• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Aspira Pleural Drainage Catheter Insertion Tray

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Aspira Pleural Drainage Catheter Insertion Traysee related information
Date Initiated by FirmApril 22, 2023
Date PostedJune 14, 2023
Recall Status1 Open3, Classified
Recall NumberZ-1915-2023
Recall Event ID 92301
510(K)NumberK110396 K212675 
Product Classification Peritoneal, drainage catheter for refractory ascites, long-term indwelling - Product Code PNG
ProductAspira Pleural Drainage Catheter Insertion Tray 15.5F, REF: 4992509, Version A, Sterile EO, Rx Only
Code Information Lot # I2614171, I2559314, I2546596, I2614170, I2538442, I2564679, I2548181, I2623231; UDI-DI: 00884450394946
FEI Number 1721504
Recalling Firm/
Manufacturer
Merit Medical Systems, Inc.
1600 W Merit Pkwy
South Jordan UT 84095-2416
For Additional Information ContactMerit Customer Service
801-253-1600
Manufacturer Reason
for Recall
There is a potential that valve assemblies will not open. preventing fluid from draining.
FDA Determined
Cause 2
Process control
ActionBeginning on April 22, 2023, Merit Medical issued an "Urgent Medical Device Recall Notice" to affected consignees via FedEx and UPS. On May 31, 2023, Merit Medical expanded the scope and notified additional consignees. Merit Medical asked consignees to take the following actions: 1. Please immediately determine if any of the devices identified in the attached Customer Response Form (CRF) are within your facility, quarantine them, and discontinue use and distribution. 2. Ensure that applicable personnel within your organization are made aware of this field action. 3. If the product has been further distributed to other facilities, institutions, or manufacturers, please ensure this notice is immediately shared with them and note the quantity distributed on the CRF. Additional distribution details may be required by health authorities. 4. For Aspira Valve Replacement Kits, please immediately return all affected product in your possession to Merit, per the instructions in the attached CRF. 5. For Aspira Insertion Trays, follow the attached instructions for placing the recall labels. At the point of use, the valve assembly in the Insertion Tray is to be discarded and replaced by one of the replacement Aspira Valve Repair Kits. 6. Please fill out, scan and email the completed Customer Response Form to Customer Service at response@merit.com within seven (7) days. All affected product shipped to you must be accounted for on the CRF. 7. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax.
Quantity in Commerce1054 units
DistributionWorldwide - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, Puerto Rico and the countries of France, United Kingdom, Mexico.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = PNG
510(K)s with Product Code = PNG
-
-