• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall medtronic

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall medtronicsee related information
Date Initiated by FirmMay 10, 2023
Date PostedJune 16, 2023
Recall Status1 Open3, Classified
Recall NumberZ-1777-2023
Recall Event ID 92328
PMA NumberP010031S674 
Product Classification Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D) - Product Code NIK
ProductCRTD COBALT XT HF MRI IS1 DF4, Model Number DTPA2D4; Implantable Cardioverter Defibrillators
Code Information GTIN 00763000178130, Lot Serial Numbers: RTG605297S, RTG605299S, RTG600588S, RTG601635S, RTG601691S, RTG601840S, RTG601857S, RTG602986S, RTG601966S, RTG601364S, RTG601621S, RTG600270S, RTG600554S, RTG601405S, RTG601699S, RTG601977S, RTG605270S, RTG600137S, RTG600138S, RTG600181S, RTG600271S, RTG600618S, RTG600620S, RTG600621S, RTG600622S, RTG600640S, RTG601395S, RTG601441S, RTG601681S, RTG601783S, RTG601896S, RTG601945S, RTG601948S, RTG602037S, RTG602040S, RTG602042S, RTG602047S, RTG602165S, RTG602166S, RTG602167S, RTG602340S, RTG605614S, RTG602039S, RTG602228S, RTG602306S, RTG602477S, RTG605996S, RTG605997S, RTG604903S, RTG600228S, RTG600332S, RTG601906S, RTG602441S, RTG605330S, RTG605332S, RTG605838S, RTG601643S, RTG606091S, RTG600317S, RTG600255S, RTG600467S, RTG604871S, RTG605042S, RTG605051S, RTG606010S, RTG601452S, RTG602204S, RTG604466S, RTG600411S, RTG600412S, RTG600465S, RTG600517S, RTG602559S, RTG602561S, RTG602563S, RTG602565S, RTG602566S, RTG602569S, RTG602571S, RTG602573S, RTG603385S, RTG603386S, RTG603400S, RTG603404S, RTG605218S, RTG601409S, RTG601410S, RTG601525S, RTG601863S, RTG601870S, RTG603636S, RTG603637S, RTG603684S, RTG605663S, RTG601684S, RTG600188S, RTG601614S, RTG602242S, RTG600594S, RTG600612S, RTG601853S, RTG602197S, RTG600663S, RTG601900S, RTG601937S, RTG603003S, RTG600478S, RTG600514S, RTG600550S, RTG600592S, RTG601343S, RTG601396S, RTG601400S, RTG601514S, RTG601688S, RTG601697S, RTG601873S, RTG601881S, RTG602215S, RTG602524S, RTG602980S, RTG602981S, RTG602434S, RTG601439S, RTG602190S, RTG600374S, RTG600653S, RTG602158S, RTG602240S, RTG602472S, RTG602474S, RTG602893S, RTG602897S, RTG602898S, RTG604872S, RTG605326S, RTG605327S, RTG605328S, RTG605329S, RTG605696S, RTG600563S, RTG601913S, RTG600264S, RTG600575S, RTG601600S, RTG601597S, RTG601617S, RTG601671S, RTG601687S, RTG601698S, RTG602533S, RTG602752S, RTG602760S, RTG602944S, RTG605214S, RTG605217S, RTG606083S, RTG606084S, RTG602882S, RTG602883S, RTG603008S, RTG603049S, RTG602966S, RTG602967S, RTG605999S, RTG602512S, RTG601540S, RTG600598S, RTG602181S, RTG602187S, RTG600401S, RTG601453S, RTG600527S, RTG601341S, RTG601342S, RTG601444S, RTG601337S, RTG600162S, RTG605259S, RTG605260S, RTG605261S, RTG605263S, RTG600466S, RTG600470S, RTG602425S, RTG605292S, RTG605293S, RTG600626S, RTG602335S, RTG603061S, RTG601672S, RTG600285S, RTG601420S, RTG605254S, RTG605255S, RTG605256S, RTG602193S, RTG602537S, RTG602539S, RTG602540S, RTG602542S, RTG602545S, RTG602546S, RTG602908S, RTG602910S, RTG602911S, RTG602912S, RTG602913S, RTG604506S, RTG605926S, RTG606032S, RTG606033S, RTG600221S, RTG600235S, RTG600248S, RTG600295S, RTG600471S, RTG600477S, RTG600571S, RTG600616S, RTG600624S, RTG600666S, RTG601388S, RTG601430S, RTG601498S, RTG601776S, RTG602195S, RTG602333S, RTG602970S, RTG606086S, RTG605285S, RTG605287S, RTG605288S, RTG605289S, RTG600454S, RTG600647S, RTG600649S, RTG600650S, RTG600651S, RTG601530S, RTG601560S, RTG601692S, RTG601696S, RTG601861S, RTG601862S, RTG601869S, RTG601886S, RTG602012S, RTG602199S, RTG602273S, RTG602302S, RTG602318S, RTG602519S, RTG602520S, RTG602521S, RTG602522S, RTG602523S, RTG602536S, RTG602549S, RTG602976S, RTG600130S, RTG601928S, RTG602458S, RTG600178S, RTG600604S, RTG602237S, RTG602231S, RTG600171S, RTG601389S, RTG602467S, RTG602886S, RTG602443S, RTG601628S, RTG602493S, RTG605275S, RTG600585S, RTG601660S, RTG605983S, RTG600464S, RTG601818S, RTG601816S, RTG601668S, RTG603034S, RTG600506S, RTG602313S, RTG605590S, RTG603065S, RTG603066S, RTG603067S, RTG603069S, RTG603071S, RTG603390S, RTG603394S, RTG603395S, RTG602358S, RTG601889S, RTG600549S, RTG602901S, RTG600254S, RTG600631S, RTG600632S, RTG601782S, RTG601794S, RTG602426S, RTG601583S, RTG601787S, RTG603360S, RTG601345S, RTG603973S, RTG604485S, RTG606123S, RTG601424S, RTG600150S, RTG601884S, RTG600635S, RTG601393S, RTG601844S, RTG602230S, RTG602361S, RTG605986S, RTG601533S, RTG600483S, RTG600489S, RTG600587S, RTG600610S, RTG600658S, RTG600660S, RTG602535S, RTG605893S, RTG605734S, RTG600521S, RTG601921S, RTG601975S, RTG602890S, RTG602958S, RTG602989S, RTG605088S, RTG605183S, RTG605184S, RTG605248S, RTG602157S, RTG600240S, RTG600263S, RTG601368S, RTG601506S, RTG601565S, RTG601580S, RTG601664S, RTG601882S, RTG600584S, RTG601894S, RTG600173S, RTG605845S, RTG603079S, RTG601821S, RTG601897S, RTG601979S, RTG602461S, RTG601879S, RTG601682S, RTG601666S, RTG601445S, RTG601887S, RTG601926S, RTG601942S, RTG602027S, RTG604879S, RTG601394S, RTG605354S, RTG601974S, RTG602173S, RTG602200S, RTG604905S, RTG600276S, RTG600277S, RTG600605S, RTG601379S, RTG601382S, RTG602134S, RTG602500S, RTG601413S, RTG601417S, RTG600251S, RTG600601S, RTG603039S, RTG602244S, RTG605721S, RTG605769S, RTG600186S, RTG600323S, RTG600350S, RTG600403S, RTG600479S, RTG601657S, RTG601874S, RTG602322S, RTG602454S, RTG602455S, RTG602957S, RTG601401S, RTG601421S, RTG601910S, RTG600648S, RTG606059S, RTG605684S, RTG605979S, RTG605841S, RTG601541S, RTG602057S, RTG601808S, RTG601981S, RTG602583S, RTG602585S, RTG602924S, RTG604592S, RTG605473S, RTG605474S, RTG600144S, RTG600315S, RTG600449S, RTG601642S, RTG602051S, RTG602144S, RTG605611S, RTG605613S, RTG605680S, RTG605840S, RTG605977S, RTG600141S, RTG601442S, RTG601488S, RTG602201S, RTG605157S, RTG601468S, RTG601967S, RTG601775S, RTG601865S, RTG602895S, RTG601391S, RTG600149S, RTG602118S, RTG602961S, RTG601398S, RTG602260S, RTG602464S, RTG602972S, RTG602205S, RTG602514S, RTG603030S, RTG600644S, RTG605066S, RTG601895S, RTG605294S, RTG605295S, RTG600519S, RTG600556S, RTG600638S, RTG601673S, RTG601885S, RTG601915S, RTG602207S, RTG602449S, RTG602450S, RTG602884S, RTG605588S, RTG601412S, RTG601415S, RTG602008S, RTG601982S, RTG602226S, RTG602925S, RTG602927S, RTG604452S, RTG605238S, RTG602965S, RTG603089S, RTG603091S, RTG604629S, RTG605224S, RTG600164S, RTG600319S, RTG601479S, RTG601685S, RTG602143S, RTG602146S, RTG602170S, RTG600272S, RTG601908S, RTG604665S, RTG600322S, RTG600593S, RTG601392S, RTG601809S, RTG601846S, RTG601999S, RTG602188S, RTG602192S, RTG602213S, RTG602315S, RTG602483S, RTG602484S, RTG602892S, RTG602971S, RTG600574S, RTG601773S, RTG601778S, RTG601830S, RTG601972S, RTG602128S, RTG602130S, RTG602183S, RTG602310S, RTG602974S, RTG603009S, RTG603059S, RTG605597S, RTG605717S, RTG602463S, RTG600293S, RTG601645S, RTG601649S, RTG601650S, RTG602937S, RTG602938S, RTG602939S, RTG602940S, RTG602941S, RTG604493S, RTG604495S, RTG601397S, RTG601890S, RTG602320S, RTG600237S, RTG601418S, RTG601899S, RTG602881S, RTG601447S, RTG602316S, RTG602465S, RTG603052S, RTG605862S, RTG600167S, RTG602171S, RTG604340S, RTG605072S, RTG600308S, RTG605821S, RTG606021S, RTG605479S, RTG605480S, RTG601859S, RTG602053S, RTG602054S, RTG605798S, RTG605881S, RTG600298S, RTG600501S, RTG601487S, RTG605591S, RTG600267S, RTG600423S, RTG601662S, RTG601670S, RTG601679S, RTG601767S, RTG601866S, RTG602433S, RTG602446S, RTG602456S, RTG602460S, RTG602496S, RTG602896S, RTG602975S, RTG602985S, RTG600202S, RTG600384S, RTG601349S, RTG601574S, RTG601612S, RTG602341S, RTG602342S, RTG602482S, RTG601346S, RTG601489S, RTG601828S, RTG602168S, RTG602236S, RTG601435S, RTG601858S, RTG602220S, RTG600579S, RTG600458S, RTG601661S, RTG601860S, RTG604567S, RTG605175S, RTG602647S, RTG602656S, RTG605274S, RTG601905S, RTG602507S, RTG600402S, RTG601365S, RTG601658S, RTG601659S, RTG601876S, RTG601877S, RTG602359S, RTG601529S, RTG602968S, RTG606045S, RTG600297S, RTG605075S, RTG605083S, RTG601615S, RTG605087S, RTG605912S, RTG600533S, RTG601386S, RTG601532S, RTG601689S, RTG601786S, RTG601891S, RTG602203S, RTG600139S, RTG600169S, RTG600261S, RTG600343S, RTG600664S, RTG600665S, RTG601376S, RTG601520S, RTG601613S, RTG602159S, RTG602487S, RTG600305S, RTG601629S, RTG601636S, RTG601647S, RTG601652S, RTG601683S, RTG601875S, RTG601917S, RTG602218S, RTG602435S, RTG602877S, RTG602885S, RTG605677S, RTG605833S, RTG605908S, RTG605909S, RTG606013S, RTG606014S, RTG606099S, RTG606100S, RTG601604S, RTG601790S, RTG600279S, RTG601446S, RTG605974S, RTG600225S, RTG600520S, RTG600555S, RTG601375S, RTG602185S, RTG602491S, RTG602492S, RTG602931S, RTG603078S, RTG603368S, RTG603019S, RTG605715S, RTG600530S, RTG600569S, RTG601599S, RTG601761S, RTG601898S, RTG602224S, RTG601414S, RTG601432S, RTG601695S, RTG602164S, RTG602448S, RTG601762S, RTG603883S, RTG602117S, RTG601619S, RTG601624S, RTG601626S, RTG601630S, RTG601632S, RTG602946S, RTG602947S, RTG602948S, RTG602949S, RTG602950S, RTG602951S, RTG602952S, RTG602953S, RTG603010S, RTG603011S, RTG603013S, RTG605305S, RTG605306S, RTG600283S, RTG601449S, RTG601515S, RTG601625S, RTG601771S, RTG601791S, RTG601827S, RTG601845S, RTG601916S, RTG601919S, RTG601920S, RTG601931S, RTG602153S, RTG602235S, RTG602453S, RTG602459S, RTG602495S, RTG602497S, RTG602887S, RTG602888S, RTG602506S, RTG602182S, RTG602466S, RTG605552S, RTG602015S, RTG600394S, RTG600568S, RTG601374S, RTG601638S, RTG601851S, RTG602544S, RTG602935S, RTG600159S, RTG600194S, RTG600324S, RTG600336S, RTG600596S, RTG601511S, RTG601994S, RTG602160S, RTG602440S, RTG602879S, RTG602956S, RTG602960S, RTG601914S, RTG602339S, RTG602502S, RTG602987S, RTG604907S, RTG606089S, RTG600581S, RTG601460S, RTG602473S, RTG602926S, RTG605775S, RTG601422S, RTG602490S, RTG600273S, RTG601383S, RTG600591S, RTG600307S, RTG600447S, RTG601474S, RTG602208S, RTG600208S, RTG606092S, RTG601832S, RTG605704S, RTG602018S, RTG605598S, RTG605759S, RTG602069S, RTG604873S, RTG600196S, RTG601339S, RTG601644S, RTG605921S, RTG600143S, RTG600259S, RTG600597S, RTG601605S, RTG601843S, RTG601856S, RTG602136S, RTG602152S, RTG602488S, RTG602988S, RTG605152S, RTG601852S, RTG602139S, RTG602338S, RTG602172S, RTG601872S, RTG602365S, RTG603037S, RTG600476S, RTG600296S, RTG600492S, RTG601327S, RTG601423S, RTG601675S, RTG601680S, RTG601849S, RTG601871S, RTG601958S, RTG602025S, RTG602969S, RTG601419S, RTG601839S, RTG601901S, RTG603036S, RTG602028S, RTG602210S, RTG602907S, RTG605249S, RTG600148S, RTG600155S, RTG600292S, RTG600351S, RTG600399S, RTG600526S, RTG601369S, RTG601387S, RTG601601S, RTG602494S, RTG602955S, RTG604959S, RTG605737S, RTG602163S, RTG604762S, RTG605762S, RTG606118S, RTG601795S, RTG601802S, RTG601815S, RTG601820S, RTG602017S, RTG602019S, RTG602078S, RTG602079S, RTG602221S, RTG602291S, RTG602297S, RTG602299S, RTG602516S, RTG602550S, RTG602567S, RTG602649S, RTG602657S, RTG602658S, RTG603162S, RTG603165S, RTG603166S, RTG603173S, RTG603177S, RTG603334S, RTG603335S, RTG603336S, RTG603337S, RTG603339S, RTG603341S, RTG603342S, RTG603343S, RTG603344S, RTG603345S, RTG603346S, RTG603347S, RTG603348S, RTG603349S, RTG603352S, RTG603357S, RTG603364S, RTG603375S, RTG603377S, RTG603378S, RTG603379S, RTG603380S, RTG603381S, RTG603382S, RTG603383S, RTG603384S, RTG603392S, RTG603393S, RTG603396S, RTG603397S, RTG603398S, RTG603401S, RTG603402S, RTG603403S, RTG603405S, RTG603406S, RTG603409S, RTG603410S, RTG603413S, RTG603414S, RTG603415S, RTG603416S, RTG603417S, RTG603418S, RTG603421S, RTG603423S, RTG603424S, RTG603425S, RTG603426S, RTG603427S, RTG603428S, RTG603429S, RTG603430S, RTG603431S, RTG603577S, RTG603586S, RTG603587S, RTG603588S, RTG603590S, RTG603591S, RTG603592S, RTG603597S, RTG603598S, RTG603599S, RTG603600S, RTG603601S, RTG603603S, RTG603604S, RTG603606S, RTG603607S, RTG603608S, RTG603609S, RTG603610S, RTG603611S, RTG603613S, RTG603614S, RTG603615S, RTG603616S, RTG603617S, RTG603618S, RTG603620S, RTG603623S, RTG603671S, RTG603695S, RTG603696S, RTG603697S, RTG603698S, RTG603699S, RTG603700S, RTG603703S, RTG603706S, RTG603724S, RTG603727S, RTG603771S, RTG603772S, RTG603773S, RTG603774S, RTG603775S, RTG603776S, RTG603777S, RTG603832S, RTG603833S, RTG603834S, RTG603841S, RTG603842S, RTG603843S, RTG603844S, RTG603847S, RTG603850S, RTG603851S, RTG603852S, RTG603853S, RTG603854S, RTG603857S, RTG603858S, RTG603859S, RTG603860S, RTG603861S, RTG603862S, RTG603863S, RTG603864S, RTG603865S, RTG603866S, RTG603867S, RTG603870S, RTG603871S, RTG603874S, RTG603878S, RTG603885S, RTG603893S, RTG603894S, RTG603897S, RTG603900S, RTG603901S, RTG603903S, RTG603914S, RTG603929S, RTG603930S, RTG603931S, RTG603940S, RTG603941S, RTG603942S, RTG603944S, RTG603972S, RTG603974S, RTG603975S, RTG603978S, RTG603979S, RTG603980S, RTG603981S, RTG603982S, RTG603984S, RTG603986S, RTG604035S, RTG604038S, RTG604240S, RTG604308S, RTG604309S, RTG604310S, RTG604311S, RTG604421S, RTG604448S, RTG604450S, RTG604453S, RTG604454S, RTG604455S, RTG604461S, RTG604467S, RTG604469S, RTG604472S, RTG604475S, RTG604482S, RTG604484S, RTG604487S, RTG604489S, RTG604491S, RTG604492S, RTG604496S, RTG604497S, RTG604501S, RTG604502S, RTG604505S, RTG604507S, RTG604518S, RTG604519S, RTG604520S, RTG604521S, RTG604522S, RTG604523S, RTG604524S, RTG604525S, RTG604527S, RTG604528S, RTG604529S, RTG604541S, RTG604551S, RTG604558S, RTG604559S, RTG604563S, RTG604565S, RTG604566S, RTG604575S, RTG604579S, RTG604585S, RTG604589S, RTG604590S, RTG604595S, RTG604596S, RTG604600S, RTG604602S, RTG604609S, RTG604628S, RTG604633S, RTG604634S, RTG604635S, RTG604637S, RTG604638S, RTG604644S, RTG604645S, RTG604646S, RTG604647S, RTG604648S, RTG604651S, RTG604652S, RTG604654S, RTG604655S, RTG604659S, RTG604661S, RTG604662S, RTG604675S, RTG604686S, RTG604687S, RTG604689S, RTG604708S, RTG604709S, RTG604722S, RTG604728S, RTG604729S, RTG604730S, RTG604731S, RTG604732S, RTG604733S, RTG604734S, RTG604735S, RTG604736S, RTG604738S, RTG604739S, RTG604740S, RTG604741S, RTG604747S, RTG604793S, RTG604795S, RTG604797S, RTG604847S, RTG604850S, RTG604851S, RTG604866S, RTG604906S, RTG604911S, RTG604912S, RTG604920S, RTG605040S, RTG605043S, RTG605049S, RTG605064S, RTG605131S, RTG605132S, RTG605170S, RTG605172S, RTG605173S, RTG605174S, RTG605182S, RTG605185S, RTG605187S, RTG605189S, RTG605200S, RTG605203S, RTG605215S, RTG605241S, RTG605242S, RTG605243S, RTG605244S, RTG605245S, RTG605247S, RTG605253S, RTG605266S, RTG605267S, RTG605268S, RTG605269S, RTG605276S, RTG605278S, RTG605279S, RTG605283S, RTG605284S, RTG605286S, RTG605290S, RTG605291S, RTG605296S, RTG605298S, RTG605301S, RTG605302S, RTG605303S, RTG605304S, RTG605307S, RTG605308S, RTG605309S, RTG605310S, RTG605336S, RTG605337S, RTG605342S, RTG605363S, RTG605364S, RTG605365S, RTG605475S, RTG605478S, RTG605527S, RTG605622S, RTG605631S, RTG605755S, RTG605924S, RTG605941S, RTG605942S, RTG602843S, RTG603127S, RTG603277S, RTG603331S, RTG603332S, RTG603333S, RTG603338S, RTG603350S, RTG603351S, RTG603353S, RTG603354S, RTG603356S, RTG603358S, RTG603359S, RTG603361S, RTG603363S, RTG603365S, RTG603366S, RTG603367S, RTG603369S, RTG603370S, RTG603371S, RTG603372S, RTG603374S, RTG603376S, RTG603387S, RTG603388S, RTG603389S, RTG603391S, RTG603399S, RTG603419S, RTG603420S, RTG603422S, RTG603581S, RTG603585S, RTG603595S, RTG603596S, RTG603602S, RTG603605S, RTG603612S, RTG603621S, RTG603622S, RTG603624S, RTG603626S, RTG603627S, RTG603628S, RTG603629S, RTG603630S, RTG603631S, RTG603632S, RTG603633S, RTG603634S, RTG603638S, RTG603639S, RTG603640S, RTG603641S, RTG603642S, RTG603643S, RTG603644S, RTG603645S, RTG603646S, RTG603647S, RTG603648S, RTG603649S, RTG603650S, RTG603651S, RTG603652S, RTG603653S, RTG603654S, RTG603655S, RTG603656S, RTG603657S, RTG603659S, RTG603660S, RTG603662S, RTG603663S, RTG603664S, RTG603665S, RTG603666S, RTG603667S, RTG603669S, RTG603670S, RTG603672S, RTG603673S, RTG603674S, RTG603675S, RTG603676S, RTG603677S, RTG603678S, RTG603679S, RTG603680S, RTG603683S, RTG603685S, RTG603686S, RTG603688S, RTG603689S, RTG603690S, RTG603691S, RTG603692S, RTG603707S, RTG603708S, RTG603709S, RTG603710S, RTG603712S, RTG603713S, RTG603715S, RTG603716S, RTG603717S, RTG603718S, RTG603719S, RTG603720S, RTG603721S, RTG603722S, RTG603723S, RTG603729S, RTG603730S, RTG603732S, RTG603734S, RTG603735S, RTG603737S, RTG603738S, RTG603740S, RTG603741S, RTG603742S, RTG603745S, RTG603751S, RTG603754S, RTG603755S, RTG603758S, RTG603759S, RTG603760S, RTG603762S, RTG603764S, RTG603765S, RTG603766S, RTG603767S, RTG603768S, RTG603769S, RTG603770S, RTG603778S, RTG603779S, RTG603780S, RTG603781S, RTG603782S, RTG603783S, RTG603784S, RTG603786S, RTG603787S, RTG603788S, RTG603789S, RTG603790S, RTG603791S, RTG603793S, RTG603794S, RTG603795S, RTG603796S, RTG603797S, RTG603798S, RTG603799S, RTG603800S, RTG603801S, RTG603802S, RTG603803S, RTG603804S, RTG603805S, RTG603807S, RTG603808S, RTG603809S, RTG603810S, RTG603811S, RTG603812S, RTG603813S, RTG603814S, RTG603816S, RTG603817S, RTG603818S, RTG603819S, RTG603820S, RTG603821S, RTG603822S, RTG603823S, RTG603824S, RTG603825S, RTG603827S, RTG603828S, RTG603829S, RTG603830S, RTG603831S, RTG603835S, RTG603836S, RTG603837S, RTG603840S, RTG603845S, RTG603846S, RTG603886S, RTG603887S, RTG603888S, RTG603889S, RTG603891S, RTG603892S, RTG603907S, RTG603909S, RTG603911S, RTG603913S, RTG603915S, RTG603919S, RTG603920S, RTG603921S, RTG603922S, RTG603923S, RTG603924S, RTG603925S, RTG603926S, RTG603927S, RTG603928S, RTG603933S, RTG603934S, RTG603936S, RTG603937S, RTG603938S, RTG603945S, RTG603946S, RTG603947S, RTG603948S, RTG603950S, RTG603952S, RTG603955S, RTG603957S, RTG603959S, RTG603960S, RTG603961S, RTG603963S, RTG603964S, RTG603965S, RTG603966S, RTG603967S, RTG603968S, RTG603969S, RTG603970S, RTG603971S, RTG603985S, RTG603987S, RTG603988S, RTG603989S, RTG603990S, RTG603991S, RTG603992S, RTG603994S, RTG603995S, RTG603996S, RTG603997S, RTG603998S, RTG603999S, RTG604000S, RTG604001S, RTG604002S, RTG604004S, RTG604005S, RTG604006S, RTG604007S, RTG604009S, RTG604011S, RTG604012S, RTG604015S, RTG604016S, RTG604017S, RTG604018S, RTG604019S, RTG604020S, RTG604026S, RTG604027S, RTG604028S, RTG604030S, RTG604031S, RTG604032S, RTG604033S, RTG604036S, RTG604037S, RTG604039S, RTG604040S, RTG604042S, RTG604043S, RTG604044S, RTG604051S, RTG604052S, RTG604177S, RTG604178S, RTG604184S, RTG604186S, RTG604190S, RTG604192S, RTG604220S, RTG604221S, RTG604225S, RTG604227S, RTG604228S, RTG604229S, RTG604230S, RTG604234S, RTG604236S, RTG604237S, RTG604241S, RTG604243S, RTG604244S, RTG604246S, RTG604247S, RTG604248S, RTG604251S, RTG604252S, RTG604253S, RTG604254S, RTG604255S, RTG604256S, RTG604257S, RTG604259S, RTG604260S, RTG604262S, RTG604264S, RTG604267S, RTG604268S, RTG604271S, RTG604272S, RTG604277S, RTG604279S, RTG604280S, RTG604281S, RTG604282S, RTG604284S, RTG604287S, RTG604288S, RTG604290S, RTG604291S, RTG604292S, RTG604293S, RTG604294S, RTG604295S, RTG604296S, RTG604297S, RTG604299S, RTG604300S, RTG604302S, RTG604303S, RTG604304S, RTG604305S, RTG604306S, RTG604307S, RTG604312S, RTG604313S, RTG604314S, RTG604315S, RTG604319S, RTG604335S, RTG604336S, RTG604337S, RTG604338S, RTG604339S, RTG604341S, RTG604342S, RTG604343S, RTG604344S, RTG604346S, RTG604347S, RTG604440S, RTG604441S, RTG604443S, RTG604444S, RTG604445S, RTG604447S, RTG604451S, RTG604456S, RTG604460S, RTG604463S, RTG604465S, RTG604468S, RTG604481S, RTG604560S, RTG604561S, RTG604587S, RTG604591S, RTG604630S, RTG604631S, RTG604632S, RTG604656S, RTG604657S, RTG604663S, RTG604664S, RTG604666S, RTG604667S, RTG604668S, RTG604669S, RTG604670S, RTG604671S, RTG604672S, RTG604673S, RTG604674S, RTG604676S, RTG604678S, RTG604679S, RTG604680S, RTG604681S, RTG604682S, RTG604684S, RTG604685S, RTG604690S, RTG604691S, RTG604692S, RTG604693S, RTG604694S, RTG604695S, RTG604696S, RTG604697S, RTG604698S, RTG604699S, RTG604700S, RTG604701S, RTG604704S, RTG604706S, RTG604707S, RTG604737S, RTG604743S, RTG604744S, RTG604746S, RTG604749S, RTG604750S, RTG604751S, RTG604752S, RTG604755S, RTG604756S, RTG604757S, RTG604759S, RTG604761S, RTG604763S, RTG604791S, RTG604794S, RTG604799S, RTG604802S, RTG604803S, RTG604806S, RTG604807S, RTG604808S, RTG604810S, RTG604811S, RTG604813S, RTG604817S, RTG604830S, RTG604832S, RTG604833S, RTG604835S, RTG604836S, RTG604841S, RTG604845S, RTG604846S, RTG604853S, RTG604854S, RTG604855S, RTG604856S, RTG604857S, RTG604858S, RTG604859S, RTG604860S, RTG604861S, RTG604862S, RTG604863S, RTG604864S, RTG604865S, RTG604869S, RTG604870S, RTG604874S, RTG604875S, RTG604876S, RTG604878S, RTG604880S, RTG604883S, RTG604884S, RTG604885S, RTG604886S, RTG604887S, RTG604888S, RTG604889S, RTG604890S, RTG604891S, RTG604892S, RTG604893S, RTG604894S, RTG604895S, RTG604896S, RTG604897S, RTG604898S, RTG604899S, RTG604900S, RTG604901S, RTG604902S, RTG604904S, RTG604908S, RTG604910S, RTG604913S, RTG604914S, RTG604916S, RTG604917S, RTG604921S, RTG604922S, RTG604923S, RTG604924S, RTG604925S, RTG604927S, RTG604929S, RTG604930S, RTG604932S, RTG604934S, RTG604935S, RTG604936S, RTG604937S, RTG604938S, RTG604939S, RTG604942S, RTG604943S, RTG604944S, RTG604945S, RTG604946S, RTG604947S, RTG604949S, RTG604952S, RTG604954S, RTG604955S, RTG604956S, RTG604957S, RTG604958S, RTG604960S, RTG604961S, RTG604962S, RTG604963S, RTG604964S, RTG604965S, RTG604966S, RTG604967S, RTG604968S, RTG604971S, RTG604972S, RTG604973S, RTG604981S, RTG604982S, RTG604983S, RTG604984S, RTG604985S, RTG604986S, RTG604987S, RTG604988S, RTG604989S, RTG604990S, RTG604991S, RTG604992S, RTG605001S, RTG605002S, RTG605012S, RTG605020S, RTG605038S, RTG605052S, RTG605053S, RTG605054S, RTG605055S, RTG605065S, RTG605081S, RTG605089S, RTG605090S, RTG605092S, RTG605098S, RTG605104S, RTG605105S, RTG605106S, RTG605110S, RTG605114S, RTG605123S, RTG605126S, RTG605138S, RTG605139S, RTG605141S, RTG605143S, RTG605144S, RTG605145S, RTG605146S, RTG605147S, RTG605149S, RTG605151S, RTG605153S, RTG605154S, RTG605158S, RTG605161S, RTG605163S, RTG605165S, RTG605166S, RTG605169S, RTG605179S, RTG605181S, RTG605205S, RTG605210S, RTG605222S, RTG605239S, RTG605246S, RTG605251S, RTG605252S, RTG605257S, RTG605262S, RTG605264S, RTG605265S, RTG605282S, RTG605300S, RTG605334S, RTG605335S, RTG605349S, RTG605351S, RTG605352S, RTG605355S, RTG605356S, RTG605357S, RTG605358S, RTG605359S, RTG605360S, RTG605362S, RTG605394S, RTG605518S, RTG605519S, RTG605520S, RTG605522S, RTG605525S, RTG605526S, RTG605532S, RTG605536S, RTG605539S, RTG605540S, RTG605541S, RTG605542S, RTG605547S, RTG605548S, RTG605553S, RTG605554S, RTG605555S, RTG605556S, RTG605558S, RTG605559S, RTG605563S, RTG605566S, RTG605574S, RTG605576S, RTG605578S, RTG605579S, RTG605580S, RTG605582S, RTG605583S, RTG605584S, RTG605585S, RTG605586S, RTG605592S, RTG605595S, RTG605601S, RTG605604S, RTG605615S, RTG605616S, RTG605651S, RTG605652S, RTG605658S, RTG605666S, RTG605670S, RTG605671S, RTG605673S, RTG605674S, RTG605682S, RTG605685S, RTG605688S, RTG605689S, RTG605690S, RTG605693S, RTG605694S, RTG605695S, RTG605698S, RTG605699S, RTG605707S, RTG605710S, RTG605714S, RTG605716S, RTG605719S, RTG605720S, RTG605722S, RTG605723S, RTG605727S, RTG605731S, RTG605732S, RTG605738S, RTG605739S, RTG605740S, RTG605757S, RTG605758S, RTG605768S, RTG605772S, RTG605774S, RTG605786S, RTG605787S, RTG605794S, RTG605809S, RTG605823S, RTG605824S, RTG605825S, RTG605829S, RTG605832S, RTG605834S, RTG605835S, RTG605843S, RTG605844S, RTG605847S, RTG605851S, RTG605859S, RTG605860S, RTG605861S, RTG605863S, RTG605866S, RTG605867S, RTG605870S, RTG605880S, RTG605882S, RTG605896S, RTG605905S, RTG605916S, RTG605918S, RTG605920S, RTG605947S, RTG605949S, RTG605966S, RTG605975S, RTG606019S, RTG606020S, RTG606048S, RTG600451S, RTG601484S, RTG600219S, RTG600450S, RTG600482S, RTG601331S, RTG601848S, RTG601427S, RTG603728S, RTG600201S, RTG601448S, RTG605150S, RTG601980S, RTG600662S, RTG602930S, RTG601332S, RTG601366S, RTG600655S, RTG600161S, RTG600282S, RTG602364S, RTG601983S, RTG601993S, RTG605919S, RTG600224S, RTG600567S, RTG605476S, RTG605477S, RTG605729S, RTG600437S, RTG600548S, RTG600577S, RTG600656S, RTG601904S, RTG602934S, RTG605575S, RTG605577S, RTG600529S, RTG601504S, RTG602162S, RTG600652S, RTG601521S, RTG604881S, RTG602421S, RTG602471S, RTG601334S, RTG601907S, RTG601943S, RTG602196S, RTG602303S, RTG602451S, RTG600570S, RTG601030S, RTG603004S, RTG600275S, RTG600586S, RTG601764S, RTG602586S, RTG602587S, RTG602590S, RTG602594S, RTG602595S, RTG604498S, RTG604499S, RTG604500S, RTG604554S, RTG604555S, RTG601956S, RTG601978S, RTG604583S, RTG600156S, RTG600543S, RTG600288S, RTG600558S, RTG601377S, RTG600615S, RTG601957S, RTG606017S, RTG600129S, RTG600291S, RTG600320S, RTG600513S, RTG601433S, RTG602006S, RTG602217S, RTG602219S, RTG602329S, RTG602331S, RTG602447S, RTG606114S, RTG601325S, RTG605865S, RTG600456S, RTG600544S, RTG600600S, RTG601929S, RTG602016S, RTG602206S, RTG602470S, RTG602504S, RTG605954S, RTG601878S, RTG602503S, RTG605776S, RTG605777S, RTG606050S, RTG601965S, RTG604915S, RTG600359S, RTG601406S, RTG601524S, RTG601674S, RTG601676S, RTG601841S, RTG602127S, RTG602239S, RTG602510S, RTG606102S, RTG600583S, RTG602035S, RTG600151S, RTG600633S, RTG601567S, RTG601855S, RTG601964S, RTG602357S, RTG602486S, RTG605133S, RTG605199S, RTG605202S, RTG605207S, RTG606044S, RTG602584S, RTG601367S, RTG601995S, RTG600145S, RTG600172S, RTG600203S, RTG601372S, RTG601572S, RTG601893S, RTG600580S, RTG600646S, RTG601634S, RTG601902S, RTG601961S, RTG602312S, RTG602498S, RTG605771S, RTG602191S, RTG600536S, RTG601428S, RTG601932S, RTG601933S, RTG601997S, RTG602899S, RTG605708S, RTG606034S, RTG601909S, RTG602137S, RTG605743S, RTG605778S, RTG602202S, RTG601968S, RTG603074S, RTG600135S, RTG600318S, RTG600582S, RTG601407S, RTG601465S, RTG601911S, RTG601912S, RTG602241S, RTG602314S, RTG604953S, RTG605103S, RTG605907S, RTG605963S, RTG606051S, RTG601455S, RTG601534S, RTG601883S, RTG604683S, RTG601998S, RTG601955S, RTG605745S, RTG601984S, RTG600245S, RTG602186S, RTG600661S, RTG601585S, RTG601462S, RTG601826S, RTG601842S, RTG600459S, RTG600460S, RTG601333S, RTG601390S, RTG602345S, RTG602999S, RTG605587S, RTG605115S, RTG606039S, RTG604603S, RTG604606S, RTG601837S, RTG602223S, RTG602360S, RTG602915S, RTG605128S, RTG605129S, RTG600226S, RTG603838S, RTG603839S, RTG600639S, RTG603072S, RTG601456S, RTG601457S, RTG601459S, RTG601640S, RTG601641S, RTG601938S, RTG601959S, RTG601996S, RTG602437S, RTG602438S, RTG602439S, RTG602475S, RTG602476S, RTG602551S, RTG602552S, RTG602889S, RTG602906S, RTG603023S, RTG603025S, RTG604569S, RTG604570S, RTG605529S, RTG600515S, RTG601854S, RTG606111S, RTG602229S, RTG602489S, RTG605379S, RTG605380S, RTG605381S, RTG605382S, RTG605383S, RTG605384S, RTG605386S, RTG605387S, RTG605388S, RTG605389S, RTG605391S, RTG606072S, RTG606073S, RTG606074S, RTG606075S, RTG606076S, RTG606077S, RTG606078S, RTG606079S, RTG606080S, RTG606081S, RTG602505S, RTG600361S, RTG606120S, RTG600328S, RTG600480S, RTG600627S, RTG601706S, RTG600133S, RTG600147S, RTG600168S, RTG600174S, RTG600187S, RTG601693S, RTG601817S, RTG602161S, RTG602990S, RTG605927S, RTG605928S, RTG605932S, RTG605978S, RTG605933S, RTG600312S, RTG600325S, RTG602319S, RTG600524S, RTG601868S, RTG605982S, RTG602216S, RTG601850S, RTG600222S, RTG600140S, RTG602481S, RTG602921S, RTG602922S, RTG602923S, RTG605338S, RTG605339S, RTG605340S, RTG605341S, RTG605343S, RTG605347S, RTG605348S, RTG601399S, RTG601438S, RTG601536S, RTG601539S, RTG601655S, RTG602060S, RTG602468S, RTG602499S, RTG602513S, RTG602894S, RTG603040S, RTG603043S, RTG600340S, RTG601330S, RTG602214S, RTG605366S, RTG605367S, RTG605368S, RTG605369S, RTG605370S, RTG605371S, RTG605372S, RTG605373S, RTG605374S, RTG605481S, RTG605482S, RTG605484S, RTG605485S, RTG605487S, RTG605488S, RTG605489S, RTG605490S, RTG605491S, RTG605493S, RTG605494S, RTG601690S, RTG602070S, RTG606071S, RTG604844S, RTG605500S, RTG605501S, RTG605503S, RTG605506S, RTG605507S, RTG605508S, RTG605509S, RTG605510S, RTG605511S, RTG605513S, RTG605516S, RTG605517S, RTG605395S, RTG605397S, RTG605398S, RTG605399S, RTG605400S, RTG605401S, RTG605402S, RTG605404S, RTG605405S, RTG605406S, RTG605407S, RTG605408S, RTG605409S, RTG605410S, RTG605411S, RTG604753S, RTG604754S, RTG603872S, RTG604624S, RTG604625S, RTG604626S, RTG604627S, RTG605444S, RTG605445S, RTG605447S, RTG605449S, RTG605450S, RTG605451S, RTG605455S, RTG605458S, RTG605459S, RTG605461S, RTG605463S, RTG605464S, RTG605465S, RTG605467S, RTG605468S, RTG605469S, RTG605470S, RTG605471S, RTG605472S, RTG605134S, RTG605135S, RTG605136S, RTG605137S, RTG605392S, RTG605393S, RTG605773S, RTG606067S, RTG606070S, RTG605361S, RTG602444S, RTG602530S, RTG602531S, RTG602532S, RTG602880S, RTG602902S, RTG602905S, RTG602991S, RTG602992S, RTG602993S, RTG602998S, RTG605617S, RTG605618S, RTG605619S, RTG605620S, RTG605621S, RTG605623S, RTG605624S, RTG605625S, RTG605628S, RTG605630S, RTG605633S, RTG605636S, RTG605637S, RTG605756S, RTG605277S, RTG600557S, RTG601564S, RTG601663S, RTG601768S, RTG601824S, RTG601847S, RTG602014S, RTG602126S, RTG602140S, RTG602198S, RTG602211S, RTG602332S, RTG602469S, RTG603408S, RTG603411S, RTG606082S, RTG601969S, RTG602508S, RTG602979S, RTG604653S, RTG600183S, RTG601526S, RTG602317S, RTG600250S, RTG601970S, RTG601971S, RTG601991S, RTG602876S, RTG603029S, RTG605836S, RTG604261S, RTG606065S, RTG606066S, RTG600578S, RTG601594S, RTG600385S, RTG601510S, RTG601519S, RTG605852S, RTG605883S, RTG601880S, RTG600160S, RTG601378S, RTG602232S, RTG602891S, RTG605712S, RTG605223S, RTG605711S, RTG600358S, RTG601340S, RTG602034S, RTG602983S, RTG605530S, RTG606109S, RTG601537S, RTG601669S, RTG600243S, RTG600258S, RTG601471S, RTG601595S, RTG601986S, RTG605108S, RTG601568S, RTG605946S, RTG606011S, RTG601798S, RTG600321S, RTG600668S, RTG602189S, RTG602933S, RTG601570S, RTG602330S, RTG602929S, RTG605805S, RTG601922S, RTG601805S, RTG602334S, RTG600218S, RTG600452S, RTG601522S, RTG601784S, RTG602123S, RTG602184S, RTG602132S, RTG602222S, RTG606040S, RTG600500S, RTG600617S, RTG601508S, RTG601622S, RTG601963S, RTG602020S, RTG602031S, RTG602354S, RTG602485S, RTG602994S, RTG601326S, RTG601927S, RTG602943S, RTG600473S, RTG600606S, RTG600607S, RTG601513S, RTG602131S, RTG602149S, RTG602305S, RTG600025S, RTG600564S, RTG605140S, RTG605957S, RTG605641S, RTG600484S, RTG600487S, RTG600539S, RTG601656S, RTG602011S, RTG602049S, RTG600195S, RTG600642S, RTG601335S, RTG602212S, RTG604948S, RTG605033S, RTG605483S, RTG605534S, RTG600234S, RTG600553S, RTG602909S, RTG602982S, RTG601892S, RTG605495S, RTG605497S, RTG605498S, RTG605499S, RTG600136S, RTG602209S, RTG602238S, RTG602478S, RTG602479S, RTG602480S, RTG601653S, RTG600176S, RTG600177S, RTG602878S, RTG603064S, RTG600158S, RTG601654S, RTG601834S, RTG601989S, RTG602346S, RTG605412S, RTG605413S, RTG605414S, RTG605415S, RTG605416S, RTG605417S, RTG605418S, RTG605419S, RTG605420S, RTG605421S, RTG605422S, RTG605423S, RTG605424S, RTG605425S, RTG605427S, RTG605428S, RTG605429S, RTG605430S, RTG605432S, RTG605281S, RTG605375S, RTG605376S, RTG605377S, RTG605378S, RTG603018S, RTG606085S, RTG604688S, RTG604867S, RTG604919S, RTG604951S, RTG604974S, RTG604975S, RTG604976S, RTG604977S, RTG605233S, RTG605271S, RTG605272S, RTG602423S, RTG603092S, RTG603407S, RTG603593S, RTG603594S, RTG606052S, RTG601491S, RTG601495S, RTG601502S, RTG601503S, RTG601573S, RTG601575S, RTG601581S, RTG601586S, RTG601602S, RTG601603S, RTG601606S, RTG601607S, RTG601610S, RTG601616S, RTG601620S, RTG601623S, RTG601627S, RTG601631S, RTG601637S, RTG601639S, RTG601646S, RTG601648S, RTG601665S, RTG601797S, RTG601806S, RTG601807S, RTG601812S, RTG601813S, RTG601976S, RTG602138S, RTG602174S, RTG602175S, RTG602176S, RTG602177S, RTG602178S, RTG602180S, RTG602321S, RTG602324S, RTG602325S, RTG602326S, RTG602427S, RTG602428S, RTG602432S, RTG602517S, RTG602525S, RTG602526S, RTG602527S, RTG602528S, RTG602529S, RTG602553S, RTG602554S, RTG602555S, RTG602556S, RTG602577S, RTG602580S, RTG602581S, RTG602582S, RTG602774S, RTG602776S, RTG602782S, RTG602788S, RTG602795S, RTG602796S, RTG602801S, RTG602804S, RTG602816S, RTG602822S, RTG602835S, RTG602837S, RTG602841S, RTG602849S, RTG602855S, RTG602867S, RTG602869S, RTG602870S, RTG602871S, RTG602872S, RTG602873S, RTG602874S, RTG602875S, RTG602900S, RTG602903S, RTG602904S, RTG602914S, RTG602916S, RTG602917S, RTG602918S, RTG602919S, RTG602920S, RTG602936S, RTG602962S, RTG602964S, RTG603868S, RTG603869S, RTG604474S, RTG604483S, RTG604639S, RTG604640S, RTG604641S, RTG604642S, RTG604643S, RTG604649S, RTG604650S, RTG604792S, RTG604848S, RTG604849S, RTG604852S, RTG604993S, RTG604996S, RTG604997S, RTG604999S, RTG605190S, RTG605192S, RTG605193S, RTG605196S, RTG605197S, RTG605226S, RTG605230S, RTG605231S, RTG605235S, RTG605236S, RTG605313S, RTG605314S, RTG605315S, RTG605316S, RTG605317S, RTG605318S, RTG605320S, RTG605321S, RTG605322S, RTG605323S, RTG605324S, RTG605325S, RTG600461S, RTG603935S, RTG605058S, RTG600313S, RTG601677S, RTG605989S, RTG602348S, RTG602351S, RTG602430S, RTG602431S, RTG602511S, RTG602534S, RTG602557S, RTG603157S, RTG603589S, RTG603848S, RTG603849S, RTG603855S, RTG603856S, RTG603007S, RTG602350S, RTG605433S, RTG605434S, RTG605436S, RTG605438S, RTG605440S, RTG605441S, RTG605442S, RTG605443S, RTG605686S, RTG605687S, RTG602234S, RTG605036S, RTG605985S, RTG600180S, RTG600182S, RTG600191S, RTG600192S, RTG600193S, RTG600205S, RTG600206S, RTG600220S, RTG600242S, RTG600284S, RTG601949S, RTG601951S, RTG601953S, RTG602142S, RTG602145S, RTG602147S, RTG602148S, RTG602150S, RTG602151S, RTG602154S, RTG602155S, RTG602156S, RTG602959S, RTG602973S, RTG603081S, RTG603082S, RTG603083S, RTG603084S, RTG603085S, RTG603086S, RTG603087S, RTG603815S, RTG604025S, RTG602515S, RTG603625S, RTG604748S; GTIN 00763000178154, Lot Serial Numbers: RTG604350S, RTG601347S, RTG601719S, RTG606249S, RTG606188S, RTG605216S, RTG603761S, RTG606215S, RTG605803S, RTG606279S, RTG605227S, RTG603290S, RTG602725S, RTG604144S, RTG604578S, RTG603259S, RTG603243S, RTG602026S, RTG604413S, RTG605452S, RTG605531S, RTG605790S, RTG604353S, RTG605610S, RTG604785S, RTG605031S, RTG604091S, RTG605004S, RTG605059S, RTG605749S, RTG605446S, RTG605929S, RTG604419S, RTG605990S, RTG603057S, RTG604480S, RTG602791S, RTG603099S, RTG603438S, RTG604719S, RTG602743S, RTG600268S, RTG604367S, RTG600097S, RTG601151S, RTG601262S, RTG603558S, RTG601024S, RTG600087S, RTG600546S, RTG601035S, RTG602765S, RTG603488S, RTG601036S, RTG600069S, RTG600393S, RTG600534S, RTG601147S, RTG601312S, RTG601712S, RTG601769S, RTG601833S, RTG602328S, RTG602371S, RTG602400S, RTG602414S, RTG602700S, RTG602772S, RTG604102S, RTG604156S, RTG604222S, RTG601143S, RTG600523S, RTG601478S, RTG600346S, RTG600377S, RTG600415S, RTG600455S, RTG601496S, RTG602775S, RTG603432S, RTG602821S, RTG600082S, RTG600106S, RTG601146S, RTG601284S, RTG601930S, RTG602095S, RTG602115S, RTG602393S, RTG601766S, RTG601918S, RTG603469S, RTG600345S, RTG600378S, RTG600531S, RTG601070S, RTG601154S, RTG601228S, RTG603880S, RTG604358S, RTG601136S, RTG602744S, RTG600348S, RTG605228S, RTG600405S, RTG601218S, RTG602073S, RTG605935S, RTG600837S, RTG600083S, RTG601164S, RTG602267S, RTG602391S, RTG603905S, RTG603569S, RTG604486S, RTG604508S, RTG602668S, RTG602518S, RTG601935S, RTG602718S, RTG600066S, RTG600067S, RTG600068S, RTG600113S, RTG600146S, RTG600838S, RTG600860S, RTG601265S, RTG601298S, RTG601302S, RTG601516S, RTG601518S, RTG601722S, RTG601723S, RTG601731S, RTG602007S, RTG602243S, RTG602246S, RTG602281S, RTG602286S, RTG602294S, RTG602543S, RTG602562S, RTG602576S, RTG602664S, RTG602695S, RTG602699S, RTG602702S, RTG602709S, RTG602756S, RTG603693S, RTG604050S, RTG605551S, RTG600871S, RTG604049S, RTG604715S, RTG605060S, RTG601252S, RTG602416S, RTG603297S, RTG603498S, RTG603499S, RTG602628S, RTG600889S, RTG601720S, RTG601721S, RTG603540S, RTG602296S, RTG603022S, RTG603035S, RTG603117S, RTG603225S, RTG603226S, RTG604092S, RTG604155S, RTG604396S, RTG604422S, RTG604438S, RTG604790S, RTG605991S, RTG602802S, RTG600030S, RTG600846S, RTG601304S, RTG601561S, RTG602124S, RTG602734S, RTG603045S, RTG603216S, RTG605660S, RTG601018S, RTG601019S, RTG603451S, RTG602390S, RTG602778S, RTG602417S, RTG600061S, RTG601159S, RTG600099S, RTG600112S, RTG600114S, RTG600676S, RTG601316S, RTG601429S, RTG601556S, RTG601618S, RTG601789S, RTG602105S, RTG602269S, RTG602300S, RTG602452S, RTG603523S, RTG605603S, RTG605968S, RTG606002S, RTG606003S, RTG606026S, RTG600033S, RTG600065S, RTG601140S, RTG601226S, RTG602263S, RTG602564S, RTG602632S, RTG602793S, RTG603198S, RTG601322S, RTG602249S, RTG602250S, RTG602560S, RTG602622S, RTG602672S, RTG603224S, RTG603317S, RTG601037S, RTG601038S, RTG602389S, RTG605543S, RTG603324S, RTG603220S, RTG600048S, RTG603571S, RTG604270S, RTG604798S, RTG601598S, RTG601686S, RTG602081S, RTG603976S, RTG604086S, RTG600119S, RTG601939S, RTG602274S, RTG602254S, RTG602457S, RTG602462S, RTG605594S, RTG604436S, RTG601048S, RTG601020S, RTG602116S, RTG603109S, RTG603318S, RTG600036S, RTG600049S, RTG600070S, RTG600092S, RTG600787S, RTG603046S, RTG603373S, RTG603487S, RTG603115S, RTG604714S, RTG600991S, RTG601477S, RTG602266S, RTG603954S, RTG604776S, RTG603219S, RTG605015S, RTG601156S, RTG602292S, RTG602404S, RTG603899S, RTG605891S, RTG605944S, RTG605019S, RTG606339S, RTG606377S, RTG601272S, RTG602671S, RTG603564S, RTG603902S, RTG604429S, RTG604509S, RTG601157S, RTG601158S, RTG601353S, RTG601493S, RTG601549S, RTG601717S, RTG601954S, RTG603483S, RTG600932S, RTG601294S, RTG602283S, RTG604394S, RTG601236S, RTG601705S, RTG602272S, RTG602295S, RTG603015S, RTG603020S, RTG603021S, RTG605888S, RTG601538S, RTG602758S, RTG602759S, RTG603474S, RTG604352S, RTG604415S, RTG601726S, RTG602268S, RTG602289S, RTG602349S, RTG602352S, RTG602723S, RTG602761S, RTG603223S, RTG603305S, RTG603306S, RTG605537S, RTG606362S, RTG603222S, RTG601221S, RTG603135S, RTG603195S, RTG603582S, RTG604077S, RTG602420S, RTG602654S, RTG603048S, RTG603904S, RTG600029S, RTG602693S, RTG603221S, RTG600040S, RTG602675S, RTG602692S, RTG600256S, RTG600341S, RTG601317S, RTG601318S, RTG601551S, RTG602639S, RTG602716S, RTG602767S, RTG603097S, RTG603126S, RTG603294S, RTG603519S, RTG604023S, RTG605735S, RTG606389S, RTG600055S, RTG600056S, RTG602109S, RTG602251S, RTG600760S, RTG601703S, RTG602429S, RTG603168S, RTG604085S, RTG604427S, RTG602290S, RTG600988S, RTG601925S, RTG602366S, RTG603468S, RTG604084S, RTG604360S, RTG604787S, RTG604831S, RTG605039S, RTG605626S, RTG600115S, RTG600757S, RTG600843S, RTG601180S, RTG601295S, RTG601309S, RTG601357S, RTG601952S, RTG602004S, RTG602636S, RTG603118S, RTG603552S, RTG603949S, RTG604773S, RTG600796S, RTG601476S, RTG603218S, RTG605061S, RTG603314S, RTG603551S, RTG606366S, RTG600887S, RTG601359S, RTG601360S, RTG601381S, RTG602284S, RTG602997S, RTG603455S, RTG603456S, RTG604071S, RTG604351S, RTG605069S, RTG601404S, RTG601408S, RTG602377S, RTG602409S, RTG602769S, RTG603522S, RTG603568S, RTG603895S, RTG604066S, RTG604354S, RTG605496S, RTG602627S, RTG601936S, RTG600095S, RTG600850S, RTG601016S, RTG601288S, RTG601296S, RTG601550S, RTG601587S, RTG602001S, RTG602367S, RTG602401S, RTG602403S, RTG602820S, RTG603159S, RTG603465S, RTG601300S, RTG600063S, RTG600808S, RTG600810S, RTG601277S, RTG602781S, RTG602832S, RTG603108S, RTG603463S, RTG601535S, RTG602750S, RTG602751S, RTG600064S, RTG601451S, RTG602602S, RTG602388S, RTG602419S, RTG602771S, RTG603193S, RTG603449S, RTG601145S, RTG601542S, RTG602021S, RTG602032S, RTG602368S, RTG602387S, RTG602770S, RTG603323S, RTG603435S, RTG604406S, RTG601770S, RTG602753S, RTG603436S, RTG604373S, RTG601577S, RTG601709S, RTG601924S, RTG602253S, RTG602732S, RTG602794S, RTG603879S, RTG604724S, RTG601144S, RTG605000S, RTG601142S, RTG604397S, RTG600367S, RTG601239S, RTG601559S, RTG603508S, RTG603557S, RTG603873S, RTG600090S, RTG600356S, RTG600363S, RTG600918S, RTG600919S, RTG601554S, RTG601184S, RTG601293S, RTG603524S, RTG604089S, RTG604169S, RTG605250S, RTG600390S, RTG600602S, RTG600811S, RTG600909S, RTG600945S, RTG600997S, RTG601023S, RTG601155S, RTG601244S, RTG601255S, RTG601256S, RTG601258S, RTG601263S, RTG601431S, RTG601527S, RTG601547S, RTG601609S, RTG601960S, RTG601987S, RTG601990S, RTG602111S, RTG602572S, RTG602574S, RTG602633S, RTG602635S, RTG602736S, RTG602779S, RTG603120S, RTG603311S, RTG603450S, RTG603460S, RTG603464S, RTG603525S, RTG603541S, RTG604060S, RTG604063S, RTG604067S, RTG604095S, RTG604171S, RTG604391S, RTG604403S, RTG604818S, RTG604823S, RTG605003S, RTG605232S, RTG605512S, RTG605820S, RTG605854S, RTG605950S, RTG600842S, RTG601443S, RTG601243S, RTG601546S, RTG602677S, RTG602712S, RTG603542S, RTG603875S, RTG604189S, RTG604361S, RTG600863S, RTG601137S, RTG603563S, RTG603877S, RTG604100S, RTG601011S, RTG601014S, RTG601015S, RTG601217S, RTG602688S, RTG602719S, RTG603466S, RTG604054S, RTG604435S, RTG604512S, RTG604723S, RTG605037S, RTG605078S, RTG601701S, RTG604365S, RTG602129S, RTG602396S, RTG602827S, RTG603476S, RTG603511S, RTG600854S, RTG600952S, RTG601004S, RTG601725S, RTG602787S, RTG605939S, RTG605955S, RTG605956S, RTG606007S, RTG601469S, RTG601234S, RTG601351S, RTG602257S, RTG603537S, RTG603916S, RTG604094S, RTG604378S, RTG604766S, RTG601219S, RTG601238S, RTG601711S, RTG601732S, RTG602399S, RTG601328S, RTG604840S, RTG601235S, RTG602679S, RTG603473S, RTG604780S, RTG602259S, RTG603301S, RTG600925S, RTG601005S, RTG601344S, RTG602370S, RTG603094S, RTG603156S, RTG603163S, RTG603458S, RTG603500S, RTG604128S, RTG604174S, RTG604370S, RTG604423S, RTG604580S, RTG604716S, RTG604827S, RTG604828S, RTG601507S, RTG601313S, RTG601242S, RTG601329S, RTG603119S, RTG604390S, RTG602407S, RTG601303S, RTG601470S, RTG603480S, RTG604720S, RTG605914S, RTG600891S, RTG601216S, RTG601287S, RTG602363S, RTG602676S, RTG602682S, RTG603096S, RTG603505S, RTG603526S, RTG604138S, RTG604212S, RTG604345S, RTG604364S, RTG604375S, RTG604381S, RTG604424S, RTG604430S, RTG604764S, RTG604767S, RTG604801S, RTG605876S, RTG605884S, RTG605988S, RTG600110S, RTG601450S, RTG603506S, RTG601047S, RTG601044S, RTG602362S, RTG602792S, RTG603492S, RTG600540S, RTG600468S, RTG601215S, RTG601240S, RTG600844S, RTG601028S, RTG601179S, RTG604088S, RTG602694S, RTG601134S, RTG601135S, RTG601497S, RTG602106S, RTG602119S, RTG602356S, RTG604170S, RTG600091S, RTG602741S, RTG604363S, RTG604405S, RTG600936S, RTG604712S, RTG601308S, RTG604235S, RTG604513S, RTG600085S, RTG604366S, RTG605027S, RTG605600S, RTG605010S, RTG600525S, RTG600927S, RTG602737S, RTG602754S, RTG604783S, RTG600491S, RTG600996S, RTG601710S, RTG602046S, RTG602122S, RTG604093S, RTG604183S, RTG601141S, RTG600674S, RTG601245S, RTG601492S, RTG601543S, RTG602022S, RTG602705S, RTG603876S, RTG604090S, RTG604768S, RTG601138S, RTG600294S, RTG600800S, RTG600859S, RTG600888S, RTG601269S, RTG601321S, RTG602099S, RTG604099S, RTG604187S, RTG605331S, RTG602698S, RTG600675S, RTG601531S, RTG602311S, RTG603544S, RTG603496S, RTG604074S, RTG604417S, RTG602757S, RTG603098S, RTG603128S, RTG603210S, RTG603214S, RTG605911S, RTG602707S, RTG604081S, RTG601227S, RTG601362S, RTG602615S, RTG603302S, RTG603510S, RTG604101S, RTG604431S, RTG604820S, RTG601772S, RTG601831S, RTG601985S, RTG602611S, RTG602613S, RTG602626S, RTG602630S, RTG604393S, RTG604824S, RTG605885S, RTG606004S, RTG606357S, RTG601315S, RTG603196S, RTG603197S, RTG603943S, RTG604476S, RTG604477S, RTG604478S, RTG600028S, RTG600037S, RTG600054S, RTG600074S, RTG600075S, RTG600076S, RTG600081S, RTG600104S, RTG600111S, RTG600116S, RTG600118S, RTG600124S, RTG600469S, RTG600474S, RTG600858S, RTG600862S, RTG600864S, RTG601173S, RTG601246S, RTG601314S, RTG601319S, RTG601320S, RTG601440S, RTG601464S, RTG601466S, RTG601505S, RTG601940S, RTG601988S, RTG602104S, RTG602279S, RTG602288S, RTG602381S, RTG602385S, RTG602413S, RTG602650S, RTG602661S, RTG602670S, RTG602684S, RTG602768S, RTG603103S, RTG603104S, RTG603228S, RTG603308S, RTG603554S, RTG603555S, RTG603556S, RTG603910S, RTG604276S, RTG605938S, RTG602601S, RTG603033S, RTG601021S, RTG601022S, RTG600801S, RTG601049S, RTG602085S, RTG602704S, RTG603298S, RTG604510S, RTG604080S, RTG602418S, RTG604533S, RTG602386S, RTG600794S, RTG602089S, RTG602090S, RTG602169S, RTG602247S, RTG602405S, RTG602406S, RTG603299S, RTG603300S, RTG603993S, RTG604064S, RTG604065S, RTG604437S, RTG604994S, RTG605029S, RTG604788S, RTG605007S, RTG605008S, RTG604078S, RTG600062S, RTG600486S, RTG600867S, RTG600877S, RTG600908S, RTG601260S, RTG602000S, RTG602023S, RTG602689S, RTG603176S, RTG600059S, RTG600060S, RTG600120S, RTG600817S, RTG600819S, RTG600834S, RTG601257S, RTG602093S, RTG602096S, RTG602112S, RTG604769S, RTG605034S, RTG605035S, RTG605627S, RTG605629S, RTG606352S, RTG600788S, RTG603101S, RTG603131S, RTG603136S, RTG603140S, RTG603149S, RTG603152S, RTG603153S, RTG600098S, RTG600100S, RTG600101S, RTG600105S, RTG600677S, RTG600678S, RTG600679S, RTG600879S, RTG600917S, RTG602065S, RTG602691S, RTG603330S, RTG602002S, RTG600885S, RTG600939S, RTG601384S, RTG602383S, RTG602690S, RTG600874S, RTG602013S, RTG602275S, RTG602558S, RTG603201S, RTG603307S, RTG604172S, RTG605014S, RTG605067S, RTG600538S, RTG601006S, RTG601306S, RTG601307S, RTG601557S, RTG600107S, RTG600108S, RTG600913S, RTG601548S, RTG602071S, RTG601270S, RTG602379S, RTG600089S, RTG600906S, RTG600923S, RTG600940S, RTG600980S, RTG601716S, RTG602113S, RTG602265S, RTG603546S, RTG603547S, RTG603906S, RTG604070S, RTG600071S, RTG600072S, RTG600073S, RTG600086S, RTG600117S, RTG600126S, RTG600537S, RTG600926S, RTG600987S, RTG601017S, RTG601201S, RTG601222S, RTG601354S, RTG601425S, RTG601713S, RTG602055S, RTG602059S, RTG602397S, RTG603452S, RTG603502S, RTG603536S, RTG604274S, RTG600079S, RTG600856S, RTG601010S, RTG602066S, RTG602067S, RTG602376S, RTG602395S, RTG602408S, RTG602685S, RTG600878S, RTG602380S, RTG601271S, RTG602038S, RTG602825S, RTG603209S, RTG602052S, RTG601356S, RTG602048S, RTG601229S, RTG601310S, RTG600784S, RTG600890S, RTG601247S, RTG601402S, RTG602783S, RTG603161S, RTG605572S, RTG600077S, RTG601467S, RTG602033S, RTG603545S, RTG603238S, RTG604041S, RTG604118S, RTG602398S, RTG603484S, RTG600310S, RTG600395S, RTG600984S, RTG601941S, RTG603489S, RTG603514S, RTG600672S, RTG600673S, RTG600911S, RTG601178S, RTG601248S, RTG601348S, RTG602024S, RTG602375S, RTG602410S, RTG602568S, RTG602575S, RTG602625S, RTG602663S, RTG603316S, RTG603320S, RTG604356S, RTG605435S, RTG603515S, RTG602030S, RTG601150S, RTG604239S, RTG602623S, RTG603481S, RTG601148S, RTG601152S, RTG601153S, RTG602074S, RTG602773S, RTG604781S, RTG600993S, RTG600388S, RTG601045S, RTG602064S, RTG600366S, RTG602372S, RTG600352S, RTG600375S, RTG602107S, RTG602245S, RTG602824S, RTG603572S, RTG605077S, RTG601361S, RTG602005S, RTG603471S, RTG603896S, RTG602029S, RTG603917S, RTG604224S, RTG601836S, RTG604400S, RTG601167S, RTG600929S, RTG601181S, RTG601759S, RTG603285S, RTG603286S, RTG600905S, RTG600979S, RTG601055S, RTG603183S, RTG603236S, RTG604998S, RTG605096S, RTG601175S, RTG601774S, RTG601033S, RTG600982S, RTG601177S, RTG600976S, RTG601733S, RTG604058S, RTG601192S, RTG604047S, RTG600881S, RTG600941S, RTG601080S, RTG602642S, RTG605117S, RTG604133S, RTG604134S, RTG600916S, RTG601108S, RTG601169S, RTG601194S, RTG605653S, RTG600965S, RTG600986S, RTG601189S, RTG601209S, RTG603246S, RTG603570S, RTG601788S, RTG603881S, RTG604130S, RTG604605S, RTG603315S, RTG601792S, RTG600893S, RTG600894S, RTG603251S, RTG601003S, RTG601050S, RTG601744S, RTG602589S, RTG602599S, RTG603287S, RTG603977S, RTG603006S, RTG600963S, RTG601060S, RTG603265S, RTG603434S, RTG600812S, RTG600922S, RTG600924S, RTG601000S, RTG601057S, RTG601084S, RTG601087S, RTG602592S, RTG603189S, RTG603245S, RTG603269S, RTG604123S, RTG604562S, RTG604571S, RTG604576S, RTG604577S, RTG600961S, RTG601199S, RTG602641S, RTG603178S, RTG600949S, RTG601165S, RTG603235S, RTG604143S, RTG604333S, RTG605148S, RTG600978S, RTG601042S, RTG601061S, RTG601170S, RTG603275S, RTG605093S, RTG600813S, RTG600866S, RTG600910S, RTG600975S, RTG601059S, RTG603248S, RTG603278S, RTG604141S, RTG600798S, RTG600942S, RTG600943S, RTG600948S, RTG600955S, RTG600968S, RTG600969S, RTG600970S, RTG601112S, RTG601200S, RTG601737S, RTG602660S, RTG602738S, RTG603171S, RTG603237S, RTG603242S, RTG603579S, RTG603890S, RTG604048S, RTG604127S, RTG604131S, RTG605080S, RTG605177S, RTG605635S, RTG601052S, RTG600848S, RTG600803S, RTG602646S, RTG604320S, RTG603549S, RTG602662S, RTG602868S, RTG602721S, RTG603576S, RTG600904S, RTG600896S, RTG600897S, RTG600953S, RTG603257S, RTG604056S, RTG604538S, RTG605111S, RTG605643S, RTG605646S, RTG600821S, RTG600956S, RTG600957S, RTG600958S, RTG601077S, RTG601081S, RTG601185S, RTG601741S, RTG601756S, RTG602728S, RTG603232S, RTG603433S, RTG603453S, RTG603792S, RTG604146S, RTG604298S, RTG604318S, RTG600902S, RTG603283S, RTG603562S, RTG602860S, RTG603000S, RTG603002S, RTG603266S, RTG603268S, RTG604059S, RTG600892S, RTG603533S, RTG603231S, RTG600900S, RTG600903S, RTG600907S, RTG600974S, RTG600990S, RTG601051S, RTG601168S, RTG601203S, RTG601739S, RTG601740S, RTG602593S, RTG603185S, RTG603280S, RTG600869S, RTG600981S, RTG600873S, RTG600895S, RTG600946S, RTG600947S, RTG600960S, RTG601062S, RTG601063S, RTG601071S, RTG601072S, RTG601074S, RTG601076S, RTG601083S, RTG601088S, RTG601183S, RTG601206S, RTG601578S, RTG601579S, RTG601738S, RTG601799S, RTG602591S, RTG602659S, RTG603137S, RTG603141S, RTG603202S, RTG603250S, RTG604145S, RTG604148S, RTG604149S, RTG604316S, RTG604317S, RTG605091S, RTG605681S, RTG600915S, RTG600967S, RTG601190S, RTG601195S, RTG603284S, RTG600876S, RTG600880S, RTG600973S, RTG601053S, RTG601056S, RTG601067S, RTG601166S, RTG601204S, RTG601747S, RTG601755S, RTG601777S, RTG601779S, RTG601780S, RTG601793S, RTG602644S, RTG602645S, RTG603233S, RTG603247S, RTG603253S, RTG603254S, RTG603274S, RTG603288S, RTG603958S, RTG604136S, RTG604137S, RTG604139S, RTG604582S, RTG601590S, RTG600914S, RTG600920S, RTG601066S, RTG601075S, RTG603273S, RTG603258S, RTG604142S, RTG601176S, RTG601026S, RTG603182S, RTG601193S, RTG600783S, RTG600839S, RTG600989S, RTG601054S, RTG601198S, RTG601223S, RTG601750S, RTG602651S, RTG602730S, RTG603144S, RTG604140S, RTG601078S, RTG601187S, RTG601743S, RTG603312S, RTG600985S, RTG601197S, RTG600849S, RTG601196S, RTG601041S, RTG600977S, RTG601582S, RTG601735S, RTG601749S, RTG602086S, RTG602604S, RTG603129S, RTG600921S, RTG600983S, RTG601804S, RTG603272S, RTG601781S, RTG600772S, RTG600773S, RTG600809S, RTG600959S, RTG601002S, RTG601160S, RTG601172S, RTG601225S, RTG601803S, RTG603190S, RTG603262S, RTG604568S, RTG600882S, RTG600883S, RTG600884S, RTG600912S, RTG601027S, RTG601058S, RTG601202S, RTG601591S, RTG601819S, RTG602598S, RTG602603S, RTG602740S, RTG603256S, RTG603279S, RTG603325S, RTG603531S, RTG603532S, RTG605082S, RTG605084S, RTG601171S, RTG600774S, RTG600782S, RTG600966S, RTG602655S, RTG600971S, RTG600972S, RTG601754S, RTG602648S, RTG602652S, RTG601068S, RTG602711S, RTG603271S, RTG600901S, RTG601205S, RTG601746S, RTG603174S, RTG604135S, RTG600951S, RTG601736S, RTG603170S, RTG603264S, RTG601129S, RTG601757S, RTG603143S, RTG603276S, RTG601186S, RTG603263S, RTG600962S, RTG603186S, RTG602726S, RTG601555S, RTG601274S, RTG603329S, RTG603445S, RTG602798S, RTG600898S, RTG601742S, RTG603229S, RTG604540S, RTG601191S, RTG603261S, RTG603281S, RTG601472S, RTG601473S, RTG602287S, RTG602851S, RTG603014S, RTG603227S, RTG604022S, RTG600462S, RTG601069S, RTG602686S, RTG603028S, RTG603031S, RTG601139S, RTG601481S, RTG601483S, RTG601485S, RTG601486S, RTG601490S, RTG602733S, RTG600795S, RTG600840S, RTG600857S, RTG600865S, RTG601249S, RTG601276S, RTG601667S, RTG601707S, RTG601718S, RTG602088S, RTG602638S, RTG602687S, RTG602720S, RTG602780S, RTG602789S, RTG602803S, RTG603105S, RTG603155S, RTG603309S, RTG603310S, RTG603462S, RTG603507S, RTG604083S, RTG604105S, RTG604392S, RTG604414S, RTG600872S, RTG602609S, RTG602610S, RTG603482S, RTG603530S, RTG604152S, RTG600780S, RTG601039S, RTG602248S, RTG603543S, RTG600933S, RTG600851S, RTG603188S, RTG601811S, RTG602863S, RTG600805S, RTG600853S, RTG600855S, RTG600964S, RTG601163S, RTG601758S, RTG602653S, RTG603882S, RTG604147S, RTG603212S, RTG601182S, RTG600096S, RTG600928S, RTG600930S, RTG600944S, RTG601231S, RTG601268S, RTG601566S, RTG602043S, RTG602133S, RTG602412S, RTG602588S, RTG602830S, RTG604072S, RTG604446S, RTG604534S, RTG604834S, RTG604073S, RTG604079S, RTG604109S, RTG604122S, RTG604124S, RTG600670S, RTG600671S, RTG604126S, RTG604150S, RTG604151S, RTG604167S, RTG604168S, RTG603326S, RTG603528S, RTG604348S, RTG604110S, RTG604111S, RTG604112S, RTG604121S, RTG601434S, RTG603785S, RTG603951S, RTG603953S, RTG604503S, RTG604504S, RTG605180S, RTG605188S, RTG605211S, RTG603200S, RTG603203S, RTG603204S, RTG603205S, RTG603295S, RTG603296S, RTG603439S, RTG603179S, RTG603561S, RTG601499S, RTG601558S, RTG602829S, RTG604384S, RTG604636S, RTG601715S, RTG603050S, RTG601475S, RTG601501S, RTG602378S, RTG602607S, RTG602763S, RTG602785S, RTG603441S, RTG603442S, RTG602764S, RTG603443S, RTG604008S, RTG601552S, RTG601207S, RTG601275S, RTG601289S, RTG601724S, RTG602394S, RTG602605S, RTG602777S, RTG603054S, RTG603194S, RTG603213S, RTG604003S, RTG604705S, RTG604718S, RTG605528S, RTG601064S, RTG602762S, RTG602995S, RTG603005S, RTG603187S, RTG603215S, RTG603230S, RTG603234S, RTG603267S, RTG603291S, RTG603293S, RTG603313S, RTG603322S, RTG603454S, RTG603461S, RTG603472S, RTG603493S, RTG603494S, RTG603503S, RTG603509S, RTG603512S, RTG603520S, RTG603521S, RTG603529S, RTG603535S, RTG603559S, RTG603560S, RTG603578S, RTG603580S, RTG603583S, RTG603584S, RTG603619S, RTG604021S, RTG604057S, RTG604061S, RTG604075S, RTG604082S, RTG604087S, RTG604098S, RTG604106S, RTG604107S, RTG604154S, RTG604157S, RTG604164S, RTG604203S, RTG604205S, RTG604273S, RTG604283S, RTG604328S, RTG604332S, RTG604355S, RTG604376S, RTG604377S, RTG604385S, RTG604404S, RTG604407S, RTG604408S, RTG604409S, RTG604410S, RTG604412S, RTG604416S, RTG604418S, RTG604420S, RTG604439S, RTG604457S, RTG604479S, RTG604514S, RTG604515S, RTG604530S, RTG604531S, RTG604532S, RTG604586S, RTG604598S, RTG604599S, RTG604607S, RTG604608S, RTG604660S, RTG604677S, RTG604713S, RTG604721S, RTG604765S, RTG604771S, RTG604772S, RTG604774S, RTG604775S, RTG604778S, RTG604782S, RTG604789S, RTG604815S, RTG604816S, RTG604821S, RTG604822S, RTG605005S, RTG605006S, RTG605026S, RTG605028S, RTG605032S, RTG605044S, RTG605045S, RTG605057S, RTG605063S, RTG605076S, RTG605079S, RTG605085S, RTG605118S, RTG605125S, RTG605219S, RTG605221S, RTG605229S, RTG605237S, RTG605240S, RTG605319S, RTG605396S, RTG605437S, RTG605439S, RTG605448S, RTG605453S, RTG605454S, RTG605456S, RTG605492S, RTG605502S, RTG605560S, RTG605562S, RTG605564S, RTG605565S, RTG605581S, RTG605612S, RTG605632S, RTG605642S, RTG605650S, RTG605655S, RTG605656S, RTG605657S, RTG605659S, RTG605672S, RTG605703S, RTG605706S, RTG605709S, RTG605752S, RTG605764S, RTG605795S, RTG605797S, RTG605799S, RTG605801S, RTG605812S, RTG605842S, RTG605848S, RTG605857S, RTG605874S, RTG605899S, RTG605901S, RTG605915S, RTG605925S, RTG605930S, RTG605931S, RTG605934S, RTG605940S, RTG605948S, RTG605951S, RTG605953S, RTG605960S, RTG605961S, RTG606203S, RTG606213S, RTG606233S, RTG606237S, RTG606292S, RTG606309S, RTG606380S, RTG606381S, RTG606385S, RTG606392S, RTG606405S, RTG602678S, RTG602683S, RTG602701S, RTG604125S, RTG606419S, RTG602382S, RTG602547S, RTG602596S, RTG602608S, RTG602614S, RTG602621S, RTG602624S, RTG602637S, RTG602640S, RTG602643S, RTG602680S, RTG602696S, RTG602708S, RTG603260S, RTG603918S, RTG604034S, RTG604432S, RTG604462S, RTG604564S, RTG604710S, RTG604814S, RTG604995S, RTG605094S, RTG605124S, RTG605234S, RTG605333S, RTG605639S, RTG605818S, RTG605900S; GTIN 00763000402211, Lot Serial Numbers: RTG601132S, RTG601279S, RTG601569S, RTG602864S, RTG604442S, RTG602823S, RTG603012S, RTG601100S, RTG601947S, RTG604557S, RTG605212S, RTG600868S, RTG602293S, RTG602766S, RTG601825S, RTG602839S, RTG603132S, RTG600804S, RTG601126S, RTG602256S, RTG602629S, RTG602748S, RTG604193S, RTG605047S, RTG601838S, RTG602091S, RTG602828S, RTG602862S, RTG603169S, RTG601220S, RTG603757S, RTG600847S, RTG602818S, RTG602344S, RTG600828S, RTG600830S, RTG601131S, RTG604188S, RTG600413S, RTG600797S, RTG600870S, RTG601025S, RTG601040S, RTG602009S, RTG602715S, RTG603534S, RTG602110S, RTG602857S, RTG600429S, RTG601114S, RTG602817S, RTG602100S, RTG603694S, RTG602717S, RTG602852S, RTG602806S, RTG605086S, RTG601291S, RTG601946S, RTG602402S, RTG600392S, RTG600823S, RTG602255S, RTG603158S, RTG605815S, RTG600778S, RTG600824S, RTG603184S, RTG602097S, RTG601094S, RTG604181S, RTG604470S, RTG601105S, RTG601095S, RTG602846S, RTG602847S, RTG600775S, RTG603749S, RTG603180S, RTG601214S, RTG601267S, RTG602277S, RTG603060S, RTG602050S, RTG600365S, RTG600410S, RTG600827S, RTG601007S, RTG601282S, RTG603138S, RTG604321S, RTG600391S, RTG601121S, RTG604286S, RTG600790S, RTG603100S, RTG601099S, RTG601115S, RTG603058S, RTG601128S, RTG602840S, RTG603122S, RTG604179S, RTG601311S, RTG602606S, RTG601104S, RTG601290S.RTG600995S, RTG602010S, RTG602276S, RTG602749S, RTG602786S, RTG602797S, RTG603016S, RTG603044S, RTG603160S, RTG603705S, RTG600421S, RTG600826S, RTG602805S, RTG603114S, RTG601097S, RTG602323S, RTG603053S, RTG601266S, RTG604180S, RTG603056S, RTG602842S, RTG603047S, RTG603702S, RTG603725S, RTG600861S, RTG603752S, RTG605186S, RTG601113S, RTG601254S, RTG601835S, RTG602285S, RTG603113S, RTG603130S, RTG600409S, RTG600439S, RTG601261S, RTG602076S, RTG603164S, RTG603024S, RTG600802S, RTG602856S, RTG602814S, RTG603062S, RTG602858S, RTG603134S, RTG601208S, RTG603181S, RTG601118S, RTG601101S, RTG605810S, RTG600420S, RTG602833S, RTG603063S, RTG601110S, RTG603112S, RTG603124S, RTG603125S, RTG605171S, RTG600357S, RTG600818S, RTG601012S, RTG601117S, RTG602298S, RTG602853S, RTG603042S, RTG604198S, RTG604200S, RTG605220S, RTG603681S, RTG603121S, RTG600825S, RTG603055S, RTG603107S, RTG602087S, RTG600371S, RTG601032S, RTG602826S, RTG601188S, RTG602578S, RTG603151S, RTG605018S, RTG600785S, RTG603111S, RTG600379S, RTG600382S, RTG600836S, RTG603032S, RTG600822S, RTG601120S, RTG601823S, RTG602812S, RTG602337S, RTG601091S, RTG601233S, RTG600397S, RTG604494S, RTG601125S, RTG605763S, RTG601093S, RTG600414S, RTG602984S, RTG603133S, RTG603147S, RTG603475S, RTG600432S, RTG602084S, RTG603154S, RTG602848S, RTG600416S, RTG601122S, RTG602101S, RTG603172S, RTG600362S, RTG600387S, RTG601259S, RTG600431S, RTG604199S, RTG603661S, RTG601116S, RTG602850S, RTG603148S, RTG603027S, RTG604471S, RTG602799S, RTG603539S, RTG603026S, RTG600832S, RTG602844S, RTG603748S, RTG601250S, RTG602807S, RTG603743S, RTG604194S, RTG605807S, RTG603175S, RTG602343S, RTG602618S, RTG602815S, RTG602068S, RTG603167S, RTG600417S, RTG603088S, RTG600433S, RTG604176S, RTG602834S, RTG600419S, RTG604329S, RTG600833S, RTG600807S, RTG603714S, RTG602729S, RTG603142S, RTG603145S, RTG600355S, RTG600422S, RTG603041S, RTG601124S, RTG601305S, RTG603726S, RTG603750S, RTG604330S, RTG604349S, RTG600389S, RTG602612S, RTG602784S, RTG600344S, RTG604250S, RTG604553S, RTG602667S, RTG603682S, RTG603704S, RTG602865S, RTG603123S, RTG602063S, RTG602597S, RTG602809S, RTG603191S, RTG602570S, RTG602003S, RTG603746S, RTG602831S, RTG604202S, RTG600337S, RTG602996S, RTG602819S, RTG600425S, RTG602062S, RTG602262S, RTG602838S, RTG604266S, RTG602056S, RTG602861S, RTG603102S, RTG603106S, RTG604175S, RTG600302S, RTG600376S, RTG600407S, RTG601123S, RTG601280S, RTG602746S, RTG602845S, RTG604249S, RTG604258S, RTG602808S, RTG600442S, RTG600396S, RTG600875S, RTG602541S, RTG600777S, RTG600814S, RTG601009S, RTG602811S, RTG603192S, RTG605099S, RTG605713S, RTG601133S, RTG601107S, RTG603038S, RTG600835S, RTG604275S, RTG603150S, RTG602044S, RTG600815S, RTG603739S, RTG604278S, RTG603736S, RTG602280S, RTG600776S, RTG603095S, RTG602098S, RTG601029S, RTG601297S, RTG603477S, RTG603731S, RTG600353S, RTG602800S, RTG603139S, RTG601119S, RTG600789S, RTG601278S, RTG600994S, RTG602727S, RTG602810S, RTG603711S, RTG603763S, RTG602859S, RTG604231S, RTG601031S, RTG602252S, RTG600427S, RTG601251S, RTG602977S, RTG600992S, RTG604191S, RTG604544S, RTG605046S, RTG605225S
Recalling Firm/
Manufacturer
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
8200 Coral Sea St Ne
Mounds View MN 55112-4391
For Additional Information ContactTechnical Services
800-929-4043
Manufacturer Reason
for Recall
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
FDA Determined
Cause 2
Under Investigation by firm
ActionMedtronic issued an URGENT MEDICAL DEVICE CORRECTION NOTICE to its consignees on 05/10/2023 by UPS 2-day delivery or email. Consignees are asked to forward this notification to those who need to be aware within the organization and to any location which the devices have been distributed. The notice explained the problem and the risk and requested the following patient management recommendations: - Prophylactic device replacement is NOT recommended. - Program all HV therapy pathways B>AX in all therapy zones to minimize the risk for this issue. - Prioritize reprogramming patients who have both a history of HV therapy and Rx1 programmed AX>B. - For remaining patients with AX>B programming in any HV therapy sequence, schedule (with appropriate discretion) the next follow-up for in-clinic reprogramming to minimize potential for reduced- or no-energy HV therapies to occur. - Contact Medtronic Technical Services (1-800-929-4043) or your local representative if one of the following is observed as these may be an indication of either a device or lead-related issue: o Reduced- or no-energy HV therapy is displayed in Episode Text (regardless of programmed pathway) o A persistent drop of approximately 50% in RA, RV and LV pacing lead impedance measurements as this may be an indication of increased potential for a future reduced- or no-energy therapy. Update 10/10/2023: Beginning 09/27/2023, Medtronic issued communication via letter to consignees regarding the availability of the revised IFU and SmartSync Software. Consignees are asked to forward this notification to those who need to be aware within the organization and to any location which the devices have been distributed. Consignees are asked to complete and return a confirmation certificate (or equivalent record) confirming they have received the notification.
Quantity in Commerce1914 units
DistributionUS Nationwide - Worldwide Distribution
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
PMA DatabasePMAs with Product Code = NIK
PMAs with Product Code = NIK
-
-