• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall medtronic

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall medtronicsee related information
Date Initiated by FirmMay 10, 2023
Date PostedJune 16, 2023
Recall Status1 Open3, Classified
Recall NumberZ-1762-2023
Recall Event ID 92328
PMA NumberP010031S550 
Product Classification Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D) - Product Code NIK
ProductCRTD DTMB1D4 AMPLIA MRI US DF4, Model Number DTMB1D4; Implantable Cardioverter Defibrillators
Code Information GTIN 00643169720718, Lot Serial Numbers: RPJ204877H, RPJ205009H, RPJ205996H, RPJ206294H, RPJ205851H, RPJ206501H, RPJ205969H, RPJ205051H, RPJ206429H, RPJ205893H, RPJ205344H, RPJ206237H, RPJ206208H, RPJ206032H, RPJ206082H, RPJ206104H, RPJ205080H, RPJ205234H, RPJ206059H, RPJ205438H, RPJ206376H, RPJ205287H, RPJ205509H, RPJ206342H, RPJ205891H, RPJ204872H, RPJ204899H, RPJ204933H, RPJ204973H, RPJ205142H, RPJ205777H, RPJ205989H, RPJ206295H, RPJ205059H, RPJ205148H, RPJ205732H, RPJ205873H, RPJ206504H, RPJ205500H, RPJ206357H, RPJ204876H, RPJ206359H, RPJ206369H, RPJ205623H, RPJ206218H, RPJ205743H, RPJ206172H, RPJ204917H, RPJ205981H, RPJ205041H, RPJ205073H, RPJ205235H, RPJ205246H, RPJ205279H, RPJ206045H, RPJ205880H, RPJ206050H, RPJ206207H, RPJ205049H, RPJ204888H, RPJ205038H, RPJ205039H, RPJ205047H, RPJ205053H, RPJ205054H, RPJ205333H, RPJ205334H, RPJ205349H, RPJ205420H, RPJ205471H, RPJ205510H, RPJ205593H, RPJ205717H, RPJ205727H, RPJ205982H, RPJ206074H, RPJ206186H, RPJ206234H, RPJ204891H, RPJ205317H, RPJ205918H, RPJ205410H, RPJ206238H, RPJ206500H, RPJ204998H, RPJ205089H, RPJ205560H, RPJ205770H, RPJ206389H, RPJ205871H, RPJ205872H, RPJ205876H, RPJ205957H, RPJ206198H, RPJ206199H, RPJ206418H, RPJ206499H, RPJ204956H, RPJ205488H, RPJ205784H, RPJ205964H, RPJ206004H, RPJ206308H, RPJ206480H, RPJ205339H, RPJ205425H, RPJ205427H, RPJ205428H, RPJ205468H, RPJ206159H, RPJ206178H, RPJ206368H, RPJ206485H, RPJ206486H, RPJ206355H, RPJ206483H, RPJ205003H, RPJ205006H, RPJ205716H, RPJ205845H, RPJ206012H, RPJ206210H, RPJ205525H, RPJ205800H, RPJ206391H, RPJ205647H, RPJ205745H, RPJ205911H, RPJ205933H, RPJ205980H, RPJ206253H, RPJ206382H, RPJ206409H, RPJ206487H, RPJ204978H, RPJ205114H, RPJ205545H, RPJ205823H, RPJ205910H, RPJ205961H, RPJ206038H, RPJ206069H, RPJ206166H, RPJ206273H, RPJ206311H, RPJ206315H, RPJ206383H, RPJ206466H, RPJ206516H, RPJ205436H, RPJ205763H, RPJ205976H, RPJ205061H, RPJ205563H, RPJ205771H, RPJ206107H, RPJ205485H, RPJ206399H, RPJ205091H, RPJ205514H, RPJ206408H, RPJ206009H, RPJ205296H, RPJ205422H, RPJ205678H, RPJ205164H, RPJ206122H, RPJ205590H, RPJ205923H, RPJ204964H, RPJ205008H, RPJ205055H, RPJ205106H, RPJ205227H, RPJ205503H, RPJ205636H, RPJ205896H, RPJ206031H, RPJ206152H, RPJ206196H, RPJ206400H, RPJ205135H, RPJ205072H, RPJ205836H, RPJ205854H, RPJ205940H, RPJ206023H, RPJ206114H, RPJ206193H, RPJ206283H, RPJ206496H, RPJ205850H, RPJ205903H, RPJ205548H, RPJ205676H, RPJ205285H, RPJ205297H, RPJ205582H, RPJ205799H, RPJ205897H, RPJ205626H, RPJ206097H, RPJ206139H, RPJ206188H, RPJ206385H, RPJ205970H, RPJ205621H, RPJ205995H, RPJ205963H, RPJ206047H, RPJ206313H, RPJ204897H, RPJ205371H, RPJ205797H, RPJ205977H, RPJ206180H, RPJ206195H, RPJ206431H, RPJ205504H, RPJ204870H, RPJ204898H, RPJ204985H, RPJ205023H, RPJ205029H, RPJ205193H, RPJ205347H, RPJ205397H, RPJ205559H, RPJ205681H, RPJ205926H, RPJ205945H, RPJ205494H, RPJ206297H, RPJ204966H, RPJ205020H, RPJ205121H, RPJ205858H, RPJ206467H, RPJ205577H, RPJ205695H, RPJ206390H, RPJ206397H, RPJ206461H, RPJ205387H, RPJ205443H, RPJ206044H, RPJ205772H, RPJ206035H, RPJ205752H, RPJ205755H, RPJ205887H, RPJ206332H, RPJ206472H, RPJ206489H, RPJ205403H, RPJ205693H, RPJ205722H, RPJ206061H, RPJ205108H, RPJ205147H, RPJ205847H, RPJ206054H, RPJ206093H, RPJ206251H, RPJ206378H, RPJ205596H, RPJ204955H, RPJ205531H, RPJ205588H, RPJ205943H, RPJ206453H, RPJ206456H, RPJ206515H, RPJ205320H, RPJ205914H, RPJ205116H, RPJ204960H, RPJ205825H, RPJ205907H, RPJ205909H, RPJ206115H, RPJ206364H, RPJ205119H, RPJ205401H, RPJ205612H, RPJ205646H, RPJ205814H, RPJ206049H, RPJ206094H, RPJ206274H, RPJ206350H, RPJ206404H, RPJ205303H, RPJ205657H, RPJ205662H, RPJ205664H, RPJ205666H, RPJ205867H, RPJ205983H, RPJ206005H, RPJ206153H, RPJ206157H, RPJ206160H, RPJ206161H, RPJ206162H, RPJ205508H, RPJ205734H, RPJ206105H, RPJ205270H, RPJ205669H, RPJ206352H, RPJ206437H, RPJ205280H, RPJ205230H, RPJ205231H, RPJ203865H, RPJ205241H, RPJ205346H, RPJ205633H, RPJ205905H, RPJ205992H, RPJ206062H, RPJ206073H, RPJ206345H, RPJ204883H, RPJ205523H, RPJ205004H, RPJ205293H, RPJ205742H, RPJ205630H, RPJ205709H, RPJ205729H, RPJ205792H, RPJ205953H, RPJ205960H, RPJ206043H, RPJ206201H, RPJ206471H, RPJ206491H, RPJ204968H, RPJ205477H, RPJ205561H, RPJ205685H, RPJ205793H, RPJ205941H, RPJ205956H, RPJ206017H, RPJ206025H, RPJ206036H, RPJ206053H, RPJ206126H, RPJ206142H, RPJ206155H, RPJ206156H, RPJ206158H, RPJ206322H, RPJ205394H, RPJ205406H, RPJ205423H, RPJ205432H, RPJ206243H, RPJ206245H, RPJ206432H, RPJ205654H, RPJ205082H, RPJ205631H, RPJ205739H, RPJ206099H, RPJ204962H, RPJ205024H, RPJ205268H, RPJ205274H, RPJ205599H, RPJ205609H, RPJ205629H, RPJ205726H, RPJ205779H, RPJ205780H, RPJ205884H, RPJ205932H, RPJ205978H, RPJ206080H, RPJ206144H, RPJ206339H, RPJ205036H, RPJ206347H, RPJ206373H, RPJ205021H, RPJ205419H, RPJ206116H, RPJ205528H, RPJ206288H, RPJ206517H, RPJ205170H, RPJ206213H, RPJ205048H, RPJ205316H, RPJ205151H, RPJ206371H, RPJ205408H, RPJ205441H, RPJ205790H, RPJ205833H, RPJ206140H, RPJ206415H, RPJ205576H, RPJ205951H, RPJ206021H, RPJ206259H, RPJ206279H, RPJ204923H, RPJ205774H, RPJ205937H, RPJ206030H, RPJ204979H, RPJ205618H, RPJ205948H, RPJ206381H, RPJ206451H, RPJ205115H, RPJ205409H, RPJ205579H, RPJ206336H, RPJ205328H, RPJ205874H, RPJ206039H, RPJ206110H, RPJ206475H, RPJ204975H, RPJ205506H, RPJ206177H, RPJ204963H, RPJ205076H, RPJ205435H, RPJ206476H, RPJ205604H, RPJ205750H, RPJ206228H, RPJ205882H, RPJ205511H, RPJ205994H, RPJ204938H, RPJ205050H, RPJ205168H, RPJ205198H, RPJ205327H, RPJ205335H, RPJ205418H, RPJ205490H, RPJ205491H, RPJ205656H, RPJ205972H, RPJ206200H, RPJ205046H, RPJ205415H, RPJ205614H, RPJ204886H, RPJ204909H, RPJ204880H, RPJ205822H, RPJ205831H, RPJ206015H, RPJ206298H, RPJ206304H, RPJ205310H, RPJ205313H, RPJ205869H, RPJ206002H, RPJ206135H, RPJ205572H, RPJ205637H, RPJ205855H, RPJ205761H, RPJ205112H, RPJ205542H, RPJ206302H, RPJ206386H, RPJ206412H, RPJ205592H, RPJ204885H, RPJ204936H, RPJ205177H, RPJ205353H, RPJ204901H, RPJ205066H, RPJ205104H, RPJ205493H, RPJ205505H, RPJ204967H, RPJ205925H, RPJ206028H, RPJ206070H, RPJ206072H, RPJ206079H, RPJ206299H, RPJ204894H, RPJ204980H, RPJ204986H, RPJ204999H, RPJ205250H, RPJ205639H, RPJ206268H, RPJ206247H, RPJ206351H, RPJ206362H, RPJ205299H, RPJ206448H, RPJ205011H, RPJ205176H, RPJ205502H, RPJ205598H, RPJ206323H, RPJ206328H, RPJ206474H, RPJ206505H, RPJ205226H, RPJ205312H, RPJ205442H, RPJ205632H, RPJ205645H, RPJ205747H, RPJ205912H, RPJ205971H, RPJ206319H, RPJ205725H, RPJ206214H, RPJ205781H, RPJ205090H, RPJ205413H, RPJ205556H, RPJ205594H, RPJ206377H, RPJ206433H, RPJ206457H, RPJ204965H, RPJ204969H, RPJ205149H, RPJ205195H, RPJ206048H, RPJ206490H, RPJ206477H, RPJ206154H, RPJ206257H, RPJ204887H, RPJ205352H, RPJ205439H, RPJ206034H, RPJ206167H, RPJ206231H, RPJ206502H, RPJ205454H, RPJ205475H, RPJ205585H, RPJ205687H, RPJ205881H, RPJ205885H, RPJ205839H, RPJ206184H, RPJ205710H, RPJ204856H, RPJ204864H, RPJ205161H, RPJ205180H, RPJ205181H, RPJ205182H, RPJ205254H, RPJ205255H, RPJ205356H, RPJ205377H, RPJ205378H, RPJ205379H, RPJ205380H, RPJ205382H, RPJ205390H, RPJ205391H, RPJ205392H, RPJ206051H, RPJ206084H, RPJ206088H, RPJ206090H, RPJ206095H, RPJ206363H, RPJ206272H, RPJ205466H, RPJ206108H, RPJ205188H, RPJ205848H, RPJ206092H, RPJ205481H, RPJ204935H, RPJ205109H, RPJ205483H, RPJ205154H, RPJ205245H, RPJ204954H, RPJ205163H, RPJ205185H, RPJ205330H, RPJ205783H, RPJ205864H, RPJ206102H, RPJ206118H, RPJ206387H, RPJ205118H, RPJ205345H, RPJ205697H, RPJ206068H, RPJ205228H, RPJ205706H, RPJ205868H, RPJ206018H, RPJ204878H, RPJ205014H, RPJ205015H, RPJ205016H, RPJ205017H, RPJ205018H, RPJ205019H, RPJ205239H, RPJ205260H, RPJ205262H, RPJ205263H, RPJ205264H, RPJ205265H, RPJ205446H, RPJ205449H, RPJ205450H, RPJ205557H, RPJ205558H, RPJ205668H, RPJ205670H, RPJ205672H, RPJ205675H, RPJ205694H, RPJ205696H, RPJ205930H, RPJ205935H, RPJ205936H, RPJ206136H, RPJ206137H, RPJ206261H, RPJ206340H, RPJ206413H, RPJ205238H, RPJ205776H, RPJ205968H, RPJ206229H, RPJ206410H, RPJ206424H, RPJ206425H, RPJ206444H, RPJ206497H, RPJ205326H, RPJ206071H, RPJ206343H, RPJ206405H, RPJ206442H, RPJ206443H, RPJ206462H, RPJ204927H, RPJ205634H, RPJ205806H, RPJ206330H, RPJ205068H, RPJ205237H, RPJ205056H, RPJ204996H, RPJ205803H, RPJ205660H, RPJ205580H, RPJ206482H, RPJ206286H, RPJ206348H, RPJ206384H, RPJ205530H, RPJ206086H, RPJ205348H, RPJ205843H, RPJ205741H, RPJ206236H, RPJ206366H, RPJ205215H, RPJ204929H, RPJ205713H, RPJ206067H, RPJ205086H, RPJ205138H, RPJ205244H, RPJ205472H, RPJ205518H, RPJ206046H, RPJ206085H, RPJ206098H, RPJ206291H, RPJ204889H, RPJ205110H, RPJ205219H, RPJ205474H, RPJ205434H, RPJ205782H, RPJ205172H, RPJ205516H, RPJ204958H, RPJ205013H, RPJ205635H, RPJ206174H, RPJ204893H, RPJ205012H, RPJ206454H, RPJ205052H, RPJ205248H, RPJ205711H, RPJ205756H, RPJ205153H, RPJ206100H, RPJ206129H, RPJ206132H, RPJ205692H, RPJ204903H, RPJ204905H, RPJ204919H, RPJ205069H, RPJ205212H, RPJ205323H, RPJ205359H, RPJ205578H, RPJ205737H, RPJ205931H, RPJ205939H, RPJ205942H, RPJ206089H, RPJ206134H, RPJ206148H, RPJ206206H, RPJ206344H, RPJ206464H, RPJ206470H, RPJ205189H, RPJ205190H, RPJ205376H, RPJ206150H, RPJ206452H, RPJ205156H, RPJ206010H, RPJ205111H, RPJ205866H, RPJ205459H, RPJ205846H, RPJ205875H, RPJ204944H, RPJ205288H, RPJ205307H, RPJ205479H, RPJ205569H, RPJ205570H, RPJ205625H, RPJ205648H, RPJ205746H, RPJ205974H, RPJ205984H, RPJ206037H, RPJ206127H, RPJ206221H, RPJ206223H, RPJ206226H, RPJ206233H, RPJ206241H, RPJ206244H, RPJ206248H, RPJ206434H, RPJ206479H, RPJ204977H, RPJ205611H, RPJ205987H, RPJ206027H, RPJ205247H, RPJ205343H, RPJ205487H, RPJ205705H, RPJ206204H, RPJ205549H, RPJ205555H, RPJ205959H, RPJ205232H, RPJ205286H, RPJ205040H, RPJ205617H, RPJ205944H, RPJ205571H, RPJ205584H, RPJ205986H, RPJ205988H, RPJ206190H, RPJ206224H, RPJ206365H, RPJ206374H, RPJ206439H, RPJ206445H, RPJ206478H, RPJ206494H, RPJ206498H, RPJ206375H, RPJ205603H, RPJ205952H, RPJ205718H, RPJ205764H, RPJ204926H, RPJ205536H, RPJ206022H, RPJ205720H, RPJ205728H, RPJ205966H, RPJ206013H, RPJ205813H, RPJ204882H, RPJ206310H, RPJ206380H, RPJ206398H, RPJ204875H, RPJ205440H, RPJ206401H, RPJ206417H, RPJ206514H, RPJ204867H, RPJ205217H, RPJ206327H, RPJ206411H, RPJ206473H, RPJ205804H, RPJ205532H, RPJ205708H, RPJ205369H, RPJ206316H, RPJ205691H, RPJ205889H, RPJ205319H, RPJ205824H, RPJ206449H, RPJ206481H, RPJ205534H, RPJ205077H, RPJ205210H, RPJ205985H, RPJ205257H, RPJ205751H, RPJ206138H, RPJ206292H, RPJ205486H, RPJ205958H, RPJ206197H, RPJ205495H, RPJ205827H, RPJ205900H, RPJ206222H, RPJ203863H, RPJ205298H, RPJ205539H, RPJ205927H, RPJ205100H, RPJ205191H, RPJ206171H, RPJ205554H, RPJ205575H, RPJ206416H, RPJ205081H, RPJ205587H, RPJ205661H, RPJ205842H, RPJ205921H, RPJ206026H, RPJ206123H, RPJ206289H, RPJ206317H, RPJ205898H, RPJ206392H, RPJ205922H, RPJ206256H, RPJ206346H, RPJ205489H, RPJ206014H, RPJ204914H, RPJ204922H, RPJ205130H, RPJ205342H, RPJ205744H, RPJ206109H, RPJ206263H, RPJ206287H, RPJ205096H, RPJ205175H, RPJ205967H, RPJ206361H, RPJ205324H, RPJ205197H, RPJ205476H, RPJ205267H, RPJ205424H, RPJ205426H, RPJ205079H, RPJ205216H, RPJ206081H, RPJ206125H, RPJ206360H, RPJ204890H, RPJ204896H, RPJ205480H, RPJ205830H, RPJ205902H, RPJ205849H, RPJ205906H, RPJ206438H, RPJ205856H, RPJ206168H, RPJ206426H, RPJ206458H, RPJ205010H, RPJ205186H, RPJ205512H, RPJ205025H, RPJ205628H, RPJ206113H, RPJ205315H, RPJ205546H, RPJ205610H, RPJ205835H, RPJ205853H, RPJ205888H, RPJ205915H, RPJ205934H, RPJ206185H, RPJ206463H, RPJ205565H, RPJ206003H, RPJ206303H, RPJ206083H, RPJ206314H, RPJ205460H, RPJ205533H, RPJ205568H, RPJ205622H, RPJ205643H, RPJ205714H, RPJ205719H, RPJ205786H, RPJ205913H, RPJ205938H, RPJ206042H, RPJ206354H, RPJ206419H, RPJ205099H, RPJ205098H, RPJ206029H, RPJ204868H, RPJ205543H, RPJ205993H, RPJ206427H, RPJ206493H, RPJ205615H, RPJ206420H, RPJ206421H, RPJ206422H, RPJ205463H, RPJ205221H, RPJ206170H, RPJ206242H, RPJ205805H, RPJ205807H, RPJ205194H, RPJ205067H, RPJ205028H, RPJ205496H, RPJ205567H, RPJ206305H, RPJ206149H, RPJ205075H, RPJ206414H, RPJ205445H, RPJ206460H, RPJ205686H, RPJ205857H, RPJ205065H, RPJ204884H, RPJ206147H, RPJ206269H, RPJ206296H, RPJ206459H, RPJ205355H, RPJ205601H, RPJ205607H, RPJ205620H, RPJ205808H, RPJ205878H, RPJ205916H, RPJ206106H, RPJ206111H, RPJ206112H, RPJ206120H, RPJ206128H, RPJ206163H, RPJ206165H, RPJ206203H, RPJ206252H, RPJ206258H, RPJ206271H, RPJ206281H, RPJ206324H, RPJ206333H, RPJ206349H, RPJ206441H, RPJ206469H, RPJ206506H, RPJ205204H, RPJ205205H, RPJ205206H, RPJ205208H, RPJ205209H, RPJ206321H, RPJ206503H, RPJ205456H, RPJ205870H, RPJ206335H, RPJ206488H, RPJ205573H, RPJ205649H, RPJ205650H, RPJ205651H, RPJ205677H, RPJ205766H, RPJ205767H, RPJ206000H, RPJ206219H, RPJ206395H, RPJ206406H, RPJ206511H, RPJ206041H, RPJ206052H, RPJ206510H, RPJ206337H, RPJ204855H, RPJ204857H, RPJ204858H, RPJ204859H, RPJ204861H, RPJ204862H, RPJ204863H, RPJ204865H, RPJ204873H, RPJ204874H, RPJ204892H, RPJ204895H, RPJ204904H, RPJ204940H, RPJ204942H, RPJ204943H, RPJ204945H, RPJ204946H, RPJ204947H, RPJ204948H, RPJ204949H, RPJ204951H, RPJ204952H, RPJ204953H, RPJ204987H, RPJ204988H, RPJ204989H, RPJ204990H, RPJ204992H, RPJ204995H, RPJ205000H, RPJ205001H, RPJ205002H, RPJ205034H, RPJ205060H, RPJ205113H, RPJ205117H, RPJ205122H, RPJ205123H, RPJ205124H, RPJ205125H, RPJ205127H, RPJ205128H, RPJ205129H, RPJ205131H, RPJ205132H, RPJ205133H, RPJ205134H, RPJ205136H, RPJ205137H, RPJ205139H, RPJ205140H, RPJ205141H, RPJ205143H, RPJ205144H, RPJ205146H, RPJ205152H, RPJ205157H, RPJ205158H, RPJ205159H, RPJ205166H, RPJ205167H, RPJ205169H, RPJ205174H, RPJ205201H, RPJ205203H, RPJ205207H, RPJ205213H, RPJ205220H, RPJ205261H, RPJ205284H, RPJ205290H, RPJ205291H, RPJ205292H, RPJ205294H, RPJ205295H, RPJ205302H, RPJ205304H, RPJ205305H, RPJ205308H, RPJ205321H, RPJ205322H, RPJ205331H, RPJ205383H, RPJ205384H, RPJ205386H, RPJ205389H, RPJ205398H, RPJ205519H, RPJ205521H, RPJ205527H, RPJ205537H, RPJ205538H, RPJ205544H, RPJ205550H, RPJ205562H, RPJ205597H, RPJ205608H, RPJ205680H, RPJ205682H, RPJ205730H, RPJ205731H, RPJ205733H, RPJ205735H, RPJ205791H, RPJ205802H, RPJ205809H, RPJ205817H, RPJ205820H, RPJ205821H, RPJ205828H, RPJ205829H, RPJ205832H, RPJ205837H, RPJ205838H, RPJ205883H, RPJ206513H, RPJ206518H, RPJ206455H; GTIN 00763000059675, Lot Serial Numbers: RPJ601555S, RPJ601599S, RPJ601649S, RPJ600734S, RPJ601458S, RPJ601526S, RPJ601520S, RPJ601638S, RPJ601642S, RPJ601172S, RPJ601273S, RPJ602085S, RPJ601218S, RPJ601351S, RPJ601296S, RPJ601501S, RPJ601843S, RPJ602037S, RPJ601230S, RPJ601265S, RPJ601272S, RPJ601367S, RPJ601380S, RPJ601384S, RPJ601389S, RPJ601617S, RPJ602036S, RPJ602171S, RPJ600687S, RPJ601586S, RPJ600429S, RPJ600565S, RPJ600036S, RPJ601840S, RPJ600499S, RPJ602091S, RPJ601305S, RPJ600597S, RPJ600688S, RPJ600714S, RPJ601040S, RPJ601041S, RPJ600556S, RPJ600743S, RPJ601342S, RPJ601364S, RPJ601425S, RPJ601979S, RPJ600407S, RPJ601761S, RPJ601418S, RPJ600033S, RPJ600697S, RPJ601574S, RPJ601722S, RPJ601727S, RPJ600443S, RPJ600465S, RPJ600955S, RPJ601031S, RPJ601039S, RPJ601340S, RPJ601276S, RPJ600492S, RPJ601024S, RPJ601452S, RPJ600678S, RPJ600321S, RPJ601028S, RPJ601597S, RPJ601763S, RPJ601817S, RPJ600028S, RPJ600170S, RPJ601802S, RPJ600853S, RPJ601730S, RPJ601200S, RPJ601333S, RPJ600045S, RPJ601846S, RPJ600527S, RPJ600677S, RPJ600904S, RPJ601078S, RPJ601156S, RPJ600113S, RPJ600218S, RPJ600056S, RPJ600654S, RPJ600680S, RPJ600016S, RPJ600017S, RPJ600021S, RPJ600022S, RPJ600023S, RPJ600188S, RPJ600189S, RPJ600190S, RPJ600191S, RPJ600192S, RPJ601622S, RPJ601635S, RPJ601828S, RPJ601833S, RPJ602092S, RPJ601121S, RPJ601175S, RPJ601356S, RPJ601528S, RPJ600158S, RPJ601650S, RPJ601651S, RPJ601726S, RPJ601739S, RPJ601740S, RPJ601741S, RPJ601767S, RPJ601768S, RPJ601797S, RPJ601832S, RPJ601953S, RPJ601980S, RPJ601989S, RPJ600435S, RPJ600888S, RPJ601023S, RPJ601071S, RPJ601080S, RPJ601168S, RPJ601169S, RPJ601215S, RPJ601324S, RPJ600356S, RPJ600756S, RPJ601771S, RPJ602131S, RPJ600358S, RPJ600406S, RPJ600860S, RPJ600882S, RPJ600942S, RPJ600948S, RPJ601001S, RPJ601093S, RPJ601482S, RPJ600890S, RPJ601489S, RPJ601721S, RPJ601894S, RPJ600575S, RPJ600595S, RPJ600672S, RPJ600720S, RPJ600771S, RPJ600848S, RPJ600992S, RPJ601018S, RPJ601043S, RPJ601074S, RPJ601229S, RPJ601312S, RPJ601359S, RPJ601363S, RPJ601409S, RPJ601414S, RPJ601437S, RPJ600634S, RPJ600276S, RPJ600446S, RPJ600757S, RPJ601598S, RPJ601707S, RPJ600363S, RPJ600548S, RPJ600579S, RPJ600766S, RPJ600792S, RPJ601056S, RPJ600354S, RPJ600493S, RPJ600511S, RPJ600546S, RPJ600648S, RPJ600752S, RPJ601008S, RPJ601151S, RPJ600901S, RPJ601038S, RPJ600598S, RPJ600735S, RPJ600966S, RPJ600043S, RPJ601396S, RPJ601753S, RPJ600095S, RPJ600332S, RPJ600897S, RPJ601103S, RPJ601170S, RPJ601551S, RPJ601697S, RPJ600040S, RPJ601536S, RPJ600394S, RPJ600593S, RPJ601743S, RPJ601809S, RPJ600900S, RPJ600986S, RPJ601449S, RPJ600409S, RPJ600581S, RPJ600582S, RPJ600666S, RPJ600696S, RPJ600715S, RPJ600979S, RPJ601085S, RPJ601142S, RPJ601268S, RPJ600207S, RPJ601600S, RPJ600011S, RPJ600161S, RPJ600348S, RPJ601304S, RPJ601503S, RPJ600110S, RPJ601593S, RPJ601626S, RPJ601251S, RPJ601416S, RPJ600099S, RPJ600107S, RPJ600585S, RPJ600127S, RPJ600160S, RPJ600177S, RPJ600283S, RPJ600682S, RPJ601048S, RPJ601058S, RPJ601184S, RPJ601563S, RPJ601854S, RPJ600387S, RPJ600943S, RPJ601007S, RPJ601072S, RPJ601493S, RPJ601531S, RPJ601633S, RPJ601728S, RPJ600122S, RPJ600126S, RPJ600823S, RPJ601097S, RPJ600502S, RPJ600649S, RPJ600709S, RPJ600932S, RPJ601591S, RPJ600244S, RPJ600613S, RPJ600964S, RPJ600432S, RPJ600434S, RPJ600359S, RPJ600037S, RPJ601214S, RPJ601480S, RPJ601587S, RPJ601629S, RPJ601746S, RPJ601810S, RPJ601981S, RPJ600178S, RPJ600692S, RPJ600707S, RPJ600971S, RPJ601279S, RPJ601467S, RPJ601154S, RPJ601297S, RPJ601386S, RPJ601462S, RPJ600134S, RPJ600137S, RPJ600767S, RPJ600881S, RPJ600884S, RPJ600962S, RPJ601106S, RPJ601341S, RPJ600664S, RPJ601241S, RPJ601538S, RPJ601825S, RPJ601826S, RPJ601930S, RPJ601954S, RPJ600052S, RPJ600498S, RPJ600655S, RPJ601286S, RPJ601931S, RPJ602032S, RPJ600451S, RPJ601644S, RPJ601851S, RPJ601865S, RPJ600368S, RPJ600827S, RPJ600905S, RPJ600907S, RPJ601264S, RPJ601627S, RPJ601640S, RPJ601950S, RPJ601116S, RPJ601126S, RPJ601530S, RPJ602165S, RPJ602024S, RPJ600128S, RPJ600186S, RPJ600352S, RPJ600448S, RPJ600468S, RPJ600791S, RPJ601255S, RPJ601256S, RPJ601369S, RPJ601755S, RPJ600395S, RPJ600450S, RPJ600518S, RPJ600539S, RPJ600586S, RPJ600807S, RPJ600923S, RPJ600929S, RPJ600994S, RPJ600998S, RPJ601066S, RPJ601114S, RPJ601338S, RPJ601486S, RPJ600312S, RPJ601764S, RPJ601185S, RPJ601391S, RPJ601673S, RPJ600370S, RPJ600384S, RPJ600421S, RPJ601263S, RPJ601298S, RPJ601368S, RPJ601161S, RPJ601176S, RPJ601227S, RPJ601529S, RPJ601787S, RPJ601724S, RPJ600418S, RPJ601036S, RPJ601094S, RPJ601095S, RPJ600223S, RPJ600329S, RPJ601919S, RPJ600377S, RPJ601303S, RPJ601309S, RPJ601491S, RPJ602160S, RPJ600333S, RPJ601834S, RPJ601177S, RPJ600184S, RPJ601756S, RPJ600152S, RPJ600306S, RPJ600408S, RPJ600626S, RPJ600770S, RPJ600830S, RPJ600831S, RPJ600871S, RPJ600912S, RPJ600996S, RPJ601127S, RPJ601148S, RPJ601159S, RPJ601194S, RPJ601274S, RPJ601277S, RPJ601415S, RPJ600415S, RPJ600673S, RPJ600129S, RPJ600326S, RPJ600341S, RPJ600439S, RPJ601928S, RPJ600960S, RPJ600976S, RPJ600292S, RPJ600101S, RPJ602055S, RPJ601772S, RPJ601848S, RPJ600419S, RPJ600939S, RPJ601374S, RPJ601845S, RPJ601912S, RPJ600327S, RPJ600530S, RPJ600273S, RPJ600117S, RPJ600645S, RPJ600788S, RPJ602202S, RPJ601575S, RPJ601618S, RPJ601643S, RPJ601674S, RPJ601693S, RPJ601720S, RPJ601819S, RPJ602108S, RPJ600381S, RPJ600437S, RPJ600487S, RPJ600508S, RPJ600630S, RPJ600913S, RPJ600937S, RPJ600997S, RPJ601005S, RPJ601029S, RPJ601050S, RPJ601163S, RPJ601284S, RPJ601321S, RPJ601329S, RPJ601404S, RPJ601417S, RPJ601502S, RPJ600096S, RPJ600156S, RPJ600753S, RPJ600042S, RPJ601182S, RPJ601434S, RPJ600067S, RPJ600246S, RPJ601893S, RPJ601895S, RPJ601896S, RPJ601639S, RPJ601645S, RPJ602116S, RPJ600462S, RPJ600773S, RPJ600867S, RPJ601030S, RPJ601220S, RPJ601270S, RPJ601379S, RPJ601441S, RPJ601523S, RPJ600441S, RPJ600256S, RPJ600026S, RPJ601115S, RPJ601694S, RPJ601732S, RPJ601791S, RPJ601866S, RPJ601974S, RPJ600544S, RPJ600685S, RPJ600898S, RPJ601224S, RPJ601731S, RPJ601939S, RPJ602120S, RPJ602137S, RPJ600503S, RPJ601549S, RPJ601603S, RPJ600440S, RPJ601346S, RPJ600012S, RPJ600089S, RPJ600542S, RPJ600591S, RPJ600703S, RPJ600899S, RPJ601336S, RPJ601339S, RPJ601393S, RPJ601397S, RPJ601401S, RPJ601387S, RPJ600215S, RPJ600242S, RPJ600366S, RPJ600416S, RPJ602013S, RPJ600427S, RPJ600551S, RPJ600713S, RPJ600728S, RPJ600864S, RPJ600869S, RPJ600982S, RPJ601016S, RPJ601070S, RPJ601424S, RPJ601488S, RPJ600100S, RPJ600759S, RPJ601250S, RPJ601710S, RPJ601808S, RPJ600642S, RPJ600814S, RPJ600181S, RPJ601535S, RPJ600185S, RPJ600621S, RPJ600705S, RPJ600829S, RPJ600969S, RPJ600927S, RPJ600555S, RPJ600833S, RPJ600895S, RPJ601204S, RPJ601376S, RPJ601423S, RPJ601460S, RPJ601687S, RPJ600817S, RPJ600868S, RPJ600952S, RPJ600175S, RPJ600182S, RPJ600147S, RPJ601785S, RPJ601786S, RPJ600509S, RPJ600780S, RPJ600959S, RPJ601620S, RPJ600250S, RPJ600216S, RPJ600300S, RPJ600686S, RPJ600870S, RPJ600915S, RPJ600918S, RPJ600972S, RPJ601047S, RPJ601076S, RPJ600222S, RPJ601869S, RPJ601580S, RPJ601748S, RPJ600320S, RPJ600589S, RPJ600646S, RPJ600698S, RPJ600772S, RPJ600886S, RPJ601133S, RPJ601134S, RPJ601299S, RPJ601378S, RPJ601395S, RPJ600105S, RPJ601890S, RPJ601601S, RPJ600205S, RPJ601572S, RPJ601641S, RPJ601998S, RPJ602097S, RPJ600919S, RPJ601075S, RPJ601181S, RPJ601461S, RPJ601206S, RPJ601257S, RPJ600532S, RPJ600606S, RPJ600241S, RPJ600374S, RPJ600708S, RPJ601569S, RPJ601917S, RPJ600232S, RPJ600299S, RPJ600797S, RPJ600887S, RPJ601209S, RPJ601210S, RPJ601320S, RPJ601959S, RPJ600148S, RPJ600217S, RPJ601517S, RPJ600041S, RPJ600422S, RPJ600460S, RPJ600783S, RPJ600802S, RPJ600865S, RPJ600961S, RPJ601032S, RPJ601063S, RPJ601104S, RPJ601118S, RPJ601125S, RPJ601293S, RPJ601360S, RPJ601481S, RPJ600145S, RPJ601758S, RPJ601465S, RPJ601483S, RPJ600903S, RPJ600950S, RPJ601358S, RPJ600578S, RPJ601193S, RPJ601197S, RPJ601198S, RPJ600090S, RPJ601014S, RPJ600639S, RPJ600717S, RPJ600856S, RPJ600917S, RPJ600014S, RPJ600411S, RPJ600514S, RPJ600704S, RPJ601656S, RPJ601660S, RPJ601792S, RPJ601811S, RPJ601812S, RPJ601813S, RPJ602061S, RPJ600055S, RPJ600102S, RPJ600346S, RPJ600355S, RPJ600558S, RPJ600560S, RPJ600681S, RPJ600995S, RPJ601015S, RPJ601084S, RPJ601147S, RPJ601234S, RPJ601261S, RPJ601300S, RPJ601302S, RPJ601345S, RPJ601349S, RPJ600077S, RPJ601090S, RPJ601589S, RPJ601473S, RPJ600048S, RPJ601861S, RPJ601844S, RPJ601961S, RPJ602031S, RPJ602152S, RPJ600594S, RPJ600954S, RPJ601077S, RPJ601447S, RPJ601463S, RPJ601544S, RPJ601662S, RPJ601818S, RPJ602019S, RPJ600650S, RPJ600769S, RPJ600967S, RPJ601111S, RPJ601123S, RPJ601136S, RPJ601254S, RPJ600157S, RPJ600267S, RPJ600727S, RPJ600024S, RPJ600238S, RPJ600457S, RPJ600500S, RPJ600505S, RPJ600603S, RPJ600850S, RPJ600859S, RPJ601665S, RPJ601799S, RPJ600496S, RPJ601332S, RPJ601568S, RPJ601750S, RPJ601823S, RPJ601935S, RPJ601062S, RPJ601235S, RPJ601278S, RPJ602034S, RPJ601280S, RPJ600455S, RPJ601314S, RPJ601790S, RPJ600945S, RPJ602140S, RPJ602141S, RPJ602011S, RPJ600115S, RPJ600334S, RPJ600282S, RPJ600293S, RPJ600491S, RPJ600605S, RPJ600668S, RPJ600221S, RPJ601444S, RPJ601445S, RPJ600916S, RPJ601547S, RPJ601784S, RPJ602080S, RPJ600417S, RPJ600729S, RPJ600953S, RPJ600193S, RPJ600793S, RPJ600891S, RPJ601079S, RPJ601155S, RPJ601957S, RPJ601160S, RPJ601186S, RPJ601189S, RPJ601191S, RPJ602176S, RPJ601319S, RPJ600392S, RPJ600804S, RPJ600212S, RPJ600656S, RPJ600684S, RPJ601153S, RPJ600653S, RPJ600338S, RPJ601773S, RPJ600803S, RPJ600924S, RPJ601192S, RPJ601852S, RPJ601107S, RPJ601451S, RPJ601615S, RPJ600124S, RPJ600689S, RPJ600902S, RPJ600131S, RPJ602115S, RPJ600351S, RPJ600281S, RPJ600150S, RPJ601242S, RPJ600314S, RPJ600313S, RPJ600754S, RPJ600159S, RPJ601259S, RPJ602038S, RPJ600535S, RPJ600670S, RPJ600763S, RPJ600874S, RPJ601000S, RPJ601431S, RPJ600577S, RPJ600778S, RPJ601938S, RPJ601655S, RPJ601915S, RPJ601916S, RPJ602096S, RPJ600981S, RPJ601216S, RPJ601373S, RPJ601510S, RPJ601821S, RPJ601822S, RPJ601835S, RPJ601870S, RPJ600910S, RPJ600957S, RPJ601225S, RPJ601477S, RPJ601478S, RPJ601484S, RPJ601516S, RPJ602149S, RPJ600305S, RPJ600013S, RPJ601929S, RPJ600199S, RPJ600200S, RPJ600482S, RPJ600612S, RPJ600745S, RPJ600746S, RPJ601723S, RPJ601233S, RPJ601269S, RPJ601271S, RPJ601282S, RPJ601429S, RPJ602002S, RPJ602003S, RPJ602005S, RPJ602182S, RPJ602183S, RPJ601067S, RPJ600958S, RPJ601347S, RPJ601579S, RPJ601609S, RPJ601873S, RPJ601937S, RPJ602033S, RPJ602168S, RPJ600467S, RPJ600476S, RPJ600495S, RPJ600525S, RPJ600563S, RPJ600570S, RPJ600725S, RPJ600798S, RPJ600821S, RPJ600825S, RPJ600826S, RPJ600835S, RPJ600845S, RPJ600854S, RPJ600893S, RPJ600908S, RPJ600928S, RPJ600933S, RPJ600935S, RPJ600973S, RPJ600980S, RPJ601049S, RPJ601179S, RPJ601180S, RPJ601311S, RPJ601440S, RPJ601464S, RPJ601497S, RPJ601500S, RPJ601509S, RPJ601888S, RPJ600968S, RPJ601139S, RPJ601402S, RPJ601777S, RPJ601672S, RPJ601678S, RPJ601682S, RPJ601714S, RPJ601836S, RPJ601874S, RPJ601878S, RPJ601889S, RPJ601975S, RPJ600385S, RPJ600388S, RPJ600657S, RPJ600663S, RPJ600930S, RPJ601017S, RPJ601061S, RPJ601064S, RPJ601087S, RPJ601289S, RPJ601291S, RPJ601512S, RPJ602126S, RPJ601947S, RPJ601976S, RPJ601978S, RPJ602098S, RPJ601621S, RPJ601897S, RPJ601952S, RPJ602042S, RPJ600748S, RPJ601178S, RPJ601240S, RPJ601420S, RPJ602178S, RPJ602093S, RPJ601675S, RPJ601676S, RPJ601841S, RPJ601004S, RPJ601144S, RPJ601606S, RPJ601403S, RPJ601548S, RPJ602018S, RPJ601578S, RPJ601666S, RPJ600765S, RPJ601034S, RPJ601960S, RPJ600574S, RPJ600623S, RPJ600691S, RPJ600779S, RPJ600787S, RPJ601013S, RPJ601122S, RPJ601239S, RPJ601456S, RPJ601470S, RPJ602101S, RPJ600947S, RPJ601948S, RPJ600084S, RPJ600162S, RPJ600050S, RPJ601774S, RPJ601681S, RPJ601882S, RPJ601789S, RPJ600515S, RPJ601044S, RPJ600271S, RPJ600063S, RPJ601754S, RPJ600403S, RPJ600914S, RPJ601222S, RPJ601348S, RPJ600896S, RPJ601365S, RPJ601701S, RPJ600442S, RPJ600811S, RPJ600813S, RPJ600883S, RPJ601232S, RPJ601408S, RPJ601412S, RPJ601611S, RPJ600252S, RPJ600800S, RPJ601435S, RPJ601962S, RPJ601692S, RPJ600566S, RPJ600034S, RPJ600195S, RPJ600197S, RPJ600278S, RPJ601045S, RPJ600562S, RPJ600892S, RPJ601357S, RPJ600284S, RPJ602049S, RPJ600452S, RPJ600796S, RPJ600855S, RPJ600872S, RPJ601288S, RPJ600198S, RPJ600315S, RPJ600062S, RPJ601632S, RPJ601659S, RPJ601680S, RPJ601698S, RPJ601780S, RPJ600353S, RPJ600635S, RPJ600711S, RPJ600726S, RPJ600819S, RPJ601105S, RPJ601124S, RPJ601157S, RPJ601188S, RPJ601247S, RPJ601450S, RPJ601775S, RPJ602196S, RPJ601831S, RPJ600059S, RPJ601469S, RPJ600832S, RPJ600975S, RPJ601083S, RPJ601527S, RPJ600240S, RPJ601422S, RPJ600130S, RPJ600977S, RPJ600978S, RPJ601069S, RPJ601566S, RPJ601588S, RPJ601716S, RPJ601719S, RPJ601839S, RPJ601855S, RPJ601858S, RPJ601859S, RPJ601863S, RPJ601994S, RPJ602072S, RPJ602073S, RPJ600425S, RPJ600456S, RPJ600470S, RPJ600507S, RPJ600521S, RPJ600523S, RPJ600557S, RPJ600561S, RPJ600600S, RPJ600601S, RPJ600602S, RPJ600629S, RPJ600644S, RPJ600744S, RPJ600768S, RPJ600842S, RPJ600844S, RPJ600846S, RPJ600885S, RPJ600889S, RPJ600949S, RPJ600956S, RPJ600990S, RPJ600991S, RPJ601052S, RPJ601055S, RPJ601065S, RPJ601073S, RPJ601086S, RPJ601183S, RPJ601217S, RPJ601236S, RPJ601350S, RPJ601372S, RPJ601388S, RPJ601390S, RPJ601392S, RPJ601394S, RPJ601398S, RPJ601406S, RPJ601448S, RPJ601472S, RPJ601479S, RPJ601487S, RPJ601511S, RPJ601514S, RPJ600529S, RPJ601540S, RPJ602147S, RPJ601150S, RPJ600398S, RPJ600875S, RPJ600479S, RPJ602157S, RPJ601515S, RPJ600295S, RPJ600545S, RPJ600571S, RPJ600633S, RPJ600736S, RPJ601002S, RPJ601226S, RPJ602159S, RPJ601554S, RPJ600480S, RPJ600524S, RPJ601162S, RPJ600339S, RPJ602025S, RPJ601541S, RPJ601543S, RPJ600799S, RPJ601201S, RPJ600187S, RPJ600206S, RPJ600297S, RPJ601550S, RPJ601167S, RPJ601534S, RPJ601779S, RPJ601881S, RPJ600690S, RPJ600506S, RPJ600880S, RPJ601089S, RPJ601344S, RPJ601876S, RPJ600367S, RPJ600371S, RPJ600372S, RPJ600379S, RPJ600380S, RPJ600389S, RPJ600390S, RPJ600485S, RPJ600504S, RPJ600512S, RPJ600553S, RPJ600554S, RPJ600568S, RPJ600569S, RPJ600604S, RPJ600609S, RPJ600669S, RPJ601577S, RPJ601590S, RPJ601315S, RPJ600516S, RPJ600428S, RPJ601742S, RPJ600694S, RPJ600786S, RPJ600430S, RPJ600965S, RPJ601059S, RPJ600810S, RPJ600909S, RPJ601494S, RPJ601498S, RPJ601637S, RPJ600118S, RPJ600534S, RPJ600985S, RPJ601442S, RPJ601521S, RPJ601699S, RPJ601798S, RPJ601803S, RPJ600590S, RPJ600636S, RPJ600777S, RPJ601203S, RPJ601326S, RPJ601986S, RPJ601573S, RPJ600851S, RPJ601323S, RPJ601327S, RPJ602082S, RPJ600039S, RPJ600180S, RPJ601688S, RPJ601696S, RPJ600974S, RPJ601459S, RPJ600733S, RPJ600776S, RPJ601684S, RPJ601685S, RPJ600652S, RPJ600712S, RPJ600046S, RPJ601553S, RPJ600459S, RPJ601485S, RPJ601914S, RPJ600311S, RPJ600651S, RPJ600836S, RPJ601051S, RPJ601096S, RPJ601212S, RPJ601371S, RPJ600138S, RPJ600820S, RPJ600213S, RPJ601751S, RPJ601208S, RPJ600194S, RPJ601145S, RPJ601570S, RPJ601571S, RPJ601592S, RPJ601605S, RPJ601747S, RPJ601793S, RPJ600478S, RPJ600528S, RPJ600724S, RPJ601141S, RPJ601196S, RPJ601199S, RPJ601253S, RPJ601343S, RPJ601382S, RPJ601436S, RPJ601492S, RPJ600399S, RPJ600404S, RPJ600461S, RPJ600475S, RPJ600543S, RPJ600564S, RPJ600615S, RPJ600700S, RPJ600782S, RPJ600805S, RPJ600806S, RPJ600839S, RPJ600940S, RPJ601468S, RPJ601519S, RPJ600402S, RPJ600401S, RPJ600693S, RPJ600749S, RPJ601679S, RPJ601718S, RPJ601745S, RPJ600345S, RPJ600447S, RPJ600592S, RPJ600627S, RPJ601190S, RPJ601237S, RPJ601439S, RPJ601857S, RPJ602021S, RPJ601088S, RPJ601381S, RPJ601717S, RPJ601704S, RPJ601337S, RPJ601733S, RPJ600208S, RPJ600710S, RPJ601433S, RPJ601219S, RPJ601623S, RPJ601653S, RPJ601667S, RPJ601946S, RPJ601474S, RPJ600093S, RPJ600774S, RPJ600310S, RPJ600051S, RPJ600202S, RPJ601856S, RPJ601862S, RPJ602081S, RPJ600405S, RPJ600541S, RPJ600547S, RPJ600599S, RPJ600723S, RPJ600840S, RPJ601238S, RPJ601366S, RPJ601507S, RPJ600393S, RPJ600449S, RPJ602048S, RPJ602127S, RPJ600665S, RPJ600816S, RPJ601158S, RPJ600931S, RPJ601244S, RPJ601248S, RPJ600473S, RPJ600494S, RPJ600637S, RPJ600785S, RPJ600824S, RPJ601636S, RPJ600335S, RPJ601195S, RPJ600230S, RPJ601713S, RPJ601668S, RPJ601670S, RPJ601671S, RPJ601711S, RPJ600730S, RPJ601354S, RPJ601355S, RPJ600834S, RPJ600794S, RPJ602139S, RPJ601594S, RPJ601663S, RPJ600436S, RPJ600522S, RPJ600583S, RPJ600747S, RPJ600847S, RPJ601140S, RPJ601432S, RPJ601595S, RPJ601752S, RPJ601664S, RPJ600400S, RPJ600484S, RPJ600852S, RPJ601322S, RPJ602099S, RPJ600526S, RPJ600211S, RPJ601518S, RPJ601602S, RPJ601782S, RPJ601820S, RPJ602057S, RPJ602117S, RPJ602136S, RPJ600453S, RPJ600458S, RPJ600531S, RPJ601022S, RPJ600501S, RPJ601137S, RPJ600083S, RPJ600911S, RPJ600520S, RPJ600737S, RPJ600815S, RPJ600946S, RPJ601054S, RPJ601330S, RPJ601228S, RPJ602103S, RPJ600674S, RPJ601446S, RPJ600741S, RPJ601223S, RPJ600114S, RPJ601453S, RPJ601677S, RPJ602026S, RPJ601310S, RPJ601805S, RPJ600618S, RPJ601037S, RPJ600317S, RPJ600094S, RPJ601708S, RPJ600742S, RPJ601019S, RPJ601164S, RPJ601421S, RPJ600762S, RPJ601009S, RPJ601292S, RPJ600253S, RPJ601576S, RPJ601749S, RPJ601352S, RPJ601091S, RPJ601331S, RPJ602052S, RPJ600233S, RPJ600517S, RPJ600641S, RPJ601027S, RPJ601060S, RPJ601213S, RPJ601616S, RPJ601652S, RPJ601174S, RPJ601524S, RPJ600383S, RPJ601669S, RPJ600647S, RPJ601377S, RPJ601949S, RPJ600721S, RPJ600072S, RPJ600146S, RPJ600285S, RPJ601046S, RPJ600433S, RPJ601325S, RPJ600550S, RPJ600559S, RPJ601152S, RPJ601838S, RPJ600587S, RPJ601295S, RPJ601411S, RPJ600070S, RPJ600234S, RPJ600073S, RPJ601795S, RPJ601082S, RPJ600097S, RPJ601700S, RPJ601725S, RPJ600658S, RPJ601128S, RPJ601246S, RPJ601252S, RPJ601266S, RPJ601267S, RPJ601427S, RPJ601130S, RPJ602077S, RPJ600537S, RPJ600659S, RPJ601557S, RPJ601582S, RPJ601583S, RPJ601607S, RPJ601608S, RPJ601610S, RPJ601715S, RPJ601734S, RPJ601735S, RPJ601736S, RPJ601816S, RPJ601827S, RPJ601877S, RPJ601902S, RPJ602014S, RPJ602015S, RPJ602039S, RPJ602066S, RPJ602071S, RPJ600092S, RPJ600255S, RPJ600362S, RPJ600369S, RPJ600373S, RPJ600375S, RPJ600386S, RPJ600414S, RPJ600423S, RPJ600469S, RPJ600471S, RPJ600472S, RPJ600481S, RPJ600483S, RPJ600489S, RPJ600497S, RPJ600536S, RPJ600538S, RPJ600610S, RPJ600611S, RPJ600675S, RPJ600676S, RPJ600719S, RPJ600731S, RPJ600739S, RPJ600801S, RPJ600812S, RPJ600818S, RPJ600822S, RPJ600838S, RPJ600849S, RPJ600862S, RPJ600863S, RPJ600873S, RPJ600878S, RPJ600879S, RPJ600938S, RPJ601010S, RPJ601025S, RPJ601026S, RPJ601033S, RPJ601092S, RPJ601143S, RPJ601171S, RPJ601187S, RPJ601207S, RPJ601249S, RPJ601260S, RPJ601318S, RPJ601335S, RPJ601400S, RPJ601407S, RPJ601443S, RPJ601454S, RPJ601471S, RPJ601495S, RPJ602114S, RPJ602204S, RPJ601522S, RPJ601539S, RPJ601475S, RPJ601612S, RPJ600643S, RPJ600866S, RPJ600906S, RPJ600984S, RPJ601099S, RPJ601455S, RPJ601313S, RPJ601880S, RPJ600347S, RPJ600454S, RPJ600858S, RPJ601245S, RPJ601362S, RPJ602087S, RPJ602088S, RPJ600029S, RPJ600750S, RPJ600761S, RPJ600061S, RPJ601283S, RPJ600983S, RPJ600789S, RPJ601630S, RPJ600679S, RPJ601117S, RPJ601308S, RPJ601410S, RPJ601619S, RPJ601119S, RPJ600706S, RPJ600718S, RPJ601120S, RPJ601778S, RPJ601824S, RPJ600738S, RPJ600740S, RPJ600877S, RPJ600989S, RPJ601057S, RPJ601113S, RPJ601135S, RPJ601370S, RPJ601413S, RPJ601457S, RPJ601525S, RPJ600993S, RPJ601925S, RPJ601982S, RPJ601205S, RPJ600519S, RPJ600683S, RPJ601109S, RPJ601375S, RPJ601132S, RPJ602076S, RPJ601849S, RPJ601850S, RPJ601006S, RPJ602110S, RPJ601258S, RPJ601307S, RPJ601867S, RPJ601562S, RPJ602050S, RPJ602051S, RPJ601101S, RPJ601301S, RPJ601995S, RPJ601996S, RPJ600174S, RPJ601426S, RPJ601419S, RPJ600258S, RPJ600274S, RPJ600513S, RPJ600695S, RPJ600699S, RPJ600790S, RPJ600934S, RPJ601900S, RPJ600987S, RPJ601100S, RPJ601102S, RPJ601898S, RPJ601899S, RPJ601905S
Recalling Firm/
Manufacturer
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
8200 Coral Sea St Ne
Mounds View MN 55112-4391
For Additional Information ContactTechnical Services
800-929-4043
Manufacturer Reason
for Recall
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
FDA Determined
Cause 2
Under Investigation by firm
ActionMedtronic issued an URGENT MEDICAL DEVICE CORRECTION NOTICE to its consignees on 05/10/2023 by UPS 2-day delivery or email. Consignees are asked to forward this notification to those who need to be aware within the organization and to any location which the devices have been distributed. The notice explained the problem and the risk and requested the following patient management recommendations: - Prophylactic device replacement is NOT recommended. - Program all HV therapy pathways B>AX in all therapy zones to minimize the risk for this issue. - Prioritize reprogramming patients who have both a history of HV therapy and Rx1 programmed AX>B. - For remaining patients with AX>B programming in any HV therapy sequence, schedule (with appropriate discretion) the next follow-up for in-clinic reprogramming to minimize potential for reduced- or no-energy HV therapies to occur. - Contact Medtronic Technical Services (1-800-929-4043) or your local representative if one of the following is observed as these may be an indication of either a device or lead-related issue: o Reduced- or no-energy HV therapy is displayed in Episode Text (regardless of programmed pathway) o A persistent drop of approximately 50% in RA, RV and LV pacing lead impedance measurements as this may be an indication of increased potential for a future reduced- or no-energy therapy. Update 10/10/2023: Beginning 09/27/2023, Medtronic issued communication via letter to consignees regarding the availability of the revised IFU and SmartSync Software. Consignees are asked to forward this notification to those who need to be aware within the organization and to any location which the devices have been distributed. Consignees are asked to complete and return a confirmation certificate (or equivalent record) confirming they have received the notification.
Quantity in Commerce2767 units
DistributionUS Nationwide - Worldwide Distribution
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
PMA DatabasePMAs with Product Code = NIK
PMAs with Product Code = NIK
-
-