• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall medtronic

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall medtronicsee related information
Date Initiated by FirmMay 10, 2023
Date PostedJune 16, 2023
Recall Status1 Open3, Classified
Recall NumberZ-1709-2023
Recall Event ID 92328
PMA NumberP980016S382 
Product Classification Implantable cardioverter defibrillator (non-CRT) - Product Code LWS
ProductICD-DR DDBC3D1 EVERA S IS1/DF1 GLOB, Model Number DDBC3D1; Implantable Cardioverter Defibrillators
Code Information GTIN 00643169720411, Lot Serial Numbers: BWG208720H, BWG208284H, BWG208405H, BWG208608H, BWG208700H, BWG208763H, BWG208301H, BWG208376H, BWG208406H, BWG208412H, BWG208520H, BWG208539H, BWG208652H, BWG208716H, BWG208296H, BWG208340H, BWG208484H, BWG208220H, BWG208234H, BWG208641H, BWG208702H, BWG208758H, BWG208430H, BWG208206H, BWG208215H, BWG208561H, BWG208216H, BWG208705H, BWG208502H, BWG208221H, BWG208236H, BWG208488H, BWG208403H, BWG208312H, BWG208557H, BWG208335H, BWG208530H, BWG208634H, BWG208745H, BWG208746H, BWG208249H, BWG208654H, BWG208291H, BWG208515H, BWG208596H, BWG208636H, BWG208637H, BWG208730H, BWG208757H, BWG208213H, BWG208347H, BWG208597H, BWG208605H, BWG208692H, BWG208514H, BWG208640H, BWG208326H, BWG208499H, BWG208223H, BWG208491H, BWG208541H, BWG208573H, BWG208304H, BWG208323H, BWG208345H, BWG208370H, BWG208672H, BWG208250H, BWG208270H, BWG208346H, BWG208389H, BWG208482H, BWG208494H, BWG208495H, BWG208497H, BWG208498H, BWG208521H, BWG208547H, BWG208396H, BWG208799H, BWG208224H, BWG208225H, BWG208258H, BWG208261H, BWG208272H, BWG208275H, BWG208276H, BWG208279H, BWG208328H, BWG208337H, BWG208343H, BWG208357H, BWG208360H, BWG208361H, BWG208416H, BWG208417H, BWG208422H, BWG208423H, BWG208462H, BWG208464H, BWG208465H, BWG208468H, BWG208469H, BWG208470H, BWG208471H, BWG208472H, BWG208581H, BWG208583H, BWG208584H, BWG208585H, BWG208586H, BWG208587H, BWG208588H, BWG208639H, BWG208647H, BWG208649H, BWG208660H, BWG208661H, BWG208663H, BWG208664H, BWG208668H, BWG208674H, BWG208770H, BWG208774H, BWG208775H, BWG208777H, BWG208778H, BWG208779H, BWG208780H, BWG208781H, BWG208421H, BWG208452H, BWG208486H, BWG208493H, BWG208504H, BWG208571H, BWG208614H, BWG208543H, BWG208549H, BWG208681H, BWG208544H, BWG208002H, BWG208212H, BWG208359H, BWG208373H, BWG208413H, BWG208447H, BWG208232H, BWG208252H, BWG208205H, BWG208386H, BWG208540H, BWG208575H, BWG208631H, BWG208769H, BWG208306H, BWG208768H, BWG208273H, BWG208689H, BWG208742H, BWG208798H, BWG208227H, BWG208796H, BWG208797H, BWG208475H, BWG208552H, BWG208365H, BWG208390H, BWG208350H, BWG208435H, BWG208562H, BWG208680H, BWG208691H, BWG208762H, BWG208208H, BWG208247H, BWG208300H, BWG208358H, BWG208397H, BWG208523H, BWG208525H, BWG208592H, BWG208599H, BWG208671H, BWG208682H, BWG208704H, BWG208707H, BWG208718H, BWG208428H, BWG208432H, BWG208648H, BWG208444H, BWG208559H, BWG208533H, BWG208644H, BWG208744H, BWG208603H, BWG208203H, BWG208260H, BWG208271H, BWG208384H, BWG208407H, BWG208489H, BWG208005H, BWG208694H, BWG208548H, BWG208650H, BWG208725H, BWG208253H, BWG208394H, BWG208446H, BWG208448H, BWG208454H, BWG208503H, BWG208579H, BWG208612H, BWG208723H, BWG208589H, BWG208748H, BWG208307H, BWG208202H, BWG208214H, BWG208283H, BWG208311H, BWG208314H, BWG208322H, BWG208352H, BWG208374H, BWG208496H, BWG208572H, BWG208594H, BWG208613H, BWG208629H, BWG208675H, BWG208727H, BWG208732H, BWG208737H, BWG208761H, BWG208756H, BWG208290H, BWG208297H, BWG208298H, BWG208342H, BWG208367H, BWG208393H, BWG208478H, BWG208616H, BWG208657H, BWG208703H, BWG208371H, BWG208210H, BWG208606H, BWG208715H, BWG208754H, BWG208353H, BWG208560H, BWG208209H, BWG208395H, BWG208240H, BWG208372H, BWG208431H, BWG208481H, BWG208551H, BWG208688H, BWG208517H, BWG208563H, BWG208404H, BWG208512H, BWG208266H, BWG208248H, BWG208665H, BWG208259H, BWG208449H, BWG208458H, BWG208511H, BWG208537H, BWG208564H, BWG208651H, BWG208710H, BWG208436H, BWG208455H, BWG208638H, BWG208235H, BWG208242H, BWG208318H, BWG208377H, BWG208415H, BWG208466H, BWG208467H, BWG208653H, BWG208246H, BWG208320H, BWG208535H, BWG208536H, BWG208626H, BWG208628H, BWG208632H, BWG208721H, BWG208793H, BWG208645H, BWG208408H, BWG208659H, BWG208501H, BWG208427H, BWG208627H, BWG208566H, BWG208595H, BWG208255H, BWG208643H, BWG208387H, BWG208233H, BWG208257H, BWG208375H, BWG208555H, BWG208726H, BWG208003H, BWG208329H, BWG208381H, BWG208383H, BWG208401H, BWG208556H, BWG208263H, BWG208302H, BWG208324H, BWG208341H, BWG208400H, BWG208696H, BWG208736H, BWG208751H, BWG208795H, BWG208568H, BWG208580H, BWG208254H, BWG208600H, BWG208348H, BWG208630H, BWG208633H, BWG208701H, BWG208461H, BWG208522H, BWG208711H, BWG208327H, BWG208743H, BWG208474H, BWG208593H, BWG208256H, BWG208268H, BWG208309H, BWG208319H, BWG208363H, BWG208364H, BWG208414H, BWG208463H, BWG208574H, BWG208611H, BWG208623H, BWG208642H, BWG208655H, BWG208666H, BWG208695H, BWG208755H, BWG208658H, BWG208379H, BWG208439H, BWG208534H, BWG208610H, BWG208617H, BWG208490H, BWG208457H, BWG208262H, BWG208382H, BWG208477H, BWG208554H, BWG208625H, BWG208391H, BWG208459H, BWG208529H, BWG208531H, BWG208226H, BWG208339H, BWG208604H, BWG208624H, BWG208699H, BWG208479H, BWG208697H, BWG208228H, BWG208418H, BWG208419H, BWG208450H, BWG208684H, BWG208294H, BWG208706H, BWG208285H, BWG208218H, BWG208378H, BWG208456H, BWG208569H, BWG208601H, BWG208618H, BWG208622H, BWG208683H, BWG208709H, BWG208747H, BWG208292H, BWG208295H, BWG208440H, BWG208558H, BWG208669H, BWG208679H, BWG208685H, BWG208687H, BWG208712H, BWG208713H, BWG208714H, BWG208719H, BWG208728H, BWG208765H, BWG208338H, BWG208546H, BWG208615H, BWG208487H; GTIN 00643169720428, Lot Serial Numbers: BWG606752S, BWG606481S, BWG606334S, BWG606796S, BWG606352S, BWG606515S, BWG606384S, BWG606530S, BWG606659S, BWG606661S, BWG606685S, BWG606508S, BWG606697S, BWG606689S, BWG606501S, BWG606355S, BWG606472S, BWG606512S, BWG606527S, BWG606535S, BWG606547S, BWG606558S, BWG606585S, BWG606588S, BWG606598S, BWG606691S, BWG606713S, BWG606732S, BWG606795S, BWG606741S, BWG606572S, BWG606345S, BWG606462S, BWG606390S, BWG606410S, BWG606421S, BWG606522S, BWG606531S, BWG606534S, BWG606589S, BWG606607S, BWG607320S, BWG606326S, BWG606347S, BWG606357S, BWG606409S, BWG606517S, BWG606519S, BWG606657S, BWG606513S, BWG606493S, BWG606363S, BWG606337S, BWG606690S, BWG606356S, BWG606411S, BWG606536S, BWG606537S, BWG606665S, BWG606596S, BWG606539S, BWG606580S, BWG606616S, BWG606358S, BWG606351S, BWG606348S, BWG606581S, BWG606584S, BWG606407S, BWG606757S, BWG606777S, BWG606794S, BWG606422S, BWG606502S, BWG606532S, BWG606615S, BWG606524S, BWG606538S, BWG606593S, BWG606672S, BWG606431S, BWG606468S, BWG606556S, BWG606586S, BWG606591S, BWG606595S, BWG606817S, BWG606629S, BWG606550S, BWG606381S, BWG606671S, BWG606336S, BWG606712S, BWG606544S, BWG606625S, BWG606751S, BWG606617S, BWG606801S, BWG606896S, BWG606571S, BWG606573S, BWG606745S, BWG606765S, BWG606774S, BWG606780S, BWG606798S, BWG606899S, BWG606553S, BWG606569S, BWG606343S, BWG606561S, BWG606557S, BWG606403S, BWG606555S, BWG606582S, BWG606749S, BWG606788S, BWG606420S, BWG606663S, BWG606701S, BWG606562S, BWG606563S, BWG606566S, BWG606567S, BWG606323S, BWG606330S, BWG606332S, BWG606339S, BWG606354S, BWG606360S, BWG606368S, BWG606369S, BWG606371S, BWG606526S, BWG606533S, BWG606546S, BWG606548S, BWG606549S, BWG606570S, BWG606574S, BWG606597S, BWG606600S, BWG606743S, BWG606748S, BWG606762S, BWG606764S, BWG606791S, BWG606523S, BWG606529S, BWG607296S, BWG607305S, BWG606623S, BWG606495S, BWG606498S, BWG606415S, BWG606327S, BWG606367S, BWG606692S, BWG606695S, BWG606399S, BWG606401S, BWG606406S, BWG606412S, BWG606413S, BWG606437S, BWG606554S, BWG606733S, BWG607421S, BWG606423S, BWG606424S, BWG606670S, BWG606677S, BWG606687S, BWG606693S, BWG606694S, BWG606699S, BWG606551S, BWG606329S, BWG606698S, BWG606325S, BWG606335S, BWG606426S, BWG606753S, BWG606362S, BWG607156S, BWG606380S, BWG606475S, BWG606487S, BWG606521S, BWG606577S, BWG606579S, BWG606746S, BWG606747S, BWG606790S, BWG606338S, BWG606425S, BWG606379S, BWG606417S, BWG606394S, BWG606395S, BWG606396S, BWG606397S, BWG606398S, BWG606477S, BWG606478S, BWG606480S, BWG606485S, BWG606489S, BWG606491S, BWG606714S, BWG606717S, BWG606721S, BWG606723S, BWG606724S, BWG606728S, BWG606612S, BWG606614S, BWG606619S, BWG606622S, BWG606630S, BWG606353S, BWG606387S, BWG606400S, BWG606414S, BWG606450S, BWG606453S, BWG606653S, BWG606675S, BWG606730S, BWG606758S, BWG606769S, BWG606775S, BWG606776S, BWG606779S, BWG606797S, BWG606800S, BWG606803S, BWG606804S, BWG606815S, BWG606832S, BWG606834S, BWG606853S, BWG606854S, BWG606859S, BWG606871S, BWG606876S, BWG606888S, BWG606889S, BWG606891S, BWG606892S, BWG606903S, BWG607009S, BWG607011S, BWG607186S, BWG607191S, BWG607192S, BWG607196S, BWG607203S, BWG607204S, BWG607276S, BWG607293S, BWG607313S, BWG607319S, BWG607323S, BWG607413S, BWG607461S, BWG607467S, BWG607473S, BWG607523S, BWG607525S, BWG607527S, BWG607531S, BWG607534S, BWG607581S, BWG607585S, BWG607593S, BWG607598S, BWG607619S, BWG607628S, BWG607632S, BWG607637S, BWG607651S, BWG606781S, BWG606783S, BWG606784S, BWG606785S, BWG606786S, BWG606497S, BWG606500S, BWG606664S, BWG606812S, BWG606823S, BWG606826S, BWG606839S, BWG606873S, BWG606877S, BWG606885S, BWG606893S, BWG606894S, BWG606902S, BWG607006S, BWG607071S, BWG607205S, BWG607208S, BWG607212S, BWG607280S, BWG607315S, BWG607404S, BWG607434S, BWG607436S, BWG607450S, BWG607455S, BWG607463S, BWG607520S, BWG607575S, BWG607580S, BWG607583S, BWG607586S, BWG607589S, BWG607591S, BWG607597S, BWG607599S, BWG607602S, BWG607610S, BWG607612S, BWG607613S, BWG607618S, BWG607634S, BWG607650S, BWG606408S, BWG606759S, BWG606760S, BWG606766S, BWG606767S, BWG606787S, BWG606802S, BWG606806S, BWG606808S, BWG606813S, BWG606819S, BWG606820S, BWG606821S, BWG606827S, BWG606836S, BWG606883S, BWG607012S, BWG607014S, BWG607195S, BWG607199S, BWG607214S, BWG607215S, BWG607278S, BWG607312S, BWG607385S, BWG607435S, BWG607441S, BWG607471S, BWG607472S, BWG607508S, BWG607519S, BWG607522S, BWG607528S, BWG607530S, BWG607532S, BWG607578S, BWG607588S, BWG607603S, BWG607604S, BWG607624S, BWG607635S, BWG606469S, BWG606668S, BWG606666S, BWG606754S, BWG606649S, BWG606709S, BWG606727S, BWG606476S, BWG606887S, BWG606383S, BWG606386S, BWG606389S, BWG606392S, BWG606393S, BWG606602S, BWG606603S, BWG606608S, BWG606609S, BWG606611S, BWG606620S, BWG606686S, BWG606688S, BWG606722S, BWG606725S, BWG606739S, BWG606740S, BWG607020S, BWG607023S, BWG607024S, BWG607046S, BWG607047S, BWG607048S, BWG607187S, BWG607188S, BWG607189S, BWG607209S, BWG607226S, BWG607227S, BWG607228S, BWG607229S, BWG607235S, BWG607277S, BWG607285S, BWG607292S, BWG607419S, BWG607495S, BWG607496S, BWG607501S, BWG607502S, BWG607509S, BWG607590S, BWG607630S, BWG607633S, BWG607638S, BWG607639S, BWG607734S, BWG607739S, BWG607742S, BWG607743S, BWG607896S, BWG607897S, BWG607898S, BWG607899S, BWG607900S, BWG607902S, BWG607912S, BWG607913S, BWG608016S, BWG608021S, BWG608024S, BWG608025S, BWG608026S, BWG608035S, BWG608036S, BWG608037S, BWG608038S, BWG608044S, BWG608046S, BWG608047S, BWG608048S, BWG606377S, BWG606456S, BWG606457S, BWG606463S, BWG606520S, BWG606621S, BWG606628S, BWG606631S, BWG606637S, BWG606640S, BWG606641S, BWG606642S, BWG606643S, BWG606645S, BWG606647S, BWG606648S, BWG606696S, BWG606700S, BWG606707S, BWG606708S, BWG606710S, BWG606715S, BWG606729S, BWG606731S, BWG607018S, BWG607019S, BWG607049S, BWG607050S, BWG607051S, BWG607052S, BWG607053S, BWG607157S, BWG607158S, BWG607160S, BWG607161S, BWG607220S, BWG607221S, BWG607283S, BWG607286S, BWG607287S, BWG607288S, BWG607291S, BWG607298S, BWG607299S, BWG607301S, BWG607302S, BWG607307S, BWG607386S, BWG607395S, BWG607396S, BWG607397S, BWG607398S, BWG607400S, BWG607446S, BWG607451S, BWG607462S, BWG607494S, BWG607498S, BWG607499S, BWG607500S, BWG607503S, BWG607504S, BWG607511S, BWG607513S, BWG607514S, BWG607515S, BWG607521S, BWG607608S, BWG607620S, BWG607621S, BWG607622S, BWG607625S, BWG607626S, BWG607627S, BWG607636S, BWG607642S, BWG607643S, BWG607649S, BWG607652S, BWG607653S, BWG607746S, BWG607747S, BWG607748S, BWG607749S, BWG607750S, BWG607751S, BWG607922S, BWG607923S, BWG607927S, BWG607928S, BWG607929S, BWG607930S, BWG607931S, BWG607932S, BWG607933S, BWG607934S, BWG607935S, BWG607936S, BWG607937S, BWG607938S, BWG607939S, BWG607940S, BWG607941S, BWG606321S, BWG606328S, BWG606331S, BWG606333S, BWG606378S, BWG606518S, BWG606373S, BWG606320S, BWG606404S, BWG606405S, BWG606416S, BWG606418S, BWG606419S, BWG606897S, BWG607402S, BWG607403S, BWG607418S, BWG606654S, BWG606900S, BWG606905S, BWG606906S, BWG607171S, BWG606576S, BWG606503S, BWG606744S, BWG606651S, BWG606488S, BWG606711S, BWG606552S, BWG606568S, BWG606799S, BWG606793S, BWG606440S, BWG606441S, BWG606442S, BWG606443S, BWG606447S, BWG606459S, BWG606587S, BWG606592S, BWG606601S, BWG606627S, BWG606662S, BWG606683S, BWG606684S, BWG606702S, BWG606703S, BWG606704S, BWG606705S, BWG606726S, BWG607007S, BWG607015S, BWG607054S, BWG607055S, BWG607066S, BWG607067S, BWG607068S, BWG607069S, BWG607070S, BWG607072S, BWG607073S, BWG607074S, BWG607162S, BWG607166S, BWG607167S, BWG607173S, BWG607190S, BWG607194S, BWG607198S, BWG607230S, BWG607231S, BWG607232S, BWG607390S, BWG607392S, BWG607439S, BWG607444S, BWG607452S, BWG607458S, BWG607497S, BWG607505S, BWG607507S, BWG607529S, BWG607577S, BWG607582S, BWG607584S, BWG607617S, BWG607644S, BWG607645S, BWG607647S, BWG607735S, BWG607736S, BWG607737S, BWG607738S, BWG607740S, BWG607741S, BWG607744S, BWG607745S, BWG607919S, BWG607920S, BWG607921S, BWG607950S, BWG606782S, BWG606516S, BWG607197S, BWG607200S, BWG606372S, BWG606375S, BWG606445S, BWG606446S, BWG606451S, BWG606452S, BWG606454S, BWG606455S, BWG606458S, BWG606460S, BWG606461S, BWG606624S, BWG606626S, BWG606322S, BWG606706S, BWG606890S, BWG606898S, BWG606341S, BWG606359S, BWG606385S, BWG606402S, BWG606429S, BWG606435S, BWG606439S, BWG606448S, BWG606449S, BWG606482S, BWG606483S, BWG606490S, BWG606494S, BWG606761S, BWG606763S, BWG606768S, BWG606770S, BWG606771S, BWG606792S, BWG606805S, BWG606809S, BWG606810S, BWG606811S, BWG606816S, BWG606822S, BWG606824S, BWG606830S, BWG606831S, BWG606835S, BWG606837S, BWG606838S, BWG606840S, BWG606852S, BWG606858S, BWG606872S, BWG606875S, BWG606879S, BWG606881S, BWG606884S, BWG606895S, BWG606901S, BWG606904S, BWG607022S, BWG607025S, BWG607206S, BWG607213S, BWG607281S, BWG607282S, BWG607284S, BWG607309S, BWG607317S, BWG607318S, BWG607388S, BWG607391S, BWG607394S, BWG607417S, BWG607420S, BWG607440S, BWG607442S, BWG607443S, BWG607445S, BWG607447S, BWG607448S, BWG607449S, BWG607454S, BWG607456S, BWG607457S, BWG607459S, BWG607460S, BWG607464S, BWG607465S, BWG607466S, BWG607468S, BWG607469S, BWG607470S, BWG607512S, BWG607516S, BWG607517S, BWG607524S, BWG607526S, BWG607533S, BWG607576S, BWG607579S, BWG607592S, BWG607594S, BWG607596S, BWG607600S, BWG607601S, BWG607605S, BWG607607S, BWG607609S, BWG607611S, BWG607615S, BWG607629S, BWG607631S, BWG607641S, BWG607646S, BWG606484S, BWG606486S, BWG606430S, BWG606432S, BWG606433S, BWG606434S, BWG606560S, BWG607010S, BWG607013S, BWG607294S, BWG607414S, BWG607510S, BWG606340S, BWG606349S, BWG606376S, BWG606388S, BWG606590S, BWG606594S, BWG606632S, BWG606633S, BWG606634S, BWG606635S, BWG606638S, BWG606639S, BWG606644S, BWG606391S, BWG606444S, BWG606565S, BWG606575S, BWG606605S, BWG606655S, BWG606660S, BWG606673S, BWG606680S, BWG606681S, BWG607016S, BWG607159S, BWG607163S, BWG607164S, BWG607165S, BWG607168S, BWG607169S, BWG607216S, BWG607217S, BWG607218S, BWG607219S, BWG607222S, BWG607223S, BWG607224S, BWG607225S, BWG607233S, BWG607234S, BWG607308S, BWG607310S, BWG607311S, BWG607314S, BWG607316S, BWG607321S, BWG607322S, BWG607437S, BWG607893S, BWG607901S, BWG607905S, BWG607906S, BWG607907S, BWG607908S, BWG607910S, BWG607911S, BWG607914S, BWG607993S, BWG607997S, BWG608007S, BWG608008S, BWG608009S, BWG608010S, BWG608011S, BWG608023S, BWG608039S, BWG608040S, BWG608041S, BWG608045S, BWG608050S, BWG608051S, BWG608052S, BWG606564S, BWG606667S, BWG606467S, BWG606361S, BWG606886S, BWG606428S, BWG606510S, BWG606511S, BWG606528S, BWG606504S, BWG606636S, BWG606646S, BWG606650S, BWG607170S, BWG607175S, BWG607393S, BWG607399S, BWG606505S, BWG606506S, BWG606507S, BWG606509S, BWG606719S, BWG606366S, BWG606674S, BWG606734S, BWG606514S, BWG606342S, BWG606492S, BWG606496S, BWG606499S, BWG606678S, BWG606772S, BWG606773S, BWG606778S, BWG606789S, BWG606807S, BWG606814S, BWG606818S, BWG606825S, BWG606828S, BWG606829S, BWG606833S, BWG606855S, BWG606856S, BWG606857S, BWG606860S, BWG606861S, BWG606867S, BWG606868S, BWG606869S, BWG606870S, BWG606874S, BWG606878S, BWG606880S, BWG606882S, BWG607008S, BWG607017S, BWG607021S, BWG607075S, BWG607174S, BWG607193S, BWG607202S, BWG607207S, BWG607210S, BWG607211S, BWG607279S, BWG607289S, BWG607290S, BWG607303S, BWG607306S, BWG607387S, BWG607389S, BWG607415S, BWG607416S, BWG607438S, BWG607453S, BWG607506S, BWG607518S, BWG607587S, BWG607595S, BWG607606S, BWG607614S, BWG607616S, BWG607623S, BWG607648S, BWG606438S, BWG606578S, BWG606656S, BWG606658S, BWG606716S, BWG606718S, BWG606720S, BWG606735S, BWG606736S, BWG606737S, BWG607295S, BWG607297S, BWG607640S, BWG607996S, BWG607654S, BWG607915S, BWG607916S, BWG607918S, BWG607924S, BWG607925S, BWG607926S, BWG607951S, BWG607952S, BWG608022S, BWG607942S, BWG607943S, BWG607944S, BWG607945S, BWG607946S, BWG607947S, BWG607948S, BWG607949S, BWG607994S, BWG607995S, BWG607998S, BWG607999S, BWG608000S, BWG608001S, BWG608002S, BWG608003S, BWG608004S, BWG608005S, BWG608006S, BWG608017S, BWG608018S, BWG608019S, BWG608020S, BWG608027S, BWG608028S, BWG608029S, BWG608030S, BWG608031S, BWG608032S, BWG608033S, BWG608034S, BWG607894S, BWG607895S, BWG607903S, BWG607904S, BWG608012S, BWG608013S, BWG608014S, BWG608015S, BWG608042S, BWG608043S, BWG608049S; GTIN 00643169980228, Lot Serial Numbers: BWG607183S, BWG607239S, BWG607250S, BWG608240S, BWG607092S, BWG607240S, BWG607329S, BWG606540S, BWG606911S, BWG607249S, BWG607253S, BWG608552S, BWG608872S, BWG609118S, BWG610086S, BWG610088S, BWG610619S, BWG610632S, BWG611384S, BWG611388S, BWG610089S, BWG606981S, BWG607657S, BWG608117S, BWG608160S, BWG607360S, BWG607363S, BWG608079S, BWG607336S, BWG607891S, BWG607328S, BWG607033S, BWG606924S, BWG607093S, BWG607348S, BWG607817S, BWG606958S, BWG606957S, BWG606961S, BWG606928S, BWG606916S, BWG607099S, BWG607032S, BWG606978S, BWG607080S, BWG607272S, BWG607366S, BWG608835S, BWG607029S, BWG607664S, BWG607843S, BWG608155S, BWG608244S, BWG608665S, BWG608693S, BWG608959S, BWG610341S, BWG606962S, BWG606983S, BWG607027S, BWG607031S, BWG607038S, BWG607040S, BWG607058S, BWG607273S, BWG610187S, BWG608565S, BWG607850S, BWG606927S, BWG606946S, BWG606954S, BWG606956S, BWG607078S, BWG607079S, BWG607083S, BWG607110S, BWG607152S, BWG606907S, BWG607892S, BWG606917S, BWG606925S, BWG607044S, BWG607094S, BWG607116S, BWG607119S, BWG607270S, BWG607352S, BWG607411S, BWG607479S, BWG607539S, BWG607560S, BWG607659S, BWG607660S, BWG607669S, BWG607706S, BWG607754S, BWG607953S, BWG608162S, BWG609636S, BWG609877S, BWG610142S, BWG606930S, BWG606932S, BWG607036S, BWG607143S, BWG607242S, BWG607369S, BWG607376S, BWG607566S, BWG607705S, BWG607769S, BWG607770S, BWG607771S, BWG607824S, BWG607844S, BWG608082S, BWG608146S, BWG608159S, BWG606955S, BWG607039S, BWG607104S, BWG607127S, BWG607244S, BWG607365S, BWG607427S, BWG607476S, BWG607571S, BWG607758S, BWG607816S, BWG607878S, BWG608088S, BWG608148S, BWG607675S, BWG606940S, BWG606935S, BWG606963S, BWG607153S, BWG607362S, BWG607426S, BWG607483S, BWG607565S, BWG608083S, BWG607130S, BWG607662S, BWG606937S, BWG607375S, BWG607487S, BWG608090S, BWG608150S, BWG608703S, BWG607370S, BWG607261S, BWG606936S, BWG607862S, BWG608518S, BWG608536S, BWG607103S, BWG607107S, BWG608878S, BWG606909S, BWG606964S, BWG607867S, BWG607966S, BWG608077S, BWG607251S, BWG606952S, BWG606985S, BWG607082S, BWG606921S, BWG607409S, BWG608233S, BWG607655S, BWG608559S, BWG607821S, BWG607076S, BWG607704S, BWG608157S, BWG608161S, BWG606920S, BWG606923S, BWG606931S, BWG606971S, BWG607087S, BWG607090S, BWG607147S, BWG607241S, BWG607252S, BWG607254S, BWG607255S, BWG607345S, BWG607346S, BWG607820S, BWG607882S, BWG608076S, BWG608163S, BWG608526S, BWG608549S, BWG608568S, BWG607045S, BWG607964S, BWG607041S, BWG608215S, BWG607574S, BWG608067S, BWG608078S, BWG609203S, BWG609225S, BWG609226S, BWG609234S, BWG609247S, BWG609459S, BWG609748S, BWG609963S, BWG609999S, BWG610047S, BWG610099S, BWG610101S, BWG610124S, BWG610204S, BWG610211S, BWG610225S, BWG610346S, BWG609239S, BWG609749S, BWG609949S, BWG610092S, BWG610360S, BWG607482S.BWG609001S, BWG610081S, BWG608245S, BWG608629S, BWG608997S, BWG608998S, BWG609002S, BWG609070S, BWG609086S, BWG609208S, BWG609282S, BWG609284S, BWG609369S, BWG609371S, BWG609647S, BWG609649S, BWG609657S, BWG609745S, BWG609750S, BWG609751S, BWG609982S, BWG610091S, BWG610093S, BWG610118S, BWG610235S, BWG610244S, BWG610354S, BWG610358S, BWG610639S, BWG608695S, BWG609106S, BWG609242S, BWG609753S, BWG609775S, BWG609842S, BWG609965S, BWG610233S, BWG610317S, BWG610355S, BWG607814S, BWG607481S, BWG607484S, BWG607665S, BWG607815S, BWG607828S, BWG607829S, BWG607830S, BWG607831S, BWG608058S, BWG608059S, BWG608062S, BWG608074S, BWG608081S, BWG608175S, BWG608850S, BWG608990S, BWG609104S, BWG609439S, BWG609440S, BWG609443S, BWG609447S, BWG609567S, BWG609603S, BWG609624S, BWG609967S, BWG609971S, BWG610213S, BWG610245S, BWG610251S, BWG610254S, BWG610261S, BWG610336S, BWG610338S, BWG610340S, BWG610345S, BWG610356S, BWG610514S, BWG610636S, BWG610637S, BWG610638S, BWG607488S, BWG607489S, BWG607573S, BWG607755S, BWG607757S, BWG608212S, BWG608807S, BWG609283S, BWG609379S, BWG609380S, BWG609461S, BWG609481S, BWG609628S, BWG609956S, BWG610056S, BWG610063S, BWG610634S, BWG609449S, BWG610082S, BWG610343S, BWG609381S, BWG607056S, BWG607480S, BWG607105S, BWG607374S, BWG608173S, BWG607670S, BWG606922S, BWG609006S, BWG607408S, BWG607042S, BWG607106S, BWG607236S, BWG607330S, BWG607353S, BWG606934S, BWG607004S, BWG607661S, BWG608102S, BWG607377S, BWG607425S, BWG607540S, BWG606914S, BWG606915S, BWG607030S, BWG607474S, BWG607475S, BWG607768S, BWG609634S, BWG608145S, BWG607034S, BWG607035S, BWG607778S, BWG607787S, BWG608515S, BWG606941S, BWG606942S, BWG606944S, BWG606980S, BWG607028S, BWG606926S, BWG607184S, BWG606966S, BWG607245S, BWG607037S, BWG607851S, BWG607854S, BWG608056S, BWG607564S, BWG608152S, BWG609069S, BWG606919S, BWG607247S, BWG607422S, BWG606912S, BWG606953S, BWG607339S, BWG608179S, BWG607372S, BWG607271S, BWG607065S, BWG608166S, BWG608193S, BWG608216S, BWG607269S, BWG607088S, BWG608174S, BWG607262S, BWG607327S, BWG607538S, BWG608151S, BWG608420S, BWG608871S, BWG609582S, BWG607120S, BWG607258S, BWG607818S, BWG607852S, BWG608071S, BWG608551S, BWG608701S, BWG606977S, BWG606979S, BWG606996S, BWG606997S, BWG606998S, BWG606999S, BWG607000S, BWG607001S, BWG607002S, BWG607003S, BWG610164S, BWG606943S, BWG606542S, BWG606541S, BWG606929S, BWG607145S, BWG607077S, BWG606969S, BWG607085S, BWG607097S, BWG608187S, BWG608147S, BWG608089S, BWG607149S, BWG607762S, BWG608085S, BWG607043S, BWG606951S, BWG607122S, BWG607656S, BWG607971S, BWG608143S, BWG607825S, BWG607958S, BWG607267S, BWG608158S, BWG607412S, BWG606968S, BWG606970S, BWG606974S, BWG606975S, BWG606976S, BWG607060S, BWG607061S, BWG607062S, BWG607063S, BWG607064S, BWG607135S, BWG607136S, BWG607138S, BWG607139S, BWG607141S, BWG607429S, BWG607430S, BWG607431S, BWG607432S, BWG607433S, BWG607707S, BWG607708S, BWG607709S, BWG607710S, BWG607711S, BWG608138S, BWG608139S, BWG608142S, BWG608227S, BWG608228S, BWG608229S, BWG608230S, BWG608231S, BWG609374S, BWG609395S, BWG609651S, BWG609662S, BWG609666S, BWG609736S, BWG609746S, BWG610342S, BWG610353S, BWG610357S, BWG610359S, BWG610505S, BWG608154S, BWG608211S, BWG608243S, BWG608633S, BWG608634S, BWG608635S, BWG608694S, BWG608806S, BWG608961S, BWG608980S, BWG608981S, BWG608987S, BWG608995S, BWG609103S, BWG609209S, BWG609216S, BWG609217S, BWG610641S, BWG610640S, BWG608329S, BWG608369S, BWG608861S, BWG609019S, BWG609024S, BWG609192S, BWG609462S, BWG607789S, BWG607790S, BWG607795S, BWG607803S, BWG607961S, BWG607965S, BWG607980S, BWG607987S, BWG607989S, BWG608061S, BWG608084S, BWG608086S, BWG608110S, BWG608118S, BWG608131S, BWG608140S, BWG608156S, BWG608168S, BWG608172S, BWG608180S, BWG608221S, BWG608297S, BWG608310S, BWG608312S, BWG608327S, BWG608328S, BWG608334S, BWG608337S, BWG608347S, BWG608349S, BWG608376S, BWG608445S, BWG608448S, BWG608464S, BWG608656S, BWG608657S, BWG608658S, BWG608660S, BWG608662S, BWG608681S, BWG608683S, BWG608686S, BWG608692S, BWG608724S, BWG608726S, BWG608816S, BWG608907S, BWG608164S, BWG608185S, BWG607761S, BWG607775S, BWG607776S, BWG607798S, BWG607800S, BWG607802S, BWG607804S, BWG607806S, BWG607807S, BWG607809S, BWG607842S, BWG607960S, BWG607962S, BWG607990S, BWG608066S, BWG608096S, BWG608105S, BWG608106S, BWG608114S, BWG608116S, BWG608122S, BWG608123S, BWG608141S, BWG608296S, BWG608313S, BWG608315S, BWG608316S, BWG608322S, BWG608323S, BWG608333S, BWG608335S, BWG608343S, BWG608384S, BWG608426S, BWG608436S, BWG608440S, BWG608441S, BWG608451S, BWG608454S, BWG608467S, BWG608468S, BWG608585S, BWG608621S, BWG608622S, BWG608630S, BWG608631S, BWG608639S, BWG608666S, BWG608713S, BWG608903S, BWG607779S, BWG607791S, BWG607808S, BWG607957S, BWG607978S, BWG608053S, BWG608055S, BWG608057S, BWG608070S, BWG608091S, BWG608108S, BWG608115S, BWG608120S, BWG608126S, BWG608191S, BWG608271S, BWG608278S, BWG608279S, BWG608288S, BWG608289S, BWG608300S, BWG608319S, BWG608320S, BWG608352S, BWG608365S, BWG608372S, BWG608375S, BWG608381S, BWG608387S, BWG608432S, BWG608438S, BWG608444S, BWG608450S, BWG608469S, BWG608470S, BWG608587S, BWG608611S, BWG608612S, BWG608680S, BWG608722S, BWG608750S, BWG608818S, BWG608843S, BWG608675S, BWG609248S, BWG609277S, BWG609279S, BWG609366S, BWG609655S, BWG609740S, BWG609744S, BWG610097S, BWG610098S, BWG606945S, BWG609480S, BWG606972S, BWG607005S, BWG607148S, BWG607142S, BWG608218S, BWG608499S, BWG608509S, BWG608873S, BWG609241S, BWG609268S, BWG610571S, BWG610614S, BWG610649S, BWG610715S, BWG611236S, BWG611275S, BWG607663S, BWG608404S, BWG608480S, BWG607666S, BWG607671S, BWG609155S, BWG609210S, BWG609776S, BWG608413S, BWG607668S, BWG608483S, BWG608679S, BWG608870S, BWG608241S, BWG608262S, BWG608272S, BWG608275S, BWG608281S, BWG608282S, BWG608293S, BWG608294S, BWG608305S, BWG608346S, BWG608358S, BWG608359S, BWG608363S, BWG607133S, BWG608996S, BWG610143S, BWG608421S, BWG609586S, BWG606973S, BWG609202S, BWG609632S, BWG609638S, BWG609639S, BWG610198S, BWG609756S, BWG610055S, BWG610158S, BWG610193S, BWG609627S, BWG607792S, BWG607797S, BWG607805S, BWG607810S, BWG607811S, BWG607812S, BWG607832S, BWG607833S, BWG607834S, BWG607835S, BWG607836S, BWG607837S, BWG607838S, BWG607839S, BWG607840S, BWG607841S, BWG607874S, BWG607875S, BWG607876S, BWG607877S, BWG607879S, BWG607880S, BWG607881S, BWG607883S, BWG607884S, BWG607885S, BWG607886S, BWG607887S, BWG608119S, BWG608121S, BWG608125S, BWG608127S, BWG608128S, BWG608129S, BWG608130S, BWG608132S, BWG608133S, BWG608134S, BWG608135S, BWG608136S, BWG608137S, BWG608682S, BWG608685S, BWG608687S, BWG608706S, BWG608707S, BWG608708S, BWG608710S, BWG608711S, BWG608712S, BWG608714S, BWG608715S, BWG608717S, BWG608719S, BWG608720S, BWG608721S, BWG608723S, BWG608725S, BWG608727S, BWG608729S, BWG608730S, BWG608731S, BWG608732S, BWG608733S, BWG608734S, BWG608735S, BWG608736S, BWG608737S, BWG608738S, BWG608742S, BWG608744S, BWG608746S, BWG608748S, BWG608751S, BWG608752S, BWG608754S, BWG609766S, BWG609976S, BWG610095S, BWG610199S, BWG608545S, BWG610327S, BWG610331S, BWG607095S, BWG607144S, BWG607331S, BWG607332S, BWG607333S, BWG607349S, BWG607569S, BWG607756S, BWG607826S, BWG607890S, BWG607970S, BWG608165S, BWG608167S, BWG608170S, BWG608171S, BWG608418S, BWG608419S, BWG608512S, BWG608513S, BWG608538S, BWG608539S, BWG608553S, BWG608554S, BWG608555S, BWG608557S, BWG608561S, BWG608562S, BWG608994S, BWG609453S, BWG609467S, BWG609531S, BWG609618S, BWG609626S, BWG609755S, BWG606982S, BWG606984S, BWG606986S, BWG606987S, BWG606988S, BWG606990S, BWG606991S, BWG606992S, BWG606993S, BWG606994S, BWG610312S, BWG610314S, BWG607667S, BWG607674S, BWG607677S, BWG607678S, BWG607679S, BWG607681S, BWG608400S, BWG608401S, BWG608479S, BWG608481S, BWG608482S, BWG608486S, BWG608487S, BWG608488S, BWG608489S, BWG608490S, BWG608491S, BWG608492S, BWG608494S, BWG608786S, BWG608828S, BWG608830S, BWG608832S, BWG608834S, BWG608841S, BWG608974S, BWG608977S, BWG608982S, BWG608983S, BWG608984S, BWG608986S, BWG609004S, BWG609010S, BWG609011S, BWG609022S, BWG609023S, BWG609152S, BWG609153S, BWG609159S, BWG609205S, BWG609206S, BWG609207S, BWG609212S, BWG609213S, BWG609219S, BWG609222S, BWG609255S, BWG609257S, BWG609267S, BWG609269S, BWG609270S, BWG609274S, BWG609275S, BWG609304S, BWG609306S, BWG609307S, BWG609312S, BWG609313S, BWG609314S, BWG609317S, BWG609318S, BWG609319S, BWG609320S, BWG609322S, BWG609325S, BWG607355S, BWG608696S, BWG608697S, BWG608827S, BWG609383S, BWG608177S, BWG608183S, BWG608476S, BWG608517S, BWG608528S, BWG608529S, BWG608531S, BWG608532S, BWG608533S, BWG608534S, BWG608569S, BWG608698S, BWG608847S, BWG608848S, BWG608937S, BWG606913S, BWG608829S, BWG609470S, BWG609941S, BWG609931S, BWG610679S, BWG611444S, BWG609281S, BWG607760S, BWG607772S, BWG607855S, BWG607857S, BWG607858S, BWG607859S, BWG607860S, BWG608246S, BWG608247S, BWG608248S, BWG608249S, BWG608250S, BWG608251S, BWG608252S, BWG608254S, BWG608255S, BWG608256S, BWG608257S, BWG608258S, BWG608260S, BWG608261S, BWG608263S, BWG608264S, BWG608265S, BWG608663S, BWG608951S, BWG608953S, BWG608954S, BWG608958S, BWG608964S, BWG608966S, BWG608973S, BWG609143S, BWG609145S, BWG609574S, BWG609576S, BWG609577S, BWG609579S, BWG609580S, BWG609672S, BWG609673S, BWG609675S, BWG607081S, BWG607491S, BWG607492S, BWG607493S, BWG607544S, BWG607545S, BWG607546S, BWG607547S, BWG607548S, BWG607552S, BWG607553S, BWG607554S, BWG607672S, BWG607673S, BWG607680S, BWG607694S, BWG607699S, BWG607700S, BWG607701S, BWG607703S, BWG607722S, BWG607723S, BWG607724S, BWG607725S, BWG607726S, BWG607727S, BWG607728S, BWG607729S, BWG607730S, BWG607731S, BWG607732S, BWG607861S, BWG607863S, BWG607967S, BWG607969S, BWG607991S, BWG607992S, BWG608195S, BWG608223S, BWG608224S, BWG608225S, BWG608226S, BWG608253S, BWG608266S, BWG608269S, BWG608460S, BWG608495S, BWG608496S, BWG608500S, BWG608510S, BWG608598S, BWG608617S, BWG608618S, BWG608619S, BWG608690S, BWG608760S, BWG608765S, BWG608768S, BWG608770S, BWG608775S, BWG608795S, BWG609136S, BWG609137S, BWG609138S, BWG609139S, BWG609393S, BWG609482S, BWG609495S, BWG609496S, BWG609590S, BWG609596S, BWG609598S, BWG609599S, BWG609600S, BWG610366S, BWG610367S, BWG610369S, BWG610370S, BWG610372S, BWG610373S, BWG610374S, BWG610375S, BWG610378S, BWG610379S, BWG610380S, BWG610381S, BWG610382S, BWG610383S, BWG610385S, BWG610669S, BWG610839S, BWG610094S, BWG610351S, BWG609747S, BWG609752S, BWG609757S, BWG610078S, BWG610971S, BWG607819S, BWG607124S, BWG607134S, BWG607822S, BWG608087S, BWG608232S, BWG608392S, BWG608393S, BWG608394S, BWG608437S, BWG608790S, BWG608890S, BWG609117S, BWG609170S, BWG609176S, BWG609408S, BWG609434S, BWG609435S, BWG609436S, BWG609442S, BWG609524S, BWG609665S, BWG609702S, BWG610309S, BWG610310S, BWG610311S, BWG610435S, BWG610436S, BWG610437S, BWG610519S, BWG611164S, BWG611215S, BWG611216S, BWG611632S, BWG611640S, BWG611641S, BWG611648S, BWG608181S, BWG608182S, BWG608702S, BWG608741S, BWG608780S, BWG608783S, BWG608784S, BWG608787S, BWG608793S, BWG609042S, BWG609059S, BWG608403S, BWG608484S, BWG609615S, BWG607687S, BWG607688S, BWG607690S, BWG607692S, BWG607693S, BWG608220S, BWG608222S, BWG608397S, BWG608504S, BWG608506S, BWG608507S, BWG608508S, BWG608590S, BWG607347S, BWG608547S, BWG609112S, BWG609387S, BWG609392S, BWG607100S, BWG607181S, BWG609558S, BWG606908S, BWG607275S, BWG607102S, BWG608411S, BWG608505S, BWG608540S, BWG608541S, BWG608543S, BWG609571S, BWG609402S, BWG609382S, BWG607123S, BWG607129S, BWG607354S, BWG607359S, BWG607361S, BWG608416S, BWG608417S, BWG609386S, BWG609426S, BWG609430S, BWG609431S, BWG609432S, BWG609446S, BWG609478S, BWG609501S, BWG609505S, BWG609506S, BWG609553S, BWG609568S, BWG609578S, BWG607268S, BWG608412S, BWG608471S, BWG608409S, BWG607150S, BWG607182S, BWG607026S, BWG607180S, BWG608405S, BWG608406S, BWG608407S, BWG608410S, BWG608427S, BWG608428S, BWG608809S, BWG608810S, BWG608811S, BWG608947S, BWG609572S, BWG609573S, BWG610181S, BWG610217S, BWG610223S, BWG607098S, BWG607243S, BWG607248S, BWG607256S, BWG607490S, BWG607853S, BWG607856S, BWG609415S, BWG607759S, BWG607773S, BWG607774S.BWG607777S, BWG607781S, BWG607782S, BWG607788S, BWG607793S, BWG607799S, BWG607801S, BWG607977S, BWG607988S, BWG608060S, BWG608064S, BWG608065S, BWG608068S, BWG608073S, BWG608093S, BWG608095S, BWG608101S, BWG608104S, BWG608124S, BWG608176S, BWG608194S, BWG608234S, BWG608236S, BWG608268S, BWG608284S, BWG608287S, BWG608291S, BWG608292S, BWG608306S, BWG608314S, BWG608318S, BWG608325S, BWG608326S, BWG608332S, BWG608338S, BWG608339S, BWG608340S, BWG608341S, BWG608342S, BWG608344S, BWG608351S, BWG608361S, BWG608371S, BWG608374S, BWG608380S, BWG608382S, BWG608385S, BWG608425S, BWG608455S, BWG608456S, BWG608457S, BWG608458S, BWG608459S, BWG608535S, BWG608546S, BWG608560S, BWG608573S, BWG608574S, BWG608615S, BWG608637S, BWG608642S, BWG608648S, BWG608653S, BWG608655S, BWG608664S, BWG608667S, BWG608672S, BWG608676S, BWG608677S, BWG608689S, BWG608699S, BWG608716S, BWG608718S, BWG608728S, BWG608745S, BWG608802S, BWG608820S, BWG608831S, BWG608839S, BWG608849S, BWG608904S, BWG608910S, BWG608926S, BWG608930S, BWG609008S, BWG609012S, BWG609200S, BWG609215S, BWG609230S, BWG608303S, BWG608308S, BWG606947S, BWG606948S, BWG606949S, BWG606950S, BWG607059S, BWG607111S, BWG607113S, BWG607114S, BWG607115S, BWG607154S, BWG607155S, BWG607340S, BWG607341S, BWG607342S, BWG607343S, BWG607344S, BWG607378S, BWG607379S, BWG607380S, BWG607382S, BWG607383S, BWG607535S, BWG607536S, BWG607537S, BWG607542S, BWG607543S, BWG607555S, BWG607556S, BWG607557S, BWG607558S, BWG607559S, BWG607561S, BWG607562S, BWG607563S, BWG607568S, BWG607570S, BWG607763S, BWG607764S, BWG607765S, BWG607766S, BWG607767S, BWG607954S, BWG607955S, BWG607956S, BWG607973S, BWG607974S, BWG607975S, BWG607976S, BWG607979S, BWG607981S, BWG607982S, BWG608103S, BWG608107S, BWG608109S, BWG608111S, BWG608113S, BWG608200S, BWG608201S, BWG608202S, BWG608203S, BWG608204S, BWG608205S, BWG608206S, BWG608207S, BWG608208S, BWG608209S, BWG608210S, BWG608429S, BWG608430S, BWG608511S, BWG608514S, BWG608519S, BWG608520S, BWG608521S, BWG608524S, BWG608525S, BWG608530S, BWG608537S, BWG608542S, BWG608837S, BWG608840S, BWG608846S, BWG608866S, BWG608867S, BWG608944S, BWG609181S, BWG609187S, BWG609191S, BWG609195S, BWG609196S, BWG609258S, BWG609261S, BWG609262S, BWG609264S, BWG609488S, BWG609581S, BWG608548S, BWG611393S, BWG611395S, BWG611396S, BWG611398S, BWG608274S, BWG608280S, BWG608283S, BWG608317S, BWG608321S, BWG608348S, BWG608373S, BWG608377S, BWG608449S, BWG608571S, BWG608572S, BWG608601S, BWG608610S, BWG608638S, BWG608759S, BWG608761S, BWG608762S, BWG608766S, BWG608782S, BWG608789S, BWG608838S, BWG608851S, BWG608855S, BWG608856S, BWG608864S, BWG608885S, BWG608886S, BWG608888S, BWG608892S, BWG608898S, BWG608902S, BWG608905S, BWG608906S, BWG608942S, BWG608945S, BWG608965S, BWG609051S, BWG609053S, BWG609066S, BWG609073S, BWG609074S, BWG609077S, BWG609078S, BWG609082S, BWG609083S, BWG609093S, BWG609098S, BWG609099S, BWG609115S, BWG609125S, BWG609128S, BWG609132S, BWG609141S, BWG609144S, BWG609160S, BWG609223S, BWG609224S, BWG609238S, BWG609240S, BWG609244S, BWG609280S, BWG609291S, BWG609292S, BWG609294S, BWG609310S, BWG609311S, BWG609316S, BWG609338S, BWG609377S, BWG609424S, BWG609433S, BWG609468S, BWG609503S, BWG609507S, BWG609521S, BWG609523S, BWG609539S, BWG609540S, BWG609547S, BWG609549S, BWG609566S, BWG609607S, BWG609640S, BWG609664S, BWG609706S, BWG609707S, BWG609726S, BWG609743S, BWG609785S, BWG609786S, BWG609787S, BWG609789S, BWG609795S, BWG609796S, BWG609815S, BWG609829S, BWG609831S, BWG609859S, BWG609860S, BWG609861S, BWG609862S, BWG609873S, BWG609874S, BWG609876S, BWG609878S, BWG609879S, BWG609895S, BWG609896S, BWG609898S, BWG609899S, BWG609900S, BWG609902S, BWG609903S, BWG609904S, BWG609906S, BWG609908S, BWG609910S, BWG609911S, BWG609912S, BWG609913S, BWG609914S, BWG609919S, BWG610017S, BWG610024S, BWG610027S, BWG610044S, BWG610050S, BWG610129S, BWG610131S, BWG610132S, BWG610133S, BWG610135S, BWG610136S, BWG610242S, BWG610305S, BWG610306S, BWG610307S, BWG610308S, BWG610368S, BWG610371S, BWG610376S, BWG610384S, BWG610392S, BWG610401S, BWG610402S, BWG610403S, BWG610413S, BWG610414S, BWG610416S, BWG610418S, BWG610425S, BWG610426S, BWG610434S, BWG610438S, BWG610439S, BWG610447S, BWG610448S, BWG610450S, BWG610462S, BWG610463S, BWG610468S, BWG610469S, BWG610470S, BWG610472S, BWG610473S, BWG610474S, BWG610477S, BWG610478S, BWG610479S, BWG610480S, BWG610481S, BWG610482S, BWG610484S, BWG610485S, BWG610498S, BWG610508S, BWG610529S, BWG610530S, BWG610546S, BWG610547S, BWG610548S, BWG610549S, BWG610550S, BWG610551S, BWG610552S, BWG610553S, BWG610554S, BWG610555S, BWG610626S, BWG610627S, BWG610670S, BWG610671S, BWG610673S, BWG610675S, BWG610676S, BWG610682S, BWG610685S, BWG610686S, BWG610687S, BWG610688S, BWG610690S, BWG610691S, BWG610692S, BWG610693S, BWG610694S, BWG610695S, BWG610697S, BWG610698S, BWG610699S, BWG610700S, BWG610702S, BWG610706S, BWG610725S, BWG610727S, BWG610732S, BWG610735S, BWG610737S, BWG610745S, BWG610748S, BWG610749S, BWG610750S, BWG610751S, BWG610752S, BWG610753S, BWG610754S, BWG610756S, BWG610757S, BWG610758S, BWG610759S, BWG610762S, BWG610783S, BWG610785S, BWG610786S, BWG610787S, BWG610789S, BWG610790S, BWG610798S, BWG610799S, BWG610800S, BWG610806S, BWG610807S, BWG610808S, BWG610812S, BWG610815S, BWG610816S, BWG610819S, BWG610820S, BWG610821S, BWG610823S, BWG610826S, BWG610832S, BWG610838S, BWG610840S, BWG610842S, BWG610845S, BWG610874S, BWG610878S, BWG610879S, BWG610880S, BWG610893S, BWG610894S, BWG610895S, BWG610953S, BWG610954S, BWG610955S, BWG610957S, BWG610958S, BWG610977S, BWG610989S, BWG610990S, BWG610998S, BWG611031S, BWG611032S, BWG611057S, BWG611071S, BWG611072S, BWG611076S, BWG611077S, BWG611101S, BWG611103S, BWG611106S, BWG611114S, BWG611117S, BWG611119S, BWG611120S, BWG611121S, BWG611123S, BWG611124S, BWG611126S, BWG611127S, BWG611130S, BWG611159S, BWG611160S, BWG611161S, BWG611162S, BWG611163S, BWG611174S, BWG611175S, BWG611202S, BWG611203S, BWG611204S, BWG611205S, BWG611206S, BWG611207S, BWG611208S, BWG611209S, BWG611210S, BWG611211S, BWG611212S, BWG611219S, BWG611229S, BWG611230S, BWG611231S, BWG611250S, BWG611321S, BWG611322S, BWG611362S, BWG611363S, BWG611364S, BWG609504S, BWG609537S, BWG609552S, BWG608773S, BWG608791S, BWG608852S, BWG608972S, BWG608991S, BWG609100S, BWG609102S, BWG609171S, BWG609315S, BWG609378S, BWG609406S, BWG609530S, BWG609544S, BWG609669S, BWG609670S, BWG609725S, BWG609845S, BWG609867S, BWG610049S, BWG610051S, BWG610284S, BWG610285S, BWG610313S, BWG610467S, BWG610497S, BWG610562S, BWG610563S, BWG610738S, BWG610739S, BWG610803S, BWG610809S, BWG610810S, BWG610846S, BWG610927S, BWG610928S, BWG610929S, BWG610948S, BWG611156S, BWG611171S, BWG611173S, BWG611248S, BWG611409S, BWG611410S, BWG611427S, BWG611429S, BWG611431S, BWG611638S, BWG611639S, BWG607733S, BWG607864S, BWG607865S, BWG607870S, BWG607871S, BWG607872S, BWG607873S, BWG607888S, BWG607889S, BWG607963S, BWG607983S, BWG607984S, BWG607985S, BWG607986S, BWG608198S, BWG608199S, BWG608219S, BWG608474S, BWG608475S, BWG608582S, BWG608700S, BWG608779S, BWG608813S, BWG608876S, BWG608877S, BWG608932S, BWG609027S, BWG609047S, BWG609052S, BWG609124S, BWG609363S, BWG609364S, BWG609365S, BWG609390S, BWG609391S, BWG609394S, BWG609396S, BWG609583S, BWG609584S, BWG609585S, BWG609604S, BWG609608S, BWG609680S, BWG609681S, BWG609767S, BWG609768S, BWG609771S, BWG609988S, BWG609990S, BWG609996S, BWG610279S, BWG610386S, BWG610387S, BWG610388S, BWG610389S, BWG610390S, BWG610391S, BWG610647S, BWG610650S, BWG610772S, BWG610775S, BWG610825S, BWG610851S, BWG610855S, BWG610867S, BWG610906S, BWG610911S, BWG609486S, BWG609489S, BWG609498S, BWG609499S, BWG607112S, BWG607264S, BWG608879S, BWG606939S, BWG607117S, BWG607108S, BWG607101S, BWG607176S, BWG607177S, BWG608485S, BWG607057S, BWG607084S, BWG607086S, BWG607128S, BWG609250S, BWG608285S, BWG607405S, BWG608993S, BWG609110S, BWG610162S, BWG608473S, BWG610147S, BWG608286S, BWG607780S, BWG607783S, BWG607784S, BWG607785S, BWG607786S, BWG607794S, BWG607796S, BWG607959S, BWG607972S, BWG608054S, BWG608063S, BWG608072S, BWG608100S, BWG608112S, BWG608242S, BWG608259S, BWG608267S, BWG608431S, BWG608433S, BWG608434S, BWG608435S, BWG608461S, BWG608462S, BWG608465S, BWG608466S, BWG608523S, BWG608589S, BWG608674S, BWG608688S, BWG608705S, BWG608868S, BWG608395S, BWG610103S, BWG611135S, BWG606989S, BWG606933S, BWG608833S, BWG609601S, BWG609602S, BWG609479S, BWG607179S, BWG608584S, BWG609252S, BWG607827S, BWG608575S, BWG608836S, BWG608899S, BWG608900S, BWG609131S, BWG609173S, BWG609441S, BWG609606S, BWG609612S, BWG609613S, BWG609616S, BWG609617S, BWG609619S, BWG609620S, BWG609621S, BWG609660S, BWG609676S, BWG609678S, BWG609703S, BWG609704S, BWG609731S, BWG609732S, BWG609733S, BWG609734S, BWG609735S, BWG609737S, BWG609738S, BWG609739S, BWG609741S, BWG609742S, BWG609797S, BWG609798S, BWG609800S, BWG609801S, BWG609803S, BWG609804S, BWG609846S, BWG609847S, BWG609848S, BWG609849S, BWG609868S, BWG609869S, BWG609870S, BWG609871S, BWG609915S, BWG609936S, BWG609937S, BWG609942S, BWG610001S, BWG610007S, BWG610018S, BWG610037S, BWG610038S, BWG610039S, BWG610040S, BWG610042S, BWG610043S, BWG610045S, BWG610048S, BWG610057S, BWG610058S, BWG610059S, BWG610060S, BWG610061S, BWG610064S, BWG610066S, BWG610073S, BWG610077S, BWG610084S, BWG610085S, BWG610087S, BWG610090S, BWG610096S, BWG610100S, BWG610102S, BWG610154S, BWG610155S, BWG610156S, BWG610221S, BWG610222S, BWG610224S, BWG610226S, BWG610227S, BWG610228S, BWG610229S, BWG610230S, BWG610231S, BWG610232S, BWG610250S, BWG610252S, BWG610253S, BWG610255S, BWG610256S, BWG610257S, BWG610258S, BWG610259S, BWG610260S, BWG610262S, BWG610264S, BWG610265S, BWG610266S, BWG610268S, BWG610269S, BWG610270S, BWG610271S, BWG610272S, BWG610273S, BWG610274S, BWG610276S, BWG610277S, BWG610278S, BWG610319S, BWG610321S, BWG610322S, BWG610323S, BWG610324S, BWG610325S, BWG610326S, BWG610329S, BWG610333S, BWG610344S, BWG610377S, BWG610410S, BWG610415S, BWG610417S, BWG610419S, BWG610420S, BWG610421S, BWG610422S, BWG610423S, BWG610428S, BWG610430S, BWG610464S, BWG610465S, BWG610466S, BWG610475S, BWG610476S, BWG610483S, BWG610489S, BWG610490S, BWG610491S, BWG610492S, BWG610493S, BWG610494S, BWG610495S, BWG610496S, BWG610556S, BWG610557S, BWG610558S, BWG610568S, BWG610572S, BWG610573S, BWG610575S, BWG610579S, BWG610597S, BWG610598S, BWG610610S, BWG610611S, BWG610613S, BWG610615S, BWG610616S, BWG610617S, BWG610618S, BWG610620S, BWG610621S, BWG610622S, BWG610623S, BWG610624S, BWG610625S, BWG610760S, BWG610763S, BWG610764S, BWG610765S, BWG610768S, BWG610776S, BWG610778S, BWG610779S, BWG610784S, BWG610791S, BWG610794S, BWG610802S, BWG610811S, BWG610813S, BWG610814S, BWG610818S, BWG610827S, BWG610844S, BWG610847S, BWG610848S, BWG610849S, BWG610850S, BWG610852S, BWG610853S, BWG610854S, BWG610856S, BWG610857S, BWG610858S, BWG610859S, BWG610860S, BWG610869S, BWG610870S, BWG610871S, BWG610916S, BWG610917S, BWG610941S, BWG610944S, BWG610945S, BWG610946S, BWG610947S, BWG610993S, BWG611006S, BWG611007S, BWG611008S, BWG611009S, BWG611010S, BWG611011S, BWG611013S, BWG611014S, BWG611015S, BWG611083S, BWG611084S, BWG611085S, BWG611086S, BWG611087S, BWG611089S, BWG611090S, BWG611093S, BWG611094S, BWG611096S, BWG611244S, BWG611296S, BWG611297S, BWG611332S, BWG611333S, BWG611334S, BWG611335S, BWG611336S, BWG611339S, BWG611341S, BWG611342S, BWG611343S, BWG611344S, BWG611345S, BWG611346S, BWG611347S, BWG611348S, BWG611349S, BWG611350S, BWG611351S, BWG611352S, BWG611353S, BWG611354S, BWG611371S, BWG611372S, BWG611373S, BWG611374S, BWG611376S, BWG611377S, BWG611379S, BWG611382S, BWG611385S, BWG611386S, BWG611387S, BWG611389S, BWG609836S, BWG608596S, BWG608597S, BWG608826S, BWG609021S, BWG609029S, BWG609032S, BWG609033S, BWG609035S, BWG609037S, BWG609039S, BWG609060S, BWG609085S, BWG609120S, BWG609122S, BWG609123S, BWG609587S, BWG609588S, BWG609589S, BWG609677S, BWG609700S, BWG609708S, BWG609763S, BWG609764S, BWG609765S, BWG609875S, BWG610125S, BWG610138S, BWG610145S, BWG610167S, BWG610178S, BWG610612S, BWG610651S, BWG610653S, BWG610654S, BWG610655S.BWG610659S, BWG610943S, BWG611052S, BWG611053S, BWG611054S, BWG611058S, BWG611059S, BWG611061S, BWG611063S, BWG611064S, BWG611065S, BWG611066S, BWG611067S, BWG611073S, BWG611074S, BWG611075S, BWG611078S, BWG611079S, BWG611080S, BWG611088S, BWG611091S, BWG611092S, BWG608472S, BWG610393S, BWG610733S, BWG607237S, BWG607238S, BWG607549S, BWG607551S, BWG607682S, BWG607683S, BWG607684S, BWG607685S, BWG607686S, BWG607689S, BWG607691S, BWG607695S, BWG607696S, BWG607697S, BWG607698S, BWG607702S, BWG607712S, BWG607713S, BWG607714S, BWG607715S, BWG607716S, BWG607718S, BWG607719S, BWG607720S, BWG607721S, BWG607753S, BWG607866S, BWG607868S, BWG607869S, BWG608092S, BWG608094S, BWG608097S, BWG608098S, BWG608099S, BWG608190S, BWG608192S, BWG608196S, BWG608197S, BWG608396S, BWG608398S, BWG608477S, BWG608501S, BWG608502S, BWG608503S, BWG608556S, BWG608558S, BWG608566S, BWG608586S, BWG608588S, BWG608599S, BWG608600S, BWG608608S, BWG608613S, BWG608614S, BWG608616S, BWG608620S, BWG608623S, BWG608624S, BWG608625S, BWG608626S, BWG608627S, BWG608628S, BWG608632S, BWG608643S, BWG608644S, BWG608645S, BWG608646S, BWG608647S, BWG608649S, BWG608650S, BWG608651S, BWG608652S, BWG608654S, BWG608659S, BWG608661S, BWG608668S, BWG608669S, BWG608670S, BWG608673S, BWG608678S, BWG608814S, BWG608815S, BWG608817S, BWG608819S, BWG608842S, BWG608845S, BWG608874S, BWG608875S, BWG608880S, BWG608893S, BWG608895S, BWG608901S, BWG608908S, BWG608920S, BWG608922S, BWG608923S, BWG608927S, BWG608928S, BWG608935S, BWG609061S, BWG609076S, BWG609126S, BWG609127S, BWG609142S, BWG609146S, BWG609151S, BWG609287S, BWG609289S, BWG609293S, BWG609295S, BWG609296S, BWG609299S, BWG609300S, BWG609301S, BWG609302S, BWG609303S, BWG609333S, BWG609334S, BWG609337S, BWG609339S, BWG609340S, BWG609341S, BWG609342S, BWG609345S, BWG609347S, BWG609359S, BWG609360S, BWG609384S, BWG609385S, BWG609388S, BWG609397S, BWG609398S, BWG609401S, BWG609407S, BWG609409S, BWG609410S, BWG609411S, BWG609413S, BWG609416S, BWG609418S, BWG609419S, BWG609421S, BWG609422S, BWG609425S, BWG609428S, BWG609429S, BWG609437S, BWG609473S, BWG609474S, BWG609483S, BWG609484S, BWG609485S, BWG609487S, BWG609491S, BWG609492S, BWG609493S, BWG609494S, BWG609500S, BWG609508S, BWG609509S, BWG609510S, BWG609511S, BWG609513S, BWG609514S, BWG609515S, BWG609516S, BWG609522S, BWG609527S, BWG609529S, BWG609533S, BWG609534S, BWG609535S, BWG609536S, BWG609541S, BWG609542S, BWG609543S, BWG609546S, BWG609548S, BWG609559S, BWG609684S, BWG609685S, BWG609689S, BWG609692S, BWG609693S, BWG609696S, BWG609697S, BWG609711S, BWG609713S, BWG609714S, BWG609719S, BWG609758S, BWG609759S, BWG609760S, BWG609761S, BWG609762S, BWG609769S, BWG609770S, BWG609774S, BWG609784S, BWG609788S, BWG609791S, BWG609792S, BWG609799S, BWG609890S, BWG610104S, BWG610105S, BWG610106S, BWG610108S, BWG610109S, BWG610111S, BWG610112S, BWG610113S, BWG610119S, BWG610234S, BWG610263S, BWG610267S, BWG610281S, BWG610288S, BWG610289S, BWG610290S, BWG610300S, BWG610304S, BWG610335S, BWG610362S, BWG610363S, BWG610364S, BWG610656S, BWG610657S, BWG610658S, BWG610668S, BWG607845S, BWG609635S, BWG607263S, BWG607265S, BWG607266S, BWG607274S, BWG608295S, BWG608298S, BWG608301S, BWG608302S, BWG608307S, BWG608353S, BWG608364S, BWG608378S, BWG608379S, BWG608399S, BWG608443S, BWG608446S, BWG608453S, BWG608463S, BWG608578S, BWG608579S, BWG608580S, BWG608603S, BWG608604S, BWG608605S, BWG608606S, BWG608607S, BWG608709S, BWG608749S, BWG608753S, BWG608755S, BWG608756S, BWG608777S, BWG608778S, BWG608781S, BWG608798S, BWG608803S, BWG608804S, BWG608805S, BWG608862S, BWG608863S, BWG608865S, BWG608883S, BWG608891S, BWG608894S, BWG608896S, BWG608913S, BWG608914S, BWG608915S, BWG608916S, BWG608917S, BWG608921S, BWG608924S, BWG608929S, BWG608934S, BWG608939S, BWG608941S, BWG608948S, BWG608955S, BWG608960S, BWG608962S, BWG608967S, BWG608969S, BWG608975S, BWG608976S, BWG608978S, BWG608979S, BWG609003S, BWG609009S, BWG609013S, BWG609017S, BWG609020S, BWG609026S, BWG609034S, BWG609036S, BWG609038S, BWG609041S, BWG609043S, BWG609049S, BWG609057S, BWG609058S, BWG609062S, BWG609063S, BWG609064S, BWG609080S, BWG609084S, BWG609087S, BWG609088S, BWG609089S, BWG609091S, BWG609108S, BWG609109S, BWG609111S, BWG609113S, BWG609114S, BWG609148S, BWG609149S, BWG609150S, BWG609154S, BWG609156S, BWG609161S, BWG609162S, BWG609163S, BWG609164S, BWG609165S, BWG609166S, BWG609193S, BWG609194S, BWG609197S, BWG609204S, BWG609211S, BWG609214S, BWG609218S, BWG609265S, BWG609271S, BWG609272S, BWG609273S, BWG609278S, BWG609285S, BWG609298S, BWG609305S, BWG609308S, BWG609309S, BWG609351S, BWG609352S, BWG609355S, BWG609356S, BWG609370S, BWG609372S, BWG609373S, BWG609403S, BWG609404S, BWG609405S, BWG609427S, BWG609448S, BWG609458S, BWG609460S, BWG609463S, BWG609465S, BWG609469S, BWG609471S, BWG609472S, BWG609476S, BWG609525S, BWG609526S, BWG609528S, BWG609538S, BWG609594S, BWG609622S, BWG609623S, BWG609625S, BWG609630S, BWG609633S, BWG609652S, BWG609653S, BWG609654S, BWG609656S, BWG609661S, BWG609674S, BWG609682S, BWG609683S, BWG609686S, BWG609687S, BWG609688S, BWG609690S, BWG609691S, BWG609694S, BWG609718S, BWG609720S, BWG609805S, BWG609806S, BWG609825S, BWG609826S, BWG609827S, BWG609850S, BWG609851S, BWG609852S, BWG609853S, BWG609854S, BWG609855S, BWG609856S, BWG609857S, BWG609858S, BWG609987S, BWG609989S, BWG610019S, BWG610021S, BWG610065S, BWG610067S, BWG610068S, BWG610069S, BWG610070S, BWG610071S, BWG610072S, BWG610074S, BWG610076S, BWG610080S, BWG610083S, BWG610115S, BWG610120S, BWG610121S, BWG610141S, BWG610144S, BWG610160S, BWG610161S, BWG610163S, BWG610165S, BWG610166S, BWG610168S, BWG610169S, BWG610170S, BWG610171S, BWG610172S, BWG610173S, BWG610174S, BWG610175S, BWG610176S, BWG610177S, BWG610179S, BWG610183S, BWG610184S, BWG610196S, BWG610197S, BWG610200S, BWG610201S, BWG610202S, BWG610203S, BWG610205S, BWG610206S, BWG610214S, BWG610215S, BWG610216S, BWG610218S, BWG610219S, BWG610220S, BWG610236S, BWG610237S, BWG610238S, BWG610239S, BWG610240S, BWG610241S, BWG610243S, BWG610246S, BWG610247S, BWG610283S, BWG610315S, BWG610316S, BWG610318S, BWG610328S, BWG610330S, BWG610350S, BWG610365S, BWG610396S, BWG610397S, BWG610398S, BWG610399S, BWG610400S, BWG610404S, BWG610406S, BWG610407S, BWG610408S, BWG610409S, BWG610411S, BWG610412S, BWG610456S, BWG610457S, BWG610459S, BWG610460S, BWG610461S, BWG610504S, BWG610506S, BWG610518S, BWG610520S, BWG610521S, BWG610522S, BWG610523S, BWG610524S, BWG610525S, BWG610526S, BWG610531S, BWG610532S, BWG610533S, BWG610534S, BWG610535S, BWG610536S, BWG610537S, BWG610538S, BWG610539S, BWG610540S, BWG610541S, BWG610542S, BWG610576S, BWG610578S, BWG610581S, BWG610583S, BWG610584S, BWG610585S, BWG610586S, BWG610587S, BWG610588S, BWG610589S, BWG610590S, BWG610591S, BWG610592S, BWG610593S, BWG610594S, BWG610595S, BWG610596S, BWG610599S, BWG610600S, BWG610601S, BWG610602S, BWG610603S, BWG610604S, BWG610605S, BWG610606S, BWG610607S, BWG610608S, BWG610689S, BWG610696S, BWG610711S, BWG610716S, BWG610717S, BWG610719S, BWG610722S, BWG610724S, BWG610726S, BWG610728S, BWG610729S, BWG610730S, BWG610731S, BWG610734S, BWG610736S, BWG610740S, BWG610741S, BWG610782S, BWG610788S, BWG610792S, BWG610793S, BWG610801S, BWG610864S, BWG610865S, BWG610898S, BWG610899S, BWG610900S, BWG610901S, BWG610902S, BWG610903S, BWG610904S, BWG610905S, BWG610907S, BWG610908S, BWG610909S, BWG610910S, BWG610912S, BWG610914S, BWG610915S, BWG610918S, BWG610919S, BWG610920S, BWG610921S, BWG610922S, BWG610923S, BWG610925S, BWG610926S, BWG610959S, BWG610960S, BWG610961S, BWG610962S, BWG610963S, BWG610964S, BWG610965S, BWG610986S, BWG610987S, BWG610988S, BWG610991S, BWG610992S, BWG610995S, BWG611035S, BWG611036S, BWG611043S, BWG611044S, BWG611045S, BWG611095S, BWG611133S, BWG611134S, BWG611136S, BWG611142S, BWG611143S, BWG611145S, BWG611220S, BWG611221S, BWG611222S, BWG611223S, BWG611224S, BWG611225S, BWG611228S, BWG611298S, BWG611370S, BWG611405S, BWG611406S, BWG611411S, BWG611412S, BWG611413S, BWG611414S, BWG611415S, BWG611416S, BWG611417S, BWG611418S, BWG611419S, BWG611420S, BWG611421S, BWG611422S, BWG611424S, BWG611425S, BWG611426S, BWG611428S, BWG611430S, BWG611432S, BWG611433S, BWG611434S, BWG611435S, BWG611436S, BWG611437S, BWG611438S, BWG611439S, BWG611440S, BWG611441S, BWG611442S, BWG611443S, BWG611445S, BWG611446S, BWG611447S, BWG611448S, BWG611449S, BWG611450S, BWG611451S, BWG611452S, BWG611453S, BWG611454S, BWG611631S, BWG611635S, BWG611636S, BWG611637S, BWG611646S, BWG611647S, BWG611649S, BWG611653S, BWG611654S, BWG611655S, BWG611656S, BWG611657S, BWG611658S, BWG608276S, BWG608277S, BWG608290S, BWG608299S, BWG608324S, BWG608336S, BWG608350S, BWG608355S, BWG608357S, BWG608362S, BWG608370S, BWG608390S, BWG608391S, BWG608414S, BWG608423S, BWG608439S, BWG608442S, BWG608452S, BWG608767S, BWG608771S, BWG608774S, BWG608776S, BWG608796S, BWG608797S, BWG608800S, BWG608808S, BWG608853S, BWG608854S, BWG608858S, BWG608909S, BWG608911S, BWG608918S, BWG608919S, BWG608925S, BWG608931S, BWG608933S, BWG608936S, BWG608949S, BWG608971S, BWG608992S, BWG609000S, BWG609028S, BWG609044S, BWG609054S, BWG609067S, BWG609068S, BWG609072S, BWG609095S, BWG609096S, BWG609097S, BWG609101S, BWG609129S, BWG609130S, BWG609135S, BWG609172S, BWG609174S, BWG609175S, BWG609177S, BWG609182S, BWG609186S, BWG609188S, BWG609189S, BWG609220S, BWG609263S, BWG609326S, BWG609327S, BWG609329S, BWG609336S, BWG609343S, BWG609344S, BWG609346S, BWG609362S, BWG609367S, BWG609368S, BWG609417S, BWG609420S, BWG609438S, BWG609444S, BWG609445S, BWG609456S, BWG609457S, BWG609502S, BWG609520S, BWG609550S, BWG609551S, BWG609554S, BWG609562S, BWG609597S, BWG609605S, BWG609659S, BWG609679S, BWG609698S, BWG609701S, BWG609716S, BWG609722S, BWG609723S, BWG609730S, BWG609802S, BWG609809S, BWG609810S, BWG609828S, BWG609833S, BWG609840S, BWG609841S, BWG609881S, BWG609916S, BWG609917S, BWG609918S, BWG609921S, BWG610028S, BWG610029S, BWG610032S, BWG610053S, BWG610054S, BWG610062S, BWG610075S, BWG610107S, BWG610110S, BWG610137S, BWG610149S, BWG610150S, BWG610151S, BWG610152S, BWG610153S, BWG610157S, BWG610192S, BWG610194S, BWG610207S, BWG610208S, BWG610209S, BWG610286S, BWG610287S, BWG610291S, BWG610292S, BWG610293S, BWG610348S, BWG610349S, BWG610361S, BWG610427S, BWG610429S, BWG610440S, BWG610441S, BWG610442S, BWG610443S, BWG610444S, BWG610445S, BWG610446S, BWG610449S, BWG610451S, BWG610458S, BWG610471S, BWG610487S, BWG610488S, BWG610499S, BWG610500S, BWG610501S, BWG610502S, BWG610503S, BWG610507S, BWG610543S, BWG610544S, BWG610545S, BWG610577S, BWG610678S, BWG610680S, BWG610681S, BWG610683S, BWG610684S, BWG610703S, BWG610712S, BWG610713S, BWG610714S, BWG610718S, BWG610721S, BWG610723S, BWG610742S, BWG610743S, BWG610769S, BWG610771S, BWG610777S, BWG610780S, BWG610781S, BWG610804S, BWG610805S, BWG610824S, BWG610828S, BWG610861S, BWG610862S, BWG610866S, BWG610868S, BWG610875S, BWG610882S, BWG610883S, BWG610967S, BWG610968S, BWG610969S, BWG610978S, BWG610982S, BWG610983S, BWG610997S, BWG611002S, BWG611003S, BWG611004S, BWG611005S, BWG611021S, BWG611022S, BWG611023S, BWG611024S, BWG611030S, BWG611033S, BWG611034S, BWG611037S, BWG611051S, BWG611055S, BWG611056S, BWG611081S, BWG611082S, BWG611125S, BWG611128S, BWG611129S, BWG611137S, BWG611138S, BWG611139S, BWG611140S, BWG611141S, BWG611213S, BWG611214S, BWG611245S, BWG611246S, BWG611249S, BWG611251S, BWG611252S, BWG611320S, BWG611365S, BWG611423S, BWG611630S, BWG611633S, BWG611642S, BWG611643S, BWG611644S, BWG611650S, BWG608270S, BWG608273S, BWG608304S, BWG608309S, BWG608311S, BWG608354S, BWG608356S, BWG608360S, BWG608366S, BWG608367S, BWG608368S, BWG608383S, BWG608386S, BWG608388S, BWG608389S, BWG608402S, BWG608408S, BWG608447S, BWG608576S, BWG608577S, BWG608581S, BWG608591S, BWG608609S, BWG608671S, BWG608684S, BWG608740S, BWG608743S, BWG608747S, BWG608757S, BWG608758S, BWG608763S, BWG608764S, BWG608769S, BWG608772S, BWG608792S, BWG608799S, BWG608801S, BWG608812S, BWG608821S, BWG608824S, BWG608825S, BWG608857S, BWG608859S, BWG608860S, BWG608882S, BWG608884S, BWG608887S, BWG608889S, BWG608897S, BWG608912S.BWG608938S, BWG608946S, BWG608950S, BWG608952S, BWG608956S, BWG608963S, BWG608970S, BWG608985S, BWG608988S, BWG608989S, BWG609007S, BWG609045S, BWG609046S, BWG609050S, BWG609055S, BWG609056S, BWG609065S, BWG609071S, BWG609075S, BWG609079S, BWG609090S, BWG609092S, BWG609105S, BWG609116S, BWG609119S, BWG609121S, BWG609133S, BWG609134S, BWG609140S, BWG609157S, BWG609158S, BWG609167S, BWG609168S, BWG609169S, BWG609178S, BWG609180S, BWG609183S, BWG609184S, BWG609185S, BWG609190S, BWG609198S, BWG609201S, BWG609227S, BWG609228S, BWG609229S, BWG609232S, BWG609233S, BWG609235S, BWG609253S, BWG609256S, BWG609259S, BWG609260S, BWG609286S, BWG609288S, BWG609290S, BWG609297S, BWG609321S, BWG609323S, BWG609324S, BWG609328S, BWG609332S, BWG609335S, BWG609348S, BWG609349S, BWG609350S, BWG609353S, BWG609354S, BWG609357S, BWG609361S, BWG609375S, BWG609376S, BWG609399S, BWG609400S, BWG609412S, BWG609414S, BWG609450S, BWG609451S, BWG609452S, BWG609455S, BWG609517S, BWG609518S, BWG609519S, BWG609555S, BWG609556S, BWG609557S, BWG609563S, BWG609564S, BWG609609S, BWG609610S, BWG609611S, BWG609641S, BWG609642S, BWG609644S, BWG609646S, BWG609648S, BWG609650S, BWG609663S, BWG609667S, BWG609668S, BWG609671S, BWG609695S, BWG609705S, BWG609710S, BWG609712S, BWG609715S, BWG609717S, BWG609721S, BWG609724S, BWG609727S, BWG609728S, BWG609729S, BWG609778S, BWG609779S, BWG609780S, BWG609781S, BWG609782S, BWG609783S, BWG609793S, BWG609794S, BWG609807S, BWG609808S, BWG609811S, BWG609812S, BWG609813S, BWG609817S, BWG609819S, BWG609820S, BWG609821S, BWG609822S, BWG609823S, BWG609824S, BWG609830S, BWG609832S, BWG609837S, BWG609838S, BWG609844S, BWG609863S, BWG609864S, BWG609866S, BWG609880S, BWG609882S, BWG609883S, BWG609884S, BWG609885S, BWG609886S, BWG609887S, BWG609888S, BWG609891S, BWG609892S, BWG609893S, BWG609924S, BWG609926S, BWG610004S, BWG610008S, BWG610009S, BWG610010S, BWG610011S, BWG610013S, BWG610014S, BWG610015S, BWG610016S, BWG610030S, BWG610034S, BWG610035S, BWG610036S, BWG610052S, BWG610079S, BWG610114S, BWG610116S, BWG610122S, BWG610126S, BWG610127S, BWG610128S, BWG610139S, BWG610140S, BWG610146S, BWG610148S, BWG610180S, BWG610182S, BWG610185S, BWG610186S, BWG610188S, BWG610189S, BWG610190S, BWG610191S, BWG610195S, BWG610210S, BWG610212S, BWG610248S, BWG610249S, BWG610294S, BWG610295S, BWG610296S, BWG610297S, BWG610298S, BWG610299S, BWG610301S, BWG610302S, BWG610303S, BWG610332S, BWG610334S, BWG610337S, BWG610339S, BWG610347S, BWG610394S, BWG610395S, BWG610405S, BWG610424S, BWG610431S, BWG610432S, BWG610433S, BWG610452S, BWG610455S, BWG610486S, BWG610509S, BWG610510S, BWG610511S, BWG610512S, BWG610513S, BWG610515S, BWG610516S, BWG610517S, BWG610559S, BWG610560S, BWG610561S, BWG610564S, BWG610565S, BWG610566S, BWG610567S, BWG610569S, BWG610570S, BWG610574S, BWG610582S, BWG610609S, BWG610628S, BWG610629S, BWG610630S, BWG610631S, BWG610672S, BWG610674S, BWG610677S, BWG610701S, BWG610704S, BWG610705S, BWG610707S, BWG610708S, BWG610709S, BWG610710S, BWG610746S, BWG610747S, BWG610761S, BWG610766S, BWG610770S, BWG610773S, BWG610774S, BWG610795S, BWG610796S, BWG610797S, BWG610817S, BWG610822S, BWG610829S, BWG610830S, BWG610831S, BWG610833S, BWG610834S, BWG610835S, BWG610836S, BWG610837S, BWG610841S, BWG610843S, BWG610863S, BWG610876S, BWG610877S, BWG610881S, BWG610884S, BWG610885S, BWG610886S, BWG610887S, BWG610888S, BWG610889S, BWG610890S, BWG610891S, BWG610892S, BWG610930S, BWG610931S, BWG610934S, BWG610936S, BWG610938S, BWG610940S, BWG610949S, BWG610950S, BWG610951S, BWG610952S, BWG610970S, BWG610972S, BWG610973S, BWG610974S, BWG610975S, BWG610976S, BWG610979S, BWG610980S, BWG610985S, BWG610994S, BWG610996S, BWG610999S, BWG611000S, BWG611001S, BWG611016S, BWG611017S, BWG611018S, BWG611019S, BWG611020S, BWG611025S, BWG611026S, BWG611027S, BWG611028S, BWG611029S, BWG611038S, BWG611039S, BWG611041S, BWG611042S, BWG611046S, BWG611048S, BWG611049S, BWG611050S, BWG611060S, BWG611062S, BWG611068S, BWG611069S, BWG611070S, BWG611097S, BWG611098S, BWG611099S, BWG611100S, BWG611102S, BWG611104S, BWG611105S, BWG611107S, BWG611108S, BWG611109S, BWG611110S, BWG611111S, BWG611112S, BWG611113S, BWG611115S, BWG611116S, BWG611118S, BWG611122S, BWG611131S, BWG611132S, BWG611144S, BWG611146S, BWG611155S, BWG611157S, BWG611158S, BWG611165S, BWG611166S, BWG611167S, BWG611176S, BWG611177S, BWG611178S, BWG611179S, BWG611180S, BWG611181S, BWG611182S, BWG611183S, BWG611232S, BWG611233S, BWG611234S, BWG611235S, BWG611237S, BWG611238S, BWG611239S, BWG611240S, BWG611241S, BWG611242S, BWG611243S, BWG611247S, BWG611253S, BWG611254S, BWG611255S, BWG611256S, BWG611257S, BWG611258S, BWG611259S, BWG611260S, BWG611261S, BWG611262S, BWG611263S, BWG611264S, BWG611265S, BWG611266S, BWG611267S, BWG611268S, BWG611269S, BWG611270S, BWG611271S, BWG611272S, BWG611273S, BWG611274S, BWG611276S, BWG611277S, BWG611282S, BWG611284S, BWG611292S, BWG611302S, BWG611305S, BWG611309S, BWG611311S, BWG611312S, BWG611313S, BWG611314S, BWG611315S, BWG611316S, BWG611317S, BWG611318S, BWG611319S, BWG611323S, BWG611324S, BWG611325S, BWG611326S, BWG611327S, BWG611328S, BWG611329S, BWG611330S, BWG611331S, BWG611355S, BWG611356S, BWG611357S, BWG611358S, BWG611361S, BWG611366S, BWG611367S, BWG611368S, BWG611369S, BWG611391S, BWG611392S, BWG611394S, BWG611397S, BWG611399S, BWG611400S, BWG611401S, BWG611402S, BWG611403S, BWG611404S, BWG608570S, BWG608592S, BWG608593S, BWG608594S, BWG608595S, BWG609014S, BWG609018S, BWG609569S, BWG609570S, BWG609897S, BWG610660S, BWG610662S, BWG610663S, BWG610664S, BWG610665S, BWG610666S, BWG610667S, BWG608498S, BWG609179S, BWG609777S, BWG609532S, BWG609560S, BWG609561S, BWG609565S, BWG609772S, BWG609773S, BWG609818S, BWG609839S, BWG609843S, BWG609872S, BWG609901S, BWG609905S, BWG609909S, BWG609920S, BWG609199S, BWG609243S, BWG609249S, BWG609251S, BWG609254S, BWG610966S, BWG609834S, BWG609814S, BWG609816S, BWG609658S, BWG610159S, BWG610453S, BWG610454S, BWG610767S, BWG610872S, BWG610873S, BWG610896S, BWG610897S, BWG611168S, BWG611169S, BWG611170S, BWG611172S, BWG611200S, BWG611226S, BWG611227S, BWG611359S, BWG611360S, BWG611408S, BWG611634S, BWG611645S, BWG610117S, BWG610633S, BWG611383S, BWG611407S, BWG611615S, BWG611617S, BWG611618S, BWG611619S, BWG611620S, BWG611621S, BWG611629S
FEI Number 2182208
Recalling Firm/
Manufacturer
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
8200 Coral Sea St Ne
Mounds View MN 55112-4391
For Additional Information ContactTechnical Services
800-929-4043
Manufacturer Reason
for Recall
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
FDA Determined
Cause 2
Under Investigation by firm
ActionMedtronic issued an URGENT MEDICAL DEVICE CORRECTION NOTICE to its consignees on 05/10/2023 by UPS 2-day delivery or email. Consignees are asked to forward this notification to those who need to be aware within the organization and to any location which the devices have been distributed. The notice explained the problem and the risk and requested the following patient management recommendations: - Prophylactic device replacement is NOT recommended. - Program all HV therapy pathways B>AX in all therapy zones to minimize the risk for this issue. - Prioritize reprogramming patients who have both a history of HV therapy and Rx1 programmed AX>B. - For remaining patients with AX>B programming in any HV therapy sequence, schedule (with appropriate discretion) the next follow-up for in-clinic reprogramming to minimize potential for reduced- or no-energy HV therapies to occur. - Contact Medtronic Technical Services (1-800-929-4043) or your local representative if one of the following is observed as these may be an indication of either a device or lead-related issue: o Reduced- or no-energy HV therapy is displayed in Episode Text (regardless of programmed pathway) o A persistent drop of approximately 50% in RA, RV and LV pacing lead impedance measurements as this may be an indication of increased potential for a future reduced- or no-energy therapy. Update 10/10/2023: Beginning 09/27/2023, Medtronic issued communication via letter to consignees regarding the availability of the revised IFU and SmartSync Software. Consignees are asked to forward this notification to those who need to be aware within the organization and to any location which the devices have been distributed. Consignees are asked to complete and return a confirmation certificate (or equivalent record) confirming they have received the notification.
Quantity in Commerce5176 units
DistributionUS Nationwide - Worldwide Distribution
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
PMA DatabasePMAs with Product Code = LWS
PMAs with Product Code = LWS
-
-