212796
14212797
14212798
14212800
14212801
14212802
14215747
NS14188216
NS14188218
NS14188219
NS14212799
14250738
14250742
14250743
14253965
14254863
14254864
14254865
14254866
14254867
14236671
14236672
14236673
14236674
14236675
14239466
14239467
14239468
14239469
14239470
14239471
14239472
14242269
14242270
14242271
14242272
NS14242268
14195265
14195266
14196667
14196668
14198997
14202554
14220941
14224113
14224114
14224115
14224116
14224117
14230375
14230376
14230379
14236670
14258545
14261429
14261430
14261431
14261432
14261433
14261434
14261435
NS14220945
NS14227791
14198998
14202555
14215746
14215748
14215749
14215750
14215751
14215752
14221686
14221687
14221688
14221689
14227792
14227793
14227794
14227795
14227796
14227797
14230373
14230374
14230377
14230378
NS14218742
NS14220943
NS14220946
NS14227798
NS14227798X
14221684
14221685
14221690
14224118
14242266
14242267
14244118
14244684
14244685
14244686
14244687
14244688
14244689
14244690
14248153
14250739
14250740
14250741
14254868
14258128
14258539
14258540
14258541
14258542
14258543
14258544
NS14258538
14220594
14220595
14273582
14273583
14273584
14297279
14297282
14232891
14273588
14276221
14276226
14276227
14297283
NS14273587
14411418
14414291
14414706
14414708
NS14414290
14414702
14414703
14414704
14414705
14414707
14420963
14420964
14420965
14420966
14236411
14276222
14276223
14276224
14289080
14289081
14289082
14289084
14289085
14294554
14236413
14307805
14307806
14307807
14307808
14307809
14307810
14311197
14311198
14311202
14315254
14236412
14237045
14237046
14237047
14239944
14248278
14320365
14320366
14320369
14320370
14285090
14288039
14297280
14302992
14302993
14302994
14302995
14302996
14302997
14307811
14311196
14311199
14311200
14311201
14315249
14315250
14315251
14315252
14315253
14315255
14289079
14289083
14294551
14294552
14294553
14294555
14294556
14294557
14248273
14248275
14250701
14327099
14327100
14327101
14327102
14327103
14327105
14327106
14329716
14329717
14329718
14329720
NS14329715
14248272
14248276
14329719
14336283
14336284
14336285
14336300
NS14329713
NS14329714
14320367
14320368
14323990
14323991
14323992
14323994
14323995
14323996
NS14323993
NS14327104
14266836
NS14324460
14339745
14339746
14343032
14343033
14343034
14343035
14343036
14343037
14343038
14343039
14320069
14345806
14345807
14345808
14345809
14345810
NS14324458
NS14345805
14312288
14345811
14345812
14354182
14354183
14354184
14354185
14354186
14354187
14356723
14357303
14363077
14363079
14363080
14363085
NS14363081
NS14363083
14357304
14357305
14357306
14357308
14357309
14360460
14360461
NS14357307
NS14360457
NS14360458
NS14360459
14336279
14336280
14336281
14336282
14339739
14339741
14339742
14339743
14339744
NS14339740
14318003
14322702
14325011
14326817
14363078
14363082
14363084
14377033
14377034
14377036
14377038
14377040
NS14377035
14328175
14360462
NS14360463
NS14360464
NS14360465
NS14370207
14377037
14377039
14381228
14381229
14381230
14381231
14381232
14381233
14381234
14385328
14385329
NS14381235
NS14381236
14338889
NS14359592
NS14370206
NS14370210
NS14370211
NS14370215
14388690
14388694
14388695
14388696
14388697
14394808
14394809
14394810
14394811
14394812
14394813
14394814
14329365
14331591
14338888
14397740
14397741
14397742
14397743
14385327
14385330
14385331
14385332
14385333
14388691
14388692
14388693
14397738
14397739
14397744
14402385
14402386
14402387
14405352
14405353
14405355
14405358
14411412
14411413
14411414
14411415
14411416
14411417
14370202
14370218
14371673
14371674
14420967
14420968
14420969
14424153
14424154
14424155
14424156
14424157
14424158
NS14386148
14448225
14448226
14448229
14448230
14451390
14451391
14451392
14451393
NS14448224
14370203
14373560
14378559
14383705
14424152
14370217
14383706
14387115
NS14373559
NS14380359
NS14386149
NS14386151
14386145
14387114
14392220
14393093
14393094
14394222
14397750
14397751
14448227
14448228
NS14448223
14378560
14387116
14401315
14406723
14411463
14411464
14413715
14451394
14451395
14451396
14453984
14453985
14453986
14453987
NS14439797
NS14452894
14413714
14420109
14420110
14420390
14427108
14427110
14461732
14461733
14461734
14461735
14461736
14427109
14427112
14461731
14464701
14464702
14464703
14464704
14464705
NS14418655
14420307
14427111
14427805
14433110
14433111
14434831
14437153
14437155
14437157
14437158
14437159
14437161
14437164
14446798
14464699
14464700
14467340
14467341
14467342
14467343
14467344
14401316
14405013
14406148
14408726
14410622
14418656
14456814
14456815
14456818
14461737
NS14410623
14434592
14445421
14492887
14495701
14495702
14495704
14495705
14431474
14469248
14469249
14469250
14469251
14469252
14483961
14483962
14486089
14486090
14486091
14486092
14486093
14486094
14486095
14492881
14492882
14442307
14450412
14454129
14420108
14433350
14434593
14435952
14438638
14442308
14478197
14478198
14478199
14478200
14478203
14483960
14483963
14483964
14483965
NS14483959
14439798
14446796
14467338
14467339
14469246
14469247
14492883
14492885
14492886
NS14485428
14446460
14448149
14498633
14498634
14498635
14498636
14498637
14501260
14501261
14501262
14501263
NS14506088
NS14506089
14397752
14438637
14440486
14440487
14440488
14440489
14442309
14451724
NS14443148
NS14446800
NS14446802
14446795
14450415
14451723
14452397
14452725
14452726
14478373
14485425
14486891
14488153
NS14511586
NS14527916
14457696
14459429
14462215
14511226
14513982
14513983
14513984
14513985
14513986
14513987
14513988
14531276
14534420
14534421
14462483
14464721
14466807
14466808
14468459
14468460
14531272
14531273
14531274
14531275
14485427
14485739
14506231
14542456
14548291
14548292
14548293
14553648
NS14518614
NS14518616
14550801
14553650
14553651
14553653
14553654
14553655
14556278
NS14553649
14506918
14513206
14530073
14556279
14556280
14472102
14473095
14534422
14534423
14534424
14534425
NS14534419
14472630
14473854
14474275
14474280
14536686
14536687
14536688
14536689
14536690
14536691
14536692
14539884
14474281
14485426
14539881
14539883
14539885
14539886
14539887
14539888
14542450
14542451
14542452
14542454
NS14539882
NS14542453
NS14542457
14474024
14474025
14474026
14474027
14474028
14474029
14474030
14478201
NS14478202
14270688
14270689
14270690
14270691
14270692
14270693
14276225
14285095
NS14290291
14454927
14457695
14506082
14506085
14511228
14511229
14511230
14511232
14580290
14489161
NS14518615
NS14572569
14577723
14577724
14577725
14577726
NS14577727
NS14579799
14506919
14518611
14521588
NS14527915
NS14536966
14511585
14514285
14518612
14518613
14521591
14523439
14527166
14530074
14539391
14539392
NS14521590
14527914
14531996
14534605
14536965
14542082
14547243
14551339
NS14573819
14555678
14597775
14597777
14602376
14602378
14604806
NS14613686
14232885
14232891
14273585
14273586
14273589
14285091
14285092
14285093
14285094
14297278
14297281
14297284
NS14232893
NS14232896
NS14239945
NS14248274
NS14248279
NS14250702
NS14276228
14359593
14359594
14370216
14402380
14402381
14402382
14402383
14402384
14405353
14405354
14405356
14405357
NS14370204
14453982
14453983
14453984
14453988
14456812
14456813
14456814
14456816
14456817
14431473
14433351
14434539
14434540
14434594
14438131
14438704
14438705
14442308
UDI:
(01)00827002083561(17)261014(10)14220596
(01)00827002082595(17)261019(10)14220940
(01)00827002474161(17)261024(10)14266927
(01)00827002474161(17)261024(10)14266928
(01)00827002474178(17)261025(10)14266929
(01)00827002474161(17)261024(10)14266930
(01)00827002131026(17)261024(10)14266931
(01)00827002082618(17)261025(10)14270156
(01)00827002474192(17)261025(10)14270688
(01)00827002083561(17)261013(10)NS14220942
(01)00827002082618(17)260917(10)14173005
(01)00827002082618(17)260928(10)14173006
(01)00827002131026(17)260909(10)14173009
(01)00827002131026(17)260909(10)14174725
(01)00827002474161(17)260909(10)14183341
(01)00827002474161(17)260923(10)14183343
(01)00827002082618(17)260909(10)14183344
(01)00827002474161(17)260923(10)14183345
(01)00827002474161(17)260917(10)14183346
(01)00827002131033(17)260909(10)14183347
(01)00827002082618(17)260917(10)14185759
(01)00827002083578(17)260909(10)14185762
(01)00827002082618(17)260926(10)14185763
(01)00827002082618(17)260915(10)14185764
(01)00827002082618(17)260909(10)14185765
(01)00827002083578(17)260909(10)NS14183348
(01)00827002083561(17)260914(10)14144260
(01)00827002083561(17)260914(10)14144261
(01)00827002474185(17)260914(10)14149518
(01)00827002083561(17)260914(10)14151428
(01)00827002474185(17)260914(10)14165309
(01)00827002474185(17)260929(10)14176799
(01)00827002083561(17)260930(10)14176801
(01)00827002083561(17)260926(10)14176802
(01)00827002474185(17)260916(10)14176804
(01)00827002083561(17)260930(10)14181260
(01)00827002083561(17)260926(10)14181262
(01)00827002083561(17)260915(10)14181263
(01)00827002474192(17)260916(10)14195340
(01)00827002082595(17)260926(10)14195341
(01)00827002082595(17)260930(10)14195342
(01)00827002083561(17)260909(10)14198726
(01)00827002082595(17)260923(10)14198727
(01)00827002082595(17)260914(10)14198728
(01)00827002474185(17)260928(10)14198729
(01)00827002083561(17)260914(10)14198730
(01)00827002082595(17)260930(10)14198731
(01)00827002082595(17)260909(10)14198732
(01)00827002083561(17)260930(10)14202396
(01)00827002083561(17)260930(10)14202397
(01)00827002083561(17)260914(10)14202398
(01)00827002082595(17)260930(10)14205449
(01)00827002083578(17)260914(10)NS14202553
(01)00827002082595(17)260914(10)14149487
(01)00827002082595(17)260910(10)14158173
(01)00827002082595(17)260914(10)14159388
(01)00827002082618(17)260909(10)14173007
(01)00827002083578(17)260929(10)14202382
(01)00827002083561(17)260926(10)14205450
(01)00827002082595(17)260930(10)14208005
(01)00827002082595(17)260930(10)14208006
(01)00827002082595(17)260923(10)14208007
(01)00827002474192(17)260923(10)14208008
(01)00827002131057(17)260923(10)14208009
(01)00827002474185(17)260928(10)14210749
(01)00827002474185(17)260923(10)14210750
(01)00827002131040(17)260917(10)14210751
(01)00827002082618(17)260917(10)NS14202380
(01)00827002082618(17)260923(10)NS14202381
(01)00827002082595(17)260916(10)14158174
(01)00827002082595(17)260915(10)14159387
(01)00827002082595(17)260930(10)14181258
(01)00827002082595(17)260926(10)14181259
(01)00827002083561(17)260917(10)14188217
(01)00827002082595(17)260922(10)14188363
(01)00827002082595(17)260929(10)14188366
(01)00827002082595(17)260930(10)14195343
(01)00827002082595(17)260923(10)14195344
(01)00827002474192(17)260917(10)14195345
(01)00827002099456(17)260924(10)14196666
(01)00827002099456(17)260923(10)14196669
(01)00827002131088(17)260926(10)14202556
(01)00827002083561(17)260930(10)14210752
(01)00827002083561(17)260926(10)14210753
(01)00827002474185(17)260928(10)14210754
(01)00827002131026(17)260917(10)14212795
(01)00827002082618(17)260917(10)14212796
(01)00827002474178(17)260917(10)14212797
(01)00827002083578(17)260930(10)14212798
(01)00827002083578(17)260928(10)14212800
(01)00827002131033(17)260926(10)14212801
(01)00827002083578(17)260926(10)14212802
(01)00827002474178(17)260917(10)14215747
(01)00827002083561(17)260917(10)NS14188216
(01)00827002083561(17)260914(10)NS14188218
(01)00827002083561(17)260916(10)NS14188219
(01)00827002129245(17)260916(10)NS14212799
(01)00827002082595(17)261102(10)14250738
(01)00827002083561(17)261102(10)14250742
(01)00827002083561(17)261102(10)14250743
(01)00827002082595(17)261102(10)14253965
(01)00827002131057(17)261024(10)14254863
(01)00827002082595(17)261102(10)14254864
(01)00827002082595(17)261102(10)14254865
(01)00827002474192(17)261014(10)14254866
(01)00827002082595(17)261014(10)14254867
(01)00827002082595(17)261025(10)14236671
(01)00827002082595(17)261021(10)14236672
(01)00827002083561(17)261014(10)14236673
(01)00827002083561(17)261022(10)14236674
(01)00827002083561(17)261014(10)14236675
(01)00827002083578(17)261022(10)14239466
(01)00827002083578(17)261022(10)14239467
(01)00827002082618(17)261024(10)14239468
(01)00827002083578(17)261022(10)14239469
(01)00827002474161(17)261022(10)14239470
(01)00827002474161(17)261022(10)14239471
(01)00827002082618(17)261022(10)14239472
(01)00827002474178(17)261022(10)14242269
(01)00827002131033(17)261022(10)14242270
(01)00827002083578(17)261029(10)14242271
(01)00827002083578(17)261029(10)14242272
(01)00827002082618(17)261024(10)NS14242268
(01)00827002083561(17)260930(10)14195265
(01)00827002083561(17)260930(10)14195266
(01)00827002099449(17)260926(10)14196667
(01)00827002099449(17)260926(10)14196668
(01)00827002474185(17)261012(10)14198997
(01)00827002131064(17)260928(10)14202554
(01)00827002083561(17)261020(10)14220941
(01)00827002082595(17)260926(10)14224113
(01)00827002082595(17)260926(10)14224114
(01)00827002083561(17)260930(10)14224115
(01)00827002083561(17)260929(10)14224116
(01)00827002083561(17)260924(10)14224117
(01)00827002082618(17)261024(10)14230375
(01)00827002082618(17)261021(10)14230376
(01)00827002083578(17)261020(10)14230379
(01)00827002082595(17)261014(10)14236670
(01)00827002474161(17)261014(10)14258545
(01)00827002474178(17)261014(10)14261429
(01)00827002474161(17)261014(10)14261430
(01)00827002131026(17)261024(10)14261431
(01)00827002082618(17)261102(10)14261432
(01)00827002082618(17)261102(10)14261433
(01)00827002083578(17)261019(10)14261434
(01)00827002083578(17)261013(10)14261435
(01)00827002083561(17)261011(10)NS14220945
(01)00827002129245(17)261013(10)NS14227791
(01)00827002131040(17)261011(10)14198998
(01)00827002131071(17)261012(10)14202555
(01)00827002474161(17)260928(10)14215746
(01)00827002083578(17)260926(10)14215748
(01)00827002083578(17)260926(10)14215749
(01)00827002083578(17)260923(10)14215750
(01)00827002474161(17)260923(10)14215751
(01)00827002474161(17)260928(10)14215752
(01)00827002082618(17)260929(10)14221686
(01)00827002131033(17)260928(10)14221687
(01)00827002083578(17)260928(10)14221688
(01)00827002129245(17)260928(10)14221689
(01)00827002131026(17)261020(10)14227792
(01)00827002131026(17)261020(10)14227793
(01)00827002082618(17)261020(10)14227794
(01)00827002082618(17)261020(10)14227795
(01)00827002474178(17)261014(10)14227796
(01)00827002083578(17)261020(10)14227797
(01)00827002083578(17)261021(10)14230373
(01)00827002083578(17)261021(10)14230374
(01)00827002083578(17)261021(10)14230377
(01)00827002083578(17)261019(10)14230378
(01)00827002082595(17)261012(10)NS14218742
(01)00827002082595(17)261011(10)NS14220943
(01)00827002082595(17)261011(10)NS14220946
(01)00827002083578(17)261014(10)NS14227798
(01)00827002083578(17)261014(10)NS14227798X
(01)00827002131026(17)260926(10)14221684
(01)00827002131026(17)260923(10)14221685
(01)00827002474161(17)260926(10)14221690
(01)00827002474192(17)260926(10)14224118
(01)00827002082618(17)261024(10)14242266
(01)00827002082618(17)261029(10)14242267
(01)00827002474161(17)261022(10)14244118
(01)00827002474161(17)261022(10)14244684
(01)00827002082618(17)261029(10)14244685
(01)00827002082618(17)261102(10)14244686
(01)00827002082618(17)261102(10)14244687
(01)00827002474161(17)261022(10)14244688
(01)00827002474178(17)261022(10)14244689
(01)00827002083578(17)261022(10)14244690
(01)00827002083578(17)261022(10)14248153
(01)00827002082595(17)261102(10)14250739
(01)00827002082595(17)261102(10)14250740
(01)00827002082595(17)261024(10)14250741
(01)00827002082595(17)261019(10)14254868
(01)00827002082595(17)261019(10)14258128
(01)00827002474161(17)261014(10)14258539
(01)00827002474161(17)261014(10)14258540
(01)00827002131026(17)261014(10)14258541
(01)00827002082618(17)261008(10)14258542
(01)00827002082618(17)261014(10)14258543
(01)00827002129245(17)261008(10)14258544
(01)00827002083578(17)261013(10)NS14258538
(01)00827002099449(17)261101(10)14220594
(01)00827002131064(17)261102(10)14220595
(01)00827002082618(17)261027(10)14273582
(01)00827002082618(17)261029(10)14273583
(01)00827002082618(17)261029(10)14273584
(01)00827002082618(17)261102(10)14297279
(01)00827002082618(17)261102(10)14297282
(01)00827002083561(17)261101(10)14232891
(01)00827002474178(17)261102(10)14273588
(01)00827002082618(17)261027(10)14276221
(01)00827002131033(17)261029(10)14276226
(01)00827002083578(17)261027(10)14276227
(01)00827002082618(17)261103(10)14297283
(01)00827002082618(17)261027(10)NS14273587
(01)00827002131026(17)261216(10)14411418
(01)00827002083578(17)261214(10)14414291
(01)00827002131026(17)261221(10)14414706
(01)00827002131026(17)270103(10)14414708
(01)00827002082618(17)261214(10)NS14414290
(01)00827002131033(17)261221(10)14414702
(01)00827002131033(17)261221(10)14414703
(01)00827002131033(17)261221(10)14414704
(01)00827002131026(17)261220(10)14414705
(01)00827002131033(17)270103(10)14414707
(01)00827002082618(17)261229(10)14420963
(01)00827002474161(17)261228(10)14420964
(01)00827002474161(17)261229(10)14420965
(01)00827002082618(17)261229(10)14420966
(01)00827002083561(17)261101(10)14236411
(01)00827002082618(17)261028(10)14276222
(01)00827002082618(17)261029(10)14276223
(01)00827002082618(17)261029(10)14276224
(01)00827002083578(17)261027(10)14289080
(01)00827002083578(17)261027(10)14289081
(01)00827002082618(17)261029(10)14289082
(01)00827002083578(17)261028(10)14289084
(01)00827002083578(17)261029(10)14289085
(01)00827002082618(17)261102(10)14294554
(01)00827002131040(17)261102(10)14236413
(01)00827002082618(17)261104(10)14307805
(01)00827002083578(17)261108(10)14307806
(01)00827002082618(17)261108(10)14307807
(01)00827002082618(17)261110(10)14307808
(01)00827002083578(17)261104(10)14307809
(01)00827002082618(17)261111(10)14307810
(01)00827002082618(17)261110(10)14311197
(01)00827002082618(17)261111(10)14311198
(01)00827002131033(17)261107(10)14311202
(01)00827002083578(17)261103(10)14315254
(01)00827002083561(17)261102(10)14236412
(01)00827002083561(17)261108(10)14237045
(01)00827002474185(17)261107(10)14237046
(01)00827002083561(17)261103(10)14237047
(01)00827002083561(17)261104(10)14239944
(01)00827002082595(17)261104(10)14248278
(01)00827002082595(17)261104(10)14320365
(01)00827002082595(17)261111(10)14320366
(01)00827002083561(17)261111(10)14320369
(01)00827002082595(17)261111(10)14320370
(01)00827002083561(17)261101(10)14285090
(01)00827002083561(17)261101(10)14288039
(01)00827002083578(17)261101(10)14297280
(01)00827002083561(17)261108(10)14302992
(01)00827002083561(17)261108(10)14302993
(01)00827002083561(17)261108(10)14302994
(01)00827002131057(17)261104(10)14302995
(01)00827002082595(17)261104(10)14302996
(01)00827002082595(17)261104(10)14302997
(01)00827002083578(17)261104(10)14307811
(01)00827002083578(17)261111(10)14311196
(01)00827002083578(17)261111(10)14311199
(01)00827002082618(17)261104(10)14311200
(01)00827002082618(17)261103(10)14311201
(01)00827002131026(17)261110(10)14315249
(01)00827002082618(17)261104(10)14315250
(01)00827002082618(17)261103(10)14315251
(01)00827002082618(17)261103(10)14315252
(01)00827002083578(17)261103(10)14315253
(01)00827002082618(17)261104(10)14315255
(01)00827002082618(17)261101(10)14289079
(01)00827002082618(17)261102(10)14289083
(01)00827002082618(17)261102(10)14294551
(01)00827002082618(17)261101(10)14294552
(01)00827002083578(17)261101(10)14294553
(01)00827002082618(17)261101(10)14294555
(01)00827002083578(17)261102(10)14294556
(01)00827002083578(17)261029(10)14294557
(01)00827002082595(17)261104(10)14248273
(01)00827002082595(17)261108(10)14248275
(01)00827002474185(17)261111(10)14250701
(01)00827002474161(17)261104(10)14327099
(01)00827002082618(17)261104(10)14327100
(01)00827002082618(17)261104(10)14327101
(01)00827002082618(17)261104(10)14327102
(01)00827002083578(17)261104(10)14327103
(01)00827002082618(17)261104(10)14327105
(01)00827002082618(17)261104(10)14327106
(01)00827002474161(17)261107(10)14329716
(01)00827002083578(17)261104(10)14329717
(01)00827002082618(17)261107(10)14329718
(01)00827002083578(17)261104(10)14329720
(01)00827002082618(17)261107(10)NS14329715
(01)00827002131071(17)261107(10)14248272
(01)00827002099449(17)261107(10)14248276
(01)00827002082618(17)261115(10)14329719
(01)00827002082618(17)261111(10)14336283
(01)00827002083578(17)261109(10)14336284
(01)00827002083578(17)261109(10)14336285
(01)00827002082618(17)261109(10)14336300
(01)00827002082618(17)261108(10)NS14329713
(01)00827002082618(17)261111(10)NS14329714
(01)00827002131040(17)261109(10)14320367
(01)00827002083561(17)261104(10)14320368
(01)00827002083578(17)261103(10)14323990
(01)00827002083578(17)261103(10)14323991
(01)00827002082618(17)261103(10)14323992
(01)00827002129245(17)261103(10)14323994
(01)00827002474161(17)261103(10)14323995
(01)00827002131026(17)261103(10)14323996
(01)00827002082618(17)261104(10)NS14323993
(01)00827002083578(17)261104(10)NS14327104
(01)00827002474185(17)261110(10)14266836
(01)00827002083561(17)261111(10)NS14324460
(01)00827002131026(17)261203(10)14339745
(01)00827002082618(17)261111(10)14339746
(01)00827002083578(17)261129(10)14343032
(01)00827002083578(17)261206(10)14343033
(01)00827002082618(17)261201(10)14343034
(01)00827002082618(17)261115(10)14343035
(01)00827002082618(17)261208(10)14343036
(01)00827002131033(17)261202(10)14343037
(01)00827002131026(17)261203(10)14343038
(01)00827002082618(17)261115(10)14343039
(01)00827002083561(17)261201(10)14320069
(01)00827002474178(17)261112(10)14345806
(01)00827002083578(17)261112(10)14345807
(01)00827002474161(17)261111(10)14345808
(01)00827002082618(17)261111(10)14345809
(01)00827002082618(17)261115(10)14345810
(01)00827002082595(17)261124(10)NS14324458
(01)00827002082618(17)261206(10)NS14345805
(01)00827002083561(17)261123(10)14312288
(01)00827002083578(17)261115(10)14345811
(01)00827002083578(17)261115(10)14345812
(01)00827002083561(17)261203(10)14354182
(01)00827002083561(17)261130(10)14354183
(01)00827002082595(17)261206(10)14354184
(01)00827002082595(17)261129(10)14354185
(01)00827002083561(17)261206(10)14354186
(01)00827002083561(17)261201(10)14354187
(01)00827002082595(17)261201(10)14356723
(01)00827002131026(17)261203(10)14357303
(01)00827002474161(17)261201(10)14363077
(01)00827002474178(17)261129(10)14363079
(01)00827002129245(17)261124(10)14363080
(01)00827002083578(17)261201(10)14363085
(01)00827002129245(17)261201(10)NS14363081
(01)00827002083578(17)261201(10)NS14363083
(01)00827002474161(17)261203(10)14357304
(01)00827002474178(17)261203(10)14357305
(01)00827002131033(17)261202(10)14357306
(01)00827002474161(17)261201(10)14357308
(01)00827002082618(17)261130(10)14357309
(01)00827002474161(17)261203(10)14360460
(01)00827002131026(17)261203(10)14360461
(01)00827002129245(17)261203(10)NS14357307
(01)00827002083578(17)261207(10)NS14360457
(01)00827002083578(17)261201(10)NS14360458
(01)00827002129245(17)261202(10)NS14360459
(01)00827002082618(17)261111(10)14336279
(01)00827002082618(17)261111(10)14336280
(01)00827002082618(17)261112(10)14336281
(01)00827002082618(17)261112(10)14336282
(01)00827002083578(17)261112(10)14339739
(01)00827002082618(17)261207(10)14339741
(01)00827002082618(17)261112(10)14339742
(01)00827002083578(17)261201(10)14339743
(01)00827002474161(17)261112(10)14339744
(01)00827002083578(17)261112(10)NS14339740
(01)00827002083561(17)261201(10)14318003
(01)00827002083561(17)261129(10)14322702
(01)00827002083561(17)261210(10)14325011
(01)00827002083561(17)261203(10)14326817
(01)00827002082618(17)261203(10)14363078
(01)00827002474161(17)261130(10)14363082
(01)00827002082618(17)261130(10)14363084
(01)00827002083578(17)261210(10)14377033
(01)00827002083578(17)261210(10)14377034
(01)00827002082618(17)261210(10)14377036
(01)00827002083578(17)261215(10)14377038
(01)00827002082618(17)261215(10)14377040
(01)00827002083578(17)261210(10)NS14377035
(01)00827002082595(17)261214(10)14328175
(01)00827002131026(17)261130(10)14360462
(01)00827002082618(17)261202(10)NS14360463
(01)00827002082618(17)261203(10)NS14360464
(01)00827002082618(17)261129(10)NS14360465
(01)00827002082595(17)261227(10)NS14370207
(01)00827002083578(17)261210(10)14377037
(01)00827002083578(17)261210(10)14377039
(01)00827002083578(17)261209(10)14381228
(01)00827002083578(17)261208(10)14381229
(01)00827002131033(17)261208(10)14381230
(01)00827002131026(17)261210(10)14381231
(01)00827002082618(17)261209(10)14381232
(01)00827002082618(17)261209(10)14381233
(01)00827002083578(17)261210(10)14381234
(01)00827002131026(17)261208(10)14385328
(01)00827002082618(17)261209(10)14385329
(01)00827002083578(17)261210(10)NS14381235
(01)00827002083578(17)261210(10)NS14381236
(01)00827002082595(17)261214(10)14338889
(01)00827002083561(17)261207(10)NS14359592
(01)00827002083561(17)261210(10)NS14370206
(01)00827002083561(17)261209(10)NS14370210
(01)00827002082595(17)261210(10)NS14370211
(01)00827002083561(17)261210(10)NS14370215
(01)00827002083578(17)261215(10)14388690
(01)00827002082618(17)261209(10)14388694
(01)00827002082618(17)261209(10)14388695
(01)00827002082618(17)261209(10)14388696
(01)00827002474178(17)261216(10)14388697
(01)00827002131026(17)261215(10)14394808
(01)00827002131033(17)261214(10)14394809
(01)00827002131026(17)261215(10)14394810
(01)00827002082618(17)261215(10)14394811
(01)00827002082618(17)261210(10)14394812
(01)00827002082618(17)261215(10)14394813
(01)00827002082618(17)261210(10)14394814
(01)00827002082595(17)261206(10)14329365
(01)00827002082595(17)261210(10)14331591
(01)00827002082595(17)261212(10)14338888
(01)00827002474161(17)261215(10)14397740
(01)00827002131026(17)261214(10)14397741
(01)00827002082618(17)261215(10)14397742
(01)00827002082618(17)261217(10)14397743
(01)00827002083578(17)261208(10)14385327
(01)00827002131033(17)261209(10)14385330
(01)00827002083578(17)261210(10)14385331
(01)00827002083578(17)261209(10)14385332
(01)00827002082618(17)261210(10)14385333
(01)00827002474161(17)261210(10)14388691
(01)00827002474161(17)261209(10)14388692
(01)00827002131026(17)261210(10)14388693
(01)00827002083578(17)261216(10)14397738
(01)00827002083578(17)261214(10)14397739
(01)00827002474178(17)261215(10)14397744
(01)00827002129245(17)261214(10)14402385
(01)00827002082618(17)261215(10)14402386
(01)00827002082618(17)261215(10)14402387
(01)00827002083578(17)261215(10)14405352
(01)00827002082618(17)261216(10)14405353
(01)00827002083578(17)261216(10)14405355
(01)00827002131033(17)261215(10)14405358
(01)00827002131026(17)261215(10)14411412
(01)00827002474161(17)261215(10)14411413
(01)00827002131026(17)261215(10)14411414
(01)00827002131026(17)261228(10)14411415
(01)00827002131026(17)261228(10)14411416
(01)00827002131026(17)261229(10)14411417
(01)00827002131064(17)261220(10)14370202
(01)00827002131071(17)261229(10)14370218
(01)00827002474185(17)261229(10)14371673
(01)00827002131040(17)261229(10)14371674
(01)00827002082618(17)261229(10)14420967
(01)00827002474178(17)261217(10)14420968
(01)00827002083578(17)261216(10)14420969
(01)00827002082618(17)261220(10)14424153
(01)00827002082618(17)261229(10)14424154
(01)00827002082618(17)261229(10)14424155
(01)00827002083578(17)261228(10)14424156
(01)00827002082618(17)261221(10)14424157
(01)00827002082618(17)261229(10)14424158
(01)00827002083561(17)261228(10)NS14386148
(01)00827002474161(17)270119(10)14448225
(01)00827002082618(17)270119(10)14448226
(01)00827002474178(17)270119(10)14448229
(01)00827002083578(17)270119(10)14448230
(01)00827002082618(17)270119(10)14451390
(01)00827002082618(17)270121(10)14451391
(01)00827002082618(17)270121(10)14451392
(01)00827002082618(17)270119(10)14451393
(01)00827002083578(17)270120(10)NS14448224
(01)00827002131064(17)261229(10)14370203
(01)00827002083561(17)261229(10)14373560
(01)00827002082595(17)270105(10)14378559
(01)00827002082595(17)270103(10)14383705
(01)00827002083578(17)261217(10)14424152
(01)00827002099456(17)270106(10)14370217
(01)00827002082595(17)270104(10)14383706
(01)00827002474185(17)270104(10)14387115
(01)00827002083561(17)270104(10)NS14373559
(01)00827002082595(17)270104(10)NS14380359
(01)00827002083561(17)270104(10)NS14386149
(01)00827002083561(17)261228(10)NS14386151
(01)00827002083561(17)270104(10)14386145
(01)00827002083561(17)270104(10)14387114
(01)00827002131064(17)270106(10)14392220
(01)00827002131040(17)270105(10)14393093
(01)00827002083561(17)270105(10)14393094
(01)00827002083561(17)270106(10)14394222
(01)00827002083561(17)270105(10)14397750
(01)00827002082595(17)270106(10)14397751
(01)00827002474161(17)270119(10)14448227
(01)00827002474161(17)270119(10)14448228
(01)00827002082618(17)270119(10)NS14448223
(01)00827002474192(17)270107(10)14378560
(01)00827002474185(17)270105(10)14387116
(01)00827002082595(17)270113(10)14401315
(01)00827002082595(17)270105(10)14406723
(01)00827002083561(17)270118(10)14411463
(01)00827002083561(17)270117(10)14411464
(01)00827002083561(17)270117(10)14413715
(01)00827002083578(17)270121(10)14451394
(01)00827002083578(17)270121(10)14451395
(01)00827002129245(17)270120(10)14451396
(01)00827002082618(17)270121(10)14453984
(01)00827002082618(17)270124(10)14453985
(01)00827002082618(17)270119(10)14453986
(01)00827002082618(17)270124(10)14453987
(01)00827002082595(17)270120(10)NS14439797
(01)00827002082618(17)270105(10)NS14452894
(01)00827002474185(17)270113(10)14413714
(01)00827002082595(17)270119(10)14420109
(01)00827002082595(17)270117(10)14420110
(01)00827002131057(17)270118(10)14420390
(01)00827002083561(17)270119(10)14427108
(01)00827002083561(17)270117(10)14427110
(01)00827002082618(17)270119(10)14461732
(01)00827002082618(17)270119(10)14461733
(01)00827002082618(17)270120(10)14461734
(01)00827002082618(17)270124(10)14461735
(01)00827002082618(17)270121(10)14461736
(01)00827002083561(17)270119(10)14427109
(01)00827002131040(17)270118(10)14427112
(01)00827002082618(17)270120(10)14461731
(01)00827002082618(17)270123(10)14464701
(01)00827002082618(17)270125(10)14464702
(01)00827002082618(17)270121(10)14464703
(01)00827002082618(17)270121(10)14464704
(01)00827002082618(17)270128(10)14464705
(01)00827002082595(17)270113(10)NS14418655
(01)00827002131064(17)270118(10)14420307
(01)00827002083561(17)270117(10)14427111
(01)00827002131064(17)270118(10)14427805
(01)00827002083561(17)270119(10)14433110
(01)00827002474185(17)270118(10)14433111
(01)00827002131088(17)270117(10)14434831
(01)00827002131071(17)270120(10)14437153
(01)00827002099456(17)270121(10)14437155
(01)00827002131071(17)270128(10)14437157
(01)00827002131088(17)270125(10)14437158
(01)00827002099449(17)270123(10)14437159
(01)00827002099456(17)270121(10)14437161
(01)00827002131088(17)270121(10)14437164
(01)00827002099456(17)270125(10)14446798
(01)00827002082618(17)270121(10)14464699
(01)00827002082618(17)270123(10)14464700
(01)00827002131026(17)270123(10)14467340
(01)00827002082618(17)270119(10)14467341
(01)00827002082618(17)270125(10)14467342
(01)00827002082618(17)270125(10)14467343
(01)00827002131026(17)270125(10)14467344
(01)00827002082595(17)270113(10)14401316
(01)00827002083561(17)270113(10)14405013
(01)00827002083561(17)270113(10)14406148
(01)00827002082595(17)270117(10)14408726
(01)00827002083561(17)270117(10)14410622
(01)00827002474192(17)270113(10)14418656
(01)00827002083578(17)270121(10)14456814
(01)00827002083578(17)270121(10)14456815
(01)00827002083578(17)270119(10)14456818
(01)00827002082618(17)270121(10)14461737
(01)00827002082595(17)270112(10)NS14410623
(01)00827002474192(17)270124(10)14434592
(01)00827002474185(17)270206(10)14445421
(01)00827002083578(17)270211(10)14492887
(01)00827002083578(17)270211(10)14495701
(01)00827002082618(17)270211(10)14495702
(01)00827002474178(17)270217(10)14495704
(01)00827002474161(17)270217(10)14495705
(01)00827002082595(17)270120(10)14431474
(01)00827002083578(17)270121(10)14469248
(01)00827002083578(17)270120(10)14469249
(01)00827002083578(17)270120(10)14469250
(01)00827002131033(17)270124(10)14469251
(01)00827002083578(17)270124(10)14469252
(01)00827002474178(17)270128(10)14483961
(01)00827002083578(17)270125(10)14483962
(01)00827002129245(17)270202(10)14486089
(01)00827002083578(17)270201(10)14486090
(01)00827002474161(17)270131(10)14486091
(01)00827002083578(17)270202(10)14486092
(01)00827002083578(17)270207(10)14486093
(01)00827002083578(17)270207(10)14486094
(01)00827002474161(17)270202(10)14486095
(01)00827002083578(17)270215(10)14492881
(01)00827002083578(17)270215(10)14492882
(01)00827002083561(17)270202(10)14442307
(01)00827002083561(17)270208(10)14450412
(01)00827002083561(17)270208(10)14454129
(01)00827002082595(17)270128(10)14420108
(01)00827002082595(17)270123(10)14433350
(01)00827002082595(17)270121(10)14434593
(01)00827002082595(17)270127(10)14435952
(01)00827002082595(17)270201(10)14438638
(01)00827002474192(17)270127(10)14442308
(01)00827002474161(17)270124(10)14478197
(01)00827002474161(17)270127(10)14478198
(01)00827002474161(17)270128(10)14478199
(01)00827002474178(17)270125(10)14478200
(01)00827002083578(17)270125(10)14478203
(01)00827002474161(17)270131(10)14483960
(01)00827002474161(17)270125(10)14483963
(01)00827002474161(17)270125(10)14483964
(01)00827002474178(17)270125(10)14483965
(01)00827002083578(17)270131(10)NS14483959
(01)00827002082595(17)270206(10)14439798
(01)00827002083561(17)270206(10)14446796
(01)00827002082618(17)270119(10)14467338
(01)00827002082618(17)270120(10)14467339
(01)00827002083578(17)270119(10)14469246
(01)00827002083578(17)270121(10)14469247
(01)00827002083578(17)270217(10)14492883
(01)00827002083578(17)270214(10)14492885
(01)00827002083578(17)270211(10)14492886
(01)00827002082595(17)270201(10)NS14485428
(01)00827002131040(17)270210(10)14446460
(01)00827002083561(17)270208(10)14448149
(01)00827002082618(17)270217(10)14498633
(01)00827002082618(17)270211(10)14498634
(01)00827002082618(17)270217(10)14498635
(01)00827002083578(17)270217(10)14498636
(01)00827002082618(17)270216(10)14498637
(01)00827002474161(17)270211(10)14501260
(01)00827002082618(17)270211(10)14501261
(01)00827002082618(17)270217(10)14501262
(01)00827002082618(17)270218(10)14501263
(01)00827002082618(17)270211(10)NS14506088
(01)00827002082618(17)270211(10)NS14506089
(01)00827002082595(17)270127(10)14397752
(01)00827002474192(17)270202(10)14438637
(01)00827002083561(17)270202(10)14440486
(01)00827002474185(17)270209(10)14440487
(01)00827002083561(17)270131(10)14440488
(01)00827002083561(17)270127(10)14440489
(01)00827002082595(17)270127(10)14442309
(01)00827002083561(17)270208(10)14451724
(01)00827002083561(17)270131(10)NS14443148
(01)00827002082595(17)270131(10)NS14446800
(01)00827002083561(17)270131(10)NS14446802
(01)00827002082595(17)270211(10)14446795
(01)00827002083561(17)270206(10)14450415
(01)00827002474185(17)270207(10)14451723
(01)00827002083561(17)270207(10)14452397
(01)00827002082595(17)270209(10)14452725
(01)00827002082595(17)270208(10)14452726
(01)00827002083561(17)270222(10)14478373
(01)00827002083561(17)270223(10)14485425
(01)00827002083561(17)270221(10)14486891
(01)00827002083561(17)270224(10)14488153
(01)00827002083561(17)270225(10)NS14511586
(01)00827002083561(17)270221(10)NS14527916
(01)00827002474192(17)270214(10)14457696
(01)00827002082595(17)270210(10)14459429
(01)00827002474185(17)270209(10)14462215
(01)00827002082618(17)270217(10)14511226
(01)00827002474161(17)270217(10)14513982
(01)00827002474161(17)270224(10)14513983
(01)00827002082618(17)270224(10)14513984
(01)00827002082618(17)270217(10)14513985
(01)00827002474178(17)270224(10)14513986
(01)00827002129245(17)270224(10)14513987
(01)00827002474161(17)270224(10)14513988
(01)00827002082618(17)270217(10)14531276
(01)00827002131026(17)270224(10)14534420
(01)00827002082618(17)270224(10)14534421
(01)00827002082595(17)270210(10)14462483
(01)00827002082595(17)270214(10)14464721
(01)00827002082595(17)270214(10)14466807
(01)00827002082595(17)270214(10)14466808
(01)00827002474192(17)270214(10)14468459
(01)00827002131040(17)270216(10)14468460
(01)00827002082618(17)270224(10)14531272
(01)00827002082618(17)270224(10)14531273
(01)00827002082618(17)270224(10)14531274
(01)00827002082618(17)270217(10)14531275
(01)00827002131057(17)270225(10)14485427
(01)00827002082595(17)270222(10)14485739
(01)00827002082595(17)270302(10)14506231
(01)00827002083578(17)270228(10)14542456
(01)00827002082618(17)270304(10)14548291
(01)00827002082618(17)270304(10)14548292
(01)00827002082618(17)270304(10)14548293
(01)00827002082618(17)270304(10)14553648
(01)00827002082595(17)270222(10)NS14518614
(01)00827002082595(17)270222(10)NS14518616
(01)00827002129245(17)270306(10)14550801
(01)00827002474161(17)270304(10)14553650
(01)00827002474178(17)270304(10)14553651
(01)00827002474161(17)270304(10)14553653
(01)00827002474161(17)270306(10)14553654
(01)00827002131026(17)270304(10)14553655
(01)00827002474178(17)270304(10)14556278
(01)00827002083578(17)270303(10)NS14553649
(01)00827002083561(17)270304(10)14506918
(01)00827002083561(17)270302(10)14513206
(01)00827002474185(17)270302(10)14530073
(01)00827002131033(17)270304(10)14556279
(01)00827002083578(17)270304(10)14556280
(01)00827002082595(17)270216(10)14472102
(01)00827002082595(17)270215(10)14473095
(01)00827002131026(17)270218(10)14534422
(01)00827002082618(17)270223(10)14534423
(01)00827002082618(17)270225(10)14534424
(01)00827002131026(17)270223(10)14534425
(01)00827002082618(17)270224(10)NS14534419
(01)00827002474185(17)270217(10)14472630
(01)00827002083561(17)270218(10)14473854
(01)00827002474192(17)270217(10)14474275
(01)00827002083561(17)270218(10)14474280
(01)00827002083578(17)270224(10)14536686
(01)00827002131033(17)270224(10)14536687
(01)00827002083578(17)270224(10)14536688
(01)00827002083578(17)270224(10)14536689
(01)00827002474178(17)270224(10)14536690
(01)00827002131033(17)270224(10)14536691
(01)00827002474178(17)270303(10)14536692
(01)00827002474161(17)270223(10)14539884
(01)00827002082595(17)270221(10)14474281
(01)00827002474185(17)270221(10)14485426
(01)00827002083578(17)270217(10)14539881
(01)00827002129245(17)270223(10)14539883
(01)00827002083578(17)270301(10)14539885
(01)00827002474161(17)270301(10)14539886
(01)00827002474161(17)270301(10)14539887
(01)00827002474161(17)270301(10)14539888
(01)00827002082618(17)270224(10)14542450
(01)00827002082618(17)270301(10)14542451
(01)00827002082618(17)270228(10)14542452
(01)00827002131033(17)270225(10)14542454
(01)00827002083578(17)270217(10)NS14539882
(01)00827002082618(17)270222(10)NS14542453
(01)00827002082618(17)270221(10)NS14542457
(01)00827002474161(17)270124(10)14474024
(01)00827002474178(17)270124(10)14474025
(01)00827002474161(17)270124(10)14474026
(01)00827002474161(17)270124(10)14474027
(01)00827002474161(17)270124(10)14474028
(01)00827002474161(17)270125(10)14474029
(01)00827002474178(17)270121(10)14474030
(01)00827002474161(17)270124(10)14478201
(01)00827002082618(17)270119(10)NS14478202
(01)00827002474192(17)261025(10)14270688
(01)00827002474192(17)261025(10)14270689
(01)00827002474185(17)261025(10)14270690
(01)00827002083561(17)261102(10)14270691
(01)00827002083561(17)261102(10)14270692
(01)00827002082595(17)261025(10)14270693
(01)00827002082618(17)261025(10)14276225
(01)00827002082595(17)261102(10)14285095
(01)00827002082618(17)261021(10)NS14290291
(01)00827002083561(17)270208(10)14454927
(01)00827002131057(17)270210(10)14457695
(01)00827002082618(17)270221(10)14506082
(01)00827002082618(17)270214(10)14506085
(01)00827002082618(17)270224(10)14511228
(01)00827002474161(17)270215(10)14511229
(01)00827002082618(17)270301(10)14511230
(01)00827002474161(17)270214(10)14511232
(01)00827002099449(17)270329(10)14580290
(01)00827002131064(17)270311(10)14489161
(01)00827002082595(17)270314(10)NS14518615
(01)00827002083561(17)270316(10)NS14572569
(01)00827002082618(17)270318(10)14577723
(01)00827002082618(17)270311(10)14577724
(01)00827002082618(17)270317(10)14577725
(01)00827002082618(17)270317(10)14577726
(01)00827002082618(17)270311(10)NS14577727
(01)00827002082618(17)270310(10)NS14579799
(01)00827002082595(17)270315(10)14506919
(01)00827002131040(17)270316(10)14518611
(01)00827002083561(17)270316(10)14521588
(01)00827002083561(17)270314(10)NS14527915
(01)00827002082595(17)270314(10)NS14536966
(01)00827002474185(17)270317(10)14511585
(01)00827002131057(17)270318(10)14514285
(01)00827002474192(17)270314(10)14518612
(01)00827002082595(17)270317(10)14518613
(01)00827002474192(17)270317(10)14521591
(01)00827002474185(17)270315(10)14523439
(01)00827002131057(17)270314(10)14527166
(01)00827002083561(17)270318(10)14530074
(01)00827002474185(17)270317(10)14539391
(01)00827002083561(17)270317(10)14539392
(01)00827002083561(17)270316(10)NS14521590
(01)00827002131040(17)270317(10)14527914
(01)00827002082595(17)270317(10)14531996
(01)00827002083561(17)270318(10)14534605
(01)00827002474185(17)270317(10)14536965
(01)00827002082595(17)270317(10)14542082
(01)00827002474192(17)270318(10)14547243
(01)00827002083561(17)270322(10)14551339
(01)00827002082595(17)270316(10)NS14573819
(01)00827002474192(17)270324(10)14555678
(01)00827002083578(17)270321(10)14597775
(01)00827002083578(17)270318(10)14597777
(01)00827002474161(17)270317(10)14602376
(01)00827002131033(17)270318(10)14602378
(01)00827002082618(17)270325(10)14604806
(01)00827002082618(17)270323(10)NS14613686
(01)00827002099456(17)261027(10)14232885
(01)00827002083561(17)261101(10)14232891
(01)00827002082618(17)261028(10)14273585
(01)00827002082618(17)261028(10)14273586
(01)00827002131026(17)261102(10)14273589
(01)00827002082595(17)261102(10)14285091
(01)00827002083561(17)261027(10)14285092
(01)00827002083561(17)261027(10)14285093
(01)00827002082595(17)261102(10)14285094
(01)00827002082618(17)261027(10)14297278
(01)00827002083578(17)261102(10)14297281
(01)00827002083578(17)261028(10)14297284
(01)00827002082595(17)261027(10)NS14232893
(01)00827002083561(17)261028(10)NS14232896
(01)00827002083561(17)261027(10)NS14239945
(01)00827002099449(17)261022(10)NS14248274
(01)00827002082595(17)261027(10)NS14248279
(01)00827002083561(17)261027(10)NS14250702
(01)00827002083578(17)261027(10)NS14276228
(01)00827002083561(17)261219(10)14359593
(01)00827002083561(17)261220(10)14359594
(01)00827002099456(17)270103(10)14370216
(01)00827002083578(17)261210(10)14402380
(01)00827002131033(17)261215(10)14402381
(01)00827002131026(17)261215(10)14402382
(01)00827002082618(17)261217(10)14402383
(01)00827002083578(17)261214(10)14402384
(01)00827002082618(17)261216(10)14405353
(01)00827002082618(17)261215(10)14405354
(01)00827002082618(17)261216(10)14405356
(01)00827002082618(17)261216(10)14405357
(01)00827002082595(17)261228(10)NS14370204
(01)00827002082618(17)270127(10)14453982
(01)00827002083578(17)270119(10)14453983
(01)00827002082618(17)270121(10)14453984
(01)00827002083578(17)270124(10)14453988
(01)00827002082618(17)270120(10)14456812
(01)00827002082618(17)270120(10)14456813
(01)00827002083578(17)270121(10)14456814
(01)00827002082618(17)270120(10)14456816
(01)00827002082618(17)270121(10)14456817
(01)00827002082595(17)270117(10)14431473
(01)00827002082595(17)270123(10)14433351
(01)00827002082595(17)270121(10)14434539
(01)00827002082595(17)270123(10)14434540
(01)00827002082595(17)270124(10)14434594
(01)00827002131057(17)270127(10)14438131
(01)00827002474185(17)270131(10)14438704
(01)00827002083561(17)270201(10)14438705
(01)00827002474192(17)270127(10)14442308
|
| FEI Number |
1820334
|
Recalling Firm/ Manufacturer |
Cook Incorporated 750 N Daniels Way Bloomington IN 47404-9120
|
| For Additional Information Contact | Cook Medical Customer Relations Department 800-457-4500 |
Manufacturer Reason for Recall | The loading cartridges included with these devices may contain a small, unintended, stainless-steel cannula. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | On May 5, 2022, the firm distributed Urgent Medical Device Recall letters to affected customers. Customers were advised that the presence of a small stainless steel cannula inside the loading cartridge may go undetected by the user.
Customers were instructed to examine their inventory and quarantine any affected product that remains unused. Immediately cease all further distribution and use of the affected products. Customers should return the product to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Please share this information with appropriate personnel, down to the user level, within your organization or with any organization where the affected devices may have been transferred.
Physicians should practice their standard patient monitoring following the procedure for an early identification of any complications to mitigate their severity.
If you have any questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235. |
| Quantity in Commerce | 106,033 |
| Distribution | Domestic distribution nationwide. Foreign distribution worldwide. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = KRD
|