• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall BD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 2 Device Recall BD see related information
Date Initiated by Firm May 25, 2022
Date Posted August 26, 2022
Recall Status1 Open3, Classified
Recall Number Z-1617-2022
Recall Event ID 90336
510(K)Number K024240  
Product Classification Culture media, non-propagating transport - Product Code JSM
Product a. BECTON DICKINSON BD Vacutainer C&S Preservative Urine Tube TUBE, URINE COLL PLAS 4ML (100/BX 10BX/CS), Model Number: BD 364951.
b. BECTON DICKINSON BD Vacutainer C&S Transfer Straw Kit TRANSFER STRAW KIT, URINE 4ML(50/BX 4BX/CS) Model Number: BD 364953.
c. BECTON DICKINSON BD Vacutainer COLLECTION KIT, URINE COMPLETE 8ML 13"X75" (50/CS) BD Model Number: 364957

Code Information GTIN: a. 00382903649518 EA; 30382903649519 PK; 50382903649513. b. 00382903649532; 30382903649533; 50382903649537. c. 00382903649570 EA; 50382903649575 CS
Recalling Firm/
Manufacturer
Mckesson Medical-Surgical Inc. Corporate Office
9954 Maryland Drive
Deep Run Iii Ste. 4000
Richmond VA 23233
For Additional Information Contact Ms. Ruby Upadhyaya
804-553-2281
Manufacturer Reason
for Recall
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
FDA Determined
Cause 2
Environmental control
Action Mckesson Letter Notifcations issued via USP by or by email if available by the following third-party recall company: Sedgwick Brand Protection (formerly Stericycle) Sedgwick, Memphis, TN 38120 sedgwick@sedgwick.com. from May 10, 2022 continuing through July 21, 2022. Letter states: If you have further distributed this product, please notify your patients at once of this product recall. Your notification to your patients may be enhanced by including a copy of this recall notification letter. ¿ This recall should be carried out to the user level. Your assistance is appreciated and necessary to prevent possible patient harm. ¿ Please complete and return the enclosed response form as soon as possible even if you do not have any product on hand. ¿ Customers should contact Sedgwick directly by phone at 1-888-943-2401 to return the recalled products or if they have additional questions regarding this notification. Representatives are available Monday Friday, 8:00 am 5:00 pm PT
Quantity in Commerce a. 25 bx b. 53 bx c. 31 cs
Distribution US Nationwide
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.