|
Class 2 Device Recall Alcor |
|
Date Initiated by Firm |
July 20, 2022 |
Create Date |
September 16, 2022 |
Recall Status1 |
Open3, Classified |
Recall Number |
Z-1751-2022 |
Recall Event ID |
90762 |
Product Classification |
Device, automated sedimentation rate - Product Code GKB
|
Product |
iSED Primary Pump Tubing Spare Part Kit- (IVD) tubing is used to move sample and wash solution from the aspiration needle, through the reading chamber, and to the waste on the iSED Automated ESR Analyzer PN: 112-13-002 |
Code Information |
Lot 4/22 |
Recalling Firm/ Manufacturer |
Alcor Scientific, Inc. 20 Thurber Blvd Smithfield RI 02917-1858
|
For Additional Information Contact |
SAME 401-737-3774
|
Manufacturer Reason for Recall |
Primary Pump Tubing Spare Part is incorrect. If the tubing is installed onto an ALCOR iSED ESR Analyzer, there is a potential of a biohazard leak from the iSEDs Probe Tube
|
FDA Determined Cause 2 |
Nonconforming Material/Component |
Action |
Alcor Scientific notified consignees via Emails on 21 July 2022 that included an Advisory Notice (AN-001). Letter states reason for recall, health risk and action to take:
" Remove from your inventory any Lot 4/22 Primary Pump Tubing Spare Part Kits (112-13-002). Destroy this tubing.
" If the tubing has been installed on any iSED ESR Analyzers, the tubing must be immediately replaced with new Primary Pump Tubing (112-13-002) from any other lot and Alcor must be notified of the locations where that tubing was installed.
" Complete this form and return via email to techservice@alcorscientific.com within 3 days.
ALCOR will replace all Primary Pump Tubing from Lot 4/22 that was shipped to your facility. |
Quantity in Commerce |
38 units |
Distribution |
US Nationwide Distribution - FL, MA, MD, PA, SD, WI
Foreign:
Croatia
Honduras
Korea
|
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|