| | Class 1 Device Recall Cobalt |  |
| Date Initiated by Firm | June 22, 2022 |
| Date Posted | August 09, 2022 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1485-2022 |
| Recall Event ID |
90565 |
| PMA Number | P980016S713 |
| Product Classification |
Implantable cardioverter defibrillator (non-CRT) - Product Code LWS
|
| Product | Cobalt Implantable Cardioverter Defibrillators: a. Cobalt XT VR ICD, Model Numbers: DVPA2D1, DVPA2D4; b. Cobalt XT DR ICD, Model Numbers: DDPA2D1, DDPA2D4 |
| Code Information |
a. Model Numbers: DVPA2D1: GTIN 00763000178451, Serial Numbers: RSC600237S, RSC600256S, RSC600277S, RSC600327S, RSC600419S, RSC601038S, RSC601061S, RSC601065S, RSC601084S, RSC601126S, RSC601134S, RSC601138S, RSC601158S, RSC601159S, RSC601175S, RSC601193S, RSC601209S, RSC601210S, RSC601213S, RSC601214S, RSC601217S, RSC601226S, RSC601242S, RSC601244S, RSC601249S, RSC601313S, RSC601314S, RSC601315S, RSC601317S, RSC601323S, RSC601330S, RSC601332S, RSC601337S, RSC601341S, RSC601342S, RSC601346S, RSC601349S, RSC601354S, RSC601358S, RSC601370S, RSC601372S, RSC601377S, RSC601379S, RSC601397S, RSC601402S, RSC601411S, RSC601420S, RSC601422S, RSC601438S, RSC601441S, RSC601446S, RSC601453S, RSC601467S, RSC601473S, RSC601476S, RSC601486S, RSC601493S, RSC601494S, RSC601500S, RSC601506S, RSC601510S, RSC601511S, RSC601516S, RSC601520S, RSC601525S, RSC601526S, RSC601532S, RSC601539S, RSC601542S, RSC601581S, RSC601622S, RSC601627S, RSC601628S, RSC601631S, RSC601635S, RSC601636S, RSC601644S, RSC601650S, RSC601651S, RSC601659S, RSC601674S, RSC601676S, RSC601678S, RSC601680S, RSC601681S, RSC601683S, RSC601690S, RSC601693S, RSC601694S, RSC601696S, RSC601697S, RSC601698S, RSC601699S, RSC601702S, RSC601704S, RSC601705S, RSC601707S, RSC601708S, RSC601712S, RSC601714S, RSC601721S, RSC601722S, RSC601726S, RSC601733S, RSC601756S; GTIN 00763000178475, Serial Numbers: RSC600021S, RSC600025S, RSC600029S, RSC600031S, RSC600037S, RSC600053S, RSC600056S, RSC600059S, RSC600062S, RSC600063S, RSC600077S, RSC600223S, RSC600227S, RSC600309S, RSC600315S, RSC600403S, RSC600405S, RSC600410S, RSC600459S, RSC600460S, RSC600463S, RSC600469S, RSC600479S, RSC600509S, RSC600518S, RSC600523S, RSC600524S, RSC600525S, RSC600526S, RSC600529S, RSC600538S, RSC600540S, RSC600542S, RSC600543S, RSC600554S, RSC600555S, RSC600560S, RSC600562S, RSC600564S, RSC600569S, RSC600571S, RSC600572S, RSC600573S, RSC600575S, RSC600577S, RSC600578S, RSC600581S, RSC600586S, RSC600588S, RSC600598S, RSC600599S, RSC600606S, RSC600610S, RSC600613S, RSC600616S, RSC600618S, RSC600619S, RSC600620S, RSC600621S, RSC600670S, RSC600671S, RSC600675S, RSC600676S, RSC600677S, RSC600683S, RSC600685S, RSC600688S, RSC600689S, RSC600691S, RSC600692S, RSC600694S, RSC600695S, RSC600697S, RSC600698S, RSC600699S, RSC600701S, RSC600704S, RSC600707S, RSC600709S, RSC600710S, RSC600711S, RSC600713S, RSC600714S, RSC600717S, RSC600718S, RSC600721S, RSC600722S, RSC600725S, RSC600726S, RSC600729S, RSC600730S, RSC600732S, RSC600733S, RSC600734S, RSC600735S, RSC600737S, RSC600739S, RSC600740S, RSC600741S, RSC600742S, RSC600745S, RSC600746S, RSC600747S, RSC600748S, RSC600749S, RSC600750S, RSC600751S, RSC600753S, RSC600754S, RSC600756S, RSC600759S, RSC600760S, RSC600761S, RSC600763S, RSC600764S, RSC600767S, RSC600768S, RSC600769S, RSC600771S, RSC600772S, RSC600773S, RSC600775S, RSC600776S, RSC600778S, RSC600779S, RSC600780S, RSC600781S, RSC600782S, RSC600783S, RSC600785S, RSC600786S, RSC600787S, RSC600788S, RSC600789S, RSC600790S, RSC600792S, RSC600793S, RSC600794S, RSC600795S, RSC600796S, RSC600797S, RSC600798S, RSC600799S, RSC600800S, RSC600804S, RSC600806S, RSC600807S, RSC600808S, RSC600809S, RSC600810S, RSC600811S, RSC600812S, RSC600814S, RSC600816S, RSC600817S, RSC600823S, RSC600825S, RSC600826S, RSC600827S, RSC600830S, RSC600831S, RSC600832S, RSC600835S, RSC600837S, RSC600840S, RSC600841S, RSC600842S, RSC600843S, RSC600844S, RSC600846S, RSC600851S, RSC600852S, RSC600857S, RSC600858S, RSC600863S, RSC600864S, RSC600865S, RSC600869S, RSC600870S, RSC600874S, RSC600876S, RSC600880S, RSC600881S, RSC600884S, RSC600890S, RSC600891S, RSC600892S, RSC600893S, RSC600894S, RSC600904S, RSC600907S, RSC600908S, RSC600910S, RSC600911S, RSC600913S, RSC600914S, RSC600915S, RSC600916S, RSC600921S, RSC600922S, RSC600926S, RSC600927S, RSC600929S, RSC600930S, RSC600958S, RSC600959S, RSC600960S, RSC600961S, RSC600963S, RSC600964S, RSC600965S, RSC600966S, RSC600967S, RSC600968S, RSC600969S, RSC600986S, RSC600987S, RSC601003S, RSC601005S.RSC601006S, RSC601007S, RSC601008S, RSC601011S, RSC601015S, RSC601016S, RSC601019S, RSC601043S, RSC601044S, RSC601048S, RSC601049S, RSC601050S, RSC601052S, RSC601053S, RSC601054S, RSC601055S, RSC601197S, RSC601256S, RSC601258S, RSC601259S, RSC601260S, RSC601263S, RSC601265S, RSC601266S, RSC601267S, RSC601268S, RSC601269S, RSC601294S, RSC601295S, RSC601296S, RSC601297S, RSC601298S, RSC601299S, RSC601300S, RSC601301S, RSC601302S, RSC601303S, RSC601304S, RSC601305S, RSC601306S, RSC601307S, RSC601326S, RSC601407S, RSC601408S, RSC601416S, RSC601417S, RSC601428S, RSC601445S, RSC601447S, RSC601448S, RSC601449S, RSC601451S, RSC601452S, RSC601461S, RSC601463S, RSC601465S, RSC601535S, RSC601540S, RSC601544S, RSC601546S, RSC601548S, RSC601549S, RSC601550S, RSC601559S, RSC601563S, RSC601565S, RSC601570S, RSC601571S, RSC601573S, RSC601574S, RSC601575S, RSC601580S, RSC601582S, RSC601584S, RSC601588S, RSC601590S, RSC601595S, RSC601597S, RSC601598S, RSC601603S, RSC601604S, RSC601607S, RSC601608S, RSC601612S, RSC601661S, RSC601670S, RSC601686S, RSC601687S, RSC601691S, RSC601700S, RSC601715S, RSC601719S, RSC601729S, RSC601730S, RSC601732S, RSC601734S, RSC601736S, RSC601737S, RSC601738S, RSC601740S, RSC601742S, RSC601744S, RSC601747S, RSC601749S, RSC601750S, RSC601753S, RSC601755S, RSC601757S; GTIN 00763000380380, Serial Numbers: RSC600057S, RSC600067S, RSC600068S, RSC600070S, RSC600072S, RSC600073S, RSC600076S, RSC600078S, RSC600083S, RSC600088S, RSC600089S, RSC600090S, RSC600092S, RSC600094S, RSC600095S, RSC600097S, RSC600101S, RSC600102S, RSC600104S, RSC600105S, RSC600106S, RSC600107S, RSC600108S, RSC600109S, RSC600114S, RSC600117S, RSC600119S, RSC600122S, RSC600124S, RSC600266S, RSC600439S, RSC600454S, RSC600456S, RSC600457S, RSC600458S, RSC600522S, RSC600527S, RSC600528S, RSC600531S, RSC600532S, RSC600533S, RSC600535S, RSC600537S, RSC600541S, RSC600546S, RSC600556S, RSC600557S, RSC600558S, RSC600561S, RSC600563S, RSC600565S, RSC600566S, RSC600567S, RSC600570S, RSC600574S, RSC600576S, RSC600582S, RSC600584S, RSC600587S, RSC600590S, RSC600591S, RSC600592S, RSC600594S, RSC600595S, RSC600596S, RSC600600S, RSC600601S, RSC600602S, RSC600603S, RSC600607S, RSC600608S, RSC600614S, RSC600829S, RSC600853S, RSC600854S, RSC600856S, RSC600859S, RSC600867S, RSC600868S, RSC600871S, RSC600872S, RSC600873S, RSC600875S, RSC600877S, RSC600879S, RSC600882S, RSC600887S, RSC600896S, RSC600897S, RSC600898S, RSC600900S, RSC600901S, RSC600903S, RSC600906S, RSC600909S, RSC600918S, RSC600928S, RSC600934S, RSC600935S, RSC600936S, RSC600937S, RSC600942S, RSC600943S, RSC600944S, RSC600945S, RSC600950S, RSC600953S, RSC600954S, RSC600955S, RSC600957S, RSC600972S, RSC600973S, RSC600974S, RSC600975S, RSC600976S, RSC600977S, RSC600979S, RSC600980S, RSC600984S, RSC600985S, RSC600991S, RSC600994S, RSC600995S, RSC600996S, RSC600997S, RSC600998S, RSC601000S, RSC601013S, RSC601279S, RSC601283S, RSC601284S, RSC601293S, RSC601562S, RSC601591S, RSC601596S, RSC601728S, RSC601754S, RSC601768S; GTIN none, Serial Numbers: RSC601291S, RSC601741S DVPA2D4: GTIN 00763000178468, Serial Numbers: RSD600321S, RSD600325S, RSD601337S, RSD602358S, RSD602590S, RSD602748S, RSD602794S, RSD602910S, RSD603138S, RSD603280S, RSD603519S, RSD603660S, RSD603697S, RSD603727S, RSD603739S, RSD603753S, RSD603762S, RSD603815S, RSD603997S, RSD604003S, RSD604056S, RSD604057S, RSD604071S, RSD604074S, RSD604075S, RSD604076S, RSD604080S, RSD604183S, RSD604184S, RSD604192S, RSD604202S, RSD604205S, RSD604221S, RSD604222S, RSD604230S, RSD604231S, RSD604245S, RSD604247S, RSD604311S, RSD604324S, RSD604327S; GTIN 00763000178482, Serial Numbers: RSD600016S, RSD600017S, RSD600018S, RSD600019S, RSD600022S, RSD600023S, RSD600030S, RSD600031S, RSD600038S, RSD600041S, RSD600042S, RSD600043S, RSD600044S, RSD600046S, RSD600047S, RSD600048S, RSD600061S, RSD600062S, RSD600063S, RSD600064S, RSD600087S, RSD600090S, RSD600091S, RSD600093S, RSD600094S, RSD600095S, RSD600099S, RSD600163S, RSD600168S, RSD600170S, RSD600171S, RSD600172S, RSD600174S, RSD600175S, RSD600176S, RSD600177S, RSD600178S, RSD600179S, RSD600180S, RSD600181S, RSD600182S, RSD600183S, RSD600295S, RSD600628S, RSD600633S, RSD600635S, RSD600636S, RSD600639S, RSD600644S, RSD600646S, RSD600647S, RSD600652S, RSD600679S, RSD600713S, RSD600717S, RSD600720S, RSD600721S, RSD600722S, RSD600727S, RSD600733S, RSD600743S, RSD600746S, RSD600748S, RSD600753S, RSD600754S, RSD600755S, RSD600759S, RSD600761S, RSD600763S, RSD600764S, RSD600765S, RSD600766S, RSD600767S, RSD600768S, RSD600769S, RSD600770S, RSD600771S, RSD600772S, RSD600773S, RSD600775S, RSD600776S, RSD600778S, RSD600779S, RSD600780S, RSD600781S, RSD600782S, RSD600783S, RSD600784S, RSD600785S, RSD600787S, RSD600788S, RSD600789S, RSD600792S, RSD600793S, RSD600794S, RSD600795S, RSD600796S, RSD600797S, RSD600798S, RSD600800S, RSD600801S, RSD600802S, RSD600803S, RSD600805S, RSD600806S, RSD600807S, RSD600808S, RSD600809S, RSD600810S, RSD600811S, RSD600812S, RSD600813S, RSD600814S, RSD600815S, RSD600817S, RSD600818S, RSD600821S, RSD600914S, RSD600917S, RSD600928S, RSD600930S, RSD600931S, RSD600934S, RSD600945S, RSD600947S, RSD600949S, RSD600950S, RSD600959S, RSD600960S, RSD600967S, RSD600968S, RSD600970S, RSD600973S, RSD601004S, RSD601005S, RSD601006S, RSD601014S, RSD601016S, RSD601019S, RSD601028S, RSD601031S, RSD601032S, RSD601033S, RSD601034S, RSD601039S, RSD601052S, RSD601054S, RSD601062S, RSD601063S, RSD601064S, RSD601065S, RSD601066S, RSD601072S, RSD601076S, RSD601077S, RSD601078S, RSD601079S, RSD601080S, RSD601081S, RSD601082S, RSD601083S, RSD601086S, RSD601087S, RSD601088S, RSD601089S, RSD601091S, RSD601092S, RSD601094S, RSD601095S, RSD601097S, RSD601098S, RSD601101S, RSD601102S, RSD601103S, RSD601104S, RSD601106S, RSD601111S, RSD601114S, RSD601115S, RSD601116S, RSD601117S, RSD601119S, RSD601121S, RSD601122S, RSD601123S, RSD601124S, RSD601125S, RSD601126S, RSD601127S, RSD601128S, RSD601129S, RSD601130S, RSD601132S, RSD601136S, RSD601138S, RSD601139S, RSD601141S, RSD601142S, RSD601143S, RSD601144S, RSD601145S, RSD601146S, RSD601147S, RSD601148S, RSD601149S, RSD601150S, RSD601151S, RSD601152S, RSD601153S, RSD601155S, RSD601156S, RSD601157S, RSD601158S, RSD601159S, RSD601160S, RSD601161S, RSD601162S, RSD601163S, RSD601164S, RSD601165S, RSD601167S, RSD601169S, RSD601170S, RSD601171S, RSD601173S, RSD601174S, RSD601176S, RSD601178S, RSD601179S, RSD601180S, RSD601181S, RSD601182S, RSD601183S, RSD601184S, RSD601186S, RSD601188S, RSD601189S, RSD601191S, RSD601192S, RSD601193S, RSD601198S, RSD601199S, RSD601200S, RSD601201S, RSD601203S, RSD601205S, RSD601206S, RSD601208S, RSD601209S, RSD601213S, RSD601214S, RSD601215S, RSD601216S, RSD601217S, RSD601218S, RSD601219S, RSD601220S, RSD601224S, RSD601225S, RSD601227S, RSD601228S, RSD601230S, RSD601232S, RSD601233S, RSD601235S, RSD601240S, RSD601241S, RSD601242S, RSD601243S, RSD601244S, RSD601246S, RSD601247S, RSD601248S, RSD601249S, RSD601250S, RSD601251S, RSD601252S, RSD601254S, RSD601258S, RSD601260S, RSD601261S, RSD601262S, RSD601263S, RSD601264S, RSD601265S, RSD601266S, RSD601269S, RSD601270S, RSD601271S, RSD601272S, RSD601278S, RSD601280S, RSD601282S, RSD601283S, RSD601285S, RSD601286S, RSD601287S, RSD601288S, RSD601289S, RSD601290S, RSD601291S, RSD601292S, RSD601294S, RSD601296S, RSD601298S, RSD601299S, RSD601300S, RSD601301S, RSD601302S, RSD601303S, RSD601304S, RSD601305S, RSD601306S, RSD601307S, RSD601309S, RSD601310S, RSD601311S, RSD601312S, RSD601313S, RSD601314S, RSD601315S, RSD601316S, RSD601318S, RSD601319S, RSD601320S, RSD601321S, RSD601322S, RSD601323S, RSD601325S, RSD601326S, RSD601327S, RSD601328S, RSD601329S, RSD601330S, RSD601331S, RSD601339S, RSD601340S, RSD601341S, RSD601342S, RSD601344S, RSD601345S, RSD601346S, RSD601347S, RSD601348S, RSD601349S, RSD601350S, RSD601351S, RSD601352S, RSD601353S, RSD601355S, RSD601356S, RSD601360S, RSD601398S, RSD601407S, RSD601408S, RSD601409S, RSD601411S, RSD601419S, RSD601420S, RSD601422S, RSD601424S, RSD601436S, RSD601439S, RSD601444S, RSD601446S, RSD601463S, RSD601470S, RSD601472S, RSD601476S, RSD601478S, RSD601480S, RSD601481S, RSD601482S, RSD601483S, RSD601484S, RSD601486S, RSD601490S, RSD601491S, RSD601493S, RSD601494S, RSD601495S, RSD601496S, RSD601497S, RSD601498S, RSD601499S, RSD601500S, RSD601501S, RSD601502S, RSD601504S, RSD601505S, RSD601506S, RSD601507S, RSD601508S, RSD601509S, RSD601512S, RSD601513S, RSD601514S, RSD601516S, RSD601517S, RSD601518S, RSD601519S, RSD601520S, RSD601521S, RSD601522S, RSD601523S, RSD601524S, RSD601525S, RSD601528S, RSD601529S, RSD601530S, RSD601531S, RSD601532S, RSD601533S, RSD601534S, RSD601535S, RSD601536S, RSD601538S, RSD601539S, RSD601546S, RSD601549S, RSD601554S, RSD601562S, RSD601565S, RSD601588S, RSD601589S, RSD601592S, RSD601593S, RSD601595S, RSD601597S, RSD601598S, RSD601599S, RSD601600S, RSD601601S, RSD601602S, RSD601604S, RSD601605S, RSD601607S, RSD601608S, RSD601609S, RSD601610S, RSD601611S, RSD601612S, RSD601613S, RSD601614S, RSD601616S, RSD601617S, RSD601618S, RSD601619S, RSD601621S, RSD601622S, RSD601624S, RSD601625S, RSD601627S, RSD601628S, RSD601630S, RSD601631S, RSD601632S, RSD601634S, RSD601635S, RSD601637S, RSD601639S, RSD601644S, RSD601645S, RSD601646S, RSD601649S, RSD601651S, RSD601653S, RSD601654S, RSD601655S, RSD601657S, RSD601658S, RSD601659S, RSD601660S, RSD601664S, RSD601666S, RSD601667S, RSD601668S, RSD601669S, RSD601672S, RSD601673S, RSD601674S, RSD601676S, RSD601677S, RSD601678S, RSD601680S, RSD601681S, RSD601682S, RSD601683S, RSD601685S, RSD601687S, RSD601690S, RSD601691S, RSD601694S, RSD601695S, RSD601700S, RSD601701S, RSD601705S, RSD601709S, RSD601711S, RSD601712S, RSD601713S, RSD601715S, RSD601716S, RSD601717S, RSD601718S, RSD601719S, RSD601720S, RSD601721S, RSD601722S, RSD601723S, RSD601724S, RSD601725S, RSD601726S, RSD601727S, RSD601728S, RSD601729S, RSD601730S, RSD601731S, RSD601732S, RSD601733S, RSD601734S, RSD601735S, RSD601736S, RSD601737S, RSD601738S, RSD601739S, RSD601741S, RSD601742S, RSD601743S, RSD601744S, RSD601745S, RSD601746S, RSD601747S, RSD601748S, RSD601749S, RSD601751S, RSD601752S, RSD601756S, RSD601757S, RSD601758S, RSD601759S, RSD601760S, RSD601761S, RSD601762S, RSD601765S, RSD601769S, RSD601772S, RSD601778S, RSD601782S, RSD601786S, RSD601788S, RSD601789S, RSD601794S, RSD601795S, RSD601798S, RSD601799S, RSD601802S, RSD601803S, RSD601804S, RSD601805S, RSD601806S, RSD601807S, RSD601808S, RSD601809S, RSD601811S, RSD601813S, RSD601816S, RSD601817S, RSD601821S, RSD601825S, RSD601827S, RSD601831S, RSD601832S, RSD601834S, RSD601835S, RSD601837S, RSD601843S, RSD601851S, RSD601854S, RSD601857S, RSD601864S, RSD601867S, RSD601875S, RSD601883S, RSD601894S, RSD601895S, RSD601898S, RSD601899S, RSD601900S, RSD601901S, RSD601905S, RSD601907S, RSD601915S, RSD601918S, RSD601919S, RSD601920S, RSD601928S, RSD601930S, RSD601942S, RSD601943S, RSD601944S, RSD601947S, RSD601955S, RSD601962S, RSD601965S, RSD601970S, RSD601982S, RSD601984S, RSD601986S, RSD601989S, RSD601990S, RSD601991S, RSD601992S, RSD601993S, RSD601994S, RSD601995S, RSD601996S, RSD601997S, RSD601998S, RSD601999S, RSD602000S, RSD602001S, RSD602002S, RSD602003S, RSD602004S, RSD602005S, RSD602006S, RSD602007S, RSD602008S, RSD602009S, RSD602016S, RSD602017S, RSD602019S, RSD602020S, RSD602021S, RSD602022S, RSD602023S, RSD602025S, RSD602026S, RSD602027S, RSD602028S, RSD602029S, RSD602030S, RSD602032S, RSD602033S, RSD602034S, RSD602046S, RSD602066S, RSD602070S, RSD602071S, RSD602072S, RSD602074S, RSD602075S, RSD602076S, RSD602077S, RSD602079S, RSD602082S, RSD602086S, RSD602088S, RSD602089S, RSD602112S, RSD602115S, RSD602116S, RSD602118S, RSD602122S, RSD602124S, RSD602160S, RSD602161S, RSD602162S, RSD602163S, RSD602164S, RSD602165S, RSD602167S, RSD602168S, RSD602169S, RSD602170S, RSD602171S, RSD602173S, RSD602177S, RSD602178S, RSD602179S, RSD602181S, RSD602183S, RSD602184S, RSD602185S, RSD602186S, RSD602187S, RSD602188S, RSD602189S, RSD602191S, RSD602193S, RSD602194S, RSD602210S, RSD602221S, RSD602223S, RSD602225S, RSD602226S, RSD602228S, RSD602230S, RSD602235S, RSD602237S, RSD602238S, RSD602247S, RSD602252S, RSD602253S, RSD602254S, RSD602255S, RSD602259S, RSD602260S, RSD602264S, RSD602265S, RSD602267S, RSD602268S, RSD602275S, RSD602276S, RSD602277S, RSD602278S, RSD602279S, RSD602284S, RSD602289S, RSD602300S, RSD602302S, RSD602316S, RSD602341S, RSD602353S, RSD602362S, RSD602364S, RSD602401S, RSD602415S, RSD602431S, RSD602435S, RSD602467S, RSD602482S, RSD602483S, RSD602488S, RSD602495S, RSD602498S, RSD602502S, RSD602503S, RSD602515S, RSD602516S, RSD602530S, RSD602531S, RSD602532S, RSD602535S, RSD602541S, RSD602543S, RSD602546S, RSD602548S, RSD602549S, RSD602551S, RSD602553S, RSD602554S, RSD602556S, RSD602559S, RSD602560S, RSD602563S, RSD602568S, RSD602571S, RSD602572S, RSD602573S, RSD602574S, RSD602578S, RSD602787S, RSD602790S, RSD602791S, RSD602793S, RSD602795S, RSD602803S, RSD602805S, RSD602806S, RSD602807S, RSD602808S, RSD602809S, RSD602810S, RSD602811S, RSD602812S, RSD602813S, RSD602814S, RSD602815S, RSD602816S, RSD602818S, RSD602819S, RSD602820S, RSD602821S, RSD602822S, RSD602824S, RSD602825S, RSD602826S, RSD602827S, RSD602828S, RSD602829S, RSD602830S, RSD602831S, RSD602832S, RSD602833S, RSD602834S, RSD602835S, RSD602836S, RSD602837S, RSD602838S, RSD602840S, RSD602841S, RSD602842S, RSD602843S, RSD602844S, RSD602845S, RSD602846S, RSD602847S, RSD602848S, RSD602849S, RSD602850S, RSD602851S, RSD602853S, RSD602854S, RSD602855S, RSD602857S, RSD602859S, RSD602860S, RSD602861S, RSD602863S, RSD602864S, RSD602945S, RSD602951S, RSD602953S, RSD602954S, RSD602958S, RSD602959S, RSD602960S, RSD602961S, RSD603075S, RSD603087S, RSD603109S, RSD603111S, RSD603119S, RSD603121S, RSD603134S, RSD603144S, RSD603148S, RSD603155S, RSD603177S, RSD603178S, RSD603179S, RSD603180S, RSD603185S, RSD603187S, RSD603188S, RSD603192S, RSD603194S, RSD603199S, RSD603200S, RSD603228S, RSD603229S, RSD603232S, RSD603238S, RSD603240S, RSD603261S, RSD603269S, RSD603273S, RSD603406S, RSD603423S, RSD603426S, RSD603427S, RSD603428S, RSD603436S, RSD603437S, RSD603439S, RSD603440S, RSD603441S, RSD603442S, RSD603443S, RSD603445S, RSD603447S, RSD603448S, RSD603449S, RSD603452S, RSD603455S, RSD603456S, RSD603457S, RSD603461S, RSD603464S, RSD603465S, RSD603467S, RSD603469S, RSD603474S, RSD603477S, RSD603479S, RSD603481S, RSD603482S, RSD603483S, RSD603484S, RSD603486S, RSD603487S, RSD603489S, RSD603494S, RSD603496S, RSD603497S, RSD603502S, RSD603503S, RSD603504S, RSD603505S, RSD603506S, RSD603507S, RSD603508S, RSD603509S, RSD603510S, RSD603511S, RSD603512S, RSD603513S, RSD603514S, RSD603515S, RSD603516S, RSD603517S, RSD603520S, RSD603521S, RSD603526S, RSD603527S, RSD603528S, RSD603529S, RSD603530S, RSD603531S, RSD603533S, RSD603536S, RSD603537S, RSD603538S, RSD603539S, RSD603540S, RSD603541S, RSD603542S, RSD603543S, RSD603544S, RSD603556S, RSD603557S, RSD603560S, RSD603561S, RSD603565S, RSD603566S, RSD603568S, RSD603569S, RSD603572S, RSD603574S, RSD603582S, RSD603583S, RSD603584S, RSD603585S, RSD603588S, RSD603590S, RSD603591S, RSD603592S, RSD603594S, RSD603596S, RSD603602S, RSD603603S, RSD603604S, RSD603605S, RSD603607S, RSD603620S, RSD603621S, RSD603639S, RSD603644S, RSD603645S, RSD603648S, RSD603649S, RSD603653S, RSD603654S, RSD603657S, RSD603663S, RSD603664S, RSD603666S, RSD603677S, RSD603681S, RSD603682S, RSD603693S, RSD603709S, RSD603714S, RSD603718S, RSD603721S, RSD603766S, RSD603767S, RSD603769S, RSD603770S, RSD603771S, RSD603772S, RSD603775S, RSD603777S, RSD603779S, RSD603781S, RSD603782S, RSD603783S, RSD603785S, RSD603786S, RSD603788S, RSD603789S, RSD603791S, RSD603792S, RSD603793S, RSD603797S, RSD603799S, RSD603802S, RSD603805S, RSD603806S, RSD603808S, RSD603812S, RSD603816S, RSD603817S, RSD603818S, RSD603824S, RSD603827S, RSD603830S, RSD603831S, RSD603832S, RSD603835S, RSD603836S, RSD603838S, RSD603839S, RSD603843S, RSD603845S, RSD603888S, RSD603890S, RSD603892S, RSD603907S, RSD603917S, RSD603918S, RSD603946S, RSD603951S, RSD603953S, RSD603959S, RSD603961S, RSD603966S, RSD603978S, RSD604011S, RSD604036S, RSD604105S, RSD604112S, RSD604116S, RSD604117S, RSD604119S, RSD604120S, RSD604125S, RSD604127S, RSD604129S, RSD604136S, RSD604137S, RSD604140S, RSD604143S, RSD604144S, RSD604149S, RSD604150S, RSD604151S, RSD604152S, RSD604156S, RSD604159S, RSD604160S, RSD604161S, RSD604162S, RSD604164S, RSD604165S, RSD604166S, RSD604171S, RSD604172S, RSD604173S, RSD604175S, RSD604176S, RSD604178S, RSD604179S, RSD604180S, RSD604206S, RSD604211S, RSD604217S, RSD604219S, RSD604229S, RSD604236S, RSD604254S, RSD604260S, RSD604264S, RSD604265S, RSD604266S, RSD604267S, RSD604268S, RSD604272S, RSD604275S, RSD604279S, RSD604284S, RSD604288S, RSD604295S, RSD604296S, RSD604298S, RSD604304S, RSD604316S, RSD604321S, RSD604322S, RSD604323S, RSD604326S, RSD604329S, RSD604339S, RSD604341S, RSD604342S, RSD604343S, RSD604344S, RSD604345S, RSD604347S, RSD604348S, RSD604349S, RSD604350S, RSD604351S, RSD604352S, RSD604356S, RSD604359S, RSD604360S, RSD604362S, RSD604363S, RSD604364S, RSD604365S; GTIN 00763000380397, Serial Numbers: RSD600050S, RSD600051S, RSD600052S, RSD600053S, RSD600054S, RSD600055S, RSD600056S, RSD600058S, RSD600066S, RSD600068S, RSD600069S, RSD600070S, RSD600072S, RSD600073S, RSD600076S, RSD600077S, RSD600078S, RSD600079S, RSD600082S, RSD600083S, RSD600084S, RSD600085S, RSD600086S, RSD600088S, RSD600101S, RSD600102S, RSD600104S, RSD600105S, RSD600107S, RSD600109S, RSD600110S, RSD600111S, RSD600113S, RSD600116S, RSD600117S, RSD600118S, RSD600119S, RSD600122S, RSD600125S, RSD600126S, RSD600128S, RSD600131S, RSD600133S, RSD600135S, RSD600140S, RSD600141S, RSD600142S, RSD600144S, RSD600146S, RSD600147S, RSD600149S, RSD600151S, RSD600152S, RSD600153S, RSD600155S, RSD600157S, RSD600160S, RSD600162S, RSD600166S, RSD600167S, RSD600173S, RSD600613S, RSD600626S, RSD600627S, RSD600631S, RSD600632S, RSD600637S, RSD600638S, RSD600640S, RSD600642S, RSD600643S, RSD600648S, RSD600649S, RSD600650S, RSD600651S, RSD600653S, RSD600655S, RSD600657S, RSD600658S, RSD600659S, RSD600660S, RSD600661S, RSD600662S, RSD600663S, RSD600665S, RSD600668S, RSD600670S, RSD600671S, RSD600673S, RSD600674S, RSD600675S, RSD600676S, RSD600677S, RSD600678S, RSD600680S, RSD600681S, RSD600682S, RSD600683S, RSD600685S, RSD600687S, RSD600688S, RSD600689S, RSD600690S, RSD600692S, RSD600693S, RSD600694S, RSD600696S, RSD600698S, RSD600699S, RSD600702S, RSD600703S, RSD600704S, RSD600706S, RSD600708S, RSD600709S, RSD600710S, RSD600714S, RSD600718S, RSD600719S, RSD600723S, RSD600724S, RSD600725S, RSD600726S, RSD600729S, RSD600730S, RSD600731S, RSD600734S, RSD600735S, RSD600738S, RSD600740S, RSD600742S, RSD600747S, RSD600749S, RSD600750S, RSD600751S, RSD600752S, RSD600756S, RSD600757S, RSD600762S, RSD600774S, RSD600786S, RSD600791S, RSD600799S, RSD600804S, RSD600816S, RSD601023S, RSD601024S, RSD601027S, RSD601029S, RSD601030S, RSD601035S, RSD601037S, RSD601038S, RSD601040S, RSD601041S, RSD601044S, RSD601046S, RSD601048S, RSD601049S, RSD601051S, RSD601055S, RSD601057S, RSD601058S, RSD601061S, RSD601068S, RSD601071S, RSD601084S, RSD601085S, RSD601090S, RSD601093S, RSD601105S, RSD601109S, RSD601110S, RSD601113S, RSD601118S, RSD601120S, RSD601137S, RSD601187S, RSD601194S, RSD601195S, RSD601196S, RSD601207S, RSD601212S, RSD601221S, RSD601223S, RSD601231S, RSD601234S, RSD601238S, RSD601253S, RSD601256S, RSD601267S, RSD601275S, RSD601281S, RSD601297S, RSD601343S, RSD601383S, RSD601393S, RSD601464S, RSD601465S, RSD601468S, RSD601469S, RSD601471S, RSD601473S, RSD601474S, RSD601479S, RSD601488S, RSD601492S, RSD601503S, RSD601510S, RSD601515S, RSD601526S, RSD601540S, RSD601623S, RSD601636S, RSD601638S, RSD601647S, RSD601648S, RSD601656S, RSD601671S, RSD601675S, RSD601688S, RSD601692S, RSD601693S, RSD601697S, RSD601698S, RSD601699S, RSD601702S, RSD601703S, RSD601704S, RSD601707S, RSD601708S, RSD601710S, RSD601755S, RSD601764S, RSD601766S, RSD601767S, RSD601768S, RSD601776S, RSD601777S, RSD601779S, RSD601780S, RSD601781S, RSD601783S, RSD601787S, RSD601790S, RSD602125S, RSD602126S, RSD602131S, RSD602133S, RSD602148S, RSD602159S, RSD602436S, RSD602437S, RSD602442S, RSD602445S, RSD602449S, RSD602506S, RSD602507S, RSD602508S, RSD602577S, RSD602950S, RSD602955S, RSD602956S, RSD602962S, RSD602964S, RSD603076S, RSD603092S, RSD603113S, RSD603143S, RSD603206S, RSD603264S, RSD603268S, RSD603275S, RSD603282S, RSD603284S, RSD603435S, RSD603450S, RSD603454S, RSD603468S, RSD603472S, RSD603475S, RSD603476S, RSD603495S, RSD603921S, RSD603931S, RSD604115S, RSD604146S, RSD604153S, RSD604157S, RSD604167S, RSD604174S, RSD604177S, RSD604198S, RSD604204S, RSD604234S, RSD604255S, RSD604256S, RSD604262S, RSD604292S; GTIN none, Serial Numbers: RSD602523S, RSD603425S, RSD603492S, RSD603950S, RSD603971S, RSD604109S; b. Model Numbers: DDPA2D1: GTIN 00763000178338, Serial Numbers: RSL600237S, RSL600432S, RSL600496S, RSL600509S, RSL600666S, RSL600825S, RSL600826S, RSL601330S, RSL601357S, RSL601362S, RSL601367S, RSL601368S, RSL601373S, RSL601385S, RSL601430S, RSL601438S, RSL601444S, RSL601482S, RSL601490S, RSL601491S, RSL601500S, RSL601501S, RSL601504S, RSL601506S, RSL601510S, RSL601520S, RSL601526S, RSL601529S, RSL601535S, RSL601538S, RSL601540S, RSL601545S, RSL601549S, RSL601559S, RSL601617S, RSL601618S, RSL601643S, RSL601651S, RSL601652S, RSL601667S, RSL601670S, RSL601673S, RSL601676S, RSL601677S, RSL601679S, RSL601680S, RSL601681S, RSL601688S, RSL601694S, RSL601708S, RSL601709S, RSL601716S, RSL601721S, RSL601737S, RSL601744S, RSL601761S, RSL601762S, RSL601767S, RSL601769S, RSL601784S, RSL601798S, RSL601800S, RSL601806S, RSL601810S, RSL601814S, RSL601821S, RSL601825S, RSL601833S, RSL601850S, RSL601856S, RSL601860S, RSL601869S, RSL601877S, RSL601942S, RSL601945S, RSL601951S, RSL601953S, RSL601954S, RSL601956S, RSL601962S, RSL601963S, RSL601965S, RSL601986S, RSL601991S, RSL601993S, RSL601994S, RSL601995S, RSL602015S, RSL602026S, RSL602027S, RSL602028S, RSL602029S, RSL602031S, RSL602033S, RSL602034S, RSL602039S, RSL602042S, RSL602049S, RSL602056S, RSL602060S, RSL602063S, RSL602066S, RSL602068S, RSL602070S, RSL602071S, RSL602072S, RSL602075S, RSL602076S, RSL602078S, RSL602079S, RSL602080S, RSL602085S; GTIN 00763000178352, Serial Numbers: RSL600037S, RSL600042S, RSL600043S, RSL600044S, RSL600063S, RSL600064S, RSL600066S, RSL600069S, RSL600077S, RSL600082S, RSL600083S, RSL600107S, RSL600124S, RSL600158S, RSL600170S, RSL600183S, RSL600408S, RSL600450S, RSL600696S, RSL600805S, RSL600807S, RSL600812S, RSL600813S, RSL600815S, RSL600819S, RSL600822S, RSL600833S, RSL600840S, RSL600846S, RSL600868S, RSL600869S, RSL600874S, RSL600875S, RSL600878S, RSL600882S, RSL600884S, RSL600894S, RSL600896S, RSL600901S, RSL600906S, RSL600910S, RSL600922S, RSL600930S, RSL600940S, RSL600951S, RSL600955S, RSL600958S, RSL600963S, RSL600972S, RSL600981S, RSL600993S, RSL600998S, RSL600999S, RSL601004S, RSL601005S, RSL601006S, RSL601007S, RSL601008S, RSL601009S, RSL601010S, RSL601012S, RSL601013S, RSL601015S, RSL601016S, RSL601017S, RSL601019S, RSL601020S, RSL601023S, RSL601024S, RSL601028S, RSL601029S, RSL601031S, RSL601034S, RSL601035S, RSL601037S, RSL601039S, RSL601040S, RSL601042S, RSL601045S, RSL601046S, RSL601047S, RSL601069S, RSL601070S, RSL601071S, RSL601072S, RSL601075S, RSL601076S, RSL601081S, RSL601082S, RSL601084S, RSL601089S, RSL601091S, RSL601098S, RSL601099S, RSL601101S, RSL601102S, RSL601108S, RSL601109S, RSL601110S, RSL601113S, RSL601118S, RSL601120S, RSL601123S, RSL601124S, RSL601126S, RSL601127S, RSL601129S, RSL601130S, RSL601131S, RSL601132S, RSL601134S, RSL601137S, RSL601138S, RSL601139S, RSL601140S, RSL601141S, RSL601142S, RSL601143S, RSL601144S, RSL601145S, RSL601146S, RSL601147S, RSL601150S, RSL601153S, RSL601155S, RSL601156S, RSL601157S, RSL601158S, RSL601161S, RSL601166S, RSL601170S, RSL601174S, RSL601175S, RSL601180S, RSL601181S, RSL601182S, RSL601185S, RSL601186S, RSL601187S, RSL601197S, RSL601199S, RSL601203S, RSL601204S, RSL601206S, RSL601207S, RSL601208S, RSL601211S, RSL601219S, RSL601229S, RSL601232S, RSL601233S, RSL601239S.RSL601241S, RSL601257S, RSL601258S, RSL601259S, RSL601261S, RSL601262S, RSL601263S, RSL601264S, RSL601265S, RSL601266S, RSL601267S, RSL601270S, RSL601271S, RSL601274S, RSL601275S, RSL601276S, RSL601283S, RSL601311S, RSL601312S, RSL601314S, RSL601320S, RSL601322S, RSL601323S, RSL601327S, RSL601329S, RSL601352S, RSL601354S, RSL601356S, RSL601358S, RSL601359S, RSL601360S, RSL601375S, RSL601388S, RSL601397S, RSL601398S, RSL601399S, RSL601400S, RSL601401S, RSL601402S, RSL601403S, RSL601404S, RSL601409S, RSL601413S, RSL601427S, RSL601432S, RSL601449S, RSL601450S, RSL601451S, RSL601452S, RSL601459S, RSL601460S, RSL601461S, RSL601463S, RSL601465S, RSL601466S, RSL601467S, RSL601468S, RSL601469S, RSL601471S, RSL601472S, RSL601473S, RSL601474S, RSL601476S, RSL601479S, RSL601483S, RSL601487S, RSL601488S, RSL601521S, RSL601528S, RSL601530S, RSL601531S, RSL601536S, RSL601537S, RSL601541S, RSL601543S, RSL601560S, RSL601561S, RSL601562S, RSL601563S, RSL601565S, RSL601566S, RSL601567S, RSL601568S, RSL601569S, RSL601574S, RSL601575S, RSL601576S, RSL601577S, RSL601578S, RSL601579S, RSL601580S, RSL601581S, RSL601583S, RSL601584S, RSL601586S, RSL601587S, RSL601589S, RSL601590S, RSL601592S, RSL601593S, RSL601596S, RSL601597S, RSL601598S, RSL601601S, RSL601602S, RSL601604S, RSL601607S, RSL601881S, RSL601885S, RSL601901S, RSL601913S, RSL601915S, RSL601922S, RSL602004S, RSL602020S, RSL602022S, RSL602024S, RSL602087S, RSL602088S, RSL602089S, RSL602090S, RSL602104S, RSL602111S, RSL602112S, RSL602113S, RSL602114S, RSL602115S; GTIN 00763000380366, Serial Numbers: RSL600127S, RSL600128S, RSL600131S, RSL600132S, RSL600134S, RSL600135S, RSL600136S, RSL600137S, RSL600138S, RSL600139S, RSL600140S, RSL600143S, RSL600144S, RSL600145S, RSL600146S, RSL600325S, RSL600476S, RSL600731S, RSL600732S, RSL600734S, RSL600735S, RSL600737S, RSL600738S, RSL600739S, RSL600740S, RSL600742S, RSL600744S, RSL600745S, RSL600747S, RSL600748S, RSL600749S, RSL600750S, RSL600752S, RSL600756S, RSL600760S, RSL600763S, RSL600766S, RSL600767S, RSL600768S, RSL600773S, RSL600774S, RSL600775S, RSL600776S, RSL600780S, RSL600783S, RSL600784S, RSL600786S, RSL600791S, RSL600792S, RSL600796S, RSL600799S, RSL600800S, RSL600829S, RSL600830S, RSL600831S, RSL600832S, RSL600834S, RSL600835S, RSL600836S, RSL600837S, RSL600838S, RSL600839S, RSL600841S, RSL600843S, RSL600844S, RSL600847S, RSL600849S, RSL600850S, RSL600851S, RSL600852S, RSL600853S, RSL600854S, RSL600856S, RSL600859S, RSL600861S, RSL600862S, RSL600863S, RSL600864S, RSL600865S, RSL600866S, RSL600867S, RSL600872S, RSL600877S, RSL600880S, RSL600881S, RSL600883S, RSL600885S, RSL600886S, RSL600887S, RSL600888S, RSL600889S, RSL600890S, RSL600892S, RSL600893S, RSL600895S, RSL600897S, RSL600898S, RSL600899S, RSL600900S, RSL600902S, RSL600903S, RSL600905S, RSL600907S, RSL600908S, RSL600909S, RSL600911S, RSL600912S, RSL600913S, RSL600914S, RSL600916S, RSL600917S, RSL600918S, RSL600920S, RSL600921S, RSL600923S, RSL600924S, RSL600925S, RSL600926S, RSL600927S, RSL600928S, RSL600929S, RSL600931S, RSL600932S, RSL600933S, RSL600936S, RSL600937S, RSL600938S, RSL600941S, RSL600942S, RSL600943S, RSL600944S, RSL600946S, RSL600947S, RSL600948S, RSL600950S, RSL600952S, RSL600953S, RSL600956S, RSL600957S, RSL600959S, RSL600960S, RSL600961S, RSL600962S, RSL600965S, RSL600966S, RSL600967S, RSL600968S, RSL600969S, RSL600971S, RSL600973S, RSL600974S, RSL600975S, RSL600977S, RSL600978S, RSL600980S, RSL600982S, RSL600983S, RSL600984S, RSL600986S, RSL600987S, RSL600988S, RSL600989S, RSL600991S, RSL600992S, RSL600994S, RSL600995S, RSL600996S, RSL600997S, RSL601073S, RSL601188S, RSL601189S, RSL601191S, RSL601194S, RSL601200S, RSL601201S, RSL601205S, RSL601215S, RSL601217S, RSL601221S, RSL601222S, RSL601223S, RSL601227S, RSL601228S, RSL601230S, RSL601237S, RSL601238S, RSL601243S, RSL601244S, RSL601247S, RSL601250S, RSL601251S, RSL601252S, RSL601254S, RSL601255S, RSL601285S, RSL601286S, RSL601288S, RSL601289S, RSL601291S, RSL601292S, RSL601315S, RSL601316S, RSL601318S.RSL601319S, RSL601324S, RSL601326S, RSL601328S, RSL601390S, RSL601391S, RSL601392S, RSL601394S, RSL601395S, RSL601396S, RSL601411S, RSL601414S, RSL601415S, RSL601417S, RSL601418S, RSL601421S, RSL601422S, RSL601424S, RSL601425S, RSL601426S, RSL601428S, RSL601547S, RSL601548S, RSL601550S, RSL601551S, RSL601553S, RSL601556S, RSL601557S, RSL601558S, RSL601879S, RSL601880S, RSL601883S, RSL601884S, RSL601886S, RSL601888S, RSL601890S, RSL601891S, RSL601892S, RSL601893S, RSL601894S, RSL601897S, RSL601903S, RSL601909S, RSL601918S, RSL601923S, RSL601924S, RSL601926S, RSL602008S, RSL602092S, RSL602093S, RSL602094S, RSL602095S, RSL602100S, RSL602106S, RSL602121S, RSL602126S, RSL602127S, RSL602133S; GTIN none, Serial Numbers: RSL601920S; DDPA2D4: GTIN 00763000178345, Serial Numbers: RSM601171S, RSM604498S, RSM604564S, RSM604889S, RSM605730S, RSM605891S, RSM606070S, RSM606236S, RSM606893S, RSM607862S, RSM607870S, RSM607872S, RSM607875S, RSM607876S, RSM607929S, RSM608125S, RSM608239S, RSM608307S, RSM608330S, RSM608341S, RSM608344S, RSM608557S, RSM608697S, RSM608724S, RSM608725S, RSM608733S, RSM608738S, RSM608740S, RSM608748S, RSM608768S, RSM608788S, RSM608789S, RSM608796S, RSM608999S, RSM609011S, RSM609013S; GTIN 00763000178369, Serial Numbers: RSM600023S, RSM600026S, RSM600027S, RSM600029S, RSM600030S, RSM600031S, RSM600032S, RSM600034S, RSM600035S, RSM600036S, RSM600041S, RSM600042S, RSM600043S, RSM600050S, RSM600051S, RSM600052S, RSM600053S, RSM600054S, RSM600056S, RSM600057S, RSM600058S, RSM600059S, RSM600060S, RSM600061S, RSM600062S, RSM600063S, RSM600064S, RSM600074S, RSM600080S, RSM600081S, RSM600082S, RSM600083S, RSM600138S, RSM600455S, RSM600511S, RSM600522S, RSM600535S, RSM600567S, RSM600598S, RSM600608S, RSM600649S, RSM600650S, RSM600653S, RSM600654S, RSM600657S, RSM600659S, RSM600662S, RSM600665S, RSM600670S, RSM600673S, RSM600677S, RSM600682S, RSM600694S, RSM600699S, RSM600700S, RSM600706S, RSM600714S, RSM600720S, RSM600724S, RSM600734S, RSM600736S, RSM600737S, RSM600741S, RSM600782S, RSM600788S, RSM600797S, RSM600799S, RSM600805S, RSM600812S, RSM600823S, RSM600825S, RSM600826S, RSM600827S, RSM600828S, RSM600829S, RSM600906S, RSM600939S, RSM600991S, RSM600993S, RSM600997S, RSM601123S, RSM601179S, RSM601194S, RSM601197S, RSM601274S, RSM601275S, RSM601304S, RSM601306S, RSM601307S, RSM601324S, RSM601328S, RSM601329S, RSM601367S, RSM601368S, RSM601380S, RSM601385S, RSM601389S, RSM601423S, RSM601441S, RSM601450S, RSM601451S, RSM601454S, RSM601455S, RSM601456S, RSM601458S, RSM601459S, RSM601460S, RSM601461S, RSM601464S, RSM601465S, RSM601466S, RSM601467S, RSM601468S, RSM601470S, RSM601473S, RSM601474S, RSM601475S, RSM601476S, RSM601477S, RSM601478S, RSM601479S, RSM601484S, RSM601488S, RSM601490S, RSM601492S, RSM601499S, RSM601504S, RSM601510S, RSM601513S, RSM601519S, RSM601523S, RSM601524S, RSM601530S, RSM601531S, RSM601537S, RSM601539S, RSM601543S, RSM601544S, RSM601545S, RSM601546S, RSM601598S, RSM601600S, RSM601604S, RSM601605S, RSM601716S, RSM601718S, RSM601720S, RSM601721S, RSM601722S, RSM601723S, RSM601726S, RSM601727S, RSM601729S, RSM601734S, RSM601739S, RSM601742S, RSM601745S, RSM601746S, RSM601758S, RSM601765S, RSM601766S, RSM601767S, RSM601779S, RSM601780S, RSM601788S, RSM601794S, RSM601805S, RSM601807S, RSM601809S, RSM601819S, RSM601826S, RSM601832S, RSM601846S, RSM601850S, RSM601853S, RSM601870S, RSM601875S, RSM601876S, RSM601883S, RSM601894S, RSM601898S, RSM601899S, RSM601901S, RSM601905S, RSM601906S, RSM601907S, RSM601908S, RSM601909S, RSM601911S, RSM601912S, RSM601913S, RSM601914S, RSM601915S, RSM601916S, RSM601918S, RSM601919S, RSM601921S, RSM601922S, RSM601923S, RSM601924S, RSM601925S, RSM601926S, RSM601927S, RSM601929S, RSM601930S, RSM601931S, RSM601932S, RSM601933S, RSM601936S, RSM601938S, RSM601940S, RSM601943S, RSM601944S, RSM601945S, RSM601949S, RSM601950S, RSM601952S, RSM601953S, RSM601955S, RSM601958S, RSM601961S, RSM601962S, RSM601965S, RSM601966S, RSM601968S, RSM601972S, RSM601977S, RSM601979S.RSM601980S, RSM601982S, RSM601984S, RSM601985S, RSM601991S, RSM601992S, RSM601994S, RSM601995S, RSM601996S, RSM601999S, RSM602000S, RSM602001S, RSM602002S, RSM602006S, RSM602007S, RSM602009S, RSM602010S, RSM602011S, RSM602013S, RSM602014S, RSM602015S, RSM602018S, RSM602019S, RSM602020S, RSM602021S, RSM602024S, RSM602025S, RSM602027S, RSM602031S, RSM602032S, RSM602033S, RSM602034S, RSM602035S, RSM602036S, RSM602037S, RSM602038S, RSM602039S, RSM602040S, RSM602041S, RSM602043S, RSM602044S, RSM602045S, RSM602047S, RSM602048S, RSM602049S, RSM602050S, RSM602051S, RSM602052S, RSM602053S, RSM602054S, RSM602055S, RSM602056S, RSM602057S, RSM602058S, RSM602059S, RSM602061S, RSM602065S, RSM602069S, RSM602072S, RSM602079S, RSM602081S, RSM602082S, RSM602083S, RSM602084S, RSM602085S, RSM602087S, RSM602093S, RSM602094S, RSM602096S, RSM602098S, RSM602101S, RSM602103S, RSM602104S, RSM602107S, RSM602108S, RSM602110S, RSM602112S, RSM602116S, RSM602121S, RSM602122S, RSM602123S, RSM602124S, RSM602128S, RSM602129S, RSM602130S, RSM602131S, RSM602132S, RSM602134S, RSM602135S, RSM602137S, RSM602138S, RSM602141S, RSM602142S, RSM602143S, RSM602144S, RSM602145S, RSM602148S, RSM602150S, RSM602151S, RSM602157S, RSM602160S, RSM602162S, RSM602163S, RSM602164S, RSM602166S, RSM602167S, RSM602169S, RSM602172S, RSM602173S, RSM602174S, RSM602175S, RSM602176S, RSM602178S, RSM602179S, RSM602181S, RSM602185S, RSM602186S, RSM602190S, RSM602191S, RSM602192S, RSM602193S, RSM602195S, RSM602198S, RSM602199S, RSM602200S, RSM602201S, RSM602202S, RSM602203S, RSM602206S, RSM602207S, RSM602210S, RSM602213S, RSM602216S, RSM602217S, RSM602218S, RSM602220S, RSM602221S, RSM602225S, RSM602226S, RSM602228S, RSM602229S, RSM602231S, RSM602233S, RSM602234S, RSM602236S, RSM602237S, RSM602244S, RSM602247S, RSM602248S, RSM602251S, RSM602252S, RSM602253S, RSM602254S, RSM602255S, RSM602256S, RSM602257S, RSM602258S, RSM602259S, RSM602260S, RSM602262S, RSM602263S, RSM602264S, RSM602266S, RSM602267S, RSM602268S, RSM602269S, RSM602270S, RSM602278S, RSM602279S, RSM602280S, RSM602283S, RSM602287S, RSM602290S, RSM602291S, RSM602292S, RSM602293S, RSM602295S, RSM602297S, RSM602298S, RSM602299S, RSM602300S, RSM602301S, RSM602302S, RSM602306S, RSM602308S, RSM602309S, RSM602310S, RSM602311S, RSM602312S, RSM602317S, RSM602318S, RSM602319S, RSM602320S, RSM602322S, RSM602325S, RSM602326S, RSM602328S, RSM602329S, RSM602331S, RSM602336S, RSM602339S, RSM602340S, RSM602341S, RSM602342S, RSM602343S, RSM602345S, RSM602347S, RSM602348S, RSM602349S, RSM602352S, RSM602354S, RSM602356S, RSM602357S, RSM602358S, RSM602359S, RSM602363S, RSM602365S, RSM602366S, RSM602368S, RSM602369S, RSM602373S, RSM602374S, RSM602376S, RSM602377S, RSM602379S, RSM602380S, RSM602381S, RSM602382S, RSM602385S, RSM602386S, RSM602387S, RSM602388S, RSM602389S, RSM602390S, RSM602391S, RSM602396S, RSM602406S, RSM602421S, RSM602423S, RSM602425S, RSM602428S, RSM602434S, RSM602441S, RSM602447S, RSM602453S, RSM602457S, RSM602459S, RSM602463S, RSM602464S, RSM602465S, RSM602466S, RSM602472S, RSM602473S, RSM602474S, RSM602475S, RSM602478S, RSM602479S, RSM602482S, RSM602483S, RSM602484S, RSM602485S, RSM602486S, RSM602488S, RSM602489S, RSM602491S, RSM602492S, RSM602495S, RSM602497S, RSM602498S, RSM602500S, RSM602501S, RSM602505S, RSM602506S, RSM602509S, RSM602511S, RSM602512S, RSM602515S, RSM602516S, RSM602517S, RSM602518S, RSM602519S, RSM602520S, RSM602523S, RSM602524S, RSM602525S, RSM602526S, RSM602527S, RSM602530S, RSM602532S, RSM602533S, RSM602535S, RSM602536S, RSM602537S, RSM602538S, RSM602541S, RSM602543S, RSM602544S, RSM602545S, RSM602546S, RSM602548S, RSM602549S, RSM602552S, RSM602553S, RSM602554S, RSM602556S, RSM602557S, RSM602559S, RSM602561S, RSM602562S, RSM602563S, RSM602565S, RSM602568S, RSM602569S, RSM602570S, RSM602572S, RSM602577S, RSM602580S, RSM602581S, RSM602589S, RSM602590S, RSM602591S, RSM602594S, RSM602595S, RSM602599S, RSM602601S, RSM602602S, RSM602605S, RSM602607S, RSM602608S, RSM602609S, RSM602610S, RSM602613S, RSM602617S, RSM602619S, RSM602623S, RSM602624S, RSM602626S, RSM602627S, RSM602632S, RSM602633S, RSM602634S, RSM602635S, RSM602636S, RSM602637S, RSM602645S, RSM602646S, RSM602647S, RSM602648S, RSM602651S, RSM602653S, RSM602673S, RSM602674S, RSM602678S, RSM602681S, RSM602682S, RSM602685S, RSM602689S, RSM602691S, RSM602693S, RSM602697S, RSM602702S, RSM602707S, RSM602709S, RSM602711S, RSM602712S, RSM602713S, RSM602716S, RSM602724S, RSM602726S, RSM602728S, RSM602730S, RSM602732S, RSM602734S, RSM602735S, RSM602736S, RSM602737S, RSM602739S, RSM602740S, RSM602741S, RSM602742S, RSM602743S, RSM602744S, RSM602745S, RSM602746S, RSM602747S, RSM602748S, RSM602749S, RSM602751S, RSM602752S, RSM602754S, RSM602755S, RSM602758S, RSM602759S, RSM602760S, RSM602761S, RSM602765S, RSM602767S, RSM602768S, RSM602769S, RSM602773S, RSM602776S, RSM602777S, RSM602782S, RSM602784S, RSM602786S, RSM602788S, RSM602791S, RSM602792S, RSM602793S, RSM602794S, RSM602795S, RSM602810S, RSM602811S, RSM602816S, RSM602817S, RSM602821S, RSM602822S, RSM602824S, RSM602828S, RSM602829S, RSM602830S, RSM602831S, RSM602833S, RSM602837S, RSM602838S, RSM602839S, RSM602841S, RSM602842S, RSM602848S, RSM602850S, RSM602851S, RSM602852S, RSM602853S, RSM602854S, RSM602856S, RSM602858S, RSM602859S, RSM602861S, RSM602862S, RSM602864S, RSM602865S, RSM602866S, RSM602867S, RSM602870S, RSM602871S, RSM602872S, RSM602873S, RSM602874S, RSM602877S, RSM602878S, RSM602880S, RSM602882S, RSM602883S, RSM602887S, RSM602889S, RSM602894S, RSM602896S, RSM602898S, RSM602899S, RSM602901S, RSM602903S, RSM602905S, RSM602907S, RSM602910S, RSM602911S, RSM602912S, RSM602916S, RSM602920S, RSM602921S, RSM602924S, RSM602927S, RSM602930S, RSM602931S, RSM602932S, RSM602933S, RSM602934S, RSM602935S, RSM602940S, RSM603035S, RSM603036S, RSM603037S, RSM603038S, RSM603039S, RSM603040S, RSM603041S, RSM603042S, RSM603043S, RSM603045S, RSM603046S, RSM603047S, RSM603048S, RSM603049S, RSM603050S, RSM603051S, RSM603052S, RSM603053S, RSM603054S, RSM603055S, RSM603056S, RSM603057S, RSM603058S, RSM603060S, RSM603062S, RSM603063S, RSM603064S, RSM603065S, RSM603066S, RSM603067S, RSM603068S, RSM603069S, RSM603070S, RSM603071S, RSM603072S, RSM603073S, RSM603074S, RSM603075S, RSM603076S, RSM603077S, RSM603078S, RSM603079S, RSM603080S, RSM603081S, RSM603083S, RSM603084S, RSM603085S, RSM603086S, RSM603087S, RSM603088S, RSM603089S, RSM603090S, RSM603091S, RSM603092S, RSM603093S, RSM603094S, RSM603095S, RSM603096S, RSM603097S, RSM603098S, RSM603099S, RSM603100S, RSM603101S, RSM603102S, RSM603103S, RSM603105S, RSM603106S, RSM603107S, RSM603108S, RSM603110S, RSM603111S, RSM603112S, RSM603113S, RSM603114S, RSM603115S, RSM603116S, RSM603117S, RSM603118S, RSM603119S, RSM603120S, RSM603121S, RSM603123S, RSM603124S, RSM603125S, RSM603126S, RSM603127S, RSM603128S, RSM603129S, RSM603130S, RSM603131S, RSM603132S, RSM603133S, RSM603134S, RSM603137S, RSM603139S, RSM603144S, RSM603145S, RSM603146S, RSM603147S, RSM603149S, RSM603150S, RSM603151S, RSM603153S, RSM603154S, RSM603155S, RSM603156S, RSM603157S, RSM603161S, RSM603162S, RSM603164S, RSM603167S, RSM603168S, RSM603169S, RSM603170S, RSM603171S, RSM603172S, RSM603177S, RSM603179S, RSM603180S, RSM603182S, RSM603183S, RSM603186S, RSM603187S, RSM603188S, RSM603189S, RSM603190S, RSM603191S, RSM603192S, RSM603194S, RSM603198S, RSM603205S, RSM603206S, RSM603211S, RSM603213S, RSM603219S, RSM603220S, RSM603221S, RSM603222S, RSM603223S, RSM603224S, RSM603225S, RSM603229S, RSM603231S, RSM603233S, RSM603236S, RSM603238S, RSM603239S, RSM603241S, RSM603248S, RSM603251S, RSM603258S, RSM603267S, RSM603275S, RSM603279S, RSM603281S, RSM603288S, RSM603307S, RSM603311S, RSM603312S, RSM603320S, RSM603338S, RSM603339S, RSM603340S, RSM603341S, RSM603342S, RSM603343S, RSM603350S, RSM603352S, RSM603353S, RSM603354S, RSM603355S, RSM603357S, RSM603362S, RSM603364S, RSM603365S, RSM603366S, RSM603367S, RSM603368S, RSM603374S, RSM603378S, RSM603383S, RSM603384S, RSM603385S, RSM603389S, RSM603390S, RSM603392S, RSM603394S, RSM603401S, RSM603423S, RSM603425S, RSM603426S, RSM603430S, RSM603431S, RSM603433S, RSM603434S, RSM603435S, RSM603439S, RSM603440S, RSM603441S, RSM603442S, RSM603443S, RSM603444S, RSM603452S, RSM603455S, RSM603463S, RSM603465S, RSM603483S, RSM603484S, RSM603486S, RSM603487S, RSM603488S, RSM603489S, RSM603495S, RSM603499S, RSM603500S, RSM603501S, RSM603502S, RSM603503S, RSM603504S, RSM603507S, RSM603508S, RSM603511S, RSM603515S, RSM603526S, RSM603533S, RSM603535S, RSM603537S, RSM603542S, RSM603545S, RSM603550S, RSM603551S, RSM603552S, RSM603556S, RSM603564S, RSM603569S, RSM603578S, RSM603579S, RSM603585S, RSM603586S, RSM603589S, RSM603594S, RSM603595S, RSM603605S, RSM603610S, RSM603613S, RSM603615S, RSM603620S, RSM603622S, RSM603626S, RSM603652S, RSM603706S, RSM603724S, RSM603752S, RSM603781S, RSM603810S, RSM603815S, RSM603820S, RSM603824S, RSM603825S, RSM603843S, RSM603848S, RSM603851S, RSM603856S, RSM603882S, RSM603893S, RSM603895S, RSM603896S, RSM603897S, RSM603898S, RSM603899S, RSM603900S, RSM603901S, RSM603902S, RSM603904S, RSM603905S, RSM603906S, RSM603907S, RSM603908S, RSM603909S, RSM603910S, RSM603912S, RSM603913S, RSM603914S, RSM603915S, RSM603916S, RSM603917S, RSM603920S, RSM603921S, RSM603922S, RSM603924S, RSM603925S, RSM603927S, RSM603928S, RSM603929S, RSM603930S, RSM603931S, RSM603932S, RSM603933S, RSM603934S, RSM603935S, RSM603936S, RSM603937S, RSM603938S, RSM603939S, RSM603940S, RSM603941S, RSM603942S, RSM603944S, RSM603945S, RSM603947S, RSM603948S, RSM603949S, RSM603950S, RSM603951S, RSM603952S, RSM603953S, RSM603954S, RSM603955S, RSM603956S, RSM603958S, RSM603959S, RSM603960S, RSM603961S, RSM603962S, RSM603964S, RSM603969S, RSM603975S, RSM603977S, RSM603979S, RSM603980S, RSM603981S, RSM603983S, RSM603986S, RSM603987S, RSM603993S, RSM604000S, RSM604005S, RSM604007S, RSM604009S, RSM604010S, RSM604019S, RSM604020S, RSM604026S, RSM604028S, RSM604029S, RSM604041S, RSM604042S, RSM604045S, RSM604047S, RSM604053S, RSM604059S, RSM604062S, RSM604063S, RSM604064S, RSM604067S, RSM604068S, RSM604072S, RSM604075S, RSM604078S, RSM604083S, RSM604084S, RSM604086S, RSM604089S, RSM604102S, RSM604105S, RSM604109S, RSM604111S, RSM604112S, RSM604117S, RSM604123S, RSM604130S, RSM604133S, RSM604140S, RSM604172S, RSM604233S, RSM604235S, RSM604236S, RSM604242S, RSM604243S, RSM604244S, RSM604247S, RSM604248S, RSM604249S, RSM604252S, RSM604256S, RSM604282S, RSM604290S, RSM604292S, RSM604294S, RSM604296S, RSM604298S, RSM604312S, RSM604316S, RSM604319S, RSM604328S, RSM604331S, RSM604336S, RSM604346S, RSM604349S, RSM604357S, RSM604365S, RSM604385S, RSM604387S, RSM604393S, RSM604442S, RSM604443S, RSM604459S, RSM604479S, RSM604527S, RSM604548S, RSM604586S, RSM604592S, RSM604605S, RSM604616S, RSM604618S, RSM604619S, RSM604645S, RSM604652S, RSM604655S, RSM604656S, RSM604659S, RSM604662S, RSM604664S, RSM604666S, RSM604671S, RSM604681S, RSM604693S, RSM604702S, RSM604712S, RSM604713S, RSM604716S, RSM604736S, RSM604737S, RSM604748S, RSM604856S, RSM604983S, RSM604986S, RSM604987S, RSM604988S, RSM605019S, RSM605020S, RSM605026S, RSM605034S, RSM605035S, RSM605042S, RSM605052S, RSM605069S, RSM605075S, RSM605087S, RSM605095S, RSM605096S, RSM605097S, RSM605098S, RSM605099S, RSM605100S, RSM605101S, RSM605102S, RSM605103S, RSM605104S, RSM605105S, RSM605106S, RSM605107S, RSM605108S, RSM605114S, RSM605116S, RSM605117S, RSM605118S, RSM605119S, RSM605120S, RSM605121S, RSM605122S, RSM605123S, RSM605124S, RSM605125S, RSM605126S, RSM605127S, RSM605128S, RSM605129S, RSM605131S, RSM605132S, RSM605133S, RSM605135S, RSM605136S, RSM605137S, RSM605138S, RSM605139S, RSM605142S, RSM605144S, RSM605145S, RSM605147S, RSM605148S, RSM605150S, RSM605161S, RSM605162S, RSM605163S, RSM605166S, RSM605205S, RSM605207S, RSM605209S, RSM605210S, RSM605211S, RSM605212S, RSM605215S, RSM605218S, RSM605219S, RSM605221S, RSM605224S, RSM605226S, RSM605227S, RSM605228S, RSM605232S, RSM605234S, RSM605235S, RSM605236S, RSM605240S, RSM605241S, RSM605246S, RSM605247S, RSM605248S, RSM605249S, RSM605250S, RSM605251S.RSM605252S, RSM605253S, RSM605254S, RSM605255S, RSM605256S, RSM605259S, RSM605260S, RSM605261S, RSM605262S, RSM605263S, RSM605264S, RSM605265S, RSM605266S, RSM605267S, RSM605268S, RSM605269S, RSM605270S, RSM605271S, RSM605272S, RSM605275S, RSM605276S, RSM605294S, RSM605296S, RSM605331S, RSM605332S, RSM605333S, RSM605334S, RSM605335S, RSM605336S, RSM605337S, RSM605338S, RSM605339S, RSM605340S, RSM605341S, RSM605342S, RSM605343S, RSM605344S, RSM605345S, RSM605346S, RSM605347S, RSM605348S, RSM605349S, RSM605350S, RSM605351S, RSM605352S, RSM605353S, RSM605355S, RSM605356S, RSM605357S, RSM605358S, RSM605359S, RSM605360S, RSM605361S, RSM605362S, RSM605363S, RSM605364S, RSM605365S, RSM605366S, RSM605367S, RSM605368S, RSM605369S, RSM605370S, RSM605371S, RSM605372S, RSM605373S, RSM605374S, RSM605375S, RSM605376S, RSM605377S, RSM605378S, RSM605379S, RSM605380S, RSM605381S, RSM605382S, RSM605383S, RSM605384S, RSM605385S, RSM605386S, RSM605387S, RSM605388S, RSM605389S, RSM605390S, RSM605391S, RSM605393S, RSM605395S, RSM605396S, RSM605397S, RSM605398S, RSM605399S, RSM605401S, RSM605402S, RSM605403S, RSM605404S, RSM605405S, RSM605406S, RSM605407S, RSM605408S, RSM605409S, RSM605410S, RSM605411S, RSM605412S, RSM605413S, RSM605414S, RSM605415S, RSM605416S, RSM605417S, RSM605418S, RSM605419S, RSM605420S, RSM605421S, RSM605422S, RSM605423S, RSM605424S, RSM605425S, RSM605426S, RSM605427S, RSM605428S, RSM605429S, RSM605430S, RSM605431S, RSM605432S, RSM605433S, RSM605434S, RSM605435S, RSM605436S, RSM605437S, RSM605438S, RSM605439S, RSM605440S, RSM605441S, RSM605442S, RSM605443S, RSM605444S, RSM605445S, RSM605446S, RSM605447S, RSM605448S, RSM605449S, RSM605450S, RSM605451S, RSM605452S, RSM605453S, RSM605454S, RSM605455S, RSM605456S, RSM605457S, RSM605458S, RSM605459S, RSM605460S, RSM605461S, RSM605462S, RSM605463S, RSM605464S, RSM605465S, RSM605466S, RSM605470S, RSM605471S, RSM605472S, RSM605473S, RSM605474S, RSM605475S, RSM605477S, RSM605479S, RSM605480S, RSM605481S, RSM605483S, RSM605484S, RSM605486S, RSM605487S, RSM605488S, RSM605489S, RSM605490S, RSM605493S, RSM605494S, RSM605495S, RSM605496S, RSM605497S, RSM605498S, RSM605499S, RSM605500S, RSM605501S, RSM605502S, RSM605503S, RSM605505S, RSM605576S, RSM605578S, RSM605657S, RSM605677S, RSM605687S, RSM605691S, RSM605699S, RSM605706S, RSM606066S, RSM606067S, RSM606068S, RSM606069S, RSM606071S, RSM606072S, RSM606076S, RSM606077S, RSM606082S, RSM606083S, RSM606084S, RSM606086S, RSM606090S, RSM606091S, RSM606092S, RSM606094S, RSM606095S, RSM606096S, RSM606097S, RSM606098S, RSM606099S, RSM606100S, RSM606102S, RSM606105S, RSM606106S, RSM606108S, RSM606111S, RSM606112S, RSM606114S, RSM606115S, RSM606116S, RSM606117S, RSM606118S, RSM606120S, RSM606121S, RSM606122S, RSM606127S, RSM606128S, RSM606129S, RSM606131S, RSM606132S, RSM606142S, RSM606144S, RSM606155S, RSM606160S, RSM606162S, RSM606163S, RSM606164S, RSM606165S, RSM606177S, RSM606178S, RSM606179S, RSM606181S, RSM606183S, RSM606189S, RSM606191S, RSM606194S, RSM606201S, RSM606203S, RSM606210S, RSM606220S, RSM606223S, RSM606232S, RSM606234S, RSM606235S, RSM606244S, RSM606246S, RSM606249S, RSM606285S, RSM606286S, RSM606366S, RSM606377S, RSM606393S, RSM606398S, RSM606408S, RSM606412S, RSM606419S, RSM606422S, RSM606423S, RSM606438S, RSM606446S, RSM606451S, RSM606452S, RSM606476S, RSM606478S, RSM606485S, RSM606496S, RSM606530S, RSM606578S, RSM606596S, RSM606699S, RSM606700S, RSM606702S, RSM606703S, RSM606704S, RSM606706S, RSM606707S, RSM606709S, RSM606712S, RSM606722S, RSM606726S, RSM606732S, RSM606738S, RSM606740S, RSM606745S, RSM606747S, RSM606768S, RSM606780S, RSM606788S, RSM606791S, RSM606802S, RSM606805S, RSM606806S, RSM606808S, RSM606816S, RSM606817S, RSM606846S, RSM606853S, RSM606861S, RSM606879S, RSM606894S, RSM606896S, RSM606897S, RSM606901S, RSM606903S, RSM606904S, RSM606924S, RSM606930S, RSM606937S, RSM606941S, RSM606942S, RSM606944S, RSM606947S, RSM606948S, RSM606949S, RSM606950S, RSM606952S, RSM606955S, RSM606958S, RSM606962S, RSM607008S, RSM607010S, RSM607022S, RSM607049S, RSM607054S, RSM607055S, RSM607073S, RSM607127S, RSM607141S, RSM607145S, RSM607149S, RSM607152S, RSM607157S, RSM607160S, RSM607162S, RSM607163S, RSM607166S, RSM607167S, RSM607180S, RSM607181S, RSM607183S, RSM607220S, RSM607263S, RSM607302S, RSM607303S, RSM607306S, RSM607308S, RSM607313S, RSM607314S, RSM607318S, RSM607319S, RSM607320S, RSM607321S, RSM607322S, RSM607323S, RSM607324S, RSM607326S, RSM607327S, RSM607328S, RSM607330S, RSM607331S, RSM607332S, RSM607333S, RSM607334S, RSM607335S, RSM607336S, RSM607338S, RSM607339S, RSM607344S, RSM607345S, RSM607346S, RSM607356S, RSM607357S, RSM607359S, RSM607363S, RSM607365S, RSM607366S, RSM607367S, RSM607368S, RSM607369S, RSM607370S, RSM607375S, RSM607376S, RSM607377S, RSM607379S, RSM607380S, RSM607382S, RSM607386S, RSM607388S, RSM607391S, RSM607393S, RSM607399S, RSM607403S, RSM607411S, RSM607428S, RSM607488S, RSM607489S, RSM607494S, RSM607495S, RSM607498S, RSM607504S, RSM607507S, RSM607512S, RSM607518S, RSM607519S, RSM607521S, RSM607523S, RSM607528S, RSM607530S, RSM607533S, RSM607535S, RSM607536S, RSM607537S, RSM607538S, RSM607539S, RSM607541S, RSM607543S, RSM607544S, RSM607545S, RSM607549S, RSM607576S, RSM607579S, RSM607728S, RSM607733S, RSM607734S, RSM607739S, RSM607743S, RSM607747S, RSM607790S, RSM607936S, RSM608061S, RSM608096S, RSM608228S, RSM608264S, RSM608265S, RSM608269S, RSM608270S, RSM608272S, RSM608274S, RSM608276S, RSM608278S, RSM608280S, RSM608282S, RSM608296S, RSM608373S, RSM608391S, RSM608392S, RSM608393S, RSM608394S, RSM608395S, RSM608396S, RSM608397S, RSM608400S, RSM608403S, RSM608404S, RSM608405S, RSM608406S, RSM608408S, RSM608411S, RSM608413S, RSM608414S, RSM608418S, RSM608419S, RSM608420S, RSM608421S, RSM608422S, RSM608435S, RSM608444S, RSM608450S, RSM608452S, RSM608456S, RSM608469S, RSM608470S, RSM608478S, RSM608488S, RSM608491S, RSM608492S, RSM608495S, RSM608498S, RSM608501S, RSM608503S, RSM608505S, RSM608507S, RSM608509S, RSM608510S, RSM608513S, RSM608514S, RSM608515S, RSM608517S, RSM608518S, RSM608520S, RSM608522S, RSM608526S, RSM608530S, RSM608531S, RSM608532S, RSM608534S, RSM608537S, RSM608538S, RSM608539S, RSM608540S, RSM608542S, RSM608560S, RSM608561S, RSM608576S, RSM608579S, RSM608582S, RSM608591S, RSM608595S, RSM608596S, RSM608600S, RSM608601S, RSM608604S, RSM608607S, RSM608608S, RSM608609S, RSM608611S, RSM608612S, RSM608613S, RSM608617S, RSM608620S, RSM608621S, RSM608622S, RSM608623S, RSM608625S, RSM608626S, RSM608628S, RSM608629S, RSM608632S, RSM608642S, RSM608648S, RSM608649S, RSM608651S, RSM608653S, RSM608656S, RSM608665S, RSM608667S, RSM608669S, RSM608675S, RSM608679S, RSM608684S, RSM608779S, RSM608854S, RSM608855S, RSM608857S, RSM608859S, RSM608864S, RSM608865S, RSM608867S, RSM608869S, RSM608873S, RSM608876S, RSM608880S, RSM608883S, RSM608886S, RSM608887S, RSM608893S, RSM608904S, RSM608907S, RSM608914S, RSM608915S, RSM608916S, RSM608917S, RSM608923S, RSM608924S, RSM608927S, RSM608930S, RSM608934S, RSM608935S, RSM608936S, RSM608937S, RSM608943S, RSM608944S, RSM608948S, RSM608952S, RSM608963S, RSM608970S, RSM608971S, RSM608979S, RSM608982S, RSM608983S, RSM608987S, RSM609022S, RSM609025S, RSM609026S, RSM609029S, RSM609030S, RSM609032S, RSM609053S, RSM609058S, RSM609065S, RSM609069S, RSM609084S, RSM609087S, RSM609088S, RSM609090S, RSM609093S, RSM609095S, RSM609097S, RSM609100S, RSM609102S, RSM609103S, RSM609105S, RSM609106S, RSM609107S, RSM609110S, RSM609117S, RSM609118S, RSM609119S, RSM609122S, RSM609123S, RSM609126S, RSM609129S, RSM609135S, RSM609196S, RSM609198S, RSM609199S, RSM609202S, RSM609203S, RSM609207S, RSM609226S, RSM609230S, RSM609246S, RSM609250S, RSM609251S, RSM609253S, RSM609254S, RSM609256S, RSM613229S; GTIN 00763000380373, Serial Numbers: RSM600038S, RSM600039S, RSM600065S, RSM600066S, RSM600067S, RSM600068S, RSM600069S, RSM600070S, RSM600071S, RSM600072S, RSM600073S, RSM600075S, RSM600077S, RSM600078S, RSM600079S, RSM600084S, RSM600085S, RSM600086S, RSM600087S, RSM600089S, RSM600090S, RSM600091S, RSM600092S, RSM600093S, RSM600094S, RSM600095S, RSM600096S, RSM600097S, RSM600098S, RSM600099S, RSM600100S, RSM600101S, RSM600102S, RSM600103S, RSM600104S, RSM600105S, RSM600106S, RSM600107S, RSM600108S, RSM600109S, RSM600376S, RSM600385S, RSM600502S, RSM600518S, RSM600519S, RSM600520S, RSM600523S, RSM600524S, RSM600525S, RSM600526S, RSM600527S, RSM600528S, RSM600529S, RSM600530S, RSM600531S, RSM600532S, RSM600533S, RSM600534S, RSM600536S, RSM600538S, RSM600539S, RSM600540S, RSM600541S, RSM600543S, RSM600544S, RSM600545S, RSM600548S, RSM600549S, RSM600551S, RSM600552S, RSM600553S, RSM600554S, RSM600555S, RSM600557S, RSM600559S, RSM600560S, RSM600561S, RSM600562S, RSM600563S, RSM600564S, RSM600565S, RSM600568S, RSM600569S, RSM600570S, RSM600571S, RSM600572S, RSM600573S, RSM600574S, RSM600575S, RSM600576S, RSM600577S, RSM600578S, RSM600579S, RSM600581S, RSM600582S, RSM600583S, RSM600585S, RSM600586S, RSM600587S, RSM600588S, RSM600589S, RSM600590S, RSM600591S, RSM600592S, RSM600593S, RSM600594S, RSM600599S, RSM600600S, RSM600602S, RSM600604S, RSM600605S, RSM600606S, RSM600607S, RSM600609S, RSM600610S, RSM600611S, RSM600612S, RSM600613S, RSM600614S, RSM600615S, RSM600617S, RSM600618S, RSM600619S, RSM600620S, RSM600621S, RSM600622S, RSM600623S, RSM600624S, RSM600626S, RSM600627S, RSM600628S, RSM600629S, RSM600632S, RSM600634S, RSM600635S, RSM600642S, RSM600645S, RSM600648S, RSM600941S, RSM601180S, RSM601181S, RSM601182S, RSM601183S, RSM601184S, RSM601185S, RSM601186S, RSM601188S, RSM601189S, RSM601190S, RSM601191S, RSM601192S, RSM601193S, RSM601195S, RSM601196S, RSM601198S, RSM601199S, RSM601200S, RSM601201S, RSM601202S, RSM601203S, RSM601204S, RSM601205S, RSM601206S, RSM601207S, RSM601208S, RSM601209S, RSM601210S, RSM601211S, RSM601212S, RSM601213S, RSM601214S, RSM601215S, RSM601216S, RSM601217S, RSM601218S, RSM601219S, RSM601220S, RSM601221S, RSM601222S, RSM601223S, RSM601224S, RSM601225S, RSM601226S, RSM601227S, RSM601228S, RSM601229S, RSM601230S, RSM601231S, RSM601232S, RSM601233S, RSM601234S, RSM601235S, RSM601236S, RSM601237S, RSM601238S, RSM601239S, RSM601240S, RSM601241S, RSM601242S, RSM601243S, RSM601244S, RSM601245S, RSM601246S, RSM601247S, RSM601248S, RSM601249S, RSM601250S, RSM601251S, RSM601252S, RSM601253S, RSM601254S, RSM601255S, RSM601256S, RSM601258S, RSM601259S, RSM601260S, RSM601261S, RSM601262S, RSM601263S, RSM601264S, RSM601265S, RSM601266S, RSM601267S, RSM601268S, RSM601269S, RSM601270S, RSM601271S, RSM601272S, RSM601273S, RSM601276S, RSM601277S, RSM601278S, RSM601279S, RSM601280S, RSM601281S, RSM601282S, RSM601283S, RSM601284S, RSM601285S, RSM601286S, RSM601287S, RSM601288S, RSM601289S, RSM601290S, RSM601291S, RSM601293S, RSM601294S, RSM601295S, RSM601297S, RSM601298S, RSM601299S, RSM601300S, RSM601301S, RSM601302S, RSM601303S, RSM601305S, RSM601308S, RSM601309S, RSM601310S, RSM601311S, RSM601312S, RSM601313S, RSM601314S, RSM601315S, RSM601316S, RSM601317S, RSM601318S, RSM601319S, RSM601320S, RSM601321S, RSM601322S, RSM601323S, RSM601326S, RSM601327S, RSM601330S, RSM601331S, RSM601332S, RSM601333S, RSM601334S, RSM601335S, RSM601336S, RSM601337S, RSM601338S, RSM601339S, RSM601340S, RSM601341S, RSM601342S, RSM601343S, RSM601344S, RSM601345S, RSM601346S, RSM601347S, RSM601348S, RSM601349S, RSM601350S, RSM601351S, RSM601352S, RSM601353S, RSM601354S, RSM601355S, RSM601356S, RSM601357S, RSM601358S, RSM601359S, RSM601360S, RSM601361S, RSM601362S, RSM601363S, RSM601364S, RSM601365S, RSM601366S, RSM601369S, RSM601370S, RSM601371S, RSM601372S, RSM601373S, RSM601374S, RSM601375S, RSM601376S, RSM601377S, RSM601378S, RSM601379S, RSM601381S, RSM601382S, RSM601383S, RSM601384S, RSM601386S, RSM601387S, RSM601388S, RSM601390S, RSM601391S, RSM601392S, RSM601394S, RSM601395S, RSM601396S, RSM601398S, RSM601399S, RSM601400S, RSM601401S, RSM601402S, RSM601403S, RSM601404S, RSM601405S, RSM601406S, RSM601407S, RSM601408S, RSM601409S, RSM601410S, RSM601411S, RSM601412S, RSM601414S, RSM601415S, RSM601416S, RSM601417S, RSM601418S, RSM601419S, RSM601420S, RSM601421S, RSM601422S, RSM601424S, RSM601425S, RSM601427S, RSM601428S, RSM601429S, RSM601430S, RSM601431S, RSM601432S, RSM601433S, RSM601434S, RSM601435S, RSM601436S, RSM601437S, RSM601438S, RSM601439S, RSM601440S, RSM601442S, RSM601443S, RSM601444S, RSM601445S, RSM601446S, RSM601448S, RSM601449S, RSM601452S, RSM601457S, RSM601462S, RSM601463S, RSM601469S, RSM601471S, RSM601480S, RSM601481S, RSM601483S, RSM601485S, RSM601486S, RSM601487S, RSM601489S, RSM601491S, RSM601493S, RSM601494S, RSM601495S, RSM601496S, RSM601497S, RSM601498S, RSM601500S, RSM601501S, RSM601502S, RSM601503S, RSM601505S, RSM601506S, RSM601507S, RSM601508S, RSM601511S, RSM601512S, RSM601514S, RSM601515S, RSM601516S, RSM601517S, RSM601518S, RSM601520S, RSM601521S, RSM601522S, RSM601525S, RSM601526S, RSM601527S, RSM601532S, RSM601533S, RSM601534S, RSM601535S, RSM601536S, RSM601538S, RSM601540S, RSM601541S, RSM601542S, RSM601715S, RSM601772S, RSM601782S, RSM601796S, RSM601800S, RSM601816S, RSM601820S, RSM601822S, RSM601829S, RSM601830S, RSM601831S, RSM601835S, RSM601837S, RSM601841S, RSM601864S, RSM601865S, RSM601866S, RSM601867S, RSM601868S, RSM601869S, RSM601871S, RSM601872S, RSM601873S, RSM601874S, RSM601877S, RSM601878S, RSM601879S, RSM601880S, RSM601881S, RSM601882S, RSM601885S, RSM601886S, RSM601887S, RSM601888S, RSM601889S, RSM601890S, RSM601891S, RSM601892S, RSM601893S, RSM601895S, RSM601896S, RSM601897S, RSM601900S, RSM601902S, RSM601903S, RSM601904S, RSM601917S, RSM601920S, RSM601928S, RSM601934S, RSM601935S, RSM601937S, RSM601939S, RSM601941S, RSM601942S, RSM601946S, RSM601947S, RSM601948S, RSM601951S, RSM601954S, RSM601956S, RSM601957S, RSM601959S, RSM601960S, RSM601963S, RSM601964S, RSM601967S, RSM601970S, RSM601971S, RSM601973S, RSM601974S, RSM601975S, RSM601976S, RSM601978S, RSM601981S, RSM601983S, RSM601987S, RSM601989S, RSM601990S, RSM601993S, RSM601997S, RSM601998S, RSM602003S, RSM602004S, RSM602005S, RSM602008S, RSM602012S, RSM602016S, RSM602017S, RSM602022S, RSM602023S, RSM602026S, RSM602028S, RSM602030S, RSM602042S, RSM602046S, RSM602060S, RSM602062S, RSM602063S, RSM602064S, RSM602066S, RSM602067S, RSM602068S, RSM602070S, RSM602071S, RSM602073S, RSM602074S, RSM602075S, RSM602076S, RSM602077S, RSM602078S, RSM602080S, RSM602086S, RSM602089S, RSM602090S, RSM602091S, RSM602092S, RSM602095S, RSM602097S, RSM602099S, RSM602100S, RSM602102S, RSM602105S, RSM602106S, RSM602109S, RSM602111S, RSM602113S, RSM602114S, RSM602115S, RSM602117S, RSM602118S, RSM602119S, RSM602120S, RSM602125S, RSM602126S, RSM602127S, RSM602133S, RSM602136S, RSM602139S, RSM602140S, RSM602146S, RSM602147S, RSM602149S, RSM602152S, RSM602153S, RSM602154S, RSM602155S, RSM602156S, RSM602158S, RSM602159S, RSM602161S, RSM602165S, RSM602168S, RSM602170S, RSM602171S, RSM602177S, RSM602180S, RSM602182S, RSM602183S, RSM602184S, RSM602187S, RSM602188S, RSM602189S, RSM602194S, RSM602196S, RSM602197S, RSM602204S, RSM602205S, RSM602208S, RSM602209S, RSM602211S, RSM602212S, RSM602214S, RSM602215S, RSM602219S, RSM602222S, RSM602223S, RSM602224S, RSM602227S, RSM602230S, RSM602232S, RSM602235S, RSM602238S, RSM602239S, RSM602240S, RSM602241S, RSM602242S, RSM602243S, RSM602245S, RSM602246S, RSM602249S, RSM602250S, RSM602261S, RSM602265S, RSM602271S, RSM602272S, RSM602273S, RSM602274S, RSM602275S, RSM602276S, RSM602277S, RSM602281S, RSM602282S, RSM602284S, RSM602285S, RSM602286S, RSM602288S, RSM602289S, RSM602294S, RSM602296S, RSM602303S, RSM602304S, RSM602305S, RSM602307S, RSM602313S, RSM602314S, RSM602315S, RSM602316S, RSM602321S, RSM602323S, RSM602324S, RSM602327S, RSM602330S, RSM602332S, RSM602333S, RSM602334S, RSM602335S, RSM602337S, RSM602338S, RSM602344S, RSM602346S, RSM602350S, RSM602351S, RSM602353S, RSM602355S, RSM602360S, RSM602361S, RSM602362S, RSM602364S, RSM602367S, RSM602370S, RSM602371S, RSM602372S, RSM602375S, RSM602378S, RSM602383S, RSM602384S, RSM602392S, RSM602393S, RSM602394S, RSM602395S, RSM602397S, RSM602398S, RSM602399S, RSM602400S, RSM602401S, RSM602402S, RSM602403S, RSM602404S, RSM602405S, RSM602407S, RSM602408S, RSM602409S, RSM602410S, RSM602411S, RSM602413S, RSM602414S, RSM602415S, RSM602416S, RSM602417S, RSM602418S, RSM602419S, RSM602420S, RSM602422S, RSM602424S, RSM602426S, RSM602427S, RSM602429S, RSM602430S, RSM602431S, RSM602432S, RSM602435S, RSM602436S, RSM602437S, RSM602438S, RSM602439S, RSM602440S, RSM602442S, RSM602443S, RSM602444S, RSM602445S, RSM602446S, RSM602448S, RSM602449S, RSM602450S, RSM602451S, RSM602452S, RSM602454S, RSM602455S, RSM602456S, RSM602458S, RSM602460S, RSM602461S, RSM602462S, RSM602467S, RSM602468S, RSM602469S, RSM602470S, RSM602471S, RSM602476S, RSM602477S, RSM602480S, RSM602481S, RSM602487S, RSM602490S, RSM602493S, RSM602494S, RSM602496S, RSM602499S, RSM602502S, RSM602503S, RSM602504S, RSM602507S, RSM602508S, RSM602510S, RSM602513S, RSM602514S, RSM602522S, RSM602528S, RSM602529S, RSM602531S, RSM602534S, RSM602539S, RSM602540S, RSM602547S, RSM602550S, RSM602555S, RSM602558S, RSM602560S, RSM602564S, RSM602566S, RSM602567S, RSM602571S, RSM602573S, RSM602574S, RSM602575S, RSM602576S, RSM602578S, RSM602579S, RSM602582S, RSM602583S, RSM602584S, RSM602586S, RSM602587S, RSM602588S, RSM602592S, RSM602593S, RSM602596S, RSM602597S, RSM602598S, RSM602600S, RSM602603S, RSM602604S, RSM602606S, RSM602611S, RSM602612S, RSM602614S, RSM602615S, RSM602616S, RSM602618S, RSM602620S, RSM602621S, RSM602622S, RSM602625S, RSM602628S, RSM602629S, RSM602630S, RSM602631S, RSM602650S, RSM602666S, RSM602686S, RSM602690S, RSM602701S, RSM602705S, RSM602718S, RSM602727S, RSM602729S, RSM602731S, RSM602733S, RSM602750S, RSM602753S, RSM602756S, RSM602757S, RSM602762S, RSM602763S, RSM602764S, RSM602770S, RSM602772S, RSM602778S, RSM602779S, RSM602780S, RSM602781S, RSM602787S, RSM602789S, RSM602790S, RSM602796S, RSM602797S, RSM602798S, RSM602799S, RSM602800S, RSM602801S, RSM602802S, RSM602803S, RSM602804S, RSM602805S, RSM602806S, RSM602807S, RSM602808S, RSM602809S, RSM602812S, RSM602813S, RSM602814S, RSM602815S, RSM602818S, RSM602819S, RSM602820S, RSM602823S, RSM602825S, RSM602826S, RSM602827S, RSM602832S, RSM602834S, RSM602835S, RSM602836S, RSM602840S, RSM602843S, RSM602844S, RSM602845S, RSM602847S, RSM602849S, RSM602857S, RSM602860S, RSM602863S, RSM602869S, RSM602875S, RSM602876S, RSM602879S, RSM602881S, RSM602884S, RSM602885S, RSM602886S, RSM602888S, RSM602890S, RSM602891S, RSM602892S, RSM602893S, RSM602895S, RSM602897S, RSM602900S, RSM602902S, RSM602904S, RSM602906S, RSM602908S, RSM602909S, RSM602913S, RSM602914S, RSM602915S, RSM602918S, RSM602919S, RSM602922S, RSM602923S, RSM602925S, RSM602926S, RSM602928S, RSM602936S, RSM602937S, RSM602938S, RSM602939S, RSM602941S, RSM602942S, RSM602943S, RSM602944S, RSM602946S, RSM602947S, RSM602948S, RSM602949S, RSM602950S, RSM602951S, RSM602952S, RSM602953S, RSM602954S, RSM602955S, RSM602956S, RSM602957S, RSM602958S, RSM602959S, RSM602960S, RSM602961S, RSM602962S, RSM602963S, RSM602964S, RSM602965S, RSM602966S, RSM602967S, RSM602968S, RSM602969S, RSM602970S, RSM602971S, RSM602972S, RSM602973S, RSM602974S, RSM602975S, RSM602976S, RSM602977S, RSM602978S, RSM602979S, RSM602980S, RSM602981S, RSM602982S, RSM602983S, RSM602984S, RSM602985S, RSM602986S, RSM602987S, RSM602988S, RSM602989S, RSM602990S, RSM602991S, RSM602992S, RSM602993S, RSM602994S, RSM602995S, RSM602996S, RSM602997S, RSM602998S, RSM602999S, RSM603000S, RSM603001S, RSM603002S, RSM603003S, RSM603004S, RSM603005S, RSM603006S, RSM603007S, RSM603008S, RSM603009S, RSM603010S, RSM603011S, RSM603012S, RSM603013S, RSM603014S, RSM603015S, RSM603016S, RSM603017S, RSM603018S, RSM603019S, RSM603020S, RSM603021S, RSM603022S, RSM603023S, RSM603024S, RSM603025S, RSM603026S, RSM603027S, RSM603028S, RSM603029S, RSM603030S, RSM603031S, RSM603032S, RSM603033S, RSM603034S, RSM603044S, RSM603059S, RSM603082S, RSM603135S, RSM603136S, RSM603138S, RSM603140S, RSM603141S, RSM603142S, RSM603143S, RSM603148S, RSM603152S, RSM603158S, RSM603159S, RSM603160S, RSM603163S, RSM603165S, RSM603166S, RSM603173S, RSM603174S, RSM603175S, RSM603176S, RSM603178S, RSM603181S, RSM603184S, RSM603185S, RSM603193S, RSM603196S, RSM603197S, RSM603199S, RSM603201S, RSM603202S, RSM603204S, RSM603207S, RSM603208S, RSM603209S, RSM603210S, RSM603212S, RSM603214S, RSM603215S, RSM603216S, RSM603217S, RSM603218S, RSM603226S, RSM603227S, RSM603228S, RSM603230S, RSM603232S, RSM603234S, RSM603235S, RSM603237S, RSM603240S, RSM603242S, RSM603243S, RSM603244S, RSM603245S, RSM603246S, RSM603247S, RSM603249S, RSM603252S, RSM603253S, RSM603254S, RSM603255S, RSM603256S, RSM603257S, RSM603259S, RSM603260S, RSM603261S, RSM603262S, RSM603263S, RSM603264S, RSM603266S, RSM603268S, RSM603270S, RSM603271S, RSM603272S, RSM603273S, RSM603274S, RSM603277S, RSM603280S, RSM603282S, RSM603283S, RSM603287S, RSM603290S, RSM603292S, RSM603293S, RSM603294S, RSM603295S, RSM603296S, RSM603297S, RSM603298S, RSM603299S, RSM603300S, RSM603301S, RSM603303S, RSM603304S, RSM603305S, RSM603306S, RSM603308S, RSM603309S, RSM603310S, RSM603313S, RSM603314S, RSM603315S, RSM603316S, RSM603317S, RSM603318S, RSM603319S, RSM603321S, RSM603322S, RSM603324S, RSM603325S, RSM603326S, RSM603327S, RSM603328S, RSM603329S, RSM603330S, RSM603331S, RSM603332S, RSM603333S, RSM603335S, RSM603358S, RSM603359S, RSM603360S, RSM603361S, RSM603369S, RSM603372S, RSM603373S, RSM603375S, RSM603381S, RSM603386S, RSM603388S, RSM603391S, RSM603396S, RSM603412S, RSM603422S, RSM603427S, RSM603472S, RSM603476S, RSM603480S, RSM603492S, RSM603494S, RSM603601S, RSM603628S, RSM603629S, RSM603630S, RSM603631S, RSM603632S, RSM603633S, RSM603634S, RSM603635S, RSM603636S, RSM603637S, RSM603638S, RSM603639S, RSM603640S, RSM603641S, RSM603642S, RSM603643S, RSM603644S, RSM603645S, RSM603646S, RSM603647S, RSM603648S, RSM603649S, RSM603650S, RSM603651S, RSM603653S, RSM603654S, RSM603655S, RSM603656S, RSM603657S, RSM603658S, RSM603659S, RSM603660S, RSM603661S, RSM603662S, RSM603663S, RSM603664S, RSM603665S, RSM603666S, RSM603667S, RSM603668S, RSM603669S, RSM603670S, RSM603671S, RSM603672S, RSM603673S, RSM603674S, RSM603675S, RSM603676S, RSM603677S, RSM603678S, RSM603679S, RSM603680S, RSM603681S, RSM603682S, RSM603683S, RSM603684S, RSM603685S, RSM603686S, RSM603687S, RSM603688S, RSM603689S, RSM603690S, RSM603691S, RSM603692S, RSM603693S, RSM603694S, RSM603695S, RSM603696S, RSM603697S, RSM603698S, RSM603699S, RSM603700S, RSM603701S, RSM603702S, RSM603703S, RSM603704S, RSM603705S, RSM603707S, RSM603708S, RSM603709S, RSM603710S, RSM603711S, RSM603712S, RSM603713S, RSM603714S, RSM603715S, RSM603716S, RSM603717S, RSM603718S, RSM603719S, RSM603720S, RSM603721S, RSM603722S, RSM603723S, RSM603725S, RSM603726S, RSM603727S, RSM603728S, RSM603729S, RSM603730S, RSM603731S, RSM603732S, RSM603733S, RSM603734S, RSM603735S, RSM603736S, RSM603737S, RSM603738S, RSM603739S, RSM603740S, RSM603741S, RSM603742S, RSM603743S, RSM603744S, RSM603745S, RSM603746S, RSM603747S, RSM603748S, RSM603749S, RSM603750S, RSM603751S, RSM603753S, RSM603754S, RSM603755S, RSM603756S, RSM603757S, RSM603758S, RSM603759S, RSM603761S, RSM603762S, RSM603763S, RSM603764S, RSM603765S, RSM603766S, RSM603767S, RSM603768S, RSM603769S, RSM603770S, RSM603771S, RSM603772S, RSM603773S, RSM603774S, RSM603775S, RSM603776S, RSM603777S, RSM603778S, RSM603779S, RSM603780S, RSM603782S, RSM603783S, RSM603784S, RSM603785S, RSM603786S, RSM603787S, RSM603788S, RSM603789S, RSM603790S, RSM603791S, RSM603792S, RSM603793S, RSM603794S, RSM603795S, RSM603796S, RSM603797S, RSM603798S, RSM603799S, RSM603800S, RSM603801S, RSM603802S, RSM603803S, RSM603804S, RSM603805S, RSM603806S, RSM603807S, RSM603808S, RSM603809S, RSM603811S, RSM603812S, RSM603813S, RSM603814S, RSM603816S, RSM603817S, RSM603818S, RSM603819S, RSM603821S, RSM603822S, RSM603823S, RSM603826S, RSM603827S, RSM603828S, RSM603829S, RSM603830S, RSM603831S, RSM603832S, RSM603833S, RSM603834S, RSM603835S, RSM603836S, RSM603837S, RSM603838S, RSM603839S, RSM603840S, RSM603841S, RSM603842S, RSM603844S, RSM603845S, RSM603846S, RSM603847S, RSM603849S, RSM603850S, RSM603852S, RSM603853S, RSM603854S, RSM603855S, RSM603857S, RSM603858S, RSM603859S, RSM603860S, RSM603861S, RSM603862S, RSM603863S, RSM603864S, RSM603865S, RSM603866S, RSM603867S, RSM603868S, RSM603870S, RSM603871S, RSM603872S, RSM603873S, RSM603874S, RSM603875S, RSM603876S, RSM603877S, RSM603878S, RSM603879S, RSM603880S, RSM603881S, RSM603883S, RSM603884S, RSM603885S, RSM603886S, RSM603887S, RSM603888S, RSM603889S, RSM603890S, RSM603891S, RSM603892S, RSM603894S, RSM603911S, RSM603926S, RSM603946S, RSM603963S, RSM603966S, RSM603967S, RSM603968S, RSM603970S, RSM603972S, RSM603974S, RSM603976S, RSM603982S, RSM603984S, RSM603985S, RSM603988S, RSM603989S, RSM603990S, RSM603991S, RSM603994S, RSM604001S, RSM604002S, RSM604004S, RSM604008S, RSM604011S, RSM604012S, RSM604014S, RSM604015S, RSM604016S, RSM604017S, RSM604018S, RSM604023S, RSM604024S, RSM604025S, RSM604030S, RSM604031S, RSM604032S, RSM604033S, RSM604034S, RSM604035S, RSM604036S, RSM604038S, RSM604043S, RSM604046S, RSM604048S, RSM604050S, RSM604055S, RSM604057S, RSM604058S, RSM604060S, RSM604061S, RSM604065S, RSM604066S, RSM604070S, RSM604071S, RSM604073S, RSM604074S, RSM604077S, RSM604080S, RSM604081S, RSM604087S, RSM604090S, RSM604091S, RSM604093S, RSM604095S, RSM604097S, RSM604098S, RSM604099S, RSM604100S, RSM604103S, RSM604106S, RSM604107S, RSM604108S, RSM604110S, RSM604113S, RSM604114S, RSM604115S, RSM604118S, RSM604119S, RSM604121S, RSM604122S, RSM604125S, RSM604126S, RSM604127S, RSM604128S, RSM604131S, RSM604132S, RSM604134S, RSM604135S, RSM604136S, RSM604137S, RSM604138S, RSM604139S, RSM604141S, RSM604142S, RSM604170S, RSM604177S, RSM604191S, RSM604192S, RSM604193S, RSM604194S, RSM604195S, RSM604196S, RSM604198S, RSM604199S, RSM604200S, RSM604201S, RSM604234S, RSM604237S, RSM604241S, RSM604250S, RSM604251S, RSM604254S, RSM604255S, RSM604276S, RSM604281S, RSM604283S, RSM604284S, RSM604285S, RSM604286S, RSM604287S, RSM604289S, RSM604295S, RSM604299S, RSM604300S, RSM604320S, RSM604354S, RSM604380S, RSM604386S, RSM604440S, RSM604451S, RSM604452S, RSM604454S, RSM604460S, RSM604471S, RSM604573S, RSM604638S, RSM604642S, RSM604644S, RSM604653S, RSM604658S, RSM604663S, RSM604665S, RSM604669S, RSM604670S, RSM604673S, RSM604674S, RSM604675S, RSM604676S, RSM604677S, RSM604725S, RSM604752S, RSM605208S, RSM605214S, RSM605217S, RSM605220S, RSM605222S, RSM605223S, RSM605225S, RSM605229S, RSM605230S, RSM605231S, RSM605238S, RSM605242S, RSM605243S, RSM605244S, RSM605245S, RSM605277S, RSM605278S, RSM605279S, RSM605280S, RSM605281S, RSM605282S, RSM605283S, RSM605284S, RSM605285S, RSM605286S, RSM605287S, RSM605288S, RSM605289S, RSM605290S, RSM605292S, RSM605293S, RSM605295S, RSM605297S, RSM605298S, RSM605299S, RSM605300S, RSM605301S, RSM605302S, RSM605303S, RSM605304S, RSM605305S, RSM605306S, RSM605307S, RSM605308S, RSM605309S, RSM605310S, RSM605311S, RSM605312S, RSM605313S, RSM605314S, RSM605315S, RSM605316S, RSM605317S, RSM605318S, RSM605320S, RSM605321S, RSM605322S, RSM605323S, RSM605324S, RSM605325S, RSM605326S, RSM605327S, RSM605328S, RSM605330S, RSM605467S, RSM605468S, RSM605469S, RSM605476S, RSM605478S, RSM605482S, RSM605485S, RSM605491S, RSM605492S, RSM605504S, RSM605658S, RSM605664S, RSM605668S, RSM605686S, RSM605694S, RSM605696S, RSM605707S, RSM605708S, RSM606126S, RSM606130S, RSM606151S, RSM606153S, RSM606154S, RSM606158S, RSM606161S, RSM606166S, RSM606170S, RSM606173S, RSM606175S, RSM606182S, RSM606184S, RSM606188S, RSM606192S, RSM606195S, RSM606198S, RSM606199S, RSM606243S, RSM606245S, RSM606269S, RSM606270S, RSM606288S, RSM606322S, RSM606395S, RSM606411S, RSM606415S, RSM606425S, RSM606447S, RSM606457S, RSM606481S, RSM606486S, RSM606501S, RSM606509S, RSM606512S, RSM606513S, RSM606515S, RSM606516S, RSM606518S, RSM606519S, RSM606520S, RSM606521S, RSM606522S, RSM606523S, RSM606524S, RSM606526S, RSM606528S, RSM606529S, RSM606531S, RSM606532S, RSM606533S, RSM606534S, RSM606535S, RSM606536S, RSM606537S, RSM606538S, RSM606539S, RSM606540S, RSM606541S, RSM606542S, RSM606543S, RSM606544S, RSM606545S, RSM606546S, RSM606547S, RSM606550S, RSM606551S, RSM606552S, RSM606553S, RSM606554S, RSM606555S, RSM606556S, RSM606557S, RSM606558S, RSM606560S, RSM606571S, RSM606573S, RSM606574S, RSM606575S, RSM606576S, RSM606577S, RSM606579S, RSM606581S, RSM606582S, RSM606584S, RSM606585S, RSM606590S, RSM606591S, RSM606592S, RSM606593S, RSM606594S, RSM606595S, RSM606597S, RSM606599S, RSM606653S, RSM606701S, RSM606710S, RSM606713S, RSM606714S, RSM606716S, RSM606720S, RSM606731S, RSM606739S, RSM606748S, RSM606749S, RSM606750S, RSM606752S, RSM606761S, RSM606763S, RSM606764S, RSM606779S, RSM606793S, RSM606807S, RSM606810S, RSM606811S, RSM606812S, RSM606813S, RSM606818S, RSM606819S, RSM606820S, RSM606823S, RSM606826S, RSM606827S, RSM606837S, RSM606845S, RSM606855S, RSM606859S, RSM606865S, RSM606867S, RSM606869S, RSM606870S, RSM606871S, RSM606872S, RSM606874S, RSM606875S, RSM606876S, RSM606877S, RSM606880S, RSM606881S, RSM606883S, RSM606884S, RSM606885S, RSM606886S, RSM606887S, RSM606889S, RSM606891S, RSM606892S, RSM606895S, RSM606898S, RSM606900S, RSM606902S, RSM606905S, RSM606908S, RSM606910S, RSM606911S, RSM606912S, RSM606914S, RSM606915S, RSM606916S, RSM606918S, RSM606919S, RSM606920S, RSM606921S, RSM606922S, RSM606925S, RSM606927S, RSM606929S, RSM606931S, RSM606932S, RSM606933S, RSM606934S, RSM606935S, RSM606936S, RSM606938S, RSM606939S, RSM606943S, RSM606945S, RSM606946S, RSM606951S, RSM606953S, RSM606954S, RSM606956S, RSM606957S, RSM606960S, RSM606963S, RSM606964S, RSM606965S, RSM606966S, RSM606969S, RSM606973S, RSM606975S, RSM606979S, RSM606980S, RSM606983S, RSM606985S, RSM606987S, RSM606991S, RSM606993S, RSM606994S, RSM606996S, RSM606997S, RSM606999S, RSM607000S, RSM607002S, RSM607003S, RSM607005S, RSM607013S, RSM607015S, RSM607017S, RSM607018S, RSM607020S, RSM607021S, RSM607026S, RSM607032S, RSM607033S, RSM607039S, RSM607040S, RSM607041S, RSM607042S, RSM607043S, RSM607048S, RSM607050S, RSM607051S, RSM607053S, RSM607056S, RSM607059S, RSM607060S, RSM607061S, RSM607062S, RSM607065S, RSM607066S, RSM607067S, RSM607069S, RSM607075S, RSM607076S, RSM607077S, RSM607079S, RSM607084S, RSM607086S, RSM607088S, RSM607093S, RSM607100S, RSM607101S, RSM607121S, RSM607123S, RSM607124S, RSM607125S, RSM607128S, RSM607129S, RSM607131S, RSM607135S, RSM607136S, RSM607138S, RSM607144S, RSM607146S, RSM607147S, RSM607148S, RSM607150S, RSM607151S, RSM607153S, RSM607158S, RSM607159S, RSM607164S, RSM607169S, RSM607170S, RSM607171S, RSM607172S, RSM607173S, RSM607174S, RSM607175S, RSM607176S, RSM607177S, RSM607178S, RSM607179S, RSM607182S, RSM607189S, RSM607316S, RSM607451S, RSM607481S, RSM607608S, RSM607610S, RSM607611S, RSM607612S, RSM607614S, RSM607615S, RSM607616S, RSM607618S, RSM607619S, RSM607621S, RSM607622S, RSM607628S, RSM607631S, RSM607632S, RSM607634S, RSM607636S, RSM607637S, RSM607638S, RSM607639S, RSM607640S, RSM607641S, RSM607643S, RSM607645S, RSM607647S, RSM607648S, RSM607649S, RSM607651S, RSM607654S, RSM607655S, RSM607656S, RSM607657S, RSM607658S, RSM607660S, RSM607661S, RSM607663S, RSM607664S, RSM607665S, RSM607666S, RSM607667S, RSM607668S, RSM607672S, RSM607675S, RSM607678S, RSM607679S, RSM607680S, RSM607682S, RSM607727S, RSM607730S, RSM607731S, RSM607735S, RSM607736S, RSM607738S, RSM607740S, RSM607742S, RSM607744S, RSM607746S, RSM607748S, RSM607750S, RSM607751S, RSM607752S, RSM607753S, RSM607755S, RSM607756S, RSM607758S, RSM607761S, RSM607762S, RSM607763S, RSM607764S, RSM607770S, RSM607771S, RSM607772S, RSM607773S, RSM607774S, RSM607775S, RSM607776S, RSM607777S, RSM607778S, RSM607780S, RSM607781S, RSM607782S, RSM607783S, RSM607784S, RSM607786S, RSM607787S, RSM607788S, RSM607789S, RSM607792S, RSM607794S, RSM607795S, RSM607798S, RSM607800S, RSM607801S, RSM607802S, RSM607803S, RSM607804S, RSM607805S, RSM607816S, RSM607818S, RSM607832S, RSM607835S, RSM607837S, RSM607839S, RSM607869S, RSM607871S, RSM607879S, RSM607882S, RSM607883S, RSM607890S, RSM607899S, RSM608005S, RSM608024S, RSM608071S, RSM608258S, RSM608316S, RSM608336S, RSM608340S, RSM608347S, RSM608355S, RSM608363S, RSM608365S, RSM608377S, RSM608387S, RSM608423S, RSM608425S, RSM608426S, RSM608428S, RSM608429S, RSM608430S, RSM608433S, RSM608434S, RSM608436S, RSM608437S, RSM608438S, RSM608439S, RSM608440S, RSM608441S, RSM608445S, RSM608446S, RSM608447S, RSM608448S, RSM608449S, RSM608451S, RSM608454S, RSM608455S, RSM608457S, RSM608458S, RSM608459S, RSM608460S, RSM608461S, RSM608462S, RSM608463S, RSM608464S, RSM608465S, RSM608467S, RSM608468S, RSM608471S, RSM608472S, RSM608473S, RSM608474S, RSM608476S, RSM608477S, RSM608479S, RSM608480S, RSM608481S, RSM608482S, RSM608483S, RSM608484S, RSM608489S, RSM608490S, RSM608567S, RSM608569S, RSM608572S, RSM608573S, RSM608574S, RSM608575S, RSM608578S, RSM608580S, RSM608581S, RSM608583S, RSM608585S, RSM608588S, RSM608589S, RSM608590S, RSM608592S, RSM608597S, RSM608599S, RSM608602S, RSM608603S, RSM608605S, RSM608606S, RSM608610S, RSM608614S, RSM608615S, RSM608616S, RSM608618S, RSM608624S, RSM608630S, RSM608631S, RSM608633S, RSM608634S, RSM608635S, RSM608636S, RSM608638S, RSM608640S, RSM608641S, RSM608643S, RSM608644S, RSM608645S, RSM608646S, RSM608647S, RSM608650S, RSM608652S, RSM608654S, RSM608655S, RSM608657S, RSM608658S, RSM608659S, RSM608660S, RSM608661S, RSM608662S, RSM608663S, RSM608664S, RSM608666S, RSM608668S, RSM608670S, RSM608671S, RSM608672S, RSM608673S, RSM608676S, RSM608677S, RSM608678S, RSM608681S, RSM608683S, RSM608722S, RSM608790S, RSM608938S, RSM608940S, RSM609017S, RSM609018S, RSM609019S, RSM609036S, RSM609037S, RSM609040S, RSM609041S, RSM609044S, RSM609045S, RSM609046S, RSM609047S, RSM609048S, RSM609049S, RSM609054S, RSM609056S, RSM609059S, RSM609063S, RSM609066S, RSM609067S, RSM609068S, RSM609071S, RSM609072S, RSM609073S, RSM609074S, RSM609076S, RSM609077S, RSM609078S, RSM609079S, RSM609080S, RSM609081S, RSM609082S, RSM609083S, RSM609086S, RSM609089S, RSM609091S, RSM609092S, RSM609096S, RSM609101S, RSM609108S, RSM609112S, RSM609113S, RSM609114S, RSM609116S, RSM609120S, RSM609121S, RSM609124S, RSM609125S, RSM609128S, RSM609130S, RSM609131S, RSM609132S, RSM609133S, RSM609134S, RSM609136S, RSM609137S, RSM609138S, RSM609139S, RSM609140S, RSM609141S, RSM609142S, RSM609143S, RSM609144S, RSM609145S, RSM609146S, RSM609147S, RSM609150S, RSM609151S, RSM609153S, RSM609155S, RSM609157S, RSM609158S, RSM609160S, RSM609162S, RSM609163S, RSM609164S, RSM609165S, RSM609166S, RSM609167S, RSM609168S, RSM609169S, RSM609171S, RSM609172S, RSM609173S, RSM609174S, RSM609175S, RSM609176S, RSM609177S, RSM609178S, RSM609179S, RSM609180S, RSM609181S, RSM609183S, RSM609184S, RSM609185S, RSM609187S, RSM609188S, RSM609189S, RSM609190S, RSM609191S, RSM609192S, RSM609193S, RSM609194S, RSM609195S, RSM609200S, RSM609201S, RSM609204S, RSM609206S, RSM609208S, RSM609209S, RSM609210S, RSM609211S, RSM609212S, RSM609213S, RSM609214S, RSM609215S, RSM609216S, RSM609217S, RSM609219S, RSM609220S, RSM609221S, RSM609223S, RSM609224S, RSM609225S, RSM609227S, RSM609228S, RSM609229S, RSM609232S, RSM609234S, RSM609236S, RSM609237S, RSM609238S, RSM609239S, RSM609240S, RSM609241S, RSM609242S, RSM609243S, RSM609244S, RSM609245S, RSM609247S, RSM609248S, RSM609249S, RSM609255S, RSM609257S, RSM609259S, RSM609260S, RSM609262S; GTIN none, Serial Numbers: RSM605206S, RSM606878S, RSM607754S, RSM609094S, RSM609098S |
Recalling Firm/ Manufacturer |
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) 8200 Coral Sea St Ne Mounds View MN 55112-4391
|
| For Additional Information Contact | Technical Services 763-723-4636 |
Manufacturer Reason for Recall | There is the potential for reduced shock energy (~79% of programmed energy) during high-voltage (HV) therapy for Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds). |
FDA Determined Cause 2 | Device Design |
| Action | Beginning 22-JUN-2022 communications to all affected consignees began. In the
United States notifications are sent via UPS 2-day delivery. For outside U.S. countries,
notifications are handled via a regionally approved method. Consignees are asked to forward this notification to those who need to be aware within the organization and to any location which the devices have been distributed. Consignees are asked to complete and return a confirmation certificate (or equivalent record) confirming they have received the notification. Consignees are asked to quarantine the subset of unused identified Cobalt and Crome devices for retrieval due to potential manufacturing non-conformance and return the inventory to Medtronic.
On approximately 08/10/2022, Medtronic notified Risk Managers of the availability of a SW update and actions specific to update the SW on the SmartSync devices, that will in turn update the patient's implanted device during their next clinic visit. In addition, on this same date, a separate communication dated August 2022 is being sent to all affected Implanting Physicians notifying them that a SW update is available and recommended actions to correct this issue with their patient's implanted device. The second notification to physicians did not request returned confirmation. |
| Quantity in Commerce | 6884 units |
| Distribution | Worldwide Distribution |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| PMA Database | PMAs with Product Code = LWS PMAs with Product Code = LWS
|
|
|
|