| | Class 2 Device Recall |  |
| Date Initiated by Firm | October 17, 2022 |
| Date Posted | November 23, 2022 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0282-2023 |
| Recall Event ID |
91055 |
| 510(K)Number | K150018 |
| Product Classification |
Confocal optical imaging - Product Code OWN
|
| Product | VS3 Iridium Visionsense Infrared (IR) Fluorescence Light Integrator (Beam Combiner)
Product Number: 161-0001 |
| Code Information |
UDI-DI: 18130400103625
Lot Numbers:
Serial Number/Lot Number
3611111
3611161
3611191
361-0012
361-0901
361-0902
361-0903
361-0904
361-0906
361-0907
361-0908
361-0909
361-0910
361-0911
361-0913
361-0914
361-0915
361-0916
361-0917
361-0918
361-0919
361-0921
361-0922
361-0923
361-0924
361-0926
361-0929
361-0931
361-0932
361-0933
361-0937
361-0938
361-0940
361-0941
3610942
361-0943
361-0944
361-0945
3610946
3610947
3610949
3610950
3610951
3610952
3610953
361-0955
3610956
3610957
3610958
3610961
3610963
3610964
3610965
361-0966
3610968
361-0970
3610971
3610972
3610973
3610974
3610975
3610977
3610978
3610979
3610996
3610998
3610999
3611000
3611004
3611006
3611007
3611008
3611009
3611011
3611012
3611014
3611015
3611016
3611019
3611025
3611027
3611029
3611031
3611033
3611034
3611035
3611036
3611039
3611041
3611042
3611044
3611046
3611048
3611049
3611052
3611053
3611055
3611056
3611057
3611058
3611059
3611060
361-1061
3611062
3611063
3611065
361-1067
3611068
3611070
3611071
3611073
3611075
3611077
3611078
3611079
3611080
3611082
3611083
3611084
3611085
3611086
3611087
3611088
3611092
3611096
3611097
3611099
3611100
3611101
3611102
361-1103
3611104
3611106
3611108
3611109
3611110
3611112
3611113
3611114
3611116
3611117
3611118
3611119
3611120
3611121
361-1122
3611123
3611124
3611125
3611126
3611128
3611129
3611130
3611131
361-1132
3611133
3611134
3611135
3611136
3611138
3611139
3611140
361-1144
3611146
361-1148
3611149
3611152
3611153
3611154
3611156
3611157
3611158
3611159
3611162
3611167
361-1168
3611170
3611171
3611172
3611173
3611174
3611175
3611176
3611177
3611178
3611180
3611181
3611182
3611188
3611189
3611190
3611194
3611195
3611196
3611197
3611199
3611200
3611201
3611218
3611219
3611223
3611224
3611225
361-1013
|
Recalling Firm/ Manufacturer |
Visionsense, Ltd. 20, Ha-Magshimim Petakh Tikva Israel
|
| For Additional Information Contact | Medtronic Customer Service 800-962-9888 |
Manufacturer Reason for Recall | Broken Optical fiber cable damage can occur through handling or bending and can lead to loss of functionality and heating/melting of the outer cable surface. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | Medtronic issued Urgent Medical Device Correction Letter on 17-OCTOBER-2022 to US consignees via UPS 2-day delivery. For outside U.S. countries, notifications are handled via a regionally approved method. Letter states reason for recall, health risk and action to take:
1.Prior to use, visually check the Beam-Combiner and MMS optical fiber cable for signs of being pinched, kinkedor damaged. Avoid folding the cables and handle with care to prevent damage.
2.After the white balance has been performed as per IFU guidance, test the EleVision IR system at the beginningof the surgical procedure as follows:
a)Turn on the laser.
b)For the laparoscopic configuration of the EleVision IR system, point the distal end of the endoscopetowards a bright surface, keeping the scope approximately 3 cm from the target. This prevents thewhite light from scattering. For the open configuration of the EleVision IR system, position the distalface of the MMS approximately 30 cm from any object. If you see a message on the screen indicatingthat the device is too close to tissue, move the device further away from the object until this messagedisappears.
c)After two (2) seconds, the monitor screen should show a message indicating that the laser is on, ablinking red dot, and a running clock. If all these indications are seen properly on the monitor screen,turn off the laser after five (5) seconds.
If the screen does not show an indication that the laser is on, a blinking red dot, and a running clock, the laser is not functioning properly and the EleVision IR system should not be used in the surgical procedure.
If any pinching, kinking, or damage of the optical fiber cable is noted, or if the system does not work properly during the above-described laser test, do not use the system. Please report a complaint and arrange for servicing of the Beam Combiner and / or MMS device by contacting Medtronic at 1-800-255-6774 or rs.serviceexperience@medtronic.com.
Additional actions:
"Immediately notify |
| Quantity in Commerce | 204 units: 93 units OUS ; 111 units US |
| Distribution | US Nationwide distribution.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = OWN
|
|
|
|