|
Class 2 Device Recall pro mattress |
![see related information](../images/related-info.gif) |
Date Initiated by Firm |
March 31, 2023 |
Date Posted |
May 01, 2023 |
Recall Status1 |
Open3, Classified |
Recall Number |
Z-1482-2023 |
Recall Event ID |
91967 |
Product Classification |
Bed, flotation therapy, powered - Product Code IOQ
|
Product |
Hill-Rom pro+ mattress, product codes (REF): a) 20877002S, b) 20877003S, c) 20877004S, d) 20877007S, e) 20877008S, f) 21649301S, g) 21649401S, h) 21649402S, i) P006800A01, j) P006800A02, k) P3255A01, l) P3255A02, m) P3255A03, n) P3255ARENT01, o) P7923A01, p) P7923A02, q) P7923A03, r) P7923A04, s) P7924A01, t) P7924A02, u) P7924A03, v) P7924A04, w) P7924ARENT01 |
Code Information |
a) 20877002S, UDI/DI N/A, Serial Numbers: W315BP5905, W315BP5907, W315BP5910, W315BP5911;
b) 20877003S, UDI/DI N/A, Serial Numbers: W301BP5612, W301BP5615;
c) 20877004S, UDI/DI N/A, Serial Numbers: W231BP4205, W238BP4341, W238BP4342, W238BP4343, W238BP4345, W238BP4347, W238BP4348, W238BP4349, W246BP4519, W246BP4525, W246BP4526, W245BP4502, W245BP4503, W246BP4510, W246BP4511, W246BP4513, W246BP4514, W246BP4520, W246BP4521, W246BP4523, W246BP4524, W252BP4622, W252BP4629, W252BP4634, W252BP4636, W246BP4505, W246BP4512, W252BP4612, W252BP4618, W252BP4620, W256BP4671, W256BP4673, W287BP5300, W287BP5301, W287BP5302, W322BP6072, W322BP6076,;
d) 20877007S, UDI/DI N/A, Serial Numbers: W299BP5609, W299BP5611;
e) 20877008S, UDI/DI N/A, Serial Numbers: W301BP5616, W301BP5621;
f) 21649301S, UDI/DI N/A, Serial Numbers: W188TP7991, W188TP7992, W188TP7993, W188TP7994, W188TP7995, W188TP7996, W188TP7997, W188TP7998;
g) 21649401S, UDI/DI N/A, Serial Numbers: W188TP8006, W188TP8007, W188TP8008, W188TP8009, W188TP8010, W188TP8011;
h) 21649402S, UDI/DI N/A, Serial Numbers: W188TP7999, W188TP8000, W188TP8002, W188TP8003, W188TP8004, W188TP8005;
i) P006800A01, UDI/DI 00887761977846, Serial Numbers: V314BP0017, V328BP0145, V328BP0146, V328BP0147, V328BP0148, V328BP0149, V328BP0150, V328BP0151, V328BP0152, V328BP0153, V328BP0154, V328BP0156, V328BP0159, V328BP0155, V328BP0157, V328BP0158, V330BP0160, V330BP0161, V330BP0163, V330BP0166, V330BP0167;
j) P006800A02, UDI/DI 00887761977839, Serial Numbers: V314BP0023, V314BP0027;
k) P3255A01, UDI/DI 00887761977860, Serial Numbers: V301BP0010, V301BP0011, V301BP0012, V314BP0018, V314BP0021, V314BP0026, V314BP0019, V314BP0032, V314BP0020, V314BP0029, V314BP0030, V314BP0031, V314BP0034, V314BP0036, V323BP0080, V325BP0109, V326BP0120, W012BP0498, W036BP0623, W036BP0624, W036BP0625, W036BP0626, W036BP0627, W036BP0628, W036BP0629, W036BP0630, W036BP0631, W036BP0633, W036BP0634, W036BP0635, W036BP0636, W036BP0637, W036BP0638, W036BP0639, W036BP0640, W036BP0641, W039BP0643, W039BP0644, W137BP2190, W137BP2191, W137BP2192, W137BP2193, W137BP2194, W137BP2195, W169BP3029, W169BP3030, W172BP3033, W172BP3034, W172BP3035, W180BP3280, W180BP3281, W183BP3371, W183BP3373, W183BP3374, W183BP3375, W183BP3376, W183BP3377, W183BP3381, W183BP3384, W186BP3390, W186BP3391, W186BP3392, W186BP3393, W186BP3394, W186BP3395, W186BP3396, W186BP3399, W186BP3400, W186BP3401, W186BP3402, W169BP3008, W183BP3288, W183BP3289, W183BP3290, W183BP3291, W183BP3292, W183BP3293, W183BP3297, W183BP3298, W183BP3299, W183BP3300, W183BP3302, W183BP3303, W183BP3304, W183BP3305, W183BP3306, W183BP3307, W183BP3331, W183BP3333, W183BP3334, W183BP3335, W183BP3336, W183BP3337, W183BP3338, W183BP3339, W183BP3342, W183BP3343, W145BP2443, W145BP2445, W145BP2446, W154BP2654, W154BP2655, W179BP3172, W210BP3863, W210BP3864, W210BP3865, W210BP3866, W210BP3867, W210BP3871, W210BP3872, W210BP3878, W210BP3879, W210BP3881, W210BP3882, W210BP3883, W210BP3886, W210BP3839, W210BP3840, W210BP3857, W210BP3858, W210BP3859, W210BP3861, W210BP3880, W210BP3888, W210BP3890, W210BP3842, W210BP3851, W210BP3860, W214BP3896, W214BP3901, W214BP3909, W214BP3910, W214BP3911, W214BP3912, W214BP3914, W214BP3915, W214BP3916, W214BP3917, W214BP3918, W214BP3919, W210BP3846, W210BP3852, W214BP3894, W214BP3895, W214BP3897, W214BP3898, W214BP3899, W214BP3900, W214BP3902, W214BP3903, W214BP3904, W214BP3905, W214BP3907, W214BP3908, W214BP3920, W214BP3921, W214BP3922, W214BP3953, W214BP3954, W204BP3757, W214BP3906, W214BP3913, W214BP3923, W214BP3925, W214BP3926, W214BP3927, W214BP3928, W214BP3929, W214BP3930, W214BP3932, W214BP3933, W214BP3934, W214BP3935, W215BP3989, W215BP3990, W215BP3991, W215BP3992, W215BP3993, W215BP3994, W215BP3995, W215BP3996, W215BP4001, W215BP4002, W215BP4003, W215BP4005, W215BP4006, W215BP4009, W215BP4011, W190BP3471, W190BP3473, W190BP3476, W190BP3477, W190BP3478, W190BP3479, W190BP3484, W190BP3512, W190BP3514, W190BP3516, W190BP3521, W194BP3544, W194BP3545, W196BP3597, W196BP3603, W196BP3605, W196BP3611, W196BP3612, W196BP3613, W196BP3614, W196BP3615, W196BP3616, W196BP3617, W196BP3618, W196BP3619, W196BP3621, W196BP3622, W196BP3623, W196BP3595, W196BP3596, W196BP3598, w196bp3599, W196BP3600, W196BP3601, W196BP3602, W196BP3604, W196BP3606, W196BP3607, W196BP3608, W196BP3610, W217BP4014, W217BP4015, W217BP4016, W217BP4017, W217BP4018, W217BP4019, W217BP4020, W217BP4021, W217BP4022, W217BP4023, W217BP4024, W217BP4025, W217BP4026, W217BP4027, W217BP4028, W217BP4030, W235BP4261, W243BP4457, W243BP4458, W243BP4459, W243BP4462, W243BP4463, W243BP4464, W243BP4465, W243BP4436, W243BP4437, W243BP4438, W243BP4439, W243BP4441, W243BP4442, W243BP4445, W243BP4446, W243BP4447, W243BP4448, W243BP4449, W243BP4451, W243BP4454, W243BP4460, W243BP4461, W243BP4466, W243BP4467, W243BP4468, W243BP4470, W243BP4471, W243BP4472, W243BP4477, W243BP4478, W245BP4501, W246BP4504, W246BP4518, W246BP4528, W246BP4529, W246BP4530, W246BP4531, W246BP4532, W246BP4533, W246BP4534, W246BP4535, W246BP4537, W246BP4538, W246BP4540, W246BP4541, W246BP4542, W246BP4543, W246BP4544, W246BP4545, W246BP4547, W246BP4548, W246BP4552, W246BP4553, W246BP4554, W246BP4555, W246BP4556, W246BP4557, W246BP4558, W246BP4559, W246BP4560, W246BP4567, W246BP4569, W246BP4570, W246BP4571, W223BP4079, W246BP4536, W246BP4550, W246BP4561, W246BP4563, W246BP4564, W246BP4565, W246BP4566, W246BP4568, W260BP4750, W260BP4751, W260BP4752, W260BP4756, W260BP4757, W260BP4758, W260BP4759, W260BP4761, W260BP4762, W260BP4763, W260BP4764, W260BP4765, W260BP4766, W260BP4767, W260BP4787, W260BP4775, W260BP4776, W260BP4777, W260BP4785, W260BP4786, W260BP4793, W260BP4794, W260TP0946, W273BP5000, W273BP5001, W273BP5002, W273BP5003, W274BP5014, W274BP5015, W291BP5346, W307BP5825, W307BP5826, W302BP5685, W306BP5747, W306BP5749, W306BP5752, W306BP5753, W306BP5754, W306BP5755, W306BP5756, W306BP5757, W307BP5775, W307BP5776, W307BP5811, W307BP5815, W307BP5780, W307BP5781, W307BP5821, W307BP5822, W313BP5831, W313BP5838, W313BP5839, W320BP5999, W320BP6000, W320BP6003, W320BP6004, W320BP6005, W320BP6006, W320BP6007, W320BP6009, W320BP6011, W320BP6013, W320BP6015, W320BP6016, W320BP6018, W316BP5984, W320BP6019, W320BP6020, W320BP6021, W320BP6022, W320BP6024, W320BP6026, W320BP6027, W320BP6028, W320BP6031, W320BP6032, W320BP6033, W320BP6034, W320BP6037, W320BP6038, W320BP6039, W320BP6041, W320BP6042, W320BP6043, W323BP6078, W323BP6079, W323BP6102, W323BP6103, W323BP6109, W323BP6110, W323BP6111, W320BP6045;
l) P3255A02, UDI/DI 00887761977853, Serial Numbers: V314BP0038, V314BP0039, V314BP0040, V314BP0022, V314BP0024, V314BP0025, V314BP0033, V314BP0035, V314BP0037, V322BP0063, V322BP0064, V322BP0065, V323BP0072, V323BP0073, V323BP0074, V323BP0081, V324BP0098, V343BP0203, V343BP0205, V330BP0164, V330BP0165, V342BP0196, V343BP0204, V343BP0206, V343BP0207, V343BP0224, V343BP0225, V343BP0226, V343BP0230, V343BP0232, V344BP0251, V350BP0280, V351BP0335, V351BP0336, V351BP0337, V351BP0338, V344BP0250, V349BP0257, V349BP0258, V349BP0259, V350BP0303, V350BP0304, V350BP0305, V350BP0306, V350BP0307, V351BP0317, V351BP0320, V351BP0321, V351BP0322, V350BP0294, V353BP0341, V353BP0342, V353BP0343, V353BP0344, V353BP0345, V353BP0346, V353BP0347, V353BP0348, V353BP0349, V353BP0350, V353BP0351, V353BP0352, V353BP0353, V353BP0354, V353BP0355, V353BP0356, V350BP0293, V351BP0308, V351BP0309, V351BP0310, V351BP0311, V351BP0312, V351BP0313, V351BP0314, V351BP0315, V351BP0316, V351BP0339, V353BP0358, V353BP0359, V353BP0360, V353BP0362, V353BP0363, V353BP0364, V353BP0365, V353BP0366, V353BP0368, V356BP0434, V356BP0435, V356BP0436, V356BP0437, V356BP0438, V356BP0439, V356BP0440, V356BP0441, V356BP0442, V356BP0443, V356BP0444, V356BP0445, V356BP0446, V356BP0447, V356BP0448, V356BP0449, V356BP0450, V356BP0452, V356BP0453, V356BP0479, V356BP0395, V356BP0406, W039BP0645, W039BP0646, W039BP0647, W039BP0648, W039BP0649, W039BP0650, W039BP0651, W039BP0652, W039BP0653, W014BP0510, W028BP0619, W043BP0699, W043BP0704, W043BP0705, W043BP0706, W043BP0707, W043BP0708, W043BP0719, W042BP0688, W042BP0689, W042BP0690, W042BP0691, W042BP0692, W042BP0693, W042BP0694, W042BP0695, W042BP0696, W042BP0697, W043BP0700, W043BP0703, W043BP0711, W043BP0712, W043BP0709, W043BP0710, W043BP0713, W043BP0723, W043BP0724, W043BP0725, W043BP0726, W043BP0727, W043BP0728, W043BP0729, W043BP0730, W043BP0731, W043BP0732, W043BP0733, W043BP0734, W043BP0735, W043BP0736, W043BP0737, W043BP0742, W043BP0738, W043BP0739, W043BP0740, W043BP0741, W046BP0744, W046BP0745, W046BP0746, W046BP0747, W046BP0748, W046BP0749, W046BP0750, W046BP0751, W046BP0752, W046BP0753, W046BP0754, W046BP0755, W046BP0756, W046BP0757, W046BP0758, W046BP0759, W046BP0760, W046BP0761, W046BP0762, W046BP0763, W046BP0765, W046BP0766, W046BP0767, W046BP0768, W046BP0769, W046BP0770, W046BP0771, W046BP0772, W046BP0773, W046BP0774, W050BP0794, W050BP0795, W050BP0796, W050BP0797, W050BP0798, W050BP0799, W050BP0800, W050BP0801, W050BP0802, W050BP0805, W050BP0806, W050BP0803, W050BP0804, W050BP0807, W050BP0809, W050BP0810, W050BP0811, W050BP0812, W050BP0813, W050BP0814, W050BP0815, W050BP0816, W050BP0817, W050BP0818, W050BP0819, W050BP0822, W050BP0788, W050BP0820, W050BP0821, W050BP0823, W050BP0824, W050BP0825, W050BP0826, W050BP0827, W050BP0828, W053BP0836, W053BP0837, W053BP0838, W053BP0839, W056BP0883, W056BP0884, W056BP0885, W056BP0886, W056BP0870, W075BP1074, W075BP1075, W075BP1064, W075BP1065, W063BP0938, W063BP0939, W063BP0940, W075BP1080, W075BP1081, W078BP1087, W078BP1088, W078BP1089, W078BP1090, W078BP1091, W078BP1093, W078BP1094, W078BP1098, W078BP1099, W078BP1100, W078BP1101, W078BP1102, W078BP1103, W078BP1104, W084BP1211, W084BP1225, W084BP1226, W084BP1227, W084BP1228, W084BP1229, W084BP1230, W084BP1231, W070BP1003, W084BP1212, W084BP1213, W084BP1214, W084BP1215, W084BP1216, W084BP1217, W084BP1221, W084BP1222, W084BP1223, W084BP1224, W088BP1233, W088BP1234, W088BP1235, W088BP1236, W088BP1237, W088BP1238, W088BP1239, W088BP1240, W088BP1245, W088BP1246, W088BP1247, W088BP1248, W088BP1249, W088BP1250, W088BP1251, W088BP1252, W088BP1253, W088BP1254, W088BP1255, W088BP1256, W088BP1257, W088BP1258, W088BP1259, W088BP1260, W088BP1261, W088BP1262, W088BP1263, W088BP1264, W088BP1267, W088BP1268, W088BP1269, W088BP1270, W088BP1271, W088BP1272, W091BP1277, W091BP1278, W091BP1279, W091BP1280, W091BP1281, W053BP0867, W091BP1286, W091BP1287, W091BP1289, W091BP1290, W091BP1291, W091BP1292, W091BP1293, W091BP1294, W091BP1295, W091BP1296, W095BP1301, W095BP1302, W095BP1303, W095BP1305, W095BP1306, W095BP1307, W095BP1308, W095BP1312, W095BP1313, W095BP1314, W095BP1315, W095BP1316, W095BP1317, W095BP1318, W095BP1319, W095BP1320, W095BP1321, W095BP1322, W095BP1323, W095BP1324, W095BP1325, W095BP1326, W095BP1327, W095BP1328, W095BP1329, W095BP1330, W095BP1331, W095BP1332, W095BP1333, W095BP1334, W095BP1335, W095BP1336, W095BP1337, W095BP1341, W095BP1343, W095BP1350, W095BP1366, W095BP1344, W095BP1351, W095BP1352, W095BP1353, W095BP1354, W095BP1357, W095BP1358, W095BP1359, W095BP1364, W095BP1365, W095BP1309, W095BP1310, W095BP1311, W095BP1340, W096BP1370, W096BP1371, W096BP1372, W096BP1375, W096BP1376, W096BP1377, W096BP1378, W096BP1379, W096BP1380, W096BP1386, W096BP1388, W096BP1389, W071BP1027, W071BP1033, W071BP1034, W084BP1203, W095BP1360, W096BP1382, W096BP1383, W096BP1387, W096BP1390, W096BP1394, W096BP1395, W096BP1396, W096BP1397, W096BP1398, W100BP1403, W100BP1404, W100BP1406, W100BP1413, W100BP1414, W100BP1415, W100BP1416, W100BP1417, W100BP1418, W100BP1419, W100BP1420, W100BP1421, W100BP1405, W100BP1407, W100BP1408, W100BP1409, W100BP1410, W110BP1590, W110BP1591, W110BP1592, W110BP1595, W110BP1596, W110BP1597, W110BP1598, W169BP2996, W169BP2997, W169BP2998, W169BP2999, W169BP3000, W169BP3001, W169BP3003, W169BP3004, W169BP3005, W169BP3009, W169BP3010, W179BP3152, W179BP3153, W179BP3154, W179BP3155, W179BP3156, W179BP3157, W179BP3158, W179BP3159, W179BP3160, W179BP3162, W179BP3163, W179BP3164, W179BP3165, W180BP3196, W180BP3197, W180BP3198, W180BP3199, W180BP3200, W180BP3201, W180BP3202, W180BP3203, W180BP3204, W180BP3205, W180BP3206, W180BP3208, W180BP3213, W180BP3214, W180BP3216, W169BP2981, W169BP3006, W169BP3007, W179BP3161, W179BP3166, W179BP3167, W179BP3168, W179BP3185, W179BP3186, W179BP3187, W179BP3188, W179BP3189, W179BP3190, W179BP3191, W179BP3192, W179BP3193, W179BP3194, W180BP3223, W180BP3229, W180BP3285, W173BP3107, W179BP3169, W179BP3177, W179BP3178, W179BP3179, W179BP3181, W179BP3182, W179BP3184, W180BP3209, W180BP3210, W180BP3211, W180BP3212, W203BP3722, W203BP3723, W203BP3724, W203BP3725, W203BP3732, W203BP3734, W203BP3736, W203BP3738, W203BP3740, W203BP3747, W203BP3748, W203BP3749, W203BP3750, W203BP3751, W204BP3752, W204BP3753, W204BP3754, W204BP3755, W204BP3756, W204BP3758, W204BP3759, W204BP3760, W204BP3761, W203BP3731, W203BP3733, W208BP3764, W208BP3765, W208BP3766, W208BP3767, W208BP3768, W208BP3769, W208BP3770, W208BP3772, W208BP3773, W208BP3774, W208BP3775, W208BP3776, W208BP3777, W208BP3780, W208BP3781, W208BP3782, W208BP3783, W208BP3784, W208BP3785, W208BP3786, W203BP3735, W203BP3741, W203BP3742, W203BP3743, W203BP3744, W203BP3745, W208BP3763, W208BP3771, W208BP3778, W208BP3787, W208BP3788, W208BP3789, W208BP3790, W208BP3792, W208BP3794, W208BP3795, W208BP3797, W208BP3798, W210BP3800, W210BP3801, W210BP3802, W210BP3803, W210BP3804, W210BP3811, W210BP3834, W203BP3737, W203BP3739, W203BP3746, W210BP3806, W210BP3807, W210BP3813, W210BP3830, W210BP3831, W210BP3832, W210BP3833, W210BP3835, W210BP3836, W210BP3853, W210BP3856, W217BP4037, W217BP4038, W217BP4039, W217BP4040, W196BP3566, W196BP3567, W196BP3568, W196BP3569, W196BP3570, W196BP3571, W196BP3572, W196BP3573, W196BP3574, W196BP3575, W196BP3576, W196BP3577, W196BP3578, W196BP3579, W196BP3580, W196BP3585, W196BP3593, W196BP3594, W196BP3609, W201BP3714, W235BP4282, W235BP4287, W235BP4292, W235BP4299, W235BP4301, W235BP4302, W235BP4304, W235BP4307, W235BP4309, W235BP4311, W235BP4313, W236BP4314, W236BP4315, W236BP4320, W236BP4321, W236BP4322, W236BP4324, W236BP4326, W236BP4329, W236BP4335, W235BP4284, W235BP4286, W235BP4289, W235BP4293, W235BP4294, W235BP4298, W235BP4300, W235BP4305, W236BP4316, W236BP4317, W236BP4318, W236BP4319, W236BP4323, W236BP4325, W236BP4328, W236BP4330, W236BP4331, W236BP4333, W236BP4334, W236BP4336, W236BP4337, W238BP4344, W138BP4343, W225BP4118, W238BP4338, W238BP4339, W238BP4340, W274BP5016, W274BP5018, W274BP5019, W307BP5777, W307BP5778, W307BP5810, W313BP5842, W313BP5850, W315BP5884, W315BP5902, W315BP5904, W316BP5988, W316BP5989, W316BP5990, W316BP5991, W316BP5995, W320BP5996, W320BP5997, W322BP6077, W323BP6082, W323BP6088, W344BP6727, W344BP6728, W344BP6729, W344BP6717, W347BP6771, W347BP6776, W347BP6777, W347BP6778, W347BP6781, W347BP6782, W347BP6783, W347BP6787, W347BP6788, W347BP6789, W347BP6790, W347BP6791, W347BP6792, W347BP6793, W347BP6802, W347BP6803, W348BP6805, W348BP6806, W348BP6807, W348BP6809, W348BP6810, W348BP6811, W348BP6812, W348BP6813, W348BP6814, W348BP6815, W348BP6816, W348BP6817, W348BP6818, W348BP6819, W348BP6820, W348BP6821, W348BP6823, W348BP6824, W348BP6825, W348BP6826, W348BP6827, W348BP6828, W348BP6830, W348BP6831, W323BP6148, W349BP6856, W349BP6864, W341BP6648, W341BP6655, W341BP6658, W341BP6659, W341BP6660, W341BP6661, W354BP6891, W354BP6892, W354BP6898, W354BP6899, W354BP6906, W354BP6908, W354BP6909, W354BP6916, W354BP6917, W354BP6919, W354BP6920, W354BP6921;
m) P3255A03, UDI/DI 00887761968110, Serial Numbers: W043BP0701;
n) P3255ARENT01, UDI/DI 00887761976887, Serial Numbers: V323BP0086, V323BP0076, V324BP0090, V324BP0095, V324BP0096, V325BP0101, V325BP0107, W110BP1599, W110BP1600, W110BP1601, W110BP1602, W110BP1603, W110BP1604, W110BP1605, W110BP1606, W110BP1607, W110BP1608, W109BP1555, W109BP1556, W109BP1578, W109BP1586, W110BP1609, W111BP1615, W111BP1616, W111BP1617, W111BP1618, W111BP1627, W111BP1628, W111BP1631, W111BP1635, W111BP1637, W111BP1638, W112BP1641, W112BP1642, W112BP1643, W112BP1644, W112BP1645, W112BP1646, W112BP1647, W112BP1648, W112BP1649, W112BP1650, W112BP1651, W112BP1655, W112BP1657, W112BP1658, W112BP1659, W111BP1622, W111BP1624, W118BP1749, W118BP1750, W118BP1751, W118BP1752, W118BP1775, W118BP1777, W118BP1778, W118BP1779, W118BP1780, W118BP1781, W118BP1782, W118BP1783, W118BP1784, W118BP1785, W071BP1032, W118BP1743, W118BP1753, W118BP1754, W118BP1755, W118BP1756, W118BP1757, W118BP1758, W118BP1759, W118BP1760, W118BP1761, W118BP1762, W118BP1763, W118BP1765, W118BP1767, W118BP1768, W118BP1769, W118BP1774, W118BP1776, W118BP1787, W118BP1789, W118BP1790, W118BP1791, W118BP1792, W119BP1816, W119BP1817, W119BP1818, W119BP1819, W119BP1820, W119BP1822, W119BP1823, W119BP1824, W120BP1850, W120BP1851, W120BP1852, W120BP1853, W120BP1854, W120BP1855, W120BP1856, W120BP1857, W120BP1858, W120BP1859, W120BP1860, W120BP1861, W120BP1862, W120BP1863, W120BP1864, W120BP1865, W120BP1869, W124BP1882, W124BP1883, W124BP1884, W124BP1885, W124BP1886, W124BP1887, W124BP1888, W124BP1889, W124BP1890, W124BP1891, W124BP1892, W102BP1445, W116BP1723, W116BP1727, W124BP1881, W124BP1893, W124BP1894, W124BP1895, W124BP1896, W124BP1897, W124BP1898, W124BP1899, W124BP1900, W124BP1901, W124BP1902, W126BP1969, W126BP1970, W126BP1971, W126BP1972, W126BP1974, W112BP1652, W113BP1702, W123BP1879, W123BP1880, W124BP1950, W124BP1956, W124BP1957, W126BP1973, W126BP1986, W126BP1987, W126BP1989, W126BP1990, W126BP1991, W126BP1992, W126BP1993, W126BP1994, W126BP1995, W126BP1996, W126BP1997, W127BP2015, W127BP2016, W127BP2017, W127BP2018, W127BP2020, W130BP2022, W130BP2023, W130BP2024, W130BP2025, W130BP2026, W130BP2027, W130BP2028, W130BP2029, W130BP2030, W130BP2031, W130BP2032, W130BP2034, W130BP2035, W131BP2059, W131BP2060, W131BP2061, W131BP2062, W131BP2063, W131BP2064, W131BP2065, W131BP2066, W138BP2245, W138BP2246, W138BP2247, W138BP2248, W138BP2249, W138BP2250, W138BP2251, W138BP2252, W138BP2253, W138BP2254, W138BP2255, W138BP2256, W140BP2323, W140BP2324, W140BP2325, W140BP2326, W140BP2327, W140BP2328, W140BP2332, W141BP2333, W141BP2337, W141BP2338, W141BP2339, W141BP2340, W141BP2341, W141BP2342, W141BP2343, W141BP2344, W141BP2336, W141BP2351, W141BP2352, W141BP2353, W141BP2354, W141BP2355, W141BP2356, W144BP2347, W144BP2383, W144BP2384, W144BP2385, W144BP2386, W144BP2387, W144BP2388, W144BP2389, W144BP2390, W144BP2391, W144BP2392, W144BP2393, W144BP2394, W144BP2395, W144BP2396, W144BP2397, W144BP2398, W111BP1626, W118BP1739, W137BP2175, W145BP2422, W141BP2335, W147BP2479, W147BP2480, W147BP2483, W147BP2485, W147BP2487, W147BP2488, W147BP2489, W147BP2491, W147BP2493, W147BP2494, W147BP2495, W147BP2496, W147BP2497, W147BP2498, W147BP2499, W147BP2500, W147BP2501, W147BP2502, W165BP2939, W165BP2941, W165BP2942, W165BP2943, W165BP2944, W165BP2945, W165BP2946, W165BP2947, W165BP2948, W165BP2953, W165BP2954, W165BP2955, W165BP2956, W165BP2957, W165BP2958, W165BP2960, W165BP2940, W165BP2949, W165BP2950, W165BP2951, W165BP2952, W165BP2959, W165BP2961, W165BP2962, W165BP2963, W165BP2964, W165BP2965, W165BP2966, W165BP2967, W165BP2968, W165BP2969, W169BP2973, W169BP2974, W169BP2975, W169BP2976, W169BP2977, W169BP2984, W169BP2986, W169BP2987, W169BP2988, W169BP2989, W169BP2990, W169BP2991, W169BP2993, W169BP2978, W169BP2979, W169BP2980, W169BP2992, W169BP2994, W169BP2995, W169BP3011, W169BP3012, W169BP3013, W169BP3014, W169BP3015, W169BP3016, W169BP3017, W169BP3018, W169BP3019, W169BP3020, W169BP3021, W169BP3022, W169BP3031, W147BP2484, W155BP2687, W169BP3023, W169BP3024, W169BP3025, W169BP3026, W169BP3027, W190BP3515, W190BP3518, W190BP3522, W214BP3936, W214BP3937, W214BP3938, W214BP3940, W214BP3941, W214BP3942, W214BP3943, W214BP3945, W214BP3946, W214BP3947, W214BP3948, W214BP3949, W214BP3951, W214BP3952, W214BP3964, W214BP3965, W214BP3966, W214BP3967, W214BP3950, W214BP3955, W214BP3956, W214BP3957, W214BP3958, W214BP3959, W214BP3961, W214BP3962, W214BP3979, W214BP3980, W214BP3981, W214BP3982, W214BP3983, W190BP3472, W194BP3538, W194BP3539, W194BP3541, W194BP3543, W194BP3546, W194BP3547, W194BP3562, W194BP3563, W194BP3564, W215BP3984, W215BP3985, W215BP3986, W215BP3987, W274BP5021, W274BP5023, W274BP5025, W274BP5026, W274BP5027, W274BP5028, W273BP5011, W273BP5012, W274BP5029, W274BP5030, W274BP5031, W274BP5032, W274BP5033, W274BP5034, W274BP5035, W274BP5036, W274BP5037, W274BP5038, W274BP5039, W274BP5040, W274BP5041, W274BP5042, W274BP5043, W274BP5044, W274BP5045, W274BP5046, W274BP5047, W274BP5049, W274BP5050, W274BP5051, W274BP5052, W274BP5053, W274BP5054, W274BP5055, W274BP5057, W274BP5058, W274BP5059, W274BP5060, W274BP5061, W274BP5062, W274BP5063, W274BP5064, W274BP5065, W274BP5066, W274BP5067, W274BP5068, W274BP5069, W274BP5070, W274BP5071, W274BP5072, W274BP5073, W274BP5074, W274BP5075, W274BP5076, W274BP5077, W274BP5078, W274BP5079, W274BP5081, W274BP5082, W277BP5115, W305BP5708, W305BP5709, W305BP5712, W305BP5714, W305BP5715, W305BP5716, W305BP5717, W305BP5718, W305BP5720, W305BP5721, W305BP5727, W305BP5728, W305BP5729, W305BP5730, W305BP5731, W305BP5732, W305BP5733, W305BP5734, W305BP5735, W305BP5736, W305BP5737, W305BP5738, W305BP5741, W305BP5744, W306BP5758, W306BP5759, W306BP5760, W327BP6280, W327BP6281, W327BP6282, W327BP6283, W327BP6284;
o) P7923A01, UDI/DI 00887761977945, Serial Numbers: V322BP0049, V322BP0050, V322BP0051, V322BP0052, V322BP0053, V322BP0054, V322BP0055, V322BP0056, V322BP0057, V322BP0058, V322BP0059, V322BP0060, V322BP0061, V322BP0062, V322BP0066, V323BP0022, V323BP0067, V323BP0068, V323BP0069, V323BP0070, V323BP0071, W014BP0511, W014BP0512, W014BP0513, W014BP0514, W015BP0516, W018BP0536, W018BP0537, W018BP0538, W018BP0539, W025BP0559, W025BP0560, W025BP0561, W025BP0563, W012BP0506, W046BP0776, W046BP0777, W046BP0778, W046BP0779, W046BP0780, W046BP0781, W046BP0782, W046BP0783, W046BP0784, W046BP0785, W056BP0871, W056BP0872, W056BP0873, W056BP0874, W056BP0875, W056BP0876, W056BP0877, W056BP0887, W056BP0888, W057BP0890, W057BP0891, W057BP0892, W057BP0893, W057BP0894, W057BP0895, W057BP0896, W057BP0897, W057BP0898, W063BP0919, W063BP0920, W063BP0921, W063BP0922, W063BP0923, W063BP0914, W063BP0924, W063BP0925, W063BP0926, W063BP0927, W063BP0928, W063BP0930, W063BP0945, W063BP0946, W063BP0947, W070BP0981, W070BP0963, W070BP0964, W070BP0965, W070BP0966, W070BP0967, W070BP0968, W070BP0969, W070BP0970, W070BP0971, W070BP0972, W070BP0973, W070BP0974, W070BP0975, W070BP0982, W070BP0984, W070BP0985, W070BP0986, W070BP0987, W070BP0988, W070BP0989, W070BP0990, W070BP0991, W070BP0992, W070BP0993, W070BP0997, W053BP0830, W053BP0831, W053BP0832, W053BP0833, W053BP0865, W053BP0868, W070BP0994, W070BP0995, W070BP0996, W070BP0998, W070BP0999, W070BP1000, W070BP1001, W070BP1002, W071BP1005, W053BP0864, W053BP0866, W071BP1007, W071BP1009, W071BP1022, W071BP1023, W071BP1024, W071BP1025, W071BP1029, W071BP1030, W071BP1031, W075BP1060, W075BP1061, W075BP1062, W140BP2322, W137BP2196, W137BP2197, W141BP2345, W141BP2346, W141BP2347, W145BP2404, W145BP2406, W145BP2407, W145BP2408, W145BP2409, W145BP2412, W145BP2413, W145BP2414, W147BP2505, W147BP2506, W147BP2509, W147BP2511, W147BP2512, W151BP2513, W151BP2515, W151BP2516, W151BP2519, W151BP2520, W151BP2514, W151BP2517, W151BP2518, W151BP2521, W151BP2522, W151BP2523, W151BP2524, W151BP2525, W151BP2526, W151BP2527, W151BP2528, W151BP2529, W151BP2530, W151BP2531, W151BP2532, W151BP2552, W151BP2555, W151BP2534, W151BP2535, W151BP2536, W151BP2537, W151BP2539, W151BP2540, W151BP2541, W151BP2542, W201BP3690, W201BP3716, W238BP4350, W238BP4352, W238BP4354, W238BP4355, W238BP4356, W238BP4357, W238BP4358, W238BP4359, W238BP4351, W223BP4091, W239BP4360, W267BP4822, W267BP4823, W267BP4825, W267BP4828, W267BP4830, W284BP5144, W305BP5697, W305BP5698, W305BP5699, W305BP5700, W305BP5701, W305BP5702, W305BP5705, W326BP6158, W326BP6170, W326BP6173, W326BP6186, W327BP6205, W354BP6924, W354BP6925, W354BP6926, W354BP6927, W354BP6928, W354BP6929, W354BP6930, W354BP6931, W354BP6932, W354BP6933, W354BP6934, W354BP6935, W354BP6936, W354BP6940, W354BP6941, W354BP6942, W355BP6944, W354BP6910, W354BP6911, W354BP6918, W354BP6939, W355BP5964, W355BP6914, W355BP6945, W355BP6946, W355BP6948, W355BP6949, W355BP6950, W355BP6951, W355BP6952, W355BP6953, W355BP6954, W355BP6955, W355BP6956, W355BP6957, W355BP6958, W355BP6959, W355BP6960, W355BP6961, W355BP6966, W355BP6976, W355BP6977, W355BP6978, W355BP6979, W355BP6983, W355BP6984, W355BP6985, W355BP6986, W355BP6987, W355BP6990, W355BP6980, W355BP6981, W355BP6982, W355BP6988, W355BP6989, W357BP6994, W357BP6996, W357BP6997, W357BP6998, W357BP7006, W357BP7007, W357BP7009, W357BP7012, W357BP7014, W357BP7015, W357BP7016, W357BP7017, W357BP7018, W357BP7019, W357BP7021, W357BP7022, W357BP7023, W357BP7024, W357BP7025, W357BP7026, W357BP7027, W341BP6647, W341BP6663, W341BP6666, W341BP6667, W341BP6668, W341BP6669, W341BP6670, W341BP6671, W341BP6674, W336BP6541, W336BP6546;
p) P7923A02, UDI/DI 00887761977938, Serial Numbers: V342BP0184, V342BP0187, V343BP0202, W137BP2178, W137BP2179, W137BP2180, W127BP2002, W127BP2003, W127BP2004, W151BP2544, W151BP2546, W151BP2547, W151BP2548, W151BP2549, W151BP2550, W153BP2590, W153BP2593, W153BP2594, W153BP2595, W153BP2597, W153BP2598, W153BP2599, W153BP2601, W153BP2602, W153BP2603, W155BP2690, W155BP2691, W155BP2692, W131BP2071, W134BP2158, W154BP2664, W154BP2666, W154BP2667, W154BP2669, W155BP2682, W302BP5660, W302BP5662, W302BP5663, W302BP5665, W302BP5678, W302BP5679, W302BP5680, W302BP5681, W302BP5682, W302BP5683, W302BP5684, W302BP5686, W305BP5687, W305BP5688, W305BP5689, W305BP5690, W305BP5691, W286BP5289, W305BP5693, W305BP5694, W305BP5695, W305BP5696, W327BP6245, W327BP6247;
q) P7923A03, UDI/DI 00887761977921, Serial Numbers: V322BP0047, V322BP0048, V342BP0188, V342BP0190, V342BP0191, V342BP0192, V342BP0193, V342BP0194, V342BP0195, V343BP0200, V343BP0201, V343BP0208, V343BP0209, V343BP0210, V343BP0211, V343BP0212, V343BP0213, V343BP0214, V343BP0215, V344BP0234, V344BP0235, V344BP0236, V344BP0237, V344BP0238, V344BP0239, V344BP0240, V344BP0241, V344BP0242, V344BP0243, V344BP0244, V344BP0245, V344BP0246, V344BP0252, V344BP0253, V349BP0260, V349BP0261, V349BP0262, V349BP0263, V349BP0264, V349BP0265, V349BP0266, V349BP0267, V343BP0218, V343BP0219, V343BP0220, V343BP0221, V343BP0222, V343BP0229, V343BP0231, V349BP0268, V349BP0269, V349BP0270, V349BP0271, V349BP0272, V350BP0273, V350BP0274, V350BP0275, V356BP0409, V356BP0411, V356BP0413, V356BP0414, V356BP0415, V356BP0416, V356BP0417, V356BP0418, V353BP0361, V353BP0367, V353BP0369, V353BP0370, V353BP0371, V353BP0373, V356BP0410, V356BP0419, V356BP0420, V356BP0421, V356BP0422, V356BP0423, V356BP0424, V356BP0425, V356BP0426, V356BP0375, V356BP0376, V356BP0377, V356BP0378, V356BP0379, V356BP0380, V356BP0381, V356BP0383, V356BP0384, V356BP0385, V356BP0386, V356BP0387, V356BP0388, V356BP0389, V356BP0390, V356BP0427, V356BP0404, W004BP0456, W004BP0457, W004BP0459, W004BP0458, W004BP0460, W004BP0461, W004BP0462, W004BP0463, W004BP0464, W004BP0465, W004BP0470, W004BP0471, W004BP0472, W004BP0466, W004BP0467, W004BP0468, W004BP0469, W004BP0473, W004BP0474, W004BP0476, W004BP0477, W004BP0478, W006BP0482, V356BP0431, W004BP0475, W006BP0479, W006BP0480, W006BP0481, W006BP0483, W006BP0484, W006BP0485, W006BP0487, W006BP0490, W006BP0494, W006BP0486, W006BP0488, W006BP0489, W006BP0491, W006BP0492, W006BP0493, W018BP0540, W018BP0541, W018BP0542, W018BP0543, W018BP0544, W019BP0545, W025BP0569, W025BP0570, W025BP0571, W025BP0572, W025BP0573, W025BP0576, W012BP0497, W014BP0508, W025BP0552, W025BP0553, W025BP0554, W025BP0555, W025BP0556, W025BP0557, W025BP0558, W025BP0574, W025BP0575, W025BP0577, W025BP0578, W028BP0580, W028BP0587, W028BP0588, W028BP0581, W028BP0582, W028BP0583, W028BP0584, W028BP0585, W028BP0586, W028BP0589, W028BP0591, W028BP0592, W028BP0600, W028BP0593, W028BP0594, W028BP0595, W028BP0596, W028BP0597, W028BP0598, W028BP0599, W028BP0601, W028BP0603, W028BP0604, W028BP0605, W028BP0606, W028BP0607, W028BP0610, W028BP0602, W028BP0608, W028BP0609, W028BP0612, W028BP0613, W028BP0614, W028BP0615, W028BP0616, W028BP0617, W028BP0618, W028BP0620, W036BP0622, W039BP0655, W039BP0656, W039BP0657, W039BP0658, W039BP0659, W039BP0660, W039BP0661, W039BP0662, W039BP0663, W039BP0664, W039BP0666, W039BP0667, W039BP0668, W039BP0670, W039BP0671, W039BP0672, W039BP0673, W039BP0674, W039BP0675, W039BP0669, W053BP0849, W053BP0850, W053BP0851, W053BP0852, W053BP0853, W053BP0854, W053BP0855, W053BP0857, W053BP0858, W053BP0859, W053BP0860, W053BP0862, W056BP0878, W056BP0880, W056BP0881, W056BP0882, W075BP1066, W075BP1067, W075BP1068, W075BP1069, W075BP1076, W075BP1077, W075BP1078, W063BP0932, W063BP0933, W063BP0935, W063BP0936, W063BP0937, W083BP1178, W083BP1179, W083BP1180, W083BP1181, W083BP1182, W083BP1183, W083BP1184, W083BP1185, W083BP1186, W083BP1188, W100BP1411, W100BP1412, W100BP1422, W102BP1424, W102BP1425, W102BP1426, W102BP1427, W102BP1428, W102BP1429, W102BP1430, W102BP1431, W102BP1432, W102BP1433, W102BP1434, W102BP1435, W102BP1436, W102BP1437, W102BP1438, W102BP1439, W102BP1440, W102BP1441, W102BP1442, W102BP1446, W102BP1447, W102BP1443, W102BP1448, W102BP1449, W102BP1450, W102BP1451, W102BP1452, W102BP1453, W102BP1454, W102BP1455, W102BP1456, W102BP1457, W102BP1458, W102BP1459, W102BP1460, W102BP1461, W102BP1462, W102BP1463, W102BP1464, W103BP1466, W103BP1467, W103BP1468, W103BP1469, W103BP1470, W103BP1471, W103BP1476, W103BP1480, W083BP1190, W103BP1472, W103BP1473, W103BP1474, W103BP1475, W103BP1477, W103BP1478, W103BP1479, W103BP1481, W103BP1482, W103BP1483, W103BP1484, W103BP1485, W103BP1486, W103BP1487, W103BP1489, W071BP1028, W081BP1150, W081BP1152, W081BP1153, W081BP1154, W081BP1155, W081BP1156, W081BP1157, W081BP1158, W081BP1159, W095BP1361, W095BP1363, W103BP1490, W103BP1491, W106BP1493, W106BP1494, W106BP1495, W078BP1123, W106BP1496, W106BP1497, W106BP1498, W106BP1499, W106BP1500, W106BP1501, W106BP1502, W106BP1503, W106BP1504, W106BP1505, W106BP1506, W106BP1507, W106BP1508, W106BP1509, W106BP1510, W106BP1511, W106BP1512, W106BP1513, W106BP1514, W106BP1515, W106BP1516, W106BP1517, W106BP1518, W106BP1519, W106BP1520, W106BP1521, W106BP1522, W106BP1523, W106BP1524, W106BP1525, W106BP1527, W106BP1536, W106BP1537, W106BP1538, W106BP1544, W106BP1545, W106BP1528, W106BP1529, W106BP1530, W106BP1531, W106BP1532, W106BP1534, W106BP1535, W106BP1539, W106BP1540, W106BP1541, w106bp1542, W106BP1543, W106BP1546, W106BP1547, W106BP1548, W106BP1549, W106BP1552, W106BP1553, W109BP1557, W109BP1558, W109BP1559, W109BP1560, W109BP1561, W109BP1562, W109BP1563, W109BP1564, W109BP1565, W109BP1567, W109BP1568, W109BP1569, W109BP1570, W109BP1571, W109BP1572, W109BP1573, W109BP1574, W109BP1575, W109BP1580, W109BP1582, W109BP1583, W109BP1584, W109BP1585, W088BP1244, W106BP1550, W106BP1551, W109BP1576, W109BP1577, W110BP1588, W110BP1589, W138BP2257, W138BP2258, W138BP2260, W138BP2261, W138BP2262, W138BP2263, W138BP2264, W138BP2224, W138BP2225, W138BP2226, W138BP2227, W138BP2228, W138BP2229, W138BP2230, W138BP2233, W138BP2234, W138BP2235, W138BP2236, W138BP2237, W138BP2259, W138BP2265, W138BP2266, W138BP2267, W138BP2268, W138BP2269, W138BP2270, W138BP2272, W138BP2273, W138BP2274, W138BP2275, W138BP2282, W138BP2283, W138BP2284, W138BP2291, W138BP2292, W138BP2293, W138BP2294, W138BP2295, W138BP2296, W138BP2297, W138BP2298, W138BP2299, W138BP2300, W138BP2301, W116BP1726, W127BP2009, W137BP2198, W137BP2199, W137BP2200, W137BP2201, W137BP2202, W137BP2203, W137BP2204, W137BP2205, W137BP2206, W137BP2207, W137BP2208, W137BP2209, W138BP2276, W138BP2277, W138BP2278, W138BP2279, W138BP2280, W138BP2281, W138BP2285, W138BP2286, W138BP2287, W138BP2288, W138BP2289, W138BP2290, W140BP2302, W140BP2303, W140BP2304, W140BP2308, W140BP2309, W140BP2310, W140BP2305, W140BP2306, W140BP2311, W140BP2312, W140BP2313, W140BP2314, W140BP2315, W140BP2316, W140BP2317, W140BP2318, W140BP2319, W140BP2320, W140BP2321, W137BP2185, W137BP2186, W137BP2187, W145BP2410, W119BP1800, W141BP2373, W141BP2374, W141BP2375, W141BP2376, W141BP2377, W141BP2378, W141BP2379, W141BP2380, W141BP2381, W141BP2382, W144BP2400, W144BP2401, W144BP2402, W145BP2405, W145BP2411, W151BP2563, W151BP2564, W151BP2566, W151BP2567, W151BP2568, W151BP2569, W151BP2570, W151BP2571, W153BP2606, W153BP2607, W153BP2608, W153BP2609, W153BP2611, W153BP2613, W153BP2614, W153BP2615, W153BP2616, W153BP2617, W153BP2618, W153BP2619, W153BP2620, W153BP2577, W153BP2612, W153BP2621, W153BP2622, W153BP2632, W153BP2633, W153BP2634, W153BP2635, W153BP2648, W153BP2649, W153BP2650, W153BP2651, W153BP2636, W153BP2637, W153BP2638, W153BP2639, W153BP2640, W153BP2641, W153BP2642, W153BP2644, W153BP2645, W153BP2646, W153BP2647, W154BP2663, W154BP2665, W154BP2668, W155BP2671, W155BP2672, W155BP2673, W155BP2674, W155BP2675, W155BP2676, W155BP2677, W155BP2678, W155BP2679, W155BP2680, W155BP2681, W155BP2685, W155BP2686, W155BP2688, W155BP2689, W155BP2699, W155BP2700, W155BP2701, W155BP2703, W155BP2707, W155BP2708, W155BP2710, W155BP2711, W155BP2712, W155BP2713, W155BP2714, W155BP2715, W155BP2716, W155BP2717, W155BP2718, W155BP2719, W155BP2720, W155BP2721, W155BP2722, W151BP2543, W154BP2657, W154BP2658, W154BP2659, W154BP2660, W154BP2661, W154BP2662, W155BP2704, W155BP2705, W155BP2706, W155BP2709, W155BP2723, W155BP2724, W155BP2725, W155BP2726, W155BP2727, W155BP2728, W155BP2729, W155BP2730, W155BP2731, W155BP2732, W155BP2733, W155BP2734, W155BP2737, W155BP2738, W155BP2745, W155BP2746, W158BP2748, W158BP2749, W158BP2750, W158BP2751, W158BP2755, W158BP2756, W155BP2736, W155BP2739, W155BP2740, W155BP2741, W155BP2742, W155BP2743, W155BP2744, W158BP2753, W158BP2754, W158BP2757, W158BP2759, W158BP2760, W158BP2761, W158BP2762, W158BP2763, W158BP2764, W158BP2766, W158BP2767, W158BP2768, W158BP2769, W158BP2770, W158BP2775, W158BP2776, W158BP2777, W158BP2778, W159BP2793, W159BP2799, W147BP2481, W147BP2482, W158BP2771, W158BP2772, W158BP2773, W158BP2774, W158BP2779, W158BP2780, W158BP2781, W159BP2782, W159BP2787, W159BP2788, W159BP2789, W159BP2790, W159BP2791, W159BP2792, W159BP2794, W159BP2795, W159BP2796, W159BP2798, W159BP2800, W159BP2802, W159BP2783, W159BP2784, W159BP2785, W159BP2786, W159BP2797, W159BP2801, W159BP2803, W159BP2804, W159BP2805, W159BP2806, W159BP2807, W159BP2808, W159BP2809, W159BP2810, W159BP2811, W159BP2812, W161BP2813, W161BP2814, W161BP2815, W161BP2816, W161BP2817, W161BP2818, W161BP2819, W161BP2820, W161BP2826, W161BP2827, W161BP2832, W161BP2821, W161BP2822, W161BP2823, W161BP2824, W161BP2825, W161BP2828, W161BP2829, W161BP2830, W161BP2831, W161BP2833, W161BP2834, W161BP2835, W161BP2842, W161BP2843, W161BP2844, W162BP2857, W162BP2858, W162BP2859, W161BP2836, W161BP2837, W161BP2838, W161BP2839, W161BP2840, W161BP2841, W161BP2845, W161BP2846, W161BP2847, W161BP2848, W161BP2849, W161BP2850, W161BP2851, W161BP2852, W161BP2949, W162BP2853, W162BP2854, W162BP2855, W162BP2856, W162BP2860, W162BP2861, W162BP2862, W162BP2863, W162BP2864, W162BP2865, W162BP2866, W162BP2867, W162BP2868, W162BP2869, W162BP2870, W162BP2871, W162BP2872, W162BP2892, W162BP2893, W162BP2894, W162BP2895, W162BP2896, W162BP2897, W162BP2908, W162BP2873, W162BP2874, W162BP2875, W162BP2876, W162BP2877, W162BP2878, W162BP2879, W162BP2880, W162BP2881, W162BP2882, W162BP2883, W162BP2884, W162BP2885, W162BP2886, W162BP2887, W162BP2888, W162BP2889, W162BP2890, W162BP2898, W162BP2899, W162BP2900, W162BP2901, W162BP2902, W162BP2903, W162BP2904, W162BP2905, W162BP2906, W162BP2907, W162BP2909, W172BP3036, W172BP3037, W172BP3038, W172BP3039, W172BP3040, W172BP3041, W172BP3042, W172BP3043, W172BP3044, W172BP3045, W172BP3046, W172BP3047, W172BP3048, W172BP3049, W172BP3050, W172BP3051, W172BP3052, W172BP3057, W172BP3058, W172BP3059, W172BP3060, W172BP3061, W172BP3067, W151BP2556, W151BP2557, W151BP2558, W153BP2600, W153BP2643, W162BP2911, W162BP2912, W172BP3053, W172BP3054, W172BP3055, W172BP3056, W172BP3062, W172BP3063, W172BP3066, W173BP3070, W173BP3071, W173BP3072, W173BP3073, W173BP3074, W173BP3075, W173BP3076, W173BP3077, W173BP3078, W173BP3079, W173BP3080, W173BP3081, W173BP3082, W173BP3083, W173BP3084, W173BP3085, W173BP3086, W173BP3087, W179BP3136, W179BP3137, W179BP3138, W179BP3139, W179BP3140, W179BP3141, W179BP3142, W179BP3143, W179BP3144, W179BP3145, W179BP3146, W179BP3147, W179BP3173, W179BP3174, W179BP3176, W183BP3308, W183BP3309, W183BP3310, W183BP3311, W183BP3312, W183BP3313, W183BP3314, W183BP3315, W183BP3316, W183BP3317, W183BP3318, W183BP3319, W183BP3320, W183BP3321, W183BP3322, W183BP3324, W183BP3332, W173BP3126, W183BP3367, W183BP3368, W183BP3369, W183BP3370, W183BP3372, W183BP3378, W183BP3379, W183BP3380, W183BP3383, W183BP3385, W183BP3386, W183BP3388, W186BP3397, W186BP3403, W186BP3404, W186BP3405, W186BP3406, W186BP3407, W186BP3408, W186BP3409, W186BP3410, W186BP3411, W186BP3412, W186BP3414, W169BP3002, W173BP3127, W183BP3294, W183BP3295, W183BP3296, W183BP3301, W183BP3323, W183BP3325, W183BP3326, W183BP3327, W183BP3328, W183BP3329, W183BP3330, W186BP3413, W186BP3415, W186BP3416, W186BP3417, W186BP3418, W186BP3419, W186BP3420, W190BP3451, W190BP3452, W190BP3453, W190BP3454, W190BP3455, W197BP3682, W197BP3683, W201BP3687, W201BP3692, W201BP3693, W201BP3694, W201BP3695, W201BP3697, W201BP3698, W201BP3699, W201BP3702, W201BP3703, W201BP3705, W201BP3709, W201BP3689, W201BP3691, W201BP3704, W201BP3706, W201BP3707, W201BP3708, W201BP3710, W201BP3711, W201BP3712, W201BP3713, W201BP3717, W201BP3715, W203BP3719, W203BP3720, W203BP3721, W217BP4029, W217BP4031, W217BP4032, W217BP4033, W217BP4034, W217BP4035, W217BP4036, W190BP3460, W194BP3551, W196BP3581, W196BP3582, W196BP3583, W196BP3584, W196BP3586, W196BP3587, W196BP3588, W196BP3589, W196BP3590, W196BP3591, W196BP3592, W214BP3960, W214BP3963, W218BP4043, W218BP4048, W218BP4049, W218BP4050, W218BP4051, W228BP4182, W228BP4185, W228BP4186, W228BP4188, W228BP4189, W228BP4196, W228BP4200, W228TP9682, W288BP4182, W228BP4183, W228BP4184, W228BP4190, W228BP4191, W228BP4193, W228BP4194, W228BP4195, W228BP4198, W228BP4199, W228BP4201, W232BP4258, W231BP4202, W231BP4203, W231BP4204, W258BP4723, W258BP4724, W258BP4725, W258BP4726, W258BP4727, W258BP4729, W258BP4730, W258BP4731, W258BP4732, W258BP4735, W258BP4736, W258BP4739, W258BP4716, W260BP4740, W260BP4742, W260BP4743, W260BP4744, W260BP4746, W260BP4749, W260BP4780, W260BP4781, W260BP4782, W260BP4783, W260BP4784, W260BP4789, W260BP4790, W260BP4791, W260BP4792, W264BP4817, W264BP4818, W264BP4819, W264BP4820, W264BP4821, W246BP4506, W246BP4507, W246BP4508, W246BP4515, W246BP4516, W246BP4517, W256BP4667, W256BP4668, W256BP4669, W256BP4674, W256BP4675, W256BP4676, W256BP4678, W256BP4679, W256BP4681, W256BP4682, W256BP4683, W256BP4684, W256BP4685, W256BP4687, W256BP4688, W256BP4689, W256BP4690, W256BP4699, W256BP4701, W256BP4702, W256BP4706, W291BP5342, W291BP5347, W291BP5349, W291BP5350, W291BP5352, W291BP5353, W291BP5354, W291BP5355, W291BP5356, W291BP5357, W291BP5358, W291BP5359, W291BP5360, W291BP5333, W291BP5366, W292BP5381, W292BP5382, W292BP5384, W292BP5403, W292BP5404, W292BP5405, W292BP5409, W292BP5412, W292BP5413, W292BP5414, W292BP5415, W292BP5416, W292BP5419, W292BP5420, W292BP5421, W292BP5423, W292BP5424, W292BP5426, W292BP5427, W292BP5428, W301BP5623, W301BP5624, W301BP5625, W301BP5626, W301BP5627, W301BP5628, W301BP5629, W301BP5630, W301BP5632, W301BP5633, W301BP5634, W298BP5594, W298BP5596, W298BP5597, W298BP5598, W298BP5599, W298BP5600, W298BP5601, W299BP5608, W301BP5635, W301BP5637, W301BP5638, W301BP5641, W302BP5648, W302BP5649, W291BP5365, W301BP5642, W302BP5646, W302BP5651, W302BP5655, W302BP5656, W302BP5657, W302BP5659, W327BP6225, W327BP6227, W327BP6242;
r) P7923A04, UDI/DI 00887761977914, Serial Numbers: V326BP0125, V326BP0129, V326BP0134, V326BP0135, V326BP0136, V326BP0137, V326BP0123, V326BP0126, V326BP0127, V326BP0128, V326BP0130, V326BP0131, V326BP0132, V326BP0133, V326BP0138, V328BP0140, V328BP0141, V328BP0142, V328BP0143, V328BP0144, W012BP0503, W012BP0505, W014BP0509, W063BP0954, W063BP0955, W063BP0956, W063BP0957, W063BP0958, W063BP0959, W063BP0960, W063BP0961, W070BP0976, W070BP0977, W070BP0978, W070BP0979, W070BP0980, W075BP1063, W078BP1105, W153BP2610, W151BP2560, W151BP2561, W151BP2562, W155BP2683, W155BP2684, W155BP2693, W155BP2694, W155BP2695, W155BP2697, W155BP2698, W179BP3148, W179BP3149, W179BP3150, W239BP4361, W239BP4362, W239BP4363, W239BP4364, W239BP4365, W239BP4366, W239BP4367, W239BP4368, W239BP4369, W239BP4370, W258BP4708, W258BP4709, W258BP4710, W258BP4711, W258BP4712, W258BP4715, W258BP4717, W258BP4719, W258BP4722, W291BP5361, W291BP5362, W291BP5363, W291BP5327, W292BP5433, W292BP5441, W292BP5442, W292BP5385, W292BP5429, W294BP5467, W294BP5468, W294BP5469, W294BP5470, W294BP5471, W294BP5472, W294BP5473, W294BP5474, W294BP5475, W294BP5476, W294BP5479, W294BP5480, W294BP5481, W294BP5484, W294BP5485, W294BP5487, W294BP5488, W294BP5489, W294BP5490, W298BP5548, W298BP5549, W327BP6206, W327BP6208, W327BP6222;
s) P7924A01, UDI/DI 00887761977907, Serial Numbers: V314BP0013, V323BP0075, V323BP0077, V323BP0078, V323BP0083, V323BP0084, V323BP0085, V323BP0079, V323BP0082, V324BP0091, V324BP0093, V324BP0094, W050BP0787, W050BP0789, W050BP0790, W050BP0791, W050BP0793, W137BP2188, W137BP2189, W190BP3523, W190BP3526, W190BP3533, W190BP3535, W190BP3456, W190BP3457, W190BP3459, W190BP3462, W190BP3463, W190BP3465, W190BP3468, W190BP3469, W190BP3470, W190BP3480, W190BP3481, W190BP3482, W190BP3492, W190BP3503, W190BP3505, W190BP3525, W190BP3531, W190BP3532, W179BP3170, W179BP3175, W180BP3219, W190BP3527, W194BP3537, W194BP3540, W194BP3542, W194BP3548, W194BP3549, W194BP3550, W194BP3552, W194BP3553, W194BP3554, W194BP3555, W194BP3556, W194BP3557, W194BP3558, W194BP3559, W194BP3561, W183BP3359, W190BP3458, W197BP3534, W197BP3625, W197BP3626, W197BP3627, W197BP3628, W197BP3629, W197BP3630, W197BP3631, W197BP3632, W197BP3633, W197BP3634, W197BP3635, W197BP3636, W197BP3637, W197BP3638, W197BP3639, W197BP3640, W197BP3641, W197BP3643, W197BP3647, W197BP3648, W197BP3649, W197BP3650, W197BP3651, W197BP3652, W197BP3653, W218BP4053, W218BP4062, W218BP4063, W218BP4064, W218BP4065, W218BP4066, W218BP4067, W218BP4068, W218BP4069, W218BP4059, W218BP4041, W218BP4042, W218BP4044, W218BP4045, W218BP4046, W218BP4047, W218BP4052, W218BP4054, W218BP4055, W218BP4056, W218BP4057, W218BP4058, W218BP4060, W218BP4061, W224BP4099, W224BP4101, W224BP4102, W224BP4103, W224BP4104, W284BP5145, W327BP6248, W328BP6330, W315BP5891, W315BP5897, W336BP6540, W336BP6542, W336BP6543, W336BP6544, W336BP6548, W336BP6549, W336BP6550, W336BP6552, W336BP6555, W336BP6556, W336BP6557, W336BP6558, W336BP6560, W336BP6561, W336BP6562, W336BP6563, W337BP6628, W337BP6633, W354BP6922,;
t) P7924A02, UDI/DI 00887761977891, Serial Numbers: V301BP0007, V301BP0008, V301BP0009, V314BP0015, V324BP0097, V324BP0099, V325BP0102, V325BP0103, V325BP0104, V325BP0105, V325BP0106, V324BP0092, V325BP0108, V325BP0110, V326BP0119, V326BP0121, V326BP0122, V326BP0124, W153BP2604, W153BP2605, W179BP3151, W307BP5796, W307BP5798, W307BP5799, W307BP5816, W307BP5818, W307BP5820, W307BP5824, W327BP6275, W327BP6279;
u) P7924A03, UDI/DI 00887761977884, Serial Numbers: V314BP0028, V322BP0045, V350BP0276, V350BP0277, V350BP0278, V350BP0279, V343BP0217, V343BP0227, V343BP0228, V343BP0233, V350BP0282, V350BP0283, V350BP0284, V350BP0286, V350BP0287, V350BP0288, V350BP0289, V351BP0324, V351BP0325, V351BP0326, V351BP0327, V343BP0216, V343BP0223, V351BP0323, V351BP0328, V351BP0329, V351BP0330, V351BP0331, V351BP0332, V351BP0333, V351BP0334, V350BP0291, V350BP0292, V350BP0295, V350BP0296, V350BP0297, V350BP0298, V350BP0300, V350BP0301, V350BP0302, V356BP0408, V356BP0428, V356BP0429, V356BP0432, V356BP0392, V356BP0393, V356BP0394, V356BP0397, V356BP0398, V356BP0399, V356BP0405, V356BP0407, V356BP0430, V356BP0451, V356BP0382, V356BP0396, V356BP0400, V356BP0401, V356BP0402, V356BP0403, W004BP0455, W015BP0517, W015BP0518, W015BP0519, W015BP0520, W015BP0522, W015BP0523, W015BP0524, W015BP0525, W015BP0526, W015BP0527, W015BP0528, W015BP0529, W015BP0530, W015BP0531, W015BP0532, W015BP0533, W015BP0534, W019BP0548, W025BP0564, W025BP0565, W025BP0566, W025BP0567, W025BP0568, W012BP0496, W012BP0499, W012BP0500, W012BP0501, W012BP0504, W053BP0840, W053BP0841, W053BP0842, W053BP0843, W053BP0844, W053BP0845, W053BP0846, W053BP0847, W053BP0848, W057BP0899, W057BP0900, W057BP0901, W057BP0902, W057BP0903, W057BP0904, W057BP0905, W057BP0906, W063BP0908, W063BP0909, W063BP0910, W063BP0911, W063BP0912, W063BP0913, W063BP0915, W063BP0916, W063BP0917, W063BP0948, W063BP0949, W063BP0950, W063BP0951, W063BP0952, W063BP0953, W075BP1070, W075BP1071, W075BP1072, W075BP1073, W078BP1106, W078BP1107, W078BP1096, W078BP1097, W078BP1108, W078BP1109, W078BP1110, W078BP1111, W078BP1112, W078BP1113, W078BP1114, W078BP1115, W078BP1116, W078BP1117, W078BP1122, W078BP1124, W078BP1125, W078BP1126, W078BP1127, W078BP1118, W078BP1119, W081BP1129, W081BP1130, W081BP1131, W081BP1132, W081BP1133, W081BP1134, W081BP1135, W081BP1136, W081BP1137, W081BP1138, W081BP1139, W081BP1140, W081BP1141, W081BP1142, W081BP1143, W081BP1144, W081BP1145, W081BP1146, W081BP1147, W081BP1148, W083BP1161, W083BP1162, W083BP1163, W083BP1164, W083BP1165, W063BP0943, W076BP1085, W083BP1166, W083BP1167, W083BP1168, W083BP1169, W083BP1170, W083BP1171, W083BP1172, W083BP1173, W083BP1174, W083BP1175, W083BP1176, w083bp1177, W063BP0942, W063BP0944, W078BP1120, W078BP1121, W071BP1008, W071BP1010, W075BP1082, W075BP1083, W078BP1095, W083BP1187, W083BP1189, W084BP1201, W084BP1205, W084BP1206, W084BP1207, W084BP1208, W084BP1209, W084BP1192, W084BP1193, W084BP1194, W084BP1195, W084BP1196, W084BP1197, W084BP1198, W084BP1199, W084BP1202, W084BP1218, W084BP1219, W084BP1220, W153BP2623, W153BP2624, W153BP2625, W153BP2626, W153BP2627, W153BP2628, W153BP2629, W153BP2630, W153BP2631, W151BP2565, W151BP2572, W151BP2573, W151BP2574, W151BP2575, W151BP2576, W151BP2577, W151BP2578, W151BP2579, W162BP2910, W162BP2913, W162BP2914, W162BP2915, W162BP2916, W162BP2917, W162BP2918, W140BP2307, W165BP2920, W165BP2921, W165BP2922, W165BP2923, W165BP2924, W165BP2925, W165BP2926, W165BP2927, W165BP2928, W165BP2929, W165BP2930, W165BP2931, W165BP2932, W165BP2933, W165BP2934, W165BP2935, W165BP2936, W165BP2937, W165BP2938, W173BP3088, W173BP3089, W173BP3094, W173BP3099, W172BP3064, W172BP3065, W173BP3090, W173BP3091, W173BP3092, W173BP3093, W173BP3095, W173BP3096, W173BP3097, W173BP3098, W173BP3100, W173BP3101, W173BP3102, W173BP3103, W173BP3104, W173BP3105, W173BP3106, W173BP3108, W173BP3109, W173BP3110, W173BP3111, W173BP3112, W173BP3113, W173BP3114, W173BP3116, W173BP3117, W173BP3118, W173BP3119, W151BP2553, W154BP2653, W173BP3115, W173BP3120, W173BP3121, W173BP3122, W173BP3123, W173BP3124, W173BP3125, W173BP3128, W173BP3129, W179BP3131, W179BP3132, W179BP3133, W179BP3134, W179BP3135, W180BP3217, W180BP3218, W180BP3220, W180BP3221, W180BP3222, W180BP3224, W180BP3225, W180BP3226, W180BP3227, W180BP3228, W180BP3230, W180BP3231, W180BP3232, W180BP3233, W180BP3234, W180BP3235, W180BP3236, W180BP3237, W180BP3238, W180BP3239, W180BP3282, W180BP3283, W180BP3284, W151BP2545, W154BP2656, W169BP2983, W179BP3130, W179BP3171, W179BP3183, W180BP3240, W180BP3241, W180BP3242, W180BP3243, W180BP3244, W180BP3245, W180BP3246, W180BP3247, W180BP3248, W180BP3249, W180BP3250, W180BP3251, W180BP3252, W180BP3253, W180BP3254, W180BP3255, W180BP3257, W180BP3258, W180BP3259, W180BP3260, W180BP3261, W180BP3262, W180BP3263, W180BP3264, W180BP3265, W180BP3266, W180BP3267, W180BP3268, W180BP3269, W180BP3270, W180BP3271, W180BP3272, W180BP3273, W180BP3274, W180BP3275, W180BP3276, W180BP3277, W180BP3286, W183BP3345, W183BP3346, W183BP3347, W183BP3348, W183BP3349, W183BP3350, W183BP3351, W183BP3352, W183BP3353, W183BP3354, W183BP3355, W183BP3356, W183BP3357, W183BP3358, W183BP3361, W183BP3362, W183BP3363, W183BP3364, W183BP3365, W180BP3278, W180BP3279, W190BP3442, W190BP3443, W190BP3444, W190BP3445, W190BP3447, W190BP3448, W190BP3449, W190BP3450, W190BP3494, W190BP3498, W197BP3644, W197BP3654, W197BP3655, W197BP3656, W197BP3657, W197BP3658, W197BP3659, W197BP3660, W197BP3664, W197BP3665, W197BP3666, W197BP3667, W197BP3668, W197BP3669, W197BP3671, W197BP3672, W197BP3673, W197BP3674, W197BP3675, W197BP3684, W197BP3685, W197BP3642, W197BP3646, W197BP3676, W197BP3677, W197BP3678, W197BP3679, W197BP3680, W197BP3681, W225BP4120, W225BP4121, W225BP4123, W225BP4124, W225BP4126, W225BP4129, W225BP4132, W225BP4134, W225BP4135, W225BP4136, W225BP4138, W225BP4139, W225BP4140, W225BP4141, W225BP4142, W225BP4143, W225BP4145, W225BP4161, W225BP4164, W225BP4165, W225BP4166, W225BP4167, W225BP4168, W225BP4169, W225BP4170, W225BP4171, W225BP4172, W225BP4146, W225BP4147, W225BP4148, W225BP4152, W225BP4153, W225BP4154, W225BP4155, W225BP4156, W225BP4157, W225BP4158, W225BP4159, W225BP4160, W225BP4162, W228BP4173, W228BP4174, W228BP4175, W228BP4177, W228BP4179, W228BP4180, W249BP4584, W249BP4590, W249BP4591, W249BP4592, W249BP4593, W249BP4594, W252BP4600, W252BP4601, W252BP4602, W252BP4603, W252BP4604, W252BP4605, W252BP4607, W252BP4608, W252BP4609, W252BP4610, W252BP4611, W252BP4613, W252BP4614, W252BP4615, W252BP4616, W252BP4617, W243BP4475, W243BP4476, W245BP4479, W245BP4480, W245BP4481, W245BP4482, W245BP4483, W245BP4484, W245BP4486, W245BP4487, W245BP4489, W245BP4490, W245BP4491, W245BP4495, W245BP4496, W245BP4497, W245BP4498, W245BP4499, W245BP4500, W252BP4621, W252BP4623, W252BP4624, W252BP4625, W252BP4627, W252BP4628, W252BP4630, W252BP4631, W252BP4632, W252BP4633, W252BP4635, W253BP4637, W253BP4638, W253BP4639, W253BP4640, W253BP4641, W253BP4642, W253BP4644, W253BP4645, W253BP4646, W253BP4647, W253BP4648, W253BP4649, W253BP4650, W253BP4651, W253BP4652, W253BP4653, W253BP4654, W253BP4655, W253BP4656, W253BP4657, W253BP4658, W253BP4659, W253BP4660, W253BP4662, W253BP4663, W253BP4664, W253BP4665, W253BP4666, W258BP4707, W260TP0947, W264BP4795, W264BP4796, W264BP4798, W264BP4799, W264BP4800, W264BP4801, W264BP4802, W264BP4803, W264BP4804, W264BP4805, W264BP4806, W264BP4807, W264BP4809, W264BP4811, W264BP4812, W264BP4813, W264BP4814, W264BP4815, W246BP4539, W249TP0471, W249TP0472, W249TP0473, W249TP0474, W249TP0475, W249TP0476, W249TP0477, W249TP0478, W249TP0479, W249TP0480, W249TP0481, W249TP0482, W249TP0483, W249TP0484, W249TP0485, W249TP0486, W249TP0487, W249TP0488, W249TP0489, W260BP4741, W260BP4755, W260BP4769, W260BP4770, W260BP4771, W260BP4772, W260BP4774, W315BP5917, W315BP5918, W315BP5919, W315BP5920, W316BP5921, W316BP5923, W316BP5928, W316BP5929, W316BP5930, W316BP5932, W316BP5934, W316BP5935, W316BP5936, W316BP5937, W313TP2836, W313TP2837, W313TP2838, W313TP2839, W316BP5925, W316BP5926, W316BP5970, W316BP5971, W316BP5972, W316BP5976, W315BP5886, W315BP5889, W315BP5890, W316BP5978, W316BP5982, W316BP5987, W323BP6089, W323BP6091, W323BP6096, W323BP6105, W323BP6107, W323BP6108, W323BP6112, W323BP6118, W323BP6120, W323BP6121, W323BP6123, W323BP6124, W323BP6125, W323BP6126, W323BP6128, W323BP6129, W323BP6131, W323BP6132, W323BP6114, W323BP6115, W323BP6116, W323BP6135, W323BP6136, W323BP6138, W323BP6139, W323BP6140, W323BP6143, W323BP6144, W323BP6145, W323BP6147, W323BP6149, W323BP6152, W323BP6153, W323BP6154, W323BP6156, W326BP6188, W326BP6190, W327BP6191, W327BP6193, W327BP6196, W327BP6198, W327BP6200, W327BP6214, W327BP6215, W327BP6219, W327BP6249, W327BP6250, W327BP6251, W327BP6255, W327BP6256, W327BP6257, W327BP6258, W327BP6262, W327BP6263, W327BP6264, W327BP6266, W327BP6268, W327BP6269, W327BP6287, W327BP6289, W327BP6291, W327BP6292, W327BP6293, W321TP3168, W321TP3169, W326BP6161, W327BP6230, W327BP6231, W327BP6233, W327BP6234, W327BP6235, W327BP6237, W327BP6238, W327BP6239, W327BP6240, W327BP6241, W327BP6243, W327BP6244, W327BP6259, W327BP6294, W327BP6295, W327BP6297, W328BP6300, W328BP6301, W328BP6302, W328BP6303, W328BP6304, W328BP6305, W328BP6306, W328BP6307, W328BP6308, W328BP6309, W328BP6311, W328BP6317, W328BP6318, W328BP6319, W328BP6320, W328BP6326, W328BP6335, W328BP6337, W328BP6323, W328BP6324, W328BP6325, W328BP6336, W328BP6338, W328BP6339, W328BP6340, W328BP6342, W328BP6344, W328BP6345, W328BP6346, W328BP6347, W328BP6348, W328BP6349, W328BP6352, W328BP6353, W328BP6354, W328BP6355, W328BP6356, W328BP6357, W328BP6358, W328BP6360, W328BP6361, W328BP6362, W330BP6376, W330BP6378, W330BP6382, W330BP6383, W330BP6385, W330BP6387, W330BP6388, W330BP6392, W330BP6393, W330BP6394, W329BP3571, W328BP6331, W328BP6334, W330BP6363, W330BP6364, W330BP6365, W330BP6366, W330BP6367, W330BP6368, W330BP6369, W330BP6371, W330BP6372, W330BP6373, W330BP6374, W330BP6396, W330BP6397, W333BP6405, W333BP6406, W333BP6408, W333BP6409, W333BP6410, W333BP6411, W333BP6412, W333BP6432, W333BP6399, W333BP6400, W333BP6401, W333BP6404, W333BP6415, W333BP6418, W333BP6420, W333BP6421, W333BP6422, W333BP6423, W333BP6424, W333BP6425, W333BP6426, W333BP6427, W333BP6431, W333BP6446, W333BP6447, W333BP6448, W333BP6450, W333BP6451, W333BP6453, W333BP6455, W333BP6460, W333BP6435, W333BP6439, W333BP6440, W333BP6442, W333BP6445, W333BP6452, W333BP6457, W333BP6458, W334BP6468, W334BP6469, W334BP6470, W334BP6471, W334BP6480, W334BP6485, W334BP6486, W334BP6487, W334BP6489, W334BP6492, W334BP6493, W334BP6496, W334BP6500, W334BP6501, W334BP6502, W334BP6504, W334BP6505, W334BP6473, W334BP6475, W334BP6476, W334BP6477, W334BP6478, W334BP6481, W334BP6509, W334BP6512, W334BP6513, W334BP6523, W334BP6524, W334BP6525, W334BP6526, W334BP6527, W334BP6528, W334BP6529, W334BP6530, W334BP6531, W334BP6532, W334BP6533, W334BP6534, W334BP6535, W334BP6536, W334BP6537, W330BP6370, W333BP6456, W333BP6459, W333BP6461, W333BP6462, W333BP6463, W333BP6465, W333BP6466, W333BP6467, W334BP6511, W334BP6514, W334BP6515, W334BP6516, W334BP6519, W337BP6576, W337BP6579, W337BP6580, W337BP6581, W337BP6582, W337BP6583, W337BP6584, W337BP6585, W315BP5894, W315BP5896, W337BP6598, W337BP6609, W337BP6567, W337BP6568, W337BP6569, W337BP6570, W337BP6571, W337BP6572, W337BP6573, W337BP6574, W337BP6575, W337BP6614, W337BP6615, W337BP6621, W337BP6622, W337BP6623, W337BP6624, W337BP6625, W337BP6626, W337BP6634, W337BP6635, W341BP6687, W337BP6565, W337BP6566, W337BP6577, W337BP6578, W337BP6629, W337BP6630, W337BP6631, W337BP6632, W337BP6636, W337BP6637, W337BP6638, W337BP6641, W337BP6642, W337BP6643, W337BP6644, W337BP6645, W341BP6646, W341BP6675, W341BP6676, W341BP6678, W341BP6679, W341BP6681, W341BP6683, W341BP6684, W344BP6688, W344BP6689, W344BP6690, W344BP6692, W344BP6694, W344BP6695, W344BP6698, W344BP6700, W344BP6701, W344BP6704, W344BP6705, W344BP6706, W344BP6707, W344BP6708, W344BP6709, W344BP6710, W344BP6713, W344BP6748, W344BP6749, W344BP6702, W344BP6703;
v) P7924A04, UDI/DI 00887761977877, Serial Numbers: V322BP0046, V330BP0162,
W291BP5345;
w) P7924ARENT01, UDI/DI 00887761976863, Serial Numbers: W112BP1660, W112BP1654, W112BP1656, W112BP1661, W112BP1662, W112BP1663, W112BP1664, W112BP1665, W112BP1666, W112BP1667, W112BP1668, W112BP1669, W112BP1670, W113BP1674, W113BP1675, W113BP1676, W113BP1677, W113BP1678, W113BP1679, W113BP1680, W113BP1681, W113BP1682, W113BP1683, W113BP1686, W113BP1687, W113BP1688, W111BP1619, W111BP1620, W111BP1621, W111BP1625, W111BP1629, W111BP1633, W111BP1639, W113BP1684, W113BP1685, W113BP1689, W113BP1690, W113BP1691, W113BP1692, W113BP1693, W113BP1694, W113BP1695, W113BP1696, W113BP1697, W113BP1698, W113BP1700, W113BP1701, W113BP1703, W113BP1672, W113BP1673, W116BP1704, W116BP1705, W116BP1706, W116BP1707, W116BP1708, W116BP1709, W116BP1710, W116BP1711, W116BP1712, W116BP1713, W116BP1714, W116BP1715, W116BP1716, W116BP1717, W116BP1718, W116BP1719, W116BP1720, W116BP1721, W116BP1722, W118BP1735, W118BP1736, W118BP1737, W118BP1738, W118BP1741, W118BP1742, W113BP1699, W116BP1724, W116BP1725, W116BP1728, W116BP1729, W116BP1730, W116BP1731, W116BP1732, W116BP1733, W118BP1744, W118BP1745, W118BP1746, W118BP1747, W118BP1748, W118BP1793, W118BP1794, W119BP1810, W119BP1811, W119BP1812, W119BP1813, W119BP1802, W119BP1805, W118BP1772, W118BP1773, W118BP1786, W118BP1788, W119BP1795, W119BP1796, W119BP1797, W119BP1798, W119BP1799, W119BP1801, W119BP1803, W119BP1804, W119BP1806, W119BP1807, W119BP1808, W119BP1809, W119BP1814, W119BP1821, W120BP1826, W120BP1827, W120BP1828, W120BP1829, W120BP1830, W120BP1831, W120BP1832, W120BP1833, W120BP1834, W120BP1835, W120BP1836, W120BP1837, W120BP1838, W120BP1839, W120BP1840, W120BP1841, W120BP1842, W120BP1843, W120BP1844, W120BP1845, W120BP1866, W120BP1867, W120BP1868, W119BP1815, W124BP1903, W124BP1904, W124BP1905, W124BP1906, W124BP1907, W124BP1908, W124BP1909, W124BP1910, W124BP1912, W124BP1918, W124BP1920, W124BP1921, W124BP1922, W124BP1923, W124BP1924, W124BP1925, W124BP1927, W124BP1928, W124BP1930, W124BP1931, W124BP1964, W118BP1740, W118BP1770, W118BP1771, W123BP1871, W123BP1872, W123BP1873, W124BP1913, W124BP1914, W124BP1915, W124BP1916, W124BP1917, W124BP1919, W124BP1932, W124BP1933, W124BP1934, W124BP1938, W124BP1942, W124BP1943, W124BP1944, W124BP1945, W126BP1966, W126BP1967, W126BP1975, W126BP1976, W126BP1977, W126BP1978, W126BP1979, W126BP1980, W126BP1981, W126BP1982, W126BP1983, W126BP1984, W126BP1985, W123BP1874, W123BP1875, W123BP1876, W123BP1877, W123BP1878, W124BP1946, W124BP1947, W124BP1954, W124BP1955, W124BP1959, W124BP1960, W124BP1961, W124BP1962, W124BP1963, W126BP1968, W131BP2067, W131BP2068, W131BP2069, W131BP2070, W131BP2072, W131BP2073, W131BP2074, W131BP2075, W131BP2076, W131BP2077, W131BP2084, W131BP2086, W131BP2087, W131BP2088, W131BP2089, W131BP2111, W131BP2112, W131BP2113, W131BP2114, W131BP2115, W131BP2117, W106BP1526, W131BP2091, W131BP2092, W131BP2093, W131BP2094, W131BP2095, W131BP2096, W131BP2097, W131BP2098, W131BP2099, W131BP2100, W131BP2101, W131BP2102, W131BP2103, W131BP2104, W131BP2105, W131BP2106, W131BP2107, W131BP2108, W131BP2109, W131BP2110, W131BP2118, W131BP2119, W131BP2120, W131BP2121, W131BP2123, W132BP2125, W132BP2127, W132BP2128, W132BP2129, W132BP2130, W132BP2131, W132BP2132, W132BP2133, W132BP2134, W132BP2135, W081BP1151, W109BP1581, W127BP2001, W127BP2014, W130BP2037, W130BP2038, W130BP2039, W130BP2040, W130BP2041, W130BP2043, W130BP2045, W130BP2047, W130BP2048, W130BP2049, W130BP2050, W130BP2051, W130BP2052, W130BP2053, W130BP2054, W130BP2055, W130BP2056, W134BP2151, W134BP2152, W134BP2153, W134BP2154, W134BP2136, W134BP2159, W134BP2160, W134BP2161, W134BP2162, W134BP2163, W134BP2164, W134BP2165, W134BP2166, W134BP2167, W134BP2168, W134BP2169, W134BP2170, W134BP2171, W134BP2172, W134BP2173, W134BP2174, W137BP2176, W137BP2177, W127BP2005, W127BP2006, W127BP2007, W127BP2008, W127BP2010, W127BP2011, W127BP2012, W127BP2013, W137BP2182, W137BP2183, W137BP2184, W137BP2211, W137BP2212, W137BP2213, W137BP2215, W137BP2216, W137BP2217, W137BP2218, W137BP2219, W137BP2220, W137BP2221, W137BP2222, W137BP2223, W138BP2231, W138BP2232, W138BP2238, W138BP2239, W138BP2240, W138BP2241, W138BP2242, W138BP2243, W138BP2244, W145BP2415, W145BP2416, W145BP2417, W145BP2418, W145BP2419, W145BP2420, W145BP2424, W145BP2425, W145BP2427, W145BP2428, W145BP2429, W145BP2430, W145BP2431, W145BP2432, W145BP2433, W145BP2434, W145BP2435, W145BP2441, W145BP2444, W145BP2448, W145BP2449, W145BP2451, W145BP2452, W145BP2453, W145BP2456, W145BP2457, W140BP2329, W140BP2330, W140BP2331, W141BP2334, W141BP2349, W141BP2350, W145BP2437, W145BP2438, W145BP2439, W145BP2442, W145BP2447, W145BP2458, W145BP2459, W145BP2460, W145BP2461, W145BP2462, W145BP2463, W145BP2464, W145BP2465, W145BP2466, W145BP2467, W145BP2469, W147BP2471, W147BP2472, W147BP2473, W147BP2474, W147BP2475, W147BP2476, W147BP2477, W147BP2478, W224BP4105, W224BP4106, W224BP4107, W224BP4108, W224BP4109, W224BP4110, W224BP4111, W214BP3944, W223BP4078, W224BP4094, W224BP4095, W224BP4096, W224BP4097, W224BP4098, W224BP4100, W224BP4112, W224BP4113, W224BP4114, W224BP4115, W224BP4116, W196BP3620, W223BP4070, W223BP4071, W223BP4072, W223BP4073, W223BP4074, W223BP4075, W223BP4076, W223BP4077, W223BP4080, W223BP4081, W223BP4082, W223BP4083, W223BP4084, W223BP4085, W223BP4086, W223BP4087, W223BP4088, W223BP4090, W223BP4092, W223BP4093, W225BP4122, W225BP4127, W225BP4130, W225BP4117, W225BP4119, W231BP4206, W231BP4207, W231BP4208, W231BP4209, W231BP4210, W231BP4211, W231BP4212, W231BP4214, W231BP4215, W231BP4216, W231BP4217, W231BP4218, W231BP4219, W231BP4220, W231BP4221, W231BP4222, W231BP4223, W231BP4224, W231BP4227, W231BP4213, W231BP4228, W231BP4229, W231BP4230, W231BP4233, W231BP4234, W231BP4241, W231BP4242, W231BP4243, W231BP4244, W231BP4245, W231BP4246, W231BP4247, W231BP4248, W231BP4249, W231BP4250, W231BP4251, W231BP4252, W231BP4253, W235BP4259, W235BP4260, W235BP4265, W235BP4266, W231BP4254, W231BP4255, W231BP4256, W231BP4257, W235BP4262, W235BP4263, W235BP4264, W235BP4270, W235BP4271, W235BP4272, W235BP4278, W235BP4279, W239BP4371, W239BP4372, W239BP4373, W239BP4374, W239BP4375, W242BP4378, W242BP4379, W242BP4380, W242BP4382, W242BP4383, W242BP4384, W242BP4385, W242BP4386, W236TP9806, W236TP9807, W236TP9808, W236TP9809, W236TP9810, W236TP9811, W236TP9812, W242BP4381, W242BP4387, W242BP4389, W242BP4391, W242BP4392, W242BP4393, W242BP4394, W242BP4395, W242BP4396, W242BP4397, W242BP4398, W242BP4399, W242BP4400, W242BP4401, W242BP4402, W242BP4403, W242BP4404, W242BP4405, W242BP4406, W242BP4407, W242BP4408, W242BP4409, W242BP4410, W243BP4423, W243TP0224, W243TP0225, W243TP0226, W235BP4291, W242BP4390, W243BP4412, W243BP4413, W243BP4414, W243BP4415, W243BP4416, W243BP4418, W243BP4419, W243BP4420, W243BP4421, W243BP4422, W243BP4425, W243BP4428, W243BP4455, W249BP4574, W249BP4575, W249BP4576, W249BP4577, W249BP4578, W249BP4579, W249BP4580, W249BP4581, W249BP4583, W249BP4585, W249BP4586, W249BP4587, W249BP4588, W252BP4595, W252BP4596, W252BP4597, W252BP4598, W252BP4599, W243BP4424, W249BP4589, W267BP4831, W267BP4832, W267BP4834, W267BP4835, W267BP4836, W267BP4838, W267BP4839, W267BP4840, W267BP4842, W267BP4844, W267BP4845, W267BP4850, W267BP4853, W267BP4855, W267BP4856, W267BP4857, W267BP4860, W267BP4861, W267BP4863, W267BP4864, W267BP4865, W267BP4866, W252BP4619, W260BP4745, W267BP4867, W267BP4868, W267BP4870, W267BP4871, W267BP4872, W267BP4873, W267BP4874, W267BP4875, W267BP4879, W267BP4880, W267BP4884, W267BP4885, W267BP4886, W267BP4888, W267BP4889, W267BP4890, W267BP4891, W267BP4892, W267BP4893, W267BP4894, W267BP4895, W267BP4896, W267BP4898, W267BP4899, W267BP4902, W270BP4923, W270BP4903, W270BP4904, W270BP4905, W270BP4906, W270BP4907, W270BP4908, W270BP4909, W270BP4910, W270BP4911, W270BP4912, W270BP4913, W270BP4914, W270BP4915, W270BP4916, W270BP4917, W270BP4918, W270BP4919, W270BP4920, W270BP4921, W270BP4922, W270BP4924, W270BP4925, W270BP4926, W270BP4927, W270BP4928, W270BP4929, W270BP4945, W270BP4946, W270BP4947, W270BP4948, W270BP4949, W270BP4950, W270BP4951, W270BP4953, W270BP4954, W270BP4955, W270BP4956, W270BP4957, W271BP4958, W271BP4964, W271BP4982, W271BP4983, W271BP4984, W271BP4985, W271BP4986, W271BP4987, W271BP4988, W271BP4989, W271BP4993, W271BP4994, W271BP4995, W271BP4996, W260BP4779, W270BP4930, W270BP4931, W270BP4932, W270BP4934, W270BP4935, W270BP4937, W270BP4938, W270BP4939, W270BP4940, W270BP4941, W270BP4942, W270BP4943, W270BP4944, W271BP4661, W271BP4960, W271BP4961, W271BP4965, W271BP4967, W271BP4969, W271BP4973, W271BP4976, W271BP4977, W271BP4978, W271BP4979, W271BP4981, W273BP5008, W273BP4998, W273BP4999, W271BP4971, W274BP5080, W277BP5083, W277BP5085, W277BP5086, W277BP5087, W277BP5088, W277BP5090, W277BP5091, W277BP5092, W277BP5093, W277BP5094, W277BP5095, W277BP5096, W277BP5097, W277BP5098, W277BP5099, W277BP5100, W277BP5101, W277BP5102, W277BP5105, W277BP5106, W277BP5110, W277BP5111, W277BP5112, W277BP5113, W277BP5114, W278BP5116, W278BP5117, W278BP5118, W278BP5119, W278BP5120, W278BP5121, W278BP5122, W278BP5123, W278BP5124, W278BP5125, W278BP5126, W278BP5127, W278BP5128, W278BP5129, W278BP5130, W278BP5131, W278BP5132, W278BP5135, W278BP5136, W278BP5137, W278BP5138, W278BP5139, W278BP5140, W278BP5141, W278BP5142, W278BP5143, W284BP5146, W284BP5147, W284BP5148, W284BP5149, W284BP5151, W284BP5155, W284BP5156, W284BP5157, W284BP5158, W284BP5160, W284BP5162, W284BP5165, W284BP5166, W284BP5167, W284BP5169, W284BP5170, W284BP5172, W284BP5173, W284BP5174, W284BP5175, W284BP5176, W284BP5178, W284BP5179, W284BP5180, W284BP5181, W284BP5182, W284BP5183, W284BP5184, W284BP5185, W284BP5186, W284BP5187, W284BP5188, W284BP5189, W284BP5190, W284BP5191, W284BP5192, W284BP5193, W284BP5194, W284BP5195, W284BP5196, W284BP5197, W284BP5198, W284BP5199, W284BP5200, W284BP5201, W284BP5202, W284BP5203, W284BP5204, W284BP5205, W284BP5206, W285BP5208, W285BP5209, W285BP5210, W285BP5214, W285BP5216, W285BP5237, W285BP5211, W285BP5212, W285BP5213, W285BP5217, W285BP5218, W285BP5219, W285BP5220, W285BP5222, W285BP5223, W285BP5224, W285BP5225, W285BP5227, W285BP5228, W285BP5229, W285BP5230, W285BP5231, W285BP5232, W285BP5233, W285BP5235, W285BP5239, W285BP5240, W285BP5241, W285BP5242, W285BP5243, W286BP5267, W286BP5268, W286BP5269, W285BP5221, W285BP5226, W285BP5234, W285BP5236, W285BP5244, W285BP5245, W285BP5246, W285BP5247, W285BP5248, W285BP5249, W285BP5250, W285BP5255, W285BP5256, W285BP5257, W285BP5261, W285BP5262, W285BP5264, W285BP5265, W286BP5270, W286BP5271, W286BP5273, W286BP5274, W286BP5276, W286BP5277, W286BP5278, W285BP5252, W285BP5253, W285BP5254, W285BP5258, W285BP5260, W286BP5279, W286BP5280, W286BP5282, W286BP5283, W286BP5284, W286BP5285, W286BP5286, W286BP5287, W286BP5288, W286BP5290, W286BP5291, W286BP5292, W286BP5293, W286BP5294, W286BP5295, W287BP5297, W287BP5298, W287BP5299, W286TP1993, W286TP1994, W286TP1995, W287BP5303, W287BP5304, W287BP5305, W287BP5306, W287BP5307, W287BP5310, W287BP5311, W287BP5312, W287BP5313, W287BP5317, W287BP5318, W287BP5319, W287BP5320, W287BP5321, W287BP5322, W291BP5329, W291BP5335, W291BP5336, W291BP5337, W291BP5338, W291BP5339, W291BP5340, W291BP5341, W291BP5374, W291BP5380, W307BP5827, W287BP5308, W287BP5309, W287BP5314, W287BP5315, W287BP5316, W291BP5343, W291BP5368, W291BP5369, W291BP5370, W291BP5372, W291BP5373, W291BP5375, W291BP5376, W291BP5377, W291BP5378, W291BP5379, W292BP5387, W292BP5388, W292BP5389, W292BP5390, W292BP5391, W292BP5395, W292BP5397, W292BP5398, W292BP5399, W292BP5400, W291BP5323, W291BP5324, W291BP5325, W291BP5326, W291BP5328, W291BP5330, W291BP5331, W291BP5332, W291BP5344, W294BP5491, W294BP5492, W294BP5493, W294BP5496, W294BP5498, W292BP5422, W292BP5431, W294BP5494, W294BP5495, W298BP5525, W298BP5526, W298BP5527, W298BP5528, W298BP5529, W298BP5530, W298BP5531, W298BP5533, W298BP5535, W298BP5536, W298BP5537, W298BP5538, W298BP5542, W298BP5543, W298BP5544, W298BP5545, W298BP5547, W298BP5550, W298BP5551, W298BP5552, W298BP5453, W298BP5464, W298BP5534, W298BP5539, W298BP5540, W298BP5546, W298BP5553, W298BP5555, W298BP5558, W298BP5559, W298BP5560, W298BP5561, W298BP5562, W298BP5563, W298BP5565, W298BP5566, W298BP5567, W298BP5568, W298BP5569, W298BP5570, W298BP5573, W298BP5574, W298BP5575, W298BP5576, W298BP5578, W298BP5579, W298BP5580, W298BP5581, W298TP2457, W298TP2458, W298TP2459, W298TP2460, W298TP2461, W298TP2462, W292BP5435, W292BP5439, W292BP5440, W292BP5444, W292BP5446, W292BP5448, W292BP5451, W292BP5454, W292BP5459, W292BP5460, W292BP5461, W292BP5462, W292BP5465, W298BP5556, W298BP5557, W298BP5571, W298BP5572, W298BP5583, W298BP5585, W298BP5586, W298BP5588, W298BP5589, W298BP5590, W298BP5591, W298BP5592, W298BP5593, W298BP5595, W299BP5602, W299BP5603, W299BP5604, W299BP5605, W299BP5606, W306BP5761, W306BP5763, W306BP5764, W306BP5765, W306BP5766, W306BP5767, W306BP5768, W306BP5769, W306BP5771, W306BP5772, W306BP5773, W307BP5779, W307BP5782, W307BP5783, W307BP5785, W307BP5786, W307BP5787, W307BP5788, W307BP5790, W307BP5793, W307BP5794, W307BP5795, W315BP5913, W315BP5914, W315BP5915, W315BP5916, W354BP6923
|
Recalling Firm/ Manufacturer |
Baxter Healthcare Corporation 1 Baxter Pkwy Deerfield IL 60015-4625
|
For Additional Information Contact |
Center for One Baxter 800-422-9837
|
Manufacturer Reason for Recall |
There is a potential for mattress cover damage (delamination) which may result in fluid absorption in the mattress (fluid ingress).
|
FDA Determined Cause 2 |
Process change control |
Action |
Baxter issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 03/31/2023 by US mail. The notice explained the issue and the risk and requested the following actions be taken:
Identify the affected product and examine the cover for damage, if damaged, discontinue use. (undamaged device may be safely used) The mattresses will be replaced. If the product was further distributed please forward a copy of the notice. For questions, contact Baxter Technical support team at 1-800-445-3720 option 2, Monday to Friday, between the hours of 8am to 6pm EST. |
Quantity in Commerce |
5549 units |
Distribution |
Worldwide distribution - US Nationwide and the countries of Canada, Qatar. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|