| | Class 2 Device Recall Visionsense |  |
| Date Initiated by Firm | October 17, 2022 |
| Date Posted | November 23, 2022 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0280-2023 |
| Recall Event ID |
91055 |
| 510(K)Number | K150018 |
| Product Classification |
Confocal optical imaging - Product Code OWN
|
| Product | VS3 Iridium Visionsense Infrared (IR) Fluorescence Miniature Microscope (MMS-IR)
Product Number: 174-0012 |
| Code Information |
UDI-DI: 18130400103489
Lot Numbes:
3784719
3784735
3784751
378-4001
378-4002
378-4003
378-4004
378-4006
378-4007
378-4008
378-4009
378-4010
378-4012
378-4013
378-4014
378-4015
378-4016
378-4017
378-4020
378-4021
378-4501
378-4502
378-4503
378-4504
378-4505
378-4506
378-4507
378-4508
378-4510
378-4511
378-4512
378-4513
378-4515
378-4529
378-4531
378-4535
378-4537
3784581
378-4582
378-4583
378-4584
378-4588
378-4589
378-4590
3784592
378-4593
378-4594
3784595
378-4596
3784599
3784601
378-4602
3784603
378-4604
3784605
3784606
3784609
3784610
3784612
3784613
3784614
3784617
378-4618
3784619
3784620
3784621
3784622
3784623
3784624
3784626
3784627
3784628
3784629
3784632
3784633
3784635
3784636
3784637
3784638
3784641
3784644
3784645
3784646
3784648
3784649
3784650
3784651
3784652
3784653
3784654
3784656
378-4657
378-4658
3784659
3784660
3784661
3784662
3784663
3784664
3784665
3784666
3784667
3784669
3784671
3784672
3784674
378-4675
3784676
378-4677
3784679
3784680
3784681
3784682
3784683
3784685
3784686
3784687
3784688
3784689
3784691
3784705
3784706
3784708
3784713
3784715
3784716
3784717
3784718
3784720
3784721
3784723
3784725
3784726
3784727
3784728
378-4729
3784731
3784732
3784733
3784734
3784736
3784738
378-4739
378-4740
378-4742
3784743
3784744
378-4745
3784746
3784752
3784753
3784755
3784757
3784758
3784759
3784760
3784761
3784762
3784763
3784764
3784765
3784766
378-4767
378-4768
378-4769
3784775
3784776
3784777
3784779
3784780
3784781
3784789
3784790
3784791
3784792
3784793
3784798
3784799
3784801
378-4805
378-4806
3784809
3784811
3784812
3784813
3784819
3784821
3784822
3784824
3784825
3784826
3784827
3784828
378-4829
3784834
3784837
3784838
3784839
3784841
378-4858
3784863
3784864
3784865
3784867
3784868
|
Recalling Firm/ Manufacturer |
Visionsense, Ltd. 20, Ha-Magshimim Petakh Tikva Israel
|
| For Additional Information Contact | Medtronic Customer Service 800-962-9888 |
Manufacturer Reason for Recall | Broken Optical fiber cable damage can occur through handling or bending and can lead to loss of functionality and heating/melting of the outer cable surface. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | Medtronic issued Urgent Medical Device Correction Letter on 17-OCTOBER-2022 to US consignees via UPS 2-day delivery. For outside U.S. countries, notifications are handled via a regionally approved method. Letter states reason for recall, health risk and action to take:
1.Prior to use, visually check the Beam-Combiner and MMS optical fiber cable for signs of being pinched, kinkedor damaged. Avoid folding the cables and handle with care to prevent damage.
2.After the white balance has been performed as per IFU guidance, test the EleVision IR system at the beginningof the surgical procedure as follows:
a)Turn on the laser.
b)For the laparoscopic configuration of the EleVision IR system, point the distal end of the endoscopetowards a bright surface, keeping the scope approximately 3 cm from the target. This prevents thewhite light from scattering. For the open configuration of the EleVision IR system, position the distalface of the MMS approximately 30 cm from any object. If you see a message on the screen indicatingthat the device is too close to tissue, move the device further away from the object until this messagedisappears.
c)After two (2) seconds, the monitor screen should show a message indicating that the laser is on, ablinking red dot, and a running clock. If all these indications are seen properly on the monitor screen,turn off the laser after five (5) seconds.
If the screen does not show an indication that the laser is on, a blinking red dot, and a running clock, the laser is not functioning properly and the EleVision IR system should not be used in the surgical procedure.
If any pinching, kinking, or damage of the optical fiber cable is noted, or if the system does not work properly during the above-described laser test, do not use the system. Please report a complaint and arrange for servicing of the Beam Combiner and / or MMS device by contacting Medtronic at 1-800-255-6774 or rs.serviceexperience@medtronic.com.
Additional actions:
"Immediately notify |
| Quantity in Commerce | 205 units: 94 units OUS; 111 units US |
| Distribution | US Nationwide distribution.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = OWN
|
|
|
|