 Class 2 Device Recall Various Laparoscopy Packssee related information
Date Initiated by FirmJune 16, 2022
Create DateSeptember 08, 2022
Recall Status1 Open3, Classified
Recall NumberZ-1694-2022
Recall Event ID 90574
Product Classification laparoscopy kit - Product Code FDE
ProductVarious Laparoscopy Packs
Code Information Tray Number, Sterilization Lot, Lot Number, Bag Serial Number 1) LSRENDOSCOPH, 2104281, 890211, 14887854 2) LSRENDOSCOPH, 2108251, 784211, 44375070 3) LSRENDOSCOPH, 2110011, 740211, 15140201 4) PLLA88I, 2104162, 895211, 14779054 5) AMGL09X, 2104281, 887211, 64029666 6) IHGB89AB, 2107293, 791211, 64104353 7) UIUA68AG, 2108202, 771211, 64233083 8) FRLC65S, 2104301, 882211, 44123696-005 9) MHLC19AH, 2105181, 863211, 63956744-006 10) CXRO12W, 2107082, 811211, 64110751-006 11) SLLP38F, 2107132, 807211, 64100877-006 12) SLLP38F, 2107132, 807211, 64100879-006 13) SLLP38F, 2107132, 807211, 64100878-006 14) CXLP87AJ, 2107162, 803211, 15011547-004 15) BHLC43J, 2108042, 784211, 44294507-005 16) BHLC43J, 2108042, 784211, 44234508-005 17) MERS50Q, 2108193, 769211, 64206337-006 18) BHRP78C, 2104131, 903211, 64073208-006 19) BHRP78C, 2104131, 903211, 64073209-006 20) BHRP78C, 2106221, 831211, 64091105-006 21) BHRP78C, 2106221, 831211, 64091104-006 22) BHRP78C, 2106233, 831211, 64090598-006 23) BHLP21C, 2107261, 800211, 44226340-005 24) BHLP21C, 2107261, 800211, 44226339-005 25) BHLP21C, 2107261, 800211, 44226341-005 26) BHLP21C, 2107261, 800211, 44226338-005 27) BHLP62E, 2107292, 799211, 44219340-005 28) BHLP62E, 2107292, 799211, 44219343-005 29) BHLP62E, 2107292, 799211, 44219342-005 30) BHLP62E, 2107292, 799211, 44219341-005 31) BHRP78C, 2108041, 789211, 64093909-006 32) BHRP78C, 2108041, 789211, 64092459-006 33) BHRP78C, 2108041, 789211, 64092460-006 34) BHLP62E, 2108242, 771211, 44336911-005 35) SARB30J, 2104291, 884211, 64040187-006 36) MCLC14F, 2106011, 853211, 44153023-005 37) MPRO72D, 2107131, 811211, 64129494-006 38) MCLC14F, 2109031, 760211, 44261569-005 39) AGLC19L, 2108191, 775211, 64186775-006
FEI Number 3003419114
Recalling Firm/
Manufacturer
American Contract Systems, Inc.
2610 Ne Industrial Dr Ste 220
Kansas City MO 64117-2648
For Additional Information Contact
816-920-5846
Manufacturer Reason
for Recall
Product was exposed to multiple sterilization cycles without validation for multiple exposures.
FDA Determined
Cause 2
Process change control
ActionThe recalling firm began notifying customers of the recall beginning June 13, 2022. Customer notification was initially made via email and an "Urgent Medical Device Recall" letter dated June 16, 2022. Customers are asked to review inventory and segregate affected product, which should be identified with the Bag Serial Number. Affected product should be destroyed, with evidence of destruction provided back to the recalling firm. A Field Action Response Form should be completed and returned to angela.bunn@owens-minor.com as soon as possible, even if the customer has no inventory. If you have any questions, please call 816-920-5846.
Quantity in Commerce39 units
DistributionDistribution within the United States to health systems and distributors in AZ, FL, IA, IL, MA, MN, MO, OH, NE, PA, RI, SD, and TX
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-