| | Class 2 Device Recall CardioMEMSTM HF System Patient Electronics System (PES) |  |
| Date Initiated by Firm | February 07, 2023 |
| Date Posted | March 03, 2023 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1218-2023 |
| Recall Event ID |
91610 |
| PMA Number | P100045 |
| Product Classification |
System, hemodynamic, implantable - Product Code MOM
|
| Product | CardioMEMS HF System Patient Electronics System (PES), Model number CM1000, The CardioMEMS HF System provides pulmonary artery (PA) hemodynamic data used for the monitoring and management of heart failure (HF) patients. The system measures changes in PA pressure which physicians use to initiate or modify heart failure treatment. |
| Code Information |
UDI/DI 05414734509169, Serial Numbers: M131100003, M131100004, M131100008, M131100012, M131100013, M131100015, M131200018, M131200022, M131200027, M131200030, M131200033, M131200034, M131200035, M131200036, M131200038, M131200040, M131200043, M131200044, M131200047, M131200048, M131200050, M131200051, M131200052, M131200053, M131200054, M131200057, M131200059, M131200063, M131200066, M131200067, M131200069, M131200070, M131200071, M131200072, M131200073, M131200074, M131200075, M131200076, M131200077, M131200078, M131200079, M131200082, M131200083, M131200085, M131200086, M131200088, M131200090, M131200092, M131200094, M131200095, M131200096, M131200097, M131200099, M131200100, M131200102, M131200103, M131200105, M131200106, M131200108, M131200109, M131200110, M131200111, M131200116, M131200117, M131200118, M131200120, M131200123, M131200128, M131200129, M131200130, M131200134, M131200135, M131200136, M131200137, M131200139, M131200140, M131200142, M131200143, M131200144, M131200145, M131200148, M131200150, M140100158, M140100160, M140100161, M140100163, M140100165, M140100166, M140100167, M140100168, M140100170, M140100176, M140200250, M140200252, M140200254, M140200256, M140200257, M140200258, M140200259, M140200260, M140200261, M140200263, M140200264, M140200266, M140200269, M140200270, M140200271, M140200274, M140200276, M140200279, M140200280, M140200287, M140200288, M140200289, M140200291, M140200293, M140200294, M140200295, M140200299, M140200300, M140200301, M140200302, M140200305, M140200307, M140200308, M140200309, M140200310, M140200311, M140200312, M140200313, M140200317, M140200318, M140200320, M140200321, M140200323, M140300331, M140300332, M140300333, M140300334, M140300335, M140300337, M140300339, M140300340, M140300341, M140300342, M140300344, M140300345, M140300348, M140300350, M140300355, M140300356, M140300357, M140300358, M140300359, M140300360, M140300361, M140300362, M140300363, M140300366, M140300369, M140300371, M140300372, M140300375, M140300379, M140300380, M140300383, M140300385, M140300391, M140300394, M140300395, M140300397, M140300398, M140300399, M140300400, M140300403, M140300404, M140300405, M140300409, M140300410, M140300411, M140300412, M140300414, M140300415, M140300416, M140300417, M140300418, M140300420, M140300421, M140300422, M140300424, M140300426, M140300428, M140300429, M140300430, M140300431, M140300432, M140300433, M140300434, M140300435, M140500487, M140500488, M140500489, M140500490, M140500491, M140500492, M140500494, M140500495, M140500496, M140500497, M140500500, M140500502, M140500503, M140500505, M140500507, M140500508, M140500509, M140500510, M140500511, M140500512, M140500515, M140500516, M140600547, M140600548, M140600549, M140600550, M140600552, M140600553, M140600554, M140600555, M140600556, M140600561, M140600562, M140600564, M140600565, M140600566, M140600567, M140600568, M140600570, M140600571, M140600572, M140600573, M140600574, M140600576, M140600577, M140600580, M140600583, M140600586, M140600587, M140600588, M140600589, M140600590, M140600591, M140600592, M140600594, M140700597, M140700598, M140700600, M140700601, M140700603, M140700604, M140700605, M140700607, M140700609, M140700610, M140700611, M140700619, M140700620, M140700621, M140700622, M140700623, M140700625, M140700626, M140700627, M140700628, M140700632, M140700633, M140700635, M140700636, M140700637, M140700638, M140700640, M140700641, M140700642, M140700643, M140700646, M140700647, M140700648, M140700649, M140700650, M140700652, M140700654, M140700657, M140700658, M140700659, M140700660, M140700661, M140700662, M140700664, M140700665, M140700668, M140700669, M140700671, M140700673, M140700674, M140700675, M140700676, M140700677, M140700678, M140700679, M140700680, M140700683, M140700684, M140700685, M140700727, M140700729, M140700730, M140700731, M140700734, M140700735, M140700737, M140700738, M140700739, M140700740, M140700743, M140700745, M140700747, M140700748, M140700749, M140700751, M140700756, M140700759, M140700760, M140700761, M140700767, M140700768, M140700769, M140700772, M140700774, M140700775, M140700776, M140700778, M140700779, M140700782, M140700785, M140700787, M140700788, M140700789, M140700791, M140700792, M140700793, M140700794, M140700795, M140700807, M140700808, M140800897, M140800900, M140800901, M140800902, M140800904, M140800909, M140800911, M140800914, M140800927, M140800928, M140800929, M140800930, M140800931, M140800933, M140800935, M140800936, M140800937, M140800938, M140800940, M140800941, M140800942, M140800943, M140800944, M140800945, M140800946, M140800947, M140800953, M140800955, M140800957, M140800958, M140800960, M140800961, M140800962, M140800963, M140800966, M140800967, M140800971, M140800972, M140800973, M140800974, M140800975, M140800976, M140800979, M140800981, M140800984, M140800985, M140800986, M140801018, M140801023, M140801025, M140801026, M140901028, M140901029, M140901030, M140901033, M140901035, M140901040, M140901041, M140901042, M140901043, M140901044, M140901045, M140901046, M140901049, M140901051, M140901052, M140901053, M140901055, M140901059, M140901062, M140901064, M140901069, M140901070, M140901071, M140901072, M140901073, M140901074, M140901075, M140901076, M140901078, M140901082, M140901084, M140901086, M140901087, M140901088, M140901089, M140901091, M140901092, M140901094, M140901095, M140901096, M140901097, M140901098, M140901099, M140901100, M140901101, M140901102, M140901103, M140901104, M140901105, M140901106, M140901107, M140901108, M140901111, M140901112, M140901113, M140901115, M140901116, M140901117, M140901119, M140901120, M140901121, M140901122, M140901123, M140901124, M140901127, M140901128, M140901129, M140901130, M140901132, M140901133, M140901135, M140901139, M140901141, M140901144, M140901145, M140901146, M140901147, M140901148, M140901150, M140901151, M140901154, M140901155, M140901156, M140901187, M140901189, M140901191, M140901192, M140901193, M140901194, M140901196, M140901200, M140901202, M140901203, M140901206, M140901207, M140901209, M140901211, M140901214, M140901222, M140901227, M140901229, M140901231, M140901232, M140901233, M140901234, M140901235, M140901238, M140901240, M140901242, M140901244, M140901246, M140901247, M140901248, M140901250, M140901252, M140901253, M140901254, M140901256, M140901258, M140901259, M140901260, M140901263, M140901267, M140901268, M140901273, M140901274, M140901280, M140901281, M140901285, M140901286, M140901289, M140901292, M140901300, M140901302, M140901304, M140901305, M140901306, M140901307, M140901308, M140901310, M140901311, M140901314, M140901315, M140901316, M140901318, M140901320, M140901321, M140901323, M140901325, M140901326, M141001328, M141001329, M141001332, M141001335, M141001336, M141001367, M141001371, M141001372, M141001373, M141001375, M141001376, M141001378, M141001379, M141001380, M141001381, M141001383, M141001385, M141001387, M141001388, M141001390, M141001391, M141001392, M141001395, M141001396, M141001400, M141001406, M141001407, M141001408, M141001410, M141001411, M141001414, M141001420, M141001421, M141001424, M141001428, M141001432, M141001433, M141001436, M141001440, M141001441, M141001447, M141001448, M141001450, M141001452, M141001453, M141001454, M141001455, M141001456, M141001458, M141001460, M141001461, M141001462, M141001465, M141001469, M141001470, M141001472, M141001473, M141001474, M141001475, M141001478, M141001480, M141001481, M141001482, M141001484, M141001485, M141001486, M141001487, M141001488, M141001489, M141001490, M141001492, M141001494, M141001498, M141001503, M141001504, M141001507, M141001510, M141001511, M141001512, M141001513, M141001514, M141001521, M141001523, M141001525, M141001526, M141001547, M141001548, M141001550, M141001551, M141001554, M141001558, M141001560, M141001561, M141001563, M141001564, M141001566, M141001567, M141001568, M141001570, M141001577, M141001578, M141001579, M141001580, M141001581, M141001582, M141001583, M141001584, M141001585, M141001586, M141001587, M141001588, M141001589, M141001590, M141001591, M141001598, M141001600, M141001601, M141001602, M141001603, M141001606, M141001613, M141001617, M141001618, M141001619, M141001620, M141001622, M141001623, M141001625, M141001629, M141001631, M141001633, M141001634, M141001636, M141001639, M141001640, M141001643, M141001646, M141001649, M141001650, M141001652, M141001653, M141001655, M141001657, M141001658, M141001659, M141001660, M141001661, M141001663, M141001668, M141001671, M141001673, M141001676, M141101707, M141101709, M141101713, M141101716, M141101717, M141101718, M141101719, M141101720, M141101721, M141101722, M141101723, M141101725, M141101726, M141101729, M141101733, M141101734, M141101737, M141101740, M141101741, M141101742, M141101744, M141101745, M141101748, M141101749, M141101759, M141101761, M141101763, M141101764, M141201767, M141201768, M141201769, M141201776, M141201777, M141201781, M141201782, M141201783, M141201784, M141201785, M141201786, M141201789, M141201796, M141201797, M141201799, M141201801, M141201802, M141201805, M141201806, M141201807, M141201808, M141201810, M141201812, M141201813, M141201814, M141201815, M141201816, M141201817, M141201818, M141201820, M141201822, M141201823, M141201825, M141201826, M141201830, M141201832, M141201833, M141201834, M141201836, M141201839, M141201841, M141201843, M141201846, M141201907, M141201909, M141201911, M141201912, M141201913, M141201914, M141201915, M141201916, M141201917, M141201919, M141201920, M141201921, M141201922, M141201924, M141201926, M141201928, M141201930, M141201934, M141201936, M141201938, M141201940, M141201941, M141201945, M141201947, M141201948, M141201949, M141201950, M141201951, M141201952, M141201953, M141201954, M141201957, M141201960, M141201961, M141201962, M141201964, M141201966, M141201969, M141201970, M141201971, M141201973, M141201975, M141201976, M141201977, M141201978, M141201979, M141201980, M141201981, M141201982, M141201984, M141201986, M141201990, M141201991, M141201996, M141201997, M141201998, M141201999, M141202000, M141202001, M141202002, M141202003, M141202005, M141202006, M141202007, M141202009, M141202011, M141202014, M141202016, M141202017, M141202018, M141202019, M141202021, M141202023, M141202024, M141202026, M141202027, M141202028, M141202031, M141202032, M141202033, M141202034, M150102068, M150102070, M150102076, M150102077, M150102079, M150102080, M150102081, M150102082, M150102083, M150102088, M150102089, M150102090, M150102091, M150102092, M150102093, M150102094, M150102095, M150102096, M150102098, M150102100, M150102101, M150102103, M150102104, M150102105, M150102106, M150102107, M150102108, M150102115, M150102117, M150102119, M150102121, M150102122, M150102125, M150102126, M150102127, M150102131, M150102134, M150102135, M150102140, M150102142, M150102147, M150102150, M150102157, M150102159, M150102160, M150102161, M150102162, M150102165, M150102167, M150102168, M150102170, M150102171, M150102172, M150102173, M150102174, M150102175, M150102178, M150102179, M150102181, M150102189, M150102195, M150102196, M150102197, M150102200, M150102201, M150102202, M150102203, M150102204, M150102205, M150102206, M150102208, M150102209, M150102210, M150102212, M150102213, M150102217, M150102218, M150102221, M150102224, M150102225, M150102226, M150102227, M150102229, M150102234, M150102235, M150102269, M150102275, M150102276, M150102277, M150102279, M150102282, M150102283, M150102285, M150102286, M150102288, M150102289, M150102294, M150102295, M150102296, M150102297, M150102303, M150102305, M150102306, M150102309, M150102311, M150102313, M150102318, M150102319, M150102320, M150102321, M150102324, M150102325, M150102326, M150102327, M150102328, M150102329, M150102331, M150102333, M150102334, M150102335, M150102336, M150102338, M150102339, M150102341, M150102342, M150102343, M150102344, M150102345, M150102346, M150102348, M150102351, M150102352, M150102354, M150102355, M150102356, M150102357, M150102358, M150102359, M150102360, M150102362, M150102364, M150102365, M150102366, M150102367, M150102370, M150102374, M150102376, M150102378, M150102379, M150102382, M150102384, M150102386, M150102417, M150102418, M150102419, M150102420, M150102421, M150102422, M150102424, M150102425, M150202430, M150202437, M150202438, M150202439, M150202440, M150202441, M150202442, M150202444, M150202445, M150202448, M150202449, M150202450, M150202452, M150202453, M150202454, M150202456, M150202457, M150202458, M150202459, M150202460, M150202461, M150202464, M150202465, M150202466, M150202468, M150202469, M150202470, M150202471, M150202472, M150202473, M150202474, M150202475, M150202479, M150202480, M150202482, M150202484, M150202485, M150202486, M150202492, M150202493, M150202494, M150202496, M150202547, M150202548, M150202550, M150202551, M150202552, M150202553, M150202554, M150202563, M150202564, M150202566, M150202567, M150202571, M150202572, M150202573, M150202574, M150202575, M150202576, M150202578, M150202579, M150202581, M150202582, M150202589, M150202592, M150202593, M150202598, M150202599, M150202600, M150202601, M150202602, M150202604, M150202605, M150202606, M150202607, M150202608, M150202609, M150202614, M150202618, M150202619, M150202620, M150202621, M150202622, M150202623, M150202625, M150202628, M150202630, M150202631, M150202632, M150202633, M150202635, M150202636, M150202637, M150202639, M150202640, M150302795, M150302796, M150302797, M150302798, M150302801, M150302807, M150302808, M150302809, M150302810, M150302811, M150302813, M150302814, M150302816, M150302818, M150302819, M150302821, M150402827, M150402829, M150402830, M150402832, M150402833, M150402834, M150402836, M150402837, M150402838, M150402839, M150402840, M150402841, M150402842, M150402843, M150402844, M150402845, M150402846, M150402849, M150402850, M150402854, M150402855, M150402857, M150402858, M150402861, M150402862, M150402864, M150402865, M150402866, M150402867, M150402868, M150402869, M150402870, M150402871, M150402872, M150402873, M150402874, M150402876, M150402877, M150402880, M150402881, M150402882, M150402883, M150402884, M150402886, M150402888, M150402890, M150402891, M150402894, M150402895, M150402896, M150402897, M150402898, M150402899, M150402900, M150402903, M150402906, M150402947, M150402948, M150402949, M150402950, M150402951, M150402953, M150402954, M150402955, M150402959, M150402961, M150402963, M150402966, M150402967, M150402968, M150402970, M150402972, M150402973, M150402974, M150402975, M150402979, M150402983, M150402984, M150402985, M150402986, M150402987, M150402990, M150402991, M150402993, M150402995, M150402996, M150402997, M150402999, M150403001, M150403004, M150403005, M150403008, M150403012, M150403013, M150403014, M150403015, M150403016, M150403017, M150403018, M150403019, M150403021, M150403023, M150403024, M150403025, M150403026, M150403027, M150403028, M150403031, M150403032, M150403033, M150403035, M150403036, M150403038, M150403039, M150403040, M150403041, M150403044, M150403045, M150403069, M150403070, M150403072, M150403074, M150403076, M150403078, M150403080, M150403082, M150403083, M150403084, M150403086, M150403087, M150403090, M150403091, M150403092, M150403093, M150403094, M150403096, M150403098, M150403100, M150403101, M150403102, M150403103, M150403104, M150403105, M150403106, M150403107, M150403113, M150403115, M150403116, M150403117, M150403119, M150403120, M150403125, M150403129, M150403130, M150403131, M150403134, M150403135, M150403136, M150403137, M150403139, M150403141, M150403143, M150403144, M150403146, M150403148, M150403149, M150403150, M150403155, M150403157, M150403165, M150403166, M150403167, M150403169, M150403171, M150403174, M150403176, M150403177, M150403178, M150403179, M150403180, M150403182, M150403183, M150403184, M150403185, M150403198, M150403199, M150403200, M150403202, M150403206, M150403207, M150403210, M150403217, M150403219, M150403220, M150403221, M150403222, M150403224, M150403227, M150403228, M150403229, M150403230, M150403231, M150403232, M150403234, M150403235, M150403238, M150403239, M150403243, M150403244, M150403245, M150403246, M150503249, M150503250, M150503253, M150503254, M150503255, M150503256, M150503257, M150503258, M150503259, M150503260, M150503261, M150503262, M150503267, M150503269, M150503272, M150503273, M150503274, M150503276, M150503277, M150503279, M150503289, M150503291, M150503294, M150503296, M150503298, M150503300, M150503301, M150503303, M150503318, M150503319, M150503324, M150503325, M150503326, M150503332, M150503333, M150503336, M150503337, M150503344, M150503348, M150503350, M150503354, M150503356, M150503357, M150503358, M150503359, M150503360, M150503363, M150503364, M150503433, M150503435, M150503436, M150503437, M150503440, M150503441, M150503443, M150503444, M150503445, M150503446, M150503447, M150503457, M150503458, M150503459, M150503460, M150503461, M150503462, M150503465, M150503467, M150503468, M150503469, M150503472, M150503473, M150503475, M150503477, M150503479, M150503481, M150503482, M150503483, M150503485, M150503486, M150503487, M150503488, M150503489, M150503490, M150503493, M150503495, M150503497, M150503499, M150503500, M150503501, M150503506, M150503517, M150503518, M150503519, M150503520, M150503524, M150503525, M150503526, M150503527, M150503531, M150503532, M150503533, M150503537, M150503538, M150503539, M150503542, M150503545, M150503546, M150503548, M150503550, M150503551, M150503553, M150503554, M150503555, M150503556, M150503557, M150503562, M150503564, M150503567, M150503568, M150503569, M150503570, M150503571, M150503573, M150503575, M150503577, M150503581, M150503582, M150503583, M150503585, M150503586, M150503587, M150503590, M150503592, M150503593, M150503595, M150503606, M150503607, M150503608, M150503609, M150503613, M150503616, M150503617, M150503618, M150503622, M150503623, M150503624, M150503625, M150503628, M150503631, M150503633, M150503634, M150503637, M150503638, M150503641, M150503642, M150503643, M150503647, M150503650, M150503651, M150503652, M150503653, M150503654, M150503655, M150503656, M150503657, M150503659, M150503660, M150503664, M150503665, M150503666, M150503667, M150503671, M150503672, M150503674, M150503676, M150503677, M150503678, M150503679, M150503683, M150503686, M150503687, M150503689, M150503690, M150503691, M150503693, M150503695, M150503696, M150503697, M150503698, M150503699, M150503702, M150503704, M150503705, M150503706, M150503708, M150503715, M150503717, M150503718, M150503721, M150503722, M150503723, M150503724, M150503725, M150503730, M150503733, M150503745, M150503746, M150503747, M150503749, M150503750, M150503752, M150503753, M150503756, M150503758, M150503759, M150503760, M150503761, M150503764, M150503765, M150503767, M150503768, M150503769, M150503770, M150503771, M150503772, M150503774, M150503775, M150503776, M150503777, M150503778, M150503779, M150503780, M150503782, M150503785, M150503787, M150503789, M150503790, M150503791, M150503792, M150503793, M150503795, M150503797, M150503800, M150503802, M150503811, M150503812, M150503813, M150503815, M150503816, M150503817, M150503820, M150503821, M150503822, M150503823, M150503824, M150503825, M150503828, M150503832, M150503835, M150603837, M150603838, M150603839, M150603841, M150603842, M150603843, M150603845, M150603846, M150603847, M150603848, M150603850, M150603851, M150603852, M150603855, M150603856, M150603857, M150603858, M150603862, M150603864, M150603865, M150603866, M150603867, M150603869, M150603871, M150603872, M150603873, M150603875, M150603876, M150603887, M150603888, M150603893, M150603894, M150603899, M150603900, M150603902, M150603906, M150603908, M150603910, M150603911, M150603913, M150603916, M150603917, M150603919, M150603920, M150603923, M150603924, M150603926, M150603928, M150603929, M150603930, M150603931, M150603936, M150603938, M150603939, M150603940, M150603941, M150603942, M150603948, M150603949, M150603950, M150603955, M150603957, M150603958, M150603960, M150603961, M150603962, M150603963, M150603964, M150603966, M150603973, M150603976, M150603978, M150603979, M150603981, M150603987, M150603990, M150603991, M150603992, M150603993, M150603994, M150603995, M150603996, M150604001, M150604004, M150604005, M150604006, M150604007, M150604008, M150604010, M150604011, M150604013, M150604014, M150604016, M150604017, M150604021, M150604022, M150604023, M150604024, M150604025, M150604029, M150604030, M150604031, M150604034, M150604036, M150604037, M150604038, M150604039, M150604047, M150604049, M150604051, M150604054, M150604055, M150604056, M150604057, M150604058, M150604060, M150604064, M150604065, M150604068, M150604073, M150604076, M150604077, M150604081, M150604082, M150604084, M150604085, M150604086, M150604090, M150604091, M150604093, M150604094, M150604099, M150604101, M150604102, M150604103, M150604105, M150604106, M150604108, M150604113, M150604114, M150604115, M150604116, M150604117, M150604119, M150604121, M150604123, M150604124, M150604127, M150604128, M150604129, M150604131, M150604132, M150604133, M150604134, M150604136, M150604137, M150604140, M150604144, M150604146, M150604152, M150604155, M150604156, M150604157, M150604163, M150604164, M150604165, M150604167, M150604169, M150604170, M150604172, M150604173, M150604174, M150604175, M150604176, M150604177, M150604178, M150604180, M150604182, M150604184, M150604185, M150604187, M150604188, M150604190, M150604191, M150604192, M150604193, M150604194, M150604195, M150604199, M150604201, M150604202, M150604204, M150604205, M150604207, M150604208, M150604210, M150604212, M150604213, M150604219, M150604221, M150604224, M150604225, M150604226, M150604229, M150604230, M150604231, M150604232, M150604233, M150604237, M150604238, M150604239, M150604242, M150604243, M150604246, M150604253, M150604255, M150604256, M150604257, M150604260, M150604263, M150604264, M150604265, M150604266, M150604269, M150604270, M150604271, M150604272, M150604274, M150604275, M150604277, M150604278, M150604282, M150604283, M150604284, M150604285, M150604289, M150604290, M150604292, M150604295, M150604299, M150604300, M150604301, M150604302, M150604303, M150604306, M150604307, M150604308, M150604310, M150604311, M150604314, M150604315, M150604316, M150604318, M150604319, M150604321, M150604323, M150604324, M150604325, M150604326, M150604327, M150604329, M150604330, M150604331, M150604333, M150604334, M150604335, M150604338, M150604340, M150604341, M150604346, M150604348, M150604351, M150604352, M150604354, M150604356, M150604358, M150604360, M150604362, M150604365, M150604371, M150604376, M150704427, M150704428, M150704429, M150704431, M150704433, M150704434, M150704435, M150704439, M150704440, M150704441, M150704446, M150704449, M150704451, M150704452, M150704455, M150704456, M150704458, M150704460, M150704463, M150704465, M150704467, M150704470, M150704471, M150704472, M150704477, M150704479, M150704480, M150704481, M150704482, M150704483, M150704485, M150704486, M150704487, M150704488, M150704491, M150704492, M150704493, M150704495, M150704496, M150704500, M150704503, M150704504, M150704505, M150704509, M150704510, M150704512, M150704513, M150704514, M150704515, M150704516, M150704517, M150704518, M150704519, M150704520, M150704525, M150704527, M150704528, M150704529, M150704530, M150704531, M150704532, M150704533, M150704534, M150704536, M150704539, M150704540, M150704542, M150704543, M150704544, M150704546, M150704547, M150704548, M150704553, M150704555, M150704556, M150704557, M150704561, M150704563, M150704565, M150704569, M150704570, M150704572, M150704580, M150704581, M150704583, M150704586, M150704587, M150704590, M150704591, M150704593, M150704594, M150704600, M150704601, M150704602, M150704603, M150704605, M150704606, M150704613, M150704614, M150704615, M150704620, M150704622, M150704623, M150704625, M150704626, M150704627, M150704632, M150704634, M150704635, M150704638, M150704639, M150704640, M150704641, M150704642, M150704643, M150704644, M150704646, M150704647, M150704648, M150704651, M150704652, M150704653, M150704654, M150704655, M150704656, M150704657, M150704658, M150704659, M150704660, M150704661, M150704663, M150704664, M150704665, M150704666, M150704729, M150704732, M150704733, M150704734, M150704736, M150704737, M150704738, M150704739, M150704743, M150704747, M150704748, M150704751, M150704752, M150704755, M150704761, M150704764, M150704768, M150704769, M150704770, M150704771, M150704772, M150704773, M150704774, M150704776, M150704777, M150704778, M150704780, M150704781, M150704782, M150704785, M150704786, M150704787, M150704788, M150704789, M150704791, M150704792, M150704793, M150704794, M150704795, M150704796, M150704797, M150704798, M150704799, M150704800, M150704801, M150704802, M150704806, M150704807, M150704811, M150704812, M150704813, M150704815, M150704818, M150704824, M150704826, M150704827, M150704829, M150704831, M150704835, M150804837, M150804838, M150804839, M150804841, M150804843, M150804845, M150804846, M150804850, M150804852, M150804854, M150804856, M150804858, M150804859, M150804862, M150804864, M150804870, M150804874, M150804875, M150804877, M150804881, M150804882, M150804883, M150804885, M150804886, M150804887, M150804890, M150804892, M150804957, M150804958, M150804961, M150804962, M150804963, M150804965, M150804966, M150804967, M150804968, M150804973, M150804974, M150804976, M150804983, M150804984, M150804988, M150804989, M150804990, M150804993, M150804994, M150804995, M150804996, M150804997, M150805001, M150805002, M150805003, M150805004, M150805005, M150805008, M150805009, M150805011, M150805012, M150805013, M150805015, M150805016, M150805030, M150805037, M150805039, M150805045, M150805047, M150805051, M150805055, M150805056, M150805060, M150805071, M150805072, M150805074, M150805076, M150805077, M150805078, M150805083, M150805086, M150805088, M150805094, M150805097, M150805098, M150805099, M150805100, M150805101, M150805102, M150805103, M150805105, M150805108, M150805109, M150805110, M150805111, M150805113, M150805114, M150805115, M150805116, M150805119, M150805120, M150805121, M150805123, M150805125, M150805127, M150805129, M150805130, M150805133, M150805135, M150805136, M150805169, M150805170, M150805171, M150805173, M150805175, M150805176, M150805177, M150805178, M150805179, M150805180, M150805182, M150805183, M150805184, M150805186, M150805187, M150805189, M150805190, M150805191, M150805192, M150805195, M150805198, M150805201, M150805202, M150805203, M150805204, M150805205, M150805206, M150805207, M150805212, M150805213, M150805217, M150805221, M150805222, M150805223, M150805226, M150905227, M150905228, M150905229, M150905230, M150905233, M150905234, M150905235, M150905236, M150905237, M150905238, M150905241, M150905242, M150905243, M150905244, M150905245, M150905247, M150905248, M150905253, M150905257, M150905259, M150905260, M150905261, M150905262, M150905263, M150905265, M150905266, M150905269, M150905272, M150905273, M150905274, M150905275, M150905277, M150905279, M150905280, M150905282, M150905283, M150905286, M150905288, M150905290, M150905297, M150905299, M150905301, M150905305, M150905306, M150905311, M150905315, M150905323, M150905326, M150905328, M150905329, M150905330, M150905331, M150905332, M150905335, M150905337, M150905338, M150905340, M150905341, M150905342, M150905343, M150905344, M150905345, M150905351, M150905353, M150905354, M150905355, M150905357, M150905358, M150905359, M150905360, M150905362, M150905364, M150905367, M150905369, M150905370, M150905371, M150905376, M150905377, M150905379, M150905380, M150905381, M150905383, M150905384, M150905386, M150905387, M150905388, M150905389, M150905390, M150905396, M150905397, M150905399, M150905402, M150905403, M150905406, M150905438, M150905439, M150905443, M150905444, M150905445, M150905448, M150905450, M150905454, M150905455, M150905456, M150905459, M150905460, M150905465, M150905466, M150905467, M150905468, M150905471, M150905473, M150905474, M150905477, M150905478, M150905479, M150905481, M150905482, M150905483, M150905485, M150905492, M150905495, M150905496, M150905500, M150905505, M150905512, M150905513, M150905515, M150905518, M150905520, M150905521, M150905525, M150905526, M150905527, M150905529, M150905533, M150905535, M150905536, M150905537, M150905540, M150905551, M150905552, M150905554, M150905555, M150905556, M150905562, M150905564, M150905566, M150905567, M150905569, M150905570, M150905571, M150905572, M150905573, M150905574, M150905576, M150905579, M150905580, M150905581, M150905585, M150905586, M150905591, M150905593, M150905594, M150905598, M150905603, M150905604, M150905605, M150905614, M150905617, M150905618, M150905625, M150905627, M150905629, M150905631, M150905634, M150905639, M150905641, M150905642, M150905644, M150905646, M150905648, M150905650, M150905652, M150905656, M150905658, M150905659, M150905661, M150905664, M150905665, M150905670, M150905673, M150905676, M151005749, M151005752, M151005757, M151005758, M151005760, M151005764, M151005766, M151005769, M151005778, M151005779, M151005781, M151005782, M151005783, M151005784, M151005785, M151005786, M151005787, M151005788, M151005790, M151005791, M151005792, M151005794, M151005796, M151005799, M151005800, M151005802, M151005804, M151005807, M151005811, M151005813, M151005814, M151005820, M151005824, M151005825, M151005826, M151005827, M151005828, M151005829, M151005831, M151005833, M151005834, M151005836, M151005837, M151005838, M151005839, M151005840, M151005846, M151005847, M151005853, M151005855, M151005856, M151005857, M151005858, M151005859, M151005860, M151005863, M151005864, M151005865, M151005866, M151005867, M151005868, M151005873, M151005877, M151005878, M151005879, M151005881, M151005882, M151005883, M151005884, M151005886, M151005888, M151005889, M151005890, M151005892, M151005894, M151005898, M151005899, M151005900, M151005904, M151005905, M151005906, M151005910, M151005912, M151005915, M151005916, M151005920, M151005922, M151005923, M151005926, M151005927, M151005928, M151005931, M151005933, M151005934, M151005935, M151005937, M151005939, M151005940, M151005945, M151005947, M151005948, M151005949, M151005950, M151005953, M151005955, M151005957, M151005958, M151005960, M151005963, M151005966, M151005967, M151005969, M151005970, M151005973, M151005974, M151005975, M151005976, M151006005, M151006011, M151006013, M151006014, M151006018, M151006019, M151006020, M151006021, M151006022, M151006026, M151006028, M151006029, M151006036, M151006046, M151006049, M151006051, M151006052, M151006062, M151006064, M151006065, M151006066, M151006067, M151006070, M151006071, M151006075, M151006076, M151006078, M151006079, M151006083, M151006085, M151006091, M151006094, M151006096, M151006097, M151006099, M151006100, M151006101, M151006102, M151006106, M151006108, M151006110, M151006113, M151006115, M151006116, M151106117, M151106122, M151106123, M151106125, M151106126, M151106127, M151106128, M151106129, M151106130, M151106131, M151106134, M151106135, M151106136, M151106145, M151106147, M151106148, M151106149, M151106150, M151106151, M151106152, M151106153, M151106154, M151106158, M151106162, M151106165, M151106167, M151106168, M151106180, M151106181, M151106182, M151106183, M151106185, M151106188, M151106190, M151106193, M151106194, M151106202, M151106203, M151106204, M151106205, M151106207, M151106209, M151106211, M151106212, M151106215, M151106217, M151106218, M151106219, M151106220, M151106221, M151106222, M151106223, M151106224, M151106225, M151106228, M151106231, M151106232, M151106233, M151106234, M151106235, M151106236, M151106238, M151106242, M151106244, M151106245, M151106249, M151106250, M151106251, M151106252, M151106254, M151106255, M151106256, M151106258, M151106259, M151106260, M151106264, M151106272, M151106275, M151106276, M151106282, M151106284, M151106288, M151106289, M151106290, M151106291, M151106292, M151106295, M151106296, M151106298, M151106307, M151106308, M151106310, M151106312, M151106313, M151106314, M151106315, M151106319, M151106320, M151106321, M151106322, M151106323, M151106324, M151106326, M151106327, M151106328, M151106329, M151106331, M151106333, M151106334, M151106335, M151106336, M151106337, M151106339, M151106340, M151106350, M151106351, M151106354, M151106356, M151106409, M151106411, M151106412, M151106414, M151106417, M151106418, M151106420, M151106424, M151106426, M151106429, M151106431, M151106432, M151106433, M151106435, M151106443, M151106446, M151106447, M151106448, M151106449, M151106456, M151106458, M151106459, M151106460, M151106461, M151106462, M151106463, M151106464, M151106465, M151206468, M151206472, M151206473, M151206474, M151206475, M151206477, M151206480, M151206481, M151206482, M151206485, M151206486, M151206487, M151206488, M151206489, M151206490, M151206491, M151206494, M151206495, M151206496, M151206497, M151206498, M151206499, M151206502, M151206507, M151206508, M151206510, M151206512, M151206515, M151206521, M151206523, M151206570, M151206572, M151206579, M151206583, M151206585, M151206586, M151206587, M151206591, M151206593, M151206595, M151206597, M151206599, M151206600, M151206602, M151206603, M151206606, M151206607, M151206609, M151206610, M151206612, M151206613, M151206614, M151206616, M151206618, M151206622, M151206623, M151206625, M160106667, M160106669, M160106673, M160106674, M160106676, M160106678, M160106681, M160106685, M160106689, M160106691, M160106692, M160106693, M160106695, M160106696, M160106697, M160106699, M160106700, M160106701, M160106709, M160106713, M160106714, M160106716, M160106721, M160106723, M160106724, M160106725, M160106726, M160106727, M160106728, M160106729, M160106732, M160106734, M160106735, M160106736, M160106739, M160106741, M160106746, M160106751, M160106759, M160106760, M160106765, M160106766, M160106771, M160106772, M160106773, M160106776, M160106778, M160106779, M160106780, M160106782, M160106783, M160206789, M160206790, M160206791, M160206794, M160206799, M160206803, M160206808, M160206814, M160206816, M160206821, M160206824, M160206825, M160206826, M160206835, M160206836, M160206837, M160206838, M160206840, M160206843, M160206909, M160206910, M160206912, M160206914, M160206915, M160206919, M160206923, M160206926, M160206927, M160206928, M160206932, M160206933, M160206934, M160206937, M160206940, M160206946, M160206952, M160206958, M160206959, M160206961, M160206968, M160206970, M160206976, M160206978, M160206982, M160206983, M160206984, M160206985, M160206987, M160206990, M160206997, M160206999, M160207004, M160207009, M160207010, M160207011, M160207013, M160207016, M160207017, M160207018, M160207019, M160207020, M160207022, M160207023, M160207027, M160207028, M160207032, M160207033, M160207039, M160207045, M160207047, M160207048, M160207049, M160207051, M160207052, M160207054, M160207059, M160207060, M160207063, M160207064, M160207065, M160207067, M160207073, M160207074, M160207086, M160307092, M160307096, M160307097, M160307102, M160307106, M160307112, M160307129, M160307131, M160307132, M160307136, M160307139, M160307146, M160307147, M160307148, M160307150, M160307153, M160307154, M160307155, M160307157, M160307158, M160307159, M160307160, M160307163, M160307165, M160307166, M160307171, M160307172, M160307175, M160307176, M160307177, M160307178, M160307181, M160307186, M160307187, M160307200, M160307240, M160307241, M160307243, M160307244, M160307247, M160307248, M160307249, M160307252, M160307256, M160307258, M160307259, M160407268, M160407269, M160407271, M160407272, M160407273, M160407275, M160407289, M160407291, M160407292, M160407294, M160407295, M160407300, M160407301, M160407302, M160407304, M160407305, M160407308, M160407309, M160407310, M160407312, M160407314, M160407315, M160407317, M160407318, M160407320, M160407322, M160407324, M160407325, M160407333, M160407335, M160407336, M160407338, M160407339, M160407343, M160407345, M160407346, M160407347, M160407348, M160407350, M160407351, M160407352, M160407353, M160407354, M160407355, M160407358, M160407365, M160407367, M160407368, M160407370, M160407373, M160407377, M160407378, M160407379, M160407380, M160407381, M160407382, M160407384, M160407385, M160407387, M160407388, M160407390, M160407391, M160407393, M160407394, M160407401, M160407406, M160407408, M160407410, M160407411, M160407413, M160407419, M160407426, M160407428, M160407431, M160407434, M160407435, M160407438, M160407440, M160407444, M160407445, M160407446, M160507448, M160507449, M160507453, M160507455, M160507462, M160507490, M160507512, M160507519, M160507527, M160507529, M160507530, M160507531, M160507533, M160507534, M160507536, M160607540, M160607543, M160607547, M160607550, M160607552, M160607553, M160607555, M160607556, M160607559, M160607561, M160607562, M160607566, M160607576, M160707582, M160707611, M160707614, M160707616, M160807654, M160807657, M160807658, M160807659, M160807661, M160807662, M160807665, M160807737, M160807740, M160807760, M160907795, M160907796, M160907807, M160907810, M160907820, FP-900056, M111201, M131200026, M140901218, M140901226, M140901228, M141001509, M141001611, M141001612, M141001614, M141001616, M141001638, M150102109, M150102139, M150102145, M150302793, M150302794, M150503278, M150503280, M150503282, M150503283, M150503285, M150503286, M150503287, M150503288, M150503307, M150503308, M150503311, M150503316, M150503714, M150603879, M150604288, M150704784, M150805019, M150805020, M150805021, M150805022, M150805023, M150805024, M150805026, M150805027, M150805028, M150805029, M150805032, M150805035, M150805036, M160106742, M160307119, M160307122, M160307124, M160307125, M160307126, M160307188, M160407267, M160507468, M160507469, M160507470, M160507471, M160507474, M160507478, M160507479, M160507481, M160507482, M160507483, M160507485, M160507486, M160507487, M160507488, M160507491, M160507492, M160507493, M160507494, M160507495, M160507496, M160507497, M160507498, M160507499, M160507500, M160507502, M160507503, M160507504, M160507506, M160507507, M160507509, M160507510, M160507515, M160507516, M160507518, M160507520, M160507522, M160507523, M160507524, M160507525, M160507526, N021703 |
| FEI Number |
3004936110
|
Recalling Firm/ Manufacturer |
St. Jude Medical 387 Technology Cir Nw Ste 500 Atlanta GA 30313-2424
|
| For Additional Information Contact | Mr. Justin Paquette 651-756-6293 |
Manufacturer Reason for Recall | Radiated Radiofrequency (RF) emissions of the
CardioMEMS Patient Electronics Systems (PES) Models CM1000, CM1010 and CM1100 and CardioMEMS Hospital Electronics Systems (HES) Model CM3000 exceed certain emissions limits and have the potential to cause interference with other medical devices, such as implantable neurostimulators, pacemakers, cardiac defibrillators, continuous glucose monitors, or other bio-wearable sensors when in close proximity to an active Hospital or Patient Electronics System (i.e., during readings). |
FDA Determined Cause 2 | Device Design |
| Action | Abbott notified consignees on about 02/07/2023 via hand-delivery of the letter within the US. The letter instructed consignees to continue to follow CardioMEMS Instructions for Use (IFU). Abbott will notify patients of the emissions issue. If you suspect a CardioMEMS Electronics System has interfered with another medical device during use, report the event to Abbott Remote Care Technical Support at 1-844-MYCMEMS (692-6367) (US). Customers were also instructed to complete and return the acknowledgement and receipt form.
Abbott notified patients on about 02/07/2023 via mailed letter. Customers were instructed that they could continue use of their CardioMEMS System and that their doctor has been notified of the issue as well. If there is suspicion that interference may be occurring during a sensor reading or any other quality issues, please report it to: Abbott Technical Support at 1-844-MYCMEMS (692-6367) (US); Monday-Friday: 8am-8pm EST |
| Quantity in Commerce | 3,177 units |
| Distribution | Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Canada, Colombia, Czech Republic, Denmark, Finland, France, Germany, Ireland, Italy, Kuwait, Netherlands, Portugal, Saudi Arabia, Spain, Sweden, Switzerland, UK, United Arab Emirates.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| PMA Database | PMAs with Product Code = MOM
|
|
|
|