• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Rusch

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Ruschsee related information
Date Initiated by FirmDecember 08, 2022
Date PostedJanuary 23, 2023
Recall Status1 Open3, Classified
Recall NumberZ-1012-2023
Recall Event ID 91301
Product Classification Tube, tracheal (w/wo connector) - Product Code BTR
ProductEndotracheal Tube labeled as the following: a. Endotracheal Tube oral/nasal Magill, High Volume, Low Pressure Cuff; b. Endotracheal Tube oral/nasal Murphy Eye, High Volume, Low Pressure Cuff; c. Flexiset Super Safety Clear Tracheal Tube, oral/nasal with Cuff and Insertion Aid; d. RUSCHELIT Safety Clear Tracheal Tube, oral/nasal, Magill; e. RUSCHELIT Safety Clear Tracheal Tube, oral/nasal, Murphy. Intended for patients requiring endotracheal intubation
Code Information Not distributed within the USA. a. Product Code, Batch Number (UDI): 112080-000065, KME21D2250 ((01)14026704195937(17)260328(10)KME21D2250), 112080-000080, KME21E1336 ((01)14026704195968(17)260428(10)KME21E1336); b. Product Code, Batch Number (UDI): 112082-000050, KME21F0726 ((01)14026704196019(17)260528(10)KME21F0726), 112082-000055, KME21E1231 ((01)14026704196026(17)260428(10)KME21E1231), 112082-000055, KME21F0945 ((01)14026704196026(17)260528(10)KME21F0945), 112082-000060, KME21E1395 ((01)14026704196033(17)260428(10)KME21E1395), 112082-000065, KME21F0383 ((01)14026704196040(17)260528(10)KME21F0383), 112082-000065, KME21F0833 ((01)14026704196040(17)260528(10)KME21F0833), 112082-000065, KME21F0931 ((01)14026704196040(17)260528(10)KME21F0931), 112082-000070, KME21E0582 ((01)14026704196057(17)260428(10)KME21E0582), 112082-000070, KME21E1267 ((01)14026704196057(17)260428(10)KME21E1267), 112082-000070, KME21E1350 ((01)14026704196057(17)260428(10)KME21E1350), 112082-000070, KME21F0435 ((01)14026704196057(17)260528(10)KME21F0435), 112082-000070, KME21E0683 ((01)14026704196057(17)260428(10)KME21E0683), 112082-000070, KME21E1183 ((01)14026704196057(17)260428(10)KME21E1183), 112082-000070, KME21E1266 ((01)14026704196057(17)260428(10)KME21E1266), 112082-000070, KME21E1289 ((01)14026704196057(17)260428(10)KME21E1289), 112082-000070, KME21F0008 ((01)14026704196057(17)260528(10)KME21F0008), 112082-000070, KME21F0836 ((01)14026704196057(17)260528(10)KME21F0836), 112082-000075, KME21D1576 ((01)14026704196064(17)260328(10)KME21D1576), 112082-000075, KME21E1185 ((01)14026704196064(17)260428(10)KME21E1185), 112082-000075, KME21E0686 ((01)14026704196064(17)260428(10)KME21E0686), 112082-000075, KME21F0438 ((01)14026704196064(17)260528(10)KME21F0438), 112082-000075, KME21D2439 ((01)14026704196064(17)260328(10)KME21D2439), 112082-000075, KME21E0521 ((01)14026704196064(17)260428(10)KME21E0521), 112082-000075, KME21E0606 ((01)14026704196064(17)260428(10)KME21E0606), 112082-000075, KME21F0319 ((01)14026704196064(17)260528(10)KME21F0319), 112082-000080, KME21D1578 ((01)14026704196071(17)260328(10)KME21D1578), 112082-000080, KME21F0651 ((01)14026704196071(17)260528(10)KME21F0651), 112082-000080, KME21E1453 ((01)14026704196071(17)260428(10)KME21E1453), 112082-000085, KME21E0900 ((01)14026704196088(17)260428(10)KME21E0900), 112082-000085, KME21F0646 ((01)14026704196088(17)260528(10)KME21F0646); c. Product Code, Batch Number (UDI): 112481-000055, KME21D1840 ((01)14026704196330(17)260328(10)KME21D1840), 112481-000055, KME21F0812 ((01)14026704196330(17)260528(10)KME21F0812), 112481-000060, KME21E1386 ((01)14026704196347(17)260428(10)KME21E1386), 112481-000070, KME21E0866 ((01)14026704196361(17)260428(10)KME21E0866), 112481-000070, KME21E1094 ((01)14026704196361(17)260428(10)KME21E1094), 112481-000070, KME21F0793 ((01)14026704196361(17)260528(10)KME21F0793), 112481-000070, KME21D1468 ((01)14026704196361(17)260328(10)KME21D1468), 112481-000070, KME21D1562 ((01)14026704196361(17)260328(10)KME21D1562), 112481-000075, KME21D1146 ((01)14026704196378(17)260328(10)KME21D1146), 112481-000075, KME21D2825 ((01)14026704196378(17)260328(10)KME21D2825), 112481-000075, KME21E1092 ((01)14026704196378(17)260428(10)KME21E1092), 112481-000075, KME21F0794 ((01)14026704196378(17)260528(10)KME21F0794), 112481-000075, KME21D1686 ((01)14026704196378(17)260328(10)KME21D1686), 112481-000080, KME21F1073 ((01)14026704196385(17)260528(10)KME21F1073), 112481-000080, KME21D1153 ((01)14026704196385(17)260328(10)KME21D1153), 112481-000080, KME21D1296 ((01)14026704196385(17)260328(10)KME21D1296), 112481-000080, KME21D1560 ((01)14026704196385(17)260328(10)KME21D1560), 112481-000080, KME21F0895 ((01)14026704196385(17)260528(10)KME21F0895), 112481-000080, KME21F0995 ((01)14026704196385(17)260528(10)KME21F0995), 112481-000085, KME21E1119 ((01)14026704196392(17)260428(10)KME21E1119), 112481-000085, KME21D0707 ((01)14026704196392(17)260328(10)KME21D0707), 112481-000085, KME21D1685 ((01)14026704196392(17)260328(10)KME21D1685), 112481-000090, KME21E1115 ((01)14026704196408(17)260428(10)KME21E1115), 112481-000090, KME21E1116 ((01)14026704196408(17)260428(10)KME21E1116); d. Product Code, Batch Number (UDI): 112480-000050, KME21D1824 ((01)14026704196200(17)260328(10)KME21D1824), 112480-000050, KME21F0153 ((01)14026704196200(17)260528(10)KME21F0153), 112480-000050, KME21F0241 ((01)14026704196200(17)260528(10)KME21F0241), 112480-000050, KME21F0354 ((01)14026704196200(17)260528(10)KME21F0354), 112480-000050, KME21F0542 ((01)14026704196200(17)260528(10)KME21F0542), 112480-000050, KME21F0425 ((01)14026704196200(17)260528(10)KME21F0425), 112480-000050, KME21F0714 ((01)14026704196200(17)260528(10)KME21F0714), 112480-000055, KME21F0437 ((01)14026704196217(17)260528(10)KME21F0437), 112480-000055, KME21D2253 ((01)14026704196217(17)260328(10)KME21D2253), 112480-000055, KME21D2511 ((01)14026704196217(17)260328(10)KME21D2511), 112480-000055, KME21D2669 ((01)14026704196217(17)260328(10)KME21D2669), 112480-000055, KME21D2826 ((01)14026704196217(17)260328(10)KME21D2826), 112480-000055, KME21E0798 ((01)14026704196217(17)260428(10)KME21E0798), 112480-000055, KME21D2901 ((01)14026704196217(17)260328(10)KME21D2901), 112480-000055, KME21E0585 ((01)14026704196217(17)260428(10)KME21E0585), 112480-000055, KME21F0436 ((01)14026704196217(17)260528(10)KME21F0436), 112480-000060, KME21D1703 ((01)14026704196224(17)260328(10)KME21D1703), 112480-000060, KME21E1148 ((01)14026704196224(17)260428(10)KME21E1148), 112480-000060, KME21D1796 ((01)14026704196224(17)260328(10)KME21D1796), 112480-000060, KME21D1823 ((01)14026704196224(17)260328(10)KME21D1823), 112480-000060, KME21D1958 ((01)14026704196224(17)260328(10)KME21D1958), 112480-000060, KME21D2255 ((01)14026704196224(17)260328(10)KME21D2255), 112480-000060, KME21E0803 ((01)14026704196224(17)260428(10)KME21E0803), 112480-000060, KME21E0865 ((01)14026704196224(17)260428(10)KME21E0865), 112480-000060, KME21E1222 ((01)14026704196224(17)260428(10)KME21E1222), 112480-000060, KME21E1100 ((01)14026704196224(17)260428(10)KME21E1100), 112480-000060, KME21F0297 ((01)14026704196224(17)260528(10)KME21F0297), 112480-000065, KME21D2714 ((01)14026704196231(17)260328(10)KME21D2714), 112480-000065, KME21F0519 ((01)14026704196231(17)260528(10)KME21F0519), 112480-000065, KME21F0780 ((01)14026704196231(17)260528(10)KME21F0780), 112480-000065, KME21F0009 ((01)14026704196231(17)260528(10)KME21F0009), 112480-000065, KME21D2776 ((01)14026704196231(17)260328(10)KME21D2776), 112480-000065, KME21D2902 ((01)14026704196231(17)260328(10)KME21D2902), 112480-000065, KME21E0802 ((01)14026704196231(17)260428(10)KME21E0802), 112480-000065, KME21E1181 ((01)14026704196231(17)260428(10)KME21E1181), 112480-000065, KME21E1347 ((01)14026704196231(17)260428(10)KME21E1347), 112480-000065, KME21E1408 ((01)14026704196231(17)260428(10)KME21E1408), 112480-000065, KME21F0010 ((01)14026704196231(17)260528(10)KME21F0010), 112480-000065, KME21E1499 ((01)14026704196231(17)260428(10)KME21E1499), 112480-000065, KME21F0427 ((01)14026704196231(17)260528(10)KME21F0427), 112480-000065, KME21F0690 ((01)14026704196231(17)260528(10)KME21F0690), 112480-000065, KME21F0830 ((01)14026704196231(17)260528(10)KME21F0830), 112480-000070, KME21D2010 ((01)14026704196248(17)260328(10)KME21D2010), 112480-000070, KME21F0116 ((01)14026704196248(17)260528(10)KME21F0116), 112480-000070, KME21E1368 ((01)14026704196248(17)260428(10)KME21E1368), 112480-000070, KME21D1574 ((01)14026704196248(17)260328(10)KME21D1574), 112480-000070, KME21D1701 ((01)14026704196248(17)260328(10)KME21D1701), 112480-000070, KME21D1797 ((01)14026704196248(17)260328(10)KME21D1797), 112480-000070, KME21D1826 ((01)14026704196248(17)260328(10)KME21D1826), 112480-000070, KME21D1879 ((01)14026704196248(17)260328(10)KME21D1879), 112480-000070, KME21D2019 ((01)14026704196248(17)260328(10)KME21D2019), 112480-000070, KME21D2258 ((01)14026704196248(17)260328(10)KME21D2258), 112480-000070, KME21D2508 ((01)14026704196248(17)260328(10)KME21D2508), 112480-000070, KME21D2538 ((01)14026704196248(17)260328(10)KME21D2538), 112480-000070, KME21D2644 ((01)14026704196248(17)260328(10)KME21D2644), 112480-000070, KME21D2596 ((01)14026704196248(17)260328(10)KME21D2596), 112480-000070, KME21D2715 ((01)14026704196248(17)260328(10)KME21D2715), 112480-000070, KME21D2753 ((01)14026704196248(17)260328(10)KME21D2753), 112480-000070, KME21D2827 ((01)14026704196248(17)260328(10)KME21D2827), 112480-000070, KME21D2904 ((01)14026704196248(17)260328(10)KME21D2904), 112480-000070, KME21E0019 ((01)14026704196248(17)260428(10)KME21E0019), 112480-000070, KME21E0499 ((01)14026704196248(17)260428(10)KME21E0499), 112480-000070, KME21E0586 ((01)14026704196248(17)260428(10)KME21E0586), 112480-000070, KME21E1500 ((01)14026704196248(17)260428(10)KME21E1500), 112480-000070, KME21F0180 ((01)14026704196248(17)260528(10)KME21F0180), 112480-000070, KME21F0278 ((01)14026704196248(17)260528(10)KME21F0278), 112480-000070, KME21F0413 ((01)14026704196248(17)260528(10)KME21F0413), 112480-000070, KME21F0544 ((01)14026704196248(17)260528(10)KME21F0544), 112480-000070, KME21F0713 ((01)14026704196248(17)260528(10)KME21F0713), 112480-000075, KME21D1827 ((01)14026704196255(17)260328(10)KME21D1827), 112480-000075, KME21E1282 ((01)14026704196255(17)260428(10)KME21E1282), 112480-000075, KME21F0183 ((01)14026704196255(17)260528(10)KME21F0183), 112480-000075, KME21F0777 ((01)14026704196255(17)260528(10)KME21F0777), 112480-000075, KME21D2700 ((01)14026704196255(17)260328(10)KME21D2700), 112480-000075, KME21D2828 ((01)14026704196255(17)260328(10)KME21D2828), 112480-000075, KME21E0331 ((01)14026704196255(17)260428(10)KME21E0331), 112480-000075, KME21E0744 ((01)14026704196255(17)260428(10)KME21E0744), 112480-000075, KME21E0903 ((01)14026704196255(17)260428(10)KME21E0903), 112480-000075, KME21E0928 ((01)14026704196255(17)260428(10)KME21E0928), 112480-000075, KME21E1014 ((01)14026704196255(17)260428(10)KME21E1014), 112480-000075, KME21E1015 ((01)14026704196255(17)260428(10)KME21E1015), 112480-000075, KME21E1047 ((01)14026704196255(17)260428(10)KME21E1047), 112480-000075, KME21E1108 ((01)14026704196255(17)260428(10)KME21E1108), 112480-000075, KME21E1146 ((01)14026704196255(17)260428(10)KME21E1146), 112480-000075, KME21E1168 ((01)14026704196255(17)260428(10)KME21E1168), 112480-000075, KME21E1190 ((01)14026704196255(17)260428(10)KME21E1190), 112480-000075, KME21E1192 ((01)14026704196255(17)260428(10)KME21E1192), 112480-000075, KME21E1209 ((01)14026704196255(17)260428(10)KME21E1209), 112480-000075, KME21E1263 ((01)14026704196255(17)260428(10)KME21E1263), 112480-000075, KME21E1268 ((01)14026704196255(17)260428(10)KME21E1268), 112480-000075, KME21E1348 ((01)14026704196255(17)260428(10)KME21E1348), 112480-000075, KME21E1394 ((01)14026704196255(17)260428(10)KME21E1394), 112480-000075, KME21F0100 ((01)14026704196255(17)260528(10)KME21F0100), 112480-000075, KME21E1501 ((01)14026704196255(17)260428(10)KME21E1501), 112480-000075, KME21F0012 ((01)14026704196255(17)260528(10)KME21F0012), 112480-000075, KME21F0237 ((01)14026704196255(17)260528(10)KME21F0237), 112480-000075, KME21F0296 ((01)14026704196255(17)260528(10)KME21F0296), 112480-000075, KME21F0433 ((01)14026704196255(17)260528(10)KME21F0433), 112480-000075, KME21F0533 ((01)14026704196255(17)260528(10)KME21F0533), 112480-000075, KME21F0712 ((01)14026704196255(17)260528(10)KME21F0712), 112480-000075, KME21F0875 ((01)14026704196255(17)260528(10)KME21F0875), 112480-000075, KME21F0941 ((01)14026704196255(17)260528(10)KME21F0941), 112480-000075, KME21F0970 ((01)14026704196255(17)260528(10)KME21F0970), 112480-000075, KME21F1005 ((01)14026704196255(17)260528(10)KME21F1005), 112480-000075, KME21F1066 ((01)14026704196255(17)260528(10)KME21F1066), 112480-000075, KME21F1108 ((01)14026704196255(17)260528(10)KME21F1108), 112480-000075, KME21F1141 ((01)14026704196255(17)260528(10)KME21F1141), 112480-000080, KME21D1880 ((01)14026704196262(17)260328(10)KME21D1880), 112480-000080, KME21F1139 ((01)14026704196262(17)260528(10)KME21F1139), 112480-000080, KME21D2256 ((01)14026704196262(17)260328(10)KME21D2256), 112480-000080, KME21D2788 ((01)14026704196262(17)260328(10)KME21D2788), 112480-000080, KME21D2789 ((01)14026704196262(17)260328(10)KME21D2789), 112480-000080, KME21D2906 ((01)14026704196262(17)260328(10)KME21D2906), 112480-000080, KME21D2938 ((01)14026704196262(17)260328(10)KME21D2938), 112480-000080, KME21E0020 ((01)14026704196262(17)260428(10)KME21E0020), 112480-000080, KME21E0500 ((01)14026704196262(17)260428(10)KME21E0500), 112480-000080, KME21F0636 ((01)14026704196262(17)260528(10)KME21F0636), 112480-000080, KME21E0468 ((01)14026704196262(17)260428(10)KME21E0468), 112480-000080, KME21E0592 ((01)14026704196262(17)260428(10)KME21E0592), 112480-000080, KME21E1223 ((01)14026704196262(17)260428(10)KME21E1223), 112480-000080, KME21E1346 ((01)14026704196262(17)260428(10)KME21E1346), 112480-000080, KME21E1371 ((01)14026704196262(17)260428(10)KME21E1371), 112480-000080, KME21E1476 ((01)14026704196262(17)260428(10)KME21E1476), 112480-000080, KME21F0151 ((01)14026704196262(17)260528(10)KME21F0151), 112480-000080, KME21F0011 ((01)14026704196262(17)260528(10)KME21F0011), 112480-000080, KME21F0186 ((01)14026704196262(17)260528(10)KME21F0186), 112480-000080, KME21F0276 ((01)14026704196262(17)260528(10)KME21F0276), 112480-000080, KME21F0333 ((01)14026704196262(17)260528(10)KME21F0333), 112480-000080, KME21F0386 ((01)14026704196262(17)260528(10)KME21F0386), 112480-000080, KME21F0419 ((01)14026704196262(17)260528(10)KME21F0419), 112480-000080, KME21F0683 ((01)14026704196262(17)260528(10)KME21F0683), 112480-000080, KME21F0705 ((01)14026704196262(17)260528(10)KME21F0705), 112480-000080, KME21F0973 ((01)14026704196262(17)260528(10)KME21F0973), 112480-000080, KME21F1074 ((01)14026704196262(17)260528(10)KME21F1074), 112480-000080, KME21F1159 ((01)14026704196262(17)260528(10)KME21F1159), 112480-000085, KME21E1189 ((01)14026704196279(17)260428(10)KME21E1189), 112480-000085, KME21E1443 ((01)14026704196279(17)260428(10)KME21E1443), 112480-000085, KME21F0235 ((01)14026704196279(17)260528(10)KME21F0235), 112480-000090, KME21D1982 ((01)14026704196286(17)260328(10)KME21D1982), 112480-000095, KME21E1409 ((01)14026704196293(17)260428(10)KME21E1409), 112480-000095, KME21E1134 ((01)14026704196293(17)260428(10)KME21E1134); e. Product Code, Batch Number (UDI): 112482-000050, KME21E0688 ((01)14026704196460(17)260428(10)KME21E0688), 112482-000050, KME21F0720 ((01)14026704196460(17)260528(10)KME21F0720), 112482-000050, KME21D1704 ((01)14026704196460(17)260328(10)KME21D1704), 112482-000050, KME21D2260 ((01)14026704196460(17)260328(10)KME21D2260), 112482-000050, KME21D2534 ((01)14026704196460(17)260328(10)KME21D2534), 112482-000050, KME21D2499 ((01)14026704196460(17)260328(10)KME21D2499), 112482-000050, KME21D2701 ((01)14026704196460(17)260328(10)KME21D2701), 112482-000050, KME21D2562 ((01)14026704196460(17)260328(10)KME21D2562), 112482-000050, KME21D2790 ((01)14026704196460(17)260328(10)KME21D2790), 112482-000050, KME21D2842 ((01)14026704196460(17)260328(10)KME21D2842), 112482-000050, KME21E0382 ((01)14026704196460(17)260428(10)KME21E0382), 112482-000050, KME21E0465 ((01)14026704196460(17)260428(10)KME21E0465), 112482-000050, KME21E0801 ((01)14026704196460(17)260428(10)KME21E0801), 112482-000050, KME21F0236 ((01)14026704196460(17)260528(10)KME21F0236), 112482-000050, KME21E1502 ((01)14026704196460(17)260428(10)KME21E1502), 112482-000050, KME21F0355 ((01)14026704196460(17)260528(10)KME21F0355), 112482-000050, KME21F0420 ((01)14026704196460(17)260528(10)KME21F0420), 112482-000050, KME21F0518 ((01)14026704196460(17)260528(10)KME21F0518), 112482-000050, KME21F0536 ((01)14026704196460(17)260528(10)KME21F0536), 112482-000050, KME21F0593 ((01)14026704196460(17)260528(10)KME21F0593), 112482-000050, KME21F1004 ((01)14026704196460(17)260528(10)KME21F1004), 112482-000055, KME21F0718 ((01)14026704196477(17)260528(10)KME21F0718), 112482-000055, KME21F0832 ((01)14026704196477(17)260528(10)KME21F0832), 112482-000055, KME21D1587 ((01)14026704196477(17)260328(10)KME21D1587), 112482-000055, KME21D1966 ((01)14026704196477(17)260328(10)KME21D1966), 112482-000055, KME21D2261 ((01)14026704196477(17)260328(10)KME21D2261), 112482-000055, KME21D2385 ((01)14026704196477(17)260328(10)KME21D2385), 112482-000055, KME21D2417 ((01)14026704196477(17)260328(10)KME21D2417), 112482-000055, KME21E0219 ((01)14026704196477(17)260428(10)KME21E0219), 112482-000055, KME21E0593 ((01)14026704196477(17)260428(10)KME21E0593), 112482-000055, KME21E0637 ((01)14026704196477(17)260428(10)KME21E0637), 112482-000055, KME21E0689 ((01)14026704196477(17)260428(10)KME21E0689), 112482-000055, KME21E0800 ((01)14026704196477(17)260428(10)KME21E0800), 112482-000055, KME21E0882 ((01)14026704196477(17)260428(10)KME21E0882), 112482-000055, KME21F1068 ((01)14026704196477(17)260528(10)KME21F1068), 112482-000055, KME21F1069 ((01)14026704196477(17)260528(10)KME21F1069), 112482-000060, KME21D2898 ((01)14026704196484(17)260328(10)KME21D2898), 112482-000060, KME21E0682 ((01)14026704196484(17)260428(10)KME21E0682), 112482-000060, KME21F0543 ((01)14026704196484(17)260528(10)KME21F0543), 112482-000060, KME21F0806 ((01)14026704196484(17)260528(10)KME21F0806), 112482-000060, KME21D2531 ((01)14026704196484(17)260328(10)KME21D2531), 112482-000060, KME21D2647 ((01)14026704196484(17)260328(10)KME21D2647), 112482-000060, KME21D2751 ((01)14026704196484(17)260328(10)KME21D2751), 112482-000060, KME21D2829 ((01)14026704196484(17)260328(10)KME21D2829), 112482-000060, KME21D2927 ((01)14026704196484(17)260328(10)KME21D2927), 112482-000060, KME21E0018 ((01)14026704196484(17)260428(10)KME21E0018), 112482-000060, KME21E0220 ((01)14026704196484(17)260428(10)KME21E0220), 112482-000060, KME21E0549 ((01)14026704196484(17)260428(10)KME21E0549), 112482-000060, KME21E0327 ((01)14026704196484(17)260428(10)KME21E0327), 112482-000060, KME21E0444 ((01)14026704196484(17)260428(10)KME21E0444), 112482-000060, KME21E0550 ((01)14026704196484(17)260428(10)KME21E0550), 112482-000060, KME21E0466 ((01)14026704196484(17)260428(10)KME21E0466), 112482-000060, KME21E0931 ((01)14026704196484(17)260428(10)KME21E0931), 112482-000060, KME21E1149 ((01)14026704196484(17)260428(10)KME21E1149), 112482-000060, KME21F0473 ((01)14026704196484(17)260528(10)KME21F0473), 112482-000060, KME21F0592 ((01)14026704196484(17)260528(10)KME21F0592), 112482-000060, KME21F0649 ((01)14026704196484(17)260528(10)KME21F0649), 112482-000060, KME21F0807 ((01)14026704196484(17)260528(10)KME21F0807), 112482-000060, KME21F0929 ((01)14026704196484(17)260528(10)KME21F0929), 112482-000060, KME21F0979 ((01)14026704196484(17)260528(10)KME21F0979), 112482-000060, KME21F1075 ((01)14026704196484(17)260528(10)KME21F1075), 112482-000060, KME21F1113 ((01)14026704196484(17)260528(10)KME21F1113), 112482-000065, KME21D2900 ((01)14026704196491(17)260328(10)KME21D2900), 112482-000065, KME21F0552 ((01)14026704196491(17)260528(10)KME21F0552), 112482-000065, KME21F0729 ((01)14026704196491(17)260528(10)KME21F0729), 112482-000065, KME21D1573 ((01)14026704196491(17)260328(10)KME21D1573), 112482-000065, KME21D1650 ((01)14026704196491(17)260328(10)KME21D1650), 112482-000065, KME21D1793 ((01)14026704196491(17)260328(10)KME21D1793), 112482-000065, KME21D1944 ((01)14026704196491(17)260328(10)KME21D1944), 112482-000065, KME21D1981 ((01)14026704196491(17)260328(10)KME21D1981), 112482-000065, KME21D2384 ((01)14026704196491(17)260328(10)KME21D2384), 112482-000065, KME21D2418 ((01)14026704196491(17)260328(10)KME21D2418), 112482-000065, KME21D2526 ((01)14026704196491(17)260328(10)KME21D2526), 112482-000065, KME21D2611 ((01)14026704196491(17)260328(10)KME21D2611), 112482-000065, KME21D2598 ((01)14026704196491(17)260328(10)KME21D2598), 112482-000065, KME21D2883 ((01)14026704196491(17)260328(10)KME21D2883), 112482-000065, KME21E0115 ((01)14026704196491(17)260428(10)KME21E0115), 112482-000065, KME21E0221 ((01)14026704196491(17)260428(10)KME21E0221), 112482-000065, KME21E0520 ((01)14026704196491(17)260428(10)KME21E0520), 112482-000065, KME21E0594 ((01)14026704196491(17)260428(10)KME21E0594), 112482-000065, KME21E0551 ((01)14026704196491(17)260428(10)KME21E0551), 112482-000065, KME21E0684 ((01)14026704196491(17)260428(10)KME21E0684), 112482-000065, KME21E0799 ((01)14026704196491(17)260428(10)KME21E0799), 112482-000065, KME21E0806 ((01)14026704196491(17)260428(10)KME21E0806), 112482-000065, KME21E1225 ((01)14026704196491(17)260428(10)KME21E1225), 112482-000065, KME21F0635 ((01)14026704196491(17)260528(10)KME21F0635), 112482-000065, KME21F0584 ((01)14026704196491(17)260528(10)KME21F0584), 112482-000065, KME21F1002 ((01)14026704196491(17)260528(10)KME21F1002), 112482-000070, KME21D2699 ((01)14026704196507(17)260328(10)KME21D2699), 112482-000070, KME21E1370 ((01)14026704196507(17)260428(10)KME21E1370), 112482-000070, KME21F0316 ((01)14026704196507(17)260528(10)KME21F0316), 112482-000070, KME21E1140 ((01)14026704196507(17)260428(10)KME21E1140), 112482-000070, KME21F1077 ((01)14026704196507(17)260528(10)KME21F1077), 112482-000070, KME21D1567 ((01)14026704196507(17)260328(10)KME21D1567), 112482-000070, KME21D1651 ((01)14026704196507(17)260328(10)KME21D1651), 112482-000070, KME21D1702 ((01)14026704196507(17)260328(10)KME21D1702), 112482-000070, KME21D1799 ((01)14026704196507(17)260328(10)KME21D1799), 112482-000070, KME21D1945 ((01)14026704196507(17)260328(10)KME21D1945), 112482-000070, KME21D1952 ((01)14026704196507(17)260328(10)KME21D1952), 112482-000070, KME21D2009 ((01)14026704196507(17)260328(10)KME21D2009), 112482-000070, KME21D2016 ((01)14026704196507(17)260328(10)KME21D2016), 112482-000070, KME21D2263 ((01)14026704196507(17)260328(10)KME21D2263), 112482-000070, KME21D2640 ((01)14026704196507(17)260328(10)KME21D2640), 112482-000070, KME21D2383 ((01)14026704196507(17)260328(10)KME21D2383), 112482-000070, KME21D2528 ((01)14026704196507(17)260328(10)KME21D2528), 112482-000070, KME21D2497 ((01)14026704196507(17)260328(10)KME21D2497), 112482-000070, KME21D2535 ((01)14026704196507(17)260328(10)KME21D2535), 112482-000070, KME21D2612 ((01)14026704196507(17)260328(10)KME21D2612), 112482-000070, KME21D2599 ((01)14026704196507(17)260328(10)KME21D2599), 112482-000070, KME21D2750 ((01)14026704196507(17)260328(10)KME21D2750), 112482-000070, KME21D2843 ((01)14026704196507(17)260328(10)KME21D2843), 112482-000070, KME21D2798 ((01)14026704196507(17)260328(10)KME21D2798), 112482-000070, KME21D2800 ((01)14026704196507(17)260328(10)KME21D2800), 112482-000070, KME21D2865 ((01)14026704196507(17)260328(10)KME21D2865), 112482-000070, KME21D2899 ((01)14026704196507(17)260328(10)KME21D2899), 112482-000070, KME21E0015 ((01)14026704196507(17)260428(10)KME21E0015), 112482-000070, KME21E0282 ((01)14026704196507(17)260428(10)KME21E0282), 112482-000070, KME21E0328 ((01)14026704196507(17)260428(10)KME21E0328), 112482-000070, KME21E0530 ((01)14026704196507(17)260428(10)KME21E0530), 112482-000070, KME21E0470 ((01)14026704196507(17)260428(10)KME21E0470), 112482-000070, KME21E0552 ((01)14026704196507(17)260428(10)KME21E0552), 112482-000070, KME21E0563 ((01)14026704196507(17)260428(10)KME21E0563), 112482-000070, KME21E1040 ((01)14026704196507(17)260428(10)KME21E1040), 112482-000070, KME21E0749 ((01)14026704196507(17)260428(10)KME21E0749), 112482-000070, KME21E0807 ((01)14026704196507(17)260428(10)KME21E0807), 112482-000070, KME21E0850 ((01)14026704196507(17)260428(10)KME21E0850), 112482-000070, KME21E0863 ((01)14026704196507(17)260428(10)KME21E0863), 112482-000070, KME21E0878 ((01)14026704196507(17)260428(10)KME21E0878), 112482-000070, KME21E0913 ((01)14026704196507(17)260428(10)KME21E0913), 112482-000070, KME21E0915 ((01)14026704196507(17)260428(10)KME21E0915), 112482-000070, KME21E0982 ((01)14026704196507(17)260428(10)KME21E0982), 112482-000070, KME21E1006 ((01)14026704196507(17)260428(10)KME21E1006), 112482-000070, KME21E1007 ((01)14026704196507(17)260428(10)KME21E1007), 112482-000070, KME21E1086 ((01)14026704196507(17)260428(10)KME21E1086), 112482-000070, KME21E1087 ((01)14026704196507(17)260428(10)KME21E1087), 112482-000070, KME21E1101 ((01)14026704196507(17)260428(10)KME21E1101), 112482-000070, KME21E1188 ((01)14026704196507(17)260428(10)KME21E1188), 112482-000070, KME21E1166 ((01)14026704196507(17)260428(10)KME21E1166), 112482-000070, KME21E1198 ((01)14026704196507(17)260428(10)KME21E1198), 112482-000070, KME21E1208 ((01)14026704196507(17)260428(10)KME21E1208), 112482-000070, KME21E1317 ((01)14026704196507(17)260428(10)KME21E1317), 112482-000070, KME21E1334 ((01)14026704196507(17)260428(10)KME21E1334), 112482-000070, KME21E1345 ((01)14026704196507(17)260428(10)KME21E1345), 112482-000070, KME21E1369 ((01)14026704196507(17)260428(10)KME21E1369), 112482-000070, KME21E1441 ((01)14026704196507(17)260428(10)KME21E1441), 112482-000070, KME21E1477 ((01)14026704196507(17)260428(10)KME21E1477), 112482-000070, KME21F0126 ((01)14026704196507(17)260528(10)KME21F0126), 112482-000070, KME21F0014 ((01)14026704196507(17)260528(10)KME21F0014), 112482-000070, KME21F0138 ((01)14026704196507(17)260528(10)KME21F0138), 112482-000070, KME21F0148 ((01)14026704196507(17)260528(10)KME21F0148), 112482-000070, KME21F0181 ((01)14026704196507(17)260528(10)KME21F0181), 112482-000070, KME21F0214 ((01)14026704196507(17)260528(10)KME21F0214), 112482-000070, KME21F0234 ((01)14026704196507(17)260528(10)KME21F0234), 112482-000070, KME21F0308 ((01)14026704196507(17)260528(10)KME21F0308), 112482-000070, KME21F0942 ((01)14026704196507(17)260528(10)KME21F0942), 112482-000070, KME21F0418 ((01)14026704196507(17)260528(10)KME21F0418), 112482-000070, KME21F0515 ((01)14026704196507(17)260528(10)KME21F0515), 112482-000070, KME21F0640 ((01)14026704196507(17)260528(10)KME21F0640), 112482-000070, KME21F0688 ((01)14026704196507(17)260528(10)KME21F0688), 112482-000070, KME21F0779 ((01)14026704196507(17)260528(10)KME21F0779), 112482-000070, KME21F0831 ((01)14026704196507(17)260528(10)KME21F0831), 112482-000070, KME21F0856 ((01)14026704196507(17)260528(10)KME21F0856), 112482-000070, KME21F0873 ((01)14026704196507(17)260528(10)KME21F0873), 112482-000070, KME21F0971 ((01)14026704196507(17)260528(10)KME21F0971), 112482-000070, KME21F1107 ((01)14026704196507(17)260528(10)KME21F1107), 112482-000070, KME21F1142 ((01)14026704196507(17)260528(10)KME21F1142), 112482-000070, KME21F1157 ((01)14026704196507(17)260528(10)KME21F1157), 112482-000075, KME21D2667 ((01)14026704196514(17)260328(10)KME21D2667), 112482-000075, KME21E0491 ((01)14026704196514(17)260428(10)KME21E0491), 112482-000075, KME21F0697 ((01)14026704196514(17)260528(10)KME21F0697), 112482-000075, KME21F0531 ((01)14026704196514(17)260528(10)KME21F0531), 112482-000075, KME21D2519 ((01)14026704196514(17)260328(10)KME21D2519), 112482-000075, KME21D1579 ((01)14026704196514(17)260328(10)KME21D1579), 112482-000075, KME21D1648 ((01)14026704196514(17)260328(10)KME21D1648), 112482-000075, KME21D1795 ((01)14026704196514(17)260328(10)KME21D1795), 112482-000075, KME21D1881 ((01)14026704196514(17)260328(10)KME21D1881), 112482-000075, KME21D1891 ((01)14026704196514(17)260328(10)KME21D1891), 112482-000075, KME21D1960 ((01)14026704196514(17)260328(10)KME21D1960), 112482-000075, KME21D2013 ((01)14026704196514(17)260328(10)KME21D2013), 112482-000075, KME21D2017 ((01)14026704196514(17)260328(10)KME21D2017), 112482-000075, KME21D2264 ((01)14026704196514(17)260328(10)KME21D2264), 112482-000075, KME21D2404 ((01)14026704196514(17)260328(10)KME21D2404), 112482-000075, KME21D2420 ((01)14026704196514(17)260328(10)KME21D2420), 112482-000075, KME21D2440 ((01)14026704196514(17)260328(10)KME21D2440), 112482-000075, KME21D2503 ((01)14026704196514(17)260328(10)KME21D2503), 112482-000075, KME21D2518 ((01)14026704196514(17)260328(10)KME21D2518), 112482-000075, KME21D2642 ((01)14026704196514(17)260328(10)KME21D2642), 112482-000075, KME21D2564 ((01)14026704196514(17)260328(10)KME21D2564), 112482-000075, KME21D2600 ((01)14026704196514(17)260328(10)KME21D2600), 112482-000075, KME21D2794 ((01)14026704196514(17)260328(10)KME21D2794), 112482-000075, KME21D2926 ((01)14026704196514(17)260328(10)KME21D2926), 112482-000075, KME21E0017 ((01)14026704196514(17)260428(10)KME21E0017), 112482-000075, KME21E0222 ((01)14026704196514(17)260428(10)KME21E0222), 112482-000075, KME21E0263 ((01)14026704196514(17)260428(10)KME21E0263), 112482-000075, KME21E0325 ((01)14026704196514(17)260428(10)KME21E0325), 112482-000075, KME21E0490 ((01)14026704196514(17)260428(10)KME21E0490), 112482-000075, KME21E0895 ((01)14026704196514(17)260428(10)KME21E0895), 112482-000075, KME21E0662 ((01)14026704196514(17)260428(10)KME21E0662), 112482-000075, KME21E1284 ((01)14026704196514(17)260428(10)KME21E1284), 112482-000075, KME21E1449 ((01)14026704196514(17)260428(10)KME21E1449), 112482-000075, KME21F0103 ((01)14026704196514(17)260528(10)KME21F0103), 112482-000075, KME21F0179 ((01)14026704196514(17)260528(10)KME21F0179), 112482-000075, KME21F0215 ((01)14026704196514(17)260528(10)KME21F0215), 112482-000075, KME21F0300 ((01)14026704196514(17)260528(10)KME21F0300), 112482-000075, KME21F0356 ((01)14026704196514(17)260528(10)KME21F0356), 112482-000075, KME21F0382 ((01)14026704196514(17)260528(10)KME21F0382), 112482-000075, KME21F0469 ((01)14026704196514(17)260528(10)KME21F0469), 112482-000075, KME21F0472 ((01)14026704196514(17)260528(10)KME21F0472), 112482-000075, KME21F0517 ((01)14026704196514(17)260528(10)KME21F0517), 112482-000075, KME21F0648 ((01)14026704196514(17)260528(10)KME21F0648), 112482-000075, KME21F0587 ((01)14026704196514(17)260528(10)KME21F0587), 112482-000075, KME21F0585 ((01)14026704196514(17)260528(10)KME21F0585), 112482-000075, KME21F0588 ((01)14026704196514(17)260528(10)KME21F0588), 112482-000075, KME21F0684 ((01)14026704196514(17)260528(10)KME21F0684), 112482-000075, KME21F0685 ((01)14026704196514(17)260528(10)KME21F0685), 112482-000075, KME21F0858 ((01)14026704196514(17)260528(10)KME21F0858), 112482-000075, KME21F0857 ((01)14026704196514(17)260528(10)KME21F0857), 112482-000075, KME21F0871 ((01)14026704196514(17)260528(10)KME21F0871), 112482-000075, KME21F0918 ((01)14026704196514(17)260528(10)KME21F0918), 112482-000075, KME21F0972 ((01)14026704196514(17)260528(10)KME21F0972), 112482-000075, KME21F1001 ((01)14026704196514(17)260528(10)KME21F1001), 112482-000075, KME21F1031 ((01)14026704196514(17)260528(10)KME21F1031), 112482-000075, KME21F1057 ((01)14026704196514(17)260528(10)KME21F1057), 112482-000075, KME21F1112 ((01)14026704196514(17)260528(10)KME21F1112), 112482-000080, KME21D1825 ((01)14026704196521(17)260328(10)KME21D1825), 112482-000080, KME21E0016 ((01)14026704196521(17)260428(10)KME21E0016), 112482-000080, KME21F0514 ((01)14026704196521(17)260528(10)KME21F0514), 112482-000080, KME21F0776 ((01)14026704196521(17)260528(10)KME21F0776), 112482-000080, KME21F1030 ((01)14026704196521(17)260528(10)KME21F1030), 112482-000080, KME21D1882 ((01)14026704196521(17)260328(10)KME21D1882), 112482-000080, KME21D1663 ((01)14026704196521(17)260328(10)KME21D1663), 112482-000080, KME21D2441 ((01)14026704196521(17)260328(10)KME21D2441), 112482-000080, KME21D2506 ((01)14026704196521(17)260328(10)KME21D2506), 112482-000080, KME21D2498 ((01)14026704196521(17)260328(10)KME21D2498), 112482-000080, KME21D2520 ((01)14026704196521(17)260328(10)KME21D2520), 112482-000080, KME21D2646 ((01)14026704196521(17)260328(10)KME21D2646), 112482-000080, KME21D2566 ((01)14026704196521(17)260328(10)KME21D2566), 112482-000080, KME21D2717 ((01)14026704196521(17)260328(10)KME21D2717), 112482-000080, KME21D2791 ((01)14026704196521(17)260328(10)KME21D2791), 112482-000080, KME21D2844 ((01)14026704196521(17)260328(10)KME21D2844), 112482-000080, KME21D2925 ((01)14026704196521(17)260328(10)KME21D2925), 112482-000080, KME21E0501 ((01)14026704196521(17)260428(10)KME21E0501), 112482-000080, KME21E1038 ((01)14026704196521(17)260428(10)KME21E1038), 112482-000080, KME21E0638 ((01)14026704196521(17)260428(10)KME21E0638), 112482-000080, KME21E0748 ((01)14026704196521(17)260428(10)KME21E0748), 112482-000080, KME21E0883 ((01)14026704196521(17)260428(10)KME21E0883), 112482-000080, KME21E0935 ((01)14026704196521(17)260428(10)KME21E0935), 112482-000080, KME21E1210 ((01)14026704196521(17)260428(10)KME21E1210), 112482-000080, KME21E1265 ((01)14026704196521(17)260428(10)KME21E1265), 112482-000080, KME21F0650 ((01)14026704196521(17)260528(10)KME21F0650), 112482-000080, KME21E1269 ((01)14026704196521(17)260428(10)KME21E1269), 112482-000080, KME21E1318 ((01)14026704196521(17)260428(10)KME21E1318), 112482-000080, KME21E1387 ((01)14026704196521(17)260428(10)KME21E1387), 112482-000080, KME21E1442 ((01)14026704196521(17)260428(10)KME21E1442), 112482-000080, KME21E1454 ((01)14026704196521(17)260428(10)KME21E1454), 112482-000080, KME21F0110 ((01)14026704196521(17)260528(10)KME21F0110), 112482-000080, KME21F0007 ((01)14026704196521(17)260528(10)KME21F0007), 112482-000080, KME21F0139 ((01)14026704196521(17)260528(10)KME21F0139), 112482-000080, KME21F0149 ((01)14026704196521(17)260528(10)KME21F0149), 112482-000080, KME21F0349 ((01)14026704196521(17)260528(10)KME21F0349), 112482-000080, KME21F0381 ((01)14026704196521(17)260528(10)KME21F0381), 112482-000080, KME21F0476 ((01)14026704196521(17)260528(10)KME21F0476), 112482-000080, KME21F0502 ((01)14026704196521(17)260528(10)KME21F0502), 112482-000080, KME21F0532 ((01)14026704196521(17)260528(10)KME21F0532), 112482-000080, KME21F0687 ((01)14026704196521(17)260528(10)KME21F0687), 112482-000080, KME21F0717 ((01)14026704196521(17)260528(10)KME21F0717), 112482-000080, KME21F0774 ((01)14026704196521(17)260528(10)KME21F0774), 112482-000080, KME21F0804 ((01)14026704196521(17)260528(10)KME21F0804), 112482-000080, KME21F0826 ((01)14026704196521(17)260528(10)KME21F0826), 112482-000080, KME21F0880 ((01)14026704196521(17)260528(10)KME21F0880), 112482-000080, KME21F0967 ((01)14026704196521(17)260528(10)KME21F0967), 112482-000080, KME21F1110 ((01)14026704196521(17)260528(10)KME21F1110), 112482-000080, KME21F0968 ((01)14026704196521(17)260528(10)KME21F0968), 112482-000080, KME21F1072 ((01)14026704196521(17)260528(10)KME21F1072), 112482-000080, KME21F1158 ((01)14026704196521(17)260528(10)KME21F1158), 112482-000085, KME21D1883 ((01)14026704196538(17)260328(10)KME21D1883), 112482-000085, KME21E1319 ((01)14026704196538(17)260428(10)KME21E1319), 112482-000085, KME21D1577 ((01)14026704196538(17)260328(10)KME21D1577), 112482-000085, KME21D1800 ((01)14026704196538(17)260328(10)KME21D1800), 112482-000085, KME21D1889 ((01)14026704196538(17)260328(10)KME21D1889), 112482-000085, KME21E0916 ((01)14026704196538(17)260428(10)KME21E0916), 112482-000085, KME21E0957 ((01)14026704196538(17)260428(10)KME21E0957), 112482-000085, KME21E1039 ((01)14026704196538(17)260428(10)KME21E1039), 112482-000085, KME21E1206 ((01)14026704196538(17)260428(10)KME21E1206), 112482-000085, KME21E1388 ((01)14026704196538(17)260428(10)KME21E1388), 112482-000085, KME21F0638 ((01)14026704196538(17)260528(10)KME21F0638), 112482-000085, KME21F0837 ((01)14026704196538(17)260528(10)KME21F0837), 112482-000085, KME21F1003 ((01)14026704196538(17)260528(10)KME21F1003), 112482-000085, KME21F1032 ((01)14026704196538(17)260528(10)KME21F1032), 112482-000085, KME21F1114 ((01)14026704196538(17)260528(10)KME21F1114), 112482-000090, KME21E1475 ((01)14026704196545(17)260428(10)KME21E1475), 112482-000090, KME21E1041 ((01)14026704196545(17)260428(10)KME21E1041), 112482-000090, KME21E1107 ((01)14026704196545(17)260428(10)KME21E1107), 112482-000090, KME21E1135 ((01)14026704196545(17)260428(10)KME21E1135), 112482-000090, KME21F0154 ((01)14026704196545(17)260528(10)KME21F0154), 112482-000090, KME21F0721 ((01)14026704196545(17)260528(10)KME21F0721), 112482-000095, KME21E0519 ((01)14026704196552(17)260428(10)KME21E0519), 112482-000095, KME21E1109 ((01)14026704196552(17)260428(10)KME21E1109), 112482-000095, KME21E1450 ((01)14026704196552(17)260428(10)KME21E1450), 112482-000095, KME21F0724 ((01)14026704196552(17)260528(10)KME21F0724)
FEI Number 3005747797
Recalling Firm/
Manufacturer
TELEFLEX LLC
3015 Carrington Mill Blvd
Morrisville NC 27560-5437
For Additional Information ContactCustomer Service
866-396-2111
Manufacturer Reason
for Recall
Potential for pilot balloon non-inflation or cuff non-deflation.
FDA Determined
Cause 2
Process control
ActionTeleflex issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 12/08/2022 by FedEx 2nd day mail. The notices explain the issues, risk, and requested the following: "Action list number 1 Medical facilities 1. We request that you immediately check your inventory for product within the scope of this FSCA. Users should cease use and distribution of affected product and immediately quarantine the affected product. 2. If you have impacted product, mark the applicable checkbox on the Acknowledgement Form (Appendix 1) and fax it to 1-855-419-8507, including Attn: Customer Service or using the email address recalls@teleflex.com. This will allow us to document the amount of product you have on hand for return. A customer service representative will contact you with a Return Goods Authorization (RGA) Number and will provide instructions for the return of products to Teleflex Medical. 3. If you do not have impacted product, mark the applicable checkbox on the Acknowledgement Form (Appendix 1) and return the form to Teleflex at the contact details provided. 4. Teleflex (or your local dealer) will issue a credit note upon receipt of the returned affected product. Action list number 2 Distributors 1. Provide this field safety notice to all customers who have received impacted product. Each of your customers is then required to complete the Acknowledgement Form and return it to you. 2. We request that you immediately check your inventory for impacted product. Cease use and distribution of impacted product and immediately quarantine the affected product. You may then return all product in scope. 3. As a distributor, you are then required to confirm to Teleflex that you have completed the field activity outlined in actions 1 and 2 of this Action List Number 2. Upon completion of your actions, please forward the completed Acknowledgement Form to the e-mail address below. Important - Please ensure you only list batch numbers in scope of this Field Safety Notice
Quantity in Commerce497 units
DistributionWorldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, NC, ND, NJ, NV, NY, OH, OK, PA, SC, SD, TX, UT, VA, WA, WI, WV and PR. The countries of Albania, Armenia, Austria, Azerbaijan, Bahamas, Barbados, Belgium, Brazil, Bulgaria, Burkina Faso, Canada, Chile, China, Croatia, Cyprus, Czech, Denmark, El Salvador, Estonia, Finland, France, French Guiana, French Polynesia, Germany, Great Britain, Greece, Hong Kong, Hungary, Iceland, Ireland, Italy Israel, Jordan, Kuwait, Lebanon, Liechtenstein, Malaysia, Mayotte, Mexico, Moldova, Morocco, Nepal, Netherlands, Norway, Panama, Philippines, Poland, Portugal, R¿union, Russia, San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Spain, South Africa, South Korea, Sweden, Switzerland, Taiwan Thailand, Turkey, Ukraine, United Arab Emirates.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-